No part of this publication may be reproduced, stored in an electronic retrieval system, or transmitted in any form
or by any means, electronic, mechanical, photocopying, recording or otherwise without the prior written consent of
the copyright owner, or as permitted by law or under license. Full acknowledgment of the source must be given.
Brecknell is a registered trade mark of the Brecknell, LLC. This publication was correct at the time of going to print
however, Brecknell, LLC reserves the right to alter without notice the specification, design, price or conditions of
supply of any product or service at any time.
All third party brands and product names used within this document are trademarks or registered trademarks of
their respective holders.
SBI521_s_en_500811.book
Page 3
Table of Contents
Manual revision history ............................................................................................................................. 4
Chapter 1 General information and warnings ......................................................................................... 5
About this Manual ..............................................................................................................5
Text Conventions ........................................................................................................ 5
Special Messages ....................................................................................................... 5
ADNovember 2012previous revision 1.3 done as a Word document. Issue AD has been converted to FrameMaker.
AEJanuary 2013update menu tables with added parameters and LED font. Add Addendum 3 - Drum Weigher
AFJanuary 2013correct some errors.
Date CreatedDetails of Changes
Legal for Trade chapter and addendum 1 and 2 were added.
Setup. Add more print formats.
4SBI521 Service Manual
Page 7
1General information and warnings
1.1About this Manual
This manual is divided into chapters by the chapter number and the large text at the top
of a page. Subsections are labeled as shown by the 1 and 1.1 headings shown above.
The names of the chapter and the next subsection level appear at the top of alternating
pages of the manual to remind you of where you are in the manual. The manual name
and page numbers appear at the bottom of the pages.
1.1.1Text Conventions
The keys used to interface with the SBI-521 are located on the front panel of the
indicator. The keystrokes are shown in BOLD incased between brackets. (e.g. [ZERO])
Displayed messages appear in bold italic type and reflect the case of the displayed
message.
1.1.2Special Messages
Examples of special messages you will see in this manual are defined below. The
signal words have specific meanings to alert you to additional information or the relative
level of hazard.
CAUTION!
This is a Caution symbol.
Cautions give information about procedures that, if not observed, could result
in damage to equipment or corruption to and loss of data.
ELECTRICAL WARNING!
THIS IS AN ELECTRICAL WARNING SYMBOL.
ELECTRICAL WARNINGS MEAN THAT FAILURE TO FOLLOW
SPECIFIC PRACTICES OR PROCEDURES MAY RESULT IN
ELECTROCUTION, ARC BURNS, EXPLOSIONS OR OTHER HAZARDS
THAT MAY CAUSE INJURY OR DEATH.
NOTE: This is a Note symbol. Notes give additional and important information, hints
and tips that help you to use your product.
SBI521 Service Manual5
Page 8
1.2Warnings
1.3Installation
DANGER: RISK OF ELECTRICAL SHOCK. BE SURE TO UPLUG THE
INDICATOR BEFORE REMOVING THE COVER OR OPENING THE UNIT.
REFER TO QUALIFIED SERVICE PERSONNEL FOR SERVICE.
l Avoid lengthy exposure to extreme heat or cold. Your scale works best when
operated at normal room temperature. Always allow the scale to acclimate to
a normal room temperature before use.
l Allow sufficient warm up time. Turn the scale on and wait for a few minutes if
possible. This will give the internal components a chance to stabilize before
weighing.
l These electronic scales are precision instruments. Do not operate near an
in-use cell phone, radio, computer or other electronic device. These devices
emit RF and can cause unstable scale readings. If your scale ever performs
poorly, try moving the scale to a different room or location.
l Avoid using in heavy vibration and airflow conditions.
l Read the weight reading in short time after loading. The output signature of
load cell and electronic circuit may be little influenced after weighing for a
long time.
1.Unpack the indicator and all components from the shipping box.
2.Remove all back panel screws to expose the indicator main board.
3.Insert the loadcell cable through the loadcell strain relief gland.
4.Connect the loadcell cable to the indicator via the loadcell connector. Refer to
section 1.4.1 for connection details.
5.Connect all other cables be used. Refer to the respective sections for
connection details.
6.Place back panel on the rear of the indicator and fasten it with the back panel
screws.
1.3.1Charge the Battery
If you have a Model SBI-521 LED or SBI-521 LCD, the battery may need to be charged.
1.Take the indicator to a power outlet located indoors.
2.Attach one end of the charger unit to the underside of the indicator and plug
the other end of the charger unit to the power outlet.
It will take approximately 12 hours to fully charge the 6 volt 2.8Ah battery.
6SBI521 Service Manual
Page 9
CAUTION: Danger of explosion if battery is incorrectly replaced. Replace only
Power
Com 1/2
Loadcell
Config/Cal Switch Seal
Power
Com 1/2Loadcell
with back cover
without back cover
with the same or equivalent type recommended by the manufacturer. Dispose
of used batteries according to the manufacturer’s instructions.
ATTENTION: Il y a danger d'explosion s'il y a remplacement incorrect de la
batterie, remplacer uniquement avec une batterie du même type ou d'un type
équivalent recommandé par le constructeur. Mettre au rebut les batteries
usagées conformément aux instructions du fabricant.
1.4Connections
The rear cover of the stainless steel enclosure must first be removed to make the
appropriate connections to the weigh platform, printer, remote display and power
supply.
To gain access to the connectors, remove the rear panel screws on the back of the
indicator.
Figure 1.1 Back View of Indicator
SBI521 Service Manual7
Page 10
Figure 1.2 Main Board
1
7
E+
SE+
E-
SE-
S+
S-
SFG
1.4.1Weigh Platform or Loadcell Connections
Connect the loadcell cable to the loadcell connector on the main board. Connection
assignments for the loadcell connector are shown in Figure 1.3. Refer to Table 1.2 on
page 9 for loadcell jumper configuration.
Figure 1.3 Loadcell Connector
Table 1.1 Loadcell Connections
Board
Designaltion
E++ excitation1
SE++ sense2
E-- excitation3 0 VDC
SE-- sense4≤0.5 VDC
S++ signal5
S-- signal6
SFGground/shield7N/A
8SBI521 Service Manual
DescriptionPinElectrical Level
5
±0.3 VDC (≤0.12A)
5
±0.3 VDC
2.5
±0.3 VDC
2.5
±0.3 VDC
Page 11
Loadcell Jumper
1
6
TX2
RX2
GND
GND
TX1
RX1
Set JP3 jumpers for either a 4-wire or 6-wire loadcell.
Table 1.2 Loadcell Jumper Connector(JP3)
Connected PinsFunction
1.4.2RS-232 Serial Port Connections
The SBI-521 indicator comes standard with two full Bi-directional RS-232 serial ports
designed for connection to either a PC or a serial printer.
4 wire loadcell is used
6 wire loadcell is used
Connect the serial printer or computer to the RS-232 terminal on the main board.
Communication terminal connections are shown in Figure 1.4. Refer to Table 1.10 and
Table 1.11 on page 12 for COM2 jumpers..
Figure 1.4 RS-232 Connector
Table 1.3 RS-232 Port Connections
DesignationDescriptionPinElectrical Level
RX1port 1 receive1-12 to +12 VDC
TX1port 1 transmit2-12 to +12 VDC
GNDground/common30 VDC
GNDground/common40 VDC
RX2port 2 receive5-12 to +12 VDC
TX2port 2 transmit6-12 to +12 VDC
SBI521 Service Manual9
Page 12
1.4.3RS-485 Serial Port Connections
1
GND
B
A
3
The SBI-521 indicator comes standard with one RS-485 network port designed for local
networks and multidrop communication links. Refer to Table 1.10 and Table 1.11 on
page 12 for COM2 jumpers.
Figure 1.5 RS-485 Connection
Table 1.4 RS-485 Port Connections
DesignationDescriptionPinElectrical Level
ARS-485 signal A10 - 5 VDC
BRS-485 signal B20 - 5 VDC
GND ground/common30 VDC
Table 1.5 Termination Resistor Jumper (JP4)
Connected PinsFunction
1-2RS485 terminal 120ohm resistor on board is disabled
2-3RS485 terminal 120ohm resistor on board is enabled
1.4.4Power Supply Connections
The SBI-521 comes standard with an external AC to DC power adapter. Simply plug
the AC adapter into the indicator DC power jack and then plug into a standard wall
outlet.
IMPORTANT: Make sure that the AC voltage appearing at the wall outlet
matches the input voltage marked on the AC adapter.
Table 1.6 Adapter Power Input Connector (ADP)
PinDefinitionIn/Out/PowerElectrical Level
1Adapter input voltage + Power input 12VDC (10.5-15VDC, ≥0.5A)
2Adapter input voltage - (GND)Power ground0Vdc
3Shell EarthN/AN/A
Table 1.7 Battery Power Input Connector (BAT)
PinDefinitionIn/Out/PowerElectrical Level
1Battery input voltage +Power input5.6-7.2Vdc (6V/2.8AH lead acid battery)
2Battery input voltage -(GND)Power ground0Vdc
10SBI521 Service Manual
Page 13
1.5Sockets and Jumpers
Figure 1.6 shows main board connectors for the loadcell, printer or computer, power
supply and battery..
Figure 1.6 SBI-521 Main Circuit Board
1.5.1Definition of Sockets and Jumpers
Table 1.8 ADP - Adapter Power Input Connector
Pin #DefinitionIn/Out/PowerElectrical Level
1Adapter input voltage +Power input
2Adapter input voltage - (GND)Power ground0Vdc
3Shell Earth
Table 1.9 Battery Power Input Connector
Pin #DefinitionIn/Out/PowerElectrical Level
1Battery input voltage +Power input5.6-7.2Vdc (6V/4.0AH lead acid battery)
2Battery input voltage - (GND)Power ground0Vdc
12Vdc(10.5-15Vdc,
≥0.5A)
SBI521 Service Manual11
Page 14
1.5.2Communication Jumpers
Jumpers on JP8.
RXD2 and TXD2 will need to be set to the desired configuration.
Table 1.10 JP8 Connector
Connected PinsFunction
COM2 is used as RS-232
COM2 is used as RS-485
COM2 is not used
COM2 is not used
Table 1.11 RXD2 and TXD2 Jumper Connectors
Connected Pins
Function
TXD2RXD2
COM2 is not used
COM2 is used as RS-232
COM2 is used RS-485
12SBI521 Service Manual
Page 15
1.6Routine Maintenance
IMPORTANT: This equipment must be routinely checked for proper operation
and calibration.
Application and usage will determine the frequency of calibration required for
safe operation.
Always turn off the machine and isolate from the power supply before starting any
routine maintenance to avoid the possibility of electric shock.
1.7Sharp Objects
Do not use sharp objects such as screwdrivers or long fingernails to operate the keys.
1.8Cleaning the Indicator
Table 1.12 Cleaning DOs and DON’Ts
DODO NOT
Wipe down the outside of standard products
with a clean cloth, moistened with water and
a small amount of mild detergent
Spray the cloth when using a proprietary
cleaning fluid
Attempt to clean the inside of the indicator
Use harsh abrasives, solvents, scouring cleaners or
alkaline cleaning solutions
Spray any liquid directly on to the display window
SBI521 Service Manual13
Page 16
2Specifications
2.1SBI-521 Series Models
ModelDescription
SBI-521LED Version, 6V4.0AH lead-acid rechargeable battery is installed
SBI-521LCD version, 6V4.0AH lead-acid rechargeable battery is installed
2.2Housing and Outline Dimension
2.2.1Housing
IP65 wash-down stainless steel housing with rotary bracket.
2.2.2Outline Dimension
l With bracket: 10.31" x 8.46" x 3.45" (262mm x 215mm x 85mm)
l Without bracket: 8.66" x 6.30" x 3.07" (220mm x 160mm x 78mm)
Figure 2.1 SBI-521Outline Drawing
14SBI521 Service Manual
Page 17
2.3Power Supply
l 100 - 240 VAC, 50/60 Hz AC adapter
l 6V 4.0AH lead-acid rechargeable battery
l Battery charging Circuit: built-in
ModelWorking Current
2.4Display
2.4.1Keypad
2.4.2Environment
Working temperature-10°C to 40°C
Storage temperature-20°C to 70°C
Humidity10 to 90% RH without condensation
ProtectionIP65
SBI-521
SBI-521
l SBI-521: 7-digits,7-segment, 0.7'' (17mm) ultra brightness LEDs with 16
annunciators
l SBI-521: 6-digit,7-segment, 1"(25mm) LCDs with 16 annunciators and
backlight
l 6 push buttons
≤100mA (use battery, no load-cell, no adaptor, LED.BRT is set to 2)
≤45mA (use battery, no load-cell, no adapter, backlight is on)
2.4.3Load Cell Excitation
Voltage5VDC
Max. Current120mA (can power 8-350 ohm bridge)
Signal connection4 or 6 lead with sense leads
Max Sensitivity-3mV/V to +3mV/V
SBI521 Service Manual15
Page 18
2.5Communication
Serial port1Full-duplex RS-232
Serial port2Full-duplex RS-232 or half-duplex RS-485
l Output code:1mV input between S+ and S- of loadcell connector will output
about 100,000 raw counts
l Hardware low pass filter and 2 programmable digital low pass filters
2.7Capacity and Division
l Programmable
l Max display range: -999,999 to 999,999
l Division number range for primary unit: 100-100,000 (division number will be
limited by REGULA setting)
l Recommended Sensitivity: >0.5uV/ display division
2.8Accuracy
≤0.01%
2.9Approvals
NTEP: Class III, certificate # 11-119
2.10Calibration Method:
l Software calibration with long-term storage in EEPROM
l Provides smooth curve fit through four points
l Calibration can be done under kg or lb weight unit with 10% -100%FS
standard weight.
16SBI521 Service Manual
Page 19
2.11Real Clock
l Built-in nonvolatile real time & date
2.12Other Main Functions
l Programmable zero range
l Programmable pre-set tare weight
l Programmable automatic zero point tracking
l Programmable motion detection window
l Programmable auto-power off time, backlight working mode (LCD version)
and adjustable LED brightness (LED Version)
l Programmable hold function with peak weight holding and dynamic weighing
l Available check weighing mode
l Available parts counting mode
l Available percentage working mode
l Available units of measure: kg, lb, lb:oz, PCS, %
l Battery voltage display and charge indication
l Programmable serial output content
SBI521 Service Manual17
Page 20
3Configuration
Class
III
Brecknell
Max
d=
To set up the indicator, you must first enter the appropriate menu mode. The front panel
keys become directional navigators to move around in the menus. See Table 3.3 for
details.
3.1Front Panel
The front panel incorporates the display and keypad.
3.1.1LED Version
18SBI521 Service Manual
Page 21
3.1.2LCD Version
Class
III
Max
Brecknell
d=
lb:oz
kg
PeakTotalNETHold
Max
d=
3.2Display
The Model SBI-521 indicator utilizes a 6-digit LCD (Liquid Crystal Display) or a 7 -digit
LED display providing the weight and system information.
3.2.1LED
The annunciators used with the LED model surround the display. The annunciator will
be lit went the mode is active.
Figure 3.1 LED Display
SBI521 Service Manual19
Page 22
Table 3.1 LED Display Annunciators and Definitions
lb:oz
kg
PeakTotalNET
Hold
LED AnnunciatorDescription
ZEROBetter known as the "Center of Zero" annunciator. It is lit whenever the displayed weight is
within ± 0.25 divisions of true zero.
MOTIONIlluminates when weight is within motion band.
PEAK/HOLD(1) flashing/not flashing: working HOLD mode and displaying number is/isn't live
(2) red/green: working in HOLD mode and HOLD type is/isn't PEAK-HOLD.
TOTALDisplay data is total times or total of weight, pieces or percentage.
NETIndicates current weight is in the net mode and the tare weight is not 0.
GROSSIndicates current weight is in the gross mode.
TAREIndicates current weight is in the tare mode.
Pcsindicates counting mode.
AC/CHGIlluminates red when battery is being charged or green if battery is fully charged or not
installed.
lbIndicates the current unit of measure is lb.
ozIndicates the current unit of measure is oz.
kgIndicates the current unit of measure is kg.
%Indicates the current unit of measure is in percentage weighing mode.
HIIlluminates when data compare is enabled and current data (weight, pieces or percent) is
over the specified upper limit.
OKIlluminates when data compare is enabled and current data (weight, pieces or percent) is
between the specified upper and lower limits.
LOIlluminates when data compare is enabled and current data (weight, pieces or percent) is
over the specified lower limit.
3.2.2LCD
The annunciators used with the LCD model are incorporated in the display. The
annunciator will be lit went the mode is active.
Figure 3.2 LCD Display
20SBI521 Service Manual
Page 23
Table 3.2 LCD Display Annunciators and Definitions
LCD AnnunciatorDescription
Better known as the "Center of Zero" annunciator. It is lit whenever the displayed weight is
within ± 0.25 divisions of true zero.
Illuminates when weight is within motion band.
HoldFlashes when hold key is pressed. Remains permanently on when the Hold function has
become activated.
PeakIndicates peak hold mode.
TotalDisplay data is total times or total of weight, pieces or percentage.
NETIndicates net mode and the tare weight is not 0.
GROSSIndicates gross mode when arrow above GROSS is illuminated.
TAREIndicates tare mode when arrow above TARE is illuminated.
Pcsindicates counting mode.
When battery is being used or charged.
lbIndicates the current unit of measure is lb.
ozIndicates the current unit of measure is oz.
kgIndicates the current unit of measure is kg.
%Indicates the current unit of measure is in percentage weighing mode.
HIIlluminates when data compare is enabled and current data (weight, pieces or percent) is
over the specified upper limit.
OKIlluminates when data compare is enabled and current data (weight, pieces or percent) is
between the specified upper and lower limits.
LOIlluminates when data compare is enabled and current data (weight, pieces or percent) is
over the specified lower limit.
3.3Keyboard
The keyboard consists of six keys, some of which have multiple functions.
Figure 3.3 SBI-521 Keypad
SBI521 Service Manual21
Page 24
Table 3.3 Function of the Keys
KeyModeConditionFunction
Weigh, count or percentPress for less than 3 secondsEnter or exit HOLD mode
Weigh, count or percentPress for more than 3 secondsEnter setup mode
Input dataPress for more than 3 seconds Input decimal point
Input dataPress for less than 3 secondsReturn to sub menu
SetupPress for more than 3 seconds Exit setup. Refer to page 54
Weigh, count or percentPress for less than 3 secondsOutput data to serial communication port
Weigh, count or percentPress for more than 3 secondsSelect working mode (weigh, count or
percent)
Input dataPress to enter dataone will be subtracted from the flashing
digit.
SetupMenu selection modeNext item of current menu
Weigh, count or percentPress for less than 3 secondsTo display net weight or gross weight
mode
Weigh, count or percentPress for more than 3 secondsTo add up current weight/pieces/
percentage to memory, display times and
total of accumulation
Input dataPress to enter dataone will be added to the flashing digit.
SetupMenu selection modeTo last item of current menu
WeighPress for less than 3 secondsChange weigh units
Count or percentPress for less than 3 secondsTo enter a piece weight or unit
percentage weight (by way of sample or
input directly)
Weigh, count or percentPress for more than 3 secondsTo input compare data (for weight, pieces
or percentage) of high and low limitation
SetupInput data modeRotate the flashed position from left to
right
Weigh, count or percentPress for less than 3 secondsTare function
Weigh, count or percentPress for more than 3 secondsInput preset tare weight
Input data or setupInput data or select menu Confirm input data or current item
selection and go to next item of current
menu or next operation
Indicator is offPower on
Weigh, count or percentPress for less than 3 secondsZero function
Weigh, count or percentPress for more than 3 secondsPower off
SetupInput data modeignore modification
Setup Menu selection modePrepare to exit from current working
mode
22SBI521 Service Manual
Page 25
3.4Navigating the Weigh Mode
3.4.1Change the Working Mode
Press the [FUNC] key for 3 seconds. Use the [FUNC] key to choose and confirm entry
into weighing mode, counting mode or percentage working mode. The count and/or
percentage mode must be set to yes in the FUNC menu within the setup mode.
Otherwise the indicator will only show the weighing mode. Refer to page 28 for details.
3.4.2Enter or Exit the HOLD Mode
Press the [HOLD] key.
3.4.3Zero
When the weight is stable and within the zero range, press the [ZERO] key to set a new
zero point. Refer to Table 3.4 for zero limitations.
3.4.4Tare
When the gross weight is larger than zero and the scale is stable, press the [TARE]
key. The indicator will show a net weight of zero and the NET annunciator will be
illuminated. Refer to Table 3.4 for tare limitations.
3.4.5Preset Tare Weight
Enter a preset tare weight by pressing the [PRESET] key for more than 3 seconds.
Pr.Tare will be displayed and the TARE annunciator flashes. Use the up and down
arrow keys to input the desired tare weight and the right arrow key to move to the next
digit. The value entered must be larger than zero and there is no limitation to preset
tare weight. After the tare weight is entered the NET annunciator will illuminate. Refer
to Table 3.4 for tare limitations.
Note: This indicator can only save one tare weight. The new tare weight will
automatically replace the old one.
3.4.6Clear Tare Weight
Remove any weight on platform and wait until the scale is stable. Press the [TARE] key.
SBI521 Service Manual23
Page 26
Table 3.4 Zero and Tare Limitations
Standard
USA ≤0noNo actionZero
Canada≤0noNo actionZero
Europe
None
(same with Europe)
Weight on
platform
>0 noTare
>0noTare
≤0noNo actionZero
>0noTareZero
≤0noNo actionZero
>0noTareZero
Data in TARE
memory unit
yesClear the tared weight
yes
yesClear the tared weight
yesNo action
yesClear the tared weightZero and clear the
yesZero and clear the
yesClear the tared weightZero and clear the
yesZero and clear the
Tare keyZero key
key function
tared weight
tared weight
tared weight
tared weight
3.4.7Output Data
When scale is stable press the [PRINT] key.
3.4.8Display Gross or Net Weight
If the tare weight is not zero, press the [N/G] key to display NET weight or GROSS
weight.
If the tare weight is zero, GROSS weight will be displayed.
3.4.9Change Weight Unit
Press the [UNIT] key to select kg, lb or lb:oz. Note: under some conditions lb:oz is not
available. Please refer the following tables.
24SBI521 Service Manual
Page 27
Table 3.5 Use kg as Primary Unit
Calibration
Division Value
kglblb:oz
Display Division Value
0.0001kg0.0001kg0.0002lbNot available
0.001kg0.001kg0.002lbNot available
0.01kg0.01kg0.02lb0.5oz
0.1kg0.1kg0.2lbNot available
1kg1kg2lbNot available
10kg10kg20 lbNot available
0.0002kg0.0002kg0.0005 lbNot available
0.002kg0.002kg0.005 lb0.1 oz
0.02kg0.02kg0.05 lb1 oz
0.2kg0.2kg0.5 lbNot available
2kg2kg 5 lbNot available
20kg20kg50 lbNot available
0.0005kg0.0005kg0.001 lbNot available
0.005kg0.005kg0.01 lb0.2 oz
0.05kg0.05kg0.1 lb2oz
0.5kg0.5kg1 lbNot available
5kg5kg10 lbNot available
50kg50kgNot availableNot available
Table 3.6 Use lb as Primary Unit
Calibration
Division Value
kglblb:oz
Display Division Value
0.0001lbNot available0.0001lbNot available
0.001 lb0.0005 kg0.001 lbNot available
0.01 lb 0.005 kg0.01 lb 0.2 oz
0.1 lb0.05 kg 0.1 lb2 oz
1 lb0.5 kg1 lbNot available
10 lb5 kg10 lbNot available
0.0002 lb0.0001 kg0.0002 lbNot available
0.002 lb0.001 kg0.002 lbNot available
0.02 lb 0.01 kg0.02 lb0.5 oz
0.2 lb0.1 kg0.2 lbNot available
2 lb1 kg2 lbNot available
20 lb10 kg20 lbNot available
0.0005 lb0.0002 kg0.0005 lbNot available
SBI521 Service Manual25
Page 28
3.4.10Check Weight
To make the weight compare function be available, CONFIG-FUNC-COMPAR item
should set to YES. The high and low limitation of weight should be set correctly
according to following steps:
1.In weighing mode, press the [DATA] key more than 3 seconds.
2.The unit setting will be shown (Unit.kg, Unit.lb). Use the [UNIT] key to change
the unit or the [TARE] key to continue.
3.After the HIGH parameter is accessed, 000000 will be displayed. Use the
[PRINT], [N/G] and [UNIT] keys to input the high weight value.
Calibration
Division Value
0.005 lb0.002 kg0.005 lb0.1 oz
0.05 lb0.02 kg 0.05 lb1 oz
0.5 lb0.2 kg0.5 lbNot available
5 lb2 kg5 lbNot available
50 lb20 kg50 lbNot available
kglblb:oz
Display Division Value
4.Press the [TARE] key to confirm the entry. The HI annunciator will be on.
5.Press the [ZERO] key to go to the low value.
6.After Low is accessed, 000000 will be displayed. Use the [PRINT], [N/G] and
[UNIT] keys to input low weight number.
7.Press the [TARE] key to confirm. The LO annunciator will be displayed.
7a.After a reasonable limitation is set and compare is active, one of annunciators
HI, OK, LO will be illuminated. The beeper will sound according to its setting in
USER-BEEP.
NOTE: If the High number value is 0 or is equal or less than low number, the
comparison will be disabled, and the input data has no limitation.
3.5Navigating the Count Mode
1.In this mode, the scale will weigh goods, calculate and display its counts after
the piece weight of the goods is obtained. The count function must be enabled
in the setup mode. The CONFIG-FUNC-COUNT item should be set to YES in
CONFIG menu.
26SBI521 Service Manual
Page 29
3.5.1Enter Count Mode
1.To enter counting mode press the [FUNC] key for more than 3 seconds.
WEIGH will be displayed.
2.Use the [N/G] or [PRINT] key to select COUNT.
3.Press [TARE] to confirm the parts counting mode. The pcs annunciator will be
illuminated.
3a.In counting mode, the function of [ZERO], [TARE], [PRINT], [HOLD], [TARE],
[N/G], [SETUP], [ON/OFF] keys are available.
3.5.2Obtain Piece Weight
There are two ways to obtain the piece weight:
Input piece weight directly
1.Press the [UNIT] key.
2.When InP.PWt is displayed, press the [TARE] key to enter into the piece
weight mode.
3.When UNIT.KG is shown, use the [PRINT] or [N/G] key to select the unit of
piece weight.
4.Use the [TARE] key to confirm and go to piece weigh entry.
5.Press the [ZERO] key to exit the piece weight mode and back to counting
mode.
6.When last stored piece weight is shown, use [PRINT], [N/G], [UNIT] key to
input new piece weight.
7.Press the [SETUP] key more than 3 seconds to input decimal point.
8.Press the [TARE] key to confirm the entry and save it.
8a.If the input piece weight is less than 0.5d, the indicator will display PWt.Er
and go back to counting mode.
Sample a Known Quantity
1.Press the [UNIT] key.
2.When InP.PWt is shown, use the [PRINT] or [N/G] key to select SPL.PWT.
3.Press the [TARE] key to weigh samples (which quantity is known) weight,
calculate piece weight.
4.When SPL.Lo is shown, remove the sample on scale and press the [TARE]
key to confirm.
5.When SPL.Hi is shown, put samples (its quantity is known) onto the scale,
Press the [TARE] key to confirm reading weight.
6.After INP.PCS is displayed, 000000 will be shown. Use the [PRINT], [N/G],
[UNIT] key to input the quantity of samples.
7.Press the [TARE] key to confirm.
SBI521 Service Manual27
Page 30
7a.If the calculated piece weight is less than 0.5d, the indicator will display
PWt.Er and go back to counting mode. Otherwise, after the reasonable piece
weight has been captured, the scale will go back to counting mode. The piece
weight will be saved and used again.
3.5.3Check Counts (counts compare) in Counting mode:
To make counts compare function be available, CONFIG-FUNC-COMPAR item should
set to YES, and high and low limitation of pieces should be set correctly according to
following steps:
1.In counting mode, press down the [DATA] key more than 3 seconds to enter the
high and low compare data.
2.When in HIGH mode, 000000 will be displayed. Use the [PRINT], [N/G], [UNIT] key to input high quantity number.
3.Press the [TARE] key to confirm the entry. The Hi annunciator will be shown
on the display.
4.When in LOW mode, 000000 will be displayed. Use the [PRINT], [N/G], [UNIT] key to input the low quantity number.
5.Press the [TARE] key to confirm the entry. The Lo annunciator will be shown
on the display.
NOTE: If the High number value is 0 or is equal or less than low number, the
comparison will be disabled.
5a.After a reasonable limitation is set and compare is active, one of annunciators
HI, OK, LO will be lit and the beeper will sound according to its setting in
USER-BEEP.
3.6Percent Weigh Mode
In this mode the scale will calculate the weight on the scale and display its percentage
after the unit-percentage-weight is obtained. NOTE: If 100% display format is set to
100%, 100.0% or 100.00% in CONFIG-FUNC-PERCEN menu item, then, the unitpercentage-weight is the weight of 1%, 0.1% or 0.01%.
To make percent weighing function be available, CONFIG-FUNC-PERCEN menu item
shouldn't be set to NONE.
3.6.1Enter Percent Weighing Mode
1.From the percent weighing mode, normal weighing or counting mode, press the
[FUNC] key for more than 3 seconds. WEIGH/COUNT will be shown.
2.Use the [PRINT] or [N/G] key to select PERCEN.
3.Press the [TARE] key to confirm and go to percent weighing mode. The last
unit-percentage-weight will be used.
28SBI521 Service Manual
Page 31
3a.In percent weigh mode, the function of [ZERO], [TARE], [PRINT], [HOLD],
[TARE], [N/G], [SETUP], [ON/OFF] keys are available.
3.6.2Obtain Percentage Weight
There are two ways to obtain the unit percent weight.
Input Weight and Percentage
The scale calculates the unit-percentage weight and weigh samples weight which
percentage is known. Input weight and its percentage from keypad, and calculate unitpercentage-weight: in percent weighing mode.
1.Press the [UNIT] key.
2.When InP.Pct is shown, press the [TARE] key to enter this mode.
3.Before entering the weight, use the [PRINT] or [N/G] key to select the
percentage from 1%, 2%, 5%, 10%, 20%, 50% and 100%. This percentage is
corresponding to the weight you will input in following steps
4.When UNIT.KG is shown, use the [PRINT] or [N/G] key to select the unit of
input weight.
5.Use the [TARE] key to confirm and go to next.
6.When last stored unit-percentage-weight data is shown, use the [PRINT], [N/G], [UNIT] keys to input new unit-percentage-weight.
7.Press the [HOLD] key more than 3 seconds to input decimal point.
8.Press the [TARE] key to confirm and save it.
9.Go back to percent weighing mode. If the calculated unit-percentage-weight is
less than 0.5d, the indicator will display Pct.Er and go back to percent
weighing mode.
Obtain Unit Percentage Weight
By weighing samples weight which percentage is known.
1.In percent weighing mode, press the [UNIT] key,
2.When InP.Pct is shown, use the [PRINT] or [N/G] key to select SPL.Pct.
3.Press the [TARE] key to weigh samples (which percentage is known) weight
and calculate piece weight.
4.When SPL.Lo is shown, remove the sample on the scale and press the
[TARE] key to confirm.
4a.Before the scale is stable, SPL.Lo will be flashed. After stability is reached, it
will go to next step.
5.When SPL.Hi is shown, place samples (its percentage is known) onto the
scale.
6.Press the [TARE] key to confirm reading weight.
6a.Before scale is stable, SPL.Hi will be flashed. After it is stability is reached, it
will go to next step.
SBI521 Service Manual29
Page 32
7.After INP.PCT is shown, 000000 (position of decimal point is determined by
CONFIG-FUNC-PERCEN setting) will be displayed
8.Use the [PRINT], [N/G], [UNIT] keys to input the percentage of samples.
9.Press the [TARE] key to confirm.
9a.If the calculated unit-percentage-weight is less than 0.5d, the indicator will
display Pct.Er and go back to percent weighing mode. Otherwise, the scale
will go back to percent weighing mode. The unit percentage weight can be
saved after the power off and can be used next time.
3.6.3Check Percent (percentage compare) in Percent Weighing Mode
To make percentage compare function be available, CONFIG-FUNC-COMPAR menu
item should set to YES, and high and low limitation of percentage should be set
correctly according to following steps:
1.In percent weighing mode, Press the [DATA] key for more than 3 seconds to
enter compare data of high and low.
2.After HIGH is shown, 000000 will be displayed. Use the [PRINT], [N/G], [UNIT] keys to input high percentage number.
3.Press the [TARE] key to confirm. The HI annunciator will be illuminated.
4.After Low is shown, 000000 will be displayed. use the [PRINT], [N/G],
[UNIT] keys to input the low percentage number.
5.Press the [TARE] key to confirm. The LO annunciator will be illuminated.
NOTE: If the High number value is 0 or is equal or less than low number, the
comparison will be disabled.
5a.After a reasonable limitation is set and compare is active, one of annunciators
HI, OK, LO will be lit and the beeper will sound according to its setting in
USER-BEEP.
3.7Hold Function
HOLD function can be used to freeze a displayed number. In this mode, the scale can
capture a dynamic number, hold a stable number, or average a unstable number and
then HOLD (freeze) this number temporarily for the user to view or record.
The HOLD function can be used in normal weighing mode, counting mode and percent
weighing mode. After entering HOLD mode, the speed of A/D converter can be
increased to 80Hz (if USER-HOLD-AD.H.SPD is set to YES) from original 10Hz for
some dynamic weighing applications.
It is possible to weigh restless weighing samples such as live animals or moving objects
within the hold function. The indicator provides special mode settings to accommodate
weight movements.
30SBI521 Service Manual
Page 33
3.7.1Activate the Hold Function
The CONFIG-FUNC-HOLD menu item must be set to YES.
Menu items of USER-HOLD-HLD.MOD /-AVG.TIM /-HLD.TIM /-DYN.RNG /-STB.TIM,
USER-OTHER-NLD.RNG need be set to reasonable value.
To speedup sampling of weight, set USER-HOLD-AD.H.SPD menu item to YES.
3.7.2Access the Hold Mode
To enter the HOLD mode, press the [HOLD] key while in the normal weighing mode,
counting mode or percent weighing mode.
3.7.3Hold Methods
l Positive Peak Number HOLD mode
l Negative Peak Number HOLD mode
l Toggle HOLD mode
l Average HOLD mode
l Auto HOLD mode
The following are details of these HOLD modes:
Positive Peak HOLD
When USER-HOLD-HLD.MOD is set to PS.PEAK, the hold mode is positive peak hold
mode.
When scale first enters this working mode, it will display the largest positive number
that is from the time of zero-point set. After entering this working mode, scale will
always catches and refresh positive larger number and display it.
To exit HOLD mode, press the [HOLD] key again.
Negative Peak HOLD
When USER-HOLD-HLD.MOD is set to NG.PEAK, the hold mode is negative peak hold
mode.
When scale first enters this working mode, it will display the largest negative number
from the time of zero-point set. After entering this working mode, scale will always
catches negative larger number and display it.
To exit HOLD mode, press the [HOLD] key again.
Toggle HOLD
When USER-HOLD-HLD.MOD is set to TOGGLE, the hold mode is toggled between
HOLD and MANUAL HOLD function.
After entering this working mode, the scale will freeze and display number if scale is
stable. Only the weight that is over USER-OTHER-NLD.RNG (zero 'dead' band) can
be held.
SBI521 Service Manual31
Page 34
To exit HOLD mode, press the [HOLD] key again. If the time of scale being unstable is
more than USER-HOLD-STB, TIM, STB.ER will be shown. Press the [TARE] key to
start averaging again or press the [HOLD] key to exit.
Average HOLD
Set the USER-HOLD-HLD.MOD parameter to AVERAG.
After entering this working mode the scale will freeze and display number if scale is
stable. If scale is not stable, but the variation is less than USER-HOLD-DYN.RNG
setting, the scale will average data in USER-HOLD-AVG.TIM, then freeze and display
the number. Only the weight that is over USER-OTHER-NLD.RNG can be frozen.
The scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TIM. If
the time of scale variation being over USER-OTHER-NLD.RNG is more than USERHOLD-STB.TIM, STB.ER will be shown.
Press [PRINT], [N/G], [UNIT] or [TARE] to start averaging again, or press the [HOLD]
key to exit.
Auto HOLD
Set the USER-HOLD-HLD.MOD parameter to AUTO. Different items can be weighed
one after another without pressing any buttons.
After entering this working mode, scale will freeze and display number if scale is stable.
If scale is not stable, but the variation is less than USER-HOLD-DYN.RNG the scale
will average data in USER-HOLD-AVG.TIM, then freeze and display the number. Only
the weight that is over USER-OTHER-NLD.RNG can be frozen. If the held weight is
moved away and a new load is put on the scale, the new load will automatically be held.
The scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TIM. If
the time of scale variation being over USER-OTHER-NLD.RNG is more than USERHOLD-STB.TIM, STB.ER will be shown. Press the [TARE] key to start averaging again
or press the [HOLD] key to exit.
3.7.4Hold Annunciators
In Positive or Negative Peak HOLD mode, the red HOLD (for LED version) or PEAK
and HOLD (for LCD version) annunciator will be illuminated.
In other HOLD modes the green HOLD (for LED version) or HOLD (for LCD version)
annunciator will be illuminated.
When the HOLD annunciator is flashing, the displayed number is live. When HOLD
annunciator is steady, the displayed number is frozen.
3.8Data Compare Function
Data compare function can be used in normal weighing mode, counting mode and
percent weighing mode. It can be called Check Weigher, Check Counts and Check
Percentage. When this function is enabled, you can set a higher and a lower limitation
of weight, counts or percentage independently. These limitations can be saved
permanently. Then, the current data of weight, counts or percentage will be compared
with the setting limitation and corresponding annunciator will be illuminated.
32SBI521 Service Manual
Page 35
3.8.1Enable Data Compare Function
To make data compare function be available, CONFIG-FUNC-COMPAR menu item
should set to YES, and high and low limitation should be set correctly according to
following steps:
3.8.2View Data Compare Function
1.From the weighing, counting or percent weighing mode, press the [DATA] key
more than 3 seconds to enter input compare data of high and low mode.
2.After HIGH is shown, the last saved high value will be displayed. Use the
[PRINT], [N/G], [UNIT] keys to enter a new high value.
3.Press the [TARE] key to confirm. The HI annunciator will be illuminated.
4.After Low is shown, the last saved low value will be displayed. Use the
[PRINT], [N/G], [UNIT] keys to enter a new low value.
5.Press the [TARE] key to confirm. The LO annunciator will be illuminated.
6.Press the [ZERO] key to exit and back to original working mode.
NOTE: If the High number value is 0 or is equal or less than low number, the
comparison will be disabled.
After a reasonable limitation is set and compare function is active, one of annunciators
HI, OK, LO will be illuminated and the beeper will sound according to its setting in
USER-BEEP.
For details refer to User Menu on page 44.
3.9Accumulation
The accumulation function can be used in weighing, counting and percent weighing
mode. When this function is enabled, you can accumulate the current net weight, piece
weigh and percentage.
NOTE: When the HOLD function is enabled and the scale is working in PEAK HOLD
mode, the accumulation function will be automatically disabled.
NOTE: The positive displayed number can only be totalled when the load on the scale
is larger than the USER-OTHER-NLD.RNG value. The number of accumulations and
total can be displayed or printed.
SBI521 Service Manual33
Page 36
3.9.1Enable Accumulation
To make the accumulation function be available the CONFIG-FUNC-ACCUMU
parameter should be set to MANUAL or AUTO.
Manual Accumulation
When CONFIG-FUNC-ACCUMU is set to MANUAL, the stable and positive displayed
net weight must be larger than the USER-OTHER-NLD.RNG setting.
Piece or percentage can be accumulated by a long press of the [TOTAL] key. The
indicator will display accumulation times first and then display total of number.
NOTE: Be sure to remove the accumulated load to avoid duplicating an accumulation.
The weight on the scale must be smaller than the setting at USER-OTHER-NLD.RNG
to continue.
Auto Accumulation
When CONFIG-FUNC-ACCUMU is set to AUTO, the stable and positive displayed net
weight must be larger than the USER-OTHER-NLD.RNG setting.
Piece or percentage can be accumulated automatically. The indicator will display
accumulation times first and then display total number of accumulations.
NOTE: Be sure to remove the accumulated load to avoid duplicating an accumulation.
The weight on the scale must be smaller than the setting at USER-OTHER-NLD.RNG
to continue.
View Total
When the displayed number is zero, press the [TOTAL] key for 3 seconds. The
indicator will display accumulation times first and then display total of number.
NOTE:
l ZERO only be active when scale is stable and weight is in SAZSM setting
range.
l TARE only be active when the scale is stable.
l When the TARE weight or ZERO weight is cleared, the indicator will display
the GROSS mode.
l NET weight is displayed when a TARE weight is entered.
3.10Display and Set Time
1.Enter the Setup Mode by pressing the [SETUP] key more than 3 seconds.
2.Use the [PRINT] or [N/G] key to select the MISC menu.
3.Press [TARE] to select the MISC menu.
4.Press the [PRINT] key to scroll to the TIME parameter.
5.Press [TARE] to display current time.
34SBI521 Service Manual
Page 37
5a.Time display Format is: txx.xx.xx(hh-mm-ss) for LED Version, 24h format
xx:xx:xx(hh-mm-ss) for LCD Version, 24h format.
6.Press the [DATA] key for 3 seconds. The first digit of the time will flash.
7.Use the [PRINT] or [N/G] key to enter the first digit.
7a.If time of no operation is more than 5s, it will automatically exit modification
mode.
8.Use the [DATA] key to move to the next digit.
9.Press the [SETUP] key to exit the time entry mode.
3.11Display and Set Date
1.Enter the Setup Mode by pressing the [SETUP] key more than 3 seconds.
2.Use the [PRINT] or [N/G] key to select the MISC menu.
3.Press [TARE] to select the MISC menu.
4.Press the [PRINT] key to scroll to the DATE parameter.
5.Press [TARE] to display current date.
5a.Date display Format is: dxx.xx.xx(yy-mm-dd) for LED Version, xx.xx.xx(yy-
mm-dd) for LCD Version.
6.Press the [DATA] key for 3 seconds. The first digit of the time will flash.
7.Use the [PRINT] or [N/G] key to enter the first digit.
7a.If time of no operation is more than 5s, it will automatically exit modification
mode.
8.Use the [DATA] key to move to the next digit.
9.Press the [SETUP] key to exit the date entry mode.
3.12Display Firmware Version
1.Enter the Setup Mode by pressing the [SETUP] key more than 3 seconds.
2.Use the [PRINT] or [N/G] key to select the MISC menu.
3.Press [TARE] to select the MISC menu.
4.Press the [PRINT] key to scroll to the VER parameter.
5.Press [TARE] to display current version.
5a.Firmware Version display Format is: Vxx.yy, xx is hardware version, yy is
software version.
6.Press the [SETUP] key to return to last menu item.
7.Press the [ON/OFF] key to prepare to exit this mode.
SBI521 Service Manual35
Page 38
3.13Display Interface Type of COM2
This function is used only to check whether the hardware and software about COM2
are matched.
1.Enter the Setup Mode by pressing the [SETUP] key more than 3 seconds.
2.Use the [PRINT] or [N/G] key to select the MISC menu.
3.Press [TARE] to select the MISC menu.
4.Press the [PRINT] key to scroll to the COM2 parameter.
5.Press [TARE] to display interface type of COM2.
6.Press the [N/G] key for more than 3 seconds to display corresponding number
of COM2 hardware type setting.
7.Press the [SETUP] key to return to last menu item.
8.Press the [ON/OFF] key to prepare to exit this mode.
3.14Display Test
1.Press the [UNIT] key more than 3 seconds to enter the SETUP mode.
2.Use the [PRINT] or [N/G] key to select TEST menu.
3.Press [TARE] to select the TEST menu.
4.Press the [PRINT] key to scroll to the DISP parameter.
5.Press [TARE] to enter the test display mode and all segments will light at first.
5a.In this mode, every press of the [N/G] key will light the next segment. Every
press of the [UNIT] key will light the next digit.
5b.Press the [TARE] key to automatically light all segments and all digits.
6.Press the [SETUP] key to return to last menu item.
7.Press the [ON/OFF] key to prepare to exit this mode.
3.15Keyboard and Buzzer Test
1.Enter the Setup Mode by pressing the [SETUP] key more than 3 seconds.
2.Use the [PRINT] or [N/G] key to select TEST menu.
3.Press [TARE] to select the TEST menu.
4.Press the [PRINT] key to scroll to the KEY parameter.
5.Press [TARE] to enter test keypad mode and key. -- will be displayed.
6.Press any key. The value of this key will be displayed and the buzzer will beep.
7.Press the [SETUP] key to return to last menu item.
8.Press the [ON/OFF] key to prepare to exit this mode.
36SBI521 Service Manual
Page 39
3.16Serial Port1/2 (COM1/2) Receiving Test
Before testing the receiving function of COM1 or COM2, a cable is need to connect a
computer to the indicator. A terminal program such as Hyper Terminal is also needed
for testing.
Note: baud rate is selected by USER-COM1/2-BAUDRT, 8N1 byte format is fixed, Hex
data (0x00 - 0xff) are used.
1.Enter the Setup Mode by pressing the [SETUP] key more than 3 seconds.
2.Use the [PRINT] or [N/G] key to select TEST-COM1.RD or TEST-COM2.RD
item.
3.Press [TARE] to enter test COM1/2 receiving function and rd1.-- or rd2.-- will
be displayed first.
3a.In this mode, received hex data (0x00 - 0xff) will be displayed on -- position.
4.Press the [SETUP] key to return to last menu item.
5.Press the [ON/OFF] key to prepare to exit this mode.
3.17Serial Port1/2(COM1/2) Transmitting Test
Before testing the transmitting function of COM1 or COM2, a cable is need to connect
a computer to the indicator. A terminal program such as Hyper Terminal is also needed
for testing.
Note: baud rate is selected by USER-COM1/2-BAUDRT, 8N1 byte format is fixed, Hex
data (0x00 - 0xff) are used.
1.Enter the Setup Mode by pressing the [SETUP] key more than 3 seconds.
2.Use the [PRINT] or [N/G] key to select TEST-COM1.TD or TEST-COM2.TD
item.
3.Press [TARE] to enter test COM1/2 receiving function and td1.-- or td2.-- will
be displayed first.
3a.In this mode, transmitted hex data (0x00 - 0xff) will be displayed on -- position,
and [HOLD], [PRINT], [N/G], [UNIT] and [TARE] keys can be used to modify
transmitted data.
4.Press the [SETUP] key to return to last menu item.
5.Press the [ON/OFF] key to prepare to exit this mode.
SBI521 Service Manual37
Page 40
4Setup Mode
CONFIGUSERCALMISCTEST
The setup menu consists of five different sub-menus. Within each sub-menu are
different menu options.
The config/cal switch must be set in the ON position in order to make changes to
specific parameters.
4.1Entering the Setup Menu
1.Power on the indicator by pressing and holding the [ON/OFF] key.
2.Press the [HOLD/SETUP] key for 3 seconds. The indicator shows "Config" to
indicate that you are in Setup Menu mode.
4.2Navigating the Setup Menu
1.Once in the “Config” menu, use the [NG/TOTAL] (up arrow) or [PRINT/FUNC]
(down arrow) key to move to the next available menu in the Setup Menu.
2.Use the [TARE/PRESET] key to access the desired menu.
3.Use the [NG/TOTAL] (up arrow) or [PRINT/FUNC] (down arrow) key to
choose a sub-menu (parameter).
4.Use the [TARE/PRESET] key to select the options within the sub-menu.
5.Use the [NG/TOTAL] (up arrow) or [PRINT/FUNC] (down arrow) key to make
a change. Press the up or down arrow until the desired choice is on the
display.
6.Press the [TARE/PRESET] key to select the
38SBI521 Service Manual
Page 41
Table 4.1 Key Navigation
[HOLD/SETUP]Access the Setup Menu.
[PRINT/FUNC]Scroll through available menus.
Choose a sub-menu
Make a change within the sub-menu
[NG/TOTAL]Scroll through available menus.
Choose a sub-menu
Make a change within the sub-menu
[UNIT/DATA]Choose a sub-menu
[TARE/PRESET]Make a change within the sub-menu
[ZERO/ON/OFF]Exit the setup mode
SBI521 Service Manual39
Page 42
4.3Setup Menu Parameters
CFG.ON
CFG.OFF
kg lb
PRIM.Ut
100 – 150,000
SECND.N
RESET
No Yes
noneUSA Canada Europe
REGULA
100 – 100,000
PRIM.NPRIM.D
0.0001 0.0002 0.0005 0.001 0.002 0. 005
0.01 0.02 0.050.10.20.5
125102050
ZRO.PNTFILTER
FUNC
HOLD COUNT PERCNT COMPAR ACCUMU WT.ADJ
Yes NoYes NoYes No
NO MANUAL AUTO
FLT1.TH FLT1.ST FLT2.TH FLT2.ST
0 - 2550 -2550 - 255
IZSM IN.IZSM OV.IZSM SAZ SM AZSM
0 - 1000 - 100
0 - 100
DSP.OVRCAL.ZRO LST.Z.TWEIGHT
WEIGHT CAL.ZRO LST.Z.T
NO100.0% 100.00%100%
0 - 100
10N.DSPUNITSMOTIONOVER.LD
No Yes
0 - 6
1 - 255
AD.FROMAD.H.SPD
Yes No
Yes NoADC COM2
1 - 64
This section provides more detailed descriptions of the selections found in the Setup
Menu.
The menu charts show the flow of the parameters and also provide a quick reference
to the parameters withing the menu.
The menu tables show the submenus, options and default parameter in LED/LCD
display format to coinside with the actual display.
4.3.1Config Menu
The figure above is an illustration of the available menus with the Config menu and the
choices within those menus. Refer to Table 4.2 for explanations of the menu choices.
1-254=filter1 used only when
vibration in ±0.5d*(1-254)
255= filter1 be always used
FLT1.ST1-648Digital filter1 intensity1-64 ADC’s data will be averaged
FLT2.TH0-2558Digital filter2 threshold0=no filter2
1-254=filter2 used only when
vibration in ±0.25d*(1-254)
255= filter2 be always used
FLT2.ST0-255240Digital filter2 intensity 0-255=weak to strong
FUNCHOLDYESYESEnable or disable hold
NO
function
COUNTYESNOEnable or disable
NO
counting function
PERCNTNONOEnable or disable percent
100%
weighing function
100.0%
100.00%
YES
NO
YES
NO
(1)NO=disable
(2)100%=enable and display format
is 100%
(3)100.0%=enable and display
format is 100.0%
(4)100.00%=enable and display
format is 100.00%
COMPARYESYESEnable or disable data
NO
comparison function
ACCUMUNONOAccumulation Mode
MANUAL
selection
AUTO
YES
NO
(1)NO=no accumulation function
(2)MANUAL=add up current number
to accumulation memory after
[TOTAL] key is pressed
(3)AUTO=automatically add up
current number to accumulation
memory after scale is stable and
weight is over (NLD.RNG)
WT.ADJYESNOEnable or disable weight
fine-tuning using keypad
in weighing mode
YES
NO
*If (REGULA) ≠none, this item is
NO
*The setting will be limited by the choice of REGULA
SBI521 Service Manual43
Page 46
4.3.2User Menu
COM2BEEPOUT2
KEY COMPAR
No Yes
NONE L.LOW IN.LMT O.HIGH OUT.LMT
COM1OUT1RESET
No Yes
BAUD.RT BYT.FMT OUT.MOD LAYOUT
1200 2400 4800 9600 19200 38400 57600
8N1 701 7E1
CONT PRINT CMD PRT.CMD STABLE
B.LINE
NONE LINE1 LINE2 LINE3 LINE4
HOLD
OTHER
HLD.MOD AVG.TIM STB.TIM HLD.TIM HLD.RNG
NONE PS.PEAK NG.PEAK TOGGLE AVERAG AUTO
1 - 60
3*AVG.TIM-255
0+65535
0-255
NLD.RNG CMD.SRC A.FF.T A.OFF.MD LCD.B LTSCAL.ID
1 - 255
NONE COM.1 COM.2 COM1-2
0 - 255
LED.BRT
0 - 255
BRT1-2-3
000000-999999
OFF DSP.TIM DSP.DAT AC.TIME AC.DATE
MULTPL SINGLE 200S 200F
1200 2400 4800 9600 19200 38400 57600
8N1 701 7E1
00-99
No Yes
CONT PRINT CMD PRT.CMD STABLE
BAUD.RT BYT.FMT OUT.MOD LAYOUT LC.ADDR EN.ADDR
MULTPL SINGLE 200S 200F
NONE LINE1 LINE2 LINE3 LINE4
B.LINE
SCAL.ID GROSS TARENET PERCNT UPCTWT COUNT
PCWT ACCUM DATETIME AD.CODE BAT.VOL STATUS
SCAL.ID GROSS TARENET PERCNT UPCTWT COUN T
PCWT ACCUM DATETIME AD.CODE BAT.VOL STATUS
The figure above is an illustration of the available menus with the USER menu and the
choices within those menus. Refer to Table 4.3 for explanations of the menu choices.
HOLDHLD.MODNONEAUTOHold mode(1)NONE=no hold function
PS.PEAK(2)PS.PEAK=Positive Peak
number Hold mode. Scale will
display and refresh the positive
peak value from last zero setting.
NG.PEAK(3)NG.PEAK=Negative PEAK
number Hold mode. It is similar to
PS.PEAK, but a negative number
is used.
TOGGLE(4)TOGGLE=Press [HOLD] key
to enter HOLD mode. If the
weight is over (NLD.RNG) and
stable, the data will be frozen
until the [HOLD] key is pressed
again to exit.
AVERAG(5) AVERAG= Average HOLD
mode. In this mode, if the weight
is over (NLD.RNG), and its
variation is less than
(DYN.RNG), the average data in
(AVG.TIM) will be frozen. Press
the [HOLD] key or (HLD.TIM)
time elapsed to exit this mode
AUTO(6)AUTO=Auto hold mode. It is
similar to AVERAG mode, but if
the one held load is removed,
and a new load that is over
(NLD.RNG) put on scale, the new
load will be automatically frozen.
AVG.TIM1-603Average data time for HOLD
1-60 seconds
mode
STB.TIM3*AVG.TI
M - 255
3*
AVG.TIM
Wait time for the scale to be
stable in HOLD mode
3*(AVG.TIM) - 255S
HLD.TIM0-655350Data HOLD time 0=data will be frozen until HOLD
key pressed.
1-65535=data frozen time is 165535s, after the time elapses,
scale will exit HOLD mode.
HLD.RNG0 - 2555Vibration range of data that can
be averaged and held in HOLD
mode
0=any data can be averaged
1-255= only the data which
vibration is in 1-255d can be
averaged and held.
OTHERNLD.RNG1-255101-255=the range of weight is 1-
255d. When current weight is
less than this value, the scale
can be regarded as empty, or the
load on scale is removed.
It must be bigger than
(CONFI.MOTION).
CMD.SRCNONECOM.1Source of the executed
command selection
(1)NONE=no any command will
be executed
COM.1(2)COM.1= command from
COM1 will be executed
COM.2(3)COM.2= command from
COM2 will be executed
COM1-2(4)COM.1.2= command from
COM1 or COM2 will be executed
A.OFF.T0-2550Auto off time0=not auto power off
1-255=auto power off after 1-255
minutes. In this period, no
operation or no weight change
A.OFF.MDOFFOFFAuto off mode(1)OFF=turn off indicator
DSP.TIM(2)DSP.TIM= display time
DSP.DAT(3)DSP.DAT=display date
AC.TIME(4)AC.TIME=turn off when only
battery is used. Display time
when AC adaptor is used
AC.DATE(5)AC.DATE=turn off when only
battery is used. Display date
when AC adaptor is used
LCD.BLT0-25530LCD backlight set(1)0=always off
(1)1=always on
(2)2=press down ZERO+UNIT
together more than 3s to turn on
or turn off
(3)3-255=auto on when key
operation or weight changing.
Auto off after 3-255s elapsed.
*Only available on LCD version
LED.BRTBRT1-2-
3-4-5
SCAL.ID 000000-
999999
BRT3LED brightness level setBRT1-BRT2-BRT3=low - high
*Only available on LED version
123456Scale ID number: 000000-
999999
NOTE: If AD.FROM = COM2, COM2 and OUT2 will be disabled.
SBI521 Service Manual51
Page 54
4.3.3Cal Menu
CAL. ON
CAL. OFF
ZEROLINEINPUT
CAL. PO CAL. P1END.YCAL .P2END. YCAL. P3
Yes NoYes No
CAL. END
Figure 4.3 CAL Menu Chart
The figure above is an illustration of the available menus with the CAL menu and the
choices within those menus. Refer to Table 4.4 for explanations of the menu choices.
Seal switch positionThe display will show whether the
seal switch is in the ON or Off
position. This parameter can’t be
changed within the software.
ZEROZero point calibrationOnly do zero point calibration, then
go to CAL.END to end (only need
where a zero shift has occured).
LINECAL.P0Line calibration point0Do zero point calibration. This point
can't be omitted.
CAL.P1Line calibration point1First weight point calibration. This
point can't be omitted and standard
weight must be over 10%FS.
END.YYESEnd calibration?YES=go to CAL.END to end
NO
NO=go to do next point calibration
CAL.P2Line calibration point2Second weight point calibration.
Standard weight must be over
10%FS and be larger than it in
CAL.P1. This point can be omitted.
END.YYESEnd calibration? YES=go to CAL.END to end
NO
NO=go to do next point calibration
CAL.P3Line calibration point3: Third weight point calibration.
Standard weight must be over
10%FS and be larger than it in
CAL.P2, this point can be omitted.
CAL.END calibration end and restart
52SBI521 Service Manual
Page 55
4.3.4Misc Menu
CODEVOLDATETIMEVERCOM2.TY
DISPCOM1.RDCOM1.TDCOM2.RDCOM2.TDKEY
Figure 4.4 MISC Menu Chart
The figure above is an illustration of the available menus with the MISC menu. There
are no programming choices within this menu. Refer to Table 4.5 for explanations of
the menu choices.
Table 4.5 MISC Menu Choices and Explanations
MISC
Submenu1Remark
CODEDisplay ADC's code, this code can be after no-filter, filter1 or filter2
VOLDisplay voltage; calibrate voltage; set full charged voltage and low battery voltage
DATEDisplay date and set date
TIMEDisplay time and set time
VERDisplay firmware version
COM2.TYDisplay interface type of COM2
4.3.5Test Menu
Figure 4.5 TEST Menu Chart
The figure above is an illustration of the available menus with the TEST menu. There
are no programming choices within this menu. Refer to Table 4.6 for explanations of
the menu choices.
Table 4.6 TEST Menu Choices and Explanations
TEST
Submenu1Remark
DISPTest LCD or LED
COM1.RDTest COM1 receiving
COM1.TDTest COM1 transmitting
COM2.RDTest COM2 receiving
COM2.TDTest COM2 transmitting
KEYTest keys and buzzer
SBI521 Service Manual53
Page 56
4.4Exit the Setup Menu
1.Power off the indicator by pressing and holding the [ZERO/off] key.
2.Move the slide switch on the rear cover back to the left and replace the metal
protective plate. Refer to section 4.2 for location.
3.Turn the indicator back on by pressing the [ON] key. The display will go
through a digit check, then settle into Normal Operating mode. All front panel
keys will now return to their normal mode of operation.
54SBI521 Service Manual
Page 57
5Calibration
The config/cal switch must be set to the ON position in order to calibrate the indicator.
NOTE: More than 10% of the full scale weight is needed for calibration.
5.1Calibration Mode
1.Access the setup mode by pressing the [SETUP] key for 3 seconds.
2.Use the [PRINT] key to select the CAL menu.
3.Press the [TARE] key to enter the calibration mode.
3a.After entering this mode, the number of calibrations will be shown first. This
number will be increment one digit after every calibration and calibration data
saved. This counter can't be modified or erased. It counts from 0000 to 9999,
when it reaches 9999, it starts over at 0000.
4.After the counter number was displayed, it will show "CAL.OFF" or "CAL-ON"
which depends on whether the sealed calibration switch is OFF or ON. If the
switch is OFF, the following steps can be done, but the result will not be saved.
5.Press the [TARE] key to go to next step.
Zero Point
1.When ZERO is displayed, remove all weight on the scale.
2.Press the [TARE] key to select zero point calibration.
3.The ZERO will flash while capturing the zero point. After getting reasonable
data, it automatically saves the change and exits to the weigh mode.
Linearity
The scale linearity can be calibrated with a zero reference and a full scale reference or
multiple linearity weight values. The linearity calibration is determined by the settings
in the CAL-LINE parameter.
1.When LINE is selected, press the [TARE] key to enter linearity calibration.
2.CAL.P0 will be displayed with a zero value. Remove all weight on the scale
and then press the [TARE] key to confirm zero point calibration. The zero
weight will flash while capturing the zero point. After getting the reasonable
zero-point data, the zero weight will become steady and then go to next step.
3.When first default standard weight is displayed after CAL.P1 being shown.
The default standard weight is 100%FS. Use the [HOLD], [PRINT], [N/G], [UNIT] keys to input the value of the loaded weight.
4.Put corresponding weight (more than 10%FS weight) onto scale.
SBI521 Service Manual55
Page 58
5.Press the [TARE] key to confirm the calibration.
5a.The indicator will flash the input standard weight. When the weight becomes
steady the indicator will automatically go to next step. If this point can't be
calibrated correctly CAL.Er will be displayed. It is possible the weight load
onto scale is too small or the input data is incorrect. Repeat the previous steps.
6.When End.y is shown and the y is flashing, it's waiting command to exit
calibration or go on next calibration point. Use the [PRINT] or [N/G] key to
select yes or no.
7.Press the [TARE] key to confirm.
7a.If yes is selected, the calibration will be saved and exit to the weigh mode.
7b.If no is selected, it will go to next step.
8.When first default standard weight is displayed after CAL.P2 being shown.
The default standard weight is 100% FS. Use the [HOLD], [PRINT], [N/G], [UNIT] keys to input the value of the loaded weight (more than 10% FS weight,
and larger than the weight used on CAL.P1) onto scale.
9.When End.y is shown and the y is flashing, it's waiting command to exit
calibration or go on next calibration point. Use the [PRINT] or [N/G] key to
select yes or no.
10.Press the [TARE] key to confirm.
11.When first default standard weight is displayed after CAL.P3 being shown.
The default standard weight is 100% FS. Use the [HOLD], [PRINT], [N/G], [UNIT] keys to input the value of the loaded weight (more than 10% FS weight,
and larger than the weight used on CAL.P2) onto scale.
12.Press the [TARE] key to confirm.
13.When calibration is complete CAL.End will be displayed. The indicator will restart and go back to the weigh mode.
5.2Weight Fine-tune
With this function, the user can adjust displayed weight a little, and no need standard
weight. But please note:
l The scale must have been calibrated previously
l The range of adjustment is "(current displayed weight) x (0.9-1.1)". it means
the range is about ±10%
l The CONFIG-REGULA =NONE and CONFIG-FUNC-WT.ADJ=YES must be
set.
1.To enter this mode, turn on indicator. After the indicator displays a 0 weight,
place weight onto scale. The indicator will display the weight.
2.Press down [TARE] and [ZERO] at same time until first digit flashes, this
means indicator has entered into "weight fine-tune" mode.
3.Use the arrow keys to input correct weight.
4.Press the [TARE] key to confirm. The active correct weight will be displayed.
56SBI521 Service Manual
Page 59
5.3Display ADC Output Code
In this mode, you can examine the stability of weighing system and increment the ADC
output code corresponding to the loaded weight.
Note:
l The increment of ADC code for full scale weight must be larger or equal to
10 times of selected display division. Otherwise, the calibration cannot be
properly completed.
m e.g. The display division is 0.1kg. Load 100kg standard weight on the
platform, the increment of ADC code should be at least more than
10x100kg/0.1kg= 10x1000=10000. In this case, the scale can be
calibrated. Otherwise, smaller division needs to be chosen.
l The variation of the ADC code should be small. Otherwise, the calibration
cannot be properly completed.
1.From the weigh mode press the [SETUP] key until CONFIG is shown.
2.Use the [PRINT] or [N/G] key to go to MISC - Code.
3.Press the [TARE] key to enter this mode and display the ADC output raw
code.
4.Press the [TARE] key to set the current code as a reference zero and then
display net code
5.Press [TARE] again to clear this reference and display gross code. From the
net code mode, press the [N/G] key to select net or gross code to display, not
clear the reference code.
6.In this mode, press the [UNIT] key to select displaying code that has been
filtered by no-filter, filter1 or filter1 and filter2 and corresponding annunciator
LO, OK, HI will be illuminated.
7.Press the [SETUP] key to return to last menu item.
8.Press the [ON/OFF] key to exit this mode
5.4Display and Calibrate Power Voltage
In this mode, you can examine the voltage of battery or the voltage from the AC adaptor
when no battery is used. You can also calibrate the displayed voltage and set the
voltage value of low battery point.
Note:
l The end customer normally has no need to set the value of low battery and
full-charged voltage, calibrate the displayed voltage, these have been done
in factory.
l The normal displaying voltage is 5.6V-7.6V
1.From the weigh mode press the [SETUP] key until CONFIG is shown.
2.Use the [PRINT] or [N/G] key to go to MISC - Vol.
3.Press the [TARE] key to enter this mode and display the battery voltage.
SBI521 Service Manual57
Page 60
If the voltage is not correct
Calibrate the voltage according to following steps:
1.Prepare a DC power supply which output voltage can be adjusted from 5V to 8V
(output current must be larger than 0.5A). Power off the indicator and unplug the
AC adaptor.
2.Connect the DC power to battery connector on main board. Adjust the voltage
to about 6V
3.Turn on the indicator with the [ON/OFF] key.
1.From the weigh mode press the [SETUP] key until CONFIG is shown.
2.Use the [PRINT] or [N/G] key to go to MISC - Vol.
3.Press the [TARE] key to enter this mode and display the battery voltage.
4.Press down the [UNIT] key until CAL.6V is shown.
5.To adjust the voltage to 6V, press the [TARE] key to confirm 6V calibration.
6.When CAL.7V is shown, adjust voltage to 7V. Press the [TARE] key to confirm
7V calibration. When CV.End is shown, that means the voltage calibration is
completed and then exit to display voltage.
7.In displaying voltage mode, press the [N/G] key more than 3 seconds.
8.When LoV. 5.6 is shown, use the [PRINT] or [N/G], [UNIT] keys to input new
value (5.6V is default, input range is 5.2V-5.8V).
9.Press the [TARE] key to confirm and exit to display voltage.
10.Press the [SETUP] key to return to last menu item.
11.Press the [ON/OFF] key to exit this mode.
58SBI521 Service Manual
Page 61
6Serial Communication
6.1Com Port 1
COM1 is a RS-232 bid-directional port. Communication wires are connected to the RS232 connector using TXD1, RXD1 and GND. Refer to page 9 for connection details.
6.2Com Port 2
COM2 can be RS232 or RS485, if used as RS232, communication wires come from
RS232 connector, and TXD2, RXD2 and GND are used, if used as RS485,
communication wires come from RS485 connector, and A and B are used (if need GND
or +5VCC1 can be used).
Please refer to page 9 for jumper connector details.
6.3Protocol
The baud rate and byte format is set by USER-COM1/2-BAUD.RT and USER-COM1/
2-BYT.FMT.
Responses to serial commands will be immediate, or within one weight measure cycle
of the scale. One second should be adequate for use as a time-out value by remote
(controlling) device.
6.4Transaction String
The length of each item in a transition string:
l Reading data --- 6bytes
l Data polarity ----1byte: "-" for negative, and followed the first digit; " " for
positive.
l Decimal point ---1byte: "."
l Measure unit ----1-5bytes:" lb"," kg","lb:oz","pcs","%". Units are always lower
case, left aligned
l Current status-- 4bytes
m If the weight is over capacity, the scale will return eight "^" characters
(the field of polarity, decimal point, weight data is filled by "^").
m If the weight is under capacity, it will return eight "_" characters (the field
of polarity, decimal point, and weight data is filled by "_").
m If the zero point is error, it will return eight "-" characters (the field of
polarity, decimal point, and weight data is filled by "-").
Useless leading 0 before digits is suppressed, reading weight is right aligned.
SBI521 Service Manual59
Page 62
Table 6.1 Symbols Used
<LF>Line Feed character (hex 0AH)
<CR>Carriage Return character (hex 0DH)
<ETX>End of Text character (hex 03H)
<SP>Space (hex 20H)
H
1H2H3 H4
Four current status bytes
<P>Polarity character: "?" or " "
W 1---W6Reading data, 1-6 bytes (six digits)
<DP>Decimal point
U
1U2 U3U4U5
Measure units, kg, lb, lb:oz, % or pcs; 2-5 bytes
<Add>Address of scale; 2 bytes (00-99)
<Prompt>Prompt characters of output content; max. 11bytes
Table 6.2 Bit Definition of H
Bit
Byte 1 (H1)Byte 2 (H2)Byte 3 (H3)Byte 4 (H4)
00 = stable0 = not under capacity 00 = compare disable
1 = not stable1 = under capacity
10 = not at zero point 0 = not over capacity
01 = lower limit
10 = ok
11= upper limit
1H2H3 H4
00 = normal weighing
01 = count weighing
10 = percent weighing
11 = other mode
1 = at zero point1 = over capacity
20 = RAM ok0 = ROM ok0 = gross weight0 = not in HOLD
1 = RAM error1 = ROM error1 = net weight1 = in HOLD
30 = eeprom OK0 = calibration ok0 = initial zero ok0 = battery ok
U<CR>55 0d<LF> U1U2 U3 U4U5<CR><LF> H1H2H3H4<CR><ETX>; simulate UNIT key
L<CR>4c 0d<LF> H1H2H3H4<CR><ETX>; simulate HOLD key
X<CR>58 0dpower off the scale, simulate OFF key
others<LF>? <CR><ETX>
62SBI521 Service Manual
Page 65
6.5.2Multiple
Output string frame
<LF><Add><Prompt><p>W
1W2W3W4W5
<Dp>W6 U1U2 U3 U4U5<CR>
---- Line number and content are determined by setting of USER-OUT1/2-xxxx
<LF><Add><Prompt>H
1H2H3 H4
<CR> ---- USER-OUT1/2-STATUS is set to YES
<LF><Add><CR> --- USER-OUT1/2-LINE is set to LINE1/2/3/4
---The number of blank lines is determined by USER-OUT1/2-LINE setting
<ETX> --- Last byte of string frame
Caution
l The decimal point position is determined by the CONFIG-PRIM.D
parameter.
l The unit position and bytes is determined by which current unit is used.
l The details of <Prompt> refer to the content in 4.3USER Submenu.
l If USER-COM2-EN.ADDR is set to NO, no <Add> will be output.
l In hold mode, if ADC conversion speed is set to high speed (80Hz), USER-
COM1/2-LAYOUT is set to MULTIPLE and many transmissions are selected
to output, the output contents from COM1 or COM2 may not catch up with
the data processed in indicator, If you want to watch real time data, you need
to select fewer output contents and set higher baud rate for COM1 or 2.
Example Layouts
When USER-OUT1/2-xxxx is set to YES (USER-COM2-EN.ADDR=No):
Weighing Mode:
SCALE ID:123456
GROSS:123lb 4.56oz
TARE:11lb 2.22oz
NET:112lb 2.34oz
ACC. N:8
TOTAL:789lb 15.2oz
DATE:2011-06-12
TIME:12:34:56
A/D CODE:1234567
VOLTAGE:6.7V
STATUS:bpq2
SBI521 Service Manual63
Page 66
Counting Mode:
SCALE ID:123456
GROSS:1234.55kg
TARE:12.15kg
NET:1222.40kg
QUANTITY:24448pcs
PIECE WT:0.05kg
ACC. N:10
TOTAL:23456pcs
DATE:2011-06-12
TIME:12:34:56
A/D CODE:1234345
VOLTAGE:6.7V
STATUS:bpq2
Percent weighing mode:
SCALE ID:123456
GROSS:12345lb
TARE:10lb
NET:12335lb
PERCENTAGE:91.4%
1% REF. WT:135lb
ACC. N:3
TOTAL:271.6%
DATE:2011-06-12
TIME:12:34:56
A/D CODE:1231234
VOLTAGE:6.7V
STATUS:bpq2
6.5.3CONFIG-AD.FROM=COM2
When CONFIG-AD.FROM is set to COM2, the original raw counts of weight are coming
from COM2. The raw counts from A/D chip on the local PCB are ignored.
NOTE:
(1) In legal trade application, CONFIG-AD.FROM=COM2 may be not allowed.
Please contact with your local legal metrology organization!
(2) In this application, Baud rate and byte format of COM2 is fixed to 9600bps and
8N1.
Total 12 bytes are in a frame string. W1W2W3 is 24bits output raw counts from ADC
(Hex format), (ST) is status byte, <XXH>is its high order 4 bits of XX and <XXL> is its
lower order 4bits of XX
Status byte definition:
BitSTHBitSTL
40= in working
1=be going to power off
5X10= no AD error
6020=not new data
7030=voltage is not low
When Indicator is going to power off, it outputs following content to COM2:
<STX> "OFF"<ETX><CR><LF>
00= AD is not over
1=AD is ove
1= AD error
1=new data
1=voltage is low
6.5.4Communication when USER-COM1/2-LAYOUT is set to 200S/200F:
Commands and response
Command:
W <CR>(57h 0dh) or P<CR> (50h 0dh), request current reading
Response:
The indicator will not respond if the scale is in motion, positive overload or negative
overload.
Figure 6.1 Set for 200S
SBI521 Service Manual65
Page 68
Note: When Units = lb:oz, response:
<STX><POL>xxxLxx.xO<G/N><STAT><CR><LF>
Or <STX><POL>xxxxxLxxO<G/N><STAT><CR><LF>
Figure 6.2 Set for 200F
Note: When Units = lb:oz, response:
<STX><POL>xxx<SP>lb<SP>xx.xoz<SP><GR/NT><CR><LF>
Or <STX><POL>xxxxx<SP>lb<SP>xxoz<SP><GR/NT><CR><LF>
Command: Z<CR> (5ah 0dh)
Response: Zero function is activated (simulate ZERO key). The indicator will not
respond if the scale is in motion, positive overload or negative overload. The indicator
will also not respond if it is not in gross mode or within the zero range specified in the
Setup Menu.
<STX> OK<CR><LF>
Command: T<CR> (54h 0dh)
Response: TARE function is activated (simulate TARE key). The indicator will not
respond if the scale is in motion, positive overload or negative overload. The indicator
will also not respond if it displaying a negative gross value.
<STX> OK <CR><LF>
Command: C/U<CR> (43h/55h 0dh)
Response: Changes units of measure (simulate UNIT key),
<STX> OK<CR><LF>
Command: G<CR> (47h 0dh)
Response: This command is sent to the indicator to revert to gross mode. The indicator
will not respond if the scale is in motion, positive overload or negative overload. The
indicator will also not respond if it is not in net mode.
<STX> OK<CR><LF>
66SBI521 Service Manual
Page 69
Command: N<CR> (4eh 0dh)
Response: This command is sent to the indicator to revert to gross mode. The indicator
will not respond if the scale is in motion, positive overload or negative overload. The
indicator will also not respond if it is not in gross mode or a tare has yet to be
established.
The Model SBI-521 must be configured to meet regulations mandated by local weights
and measures authorities.
7.1Configuration/Calibration Switch
The configuration/calibration switch is used to protect the indicator from being
configured or calibrated under legal for trade conditions. When set to legal for trade and
sealed with a lead seal, the indicator can only be configured and calibrated by
authorized personnel.
The switch is located under the triangle plate on the back panel. Remove the three
screws which hold the cover on.
68SBI521 Service Manual
The switch must be set in the ON position (switch is to the left).
Page 71
7.2Audit Trail Parameters
Two separate incrementing, non-resetable audit trail parameters are used by the Model
SBI-521 to indicate changes to various parameters or calibration.
7.2.1View Configuration Counter
1.Press and hold the [SETUP] key for 3 seconds. CONFIG is displayed
2.Press the [TARE] key.
2a.The display will briefly show the number of times the configuration has been
changed and saved up to 999.
To Exit:
1.Press the [ON/OFF] key. Exit? is displayed.
2.Press the [TARE] key to return to weigh mode.
7.2.2View Calibration Counter
1.Press and hold the [SETUP] key for 3 seconds. CONFIG is displayed.
2.Press the [FUNC] key twice. Cal is displayed.
3.Press the [TARE] key.
3a.The display will briefly show the number of times the configuration has been
changed and saved up to 999.
To Exit:
1.Press the [ON/OFF] key. Exit? is displayed.
2.Press the [TARE] key to return to the weigh mode.
SBI521 Service Manual69
Page 72
8Troubleshooting
This chapter gives explanations on commonly seen errors, display characters and
display symbols.
8.1Display Characters
ASCII LCD/LED ASCII LCD/LED ASCII LCD/LED
0AN
1BO
2CP
3DQ
4ER
5FS
6GT
7HU
8IV
9JW
KX
LY
MZ
70SBI521 Service Manual
Page 73
8.2Display Symbols
SymbolDescription
0------Zero is over the setting range
0_______Zero point is below the setting range
Ad------Signal to ADC is over maximum range
Ad______Signal to ADC is below minimum range
--------Weight is over upper limitation or display data is over limitation
________Weight is below lower limitation
EEP.E1CONFIG or CAL parameters are not correctly set
EEP.E2USER parameter is not correctly set
Lo.bAtBattery voltage is lower than setting
CAP.- - -Next displaying content is capacity
CAP.ERParameters about Capacity is not correct
CAL.PxCalibration on point (x)
CAL.OFFCalibration seal switch is on OFF position
CAL.ONCalibration seal switch is on ON position
CAL.ErCalibration error, maybe input data or loaded weight is incorrect, unstable, non-linear
CAL.EndEnd calibration
OFFPower OFF the indicator
STB.ERUnstable time is larger than setting of USER-OTHER-NLD.RNG
AC.xxxxAccumulation times is xxxx
PR.TAREPreset TARE weight
COMPInput COMPARE data mode
HIGHInput HIGH limitation data of Comparison
LOWInput LOW limitation data of Comparison
SPL.LoSample load weight of low point.
SPL.HISample load weight of high point.
SPL.PWTSample goods weight to calculate piece weight
INP.PCSInput number of pieces being counted
UNIT.KGUnit kg is selected
UNIT.LBUnit lb is selected
PWT.ERPiece weight error, sample too small (<0.5d)
SPL.PCTSample goods weight to calculate
INP.PCTInput percentage of weighted goods
PCT.ERUnit percentage weight is too small (<0.5d)
SBI521 Service Manual71
Page 74
8.3Error Messages and Troubleshooting
SymptomProbable CauseRemedy
Ad------Loadcell wires to indicator are incorrectly
Ad______
0------Weight reading exceeds Power On Zero limit.Make sure scale platform is empty. Perform zero
0______Weight reading below Power On Zero limit.Install platform on scale. Perform zero calibration.
--------Weight reading exceeds Overload limit or the
________Weight reading below Under load limit.Install platform on scale.Perform zero calibration.
EEP.E1CONFIG or CAL parameters are not correctly
EEP.E2USER parameter is not correctly set.Re-set items in USER.
CAP.ERCapacity parameters are not correct.Set PRIM.N/PRIM.d/SECND.n to correct number,
CAL.ErCalibration error. Input data or loaded weight is
PWT.ERPiece weight is error, it's too small (<0.5d), The
PCT.ERUnit-Percentage -Weight is error, it's too small
STB.ERUSER-HOLD-STB.TIM is too short, USER-
Will not turn on.Power cord not plugged in or properly
Cannot zero the display
or will not zero when
turned on.
Cannot display weight in
desired weighing unit.
Battery symbol is empty
or Lo.bAt is shown
connected, shorted, opened, ADC or loadcell(s)
are damaged.
weight value cannot be displayed in the current
unit of measure because it exceeds 6 digits.
set.
too small, too big, unstable, non-linear.
weight on the platform is too small to define a
valid reference weight.
(the weight of 1%, 0.1%, or 0.01%-determined
by CONFIG-FUNC-PERCNT is less than 0.5d).
HOLD-HLD.RNG is too small, other failure.
connected. Power outlet not supplying
electricity. Battery discharged. Other failure.
Load on scale exceeds allowable limits. Load
on scale is not stable. Load cell damage.
Unit not set to enable or d≥5oz, when unit is
lb:oz.
Batteries are discharged.Charge batteries.
Make sure wires are ok and correctly connected.
Replace loadcell or ADC chip, Service required.
calibration.
Reduce load on scale until weight value can be
displayed. Use a more appropriate unit of measure.
Re-set some parameters of CONFIG or USER.
Re-set items in CONFIG, do calibration.
make sure capacity not more than 6 digit.
Input correct data, load correct weight onto
platform, Service required.
Use a greater weight for the sample.
Use more weight for the sample.
Set USER-HOLD-STB.TIM longer, or set USERHOLD-HLD.RNG bigger. Service required.
Check power cord connections. Make sure power
cord is plugged into the power outlet. Check power
source. Replace batteries. Service required.
Remove load on scale. Wait for load to become
stable. Service required.
Enable unit in CONFIG-UNITS.
72SBI521 Service Manual
Page 75
Addendum 1: Quick Setup and Configuration
9.1Configure the SBI-521 to a Floor Scale
9.1.1Setting the Capacity and Count by
1.Refer to the next page for wiring and of sense line jumpers placement.
2.Press the [HOLD/SETUP] key for 3 seconds. The indicator shows Config to
indicate that you are in Setup Menu mode.
3.Press the [TARE] key once. (Tare is also used as an Enter key).
4.The display will show the Audit Trail counter then cycle to CFG.ON.
4a.If the display shows CFG.OFF, then remove the CAL seal on the back of the
indicator. Move the switch to the opposite position. Cycle power on the indicator
and go back to step 1.
5.Press the [TARE] key once. Display will show RESET.
6.Press the [PRINT] key until the display shows PRIM.N.
Table 9.1 Reference to Number of Divisions and Division Size
Capacity
Required
20004000.5
500050001
1000050002
7.Press the [TARE] key once. The display will show 003000 (this is the
number of divisions).
8.Press the [UNIT] key 3 times and the 3 should be flashing.
9.Press the [N/G] key 2 times until the 5 is displayed. (example for 5000 x 1 lb)
10.Press the [TARE] key once.
11.Press the [PRINT] key once until the display shows PRIM.D.
12.Press the [TARE] key once. Make sure the display shows 1.
13.Press the [TARE] key once. The display should show PRIMM.D.
14.Press the [PRINT] key once until the display shows PRIM.UT.
PRIM.NPRIM.D
15.Press the [TARE] key once. The display should show KG.
16.Press the [PRINT] key once.
17.Press the [TARE] key once.
18.Press the red [ZERO] key 2 times.
SBI521 Service Manual73
Page 76
19.The display will cycle power. After the indicator reboots, the display will show
1
7
E+
SE+
E-
SE-
S+
S-
SFG
the software version and capacity followed by the weighing mode.
9.1.2Loadcell Connection and Jumpers
Table 9.2 Loadcell Jumper Connector(JP3)
Connected PinsFunction
Table 9.3 Loadcell Terminal Block (typical Brecknell color code)
PinWire Color
4 wire loadcell is used
6 wire loadcell is used
1Red
2N/A
3Black
4N/A
5Green
6White
7Enclosure ground lug
Figure 9.1 Loadcell Connector
74SBI521 Service Manual
Page 77
4-wire
jumper
configuration
Figure 9.2 Loadcell Jumper Location
SBI521 Service Manual75
Page 78
Addendum 2: SBI-521 Calibration
1.Access the setup mode by pressing the [HOLD/SETUP] key for 3 seconds.
The indicator shows Config to indicate that you are in Setup Menu mode.
2.Press the [PRINT] key to select the CAL menu.
3.Press the [TARE] key to enter the calibration menu.
3a.After entering this mode, the number of calibrations will be shown first. This
number will increment one digit after every calibration and calibration data
saved.
3b.After the counter number above was displayed, it will show CAL.OFF or
CAL.ON which depends on whether the sealed calibration switch is OFF or
ON. If CAL OFF is displayed, you can still do calibration but the result will not
be saved. To turn CAL ON, power down the unit and flip the switch located on
the back of the indicator under the triangle stainless steel plate. Power back on
and follow the steps above.
4.Press the [TARE] key to go to next step.
5.Press the [PRINT] key to display LINE.
6.When LINE is selected, press the [TARE] key to enter linearity calibration.
7.CAL.PO will be displayed with a zero value. Remove all weight on the scale
and then press the [TARE] key to confirm zero point calibration. The zero
weight will flash while capturing the zero point. After getting the reasonable
zero-point data, the zero weight will become steady. Proceed to the next step.
8.When first default standard weight is display after CAL.P1 being shown. The
default standard weight is 100% of full scale. Using the [HOLD], [PRINT], [N/G], [UNIT] keys to input the value of the loaded weight. Minimum test
weight value 10% of full scale.
9.Put corresponding weight onto the scale.
10.Press the [TARE] key to confirm the calibration.
11.If calibration was successful, display will show END.Y and the Y is flashing. It
is waiting for a command to go onto the next calibration point or exit the
calibration mode. Use the [PRINT] or [N/G] key to select yes or no.
12.Press the [TARE] key to confirm.
12a. If yes is selected, the calibration will be saved and exit to the weigh mode.
12b. If no is selected, it will go to the next step.
13.When first default standard weight is displayed after Cal.P2 being shown.
The default standard weight is 100% of full scale. Use the [HOLD], [PRINT], [N/G], [UNIT] keys to input the value of the loaded weight. (More than 10% of
full scale weight, and larger than the weight used on CAL.P1) onto scale.
14.If calibration was successful, display will show END.Y and the Y is flashing. It
is waiting for a command to go onto the next calibration point or exit the
calibration mode. Use the [PRINT] or [N/G] key to select yes or no.
15.Press the [TARE]
15a. If yes is selected, the calibration will be saved and exit to the weigh mode.
76SBI521 Service Manual
key to confirm
Page 79
15b. If no is selected, it will go to the next step.
16.When first default standard weight is displayed after Cal.P3 being shown.
The default standard weight is 100% of full scale. Use the [HOLD], [PRINT], [N/G], [UNIT] keys to input the value of the loaded weight (more than 10% of
full scale weight and larger than the weight used on CAL.P2) onto scale
17.Press the [TARE] key to confirm.
18.When calibration is complete CAL.END will be displayed. The indicator will
restart and go back to the weigh mode.
SBI521 Service Manual77
Page 80
Addendum 3: Drum Weigher Setup
1.Turn on the indicator and go in to the Config setting by pressing and holding
down the [SETUP] key until CONFiG is displayed.
2.Press the [TARE] key and the display will show CFG. ON. If CFG.OFF is
displayed, move the seal switch under the triangle plate on the back panel.
3.Press the [TARE] key and the display will show RESET.
4.Press the [PRINT/FUNC] key and the display will show REGULA.
5.Press the [TARE] key and if the indicator is set at factory default, the display
should show USA.
6.Press the [PRINT/FUNC] key until NONE is displayed.
7.Press the [TARE] key and REGULA should be shown on the display.
8.Set your PRIM. N, PRIM. D, PRIM. UT, SECND. N, UNITS, MOTION,
OVER.LD as require for the capacity and division size in Appendix 1.
9.Press the [PRINT/FUNC] key until ZRO. PNT is displayed.
10.Press the [TARE] key and the display will show IZSM.
11.Press the [TARE] key. Change the value to be 000 by using your directional
arrow keys on the keypad.
12.Press the [TARE] to confirm your setting. The display will show IZSM.
13.Press the [PRINT/FUNC] key and the display will show IN.IZSM.
14.Press the [TARE] key and toggle by using the [PRINT/FUNC] key until LAST.Z.T is shown on the display.
15.Press the [TARE] to confirm the setting. The display will show IN.IZSM.
16.Press the [PRINT/FUNC] key and the display will show OV.IZSM.
17.Press the [PRINT/FUNC] key until the display shows LAST.Z.T.
18.Press the [TARE] key and the display will show OV.IZSM.
19.Press the [PRINT/FUNC] key and the display will show SAZSM.
20.Press the [TARE] key and the display will show a number. Change this
number to 000 by using your directional arrow keys on the keypad.
21.Press the [TARE] key to confirm the setting and the display will show SAZSM.
22.Press the [PRINT/FUNC] key and the display will show AZSM.
23.Press the [TARE] key and the display will show a number. Change this
number to 000 by using the directional arrow keys on the keypad.
24.Press the [TARE] key to confirm the setting and the display will show AZSM.
25.Press the [ZERO/OFF/ON] key and the display will show Exit.?.
26.Press the [TARE] key to confirm the setting and exit the configuration setting.
27.Proceed to Calibration (Appendix 2) and calibrate as required.