HYGROPALM HP31/32/GTS
Short Instruction Manual
1 GENERAL DESCRIPTION
The HP3x-Series handheld instrument is a power full, multifunctional device which measures,
displays and logs temper ature, humidity and calculated parameter s. The series consist of three
different devices. Ensure devices have the latest compatible rmware.
Order Code HP31 HP32 HP-GTS
Compatible
probes
2 DEVICE OVERVIEW
Fixed probe HC2 and HC2A-S
probes. Probes must be
ordered separately
[POWER] Switches the HP32 on or of f.
Fixed probe
3 GENERAL OPERATION
3.1 SOFT WARE HW4 LITE INSTALLATION
Data analysis, and general setting of the device can be easily done with the f ree HW4 Software.
To get the software, download the latest HW4 Software from the Rotronic Homepage. While
the soft ware always remains the same, the corresponding key unlocks the desired version. For
the free Lite version, no key is needed and up to 3 Rotronic devices can be connected. Probes
count as one device.
3.2 BATTERY, STAND-BY AND AUTO SWITCH
The HP3x has an integrated lithium battery. Simply connect the device to an USB port with the
delivered cable to charge it. The device has a stand-by and an auto switch off feature. After
5 minutes without user interaction the display is switched off, after 10 minutes the device is
switched off to save battery. This feature does not affect data logging operation.
3.3 DISPLAY AND HOME SCREEN
Status Bar
• Time and date
• Hold indicator
• Log indicator
• Battery status
Measured values
Action Buttons labels
[LEFT]
[RIGHT]
[UP] • Cursor up
[OK] • Opens the main menu / submenus
[DOWN] Cursor down
Action buttons, their action are indicated by
the on-screen labels.
• Parameter selection (home screen)
• Conrm entries
HP32 with HC2S-S probe (order separately)
Probe connector
Tripod socket
Reset button
Integrated magnet
Lanyard xation
USB port
3.4 MASTER RESET
There are a few instances in which the HP3x device may need to be reset. These include the
following:
• To restore to factor y default settings.
• The HP3x is not functioning properly (value are frozen).
Performing a reset will erase user data. Performing a Master Reset: Press and hold the reset
button on the backside of the device with a thin pen or an ofce clammy for 3 sec.
4 GENERAL SETTINGS
4.1 LANGUAGE
Main menu [OK] Settings Language
Available languages: English, Deutsch, Français, Italiano
4.2 DATE/TIME
Main menu [OK] Settings Date / Time
The following settings can also be set beside the date and time.
• Date format: DIN, US, ISO
• Time format: 12 h, 24 h
The date and time must be set correctly for proper operation of the device.
Please note that if the bat tery is completely discharge d, the handheld device must be time- chronized again with the HW4.
4.3 UNITS
Main menu [OK] Settings Units
Your options are as follow:
• Distance: m, f t
• Temperature: °C, °F
• Pressure: hPa, mmHg
Belt Buckle Set
Make sure that the plug head is not plugged
in with reversed polarity.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
• Weight: kg, lbs

4.4 CALCULATED PSYCHROMETRIC PAR AMETERS
Beside relative humidity and temperature, the HP32 can display 2 of the following 10 psychrometric parameters:
• Dew / Frost point (Dp/Fp)
• Wet bulb temperature (Tw)
• Enthalpy (H)
• Vapor concentration (Dv)
• Specic humidity (Q )
• Mixing ration by weight (R)
• Vapor pressure (E)
• Saturation vapor pressure (Ew)
• Saturation vapor density (SVD)
Select the parameters that you want to log and/or see on the home screen within:
Main menu [OK] Settings Displayed calculated parameters
You can change the order of values on the home screen by [UP] on the home screen.
8 ADJUSTMENT
The HP3X has a humidity and temperature adjustment function so that you can adjust directly
with the handheld instrument. The instrument can make a 1-point adjustment for temperature
and a 1-point or 3-point adjustment for humidity.
Main Menu [OK] Adjustment
Select Humidity or Temperature with the arrow keys and conrm with OK.
Example of sequence for 3-point humidity adjustment:
Reference value 1 10 %rh
Reference value 2 35 %rh
Reference value 3 80 %rh
The value of the reference can be entered with the keypad. The attached probe shows the
current measured value. After entering the rst reference value (10 %rh), press OK to conrm
“Save & Continue” and enter the next reference value (35 %rh).
You can change the atmospheric pressure constant used in calculations in:
Main menu Settings Atmospheric pressure
5 SAVING DATA
5.1 HOLD/CAPTURE AND SPOT MEASUREMENTS
Freeze the displayed values on the home screen by pressing the [Hold/Capture] action button
on the home screen. Those values can then be:
• Saved as a spot measurement with [Capture]
• Released again with [Release]
5.2 DATA LOGGING
Program or start the automatic logging at regular interval with the [Log] action button. Dene
the storage location, the interval and, if you want, a star t and stop time, then start the logging
operation by pressing the [Start] action button.
Your device will be automatically switched on at the given start time to start the logging.
The main menu is not accessible while data logging is active. Spot measurements are,
however, still possible.
You must name storage locations before use. Simply use the [Rename] action button
when selecting a storage location to assign it a name.
6 VIEWING SAVED DATA
6.1 SPOT MEASUREMENTS
Review the spot measurements by selecting their storage location within:
Main Menu [OK] Memory – Capture
6.2 LOGGED DATA
Review the logged data by selecting its storage location within:
Main Menu [OK] Memory – Log
Switch between graphic displays and MIN / MAX / AVG values of the 4 parameters with [UP]
and [DOWN].
7 ALARM
Set up upper, lower limits as well as hysteresis for every displayed parameter in:
Main Menu [OK] Alarms
When all three reference values have been entered, press OK to conrm “Save & Continue”
again and adjustment is nished. The last adjustment date is now listed under Adjustment.
9 TECHNICAL DATA
HP31 HP32 HP-GTS
Compatible probes Fixed HC2 and HC2A-S
Application range -10…60 °C | 0…100 %rh (non-cond.)
Probe application range -10…60 ° C Depends on probe
Accuracy @ 23 °C ±2 %rh / ±0.3 °C ±2 %rh / ±0.3 °C
Functions Spot measurements / Interval logging
Logging interval 1 s…24 h
Memory capacity 8 x 8000 data points
Psychrometric
calculations
Display refresh rate 1 s (typical)
Integrated Batter y 48 h – Logging every 30 s, screen off
PC Interface Micro-USB port, HW4 compatible
Housing Material ABS, TPR
IP Protection IP65
Dimension 160 x 63 x 30 mm
Weight 185 g
10 DELIVERED
• HygroPalm HP-31/32/GTS
• Soft case
• Micro-USB service cable (AC0006)
• Belt buckle set, lanyard and certicate
MIN / MAX / AVG / MKT
Dew / Frost point (Dp /Fp)
Wet bulb temperature (Tw)
Enthalpy (H)
Vapor concentration (Dv)
Specic humidity (Q )
Mixing ratio by weight (R)
Vapor pressure (E)
Saturation vapor pressure (Ew)
Saturation vapor density (SVD)
10 h – Screen always on (typical)
probes. Probes
must be ordered
separately
for example HC2A-S
±0.8 %rH / ±0.1 °C
Fixed
-10…60 ° C
Broken alarm limits are shown in red on the home screen.
To enable an acoustic signal for an alarm:
Main Menu [OK] Settings Buzzer