Rosemount 2088 Specifications

Product Data Sheet
September 2017
00813-0100-4690, Rev PE
Rosemount™ 2088 Absolute and Gage Pressure Transmitter
Performance of 0.065% with high accuracy option
Protocols available include 4–20 mA HART
Absolute and gage pressure ranges up to 4,000 psi (276 bar)
Rangeability of 50:1
®
and 1–5 Vdc HART Low Power
Rosemount 2088
September 2017

Rosemount 2088 Pressure Transmitter product offering

Proven reliability for gage and absolute applications

Available protocols include 4–20 mA HART and 1–5 Vdc HART Low Power
Fully configurable LCD display to display process variable, percent of range, and diagnostic messages
Lightweight, compact design enables easy installation
Choice of stainless steel or Alloy C-276 wetted materials

Unlock the value of devices with the Emerson™ Wireless THUM™ Adapter

Gain access to field intelligence and improve quality, safety, availability, operations, and maintenance costs
Remotely manage devices and monitor health
Enable new wireless measurement points
Utilize existing loop power

Proven, reliable, and innovative DP Level technologies

Connect to virtually any process with a comprehensive offering of process connections, fill fluids, direct
mount or capillary connections and materials
Quantify and optimize total system performance with QZ option

Instrument manifolds – quality, convenient, and easy

Designed and engineered for optimal performance with Rosemount transmitters
Save installation time and money with factory assembly
Offers a variety of styles, materials, and configurations
Contents
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2
Dimensional drawings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Emerson.com/Rosemount
September 2017

Ordering information

Rosemount 2088 In-line Pressure Transmitter

Rosemount 2088
Configuration
4–20 mA HART
Rosemount 2088 with Selectable HART
1–5 Vdc Low Power
Rosemount 2088 with Selectable HART
Transmitter output code
S
N
Additional information
Specifications: page 7 Product Certifications: page 11 Dimensional drawings: page 16
Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See page 9 for more information on material selection.
Table 1. Rosemount 2088 Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Model Product description
2088 Pressure Transmitter
Measurement type
A Absolute
G Gage
Pressure ranges
Rosemount 2088G Rosemount2088A
1 –14.7 to 30 psi (–1,01 to 2,1 bar) 0 to 30 psi (0 to 2,1 bar)
2 –14.7 to 150 psi (–1,01 to 10,3 bar) 0 to 150 psi (0 to 10,3 bar)
3 –14.7 to 800 psi (–1,01 to 55,2 bar) 0 to 800 psi (0 to 55,2 bar)
4 –14.7 to 4,000 psi (–1,01 to 275,8 bar) 0 to 4,000 psi (0 to 275,8 bar)
Transmitter ou tput
S 4–20 mA dc/Digital HART Protocol
N 1–5 Vdc Low Power/Digital HART Protocol
Materials of construction
Process connection Isolating diaphragm Fill fluid
22 316L SST 316L SST Silicone
33 Alloy C-276 Alloy C-276 Silicone
2B 316L SST 316L SST Inert
(1)
(2)
Emerson.com/Rosemount
3
Rosemount 2088
September 2017
Table 1. Rosemount 2088 Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Process connection
A
B
D
C
(3)
(3)(4)
(3)(4)
1
/2–14 NPT female
DIN 16288 G 1/2 male
M20 1.5 male
RC 1/2 female
Conduit entry
1
2
4
(3)
(3)(5)
1
/2–14 NPT
M20 1.5
G 1/2
Options (Include with selected model number)
Extended product warranty
WR3 3-year limited warranty
WR5 5-year limited warranty
Diaphragm seal assemblies
(6)(7)
S1 Assemble to one Rosemount 1199 Diaphragm Seal
Display and interface
M4 LCD display with local operator interface
M5 LCD display, configured for engineering units
Configuration buttons
D4 Analog zero and span
DZ digital zero trim
Mounting brackets
B4 SST mounting bracket with SST bolts
Product certifications
C6 Canada Explosionproof, Intrinsic Safety, Division 2, and Dust-Ignitionproof
E2 INMETRO Flameproof
E3 China Flameproof
(3)(8)
E4
E5 USA Explosionproof (XP) and Dust-Ignitionproof (DIP)
E7 IECEx Flameproof
ED AT EX Fla meproof
EM Technical Regulations Customs Union (EAC) Flameproof
(3)
I1
I2 INMETRO Intrinsic Safety
Japan Flameproof
ATEX Intrinsic Safety
4
Emerson.com/Rosemount
September 2017
Rosemount 2088
Table 1. Rosemount 2088 Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
I3 China Intrinsic Safety
I5 USA Intrinsic Safety (IS) and Nonincendive (NI)
I7 IECEx Intrinsic Safety
IM Technical Regulation Customs Union (EAC) Intrinsic Safety
K1 ATEX Flameproof, Intrinsic Safety, Type n, Dust
K2 INMETRO Flameproof, Intrinsic Safety
K5 USA Explosionproof, Dust-Ignitionproof, Intrinsic Safety, and Nonincendive
(3)
K6
K7 IECEx Flameproof, Intrinsic Safety, Type n, Dust
Canada Explosion-Proof, Intrinsic Safety, Division 2, Dust-ignitionproof, and ATEX Flameproof, Intrinsic Safety
KB
USA Explosionproof, Dust-Ignitionproof, Intrinsic Safety, Nonincendive, and Canada Explosionproof, Dust-Ignitionproof, Intrinsic Safety, Division 2
KM Technical Regulation Customs Union (EAC) Flameproof and Intrinsic Safety
(3)
KH
N1
ATEX Flameproof, Intrinsic Safety, and USA Explosion-Proof, Dust Ignitionproof, Intrinsic Safety, and Nonincendive
(3)
ATEX Type n
N3 China Type n
N7 IECEx Type n
(3)
ND
ATE X Dust
NK IECEx Dust
Shipboard approvals
SBS American Bureau of Shipping (ABS) Type Approval
SBV Bureau Veritas (BV) Type Approval
SDN Det Norske Veritas (DNV) Type Approval
SLL Lloyd's Register (LR) Type Approval
Pressure testing
P1 Hydrostatic testing
Terminal blocks
T1 Transient protection
Special cleaning
P2 Cleaning for special service
Calibration certificate
Q4 Calibration certificate
Quality calibration certificate traceability certification
Q8 Material traceability certification per EN 10204 3.1
Q15 Certificate of compliance to NACE® MR0175/ISO 15156 for wetted materials
Q25 Certificate of compliance to NACE MR0103 for wetted materials
Emerson.com/Rosemount
5
Rosemount 2088
September 2017
Table 1. Rosemount 2088 Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Digital signal
(3)
C4
CN
C5
(9)(10)
C7
(10)
C8
Conduit plug
NAMUR alarm and saturation levels, high alarm
(3)
NAMUR alarm and saturation levels, low alarm
(9)(10)
Custom alarm and saturation levels, high alarm (requires C9 and Configuration Data Sheet)
Custom alarm and saturation levels, low alarm (requires C9 and Configuration Data Sheet)
Low alarm (Standard Rosemount alarm and saturation levels)
(11)
DO 316 SST conduit plug
Configuration
C9 Software configuration
Manifold assemblies
(6)(7)
S5 Assemble to Rosemount 306 Integral Manifold
Calibration accuracy
(12)
P8 0.065% accuracy to 10:1 turndown
Water approval
(13)
DW NSF drinking water approval
Surface finish
Q16 Surface finish certification for sanitary remote seals
Toolkit total system performance reports
QZ Remote seal system performance calculation report
HART Revision configuration
(14)
HR 5
HR7
Configured for HART Revision 5
(15)
Configured for HART Revision 7
(10)
Typical model number: 2088 G 2 S 22 A 1 B4 M5
1. HART Revision 5 is the default HART output. The Rosemount2088 with selectable HART can be factory or field configured to HART Revision 7. To order HART Revision 7 factory configured, add option code HR7.
2. Materials of construction comply with recommendations per NACE MR0175/ISO 15156 for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
3. Not available with Low Power transmitter output code N.
4. Not available with Alloy C-276, materials of construction code 33.
5. Consists of a
6. Use
7. “Assemble-to” items are specified separately and require a completed model number.
8. Only available with conduit thread code 4.
9. Only available with 4–20 mA HART output (output code A).
10. Select configuratio n buttons (option code D4 or DZ) or local operator interface (option code M4) if local configuration buttons are required.
11. Transmitter shipped with 316SST conduit plug (uninstalled) in place of standard carbon steel conduit plug.
12. Requires transmitter output code S with either materials of construction code 22 or 23.
13. Requires materials of construction code 22 with process connection code A.
14. Configures the HART output to HART Revision 5. The device can be field configured to HART Revision 7 if needed.
15. Configures the HART output to HART Revision 7. The device can be field configured to HART Revision 5 if needed.
6
1
/2–14 NPT housing with a G1/2 adapter.
1
/2–14 NPT female process connection code A.
Emerson.com/Rosemount