7.2.2Common Mode Output Voltage Test............................................................................. 47
4User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
1R&S ScopeSuite Overview
The R&S ScopeSuite software is used with R&S RTO/RTO6/RTP oscilloscopes. It can
be installed on a test computer or directly on the oscilloscope. For system requirements, refer to the Release Notes.
R&S ScopeSuite Overview
The R&SScopeSuite main panel has several areas:
●
"Settings": connection settings to oscilloscope and other instruments also default
report settings
●
"Compliance Tests": selection of the compliance test
●
"Demo": accesses demo test cases that can be used for trying out the software
without having a connection to an oscilloscope
●
"Help": opens the help file, containing information about the R&S ScopeSuite configuration
●
"About": gives information about the R&SScopeSuite software
●
"Tile View": allows a personalization of the compliance test selection
You can configure which tests are visible in the compliance test section and which
are hidden, so that only the ones you use are displayed.
► To hide a test from the "Compliance Tests" view, do one of the following:
5User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
a) Right-click on the compliance test you want to hide.
The icon of the test changes, see Figure 1-1. Now with a left click you can hide
the test.
Figure 1-1: Unpin icon
b) Click on "Title View" to show a list of the available test cases. By clicking a test
case in the show list, you can pin/unpin it from the main panel.
R&S ScopeSuite Overview
6User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
2Preparing the Measurements
2.1Test Equipment
PCIe 1.1 / PCIe 2.0 Test Equipment
For PCIe compliance tests, the following test equipment is needed:
●
R&S RTO/RTO6/RTP oscilloscope with at least 6 GHz bandwidth, 2 channels and
20 GS/s available
●
Probes/cables:
–Two SMP to SMA cables or
–Two single-ended probes with at least 6GHz bandwidth bandwidth or
–One differential probe with at least 6GHz bandwidth bandwidth
●
R&S RTO/RTO6/RTP-K81 PCIe compliance test option (required option, installed
on the oscilloscope)
●
Recommended test fixture:
–For Add-In card testing: PCI Express Compliance Base Board (CBB2.0) from
PCI-SIG
–For System Board testing: PCI Express Compliance Load Board (CLB2.0) from
PCI-SIG
●
The free-of-charge R&S ScopeSuite software, which can be installed on a computer or directly on the oscilloscope.
Preparing the Measurements
Test Equipment
PCIe 3.0 Test Equipment
For PCIe compliance tests, the following test equipment is needed:
●
R&S RTP oscilloscope with at least 16 GHz bandwidth, 2 channels and 40 GS/s
available
●
Probes/cables:
–Two SMP to SMA cables or
–Two single-ended probes with at least 6GHz bandwidth bandwidth or
–One differential probe with at least 6GHz bandwidth bandwidth
–Two R&S RTP-ZMA40 Amplifier probe tips.
●
R&S RTP -K83 PCIe compliance test option (required option, installed on the oscilloscope)
●
Recommended test fixture:
–For Add-In card testing: PCI Express Compliance Base Board (CBB3.0) from
PCI-SIG
–For System Board testing: PCI Express Compliance Load Board (CLB3.0) from
PCI-SIG
7User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
●
The free-of-charge R&S ScopeSuite software, which can be installed on a computer or directly on the oscilloscope.
2.2Installing Software and License
The preparation steps are performed only once for each computer and instrument that
are used for testing.
Uninstall older versions of the R&S ScopeSuite
If an older version of the R&S ScopeSuite is installed, make sure to uninstall the old
version before you install the new one. You can find the version number of the current
installation in "Help" menu > "About". To uninstall the R&S ScopeSuite, use the Windows " Control Panel" > "Programs".
For best operation results, we recommend that the installed firmware versions of the
R&S ScopeSuite and the oscilloscope are the same.
Preparing the Measurements
Setting Up the Network
To install the R&S ScopeSuite
1. Download the latest R&S ScopeSuite software from the "Software" section on the
Rohde & Schwarz R&S RTO/RTO6/RTP website:
For system requirements, refer to the Release Notes.
To install the license key on the R&S RTO/RTO6/RTP
► When you got the license key of the compliance test option, enable it on the oscil-
loscope using [Setup] > "SW Options".
For a detailed description, refer to the R&S RTO/RTO6/RTP user manual, chapter
"Installing Options", or to the online help on the instrument.
2.3Setting Up the Network
If the R&S ScopeSuite software runs on a test computer, the computer and the testing
oscilloscope require a LAN connection.
There are two ways of connection:
●
LAN (local area network): It is recommended that you connect to a LAN with DHCP
server. This server uses the Dynamic Host Configuration Protocol (DHCP) to
assign all address information automatically.
8User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
Preparing the Measurements
Connecting the R&S
If no DHCP server is available, or if the Tabor WX2182B or WX2182C is used for
automatic testing, assign fixed IP addresses to all devices.
●
Direct connection of the instruments and the computer or connection to a switch
using LAN cables: Assign fixed IP addresses to the computer and the instruments
and reboot all devices.
To set up and test the LAN connection
1. Connect the computer and the instruments to the same LAN.
2. Start all devices.
3. If no DHCP server is available, assign fixed IP addresses to all devices.
4. Ping the instruments to make sure that the connection is established.
5. If VISA is installed, check if VISA can access the instruments.
a) Start VISA on the test computer.
b) Validate the VISA address string of each device.
See also:
●
Chapter 2.5, "Connecting the R&SRTO/RTO6/RTP", on page 9
RTO/RTO6/RTP
2.4Starting the R&SScopeSuite
To start the R&S ScopeSuite on the test computer or on the oscilloscope:
► Double-click the R&S ScopeSuite program icon.
To start the R&S ScopeSuite on the instrument, in the R&S RTO/RTO6/RTP firmware:
► In the "Apps" dialog, open the "Compliance" tab.
2.5Connecting the R&S RTO/RTO6/RTP
If the R&S ScopeSuite is installed directly on the instrument, the software detects the
R&S RTO/RTO6/RTP firmware automatically, and the "Oscilloscope" button is not
available in the R&S ScopeSuite.
If the R&S ScopeSuite software runs on a test computer, the computer and the testing
oscilloscope require a LAN connection, see Chapter 2.3, "Setting Up the Network",
on page 8. The R&S ScopeSuite software needs the IP address of the oscilloscope to
establish connection.
1. Start the R&S RTO/RTO6/RTP.
2. Start the R&S ScopeSuite software.
9User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
3. Click "Settings" > "Oscilloscope".
Preparing the Measurements
Connecting the R&S RTO/RTO6/RTP
4. Enter the IP address of the oscilloscope.
To obtain the IP address: press the Rohde & Schwarz logo at the top-right corner
of the oscilloscope's display.
5. Click "Get Instrument Information".
The computer connects with the instrument and gets the instrument data.
10User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
If the connection fails, an error message is shown.
2.6Connecting the Arbitrary Waveform Generator
Automatic test execution is possible with all instruments that are listed in the R&S ScopeSuite, in the "Instrument Settings" dialog box. In automatic mode, the R&S ScopeSuite configures the instrument and ensures that the AWG sends the required waveforms. Automatic mode requires a LAN connection and the installation of a VISA implementation (R&S VISA, see www.rohde-schwarz.com/rsvisa) on the computer that is
running the R&S ScopeSuite. If the R&S ScopeSuite is installed on the R&S RTO/
RTO6/RTP, no installation is needed because VISA is already installed on the instrument. If the Tabor WX2182B or WX2182C is used for automatic testing, fixed IP
addresses are required.
For manual test execution, it is recommended to use one of the listed AWGs, but you
can also use another AWG. In manual mode, you connect the AWG to the test board
and configure the instrument manually. VISA is not required. The R&S ScopeSuite
uses VISA if it is installed, otherwise it uses the VXI-11 protocol.
Preparing the Measurements
Connecting the Arbitrary Waveform Generator
To configure the arbitrary waveform generator for automatic testing
1. Connect the computer and the AWG.
2. Set up the LAN connection. See Chapter 2.3, "Setting Up the Network", on page 8.
3. In the R&S ScopeSuite, click "Instruments".
Alternatively, you can select the "Instrument" tab in the test case configuration dialog.
4. Click the "AWG" tab.
5. Select "Operating mode" = "Automatic".
6. Select a supported "AWG Type" and enter its IP address.
For a list of the supported AWGs, see chapter "Test Equipment".
11User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
Preparing the Measurements
Report Configuration
7. Click "Get Instrument Information".
The computer or R&S RTO/RTO6/RTP connects with the instrument and retrieves
the instrument data.
8. If the connection to the arbitrary waveform generator failed, check if the IP address
is assigned correctly.
To configure the AWG for manual testing
► In the "AWG" tab, enable the "Manual" operating mode.
2.7Report Configuration
In the "Report Configuration" menu, you can select the format of the report and the
details to be included in the report. You can also select an icon that is displayed in the
upper left corner of the report.
Also, you can enter common information on the test that is written in the "General Information" section of the test report.
12User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
Preparing the Measurements
Report Configuration
13User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
3Performing Tests
3.1Starting a Test Session
Performing Tests
Starting a Test Session
After you open a compliance test, the "Session Selection" dialog appears. In this dialog, you can create new sessions, open or view existing report.
The following functions are available for handling test sessions:
FunctionDescription
"Add"Adds a new session
"Open"Opens the selected session
"Remove"Removes the selected session
"Rename"Changes the "Session Name"
"Comment"Adds a comment
"Show report"Generates a report for the selected session
To add a test session
1. In the R&S ScopeSuite window, select the compliance test.
2. In the "Session Selection" dialog press "Add".
3. If necessary change the "Session Name"
14User Manual 1333.2299.02 ─ 05
R&S®RTO-K81, R&S®RTO6-K81, R&S®RTP-K81/-K83
To open a test session
1. In the R&S ScopeSuite window, select the compliance test.
2. In the "Session Selection" dialog, select the session you want to open and double
click on it.
Alternatively, select the session and press "Open".
To show a report for a test session
1. In the R&S ScopeSuite window, select the compliance test.
2. In the "Session Selection" dialog, select the session you want the report for and
press "Show report".
3.2Configuring the Test
Performing Tests
Configuring the Test
1. In the R&S ScopeSuite window, select the compliance test to be performed:
●"PCIe"
2. Open a test session, see Chapter 3.1, "Starting a Test Session", on page 14.
3. Adjust the "Properties" settings for the test cases you want to perform.
4. Click "Limit Manager" and edit the limit criteria, see Chapter 3.2.1.1, "Limit Man-
ager", on page 17.
5. If you want to use special report settings the "Report Config" tab to define the format and contents of the report. Otherwise the settings defined in "RSScopeSuite" >
"Settings" > "Report" are used. See Chapter 2.7, "Report Configuration",
on page 12.
6. Click "Test Checked"/"Test Single" and proceed as described in the relevant test
case chapter.
15User Manual 1333.2299.02 ─ 05