RIDGID SR-20 User Manual [en, ru, de, es, fr, it, pl]

SeekTech
SR-20
GB p. 1 DE FR NL
p. 28 p. 55 p. 82
IT p. 109
ES PT
DA
NO
PL p. 245
RU
RIDGE TOOL COMPANY
GB
SeekTech SR-20
Operating Instructions
General Safety Information
WARNING! Read these instructions and the accompanying safety booklet carefully before using
this equipment. If you are uncertain about any aspect of using this tool, contact your RIDGID distributor for more information.
Failure to understand and follow all instructions may result in electric shock, re, and/or serious personal injury.
SAVE THESE INSTRUCTIONS!
seekTech sR-20
DANGER
The SR-20 is a diagnostic tool that senses electromagnetic
elds emitted by objects underground. It is meant to aide the user in locating these objects by recognizing characteristics of the eld lines and displaying them on the screen. As electromagnetic eld lines can be distorted and interfered with, it is important to verify the location of underground objects before digging.
Several utilities may be underground in the same
area. Be sure to follow local guidelines and one-call service procedures.
Exposing the utility is the only way to verify its
existence, location, and depth.
Ridge Tool Co., its aliates and suppliers, will not be
liable for any injury or any direct, indirect, incidental or consequential damages sustained or incurred by reason of the use of the SR-20.
In any correspondence, please give all the information shown on the nameplate of your tool including model number and serial number.
CAUTION: Remove batteries entirely before shipping.
If you have any questions regarding the service or repair of this machine, contact your Ridgid distributor, your local Ridgid oce or Ridge Tool Europe at info.europe@ridgid. com
Ridge Tool Company
1
seekTech sR-20
SR-20 Components
Serial Port Connector
USB Connector
Serial Number Label
Icon Reference
Note: USB/ Serial Ports are for loading new software
Battery Compartment
Display Screen
Keypad
Handle
Speaker
Antenna Mast
Upper Antenna
Node
Folding Joint
Guidance Antennas
Lower Antenna
Node
Folding Mast Snap
Figure 1: SR-20 Components
2
Ridge Tool Company
seekTech sR-20
Introduction to the SR-20
Getting Started
Installing/Changing Batteries
Figure 2: Battery Case
CAUTION: Do not allow debris or moisture into battery
compartment. Debris or moisture may short the battery contacts, leading to rapid discharge of the batteries, which could result in electrolyte leakage or risk of re.
Folding Mast
SR-20 Modes
The SR-20 operates in three distinct modes. They are:
1. Active Line Trace Mode, used when a chosen frequency can be put onto a long conductor using a Line Transmitter, for locating conductive pipes, lines, or cables.
2.
Passive Trace Mode, used for tracing electrical lines
that are already carrying 60 Hz current (U.S.), 50 Hz current (Europe), or radio frequencies.
Sonde Mode, used for locating Sondes in pipes,
3. conduits, or tunnels that are non-conductive or cannot otherwise be traced.
IMPORTANT! Do not snap or whip the SR-20 mast to open or close it. Open it and close it by hand only.
NOTE: Avoid dragging the lower antenna node on the ground while locating with the SR-20. It may cause signal noise which will interfere with results, and may eventually damage the antenna.
Release Button
Figure 3: Folding Antenna Mast and Release Button
Ridge Tool Company
3
seekTech sR-20
Current Strength/ Signal Angle
Battery Level
Crosshairs (Map Center)
Active View Area
Frequency
Mode
Depth/
Distance
Proximity
Signal
Guidance Arrows
Tracing Line
Signal Strength
Display Elements
The “basic features” of the SR-20 are on by default. Features can be turned o or hidden to make the display clearer when doing basic locating in uncomplicated situations.
Common Display Elements
Current Strength/ Signal Angle
Frequency Mode
Depth/ Distance
Figure 4: Common Display Elements
The display screen in Active Line Trace, Passive Line Trace or Sonde mode will show the following features:
Active View Area – The area inside the circle on the
• SR-20 display where the Tracing Line, Guidance Arrows, and crosshairs are displayed.
mA Current Strength – Proportional to current on
the line. Switches to Signal Angle when Signal Angle is greater than 35°.
Signal Angle – Field tilt from the horizontal; angle
• toward the eld’s center; numeric value displayed in degrees.
Battery Level – Indicates level of remaining
battery capacity.
Measured Depth/Distance – Displays the measured
• depth when receiver is touching the ground directly over signal source. Displays computed distance when the antenna mast is pointed at a signal source in some other manner. Displays feet/inches (U.S.A. default) or meters (European default).
Battery Level
Crosshairs (Map Center)
Active View Area
Display Elements: Active Line Trace Mode
Proximity Signal
Tracing Line
Guidance Arrows
Signal Strength
Figure 5: Display Elements (Line Trace Mode)
In Active Line Trace Mode, the following features will also be displayed:
Proximity Signal – Numerical indication showing how
• close the signal source is to the locator. Displays from 1 to
999. (Line Trace modes only)
Signal Strength – Strength of signal as sensed by the
• lower Omnidirectional antenna.
Tracing Line – The Tracing Line represents the
approximate axis of the detected eld. It represents detected distortion in the eld by appearing less focused. (See page 22 for information on setting the sensitivity and how to enable or disable the distortion response in the Tracing Line.)
Distortion Line – If the normal distortion
response of the Tracing Line is disabled, a second line is shown, which represents the signal from the upper antenna node. By comparing the two lines, the user can estimate the degree of distortion present in a signal.
Guidance Arrows – The Guidance Arrows serve to
steer the operator toward the center of the detected eld, by showing when the signals reaching the left and right.
Mode – Icon for Sonde
Line Trace)
, or Radio Frequency mode.
, Line Trace , Power (Passive
Frequency – Shows current frequency setting in hertz or kilohertz.
+ Crosshairs (Map Center) – shows operator’s position
• relative to the target center.
4
Ridge Tool Company
seekTech sR-20
Display Elements: Passive Trace Mode
The screen elements in Passive Trace Mode are the same as those seen in Active Line Trace mode.
Display Elements: Sonde Mode
Pipe Direction
Signal Strength
Figure 6: Display Elements: Sonde Mode
In Sonde mode, the screen elements include several features that are unique to Sonde locating.
| | Pipe Direction – Represents the approximate
• direction of Sonde.
Sonde Icon – Appears when approaching the
• location of a Sonde.
Equator – Represents the mid-line of the Sonde’s
eld perpendicular to the axis of the Poles.
Pole Icon
Zoom Ring
Equator
Sonde Icon
Default Frequencies
Currently available frequencies in default setting include:
Sonde Mode
• 512 Hz
Active Line Trace Mode:
• 128 Hz
• 1 kHz
• 8 kHz
• 33 kHz
Passive Line Trace Mode:
th
• 50 Hz (9
)
• < 4 kHz
Radio Frequency
• 4 kHz—15 kHz (L)
• > 15 kHz (H)
Pole Icon – Represents the location of either of the
• two Poles of the Sonde’s dipole eld.
Zoom Ring – Appears when the locator moves
close to a Pole.
The use of these features is described in the Active Line Tracing, Passive Line Tracing, and Sonde Locating sections.
Ridge Tool Company
5
seekTech sR-20
Keypad
Up Key Menu Navigation/Signal Focus/ Proximity Threshold Control
Volume Controle Key
Light Sensor
Frequency Key
Figure 7: Keypad
Power On/O Key – Powers SR-20 on.
Up and Down Keys – Used for locating choices during
menu selection.
Signal Focus – If activated, the Up and Down Keys
will change the Signal Focus setting up and down. A long
press (greater than ½ second) on these keys will adjust the Proximity Threshold, while a quick press will adjust the Signal Focus.
Select Key Audio Tone Reset/Menu Item Select
Menu Key
Power ON/ OFF Key
Down Key Menu Navigation/Signal Focus/ Proximity Threshold Control
Operation Time
Using alkaline cells, typical operation time is from about 12 to 24 hours depending on sound volume and how often the backlight is on. Other factors that aect the operation time will include chemistry of the battery (many of the new high performance batteries, such as the “Duracell® ULTRA” last 10%-20% longer than conventional alkaline cells under high demand applications). Operation at lower temperatures will also reduce battery life.
To preserve battery life, the SR-20 will automatically shut down after 1 hour of no key presses. Simply power the unit on to resume use.
Low Battery Warning
When the battery gets low, a battery icon will periodically appear in the map area on the screen.
Select Key – Used to make a choice during Menu
• selection; in normal operation, to force a Measured Depth reading and recenter audio tone.
Menu Key – Used to display a “tree” of choices including
• frequency selections, display element choices, brightness and contrast, and restoring default settings. In a menu, will move up one level.
Volume Control Key – Used to raise or lower the volume
setting.
Frequency Key – Used to set the In-Use Frequency of the
SR-20 from the set of Checked-Active frequencies. The list of frequencies that have been set to Checked-Active status can be modied via the Menu Key. Frequencies are grouped into four sets: Sonde Frequencies (
(
), Power Frequencies ( ) and Radio Frequencies ( ).
Each press cycles to the next Checked-Active frequency.
Light Sensor – In Automatic mode, the light sensor controls
• when the backlight goes on or o depending on ambient light.
), Line Trace Frequencies
Figure 8: Low-Battery Warning
Just before complete shut down there will be a non-interruptible power down sequence. An extended buzz will sound when the SR-20 is about to go into shutdown sequence.
NOTE: Voltage on rechargeable batteries may sometimes drop so quickly that the unit will just shut down. The unit will power down and restart. Just replace the batteries and power the unit back on.
6
Ridge Tool Company
seekTech sR-20
Starting Up
After pressing the Power Key logo displays, and the software version number will appear on the left of the screen.
Figure 9: Start-up Screen
on the keypad, the RIDGID
Set Up
Once the SR-20 is up and running the next step is to set up the frequencies needed that match the transmitter or line to be located.
Checked-Active frequencies are already selected for use and
appear in sequence by pressing the Frequency Key . (For example, the line trace frequency of 33 kHz is available by pressing the Frequency Key.)
Activating Frequencies
Frequencies can be chosen for the set of Checked-Active frequencies so they will be available using the Frequency
Key
.
Each frequency is activated by choosing it from a list in the Main Menu (See Figure 13). Frequencies are grouped by category:
onde
S
Active Line Trace Passive Line Trace
Radio
1. Push the Menu Key
:
Figure 10: Frequency Key
Figure 11: Line Trace Frequency Selected with Frequency Key
Figure 12: Menu Key
The Main Menu screen is then activated:
Figure 13: Main Menu
2. Using the Up and Down Keys, highlight the frequency desired (Figure 14). In this example, the
operator is activating a 128 Hz frequency.
Figure 14: Highlighting a Desired Frequency (128 Hz)
Ridge Tool Company
7
seekTech sR-20
3. Press the Select Key
box for each frequency to be used.
Figure 15: Select Key
Figure 16: Desired Frequency Checked
4. Frequencies that have been selected for use will show a check in the box next to them.
Press the Menu Key
5. and exit.
(shown below) to check the
again to accept the choice
Sounds of the SR-20
The sound level is driven by the proximity to the target. The closer to the target, the higher the sound pitch will be. A rising tone indicates increasing signal.
In Active Line Trace or Passive Line Trace mode, sound is on one continuous curve and does not rescale.
When there is no distortion present, the sound of the SR-20 is a clear warbling sound when on the left side of the detected eld, with a slight click added when on the right side of the detected eld. If distortion is detected a sound similar to AM radio static sound can be heard, which gets stronger as the degree of distortion increases. If the distortion response feature is disabled, the static sound does not occur.
In Sonde Mode, the pitch will “ratchet” upward. That is, it will rise and then rescale (fall) in pitch while approaching the Sonde. Moving away from the Sonde, it will drop to a lower pitch and remain there as long as one moves away from the Sonde.
If desired, force the sound to recenter at a medium level (in any mode) by pressing the Select Key during operation.
Key Items in Using the SR-20
SIGNAL STRENGTH represents the strength of the eld being detected by the lower antenna node of the SR-20, converted mathematically for scalability. In a clear and undistorted eld, you can locate based on Signal Strength alone.
PROXIMITY SIGNAL reects the proximity of the locator to the target utility; the closer the locator moves to the center of the detected eld, the higher the Proximity Signal number gets. The Proximity Signal is calculated from the ratio of the signals received at the lower and upper antennas, adjusted for scalability.
Figure 17: Menu Key
The Main Menu lists all frequencies available for activation. For information on adding additional frequencies to the Main Menu so they can be chosen for activation, see “Frequencies Selection Control” on page 22.
8
Ridge Tool Company
DISTORTION is the degree to which the eld detected is deformed from the simple circular shape of an ideal magnetic eld caused by current in a long conductor. If multiple elds are present, the detected eld is pushed or pulled out of shape and the dierent antennas will pick up dierent eld strengths. Distortion is reected by the Tracing Line growing unfocused instead of sharp on the display screen.
GUIDANCE ARROWS are driven by the signals received at the side-wheel antennas of the SR-20. When the elds detected by these side antennas are equal, the arrows will center. If one is receiving a stronger eld signal than the other, the arrows will point toward the probable center of the target conductor.
Line Tracing with the SR-20
Active Line Tracing
In active line tracing, underground lines are energized with a Line Transmitter.
Line transmitters energize lines by direct connection with clips, by directly inducing the signal using a clamp, or by inducing the signal using inductive coils built into the transmitter.
seekTech sR-20
Figure 18: Line Trace Frequency Chosen with the Frequency Key
WARNING: Connect the ground lead and the power lead
of the transmitter before powering the transmitter on, to avoid electric shock.
1.
Energize the target conductor according to the
transmitter manufacturer’s instructions. Select the transmitter frequency. Set the frequency used on the SR-20 to the same frequency used on the transmitter using the Frequency Key. Be sure the frequency has a line trace icon
Direct Connect Method: The transmitter is attached by direct metal-to-metal connection to the target conductor at some access point such as a valve, a meter, or other point. Important: The connection between the transmitter and the conductor must be a clean, rm connection. The transmitter is also connected to a ground stake providing a strong open path to ground. Important: A weak ground connection is the most frequent cause of a poor tracing circuit. Make sure the transmitter is well connected to ground, and has enough exposure to the ground to allow current to ow through the circuit.
Inductive Clamp Mode: The transmitter is connected to an inductive clamp which is then closed around a pipe or cable. The transmitter energizes the clamp, which then induces a current in the conductor.
Inductive Mode: The transmitter is placed over the conductor, at right angles to it. There is no direct connection; the internal coils of the transmitter generate a strong eld through the ground which induces a current in the underground conductor of interest. Important: If the transmitter is too close to the SR-20 in this mode, it can cause “air-coupling” which means the locator is reading on the transmitter’s eld, not on the target conductor.
.
(This screen will ash briey when a
new frequency is chosen)
2.
Observe the Proximity Signal to ensure that the
receiver is picking up the transmitted signal. The Proximity Signal should peak over the line and drop o on either side.
3.
When tracing, the direction the pipe or cable is
running will be shown on the screen by the Tracing Line. The Tracing Line will be a clear, single line if the
eld being detected is undistorted.
Tracing Line
Figure 19: Tracing Line Showing Low Distortion
4. If other elds are interfering in some way, the distortion caused by those elds will be reected by a blurring of the Tracing Line. This alerts the operator that the apparent axis of the line may be inuenced by other elds, and requires careful evaluation. The more distorted the detected eld, the broader the cloud around the Tracing Line will be.
The Tracing Line has three important functions.
It represents the location, and the direction, of the signal being traced. It reects changes in direction of the target utility — when the utility makes a turn, for example. And it helps recognize signal distortion. It does this by becoming cloudier as distortion increases.
Ridge Tool Company
9
seekTech sR-20
Tracing Line
Figure 20: Tracing Line Showing High Distortion
Use the Guidance Arrows, Proximity Number, Signal Strength, and Tracing Line to guide the line trace. These
pieces of information are generated from discrete signal characteristics to help the operator understand the quality of the locate. An undistorted signal emitted from a line is strongest directly over that line. (Note: Unlike the Signal Trace lines, the guidance arrows require that the user orient the locator so that the guidance arrows point 90 degrees to the Signal Trace line. (See Figure 21).
WARNING: Care should be taken to watch for signal
interference that may give inaccurate readings. The Tracing Line is only representative of the position of the buried utility if the eld is UNDISTORTED. Do NOT base a locate solely on the Tracing Line.
Always cross check the locate by ensuring that:
The Tracing Line shows little or no distortion response
(blurriness).
The Proximity Signal and the Signal strength maximize
when the Tracing Line crosses the map center.
The Measured Depth increases appropriately as the
• unit is raised vertically and the Tracing Line remains aligned.
Measured Depth readings should be taken as estimates and actual depths should be independently veried by potholing or other means prior to digging.
5.
Note that an undistorted line will also be clear
rather than blurred on the screen, and the sound accompanying the image will have no “static” in it.
6.
Condence in the accuracy of a locate can be
increased by maximizing the Proximity Signal (and/ or Signal Strength), balancing the Guidance Arrows and centering the Tracing line on the screen. Conrm a locate by testing whether the Measured Depth reading is stable and reasonable. (See page 12.)
Tracing Line
Figure 21: High Probability Locate
As always, the only way to be certain of the location of a utility is through visual conrmation by exposing the utility. The accuracy of position and depth measurement improves as the SR-20 lower antenna node is placed closer and closer to the target utility. Rechecking the Measured Depth and position periodically during the excavation process can help avoid damage to a target utility and may identify additional utility signals that were not noticed prior to excavation.
When line tracing, it is important to remember that tees, curves, other conductors in the vicinity, and nearby masses of metal can add distortion to the eld, requiring closer scrutiny of the data to determine the true path of the target utility.
See below for tips on improving the signal.
Circling the last location of a clear signal at a distance of about 20 feet (6.5 m) can clarify if the distortion is coming from a local turn or tee in the line, and enable the operator to again pick up the line nearby.
If the signal is clear, the SR-20 will often show a straight signal line with very little distortion right up to a 90-degree tee, show a small amount of distortion as it follows around the curve, and then show a clear signal again as it resumes its travel after the tee. It shows very clearly when the line is turning.
10
Ridge Tool Company
Operating Tips for Active Line Tracing
• The SR-20 quickly identies distorted elds. If the guidance arrows are centered on the screen, and the Trace Line is not centered (or if the Proximity Signal number and Signal Strength are not maximized), then distortion is creating a complex non-circular eld.
To improve the tracing circuit:
a) Try changing the frequency used to a lower one. b) Move the ground stake position away from the line
to be traced. Use a larger ground contact surface (e.g., a shovel blade).
Make sure that the line is not commonly bonded to
c)
another utility. (Undo common bonds only if safe to do so).
Move the transmitter to a dierent point on the line,
d)
if possible.
If the Tracing Line will not center or if it moves across
the screen erratically, then the SR-20 may not be receiving a clear signal. The Measured Depth and the Proximity Signal may also be unstable under these circumstances.
a)
Check the transmitter to be sure that it is operating
and well grounded. Good connection and good grounding overcome low current problems.
b
) Test the circuit by pointing the lower antenna at
either transmitter lead.
Check that the SR-20 and transmitter are operating
c)
on the same frequency.
Try dierent frequencies, starting with the lowest,
d)
until the line can be picked up dependably. Using lower frequencies can overcome bleed over problems.
Relocate the ground connection for a better circuit.
e)
Ensure there is enough contact (ground stake is suciently deep) especially in dryer soils.
) In extremely dry soil, wetting the area around the
f
ground stake will improve the circuit. Be aware the moisture will dissipate and evaporate, reducing the quality of the circuit over time.
Using the numeric Signal Angle Indicator is another
way to check for distorted signals.
seekTech sR-20
M
ove the SR-20 perpendicularly to both sides of the
traced line until the numeric Signal Angle indicator reads
degrees. Be sure to keep the lower Omnidirectional
45 antenna node at the same height, and the locator mast vertical. If there is little or no distortion the traced line should be in the middle and the distance to each 45
degree point should be approximately the same on either side. If the signal is undistorted, then the distance from the line center to the 45° point is approximately equal to the depth.
A
nother variation of this technique is to move the same
distance to the right and left of the traced line, say
inches (60 cm) and check that the Signal Strength
24 readings are similar.
= Same distance
Ground
Energized Pipe
Figure 22: Checking for Distortion
• While tracing, the Proximity Signal and Signal Strength should maximize, and the Measured Depth minimize, at the same place where the guidance arrows center on the display. If this is not the case, the utility may be changing direction or other coupled signals may be present.
Higher frequencies bleed over to adjacent utilities more
• readily, but may be needed to overcome breaks in tracer wires or go over insulating couplers. If the line is ungrounded at the far end, higher frequencies may be the only means to make the line traceable.
When using the transmitter inductively, be sure to begin
• the locate about 30 feet (10 m) away to avoid “direct coupling” (also know as air coupling).
Ridge Tool Company
11
seekTech sR-20
• While tracing, the mapping display operates best under the following conditions:
1.
The line is level
2. The SR-20 Locator is above the target utility elevation
3. The SR-20 antenna mast is held approximately
vertical
If these conditions are not met, pay close attention to maximizing Signal Strength.
In general, if the SR-20 is used in a zone over the target line within a sweep area of about two “depths” of the line, the map will be useful and accurate. Be aware of this when using the map if the target or line is very shallow. The width of the useful search area for the map can be small if the line is extremely shallow.
Measuring Depth (Line Tracing Modes)
The SR-20 calculates Measured Depth by comparing the strength of the signal at the lower antenna to that at the upper antenna.
Measured Depth is measured correctly in an undistorted eld when the bottom antenna is touching the ground directly above the signal source and the antenna mast is vertical.
To measure depth, place the locator on the ground,
1.
directly above the Sonde or the line.
Current and Signal Angle Reading
The Current Strength (mA) and Signal Angle indicator ( ) in the upper right corner of the screen will display the current detected on the traced line, in milliamps, when the computed angle to the center of the detected eld is less than 35° and the SR-20 crosses the center of the eld as sensed by the guidance arrows.
When moving across the center of the eld the current display will “latch” the displayed current value (retain it in the display) until the guidance arrows reverse again, at which point the latched display will be updated. The update and latch cycle occurs whenever the guidance arrows reverse.
When the angle to the center exceeds 35°, the Signal Angle indicator will again replace the Current indicator, and the display will show the computed angle to the center of the detected eld.
Clipping (Tracing Modes)
Occasionally the Signal Strength will be strong enough that the receiver will be unable to process the whole signal, a condition known as “clipping”. When this occurs, a warning
symbol signal is particularly strong. If clipping persists, remedy it by increasing the distance between the antennas and the target line OR by reducing the strength of the current from the transmitter.
will appear on the screen. It means that the
2.
Measured Depth will be shown in the lower left hand
corner.
A Measured Depth reading can be forced by pressing
3.
the Select Key.
Measured Depth will be accurate only if the signal is
4.
undistorted and the antenna mast is held vertical.
Testing for the consistency of the Measured Depth reading can be done by raising the SR-20 a known distance (say, 12 inches (33 cm)) and observing whether the Measured Depth indicator increases by the same amount. Small variation is acceptable, but if the Measured Depth does not change, or changes drastically, it is an indication of a “distorted” eld, or very low current on the line.
NOTE: In Active Line Trace or Passive Line Trace modes, pressing and holding the Select Key will force a Measured Depth reading and will force the Signal Angle indicator to change to Current. If sound is set on, it will also recenter the audio tone.
NOTE: Measured Depth Display is disabled under clipping conditions.
12
Ridge Tool Company
seekTech sR-20
Current Strength
Current Strength Signal Angle
Figure 23: Screen Display in Dierent Locations (Line Tracing)
Ridge Tool Company
13
seekTech sR-20
Passive Line Tracing
In passive mode, the SR-20 is looking for electromagnetic “noise” that has found its way onto a buried utility line by any available means.
Electromagnetic signals can get onto buried utility lines in a variety of ways.
The most common reason is by means of direct connection to some signal source. All operating electronic devices that are connected to AC power will radiate a certain amount of electronic “noise” back onto the power lines they are connected to.
In some areas for example, buried utilities act as antennas for high powered, low frequency radio transmissions (submarine navigational and communication signals in the UK for example) and will reradiate these signals. These reradiated signals can be very useful for locating.
In short, frequencies can show up on buried conductors in numerous ways, and these can be picked up passively, if the elds are strong enough.
Select a Passive Line Trace Frequency (
1.
or icon).
The 9x multiple is the setting most commonly used to
locate 50/60 Hz signal. In well-balanced high voltage electric distribution systems, the 5x multiple may work better. The 100 Hz (in 50 Hz countries) and 120 Hz (in 60
Hz countries) frequency settings are particularly useful for pipelines that have been equipped with cathodic protection using rectiers.
As in Active Line Tracing, the Tracing Line will reect distortion in the detected eld by appearing unfocused or cloudy in proportion to the distortion. This “distortion response” is useful in recognizing when the eld being traced is being distorted by other elds of metallic objects in the vicinity.
3.
There are also two additional radio frequency
bands
• 4kHz to 15kHz (LF)
• > 15kHz (HF)
The Radio Frequency and <4 kHz bands can be useful in discriminating when tracing in a noisy environment. They are also very helpful in nding lines on blind searches. When searching over a wide area where the location of targets is unknown, one useful approach is to have multiple frequencies selected for use and to check the area at a number of frequencies in sequence looking for meaningful signals.
to help locate lines passively. They are:
9th
Figure 24: 60
2. The SR-20 has multiple Passive Line Trace frequency settings. Power frequencies (identied
with the power icon generated as the result of power transmissions, usually 50 or 60 Hz. To reduce the eects of inherent noise from line-load or neighboring devices the SR-20 can be set to locate various multiples (or harmonics) of the base 50/60 Hz frequency up to 4,000 Hz.
Hz Passive Trace Frequency
) are used to locate signals
In general, directly connected Active Line Tracing is more reliable than Passive Line Tracing.
WARNING: In Passive Line tracing, or when signals are
extremely weak, the Measured Depth will generally read too DEEP and the actual buried depth may be MUCH shallower.
14
Ridge Tool Company
seekTech sR-20
Operating Tips for Passive Line Tracing
1. In Passive Locating if you are looking for a known line, be sure you are using the best frequency for the line in question. This may be, for example, be 50 Hz (1) for a power line, or it may turn out that 50 Hz (9) produces a more reliable response on a particular line.
2.
If seeking a cathode-protected pipe in Passive Mode,
use higher-frequency (greater than 4 kHz) to pick up harmonics.
Remember that pipes can carry currents that will
3. show up on a Passive Trace as well as cables will; the only guarantee of a locate is inspection.
In general, Passive Trace locating is less reliable than
4. Active Line Tracing because Active Line Tracing oers the positive identication of the signal from the transmitter.
Especially in Passive Line Tracing, knowing that you
5. have found something is not the same as knowing what you have found. It is essential to use all the indicators available, such as Measured Depth, Signal Strength, etc., to conrm a locate. If it is possible to nd part of a passively-located cable, it can then be energized using a transmitter and positively traced.
While Passive Line Trace is most often used on
6. 50/60 Hz power lines, other cables such as phone lines, CATV lines, etc., can be energized by transient radio frequencies in the region and may appear on Passive Line Trace searches.
Sonde Locating
The SR-20 can be used to locate the signal of a Sonde (transmitter).
IMPORTANT! Signal strength is the key factor in determining the Sonde’s location. Take care to maximize the Signal Strength prior to marking an area for excavation. The following assumes that the Sonde is in a horizontal pipe, the ground is approximately level and the SR-20 is held with the antenna mast vertical.
The eld of a Sonde is dierent in form from the circular eld around a long conductor such as a pipe or cable. It is a dipole eld like the eld around the Earth, with a north Pole and a south Pole.
Figure 25: Earth’s Dipole Field
In the Sonde’s eld, the SR-20 will detect the points at either end where the eld lines curve down toward the vertical, and it will mark these points on the map display with a “Pole” icon ( Sonde, centered between the Poles, known as the “Equator”, much like the Equator on a map of the Earth if the planet were viewed sideways (See Figure 25).
). The SR-20 will also show a line at 90 degrees to the
Note that because of the SR-20’s Omnidirectional antennas, the signal stays stable regardless of orientation. This means the signal will increase smoothly when approaching the Sonde, and decrease smoothly moving away.
NOTE: A Pole is found where eld lines turn vertical. The Equator occurs when the eld lines are horizontal.
Ridge Tool Company
15
seekTech sR-20
Pole
Figure 26: Dipole Field
When locating a Sonde, rst set up the locate:
Activate the Sonde before putting it in the line. Select
the same Sonde frequency on the SR-20 and make sure it is receiving the signal.
After the Sonde has been sent into the pipe, go to the suspected Sonde location. If the direction of the pipe is unknown, push the Sonde a shorter distance into the line (~15 feet (5 m) from the access is a good starting point).
Pole
Equator
Location Methods
There are three major parts to locating a Sonde. The rst step is to localize the sonde. The second part is pinpointing. The third is verifying its location.
Step 1: Localize the sonde
• Hold the SR-20 so the antenna mast is pointing outward. Sweep the antenna and listen to the sound, it will be highest when the antenna mast is pointing in the direction of the Sonde.
• Lower the SR-20 to its normal operating position (antenna mast vertical) and walk in the direction of the Sonde. Approaching the Sonde, the Signal Strength will increase and the audio tone will rise in pitch. Use the Signal Strength and the sound to maximize the signal.
Maximize the Signal Strength. When it appears to
• be at its highest point, place the SR-20 close to the ground over the high signal point. Be careful to hold the receiver at a constant height above the ground as distance aects Signal Strength.
Note the Signal Strength and move away from the
• high point in all directions to verify that the Signal Strength drops signicantly on all sides. Mark the point with a yellow Sonde Marker.
16
Figure 27: Poles and Equator of a Sonde
If while “getting closer” the Equator appears on the screen, follow it in the direction of an increasing Signal Strength to localize the Sonde.
Ridge Tool Company
seekTech sR-20
Step 2: Pinpoint the Sonde
The Poles should appear on either side of the maximum signal point, an equal distance on either side if the Sonde is level. If they are not visible on the screen at the point of maximum Signal Strength, move from the maximum point perpendicular to the dotted line (Equator) until one appears. Center the locator over the Pole.
Where the Poles occur depends on the Sonde’s depth. The deeper the Sonde, the further away from it the Poles will be.
The dotted line represents the Equator of the Sonde. If the Sonde is not tilted, the Equator will intersect the Sonde at maximum Signal Strength and minimum Measured Depth.
NOTE: being on the Equator does not mean that the locator
is over the Sonde. Always verify the locate by maximizing Signal Strength and marking both Poles.
Mark the rst Pole location found with a red triangular
• Pole marker. After centering on the Pole, a double line indicator represents how the Sonde is lying underground, and in most cases also represents the pipe’s approximate direction.
When the locator gets close to a Pole, a zoom ring will
• appear centered on the Pole.
The second Pole will be a similar distance from the
• Sonde location in the opposite direction. Locate it in the same manner and mark it with a red triangular marker.
If the Sonde is level, the three markers should be
• aligned and the red Pole markers should be similar distances from the yellow Sonde marker. If they are not, a tilted Sonde may be indicated. (See “Tilted Sonde”) It is generally true that the Sonde will be on the line between the two Poles, unless there is extreme distortion present.
Step 3: Verify the locate
• It is important to verify the Sonde’s location by crosschecking the receiver’s information and maximizing Signal Strength. Move the SR-20 away from the maximum Signal Strength, to make sure that the signal drops o on all sides. Make sure to move the unit far enough to see a signicant signal drop in each direction.
Figure 28: Sonde Locate: Equator
• Double check the two Pole locations.
• Notice that the Measured Depth reading at the maximum Signal Strength location is reasonable and consistent. If it seems far too deep or too shallow, recheck that there is an actual maximum Signal Strength at that location.
Notice that the poles and the point of highest Signal
• Strength lie on a straight line.
IMPORTANT! Remember that being on the Equator does not mean one is over the Sonde. Seeing two Poles aligned on the display is not a substitute for centering over each Pole separately and marking their locations as described above.
If the Poles are not visible, extend the search. For best accuracy the SR-20 should be held with the mast
oriented vertically. The antenna mast must be vertical when marking the Poles and Equator, or their locations will be less accurate.
Ridge Tool Company
17
seekTech sR-20
Tilted Sondes
If the Sonde is tilted, one Pole will move closer to the Sonde and the other farther away.
If the Sonde is vertical what is seen on the screen is a single Pole at the point of maximum Signal Strength. (The Ridgid Floating Sonde is designed to have a single Pole “visible” and is weighted to maintain the Sonde on a vertical axis.) Maximizing the Signal Strength will still guide to the best location for the Sonde.
Floating Sondes
Some Sondes are designed to be ushed or to drift down a pipe pushed by water ow. The only guarantee of having located a oating Sonde is maximizing the Signal Strength and double checking that the signal falls away on every side of the maximum signal location.
Measuring Depth (Sonde Mode)
The SR-20 calculates Measured Depth by comparing the strength of the signal at the lower antenna to the upper antenna. Measured Depth is approximate; it will usually reect the physical depth when the mast is held vertical and the bottom antenna is touching the ground directly above the signal source, assuming no distortion is present.
1.
To measure depth, place the locator on the ground,
directly above the Sonde or the line.
2.
Measured Depth will be shown in the lower left hand
corner of the SR-20’s display screen.
A Measured Depth reading can be forced by pressing
3. the Select Key during a locate.
Measured Depth will be accurate only if the signal is
4. undistorted.
Clipping (Sonde Mode)
Occasionally the Signal Strength will be strong enough that the receiver will be unable to process the entire signal, a condition known as “clipping”. When this occurs, a warning
symbol signal is particularly strong.
will appear on the screen. It means that the
NOTE: Measured Depth Display is disabled under clipping conditions.
18
Ridge Tool Company
seekTech sR-20
Figure 29: Screen Display in Dierent Locations (Sonde)
Ridge Tool Company
19
seekTech sR-20
Normal
Maximum Signal Strength
Tilted
Figure 30: Tilted Sonde, Poles, and Equator
Note the right-hand Pole is closer to the Equator, due to tilt.
20
Ridge Tool Company
seekTech sR-20
Menus and Settings
Pressing the Menu Key brings up a series of choices (see Figure 31).
Auto Menu Exit Countdown Timer
In sequence from the top of the menu down, the Main Menu presents the following items:
1.
Currently Available Sonde Frequencies
(Checked-Active or not).
Currently Available Active Line Trace
2. Frequencies (Checked-Active or not).
Currently Available Passive Line Trace
3. Frequencies (Checked-Active or not).
Figure 31: Main Menu
® Auto Menu Exit Count-down Timer
While traversing the menu tree, a counter appears at the bottom of the screen counting down.
Currently Available Frequencies
Frequencies that have been set to “Checked-Active” status appear with a check box next to them.
x9
NOTE: Superscripts indicate harmonics; e.g., 60
= 540 Hz
and 50 Hzx9 = 450 Hz.
Change of Depth Units
Back Light Control
A light detector built into the upper left corner of the keypad senses low light levels. The backlight can be forced on by blocking the light to this sensor.
LCD Contrast
When this is selected by pressing the Select Key, the contrast can be adjusted. Use the Up and Down Keys to make the screen lighter or darker.
Use the Menu Key to save the setting and exit. In this menu, one can also exit by pressing the Select Key to save the setting and exit.
Currently Available Radio Frequencies (Low and
4. High) (Checked-Active or not).
Depth Measurement Units Setting
5.
Backlight Control
6.
LCD Contrast Control
7.
8.
Display Elements Control (Sub-menus
will display when selected for Sonde or line tracing modes.)
9.
Frequency Selection Control (Sub-
menus will display for categories of frequencies that can be selected.)
Information Menu including software version
10.
and unit serial number (sub-menu for restoring factory defaults will display on Information screen).
See the Menu Tree on page 24 for a complete list.
Display Elements Menu
Advanced features of the SR-20 can be enabled by using the Menu Key to show the menu tree.
The SR-20 is shipped with some of the elements switched o for simplicity. Use the Select Key to check or uncheck the box next to a display element.
“Race Track “ with Watermark and Pointer
Signal Focus Control
Guidance Arrows
No Signal Icon (Suppression)
Auto Menu Exit Countdown Timer
Distortion Line On/O
Current Strength/ Signal Angle
Proximity Threshold Control Tracing Line Distortion
Ties Audio to Signal Strength
Center Signal Strength Option
Signal Strength On/O
Figure 32: Screen Elements (Line Trace Modes)
Ridge Tool Company
21
seekTech sR-20
Optional Features
Optional Features in the Display Elements Menu include:
Race Track and Watermark
This provides an additional, visual way to track the maximum signal. If you are trying to trace a line by noticing its highest Signal Strength level, Watermark serves as a visual aid.
No-Signal Icon (Suppression)
Center Signal Strength Option
Selecting this option in the Menu Selection screen will force the number representing Signal Strength to be displayed in the center of the display area anytime when a Proximity Signal is not available.
Proximity Threshold Control
This helps to constrain the locating to a certain range from the instrument. If the Measured Depth of the target is greater than the user-selected threshold value, the Proximity Signal will read zero. If the Measured Depth is less than the threshold that has been set, the SR-20 will display a Proximity Signal value. (Line Trace Mode only.)
Signal Focus Control
The Signal Focus Control feature essentially acts something like a magnifying glass on the signal. It reduces the sample bandwidth of the signal that the receiver examines, and gives a display based on a more sensitive read of the incoming signals. The tradeo in using the Signal Focus Control setting is that the display, while more precise, will update more slowly.
he Signal Focus Control can be set at 4 Hz (wide), 2 Hz, 1 Hz,
T .5 Hz, and .25 Hz (narrow). The narrower the selected bandwidth used, the greater detection distance and precision the receiver will show, but with a lower update rate of data on the display.
Signal Focus Control
Figure 34: Signal Focus Control
When it is selected on, the Signal Focus Control is changed to narrower or wider settings using the Up (narrower) and Down (wider) Keys.
Proximity Threshold Control
Figure 33: Proximity reshold Control
When it is activated, the Proximity Threshold is controlled by a long press (greater than ½ second) on the Up Key to set a higher threshold, or by the Down Key to lower the threshold.
The settings on the Proximity Threshold control the depth thresholding of the Proximity Signal as follows.
(Lowest) Signal Strength mode. Moves Signal Strength to screen center, map display suppressed, allows negative depth to display. Audio signal reects Signal Strength.
m/3 m/10 m/30 m) Displays Proximity Threshold for
(1 detections where Measured Depth is Xm or less.
(Highest) Wide-open Proximity Mode. No threshold, no suppression, allows negative depth display.
The Proximity Threshold Control is particularly valuable if you need to eliminate signals from outside a well-dened distance for clarity.
Signal Focus Control is useful when you need to focus in on a particular signal with detail.
Sound Muting > 99’
This option enables the automatic muting of the sound when the Measured Depth is greater than the setting of the Proximity Threshold setting.
Tracing Line Response
The Tracing Line distortion response checkbox sets the sensitivity of the Target Line’s distortion display to low, medium, or high, or disables it altogether. The higher the setting, the more sensitive the “distortion cloud” around the Tracing Line becomes.
If the distortion response is disabled, the Tracing Line will become a single solid line.
Frequencies Selection Control
Additional available frequencies on the Master Frequency Menu can be added to the Main Menu list of available frequencies by going to the Frequency Selection Control
sub-menu Highlight the category of the desired frequency (Figure 35).
Press the Select Key
and selecting the desired mode.
.
22
Ridge Tool Company
Figure 35: Selecting a Frequency Category
Then use the Up and Down Keys to scroll through the available frequencies. Highlight the desired frequency to add it to the currently available list .
Checking a frequency (using the Select Key) will enable it to be included in the “Currently Available” list of frequencies on the Main Menu.
Selected frequencies in the Checked-Active set can be switched while the SR-20 is in use, by pressing the Frequency Key. The SR-20 will cycle down the list through the set of active frequencies from low to high, group by group, and repeat. Unchecking a frequency in the Main Menu will deactivate it, and it will then not appear when pressing the Frequency Key.
seekTech sR-20
Restore Factory Defaults
Pressing Select a second time will display the Restore Factory Defaults option.
Use the Up and Down Keys to highlight either the “check” symbol to restore factory defaults, or the “X” symbol to NOT restore them.
Pressing the Menu Key without changing either checkbox will exit the option and leave things as they were.
Information Screen and Restoring Defaults
Information Screen
The information screen appears at the bottom of the menus choices list. Pressing the Select Key displays information about your locator, including software version, serial number of the receiver, and its calibration date (Figure 36).
Figure 36: Information Screen
Ridge Tool Company
23
seekTech sR-20
Menu Tree
Activated Frequencies
Sonde Line Trace Power (Passive Trace) Radio
Units of Measure
eet/Meters
F
Backlight Options On/O
LCD Contrast I
Display Elements Select (
Trace Mode Sonde Mode
Watermark
Signal Focus Setting
No-Signal Indicator
Sound Signals
Center Signal Strength*
Signal Strength
Proximity Threshold*
/Auto
ncrease/Decrease
Check On/O)
SR-20 Maintenance
Transportation and Storage
Before transporting, make sure that the unit is o to preserve battery power.
When transporting, make sure that the unit is secure and does not bounce around or get bumped by loose equipment.
The SR-20 should be stored in a cool dry place.
NOTE: If storing the SR-20 for an extended period, remove the batteries completely. If shipping the SR-20, remove the batteries entirely from the unit.
Maintenance and Cleaning
1. Keep the SR-20 clean with a damp cloth and some mild detergent. Do not immerse in water.
2. When cleaning, do not use scraping tools or abrasives as they may permanently scratch the display. NEVER USE SOLVENTS to clean any part of the system. Substances like acetone and other harsh chemicals can cause cracking of the Case.
Signal Angle Indicator
Distortion Line*
Tracing Line Distortion Response*
Sound Mute > 99’
Guidance Arrows*
*=Line Trace Display Only
Frequency Select (Check On/O)
onde
S
16 Hz, 512 Hz, 640 Hz, 16 kHz, 33 kHz
Line Trace
128 Hz, 1 kHz, 8 kHz, 33 kHz
Power
50 Hzx1, 50 Hzx5, 50 Hzx9, 60 Hz 100 Hz, 120 Hz, <4 kHz
RF
Low (4-15 kHz)
High (>15 kHz)
Information Menu
Restore Default Settings (Check Yes/No)
x1
, 60 Hzx5, 60 Hzx9,
Figure 37: Menu Tree
Locating Faulty Components
For troubleshooting suggestions, please refer to the troubleshooting guide.
Service and Repair
IMPORTANT! Instrument should be taken to a RIDGID
Independent Authorized Service Center or returned to the factory. Remove batteries before shipping.
All repairs made by Ridge service facilities are warranted against defects in material and workmanship.
If you have any questions regarding the service or repair of this machine, contact your RIDGID distributor, local RIDGID oce or Ridge Tool Europe at info.europe@ridgid.com.
24
Ridge Tool Company
Icons and Symbols
DISPLAY ICONS DISPLAY ICONS (Con’t) MENU ICONS
Sonde Frequency
Active Trace Frequency
Signal Angle Indicator
Audio Level
Radio Frequency
Passive Line Trace Frequency
Battery Level
Measured Distance/Depth
Signal Strength
Factory Default Reset
Menu Check Box
Go Up One Level (press menu key)
No Sonde Present
No Power Present
No Trace Present
No RF Present
Display Elements
Frequency Selection Control
Line Direction Gradient
Tracing Line
Pass Banwidth
Distortion Line
Equator
Pipe Direction
KEYPAD ICONS
Menu Navigation/Signal Focus/Proximity Threshold Control
Menu Navigation/Signal Focus/Proximity Threshold Control
Menu Selection SondeMode: Force Depth/Re-center Audio Line Trace Mode: Force Depth, Force current, Re-center Audio Signal Strength Proximity Setting: Force Map On
Power ON / OFF Key
Menu Key
Frequency Key
Sound Key
Proximity Threshold Control
Milliamp, Current
Menu Timeout Counter
Pole Icon
Proximity Signal
Tools Menu
Backlight settings
Screen Contrast Adjust
Depth Greater Than 3 Feet/1 Meter Threshold
Depth Greater Than 10 Feet/3 Meter Threshold
Depth Greater Than 30 Feet/10 Meter Threshold
Depth Greater Than 99 Feet/30 Meter Threshold
Low Battery Warning (flashing)
Information Screen
Level Pointer (Signal Strength)
Watermark (Signal Strength)
KEYPAD ICONS
Menu Navigation/Signal Focus/Proximity Threshold Control
Menu Selection Sonde Mode: Force Depth/Recenter Audio Line Trace Mode: Force Depth, Force current, Recenter Audio Signal Strength Proximity Setting: Force Map On
Menu Navigation/Signal Focus/Proximity Threshold Control
DISPLAY ICONS
Sonde Frequency
DISPLAY ICONS (Con’t)
Proximity Signal
Power ON / OFF Key
Menu Key
Frequency Key
Sound Key
seekTech sR-20
MENU ICONS
Factory Default Reset
Active Trace Frequency
Radio Frequency
Passive Line Trace Frequency
Measured Distance/Depth
Signal Angle Indicator
Milliamp, Current
Proximity Threshold Control
Pole Icon
Tracing Line
Distortion Line
Equator
Signal Strength
Audio Level
Battery Level
Low Battery Warning (ashing)
Level Pointer (Signal Strength)
Watermark (Signal Strength)
No Sonde Present
No Power Present
No Trace Present
No RF Present
Pass Banwidth
Menu Check Box
Tools Menu
Backlight Settings
Screen Contrast Adjust
Display Elements
Frequency SelectionControl
Information Screen
Menu Timeout Counter
Go Up One Level (press menu key)
Pipe Direction
Depth Greater Than 3 Feet /1 Meter Threshold
Depth Greater Than 10 Feet /3 Meter Threshold
Depth Greater Than 30 Feet /10 Meter Threshold
Depth Greater Than 99 Feet /30 Meter Threshold
Line Direction Gradient
Figure 38: Icons and Symbols
Ridge Tool Company
25
seekTech sR-20
Trouble Shooting Guide
PROBLEM PROBABLE FAULT LOCATION
SR-20 locks up during use.
SR-20 will not pick up the signal.
While tracing, lines are “jumping” all over the screen in the mapping display.
While locating a Sonde, lines are “jumping” all over the screen.
Distance between Sonde and either Pole is not equal.
Power the unit o, and then back on. Remove the batteries if the unit will not switch o. If batteries are low, replace them.
Check that the correct mode and frequency is set. Examine circuit for possible improvements. Relocate transmitter, change grounding, frequency, etc.; modify Proximity Threshold (page 22) and/or Signal Focus Control settings (page 22).
This indicates that the SR-20 is not picking up the signal or there is interference.
Make sure that the transmitter is well connected and grounded. Point the SR-20 at either lead to be sure that there is a complete circuit.
Try a higher frequency, or connecting to a dierent point in the line, or switching to inductive mode.
Try to determine the source of any noise and eliminate it. (Bonded grounding, etc.)
Check SR-20 batteries are fresh and fully charged.
Check the batteries in the Sonde to see that they are working.
Sonde may be too far away; try starting with it closer in if possible, or do an area search.
Verify signal by placing lower antenna close to Sonde. Note – Sondes have diculty emitting signals through cast iron and ductile iron lines.
Increase Proximity Threshold and try lower settings of Signal Focus Control to improve “focus” on weaker signals.
Sonde may be tilted or there may be a cast-iron-to-plastic transition.
Unit acts erratic, won’t power down.
Display appears completely dark, or completely light when it is turned on.
There is no sound. Adjust the sound level in the sound menu. Verify Proximity Signal is greater than zero.
SR-20 will not power on. Check orientation of batteries.
26
Batteries may be low. Replace with fresh batteries and power on.
Power the unit o and then back on.
Adjust the LCD screen contrast.
Check that the batteries are charged.
Check to see that the battery contacts are OK.
Unit may have blown a fuse. (Factory service is required.)
Ridge Tool Company
seekTech sR-20
Specications
• Weight w/ batteries .......... 4 lbs. (1.8 kg)
• Weight w/o batteries ....... 3.3 lbs. (1.5 kg)
Dimensions
Length .................................. 11.2” (28.4 cm)
Width ..................................... 4.3” (1.3 m)
Height .................................... 31.1” (79 cm)
Power Source
4 C-size batteries, 1.5V Alkaline (ANSI/NEDA 14A, IEC
• LR14) or 1.2V NiMH or NiCad rechargeable batteries
Power Rating: 6V, 550mA
• Signal Strength
Non-linear in function. 2000 is 10x higher than 1000,
3000 is 10x higher then 2000, etc.
Operating Environment
Temperature ........................ -4°F to 122°F (-20°C to 50°C)
Humidity ............................... 5% to 95% RH
Storage Temperature ....... -4°F to 140°F (-20°C to 60°C)
Default Settings
• Depth units = Meter & Centimeter
• Volume = 2 (two settings above mute)
• Backlight = Auto
• Proximity Threshold = 30 feet (10m)(Trace)
• 33 kHz (Active Line Trace Mode)
Standard Equipment
Item Cat. #
• SR-20 Locator 21943
Markers and Mast Holder 12543
Operator’s Manual
• 4 C-cell batteries (Alkaline)
• Training Video (DVD)
Optional Equipment
• Additional Sonde Markers 12543
• ST-305 Transmitter 21948
ST-510 Transmitter 21953
Inductive Clamp (4.75”) 20973
Remote Sonde 16728
Float Sonde (2pcs) 19793
Ridge Tool Company
27
seekTech sR-20
DE
SeekTech SR-20
Bedienungsanleitung
Allgemeine
GEFAHR
Das SR-20 ist ein Diagnosegerät, das elektromagnetische
Felder ermittelt, die von unterirdischen Objekten ausgehen. Es soll dem Benutzer helfen, diese Objekte aufzunden, indem es Merkmale der Feldlinien erkennt und auf dem Bildschirm darstellt. Da elektromagnetische Feldlinien abgelenkt und gestört werden können, ist es wichtig, die Lage unterirdischer Objekte zu verizieren, bevor man mit Grabungen beginnt.
Sicherheitsinformationen
WARNUNG! Lesen Sie diese Anweisungen und die begleitende Sicherheitsbroschüre sorgfältig,
bevor Sie dieses Gerät benutzen. Bei Unklarheiten wenden Sie sich bitte an Ihre RIDGID Vertriebsstelle, die Sie näher informiert.
Unkenntnis und Nichtbefolgung der Anweisungen können zu elektrischen Schlägen, Feuer und/oder schweren Verletzungen führen.
BEWAHREN SIE DIESE ANWEISUNGEN AUF!
VORSICHT: Vor dem Transport die Batterien vollständig
entfernen.
Sollten Sie Fragen bezüglich Wartung oder Reparatur dieses Gerätes haben, wenden Sie sich an Ihren Händler, Ihre zuständige RIDGID Niederlassung oder an Ridge Tool Europe unter info.europe@ridgid.com
Mehrere Versorgungsleitungen können sich unter
der Erde im selben Bereich benden. Befolgen Sie unbedingt die geltenden Richtlinien und Benach­richtigungsverfahren.
Das Freilegen der Versorgungsleitung ist die
einzige Möglichkeit, ihre Existenz, Lage und Tiefe zu verizieren.
Ridge Tool Co., ihr angegliederte Unternehmen
und Lieferanten haften nicht für Verletzungen oder direkte, indirekte, Neben- oder Folgeschäden, zu denen es aufgrund der Benutzung des SR-20 kommt.
Bitte geben Sie bei jeder Korrespondenz alle auf dem Typenschild Ihres Gerätes angegebenen Informationen an, einschließlich Modell- und Seriennummer.
28
Ridge Tool Company
SR-20 Komponenten
Serieller Anschluss
USB-Anschluss
Typenschild
Symbolerläuterung
HINWEIS: USB-/ Serielle Schnittstellen dienen zum Laden neuer Software
seekTech sR-20
Batteriefach
Display
Bedienfeld
Handgri
Lautsprecher
Antennenmast
Oberer Antennen-
knoten
Klappgelenk
Führungs-
antennen
Unterer Antennen-
knoten
Einrastvorrichtung für Klappmast
Abbildung 1: SR-20 Komponenten
Ridge Tool Company
29
Loading...
+ 271 hidden pages