Rice Lake IQ Plus 210 User Manual

Page 1
®
IQ plus 210
Digital Weight Indicator
Version 1.02
Installation Manual
64640
Page 2
Page 3
Contents
About This Manual................................................................................................................................... 1
1.0 Introduction.................................................................................................................................. 1
1.1 Operating Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 LED Annunciators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.4 Indicator Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.0 Setup and Service Information.................................................................................................... 3
2.1 Enclosure Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2.1 Cable Grounding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2.2 Load Cells. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2.3 Setting the Load Cell Compensation Jumper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2.4 Serial Communications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.5 Digital Inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Enclosure Reassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.4 Board Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.5 IQ plus 210 Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.0 Configuration.............................................................................................................................. 10
3.1 Configuration Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1.1 Revolution Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1.2 EDP Command Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1.3 Front Panel Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Menu Structures and Parameter Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.1 Configuration Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.2 Format Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2.3 Calibration Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2.4 Serial Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2.5 Digital Input Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2.6 Default Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2.7 Version Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.0 Calibration.................................................................................................................................. 20
4.1 Front Panel Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.2 EDP Command Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.3 Revolution
5.0 EDP Commands.......................................................................................................................... 22
5.1 The EDP Command Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1.1 Key Press Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1.2 Reporting Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1.3 The RESETCONFIGURATION Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1.4 Parameter Setting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1.5 Normal Mode Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2 Saving and Transferring Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2.1 Saving Indicator Data to a Personal Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2.2 Downloading Configuration Data from PC to Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Technical training seminars are available through Rice Lake Weighing Systems.
Course descriptions and dates can be viewed at www.rlws.com or obtained by
calling 715-234-9171 and asking for the training department.
Copyright © 2003 Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America.
Specifications subject to change without notice.
Version 1.0, June 2003
Page 4
ii
6.0 Appendix.................................................................................................................................... 25
6.1 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.1.1 Displayed Error Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.1.2 Using the XE EDP Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.2 Status Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.2.1 Using the P EDP Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.2.2 Using the ZZ EDP Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.3 Continuous Output (Stream) Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.4 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.4.1 IQ plus 210 Indicator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.4.2 RL2100 Bench Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
IQ plus 2100 Limited Warranty.............................................................................................................. 28
IQ plus 210 Installation Manual
Page 5
About This Manual
This manual is intended for use by service technicians responsible for installing and servicing IQ plus
digital bench scales. Configuration and calibration of the indicator can be
accomplished using the indicator front panel keys, the EDP command set, or Version 2.3 or later of the
Revolution
configuration utility. See Section 3.1 on
page 10 for information about configuration methods.
®
2100
1.0 Introduction
Warning
Some procedures described in this manual require work inside the indicator enclosure. These procedures are to be performed by qualified service personnel only
.
Authorized distributors and their employees can view or download this manual from the Rice Lake Weighing Systems distributor site
www.rlws.com
at
.
The IQ plus 2100 digital bench scale consists of an IQ plus 210 digital weight indicator and an RL2100 bench scale. The indicator can be mounted to the bench scale using the 12-inch column or the attachment bracket, or can be separately mounted on the tilt stand.
The IQ plus 210 is a single-channel digital weight indicator housed in a NEMA 4X/IP66-rated stainless steel enclosure. The indicator front panel consists of a large (.8 in, 20 mm), six-digit, seven-segment LED display, seven LED annunciators, and two piezo switches used to zero the scale and select the displayed units. Features include:
Drives up to four 350
Supports 4- and 6-wire load cell connections
•Two configurable digital inputs
Single serial port supports full duplex, RS-232 communications at up to 9600 bps
•Available in 115 VAC and 230 VAC versions
load cells
The RL2100 bench scale is available in the following configurations:
5, 10, and 25 lb (2.5, 5, and 10 Kg) capacities with a 10" x 10" (25cm x 25cm) platform
50 and 100 lb (25 and 50 Kg) capacities with a 12" x 12" (30cm x 30cm) platform
The RL1042 anodized aluminum load cell is standard in the RL2100 bench scale; the RL1380 stainless steel load cell is available as an option. A stainless steel clamshell enclosure is also available for load cell protection.
1.1 Operating Modes
The IQ plus 210 has two modes of operation:
Normal (weighing) mode
Normal mode is the “production” mode of the indicator. The indicator displays the gross weight, and the LED annunciators described on page 2 indicate scale status and the type of weight value displayed. Once configuration is complete and a legal seal is affixed to the back of the indicator, this is the only mode in which the IQ plus 210 can operate.
Setup mode
Most of the procedures described in this manual require the indicator to be in setup mode, including configuration and calibration. To enter setup mode, remove the large fillister head screw from the enclosure backplate. Insert a screwdriver or a similar tool into the access hole and press the setup switch once. The indicator display changes to show the word
CONFIG
.
Introduction
1
Page 6
2
1.2 Front Panel
Figure 1-1 shows the IQ plus 210 LED annunciators, buttons, and normal mode ke y functions.
DIGITAL WEIGHT INDICATOR
Neg
Zero
Motion
ZERO
ZERO
Set scale weight to zero
Figure 1-1. IQ plus 210 Front Panel, Showing LED Annunciators and Normal Mode Key Functions
Change displayed units
UNITS
UNITS
kg g lb oz
1.3 LED Annunciators
The IQ plus 210 display uses a set of seven LED annunciators to provide additional information about the value being displayed:
Neg annunciator lights to show that the displayed
• value is negative.
Zero : Weight is within 0.25 graduations of zero. This annunciator lights when the scale is zeroed.
Motion : Scale in motion outside the specified motion band. Zeroing and print operations can only be done when the motion LED is off.
lb, kg, oz, and g annunciators indicate the units
• associated with the displayed value: lb=pounds, kg=kilograms, oz=ounces, g=grams. lb and oz annunciators are both lit when the displayed weight is in pounds and ounces.
1.4 Indicator Operations
Basic IQ plus 210 operations are summarized below.
Zero Scale
1. Remove material from the scale and wait for
Motion
the
2. Press the lights to indicate the scale is zeroed.
Change Display Units
Press the
UNITS
between primary and alternate units. The LED annunciators on the right side of the display show the current displayed units. If more than one alternate unit is configured, the display cycles through the alternate units before returning to primary units; if no alternate units are configured, the display does not change when the button is pressed.
LED to go out.
ZERO
button. The
annunciator
Zero
button to change the displayed units
IQ plus 2100 Installation Manual
Page 7
Cord grip
Insulated cable
Foil (silver side out)
Grounding clamp
Shield wire (cut)
Length of foil before folding back on cable insulation
Cut insulation here for foil-shielded cables
Braid
Cut insulation here for braided cables
NOTE: Install lockwashers first, against enclosure, under grounding clamp
2.0 Setup and Service Information
This section describes setup and service procedures for the IQ plus 2100 digital bench scale, including installation and maintenance information, replacement parts lists, and assembly drawings. See Sections 2.1 through 2.5 for IQ plus 210 indicator information; see Sections 6.6 and 6.7 for bench scale information.
2.1 Enclosure Disassembly
The indicator enclosure must be opened to connect cables after replacing the load cell, or when connecting serial communications cables or digital inputs to the scale.
The IQ plus 210 has no on/off switch.
Warning
Caution
• Use a wrist strap to ground yourself and protect components from electrostatic discharge (ESD) when working inside the indicator enclosure.
• This unit uses double pole/neutral fusing which could create an electric shock hazard. Procedures requiring work inside the indicator must be performed by qualified service personnel only.
• The supply cord serves as the power disconnect for the IQ plus 2100. The power outlet supplying the indicator must be installed near the unit and be easily accessible
Ensure power to the indicator is disconnected, then remove the screws that hold the backplate to the enclosure body, then lift the backplate away from the enclosure and set it aside.
2.2 Cable Connections
The IQ plus 210 provides three cord grips for cabling into the indicator: two for the power cord and load cell cabling, the third for communications and digital input cables. The free cord grip comes with a plug installed to prevent moisture from entering the enclosure. If your application requires serial communication or digital input cabling, remove the plug and install cables as described in Sections 2.2.4 and 2.2.5 on page 5.
Before opening the unit, ensure the power cord is disconnected from the power outlet.
•Route cables through cord grips and grounding clamps to determine cable lengths required to reach cable connectors. Mark cables to remove insulation and shield as described below:
•For cables with foil shielding, strip insulation and foil from the cable half an inch (15 mm) past the grounding clamp (see Figure 2-1). Fold the foil shield back on the cable where the cable passes through the clamp. Ensure silver (conductive) side of foil is turned outward for contact with the grounding clamp.
•For cables with braided shielding, strip cable insulation and braided shield from a point just past the grounding clamp. Strip another half inch (15 mm) of insulation only to expose the braid where the cable passes through the clamp (see Figure 2-1).
IMPOR
• yellow shield wire 3/4" past the grounding clamp. Fold wire back and secure between the cable and clamp. Shield wire function is provided by contact between the cable shield and the grounding clamp.
•Route stripped cables through cord grips and clamps. Ensure shields contact grounding clamps as shown in Figure 2-1. Tighten grounding clamp nuts.
Finish installation using cable mounts and ties to secure cables inside of indicator enclosure.
TANT!
For load cell cables, strip the
2.2.1 Cable Grounding
Except for the power cord, all cables routed through the cord grips should be grounded against the indicator enclosure. Do the following to ground shielded cables:
•Use the lockwashers, clamps, and kep nuts provided in the parts kit to install grounding clamps on the studs adjacent to the cord grips. Install grounding clamps only for cord grips that will be used; do not tighten nuts.
Figur e 2-1. Grounding Clamp Attachment for Foil-Shielded
and Braided Cabling
Setup and Service Information
3
Page 8
To setup switch
Microprocessor
J1
GND
RESET
A/D Converter
GND
11
J3
J2
DIGIN 2 DIGIN 1
UNITS
Button
J4
DIGITAL
+5V
TEST
VR1 VR2
JP2 JP1
LOAD CELL CONNECTOR
1 23456
–SIG
+SIG
+SENSE
ANALOG
+5V
TEST
+EXC
–SENSE
J5
–EXC
4
GND
JP1
ON OFF
Load cell compensation jumper
VR3
TxD
1
GND
2
RxD
3
J7
TP1
+
BR1
TP2
Transformer
SERIAL COMM
J8
ZERO
Button
J6
Blue wire
Power
Input
F1 F2
Brown wire
Figure 2-2. IQ plus 210 CPU and Power Supply Board
2.2.2 Load Cells
Wire the load cell cable to connector J5 as shown in Table 2-1. If using 6-wire load cell cable (with sense wires), remove jumpers JP1 and JP2 (see Figure 2-2). For 4-wire installation, leave jumpers JP1 and JP2 on.
NOTE: All models come with ten feet of color-coded
load cell cable. Do not cut this cable! The load cell is temperature-compensated for an exact cable length of ten feet. Cutting the load cell cable voids the load cell warranty.
When connections are complete, use two cable ties to secure the load cell cable to the inside of the enclosure.
Load Cell Wire Color
J5 Pin
Function
1 +SIG Red White 2 –SIG White Red 3 +SENSE Blue — 4 –SENSE Brown — 5 +EXC Green Green 6 –EXC Black Black
NOTES:
1.Use grounding procedure described in Section 2.2.1 on page 3 to attach shield wire to backplate.
2.For 6-wire connections, remove jumpers JP1 and JP2.
Table 2-1. J5 Pin Assignments
RL1040 RL1042
RL1380
2.2.3 Setting the Load Cell Compensation Jumper
The load cell compensation jumper (above the
ZERO
button location on the CPU board; see Figure 2-2) must be set for the type of load cell connected to the indicator:
•For load cells with balanced bridges, set the jumper in the OFF position. Balanced load cells include the RL1040, RL1250, RL1260, RL1380, and RL1385.
•For load cells with unbalanced bridges, follow the procedure below to determine the correct jumper position. Examples of unbalanced cells include the RL1042 and RL1010.
To determine the correct jumper position for unbalanced cells, do the following;
1. Disconnect load cell from indicator.
2. Use an ohmmeter to measure +EXC to +SIG and +EXC to –SIG. Measured values between the excitation line and each of the signal lines should
be within 2–3
.
3. Next, measure –EXC to +SIG and –EXC to –SIG. Measured values between the excitation line and each of the signal lines should be within
.
2–3
4. If the +EXC measurements (step 2) are
5% larger than the –EXC measurements (step 3), set the compensation jumper in the ON position. If the +EXC measurements are < 5% greater (or are less) than the –EXC measurements, set the jumper in the OFF position.
IQ plus 2100 Installation Manual
Page 9
2.2.4 Serial Communications
To attach serial communications cables, connect communications cables to connector J7 as shown in Table 2-2. Use cable ties to secure serial cable to the inside of the enclosure.
Use the SERIAL menu to configure serial communications. See Section 3.2.4 on page 18 for configuration information.
J7 Pin Label Function
1 TxD RS-232 TxD 2 GND RS-232 Ground 3 RxD RS-232 RxD
Table 2-2. J7 Pin Assignments
2.2.5 Digital Inputs
Digital inputs (connectors J2 and J3) can be used to perform remote ZERO and UNITS key presses or to send serial data to a printer (remote PRINT key function). The inputs are active (on) with lo w v oltage (0 VDC) and can be driven by TTL or 5V logic without additional hardware. Use the DIG IN menu to configure the digital inputs. See Section 3.2.5 on page 19 for information about configuring the digital inputs.
2.4 Board Removal
If you must remove the IQ plus 210 CPU board, use the following procedure:
1. Disconnect power to the indicator. Remove backplate as described in Section 2.1 on page 3.
2. Remove connections to J5 (load cell cable), J7 (serial communications), J2 and J3 (digital inputs), J4 and J6 (piezo button inputs), and J1 (setup switch). Remove blue and brown power input wires at J8. See Figure 2-2 on page 4 for connector locations.
3. Remove the five nuts from the CPU board, then lift the board out of the enclosure.
To replace the CPU board, reverse the above procedure. Be sure to reinstall cable ties to secure all cables inside the indicator enclosure.
2.3 Enclosure Reassembly
Once cabling is complete, position the backplate over the enclosure and reinstall the backplate screws. Use the torque pattern shown in Figure 2-3 to prevent distorting the backplate gasket. Torque screws to 15 in-lb (1.7 N-m).
Torque pattern
1
To rque backplate screws to 15 in-lb (1.7 N-m)
5
Fillister head screws
3
Figure 2-3. IQ plus 210 Enclosure Backplate
10
Setup switch access screw
7
8
9
4
6
2
Setup and Service Information
5
Page 10
6
2.5 IQ plus 210 Replacement Parts
Table 2-3 lists replacement parts for the IQ plus 210, including all parts referenced in Figures 2-4 through 2-8.
Ref
Number
1 14626 Kep nuts, 8-32NC hex (3) Figure 2-4 on page 7, Figure 2-7 on page 8 2 53651 Display and CPU board assembly, 115 VAC (1) Figure 2-5 on page 8
3 15365 Board mounting spacers (5) 4 52483 Enclosure backplate (1) Figure 2-4 on page 7 5 15626 Cable grips, PG9 (2) Figure 2-8 on page 9 6 30375 Nylon seal rings for cable grips (3) 7 15627 Locknuts, PCN9 (3) 8 19538 Cable grip plugs (2)
9 45042 Sealing washers (10) Figure 2-4 on page 7 10 44676 Sealing washer for setup switch access screw (1) 11 42640 Setup switch access screw, 1/4 x 28NF x 1/4 (1) 12 41965 Power cord assembly, 115VAC (1) Figure 2-8 on page 9
13 41964 Line filter assembly (1) Figure 2-6 on page 8 14 14621 Kep nuts, 6-32NC hex (5) Figure 2-5 on page 8 20 15134 Lock washers, internal tooth, No. 8 , Type A (4) Figure 2-4 on page 7, Figure 2-7 on page 8 22 52853 Overlay membrane panel (1) Figure 2-8 on page 9 23 52482 Enclosure (1) 24 14862 Screws, 8-32NC x 3/8 (8) Figure 2-4 on page 7 26 45043 Ground wire, 4-in., No. 8 (1) 27 39037 Backplate gasket (1) 28 49910 Setup switch assembly (1) Figure 2-6 on page 8 29 16892 Ground/Earth Label (1) 30 15650 Cable tie mounts (1) 31 45302 Line filter standoffs (2) 33 15131 Lock washers, external tooth, No. 6, Type A (10) Figure 2-5 on page 8 34 61113 Piezo switches (2) Figure 2-8 on page 9 35 30623 Fillister head screws, 8-32NC x 7/16 (2) Figure 2-4 on page 7
53848 200 mA TR5 subminiature fuses (2), 115 VAC F1 and F2 in Figure 2-2 on page 4
Caution
PN Description (Quantity) Figure
53650 Display and CPU board assembly, 230 VAC (1)
45254 Power cord assembly, 230VAC (1)
53881 100 mA TR5 subminiature fuses (2), 230 VAC
(See Caution below)
For protection against risk of fire, replace fuses only with same type and rating fuse.
See Section 6.4 on page 27 for complete fuse specifications.
IQ plus 2100 Installation Manual
Table 2-3. Replacement Parts
Page 11
4
27
26
B
9/10X
24/8X
A
11
10
See Line Filter
and
Ground Post
Assembly
35/2X
Figure 2-4. IQ plus 210 Backplate Assembly
Setup and Service Information
7
Page 12
8
To setup switch
14/5X
2
See J8 detail below
3/5X
33/10X
31/2X
13
Figure 2-5. IQ plus 210 Enclosure and CPU Board
30
AC input to
CPU board J8
29
Figure 2-6. Line Filter Assembly
Blue wire
28
Setup switch connection to J1 on CPU board
Cable tie
Brown wire
AC power in from line filter
Line Filter Connection to J8
To backplate
ground post
1/3X/A
20/4X/B
Figure 2-7. Ground Post Assembly
To line filter
To power cord
IQ plus 2100 Installation Manual
Page 13
12
22
23
6/2X
8
5/2X
7/2X
34/2X
To J6 on CPU board
To J4 on CPU board
Figure 2-8. IQ plus 210 Enclosure and Overlay
Setup and Service Information
9
Page 14
3.0 Configuration
To configure the IQ plus 210 indicator, the indicator must be placed in setup mode. The setup switch is accessed by removing the large fillister head screw on the enclosure backplate. Insert a screwdriver into the access hole and press the switch to enter setup mode.
When the indicator is placed in setup mode, the word
CONFIG is shown on the display. The CONFIG menu is
the first of the main menus used to configure the indicator. Detailed descriptions of these menus are given in Section 3.2. When configuration is complete, press the setup switch again to exit setup mode, then replace the setup switch access screw.
3.1 Configuration Methods
The IQ plus 210 indicator can be configured by using the front panel buttons to navigate through a series of configuration menus or by sending commands or configuration data to the EDP port. Configuration using the menus is described in Section 3.1.3.
Configuration using the EDP port can be accomplished using either the EDP command set described in
Section 5.0 or Version 2.3 or later of the Revolution configuration utility .
3.1.1 Revolution Configuration
The Revolution configuration utility provides the preferred method for configuring the IQ plus 210 indicator. Re v olution runs on a personal computer to set configuration parameters for the indicator. When Revolution configuration is complete, configuration data is downloaded to the indicator.
To use Rev olution, do the following:
1. Install Revolution on an IBM-compatible per­sonal computer running Windows
Windows 95. Minimum system requirements are 8MB of extended memory and at least 5MB of available hard disk space.
2. With both indicator and PC powered off, connect the PC serial port to the RS-232 pins on the indicator EDP port.
3. Power up the PC and the indicator. Use the setup switch to place the indicator in setup mode.
4. Start the Revolution program.
Figure 3-1 shows an example of one of the Revolution configuration displays.
Revolution provides online help for each of its configuration displays. Parameter descriptions provided in this manual for front panel configuration can also be used when configuring the indicator using Revolution: the interface is different, but the parameters set are the same.
3.1.2 EDP Command Configuration
The EDP command set can be used to configure the IQ plus 210 indicator using a personal computer, terminal, or remote keyboard. Like Revolution, EDP command configuration sends commands to the indicator serial port; unlike Revolution, EDP commands can be sent using any external device capable of sending ASCII characters over a serial connection.
EDP commands duplicate the functions available using the indicator front panel and provide some functions not otherwise available. EDP commands can be used to simulate pressing front panel buttons, to configure the indicator, or to dump lists of parameter settings. See Section 5.0 on page 22 for more information about using the EDP command set.
®
3.11 or
Figure 3-1. Sample Revolution Display
Revolution supports both uploading and downloading of indicator configuration data. This capability allows configuration data to be retrieved from one indicator, edited, then downloaded to another.
10 IQ plus 2100 Installation Manual
Page 15
3.1.3 Front Panel Configuration
DIGITAL WEIGHT INDICATOR
ZERO/ENTER
Move down in menu or
enter parameter value and
return to menu level above
UNITS/RIGHT
Move right in menu /
next parameter value
kg g lb oz
ZERO
UNITS
Zero
Motion
Neg
The IQ plus 210 indicator can be configured using a series of menus accessed through the indicator front panel when the indicator is in setup mode. Table 3-1 summarizes the functions of each of the main menus.
Menu Menu Function
CONFIG Configuration Configure grads, zero tracking, zero range, motion band, overload, and digital filtering
parameters. FORMAT Format Set format of primary units, select alternate display units. CALIBR Calibration Calibrate indicator. See Section 4.0 on page 20 for calibration procedures. SERIAL Serial Configure serial port. DIG IN Digital Input Assign digital input functions. DEFLT Default Restore default configuration. VERS Version Display installed software version number.
Table 3-1. IQ plus 210 Menu Summary
Figure 3-2. Front Panel Key Functions in Setup Mode
Configuration 11
Page 16
Front Panel Menu Navigation
The front panel buttons are used to navigate through the menus in setup mode (see Figure 3-2). The
UNITS
(RIGHT) button scrolls right (horizontally) on the same menu level; the
ZERO (ENTER) button moves down
(vertically) to different menu levels and serves as an Enter key for selecting parameter values within the menus.
Most of the configuration menus have three or four levels. Figure 3-3 shows the general structure of the IQ plus 210 configuration menus. Note the following:
On the first (main) menu level, press RIGHT to scroll through the menus. Pressing RIGHT from the VERS menu wraps around to the CONFIG menu. Press ENTER from any of the main menus to move down to the first parameter for that menu.
•On the middle (parameter) menu levels, press
RIGHT to scroll through the parameter prompts
for the menu. When the last parameter is displayed, pressing RIGHT returns to the level above. Press ENTER from any parameter prompt
to move down to the values prompts for that parameter.
•On the bottom (parameter value) menu level,
RIGHT to scroll through the values for the
press parameter. Pressing RIGHT from the last value wraps around to first value. Press ENTER to select the value and return to the parameter prompt.
Figure 3-4 on page 13 shows an example of the navigation used to select a parameter value under the SERIAL menu.
To edit numerical values, use the select the digit; use the
ENTER button to increment or
RIGHT button to
decrement the value of the flashing digit (see Figure 3-5 on page 13). Once all digits have been edited, press
ENTER to save the value and return to the parameter
prompt.
Menu Navigation
General Structure
RIGHT RIGHT RIGHT
ENTER
Pressing
RIGHT RIGHT
Parameter 1
ENTER
RIGHT
Value 1
RIGHT
Parameter 2 Parameter 3 Parameter 4
RIGHT RIGHT
Value 2 Value 3 Value 4
RIGHT RIGHT RIGHT
SERIALCALIBRCONFIG FORMAT
at the last parameter prompt returns to the menu prompt
RIGHT
XXXXXXXXXXXXXX
RIGHT
RIGHT
VERSDEFLTDIG IN
ZERO
(ENTER)
UNITS
(RIGHT)
RIGHT
Press
ENTER
12 IQ plus 2100 Installation Manual
at any parameter value to select the value and return to the parameter prompt
Figure 3-3. General Menu Structure
Page 17
PRNFRQ
DEMAND
AUTO1
BITS TERMIN
BAUD
CR
9600
7ODD
7EVEN
8NONE
CR/LF
48002400
1200
AUTO2
STREAM
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIG FORMAT
RIGHT
RIGHT RIGHT RIGHT
ENTER
Menu Navigation
Serial Menu Example
RIGHT
RIGHT RIGHT
RIGHT
ENTER
RIGHT RIGHT
RIGHT
With any of the BAUD parameter values displayed, press
ENTER
to select the value and return to the BAUD parameter prompt
RIGHT
Navigation functions same as BAUD parameter
When displaying last parameter prompt, press
RIGHT
to return to the
first-level menu prompt
(ENTER)
ZERO
(RIGHT)
UNITS
0 0 0 0 0 0
To edit numeric values, press the button to select the leftmost digit. Each digit flashes when selected: Press increment the value of the selected digit; press the
RIGHT
Once the last digit has been edited, press again and the digit stops flashing. Press to save the value entered and return to the menu level above, or press starting with the leftmost digit.
Figure 3-5. Editing Procedure for Numeric Values
button to move right to the next digit.
RIGHT
Figure 3-4. Setup Mode Menu Navigation
UNITS (RIGHT)
ZERO (ENTER)
to edit the value again,
RIGHT
ENTER
to
Configuration 13
Page 18
3.2 Menu Structures and Parameter Descriptions
The following sections provide graphic representations of the IQ plus 210 menu structures. In the actual menu structure, the values under each parameter are arranged horizontally . To save page space, menu choices are shown in vertical columns. The factory def ault setting appears at the top of each column.
Most menu diagrams are accompanied by a table that describes all parameters and parameter values associated with that menu. Default parameter values are shown in bold type.
3.2.1 Configuration Menu
GRADS
5000
number
SERIALCALIBRCONFIG FORMAT
ZTRKBN ZRANGE
OFF
0.5D 1D
3D
100%
1.9%
XXXXXXXXXXXXXX
MOTBAN
1D
2D 3D
5D
10D
20D
OFF
VERSDEFLTDIG IN
OVRLOA
FS+2%
FS+1D FS+9D
Figure 3-6. Configuration Menu
CONFIG Menu
Parameter Choices Description
Level 2 submenus
GRADS 5000
number
Graduations. Specifies the number of full scale graduations. The value entered must be in the range 1–10 000 and should be consistent with legal requirements and environmental limits on system resolution.
FS
FILTER
4
8
16 4RT 8RT
16RT
1 2
To calculate GRADS, use the formula, GRADS = Capacity / Display Divisions. Display divisions for primary and secondary units are specified on the FORMAT menu.
ZTRKBN OFF
0.5D 1D
Zero track band. Automatically zeroes the scale when within the range specified, as long as the input is within the configured zero range (ZRANGE parameter). Selections are ± display divisions. Maximum legal value varies depending on local regulations.
3D
ZRANGE 100%
1.9%
Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is ± 1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at standstill to zero the scale. Use 1.9% for legal-for-trade applications.
MOTBAN 1D
2D 3D 5D 10D 20D
Motion band. Sets the level, in display divisions, at which scale motion is detected. If motion is not detected for 1 second or more, the standstill symbol lights. Some operations, including print, tare, and zero, require the scale to be at standstill. Maximum legal value varies depending on local regulations.
If OFF is selected, ZTRKBN should also be set to OFF.
OFF
14 IQ plus 2100 Installation Manual
Table 3-2. Configuration Menu Parameters
Page 19
CONFIG Menu
KG GOZLB
PRIMAR ALTUNT
DSPDIVUNITS DECPNT
OZ
KG
88888.8
8888.88
1D
5D
2D
LB
88.8888
8.88888
888880
888888
OFF
ON
LB/OZ
G
LB/OZ
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIG FORMAT
OFF
ON
OFF
ON
OFF
ON
OFF
ON
888 88
88 88.8
If UNITS=LB/OZ,
DECPNT selections are:
888.888
Parameter Choices Description
OVRLOA FS+2%
FS+1D
Overload. Determines the point at which the display blanks and an out-of-range error
message is displayed. Maximum legal value varies depending on local regulations. FS+9D FS
FILTER 4
8 16 4RT 8RT 16RT 1
Digital filtering. Selects the digital filtering rate used to reduce the effects of mechanical
vibration from the immediate area of the scale.
Choices indicate the number of A/D conversions that are averaged to obtain the
displayed reading. A higher number gives a more accurate display by minimizing the
effect of a few noisy readings, but slows down the settling rate of the indicator. Values
with the RT suffix provide a raised threshold for filtering larger noise spikes. 2
Table 3-2. Configuration Menu Parameters (Continued)
3.2.2 Format Menu
Figure 3-7. Format Menu
Configuration 15
Page 20
FORMAT Menu
Parameter Choices Description
Level 2 submenus
PRIMAR UNITS
DECPNT DSPDIV
ALTUNT LB
KG OZ G LB/OZ
Level 3 submenus
Primary Units (PRIMAR Parameter)
DECPNT 888.888
8888.88
88888.8 888888 888880
8.88888
88.8888
or
88 88.8 888 88
DSPDIV 1D
2D 5D
UNITS LB
KG OZ G LB/OZ
Alternate Units (ALTUNT Parameter)
LB KG OZ G LB/OZ
ON
OFF
Specifies the decimal position, display divisions, and units used for the primary units. See Level 3 submenu parameter descriptions.
Specifies which alternate units can be displayed by pressing the UNITS. See Level 3 submenu parameter descriptions.
NOTE: The LB/OZ setting is not a Legal-for-Trade setting.
Decimal point location. Specifies the location of the decimal point or dummy zeroes in the primary unit display. Value should be consistent with local legal requirements.
If LB/OZ is selected as the primary unit, DECPNT selections are 88 88.8 and 888 88.
Display divisions. Selects the minimum division size for the primary units displayed weight.
Specifies primary units for displayed and printed weight. Values are: LB=pound; KG=kilogram; OZ=ounce; G=gram; LB/OZ=pounds and ounces.
Alternate units. Determines which units are displayed when the UNITS button is pressed. The value for the primary unit (selected on the PRIMAR parameter) is always ON.
NOTE: The LB/OZ setting is not a Legal-for-Trade setting.
16 IQ plus 2100 Installation Manual
Table 3-3. Format Menu Parameters
Page 21
3.2.3 Calibration Menu
WZERO
*CAL*
Display and edit
zero calibration
A/D count value
Display and edit
test weight value
WVAL
*CAL*
Display and edit
span calibration
A/D count value
WSPAN
*
CAL*
Press Enter to
remove offset from
zero and span
calibrations
REZERO
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIG FORMAT
See Section 4.0 on page 20 for calibration procedures.
Figure 3-8. Calibration Menu
CALIBR Menu
Parameter Choices Description
Level 2 submenus
WZERO Display and edit the zero calibration A/D count value. WVAL Display and edit the test weight value. WSPAN Display and edit the span calibration A/D count value. REZERO Press ENTER to remove an offset value from the zero and span calibrations.
Use this parameter only after WZERO and WSPAN have been set. See Section 4.1 on page 20 for more information about using this parameter.
Table 3-4. Calibration Menu Parameters
Configuration 17
Page 22
3.2.4 Serial Menu
9600
1200 2400 4800
BITSBAUD
8NONE
7EVEN
7ODD
SERIALCALIBRCONFIG FORMAT
TERMIN
CR/LF
CR
ECHO
OFF
ON
XXXXXXXXXXXXXX
PRNFRQ
DEMAND
AUTO1
AUTO2
STREAM
Figure 3-9. Serial Menu
SERIAL Menu
Parameter Choices Description
BAUD 9600
1200 2400 4800
BITS 8NONE
7EVEN 7ODD
TERMIN CR/LF
CR
ECHO OFF
ON
PRNFRQ DEMAND
AUTO1 AUTO2 STREAM
Baud rate. Selects the transmission speed for the serial port.
Selects number of data bits and parity of data transmitted from the serial port.
Termination character. Selects termination character for data sent from the serial port.
This command enables or disables echoing of the serial commands sent to the indicator.
Print frequency. Specifies when the indicator sends data to the serial port: DEMAND: When a PRINT digital input is activated or the KPRINT EDP command received.
AUTO1: Output enabled by scale in motion; transmitted when scale returns to standstill. AUTO2: Output enabled by scale at center of zero; transmitted when a positive scale weight
is at standstill.
STREAM: Continuous output. See Section 6.3 on page 26 for output format.
Weight data is sent to the serial port in displayed units except when other units are specified using one of the XGx EDP commands (see Section 5.1.5 on page 24).
VERSDEFLTDIG IN
If DEMAND, AUTO1, or AUTO2, is selected for this parameter, serial output is formatted as follows:
wwwwwww uu GROSS
where wwwwww is a 7-digit weight field (leading zeroes suppressed, including decimal point and minus sign, if required), uu is the units designator (lb, kg, oz, or G).
If pounds/ounces (LB/OZ) is specified as the primary units, the serial output is formatted as shown below:
xxx lb yyyy oz GROSS
where xxx is the three-digit pounds weight (including minus sign, if required) followed by lb and a space; and yyyy is the four-digit ounce weight (including decimal point) followed by oz.
18 IQ plus 2100 Installation Manual
Table 3-5. Serial Menu Parameters
Page 23
3.2.5 Digital Input Menu
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIG FORMAT
DIGIN1
OFF
DIGIN2
ZERO UNITS PRINT
OFF
ZERO UNITS PRINT
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIG FORMAT
RESET
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIG FORMAT
Software
version
Figure 3-10. Digital Input Menu
DIG IN Menu
Parameter Choices Description
Level 2 submenus
DIGIN1 DIGIN2
OFF
ZERO UNITS PRINT
Specifies the function activated by digital inputs 1 and 2.
Table 3-6. Digital Input Menu Parameters
3.2.6 Default Menu
The DEFLT menu is used to reset indicator configuration to the default values. There are no parameters associated with the DEFLT menu. From the DEFLT menu, press parameter. Press
ENTER again to reset the indicator to the default configuration values or press RIGHT (UNITS)
to cancel the indicator reset and return to the menu level above.
ENTER (ZERO) to move down to the RESET
NOTE: All load cell calibration settings are lost
when the RESET function is performed.
Figure 3-11. Default Menu
3.2.7 Version Menu
The VERS menu is used to check the software version installed in the indicator. There are no parameters associated with the VERS menu: when selected, the indicator displays the installed software version number.
Figure 3-12. Version Menu
Configuration 19
Page 24
4.0 Calibration
The IQ plus 210 can be calibrated using the front panel, EDP commands, or the Revolution™ configuration utility. Each method consists of the following steps:
Zero calibration
Entering the test weight value
Span calibration
Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods.
SERIALCALIBRCONFIG FORMAT
WZERO
*CAL*
Display and edit
zero calibration
A/D count value
WVAL
Display and edit
test weight value
WSPAN
*CAL*
Display and edit
span calibration
A/D count value
Figure 4-1. Calibration (CALIBR) Menu
4.1 Front Panel Calibration
To calibrate the indicator using the front panel, do the following:
1. Place the indicator in setup mode (display reads
CONFIG) and remove all weight from
the scale platform. If your test weights require hooks or chains, place the hooks or chains on the scale for zero calibration.
2. Press the display reads CALIBR (see Figure 4-1). Press
ENTER (ZERO) to go to zero calibration
WZERO).
(
3. With calibrate zero. The indicator displays °CAL° while calibration is in progress. When complete, the A/D count for the zero calibration is displayed. Press ENTER again to save the zero calibration value and return to the
WZERO prompt or use the procedure shown in
Figure 4-2 to edit the value. When done, press
RIGHT to go to the WVAL prompt.
4. With the test weight value. If the value is equal to the test weight you are using, press ENTER again to save the value and return to the WVAL prompt; if the value is incorrect, use the procedure shown in Figure 4-2 to edit the value. When done, press RIGHT to go to the
WSPAN prompt.
RIGHT (UNITS) button until the
WZERO displayed, press ENTER to
WVAL displayed, press ENTER to show
XXXXXXXXXXXXXX
REZERO
*CAL*
Press Enter to
remove offset from
zero and span
calibrations
NOTE: When calibrating the indicator with
LB/OZ as the primary unit, enter the WVAL value in ounces. For example, to calibrate a 5 lb scale using LB/OZ, enter 80 (5 lb x 16 oz/lb) as the WVAL value.
VERSDEFLTDIG IN
5. With WSPAN displayed, place test weights on
the scale then press ENTER to calibrate span. The indicator displays °CAL° while calibration is in progress. When complete, the A/D count for the span calibration is displayed. Press
ENTER again to save the span calibration value
and return to the WSPAN prompt or use the procedure shown in Figure 4-2 to edit the value. When done, press RIGHT to go to the
REZERO prompt.
0 0 0 0 0 0
To edit numeric values, press the button to select the leftmost digit. Each digit flashes when selected: Press increment the value of the selected digit; press the
RIGHT
Once the last digit has been edited, press again and the digit stops flashing. Press to save the value entered and return to the menu level above, or press starting with the leftmost digit.
button to move right to the next digit.
RIGHT
UNITS (RIGHT)
ZERO (ENTER)
RIGHT
ENTER
to edit the value again,
to
20 IQ plus 2100 Installation Manual
Figure 4-2. Editing Procedure for Numeric Values
Page 25
6. The rezero function is used to remove a calibration offset when hooks or chains are used to hang the test weights.
If no other apparatus was used to hang the
test weights during calibration, remove the test weights and press
RIGHT to return to
the CALIBR menu.
If hooks or chains were used during
calibration, remove these and the test weights from the scale. With all weight removed, press
ENTER to rezero the scale.
This function adjusts the zero and span calibration values. The indicator displays
°CAL° while the zero and span calibrations
are adjusted. When complete, the adjusted A/D count for the zero calibration is displayed. Press ENTER again to save the rezero value and return to the REZERO prompt or use the procedure shown in Figure 4-2 to edit the value. When done, press RIGHT to return to the CALIBR menu.
7. Press the setup switch to exit setup mode.
4.2 EDP Command Calibration
To calibrate the indicator using EDP commands, the indicator EDP port must be connected to a terminal or personal computer. (See Section 2.3.3 on page 5 for EDP port pin assignments; see Section 5.0 on page 22 for more information about using EDP commands.) Once the indicator is connected to the sending device, do the following:
1. Place the indicator in setup mode (display
CONFIG) and remove all weight from
reads the scale platform. If your test weights require hooks or chains, place the hooks or chains on the scale for zero calibration.
2. Send the WZERO EDP command to calibrate zero. The indicator displays calibration is in progress.
3. Send the WVAL command to enter the test weight value in the following format:
WVAL=nnnnnn<CR>
°CAL° while
4.3 Revolution™ Calibration
To calibrate the indicator using Revolution, the indicator must be in setup mode with the EDP port connected to a PC running the Revolution configuration utility. Use the following procedure to calibrate the indicator:
1. Select main menu.
2. On the Weight Calibration display, select the indicator model ( port then click OK.
3. Revolution uploads calibration data from the indicator then presents the information in a display like that shown in Figure 4-3.
4. Enter the span calibration then click OK.
5. The Zero Calibration dialog box prompts you to remove all weight from the scale. Clear the scale and click
NOTE: If your test weights require hooks or chains, place the hooks or chains on the scale for zero calibration.
6. When zero calibration is complete, the Span Calibration dialog box prompts you to place test weights on the scale for span calibration. Place tests weights on the scale then click OK.
7. The Rezero dialog box prompts you to remove all weight, from the scale (including hooks or chains). Remove the weights then click
8. When calibration is complete, the
Settings fields of the Weight Calibration
display are filled in. Click Exit to save the new values and return to the Rev olution main menu; to restore the previous calibration values, click
Restore Settings.
Calibrate Indicator from the Revolution
IQ+210) and communications
Value of Test Weight to be used for
OK to begin zero calibration.
OK.
New
4. Place test weights on scale equal to the specified WVAL.
5. Send the WSPAN EDP command to calibrate span. The indicator displays
°CAL° while
calibration is in progress.
6. To remove an offset value, clear all weight from the scale, including hooks or chains used to hang test weights, then send the REZERO EDP command. The indicator displays
°CAL°
while the zero and span calibrations are adjusted.
7. Send the KEXIT EDP command to exit setup mode.
Figure 4-3. Revolution Calibration Display
Calibration 21
Page 26
5.0 EDP Commands
The IQ plus 210 indicator can be controlled by a personal computer or remote keyboard connected to the indicator serial port. Control is provided by a set of EDP commands that can simulate front panel key press functions, display and change setup parameters, and perform reporting functions. The EDP port provides the capability to print configuration data or to save that data to an attached personal computer. This section describes the EDP command set and procedures for saving and transferring data using the serial port.
5.1 The EDP Command Set
The EDP command set can be divided into five groups: key press commands, reporting commands, the RESETCONFIGURATION special function command, parameter setting commands, and transmit weight data commands.
When the indicator processes an EDP command, it responds with the message verifies that the command was received and has been executed. If the command is unrecognized or cannot be executed, the indicator responds with
The following sections list the commands and command syntax used for each of these groups.
5.1.1 Key Press Commands
Key press EDP commands (see Table 5-1) simulate pressing the buttons on the front panel of the indicator.
Command Function
KZERO Press the ZERO button KUNITS Press the UNITS button KPRIM Display primary units KPRINT Send demand data to serial port KRIGHTARROW In setup mode, move right in the menu
(press the UNITS button)
KENTER In setup mode, press the ENTER
(ZERO) button
KEXIT In setup mode only, exit setup mode
Table 5-1. EDP Key Press Commands
OK. The OK response
??.
5.1.2 Reporting Commands
Reporting commands (see Table 5-2) send specific information to the EDP port. These commands can be used in both setup mode and normal mode.
Command Function
DUMPALL List all parameter values VERSION
PWrite current displayed weight. See
ZZ Write current weight and annunciator
RS Reset the indicator SWrite one frame of stream format
5.1.3 The RESETCONFIGURATION Command
Write IQ plus 210 software version
Section 6.2 on page 26 for more information.
status. See Section 6.2 on page 26 for more information.
Table 5-2. EDP Reporting Commands
The RESETCONFIGURATION command can be used to restore all configuration parameters to their default values.
This command is equivalent to using the RESET function on the DEFLT menu.
NOTE: All load cell calibration settings are lost when the
RESETCONFIGURATION command is run.
5.1.4 Parameter Setting Commands
Parameter setting commands allow you to display or change the current value for a particular configuration parameter (Tables 5-3 through 5-8).
Current configuration parameter settings can be displayed in either setup mode or normal mode using the following syntax:
command<ENTER>
Most parameter values only can be changed in setup mode. Use the following command syntax when changing parameter values:
command=value<ENTER>
where value is a number or a parameter value. Use no spaces before or after the equal (=) sign. If you type an incorrect command or value, the display reads
??.
Changes to the parameters are saved as they are entered but typically do not take effect until you exit setup mode.
For example, to set the motion band parameter to 5, type the following:
MOTBAND=5D<ENTER>
22 IQ plus 2100 Installation Manual
Page 27
Command Description Values
GRADS Graduations 1–10 000 ZTRKBND Zero track band OFF, 0.5D, 1D, 3D ZRANGE Zero range 1.9%, 100% MOTBAND Motion band 1D, 2D, 3D, 5D, 10D, 20D, OFF OVRLOAD Overload FS+2%, FS+1D, FS+9D, FS FILTER Digital filtering 1, 2, 4, 8, 16, 4RT, 8RT, 16RT
Table 5-3. CONFIG EDP Commands
Command Description Values
PRI.DECPNT Primary units decimal position 8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888, 888880
(88 88.8 or 888 88 in LB/OZ display mode) PRI.DSPDIV Primary units display divisions 1D, 2D, 5D PRI.UNITS Primary units LB, KG, OZ, G, LB/OZ ALT.LB
ALT.KG ALT.OZ ALT.G ALT.LBOZ
Enable alternate units ON, OFF
Table 5-4. FORMAT EDP Commands
Command Description Values
WZERO Zero calibration — WVAL Test weight value test_weight_value WSPAN Span calibration — REZERO Rezero — LC.CD Set deadload count value LC.CW Set span count value
Table 5-5. CALIBR EDP Commands
Command Description Values
EDP.BAUD Serial port baud rate 1200, 2400, 4800, 9600 EDP.BITS Serial port data bits/parity 8NONE, 7EVEN, 7ODD EDP.TERMIN Serial port termination character CR/LF, CR EDP.ECHO Serial port echo command OFF, ON PRNFREQ Print frequency DEMAND, AUTO1, AUTO2, STREAM
Table 5-6. SERIAL EDP Commands
Command Description Values
DIGIN1 DIGIN2
Digital input function OFF, ZERO, UNITS, PRINT
Table 5-7. DIG IN EDP Commands
EDP Commands 23
Page 28
5.1.5 Normal Mode Commands
The serial transmit weight data commands (see Table 5-8) transmit data to the serial port on demand. The transmit weight data commands are valid only in normal operating mode.
Command Description Response Format
SX Start serial streaming OK or ?? EX Stop serial streaming OK or ?? RS Reset system — XG Transmit gross weight in displayed units nnnnnn UU
where nnnnnn is the weight value, UU is the units.XGL Transmit gross weight in pounds XGK Transmit gross weight in kilograms XGO Transmit gross weight in ounces XGG Transmit gross weight in grams XGC Transmit gross weight in pounds and ounces XE Query system error conditions nnnnn nnnnn
See Section 6.1.2 on page 25 for detailed information about
the XE command response format.
Table 5-8. Normal Mode EDP Commands
5.2 Saving and Transferring Data
Connecting a personal computer to the IQ plus 210 EDP port allows you to save indicator configuration data to the PC or to download configuration data from the PC to an indicator. The following sections describe the procedures for these save and transfer operations.
5.2.1 Saving Indicator Data to a Personal Computer
Configuration data can be saved to a personal computer connected to the EDP port. The PC must be running a
communications program such as PROCOMMPLUS See Section 2.3.3 on page 5 for information about serial communications wiring and EDP port pin assignments.
When configuring the indicator, ensure that the values set for the BAUD and BITS parameters on the SERIAL menu match the baud rate, bits, and parity settings configured for the serial port on the PC.
To save all configuration data, send the DUMPALL EDP command to the indicator. The IQ plus 210 responds by sending all configuration parameters to the PC as ASCII-formatted te xt.
5.2.2 Downloading Configuration Data from PC to Indicator
Configuration data saved on a PC or floppy disk can be downloaded from the PC to an indicator. This procedure is useful when a number of indicators with similar configurations are set up or when an indicator is replaced.
To download configuration data, connect the PC to the EDP port as described in Section 5.2.1. Place the
®
.
indicator in setup mode and use the PC communications software to send the saved configuration data to the indicator. When transfer is complete, calibrate the indicator as described in Section 4.0 on page 20.
NOTES:
•Calibration settings are included in the configuration data downloaded to the indicator. If the receiving indicator is a direct replacement for another IQ plus 210 and the attached scale is not changed, recalibration is not required.
•When downloading configurations that include changed serial communications settings, edit the data file to place the serial communications changes at the end of the file. Communication between the PC and indicator will be lost once the indicator receives settings for baud rate (BAUD parameter) or data bits and parity (BITS parameter) that do not match those configured for the PC.
24 IQ plus 2100 Installation Manual
Page 29
6.0 Appendix
6.1 Error Messages
The IQ plus 210 indicator provides a number of error messages. When an error occurs, the message is sho wn on the indicator LED display . Error conditions can also be check ed remotely by using the XE EDP command as described in Section 6.1.2.
Error Message Description Solution
E A/D A/D physical error Call Rice Lake Weighing Systems (RLWS) Service.
EEEROM EEPROM physical error
EVIREE Virgin EEPROM Use the DEFLT menu to restore defaults, then recalibrate load cells. EPCKSM Parameter checksum error EACKSM A/D calibration checksum error A/D converter requires recalibration. Call RLWS Service. ELCKSM Load cell calibration checksum error Recalibrate load cells.
EIDATA Internal RAM checksum error Call RLWS Service.
INOVFL Internal overflow error Remove weight from scale. If error recurs, check configuration, check
scale for damage.
OVERFL Overflow error Weight value too large to be displayed.
------
______ A/D underrange A/D reading < –4 mV. Check scale for binding or damage.
Gross > overload limit Gross value exceeds overload limit. Check configuration.
Table 6-1. IQ plus 210 Error Messages
6.1.1 Displayed Error Messages
The IQ plus 210 provides a number of front panel error messages to assist in problem diagnosis. Table 6-1 lists these messages and their meanings.
6.1.2 Using the XE EDP Command
The XE EDP command can be used to remotely query the IQ plus 210 for the error conditions shown on the front panel. The XE command returns two 5-digit numbers in the format:
xxxxx yyyyy
where xxxxx contains a decimal representation of any existing error conditions as described in Table 6-2.
If more than one error condition exists, the number returned is the sum of the values representing the error conditions. For example, if the XE command returns the number 528, this value represents the sum of an A/D physical error (512) and an A/D calibration checksum error (16).
The second number returned (
yyyyy) uses the same bit
assignments as shown in Table 6-2 to indicate whether the test for the error condition was run. For example, the value
yyyyy = 51807 represents the decimal
equivalent of the binary value 1100 1110 0111 1111. Using the bit assignments in Table 6-2, this value indicates all tests were run.
Error
Code
0 No error 0000 0000 0000 0000 1 EEPROM physical error 0000 0000 0000 0001 2Virgin EEPROM 0000 0000 0000 0010 4 Parameter checksum error 0000 0000 0000 0100 8 Load cell calibration
checksum error
16 A/D calibration checksum
error 32 not assigned 0000 0000 0010 0000 64 Internal RAM checksum error 0000 0000 0100 0000
128 not assigned 0000 0000 1000 0000 256 not assigned 0000 0001 0000 0000
512 A/D physical error 0000 0010 0000 0000 1024 not assigned 0000 0100 0000 0000 2048 Internal overflow error 0000 1000 0000 0000 4096 not assigned 0001 0000 0000 0000 8192 not assigned 0010 0000 0000 0000
16384 A/D underrange 0100 0000 0000 0000 32768 Gross > overload limit 1000 0000 0000 0000
Description Binary Value
0000 0000 0000 1000
0000 0000 0001 0000
Table 6-2. Error Codes Returned on XE Command
Appendix 25
Page 30
6.2 Status Messages
Two EDP commands, P and ZZ, can be used to provide status about the indicator. These commands are described in the following sections.
6.2.1 Using the P EDP Command
The P EDP command returns the current displayed weight value to the EDP port. If the indicator is in an underrange or overload condition, the weight value is replaced with
6.2.2 Using the ZZ EDP Command
The ZZ EDP command can be used to remotely query which annunciators are currently displayed on the indicator front panel. The ZZ command returns the currently displayed weight and a decimal number representing the LED annunciators currently lit. The format of the returned data is:
wwwwww zzz
where wwwwww is the current displayed weight, zzz is the annunciator status value (see Table 6-3). If more than one annunciator is lit, the second number returned is the sum of the values representing the activ e annunciators.
&&&&&& (overload) or :::::: (underrange).
For example, if the annunciator status v alue returned on the ZZ command is 136, the center of zero and lb annunciators are lit: 136 represents the sum of the values for the center of zero annunciator (128) and the lb annunciator (8).
Decimal
Value
1 Reserved 2 Negative value 4oz
8lb 16 g 32 kg 64 Motion
128 Center of zero
Table 6-3. Status Codes Returned on the ZZ Command
Annunciator
6.3 Continuous Output (Stream) Format
Figure 6-1 shows the continuous output format sent to the IQ plus 210 EDP or printer port when the STREAM parameter (SERIAL menu) is set to either EDP or PRN.
<STX>
ASCII 02 (decimal)
Polarity: <Space> = Positive <–> = Negative <^> = Overload <]> = Underrange
Weight data: 7 digits, right-justified, with decimal point, leading zero suppression.
Overload = Underrange = Display overflow = OVERFL
<POL> <wwwwwww> <UNIT>
^^^^^^^
] ] ] ] ] ] ]
Figure 6-1. Continuous Output Data Format
<G>
G = Gross
L = pounds K = kilogram G = grams O = ounces <space> = lb/oz
<S>
Status: <space> = valid I = Invalid M = In motion O = Over/under range
<TERM>
<CR> <LF>
or <CR>
ASCII 13, 10
(decimal)
26 IQ plus 2100 Installation Manual
Page 31
6.4 Specifications
6.4.1 IQ plus 210 Indicator Power
Line Voltages 115 or 230 VAC Frequency 50 or 60 Hz Power Consumption 70 mA @ 115 VAC (8W)
35 mA @ 230 VAC (8W)
Fusing
115 VAC 2 x 200 mA TR5 subminiature fuses
Wickmann Time-Lag 19374 Series UL Listed, CSA Certified
230 VAC 2 x 100 mA TR5 subminiature fuses
Wickmann Time-Lag 19372 Series UL Recognized, VDE Approved
Analog Specifications
Full Scale Input Signal Up to 35 mV Excitation Voltage 10 ± 0.25 VDC,
4 x 350 load cells
Analog Signal
Input Range 0.6 mV/V – 4.5 mV/V
Analog Signal
Sensitivity 0.3 µV/graduation minimum,
Input Impedance 200 M, typical Display Resolution 10 000 dd Input Sensitivity 155 nV per internal count System Linearity Within 0.01% of full scale Calibration Method Software, constants stored in Common Mode
Voltage –1.05 to +0.95 V, referred to earth
RFI Protection Signal, excitation, and sense lines
Digital Specifications
Digital Inputs 2 inputs, TTL or switch closure, Digital Filter Software selectable
Serial Communications
Serial Port Full duplex RS-232; 9600, 4800,
Operator Interface
Display 6-digit LED display. 7-segment, .8 in LED annunciators Negative value, center of zero, motion, Keypad 2 piezo switches
1.5 µV/grad recommended
EEPROM
protected by capacitor bypass
active-low
2400, 1200 bps; 7 or 8 data bits; even, odd, or no parity
(20 mm) digits kg, g, lb, oz
Enclosure
Enclosure Dimensions 9.5 in x 6 in x 2.75 in
24 cm x 15 cm x 7 cm Weight 6.1 lb (2.8 Kg) Rating/Material NEMA 4X/IP66, stainless steel
Approvals
O
C
N
NTEP
F
L
E
A
R
N
E
O
I
N
T
C
A
E
N
O
N
S
CoC Number 00-045
E
W
R
E
U
I
S
G
A
H
E
T
S
M
A
D
N
Accuracy Class III/III L
C
LISTED
U
L
®
UL
US
File Number E151461
Measurement Canada Approved
Approval AM-5374 Accuracy Class III/III HD
6.4.2 RL2100 Bench Scale Load Cells
RL1042 aluminum load cell (standard) RL1380 stainless steel load cell (optional) Rated Output 2.0 mV/V Maximum Overload 150% Cable 10 ft (3.1 m) Output Impedance 350Ω, nominal Operating Temperature–10 to +40°C (legal);
–10 to +50°C (industrial) Storage Temperature –25 to +70°C
Dimensions & Capacities
Base Dimension Capacity
10 in x 10 in (254 mm x 254 mm)5 lb (2.5 Kg)
10 lb (5 Kg)
25 lb (10 Kg) 12 in x 12 in (305 mm x 305 mm)50 lb (25 Kg)
100 lb (50 Kg)
Approvals
O
C
N
NTEP
F
L
E
A
R
N
E
O
I
N
T
C
A
E
N
O
N
S
CoC Number 95-072A2
E
W
R
E
U
I
S
G
A
H
E
T
S
M
A
D
N
Accuracy Class III
Measurement Canada Approved
Approval AM-5082 Rev 1
Environmental
Operating Temperature–10 to +40°C (legal);
–10 to +50°C (industrial) Storage Temperature –25 to +70°C Humidity 0–95% relative humidity Altitude 2000 m (6500 ft) maximum
Appendix 27
Page 32
IQ plus 2100 Limited Warranty
Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for two years.
RLWS warrants that the equipment sold hereunder will conform to the current written specifications authorized by RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipment fails to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within the warranty period subject to the following conditions:
Upon discovery by Buyer of such nonconformity, RLWS will be given prompt written notice with a detailed explanation of the alleged deficiencies.
Individual electronic components returned to RLWS for warranty purposes must be packaged to prevent electrostatic discharge (ESD) damage in shipment. Packaging requirements are listed in a publication, Protecting Your Components From Static Damage in Shipment, available from RLWS Equipment Return Department.
Examination of such equipment by RL WS confirms that the nonconformity actually e xists, and was not caused by accident, misuse, neglect, alteration, improper installation, improper repair or improper testing; RLWS shall be the sole judge of all alleged non-conformities.
Such equipment has not been modified, altered, or changed by any person other than RLWS or its duly authorized repair agents.
•RLWS will have a reasonable time to repair or replace the defective equipment. Buyer is responsible for shipping charges both ways.
In no event will RLWS be responsible for tra vel time or on-location repairs, including assembly or disassembly of equipment, nor will RLWS be liable for the cost of any repairs made by others.
THESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WITHOUT
LIMITATION NOR DISTRIBUTOR WILL, IN ANY EVENT , BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
RLWS
OR
REPLACEMENT OF SUCH GOODS. IN ACCEPTING THIS WARRANTY , THE BUYER WAIVES ANY AND ALL OTHER
CLAIMS
S
HOULD THE SELLER BE OTHER THAN RLWS, THE BUYER AGREES TO LOOK ONLY TO THE SELLER FOR
WARRANTY
N
O TERMS, CONDITIONS, UNDERSTANDING, OR AGREEMENTS PURPORTING TO MODIFY THE TERMS OF THIS
WARRANTY OFFICER
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NEITHER RLWS
AND BUYER AGREE THAT RLWS’S SOLE AND EXCLUSIVE LIABILITY HEREUNDER IS LIMITED TO REPAIR
TO WARRANTY.
CLAIMS.
SHALL HAVE ANY LEGAL EFFECT UNLESS MADE IN WRITING AND SIGNED BY A CORPORATE
OF RLWS AND THE BUYER.
© 20 Rice Lake Weighing Systems, Inc. Rice Lake, WI USA. All Rights Reserved.
RICE LAKE WEIGHING SYSTEMS • 230 WEST COLEMAN STREET • RICE LAKE, WISCONSIN 54868 • USA
28 IQ plus 2100 Installation Manual
Page 33
Page 34
230 W. Coleman St. • Rice Lake, WI 54868 • USA
U.S. 800-472-6703 • Canada/Mexico 800-321-6703
International 715-234-9171
www.ricelake.com
mobile: m.ricelake.com
© 2011 Rice Lake Weighing Systems PN 64640 06/11
Loading...