IQ plus 2100 Limited Warranty.............................................................................................................. 28
IQ plus 210 Installation Manual
Page 5
About This Manual
This manual is intended for use by service technicians
responsible for installing and servicing IQ plus
digital bench scales.
Configuration and calibration of the indicator can be
accomplished using the indicator front panel keys, the
EDP command set, or Version 2.3 or later of the
Revolution
™
configuration utility. See Section 3.1 on
page 10 for information about configuration methods.
®
2100
1.0Introduction
Warning
Some procedures described in this
manual require work inside the indicator
enclosure. These procedures are to be
performed by qualified service personnel
only
.
Authorized distributors and their employees
can view or download this manual from the
Rice Lake Weighing Systems distributor site
www.rlws.com
at
.
The IQ plus 2100 digital bench scale consists of an IQ
plus 210 digital weight indicator and an RL2100 bench
scale. The indicator can be mounted to the bench scale
using the 12-inch column or the attachment bracket, or
can be separately mounted on the tilt stand.
The IQ plus 210 is a single-channel digital weight
indicator housed in a NEMA 4X/IP66-rated stainless
steel enclosure. The indicator front panel consists of a
large (.8 in, 20 mm), six-digit, seven-segment LED
display, seven LED annunciators, and two piezo
switches used to zero the scale and select the displayed
units. Features include:
Ω
•Drives up to four 350
•Supports 4- and 6-wire load cell connections
•Two configurable digital inputs
•Single serial port supports full duplex, RS-232
communications at up to 9600 bps
•Available in 115 VAC and 230 VAC versions
load cells
The RL2100 bench scale is available in the following
configurations:
•5, 10, and 25 lb (2.5, 5, and 10 Kg) capacities with a
10" x 10" (25cm x 25cm) platform
•50 and 100 lb (25 and 50 Kg) capacities with a 12" x
12" (30cm x 30cm) platform
The RL1042 anodized aluminum load cell is standard
in the RL2100 bench scale; the RL1380 stainless steel
load cell is available as an option. A stainless steel
clamshell enclosure is also available for load cell
protection.
1.1Operating Modes
The IQ plus 210 has two modes of operation:
Normal (weighing) mode
Normal mode is the “production” mode of the
indicator. The indicator displays the gross weight,
and the LED annunciators described on page 2
indicate scale status and the type of weight value
displayed. Once configuration is complete and a
legal seal is affixed to the back of the indicator, this
is the only mode in which the IQ plus 210 can
operate.
Setup mode
Most of the procedures described in this manual
require the indicator to be in setup mode, including
configuration and calibration.
To enter setup mode, remove the large fillister head
screw from the enclosure backplate. Insert a
screwdriver or a similar tool into the access hole
and press the setup switch once. The indicator
display changes to show the word
CONFIG
.
Introduction
1
Page 6
2
1.2Front Panel
Figure 1-1 shows the IQ plus 210 LED annunciators, buttons, and normal mode ke y functions.
DIGITAL WEIGHT INDICATOR
Neg
Zero
Motion
ZERO
ZERO
Set scale weight to zero
Figure 1-1. IQ plus 210 Front Panel, Showing LED Annunciators and Normal Mode Key Functions
Change displayed units
UNITS
UNITS
kg
g
lb
oz
1.3LED Annunciators
The IQ plus 210 display uses a set of seven LED
annunciators to provide additional information about
the value being displayed:
Neg annunciator lights to show that the displayed
•
value is negative.
•Zero : Weight is within 0.25 graduations of zero.
This annunciator lights when the scale is zeroed.
•Motion : Scale in motion outside the specified
motion band. Zeroing and print operations can only
be done when the motion LED is off.
lb, kg, oz, and g annunciators indicate the units
•
associated with the displayed value: lb=pounds,
kg=kilograms, oz=ounces, g=grams. lb and oz
annunciators are both lit when the displayed weight
is in pounds and ounces.
1.4Indicator Operations
Basic IQ plus 210 operations are summarized below.
Zero Scale
1. Remove material from the scale and wait for
Motion
the
2. Press the
lights to indicate the scale is zeroed.
Change Display Units
Press the
UNITS
between primary and alternate units. The LED
annunciators on the right side of the display show the
current displayed units. If more than one alternate unit
is configured, the display cycles through the alternate
units before returning to primary units; if no alternate
units are configured, the display does not change when
the button is pressed.
LED to go out.
ZERO
button. The
annunciator
Zero
button to change the displayed units
IQ plus 2100 Installation Manual
Page 7
Cord grip
Insulated cable
Foil (silver side out)
Grounding clamp
Shield wire (cut)
Length of foil before folding
back on cable insulation
Cut insulation here
for foil-shielded cables
Braid
Cut insulation here
for braided cables
NOTE: Install lockwashers
first, against enclosure,
under grounding clamp
2.0Setup and Service Information
This section describes setup and service procedures for the IQ plus 2100 digital bench scale, including installation
and maintenance information, replacement parts lists, and assembly drawings. See Sections 2.1 through 2.5 for IQ
plus 210 indicator information; see Sections 6.6 and 6.7 for bench scale information.
2.1Enclosure Disassembly
The indicator enclosure must be opened to connect
cables after replacing the load cell, or when connecting
serial communications cables or digital inputs to the
scale.
The IQ plus 210 has no on/off switch.
Warning
Caution
• Use a wrist strap to ground yourself and protect
components from electrostatic discharge (ESD)
when working inside the indicator enclosure.
• This unit uses double pole/neutral fusing which
could create an electric shock hazard. Procedures
requiring work inside the indicator must be
performed by qualified service personnel only.
• The supply cord serves as the power disconnect for
the IQ plus 2100. The power outlet supplying the
indicator must be installed near the unit and be
easily accessible
Ensure power to the indicator is disconnected, then
remove the screws that hold the backplate to the
enclosure body, then lift the backplate away from the
enclosure and set it aside.
2.2Cable Connections
The IQ plus 210 provides three cord grips for cabling
into the indicator: two for the power cord and load cell
cabling, the third for communications and digital input
cables. The free cord grip comes with a plug installed to
prevent moisture from entering the enclosure. If your
application requires serial communication or digital
input cabling, remove the plug and install cables as
described in Sections 2.2.4 and 2.2.5 on page 5.
Before opening the unit, ensure the
power cord is disconnected from the
power outlet.
•Route cables through cord grips and grounding
clamps to determine cable lengths required to
reach cable connectors. Mark cables to remove
insulation and shield as described below:
•For cables with foil shielding, strip insulation
and foil from the cable half an inch (15 mm)
past the grounding clamp (see Figure 2-1). Fold
the foil shield back on the cable where the cable
passes through the clamp. Ensure silver
(conductive) side of foil is turned outward for
contact with the grounding clamp.
•For cables with braided shielding, strip cable
insulation and braided shield from a point just
past the grounding clamp. Strip another half
inch (15 mm) of insulation only to expose the
braid where the cable passes through the clamp
(see Figure 2-1).
IMPOR
•
yellow shield wire 3/4" past the grounding
clamp. Fold wire back and secure between the
cable and clamp. Shield wire function is
provided by contact between the cable shield
and the grounding clamp.
•Route stripped cables through cord grips and
clamps. Ensure shields contact grounding clamps
as shown in Figure 2-1. Tighten grounding clamp
nuts.
•Finish installation using cable mounts and ties to
secure cables inside of indicator enclosure.
TANT!
For load cell cables, strip the
2.2.1Cable Grounding
Except for the power cord, all cables routed through the
cord grips should be grounded against the indicator
enclosure. Do the following to ground shielded cables:
•Use the lockwashers, clamps, and kep nuts
provided in the parts kit to install grounding
clamps on the studs adjacent to the cord grips.
Install grounding clamps only for cord grips that
will be used; do not tighten nuts.
Figur e 2-1. Grounding Clamp Attachment for Foil-Shielded
and Braided Cabling
Setup and Service Information
3
Page 8
To setup switch
Microprocessor
J1
GND
RESET
A/D Converter
GND
11
J3
J2
DIGIN 2 DIGIN 1
UNITS
Button
J4
DIGITAL
+5V
TEST
VR1VR2
JP2
JP1
LOAD CELL CONNECTOR
1 23456
–SIG
+SIG
+SENSE
ANALOG
+5V
TEST
+EXC
–SENSE
J5
–EXC
4
GND
JP1
ON
OFF
Load cell
compensation
jumper
VR3
TxD
1
GND
2
RxD
3
J7
TP1
+
BR1
–
TP2
Transformer
SERIAL COMM
J8
ZERO
Button
J6
Blue wire
Power
Input
F1F2
Brown wire
Figure 2-2. IQ plus 210 CPU and Power Supply Board
2.2.2Load Cells
Wire the load cell cable to connector J5 as shown in
Table 2-1. If using 6-wire load cell cable (with sense
wires), remove jumpers JP1 and JP2 (see Figure 2-2).
For 4-wire installation, leave jumpers JP1 and JP2 on.
NOTE: All models come with ten feet of color-coded
load cell cable. Do not cut this cable! The load cell is
temperature-compensated for an exact cable length of
ten feet. Cutting the load cell cable voids the load cell
warranty.
When connections are complete, use two cable ties to
secure the load cell cable to the inside of the enclosure.
1.Use grounding procedure described in Section 2.2.1
on page 3 to attach shield wire to backplate.
2.For 6-wire connections, remove jumpers JP1 and JP2.
Table 2-1. J5 Pin Assignments
RL1040
RL1042
RL1380
2.2.3Setting the Load Cell Compensation Jumper
The load cell compensation jumper (above the
ZERO
button location on the CPU board; see Figure 2-2) must
be set for the type of load cell connected to the
indicator:
•For load cells with balanced bridges, set the
jumper in the OFF position. Balanced load cells
include the RL1040, RL1250, RL1260, RL1380,
and RL1385.
•For load cells with unbalanced bridges, follow the
procedure below to determine the correct jumper
position. Examples of unbalanced cells include
the RL1042 and RL1010.
To determine the correct jumper position for
unbalanced cells, do the following;
1. Disconnect load cell from indicator.
2. Use an ohmmeter to measure +EXC to +SIG and
+EXC to –SIG. Measured values between the
excitation line and each of the signal lines should
Ω
be within 2–3
.
3. Next, measure –EXC to +SIG and –EXC to
–SIG. Measured values between the excitation
line and each of the signal lines should be within
Ω
.
2–3
≥
4. If the +EXC measurements (step 2) are
5%
larger than the –EXC measurements (step 3), set
the compensation jumper in the ON position. If
the +EXC measurements are < 5% greater (or are
less) than the –EXC measurements, set the
jumper in the OFF position.
IQ plus 2100 Installation Manual
Page 9
2.2.4Serial Communications
To attach serial communications cables, connect
communications cables to connector J7 as shown in
Table 2-2. Use cable ties to secure serial cable to the
inside of the enclosure.
Use the SERIAL menu to configure serial
communications. See Section 3.2.4 on page 18 for
configuration information.
J7 PinLabelFunction
1TxDRS-232 TxD
2GNDRS-232 Ground
3RxDRS-232 RxD
Table 2-2. J7 Pin Assignments
2.2.5Digital Inputs
Digital inputs (connectors J2 and J3) can be used to
perform remote ZERO and UNITS key presses or to
send serial data to a printer (remote PRINT key
function). The inputs are active (on) with lo w v oltage (0
VDC) and can be driven by TTL or 5V logic without
additional hardware. Use the DIG IN menu to configure
the digital inputs. See Section 3.2.5 on page 19 for
information about configuring the digital inputs.
2.4Board Removal
If you must remove the IQ plus 210 CPU board, use the
following procedure:
1. Disconnect power to the indicator. Remove
backplate as described in Section 2.1 on
page 3.
2. Remove connections to J5 (load cell cable), J7
(serial communications), J2 and J3 (digital
inputs), J4 and J6 (piezo button inputs), and J1
(setup switch). Remove blue and brown power
input wires at J8. See Figure 2-2 on page 4 for
connector locations.
3. Remove the five nuts from the CPU board, then
lift the board out of the enclosure.
To replace the CPU board, reverse the above procedure.
Be sure to reinstall cable ties to secure all cables inside
the indicator enclosure.
2.3Enclosure Reassembly
Once cabling is complete, position the backplate over
the enclosure and reinstall the backplate screws. Use
the torque pattern shown in Figure 2-3 to prevent
distorting the backplate gasket. Torque screws to 15
in-lb (1.7 N-m).
Torque pattern
1
To rque backplate screws to 15 in-lb (1.7 N-m)
5
Fillister head screws
3
Figure 2-3. IQ plus 210 Enclosure Backplate
10
Setup switch access screw
7
8
9
4
6
2
Setup and Service Information
5
Page 10
6
2.5IQ plus 210 Replacement Parts
Table 2-3 lists replacement parts for the IQ plus 210, including all parts referenced in Figures 2-4 through 2-8.
Ref
Number
114626Kep nuts, 8-32NC hex (3)Figure 2-4 on page 7, Figure 2-7 on page 8
253651Display and CPU board assembly, 115 VAC (1)Figure 2-5 on page 8
315365Board mounting spacers (5)
452483Enclosure backplate (1)Figure 2-4 on page 7
515626Cable grips, PG9 (2)Figure 2-8 on page 9
630375Nylon seal rings for cable grips (3)
715627Locknuts, PCN9 (3)
819538Cable grip plugs (2)
945042Sealing washers (10)Figure 2-4 on page 7
1044676Sealing washer for setup switch access screw (1)
1142640Setup switch access screw, 1/4 x 28NF x 1/4 (1)
1241965Power cord assembly, 115VAC (1)Figure 2-8 on page 9
1341964Line filter assembly (1)Figure 2-6 on page 8
1414621Kep nuts, 6-32NC hex (5)Figure 2-5 on page 8
2015134Lock washers, internal tooth, No. 8 , Type A (4)Figure 2-4 on page 7, Figure 2-7 on page 8
2252853Overlay membrane panel (1)Figure 2-8 on page 9
2352482Enclosure (1)
2414862Screws, 8-32NC x 3/8 (8)Figure 2-4 on page 7
2645043Ground wire, 4-in., No. 8 (1)
2739037Backplate gasket (1)
2849910Setup switch assembly (1)Figure 2-6 on page 8
2916892Ground/Earth Label (1)
3015650Cable tie mounts (1)
3145302Line filter standoffs (2)
3315131Lock washers, external tooth, No. 6, Type A (10)Figure 2-5 on page 8
3461113Piezo switches (2)Figure 2-8 on page 9
3530623Fillister head screws, 8-32NC x 7/16 (2)Figure 2-4 on page 7
—53848200 mA TR5 subminiature fuses (2), 115 VACF1 and F2 in Figure 2-2 on page 4
Caution
PNDescription (Quantity)Figure
53650Display and CPU board assembly, 230 VAC (1)
45254Power cord assembly, 230VAC (1)
53881100 mA TR5 subminiature fuses (2), 230 VAC
(See Caution below)
For protection against risk of fire, replace fuses only with same type and rating fuse.
See Section 6.4 on page 27 for complete fuse specifications.
IQ plus 2100 Installation Manual
Table 2-3. Replacement Parts
Page 11
4
27
26
B
9/10X
24/8X
A
11
10
See Line Filter
and
Ground Post
Assembly
35/2X
Figure 2-4. IQ plus 210 Backplate Assembly
Setup and Service Information
7
Page 12
8
To setup switch
14/5X
2
See J8 detail below
3/5X
33/10X
31/2X
13
Figure 2-5. IQ plus 210 Enclosure and CPU Board
30
AC input to
CPU board J8
29
Figure 2-6. Line Filter Assembly
Blue wire
28
Setup switch connection
to J1 on CPU board
Cable tie
Brown wire
AC power in from line filter
Line Filter Connection to J8
To backplate
ground post
1/3X/A
20/4X/B
Figure 2-7. Ground Post Assembly
To line filter
To power cord
IQ plus 2100 Installation Manual
Page 13
12
22
23
6/2X
8
5/2X
7/2X
34/2X
To J6 on CPU board
To J4 on CPU board
Figure 2-8. IQ plus 210 Enclosure and Overlay
Setup and Service Information
9
Page 14
3.0Configuration
To configure the IQ plus 210 indicator, the indicator
must be placed in setup mode. The setup switch is
accessed by removing the large fillister head screw on
the enclosure backplate. Insert a screwdriver into the
access hole and press the switch to enter setup mode.
When the indicator is placed in setup mode, the word
CONFIG is shown on the display. The CONFIG menu is
the first of the main menus used to configure the
indicator. Detailed descriptions of these menus are
given in Section 3.2. When configuration is complete,
press the setup switch again to exit setup mode, then
replace the setup switch access screw.
3.1Configuration Methods
The IQ plus 210 indicator can be configured by using
the front panel buttons to navigate through a series of
configuration menus or by sending commands or
configuration data to the EDP port. Configuration using
the menus is described in Section 3.1.3.
Configuration using the EDP port can be accomplished
using either the EDP command set described in
Section 5.0 or Version 2.3 or later of the Revolution
configuration utility .
3.1.1Revolution Configuration
The Revolution configuration utility provides the
preferred method for configuring the IQ plus 210
indicator. Re v olution runs on a personal computer to set
configuration parameters for the indicator. When
Revolution configuration is complete, configuration
data is downloaded to the indicator.
To use Rev olution, do the following:
1. Install Revolution on an IBM-compatible personal computer running Windows
Windows 95. Minimum system requirements
are 8MB of extended memory and at least
5MB of available hard disk space.
2. With both indicator and PC powered off,
connect the PC serial port to the RS-232 pins
on the indicator EDP port.
3. Power up the PC and the indicator. Use the
setup switch to place the indicator in setup
mode.
4. Start the Revolution program.
Figure 3-1 shows an example of one of the Revolution
configuration displays.
Revolution provides online help for each of its
configuration displays. Parameter descriptions provided
in this manual for front panel configuration can also be
used when configuring the indicator using Revolution:
the interface is different, but the parameters set are the
same.
™
3.1.2EDP Command Configuration
The EDP command set can be used to configure the IQ
plus 210 indicator using a personal computer, terminal,
or remote keyboard. Like Revolution, EDP command
configuration sends commands to the indicator serial
port; unlike Revolution, EDP commands can be sent
using any external device capable of sending ASCII
characters over a serial connection.
EDP commands duplicate the functions available using
the indicator front panel and provide some functions
not otherwise available. EDP commands can be used to
simulate pressing front panel buttons, to configure the
indicator, or to dump lists of parameter settings. See
Section 5.0 on page 22 for more information about
using the EDP command set.
®
3.11 or
Figure 3-1. Sample Revolution Display
Revolution supports both uploading and downloading
of indicator configuration data. This capability allows
configuration data to be retrieved from one indicator,
edited, then downloaded to another.
10IQ plus 2100 Installation Manual
Page 15
3.1.3Front Panel Configuration
DIGITAL WEIGHT INDICATOR
ZERO/ENTER
Move down in menu or
enter parameter value and
return to menu level above
UNITS/RIGHT
Move right in menu /
next parameter value
kg
g
lb
oz
ZERO
UNITS
Zero
Motion
Neg
The IQ plus 210 indicator can be configured using a series of menus accessed through the indicator front panel
when the indicator is in setup mode. Table 3-1 summarizes the functions of each of the main menus.
MenuMenu Function
CONFIGConfigurationConfigure grads, zero tracking, zero range, motion band, overload, and digital filtering
parameters.
FORMATFormatSet format of primary units, select alternate display units.
CALIBRCalibrationCalibrate indicator. See Section 4.0 on page 20 for calibration procedures.
SERIALSerialConfigure serial port.
DIG INDigital InputAssign digital input functions.
DEFLTDefaultRestore default configuration.
VERSVersionDisplay installed software version number.
Table 3-1. IQ plus 210 Menu Summary
Figure 3-2. Front Panel Key Functions in Setup Mode
Configuration11
Page 16
Front Panel Menu Navigation
The front panel buttons are used to navigate through the
menus in setup mode (see Figure 3-2). The
UNITS
(RIGHT) button scrolls right (horizontally) on the same
menu level; the
ZERO (ENTER) button moves down
(vertically) to different menu levels and serves as an
Enter key for selecting parameter values within the
menus.
Most of the configuration menus have three or four
levels. Figure 3-3 shows the general structure of the IQ
plus 210 configuration menus. Note the following:
•On the first (main) menu level, press RIGHT to scroll
through the menus. Pressing RIGHT from the VERS
menu wraps around to the CONFIG menu. Press
ENTER from any of the main menus to move down
to the first parameter for that menu.
•On the middle (parameter) menu levels, press
RIGHT to scroll through the parameter prompts
for the menu. When the last parameter is
displayed, pressing RIGHT returns to the level
above. Press ENTER from any parameter prompt
to move down to the values prompts for that
parameter.
•On the bottom (parameter value) menu level,
RIGHT to scroll through the values for the
press
parameter. Pressing RIGHT from the last value
wraps around to first value. Press ENTER to
select the value and return to the parameter
prompt.
Figure 3-4 on page 13 shows an example of the
navigation used to select a parameter value under the
SERIAL menu.
To edit numerical values, use the
select the digit; use the
ENTER button to increment or
RIGHT button to
decrement the value of the flashing digit (see Figure 3-5
on page 13). Once all digits have been edited, press
ENTER to save the value and return to the parameter
prompt.
Menu Navigation
General Structure
RIGHTRIGHTRIGHT
ENTER
Pressing
RIGHTRIGHT
Parameter 1
ENTER
RIGHT
Value 1
RIGHT
Parameter 2Parameter 3Parameter 4
RIGHTRIGHT
Value 2Value 3Value 4
RIGHTRIGHTRIGHT
SERIALCALIBRCONFIGFORMAT
at the last parameter prompt returns to the menu prompt
RIGHT
XXXXXXXXXXXXXX
RIGHT
RIGHT
VERSDEFLTDIG IN
ZERO
(ENTER)
UNITS
(RIGHT)
RIGHT
Press
ENTER
12IQ plus 2100 Installation Manual
at any parameter value to select the value and return to the parameter prompt
Figure 3-3. General Menu Structure
Page 17
PRNFRQ
DEMAND
AUTO1
BITSTERMIN
BAUD
CR
9600
7ODD
7EVEN
8NONE
CR/LF
48002400
1200
AUTO2
STREAM
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIGFORMAT
RIGHT
RIGHTRIGHTRIGHT
ENTER
Menu Navigation
Serial Menu Example
RIGHT
RIGHTRIGHT
RIGHT
ENTER
RIGHTRIGHT
RIGHT
With any of the BAUD parameter values displayed, press
ENTER
to select the value and return to the BAUD parameter prompt
RIGHT
Navigation functions same
as BAUD parameter
When displaying last
parameter prompt, press
RIGHT
to return to the
first-level menu prompt
(ENTER)
ZERO
(RIGHT)
UNITS
00 0 0 0 0
To edit numeric values, press the
button to select the leftmost digit. Each digit flashes
when selected: Press
increment the value of the selected digit; press the
RIGHT
Once the last digit has been edited, press
again and the digit stops flashing. Press
to save the value entered and return to the menu level
above, or press
starting with the leftmost digit.
Figure 3-5. Editing Procedure for Numeric Values
button to move right to the next digit.
RIGHT
Figure 3-4. Setup Mode Menu Navigation
UNITS (RIGHT)
ZERO (ENTER)
to edit the value again,
RIGHT
ENTER
to
Configuration13
Page 18
3.2Menu Structures and Parameter Descriptions
The following sections provide graphic representations of the IQ plus 210 menu structures. In the actual menu
structure, the values under each parameter are arranged horizontally . To save page space, menu choices are shown in
vertical columns. The factory def ault setting appears at the top of each column.
Most menu diagrams are accompanied by a table that describes all parameters and parameter values associated with
that menu. Default parameter values are shown in bold type.
3.2.1Configuration Menu
GRADS
5000
number
SERIALCALIBRCONFIGFORMAT
ZTRKBNZRANGE
OFF
0.5D
1D
3D
100%
1.9%
XXXXXXXXXXXXXX
MOTBAN
1D
2D
3D
5D
10D
20D
OFF
VERSDEFLTDIG IN
OVRLOA
FS+2%
FS+1D
FS+9D
Figure 3-6. Configuration Menu
CONFIG Menu
ParameterChoicesDescription
Level 2 submenus
GRADS5000
number
Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1–10 000 and should be consistent with legal requirements and environmental
limits on system resolution.
FS
FILTER
4
8
16
4RT
8RT
16RT
1
2
To calculate GRADS, use the formula, GRADS = Capacity / Display Divisions.
Display divisions for primary and secondary units are specified on the FORMAT menu.
ZTRKBNOFF
0.5D
1D
Zero track band. Automatically zeroes the scale when within the range specified, as long
as the input is within the configured zero range (ZRANGE parameter). Selections are ±
display divisions. Maximum legal value varies depending on local regulations.
3D
ZRANGE100%
1.9%
Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is
± 1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use 1.9% for legal-for-trade applications.
MOTBAN1D
2D
3D
5D
10D
20D
Motion band. Sets the level, in display divisions, at which scale motion is detected. If
motion is not detected for 1 second or more, the standstill symbol lights. Some
operations, including print, tare, and zero, require the scale to be at standstill. Maximum
legal value varies depending on local regulations.
If OFF is selected, ZTRKBN should also be set to OFF.
OFF
14IQ plus 2100 Installation Manual
Table 3-2. Configuration Menu Parameters
Page 19
CONFIG Menu
KGGOZLB
PRIMARALTUNT
DSPDIVUNITSDECPNT
OZ
KG
88888.8
8888.88
1D
5D
2D
LB
88.8888
8.88888
888880
888888
OFF
ON
LB/OZ
G
LB/OZ
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIGFORMAT
OFF
ON
OFF
ON
OFF
ON
OFF
ON
888 88
88 88.8
If UNITS=LB/OZ,
DECPNT selections are:
888.888
ParameterChoicesDescription
OVRLOAFS+2%
FS+1D
Overload. Determines the point at which the display blanks and an out-of-range error
message is displayed. Maximum legal value varies depending on local regulations.
FS+9D
FS
FILTER4
8
16
4RT
8RT
16RT
1
Digital filtering. Selects the digital filtering rate used to reduce the effects of mechanical
vibration from the immediate area of the scale.
Choices indicate the number of A/D conversions that are averaged to obtain the
displayed reading. A higher number gives a more accurate display by minimizing the
effect of a few noisy readings, but slows down the settling rate of the indicator. Values
with the RT suffix provide a raised threshold for filtering larger noise spikes.
2
Table 3-2. Configuration Menu Parameters (Continued)
3.2.2Format Menu
Figure 3-7. Format Menu
Configuration15
Page 20
FORMAT Menu
ParameterChoicesDescription
Level 2 submenus
PRIMARUNITS
DECPNT
DSPDIV
ALTUNTLB
KG
OZ
G
LB/OZ
Level 3 submenus
Primary Units (PRIMAR Parameter)
DECPNT888.888
8888.88
88888.8
888888
888880
8.88888
88.8888
or
88 88.8
888 88
DSPDIV1D
2D
5D
UNITSLB
KG
OZ
G
LB/OZ
Alternate Units (ALTUNT Parameter)
LB
KG
OZ
G
LB/OZ
ON
OFF
Specifies the decimal position, display divisions, and units used for the primary units. See
Level 3 submenu parameter descriptions.
Specifies which alternate units can be displayed by pressing the UNITS. See Level 3
submenu parameter descriptions.
NOTE: The LB/OZ setting is not a Legal-for-Trade setting.
Decimal point location. Specifies the location of the decimal point or dummy zeroes in the
primary unit display. Value should be consistent with local legal requirements.
If LB/OZ is selected as the primary unit, DECPNT selections are 88 88.8 and 888 88.
Display divisions. Selects the minimum division size for the primary units displayed weight.
Specifies primary units for displayed and printed weight. Values are: LB=pound;
KG=kilogram; OZ=ounce; G=gram; LB/OZ=pounds and ounces.
Alternate units. Determines which units are displayed when the UNITS button is pressed.
The value for the primary unit (selected on the PRIMAR parameter) is always ON.
NOTE: The LB/OZ setting is not a Legal-for-Trade setting.
16IQ plus 2100 Installation Manual
Table 3-3. Format Menu Parameters
Page 21
3.2.3Calibration Menu
WZERO
*CAL*
Display and edit
zero calibration
A/D count value
Display and edit
test weight value
WVAL
*CAL*
Display and edit
span calibration
A/D count value
WSPAN
*
CAL*
Press Enter to
remove offset from
zero and span
calibrations
REZERO
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIGFORMAT
See Section 4.0 on page 20 for calibration procedures.
Figure 3-8. Calibration Menu
CALIBR Menu
ParameterChoicesDescription
Level 2 submenus
WZERO—Display and edit the zero calibration A/D count value.
WVAL—Display and edit the test weight value.
WSPAN—Display and edit the span calibration A/D count value.
REZERO—Press ENTER to remove an offset value from the zero and span calibrations.
Use this parameter only after WZERO and WSPAN have been set. See Section 4.1 on
page 20 for more information about using this parameter.
Table 3-4. Calibration Menu Parameters
Configuration17
Page 22
3.2.4Serial Menu
9600
1200
2400
4800
BITSBAUD
8NONE
7EVEN
7ODD
SERIALCALIBRCONFIGFORMAT
TERMIN
CR/LF
CR
ECHO
OFF
ON
XXXXXXXXXXXXXX
PRNFRQ
DEMAND
AUTO1
AUTO2
STREAM
Figure 3-9. Serial Menu
SERIAL Menu
ParameterChoicesDescription
BAUD9600
1200
2400
4800
BITS8NONE
7EVEN
7ODD
TERMINCR/LF
CR
ECHOOFF
ON
PRNFRQDEMAND
AUTO1
AUTO2
STREAM
Baud rate. Selects the transmission speed for the serial port.
Selects number of data bits and parity of data transmitted from the serial port.
Termination character. Selects termination character for data sent from the serial port.
This command enables or disables echoing of the serial commands sent to the indicator.
Print frequency. Specifies when the indicator sends data to the serial port:
DEMAND: When a PRINT digital input is activated or the KPRINT EDP command received.
AUTO1:Output enabled by scale in motion; transmitted when scale returns to standstill.
AUTO2:Output enabled by scale at center of zero; transmitted when a positive scale weight
is at standstill.
STREAM:Continuous output. See Section 6.3 on page 26 for output format.
Weight data is sent to the serial port in displayed units except when other units are specified
using one of the XGx EDP commands (see Section 5.1.5 on page 24).
VERSDEFLTDIG IN
If DEMAND, AUTO1, or AUTO2, is selected for this parameter, serial output is formatted as
follows:
wwwwwww uu GROSS
where wwwwww is a 7-digit weight field (leading zeroes suppressed, including decimal point and
minus sign, if required), uu is the units designator (lb, kg, oz, or G).
If pounds/ounces (LB/OZ) is specified as the primary units, the serial output is formatted as
shown below:
xxx lb yyyy oz GROSS
where xxx is the three-digit pounds weight (including minus sign, if required) followed by lb and
a space; and yyyy is the four-digit ounce weight (including decimal point) followed by oz.
18IQ plus 2100 Installation Manual
Table 3-5. Serial Menu Parameters
Page 23
3.2.5Digital Input Menu
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIGFORMAT
DIGIN1
OFF
DIGIN2
ZERO
UNITS
PRINT
OFF
ZERO
UNITS
PRINT
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIGFORMAT
RESET
XXXXXXXXXXXXXX
VERSDEFLTDIG IN
SERIALCALIBRCONFIGFORMAT
Software
version
Figure 3-10. Digital Input Menu
DIG IN Menu
ParameterChoicesDescription
Level 2 submenus
DIGIN1
DIGIN2
OFF
ZERO
UNITS
PRINT
Specifies the function activated by digital inputs 1 and 2.
Table 3-6. Digital Input Menu Parameters
3.2.6Default Menu
The DEFLT menu is used to reset indicator configuration to the default values. There are no parameters
associated with the DEFLT menu. From the DEFLT menu, press
parameter. Press
ENTER again to reset the indicator to the default configuration values or press RIGHT (UNITS)
to cancel the indicator reset and return to the menu level above.
ENTER (ZERO) to move down to the RESET
NOTE: All load cell calibration settings are lost
when the RESET function is performed.
Figure 3-11. Default Menu
3.2.7Version Menu
The VERS menu is used to check the software version installed in the indicator. There are no parameters
associated with the VERS menu: when selected, the indicator displays the installed software version number.
Figure 3-12. Version Menu
Configuration19
Page 24
4.0Calibration
The IQ plus 210 can be calibrated using the front panel, EDP commands, or the Revolution™ configuration utility.
Each method consists of the following steps:
•Zero calibration
•Entering the test weight value
•Span calibration
•Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods.
SERIALCALIBRCONFIGFORMAT
WZERO
*CAL*
Display and edit
zero calibration
A/D count value
WVAL
Display and edit
test weight value
WSPAN
*CAL*
Display and edit
span calibration
A/D count value
Figure 4-1. Calibration (CALIBR) Menu
4.1Front Panel Calibration
To calibrate the indicator using the front panel, do the
following:
1. Place the indicator in setup mode (display
reads
CONFIG) and remove all weight from
the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.
2. Press the
display reads CALIBR (see Figure 4-1). Press
ENTER (ZERO) to go to zero calibration
WZERO).
(
3. With
calibrate zero. The indicator displays °CAL°
while calibration is in progress. When
complete, the A/D count for the zero
calibration is displayed. Press ENTER again to
save the zero calibration value and return to the
WZERO prompt or use the procedure shown in
Figure 4-2 to edit the value. When done, press
RIGHT to go to the WVAL prompt.
4. With
the test weight value. If the value is equal to the
test weight you are using, press ENTER again
to save the value and return to the WVAL
prompt; if the value is incorrect, use the
procedure shown in Figure 4-2 to edit the
value. When done, press RIGHT to go to the
WSPAN prompt.
RIGHT (UNITS) button until the
WZERO displayed, press ENTER to
WVAL displayed, press ENTER to show
XXXXXXXXXXXXXX
REZERO
*CAL*
Press Enter to
remove offset from
zero and span
calibrations
NOTE: When calibrating the indicator with
LB/OZ as the primary unit, enter the WVAL value
in ounces. For example, to calibrate a 5 lb scale
using LB/OZ, enter 80 (5 lb x 16 oz/lb) as the
WVAL value.
VERSDEFLTDIG IN
5. With WSPAN displayed, place test weights on
the scale then press ENTER to calibrate span.
The indicator displays °CAL° while calibration
is in progress. When complete, the A/D count
for the span calibration is displayed. Press
ENTER again to save the span calibration value
and return to the WSPAN prompt or use the
procedure shown in Figure 4-2 to edit the
value. When done, press RIGHT to go to the
REZERO prompt.
00 0 0 0 0
To edit numeric values, press the
button to select the leftmost digit. Each digit flashes
when selected: Press
increment the value of the selected digit; press the
RIGHT
Once the last digit has been edited, press
again and the digit stops flashing. Press
to save the value entered and return to the menu level
above, or press
starting with the leftmost digit.
button to move right to the next digit.
RIGHT
UNITS (RIGHT)
ZERO (ENTER)
RIGHT
ENTER
to edit the value again,
to
20IQ plus 2100 Installation Manual
Figure 4-2. Editing Procedure for Numeric Values
Page 25
6. The rezero function is used to remove a
calibration offset when hooks or chains are
used to hang the test weights.
•If no other apparatus was used to hang the
test weights during calibration, remove the
test weights and press
RIGHT to return to
the CALIBR menu.
•If hooks or chains were used during
calibration, remove these and the test
weights from the scale. With all weight
removed, press
ENTER to rezero the scale.
This function adjusts the zero and span
calibration values. The indicator displays
°CAL° while the zero and span calibrations
are adjusted. When complete, the adjusted
A/D count for the zero calibration is
displayed. Press ENTER again to save the
rezero value and return to the REZERO
prompt or use the procedure shown in
Figure 4-2 to edit the value. When done,
press RIGHT to return to the CALIBR
menu.
7. Press the setup switch to exit setup mode.
4.2EDP Command Calibration
To calibrate the indicator using EDP commands, the
indicator EDP port must be connected to a terminal or
personal computer. (See Section 2.3.3 on page 5 for
EDP port pin assignments; see Section 5.0 on page 22
for more information about using EDP commands.)
Once the indicator is connected to the sending device,
do the following:
1. Place the indicator in setup mode (display
CONFIG) and remove all weight from
reads
the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.
2. Send the WZERO EDP command to calibrate
zero. The indicator displays
calibration is in progress.
3. Send the WVAL command to enter the test
weight value in the following format:
WVAL=nnnnnn<CR>
°CAL° while
4.3Revolution™ Calibration
To calibrate the indicator using Revolution, the
indicator must be in setup mode with the EDP port
connected to a PC running the Revolution configuration
utility. Use the following procedure to calibrate the
indicator:
1. Select
main menu.
2. On the Weight Calibration display, select the
indicator model (
port then click OK.
3. Revolution uploads calibration data from the
indicator then presents the information in a
display like that shown in Figure 4-3.
4. Enter the
span calibration then click OK.
5. The Zero Calibration dialog box prompts you
to remove all weight from the scale. Clear the
scale and click
NOTE: If your test weights require hooks or
chains, place the hooks or chains on the scale
for zero calibration.
6. When zero calibration is complete, the Span
Calibration dialog box prompts you to place
test weights on the scale for span calibration.
Place tests weights on the scale then click OK.
7. The Rezero dialog box prompts you to remove
all weight, from the scale (including hooks or
chains). Remove the weights then click
8. When calibration is complete, the
Settings fields of the Weight Calibration
display are filled in. Click Exit to save the new
values and return to the Rev olution main menu;
to restore the previous calibration values, click
Restore Settings.
Calibrate Indicator from the Revolution
IQ+210) and communications
Value of Test Weight to be used for
OK to begin zero calibration.
OK.
New
4. Place test weights on scale equal to the
specified WVAL.
5. Send the WSPAN EDP command to calibrate
span. The indicator displays
°CAL° while
calibration is in progress.
6. To remove an offset value, clear all weight
from the scale, including hooks or chains used
to hang test weights, then send the REZERO
EDP command. The indicator displays
°CAL°
while the zero and span calibrations are
adjusted.
7. Send the KEXIT EDP command to exit setup
mode.
Figure 4-3. Revolution Calibration Display
Calibration21
Page 26
5.0EDP Commands
The IQ plus 210 indicator can be controlled by a
personal computer or remote keyboard connected to the
indicator serial port. Control is provided by a set of
EDP commands that can simulate front panel key press
functions, display and change setup parameters, and
perform reporting functions. The EDP port provides the
capability to print configuration data or to save that data
to an attached personal computer. This section
describes the EDP command set and procedures for
saving and transferring data using the serial port.
5.1The EDP Command Set
The EDP command set can be divided into five groups:
key press commands, reporting commands, the
RESETCONFIGURATION special function
command, parameter setting commands, and transmit
weight data commands.
When the indicator processes an EDP command, it
responds with the message
verifies that the command was received and has been
executed. If the command is unrecognized or cannot be
executed, the indicator responds with
The following sections list the commands and
command syntax used for each of these groups.
5.1.1Key Press Commands
Key press EDP commands (see Table 5-1) simulate
pressing the buttons on the front panel of the indicator.
CommandFunction
KZEROPress the ZERO button
KUNITSPress the UNITS button
KPRIMDisplay primary units
KPRINTSend demand data to serial port
KRIGHTARROWIn setup mode, move right in the menu
(press the UNITS button)
KENTERIn setup mode, press the ENTER
(ZERO) button
KEXITIn setup mode only, exit setup mode
Table 5-1. EDP Key Press Commands
OK. The OK response
??.
5.1.2Reporting Commands
Reporting commands (see Table 5-2) send specific
information to the EDP port. These commands can be
used in both setup mode and normal mode.
CommandFunction
DUMPALLList all parameter values
VERSION
PWrite current displayed weight. See
ZZWrite current weight and annunciator
RSReset the indicator
SWrite one frame of stream format
5.1.3The RESETCONFIGURATION Command
Write IQ plus 210 software version
Section 6.2 on page 26 for more
information.
status. See Section 6.2 on page 26 for
more information.
Table 5-2. EDP Reporting Commands
The RESETCONFIGURATION command can be used
to restore all configuration parameters to their default
values.
This command is equivalent to using the RESET
function on the DEFLT menu.
NOTE: All load cell calibration settings are lost when the
RESETCONFIGURATION command is run.
5.1.4Parameter Setting Commands
Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 5-3 through 5-8).
Current configuration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:
command<ENTER>
Most parameter values only can be changed in setup
mode. Use the following command syntax when
changing parameter values:
command=value<ENTER>
where value is a number or a parameter value. Use no
spaces before or after the equal (=) sign. If you type an
incorrect command or value, the display reads
??.
Changes to the parameters are saved as they are entered
but typically do not take effect until you exit setup
mode.
For example, to set the motion band parameter to 5,
type the following:
EDP.BAUDSerial port baud rate1200, 2400, 4800, 9600
EDP.BITSSerial port data bits/parity8NONE, 7EVEN, 7ODD
EDP.TERMINSerial port termination characterCR/LF, CR
EDP.ECHOSerial port echo commandOFF, ON
PRNFREQPrint frequencyDEMAND, AUTO1, AUTO2, STREAM
Table 5-6. SERIAL EDP Commands
CommandDescriptionValues
DIGIN1
DIGIN2
Digital input functionOFF, ZERO, UNITS, PRINT
Table 5-7. DIG IN EDP Commands
EDP Commands23
Page 28
5.1.5Normal Mode Commands
The serial transmit weight data commands (see Table 5-8) transmit data to the serial port on demand. The transmit
weight data commands are valid only in normal operating mode.
CommandDescriptionResponse Format
SXStart serial streamingOK or ??
EXStop serial streamingOK or ??
RSReset system—
XGTransmit gross weight in displayed unitsnnnnnn UU
where nnnnnn is the weight value, UU is the units.XGLTransmit gross weight in pounds
XGKTransmit gross weight in kilograms
XGOTransmit gross weight in ounces
XGGTransmit gross weight in grams
XGCTransmit gross weight in pounds and ounces
XEQuery system error conditionsnnnnn nnnnn
See Section 6.1.2 on page 25 for detailed information about
the XE command response format.
Table 5-8. Normal Mode EDP Commands
5.2Saving and Transferring Data
Connecting a personal computer to the IQ plus 210
EDP port allows you to save indicator configuration
data to the PC or to download configuration data from
the PC to an indicator. The following sections describe
the procedures for these save and transfer operations.
5.2.1Saving Indicator Data to a Personal Computer
Configuration data can be saved to a personal computer
connected to the EDP port. The PC must be running a
communications program such as PROCOMMPLUS
See Section 2.3.3 on page 5 for information about serial
communications wiring and EDP port pin assignments.
When configuring the indicator, ensure that the values
set for the BAUD and BITS parameters on the SERIAL
menu match the baud rate, bits, and parity settings
configured for the serial port on the PC.
To save all configuration data, send the DUMPALL
EDP command to the indicator. The IQ plus 210
responds by sending all configuration parameters to the
PC as ASCII-formatted te xt.
5.2.2Downloading Configuration Data from PC to
Indicator
Configuration data saved on a PC or floppy disk can be
downloaded from the PC to an indicator. This
procedure is useful when a number of indicators with
similar configurations are set up or when an indicator is
replaced.
To download configuration data, connect the PC to the
EDP port as described in Section 5.2.1. Place the
®
.
indicator in setup mode and use the PC
communications software to send the saved
configuration data to the indicator. When transfer is
complete, calibrate the indicator as described in
Section 4.0 on page 20.
NOTES:
•Calibration settings are included in the
configuration data downloaded to the indicator.
If the receiving indicator is a direct replacement
for another IQ plus 210 and the attached scale is
not changed, recalibration is not required.
•When downloading configurations that include
changed serial communications settings, edit the
data file to place the serial communications
changes at the end of the file. Communication
between the PC and indicator will be lost once
the indicator receives settings for baud rate
(BAUD parameter) or data bits and parity (BITS
parameter) that do not match those configured
for the PC.
24IQ plus 2100 Installation Manual
Page 29
6.0Appendix
6.1Error Messages
The IQ plus 210 indicator provides a number of error messages. When an error occurs, the message is sho wn on the
indicator LED display . Error conditions can also be check ed remotely by using the XE EDP command as described
in Section 6.1.2.
Error MessageDescriptionSolution
E A/DA/D physical errorCall Rice Lake Weighing Systems (RLWS) Service.
EEEROMEEPROM physical error
EVIREEVirgin EEPROMUse the DEFLT menu to restore defaults, then recalibrate load cells.
EPCKSMParameter checksum error
EACKSMA/D calibration checksum errorA/D converter requires recalibration. Call RLWS Service.
ELCKSMLoad cell calibration checksum errorRecalibrate load cells.
INOVFLInternal overflow errorRemove weight from scale. If error recurs, check configuration, check
scale for damage.
OVERFLOverflow errorWeight value too large to be displayed.
------
______A/D underrangeA/D reading < –4 mV. Check scale for binding or damage.
Gross > overload limitGross value exceeds overload limit. Check configuration.
Table 6-1. IQ plus 210 Error Messages
6.1.1Displayed Error Messages
The IQ plus 210 provides a number of front panel error
messages to assist in problem diagnosis. Table 6-1 lists
these messages and their meanings.
6.1.2Using the XE EDP Command
The XE EDP command can be used to remotely query
the IQ plus 210 for the error conditions shown on the
front panel. The XE command returns two 5-digit
numbers in the format:
xxxxx yyyyy
where xxxxx contains a decimal representation of any
existing error conditions as described in Table 6-2.
If more than one error condition exists, the number
returned is the sum of the values representing the error
conditions. For example, if the XE command returns
the number 528, this value represents the sum of an
A/D physical error (512) and an A/D calibration
checksum error (16).
The second number returned (
yyyyy) uses the same bit
assignments as shown in Table 6-2 to indicate whether
the test for the error condition was run. For example,
the value
yyyyy = 51807 represents the decimal
equivalent of the binary value 1100 1110 0111 1111.
Using the bit assignments in Table 6-2, this value
indicates all tests were run.
Two EDP commands, P and ZZ, can be used to provide
status about the indicator. These commands are
described in the following sections.
6.2.1Using the P EDP Command
The P EDP command returns the current displayed
weight value to the EDP port. If the indicator is in an
underrange or overload condition, the weight value is
replaced with
6.2.2Using the ZZ EDP Command
The ZZ EDP command can be used to remotely query
which annunciators are currently displayed on the
indicator front panel. The ZZ command returns the
currently displayed weight and a decimal number
representing the LED annunciators currently lit. The
format of the returned data is:
wwwwww zzz
where wwwwww is the current displayed weight, zzz is the
annunciator status value (see Table 6-3). If more than
one annunciator is lit, the second number returned is the
sum of the values representing the activ e annunciators.
&&&&&& (overload) or :::::: (underrange).
For example, if the annunciator status v alue returned on
the ZZ command is 136, the center of zero and lb
annunciators are lit: 136 represents the sum of the
values for the center of zero annunciator (128) and the
lb annunciator (8).
Decimal
Value
1Reserved
2Negative value
4oz
8lb
16g
32kg
64Motion
128Center of zero
Table 6-3. Status Codes Returned on the ZZ Command
Annunciator
6.3Continuous Output (Stream) Format
Figure 6-1 shows the continuous output format sent to the IQ plus 210 EDP or printer port when the STREAM
parameter (SERIAL menu) is set to either EDP or PRN.
Weight data: 7 digits, right-justified, with
decimal point, leading zero suppression.
Overload =
Underrange =
Display overflow = OVERFL
<POL> <wwwwwww> <UNIT>
^^^^^^^
] ] ] ] ] ] ]
Figure 6-1. Continuous Output Data Format
<G>
G = Gross
L = pounds
K = kilogram
G = grams
O = ounces
<space> = lb/oz
<S>
Status:
<space> = valid
I = Invalid
M = In motion
O = Over/under range
<TERM>
<CR> <LF>
or <CR>
ASCII 13, 10
(decimal)
26IQ plus 2100 Installation Manual
Page 31
6.4Specifications
6.4.1IQ plus 210 Indicator
Power
Line Voltages115 or 230 VAC
Frequency50 or 60 Hz
Power Consumption70 mA @ 115 VAC (8W)
35 mA @ 230 VAC (8W)
Fusing
115 VAC2 x 200 mA TR5 subminiature fuses
Wickmann Time-Lag 19374 Series
UL Listed, CSA Certified
230 VAC2 x 100 mA TR5 subminiature fuses
Wickmann Time-Lag 19372 Series
UL Recognized, VDE Approved
Analog Specifications
Full Scale Input Signal Up to 35 mV
Excitation Voltage10 ± 0.25 VDC,
4 x 350Ω load cells
Analog Signal
Input Range0.6 mV/V – 4.5 mV/V
Analog Signal
Sensitivity0.3 µV/graduation minimum,
Input Impedance200 MΩ, typical
Display Resolution10 000 dd
Input Sensitivity155 nV per internal count
System LinearityWithin 0.01% of full scale
Calibration MethodSoftware, constants stored in
Common Mode
Voltage–1.05 to +0.95 V, referred to earth
RFI ProtectionSignal, excitation, and sense lines
Digital Specifications
Digital Inputs2 inputs, TTL or switch closure,
Digital FilterSoftware selectable
Serial Communications
Serial PortFull duplex RS-232; 9600, 4800,
Operator Interface
Display6-digit LED display. 7-segment, .8 in
LED annunciatorsNegative value, center of zero, motion,
Keypad2 piezo switches
1.5 µV/grad recommended
EEPROM
protected by capacitor bypass
active-low
2400, 1200 bps; 7 or 8 data bits;
even, odd, or no parity
(20 mm) digits
kg, g, lb, oz
Enclosure
Enclosure Dimensions 9.5 in x 6 in x 2.75 in
24 cm x 15 cm x 7 cm
Weight6.1 lb (2.8 Kg)
Rating/MaterialNEMA 4X/IP66, stainless steel
Approvals
O
C
N
NTEP
F
L
E
A
R
N
E
O
I
N
T
C
A
E
N
•
•
O
N
S
CoC Number00-045
E
W
R
E
U
I
S
G
A
H
E
T
S
M
A
D
N
Accuracy Class III/III L
C
LISTED
U
L
®
UL
US
File NumberE151461
Measurement Canada Approved
ApprovalAM-5374
Accuracy ClassIII/III HD
6.4.2RL2100 Bench Scale
Load Cells
RL1042 aluminum load cell (standard)
RL1380 stainless steel load cell (optional)
Rated Output2.0 mV/V
Maximum Overload150%
Cable10 ft (3.1 m)
Output Impedance350Ω, nominal
Operating Temperature–10 to +40°C (legal);
–10 to +50°C (industrial)
Storage Temperature –25 to +70°C
Dimensions & Capacities
Base DimensionCapacity
10 in x 10 in (254 mm x 254 mm)5 lb (2.5 Kg)
10 lb (5 Kg)
25 lb (10 Kg)
12 in x 12 in (305 mm x 305 mm)50 lb (25 Kg)
100 lb (50 Kg)
Approvals
O
C
N
NTEP
F
L
E
A
R
N
E
O
I
N
T
C
A
E
N
•
•
O
N
S
CoC Number95-072A2
E
W
R
E
U
I
S
G
A
H
E
T
S
M
A
D
N
Accuracy ClassIII
Measurement Canada Approved
ApprovalAM-5082 Rev 1
Environmental
Operating Temperature–10 to +40°C (legal);
–10 to +50°C (industrial)
Storage Temperature –25 to +70°C
Humidity0–95% relative humidity
Altitude2000 m (6500 ft) maximum
Appendix27
Page 32
IQ plus 2100 Limited Warranty
Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a
Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the
Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials
and workmanship for two years.
RLWS warrants that the equipment sold hereunder will conform to the current written specifications authorized by
RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipment fails
to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within the warranty
period subject to the following conditions:
•Upon discovery by Buyer of such nonconformity, RLWS will be given prompt written notice with a detailed
explanation of the alleged deficiencies.
•Individual electronic components returned to RLWS for warranty purposes must be packaged to prevent
electrostatic discharge (ESD) damage in shipment. Packaging requirements are listed in a publication, ProtectingYour Components From Static Damage in Shipment, available from RLWS Equipment Return Department.
•Examination of such equipment by RL WS confirms that the nonconformity actually e xists, and was not caused by
accident, misuse, neglect, alteration, improper installation, improper repair or improper testing; RLWS shall be
the sole judge of all alleged non-conformities.
•Such equipment has not been modified, altered, or changed by any person other than RLWS or its duly authorized
repair agents.
•RLWS will have a reasonable time to repair or replace the defective equipment. Buyer is responsible for shipping
charges both ways.
•In no event will RLWS be responsible for tra vel time or on-location repairs, including assembly or disassembly of
equipment, nor will RLWS be liable for the cost of any repairs made by others.