Operating Instructions for Quick Drying Machine
Type TG200
Retsch GmbH, 42781 Haan, Retsch-Allee 1-5, Germany, 2/25/2013 0003
Page 2
Preliminary notes on these
Injury to persons or
Damage to the machine
Follow the operating instructions
the Retsch agent in your country
your supplier
Retsch GmbH directly
operating instructions
These operating instructions for the Quick Drying Machine, type TG 200, give all the necessary information for the areas mentioned in the contents. The instructions given here are aimed at the target group(s) defined for each area, in order to ensure safe handling of the TG200 and its use for the intended purpose. A knowledge of the relevant section in the instructions is the prerequisite for safe handling and use for the intended purpose by the particular target group(s). This technical documentation is a reference work and contains learning instructions. The individual sections are complete in themselves. These operating instructions do not include instructions for repairs. If repairs are necessary please contact your supplier or Retsch GmbH directly.
Warning instructions
Warnings are indicated with the following signs:
Repairs
These operating instructions do not include instructions for repairs. For your own safety repairs should be carried out only by Retsch GmbH, an authorised agent or service technicians.
In this case please inform:
Your service address is as follows:
2/25/2013 Retsch GmbH 2 0002
Page 3
Preliminary notes on these operating instructions ....... 2
Warning instructions ................................................. 2
Repairs ................................................................... 2
Safety .......................................................................... 4
Safety instructions .................................................... 4
Safety instructions, summary, part 1 .......................... 5
Safety regulations, summary, part 2 ........................... 6
Confirmation ................................................................ 7
Technical data .............................................................. 8
Use according to the intended purpose ........................ 8
Emission ................................................................. 9
Heating power ......................................................... 9
Fan power ............................................................... 9
Rated power ............................................................ 9
Systems of protection ............................................... 9
Machine dimensions .................................................. 9
Required floor space ................................................. 9
Transport and installation ............................................ 10
Packing material ....................................................... 10
Transport ................................................................ 10
Temperature variations ............................................. 10
Intermediate storage ................................................ 10
Parameters for the place of installation ....................... 11
Installation .............................................................. 11
Electrical connection ................................................. 11
Operation ..................................................................... 12
Operating elements and operation .............................. 12
Operating the TG200 ................................................ 13
Switching on / off ..................................................... 13
Starting – interruption - stopping ............................... 13
Setting the time ....................................................... 14
Setting the heating power temp °C/°F ........................ 14
Triggering the bimetallic element switch in the heater ... 15
Setting the air flow ................................................... 15
Interval – continuous operation .................................. 15
Drying mode ............................................................ 16
Storage and recalling drying parameters ..................... 16
Signal tone for end of drying process .......................... 16
Inserting the drying vessels ......................................... 17
6 litre drying vessels of glass or stainless steel ............ 17
Verwendung der weicheren Dichtung im Spanndeckel .. 17 Use and installation of the 63µm disk as additional filter 17
Inserting the clamping lid with filter element ............... 18
Attachment with 3 x 300 ml glass vessels .................... 18
Working instructions .................................................... 19
General ................................................................... 19
Mode of operation of the TG200 ................................. 19
Attainable temperatures at the measuring point ........... 20
Operating conditions 1: ............................................. 20
Operating conditions 2: ............................................. 20
Operating conditions 3: ............................................. 20
General ........................................................................ 21
Cleaning .................................................................. 21
Maintenance ............................................................ 21
Replacing fuses ........................................................ 22
Accessories .............................................................. 22
Wearing parts .......................................................... 23
Fault codes .............................................................. 23
Copyright ................................................................ 23
Alterations ............................................................... 23
Appendix ............................................... following pages
2/25/2013 Retsch GmbH 3 0002
Page 4
Safety
We exclude any liability for claims for damages in any form whatsoever for injury to persons or damage to the machine caused through non-observance of the following safety instructions.
The TG200 is an ultra-modern, highly efficient product from Retsch GmbH, and corresponds to state-of-the-art technology. When using the machine according to the intended purpose and with a knowledge of the technical documentation described here it is completely reliable and safe to operate.
Safety instructions
As the operating authority you must ensure that all persons authorised to work on the TG 200:
are aware of, and have understood all the safety
regulations,
are aware of all the handling instructions and regulations
for the target group of relevance to them before starting work,
have access to the technical documentation for this
machine at all times without problems.
Before beginning work on the TG200 new personnel must
have been made familiar with safe handling of the machine and its operation according to the intended purpose, either by verbal instruction from a competent person and/or through reading this technical documentation.
Improper operation can cause injury to persons and
damage to the machine. You are responsible for your own safety and that of your employees.
Make sure that no unauthorised persons have access to
the TG200.
For your own protection have knowledge of the instructions for operating the TG200 confirmed by your employees. A draft of a suitable form is given on the page following the section on safety.
2/25/2013 Retsch GmbH 4 0002
Page 5
Safety instructions, summary, part 1
We exclude any liability for claims for damages in any form whatsoever for injury to persons or damage to the machine caused through non-observance of the following safety instructions.
Do not make any alterations to the machine and use only spare parts and accessories approved by Retsch.
Otherwise the Declaration of Conformity with European directives made by Retsch will lose its validity. Furthermore this will also lead to the loss of any guarantee claims.
Do not dry explosives or easily flammable materials, or materials with a flashpoint below 180°C in the TG200. Risk of fire and explosion.
Do not operate the TG200 in the vicinity of easily flammable materials. Risk of fire and explosion.
You must check to what extent toxic danger exists through the charged material. Risk to health through vapours or residual dusts which are not retained by the filters.
Do not operate the TG200 without supervision. Risk of fire and explosion.
Materials should be dried only in the appropriate time, at the appropriate temperature. Take into consideration the melting point and temperature sensitivity of the material. The material to be dried can be destroyed through overheating.
Please keep the packing material for the duration of the guarantee period, since, in case of a complaint and return of the machine in unsuitable packing material your guarantee claim will be at risk.
The TG200 must not be jolted, shaken or thrown during transport. Otherwise the electronic and mechanical components may be damaged.
If subjected to high temperature variations (e.g. during transport by aircraft) the TG200 must be protected against condensed water. Otherwise the electronic components may be damaged.
If the delivery is incomplete and/or transport damaged is found, you must inform the transporter and Retsch GmbH immediately (within 24 hrs). Later complaints may possibly be no longer considered.
If the temperature drops below, or exceeds the ambient temperature the electrical and mechanical components may be damaged; the performance data are then altered to an unknown extent.
At high atmospheric humidity the electrical and mechanical components may be damaged and the performance data are altered to an unknown extent.
Maintain a safety distance of 100 mm all round. The hot air outlet can ignite easily flammable objects. Risk of fire and explosion.
If the values on the rating plate are not observed the electrical and mechanical components may be damaged.
Safety instructions
Use according to the intended purpose
Packing materials
Transport
Temperature variations
Supply schedule
Ambient temperature:
Atmospheric humidity:
Installation
Electrical connection / connecting the power supply
2/25/2013 Retsch GmbH 5 0002
Page 6
Safety regulations, summary, part 2
Do not dry explosives, easily flammable materials or materials with a flashpoint below 180°C in the TG200. Risk of fire and explosion.
Do not operate the TG200 in the vicinity of easily flammable materials. Risk of fire and explosion.
You must check to what extent toxic danger exists through the charged material. Risk to health through vapours or residual dusts which are not retained by the filters.
The bimetallic element switch is triggered when a temperature > 150°C/302°F has been reached. This can result from too low fan power and at the same time too high stagnation pressure caused by the material to be dried. Before restarting the machine the temperature should be reduced by 5 – 10°C. Otherwise the heater may be damaged.
We recommend that protective gloves are worn when removing the vessels and the clamping lid at the end of the drying process. Risk of burns.
We recommend that protective gloves are worn when removing the vessels and clamping lid at the end of the drying process. Risk of burns.
When using the clamping lid with filter element the TG200 may be operated only up to a temperature of max. +90°C/+194°F.
Otherwise the filter fabric will burn. Risk of fire.
Do not clean the TG200 under running water. Danger to life through current surge.
If the filter element and filter bags are contaminated too much the efficiency of the TG200 will be reduced to an unknown extent.
These operating instructions do not include any repair instructions. For your own safety repairs should be carried out only by Retsch GmbH, an authorised agent or service technician.
Setting the heating power – temp °C/°F
Triggering the bimetallic element switch in the heater, automatic switch off when overheated
Inserting the drying vessels
Inserting the clamping lid with filter element
Cleaning / maintenance
Wearing parts
2/25/2013 Retsch GmbH 6 0002
Page 7
Confirmation
I have noted the sections on Notes on Operating
Instructions and safety.
__________________________________
Signature of operator
__________________________________
Signature of service technician
2/25/2013 Retsch GmbH 7 0002
Page 8
Technical data
Do not make any alterations to the machine and use only spare parts and accessories approved by Retsch.
Otherwise the Declaration of Conformity with European directives made by Retsch will lose its validity.
Furthermore this will also lead to the loss of any guarantee claims.
Do not dry explosives, easily flammable materials or materials with a f lashspoint below 180°C in the TG200. Risk of fire and explosion.
Do not operate the TG200 in the vicinity of easily flammable materials.
Risk of fire and explosion
You must check to what extent toxic danger exists through the charged material.
Risk to health through vapours or residual dusts which are not retained by the filters.
Do not operate the TG200 without supervision.
Risk of fire and explosion.
Materials should be dried only in the appropriate time, at the appropriate temperature. Take into consideration the melting point and temperature sensitivity of the material.
The material to be dried can be destroyed through overheating.
Machine type designation: TG200
Use according to the intended purpose
Depending on the specific properties of the material to be dried, materials with a moisture content of max. approx. 80% can be dried. The size of the batch is normally 1000 ml. In some cases, e.g. with lightweight, high volume materials, the complete vessel volume can be utilized. The minimum grain size should not be less than 63 µm. The TG200 is not designed as a production machine or for continuous operation but as a laboratory machine, intended for one shift operation of 8 hours.
2/25/2013 Retsch GmbH 8 0002
Page 9
Emission
Noise measurement according to DIN 45635-031-01-KL3 The noise characteristic values depend on the set fan power and the type of material to be dried.
Example:
Acoustic power level LWA = 88 dB(A) Emission value related to working area L
pAeq
= 75 dB(A)
Operating conditions:
Material to be dried = clay, filling level 225 mm Max. heating power, max. fan power
Noise characteristic values :
Heating power
The heating power depends on the type and quantity of material to be dried and the set fan power. The nominal value is 2 x 1000 W. Basic heating takes place through the power output of the radial fan, depending on the air flow.
Fan power
The fan power depends on the type and quantity of the material to be dried and the accessories used. The value is approx. 185 m³/hr at no-load operation without material to be dried.
Rated power
At: 240V = 2990 VA 230V = 2865 VA 220V = 2740 VA 200V = 2490 VA
Systems of protection
IP43 or IP20 without drying vessel and lid in position
Machine dimensions
Height: approx. 1000 mm, width: 400 mm, depth: 480 mm Weight: approx. 21 kg without drying vessel
Required floor space
Width 400 mm x depth 540 mm Safety distance of 100 mm from the rear panel is necessary.
2/25/2013 Retsch GmbH 9 0002
Page 10
Transport and installation
Please keep the packing material for the duration of the guarantee period, since, in case of a complaint and return of the machine in unsuitable packing material your guarantee will be at risk.
The TG200 must not be jolted, shaken or thrown during transport. Otherwise the electronic and mechanical components may be damaged.
If subjected to high temperature variations (e.g. during transport by aircraft) the TG200 must be protected against condensed water. Otherwise the electronic components may be damaged.
Packing material
The packing material is adapted to the transport route and conforms to generally applicable packing directives.
Transport
Temperature variations
Intermediate storage
Make sure that the TG200 is also kept dry during intermediate storage.
2/25/2013 Retsch GmbH 10 0002
Page 11
Parameters for the place of installation
If the temperature drops below or exceeds the ambient temperature the electrical and mechanical components may be damaged. The performance data are altered to an unknown extent.
At high atmospheric humidity the electrical and mechanical components may be damaged. The performance data are altered to an unknown extent.
Installation
Install the TG200 only on a robust laboratory bench.
Maintain a safety distance of 100 mm all round.
The hot air outlet can ignite easily flammable objects.
Risk of fire and explosion.
If the values on the rating plate are not observed the electrical and mechanical components may be damaged.
Ambient temperature:
5°C to 40°C
Atmospheric humidity:
Maximum relative humidity 80% at temperatures up to 31°C, linear decrease up to 50% relative humidity at 40°C.
Installation height:
Max. 2000 m above sea level.
Electrical connection
The voltage and frequency of the TG200 is given on the
rating plate.
Ensure that the values agree with the existing mains
supply.
Connect the TG200 to the mains supply using the
connection cable supplied with the machine.
An external fuse of min. 20A should be installed when
connecting the power supply cable to the mains, in accordance with the regulations at the place of installation.
2/25/2013 Retsch GmbH 11 0002
Page 12
Operation
A Cold machine plug combination with on-off
switch
B Power supply cable
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
1
Key for recalling preselected drying parameters P1-P9 and on
2
The selected programme P1-P9 / on is displayed here.
3
Key for activating the setting mode for the preselected programme locations P1-P9
4
This key reduces the temperature 000-150°C or 032-302°F For switching over from °C to °F or vice-versa press keys “4+6” simultaneously.
5
Display shows preselected temperature and, after starting the TG200, the actual measured temperature.
6
This key increases the temperature 000-150°C or 032-302°F For switching over from °C to °F or vice vice-versa press keys “4+6” simultaneously.
7
This key reduces the fan power, 99 – 10
8
Display shows preselected fan power 10 – 99
9
This key increases the fan power, 10 – 99
10
This key switches ON interval operation, left LED lights up. 10 sec pause-10 sec operation
11
This key switches OFF interval operation, right LED lights up
12
This key reduces the drying period, 99 – 01min. or - - continuous operation
13
Display shows the preselected drying period 01 – 99min. or - - continuous operation
14
This key increases the drying period, 01 – 99min. or - - continuous operation
15
START key, starts the drying process and the green LED lights up
Press START key twice for quick start with max. heating power + max. fan power; green LED lights up.
16
1x STOP key, heater is switched off Red LED over key 16 flashes 2x STOP key, stops and ends the drying process and the red LED lights up.
Operating elements and operation
2/25/2013 Retsch GmbH 12 0002
Page 13
Operating the TG200
E
The main switch E is located on the right hand side of the rear panel of the TG200
Switch on the main switch ON lights up in the display drying mode 000 lights up in the display temp °C/°F 10 lights up in the display air flow LED interval off lights up
Two dashes (- -) light up in the display time The values appear at the first start. If the values are
changed they are retained at the end of the programme and appear when switching on again and pressing the START key.
The TG200 is now ready for operation, without interval, for continuous operation without heating and with minimum fan power.
16a
15 16
Starting:
Press START / key 15
Green LED over key 15 lights up TG200 starts with specified values. Red LED over key16a flashes as long as a temperature has not
been specified.
Specify temperature, red LED is extinguished, green LED lights
up.
During the drying period the temperature is kept constant within a tolerance. This means that the heater is switched off when the specified temperature has been reached and is switched on again when the temperature drops below the specified value. A temperature difference of +/- 2°C or +/-50°F is possible here.
Quick start with full power:
Press START / key 15 twice
Green LED over key 15 lights up
TG200 starts with full heating power and fan power for an
indefinite running period
The measured temperature is displayed
Stopping:
Press STOP / key 16 – heater is switched off Red LED over key 16 flashes Press STOP / key 16 a second time – everything is switched off Red LED over key 16 lights up The complete display except on is extinguished Activate LED displays = press key 15 once New values can be entered Press START / key 15 a second time
Function is carried out as with Starting
Switching on / off
Starting – interruption - stopping
2/25/2013 Retsch GmbH 13 0002
Page 14
Setting the time
12 13 14
When the TG200 is switched on the drying period is still indefinite = 2 dashes in display 13.
Switch on time 1 – 99 min. or - - continuous operation:
Key 12 reduces the time down to 1 min.
If the time drops below this value two dashes appear = indefinite time
Key 14 increases the time up to 99 min. If this time is exceeded two dashes appear = indefinite time
4 5 6
When the TG200 is switched on the heater is switched off. 000°C/032°F is shown in the display 5.
This is indicated by flashing of the red LED over the STOP key. The temperature is measured in °C or °F and shown in the display
5.
The temperature display can be changed over from centigrade to Fahrenheit with keys 4+6. Not during the
performance possibly.
Press keys 4+6 simultaneously An F for Fahrenheit appears briefly in the display 5 Press keys 4+6 simultaneously A C for centigrade appears briefly in the display 5
Switching on the heater 000 – 150°C / 000 – 302°F:
Key 4 reduces the temperature down to 000°C/032°F. Temperature 000°C/032°F = heater is switched off, red LED
over START flashes.
Key 6 increases the temperature to 150°C/302°F.
As soon as the temperature is increased with key 6 the heater switches on, the red LED over STOP is extinguished and the green LED lights up. The attainable heating power depends on the material to be dried, the filling quantity and the fan power air flow. This can only be determined empirically.
Do not dry explosives, easily flammable materials or materials with a flashpoint below 180°C in the TG200. Risk of fire and explosion.
Do not operate the TG200 in the vicinity of easily flammable materials.
Risk of fire and explosion.
You must check to what extent toxic danger exists through the charged material.
Risk to health through vapours or residual dusts which are not retained by the filters.
Setting the heating power temp °C/°F
2/25/2013 Retsch GmbH 14 0002
Page 15
Triggering the bimetallic element switch in the
2 5 8 16
Display 2 shows fault code F2 Temperature is reduced, display 5 Fan power is set to max. 99, display 8 Red LED over STOP key 16 lights up
The cooling cycle is started.
The bimetallic element switch is triggered when a temperature > 150°C/302°F has been reached. This can result from too low fan power and at the same time too high stagnation pressure caused by the material to be dried. Before restarting the machine the temperature should be reduced by 5 – 10°C.
Otherwise the heater may be damaged.
12 13 14
Reducing:
Key 12 reduces the fan power down to the minimum power step,
10 is indicated in display 13.
Keep key 12 pressed; after 5 sec. quicker adjustment is possible
Increasing:
Key 14 increases the fan power up to the maximum power step,
99 is indicated in display 13.
Keep key 14 pressed; after 5 sec. quicker adjustment is possible
It is not possible for the value to drop below 10 or to exceed 99.
10 11
At the time of switching on the TG200 interval operation is
switched off. Not during the performance possibly. When Interval is
switched on the air flow can no longer be changed.
Switching on interval:
Press key 10 LED over key 10 lights up Pause for 10 sec with fan switched off; heater is switched off Operation for 10 sec with reselected fan power; heater is
switched on again
Switching off interval:
Press key 11
LED over key11 lights up
heater
The heater is fitted with a bimetallic element switch. If a fault occurs this switches off the heater at 160°C/320°F +/­6%. The fan power is immediately set to maximum and the heater switched off. After a cooling cycle has taken place the TG200 must be restarted.
Setting the air flow
Interval – continuous operation
2/25/2013 Retsch GmbH 15 0002
Page 16
Drying mode
2 5 8 10 11 13
You can store, change or overwright the drying parameters temp – air flow – interval and time at any time in the programme
positions P1 to P9.
After switching on the TG200 the indication “on” appears in the display 2. By pressing the key PROG the next programme position P1 to P9 can be reached. After programme position P9 on reappears in the display. In the PROG mode all keys except PROG, START and STOP are inhibited.
on = free adjustment of the drying parameters 5/8/10-11
and 13 is possible, or
P1 to P9 = parameters can be saved and recalled here Start = drying process is started with the parameters
stored in programme position P1 to P9.
Allocation of storage location
Press PROG key until required storage location P1-P9 has
been reached.
Press SET key – all displays flash Set drying parameters 5/8/10-11/13 Press SET key; values are stored
Flashing of the display stops, the adjustment inhibitor is activated and sieve parameters are stored. In the PROG mode all keys except PROG, START and STOP are inhibited.
If the drying process is brought to an end within a definite time without interruption, the end is signalled by an acoustic signal tone which is sounded 5 times.
Storage and recalling drying parameters
2/25/2013 Retsch GmbH 16 0002
Signal tone for end of drying process
Page 17
Inserting the drying vessels
1. 2. 3. 4.
5.
The drying vessels of 6 litres capacity are not included in the items supplied with the TG200. They are available as separate accessories.
Art. No. 72.783.0001 – glass vessel
Art. No. 72.783.0002 - vessel of stainless steel
1. Place clamping lid in position, in doing so press the red
lever upwards – the clamping lid can now be moved
2. fits with the top edge of the support rod
3. Fill material to be dried. Place vessel in position
4. Tighten clamping lid
5. Press green lever downwards several times
We recommend that protective gloves are worn when removing the vessel and the clamping lid after the drying process.
Risk of burns.
D W
Bei sehr feinen staubigen Materialien kann die weichere im Lieferumfang enthaltene Dichtung eingesetzt werden.
Dichtung D im Deckel entnehmen und durch die weichere Dichtung
W ersetzen.
6. 7. 8.
The disk are not included in the items supplied with the TG200. They are available as separate accessories.
Art. No 03.648.0009 – sieve bottom Conidur 63 µm
6. Remove clamping elements and seal
7. Insert sieve bottom, Conidur 63 µm, centrally
8. Replace seal
9. Insert clamping elements and clamping lid. In doing so
press the green lever downwards several times.
9.
6 litre drying vessels of glass or stainless steel
Verwendung der weicheren Dichtung im Spanndeckel
Use and installation of the 63µm disk as additional filter
2/25/2013 Retsch GmbH 17 0002
Page 18
Inserting the clamping lid with filter element
10. 11. 12. 13.
14. 15.
The quick clamping lid with replaceable filter element is not included in the items supplied with the TG200. It is available as a separate accessory.
Art. No 72.643.0001 – quick clamping lid with replaceable filter
element
10. Turn clamping elements backwards
11. Remove clamping elements
12. Insert clamping elements in the quick clamping lid
13. Turn clamping elements to locking position
14. Place quick clamping lid with clamping elements in position
and tighten
15. Press green lever downwards several times
We recommend that protective gloves are worn when removing the vessel and clamping lid at the end of the drying process.
Risk of burns.
When using the clamping lid with filter element the TG200 may be operated only up to a temperature of max. +90°C/+194°F.
Otherwise the filter fabric will burn. Risk of burns.
16. 17. 18. 19.
20. 21.
The glass drying vessels, of 3 x 300 ml capactiy, are not included in the items supplied with the TG200. They are available as a separate accessory.
Art. No 72.002.0005 – glass vessels 3x 300ml
16. Insert intermediate ring
17. Place clamping lid in position
18. Position clamping elements and tighten
19. Press green lever downwards several times
20. Fill glass vessels and place in position
21. Clamp tightly by turning clockwise
We recommend that protective gloves are worn when removing the vessel and clamping lid after the drying process.
Risk of burns.
When using the lid with filter element the TG200 may be operated only up to a temperature of max. +90°C/+194°F.
Otherwise the filter fabric will burn. Risk of fire.
Attachment with 3 x 300 ml glass vessels
2/25/2013 Retsch GmbH 18 0002
Page 19
Working instructions
Schematic diagram
The TG200 is an ultra-modern, highly efficient product from the firm Retsch GmbH. Through the fluidized bed technology (see schematic diagram) and the generously dimensioned drying vessels materials can be dried effectively in the TG200 within a few minutes. The air throughput and heating power can be set separately, the temperature reached depending on the air throughput and the material to be dried. In this process agglomeration, as with conventional drying processes, is avoided to a considerable extent; further processing is thus facilitated.
A large selection of accessories makes it possible to use the machine for many different applications (see section on accessories).
TM
Air filtered on the suction side (filter quality – pore size 30 µm) is passed through a radial fan into the heating chamber. The heated air then enters the drying vessel and brings about continuous circulation of the material to be dried through the fluidized bed technology in the drying vessel. See schematic diagram.
The hot air removes moisture from the material to be dried and escapes through the filter bag or filter element (30 µm pore size).
The temperature is measured below the drying vessel in the air stream. See photo – temperature measuring point TM.
On completion of the drying process the vessel with dried material can be easily removed.
When removing the lid and drying vessel hold this only by the plastic grips or clamping elements provided. We recommend the use of protective gloves.
Risk of burns on hot metal and glass components.
General
Mode of operation of the TG200
2/25/2013 Retsch GmbH 19 0002
Page 20
Attainable temperatures at the measuring point
TM
Max. temperature set at 150°C or 302°F
Ambient temperature of 23°C / 73.4°F
Time 8 min
Air flow
Attainable
temp °C (±10%)
Attainable
temp °F (±10%)
99
ca. 87°C
ca. 188°F
60
ca. 93°C
ca. 199°F
40
ca. 100°C
ca. 212°F
20
ca. 128°C
ca. 262°F
10
ca. 150°C
ca. 302°F
Max. temperature set at 150°C or 302°F
Ambient temperature of 23°C / 73.4°F
Time 8 min
Air flow
Attainable
temp °C (±10%)
Attainable
temp °F (±10%)
99
ca. 108°C
ca. 226°F
60
ca. 115°C
ca. 239°F
40
ca. 124°C
ca. 255°F
20
ca. 150°C
ca. 302°F
Max. temperature set at 150°C or 302°F
Ambient temperature of 23°C / 73.4°F
Time 8 min
Air flow
Attainable
temp °C (±10%)
Attainable
temp °F (±10%)
99
ca. 88°C
ca. 190°F
40
ca. 98°C
ca. 208°F
30
ca. 106°C
ca. 222°F
20
ca. 120°C
ca. 248°F
10
ca. 150°C
ca. 302°F
Since the attainable temperature depends on the set fan power, the charged material to be dried and the type of clamping lid, the table below serves as a guide. The temperature in the TG200 is measured at the measuring point
TM.
Operating conditions 1:
TG200 without material to be dried, with basic equipment – clamping lid with filter bag and 6 ltr glass vessel as separate accesory, Art no. 72.783.0001
Operating conditions 2:
TG200 without material to be dried and with attachment with 3 glass vessels, each of 0.3 ltr, as separate accessory - Art. No.
72.002.0005
Operating conditions 3:
TG200 without material to be dried and with as separate accessory - Art. No. 72.643.0001
2/25/2013 Retsch GmbH 20 0002
Page 21
General
Do not clean the TG200 with running water.
Danger to life through current surge.
F1
The 3 filters F1 on the rear side of the machine must be replaced after approx. 100 hrs. The maintenance interval is shortened depending on the dustiness of the surroundings. This must be assessed empirically by the operator.
If the filter elements and bags are contaminated too much the efficiency of the TG200 will be reduced to an unknown extent.
Cleaning
TG200
The TG200 should be cleaned only with brushes, and possibly with an industrial vacuum cleaner or with compressed air.
The system of protection for the TG200 with drying vessel and clamping lid is IP43.
Without vessel and lid this is reduced to IP22.
Filter bags
The filter bags can be washed out under running water.
Filter elements and filter on the machine rear side
These must be replaced when contaminated.
Drying vessels 6 and 0.3 litres
These can be cleaned under running water or in an ultrasonic bath.
Maintenance
2/25/2013 Retsch GmbH 21 0002
Page 22
Replacing fuses
GS
F3 F4 F2 F1
Disconnect mains plug Remove 5 housing screws GS Place upper part of housing carefully on its side Remove fuse holders 1 - 4 Check fuses and replace if necessary Fuse F1 = 6.3 A mT (dia. 5x20) for fan Fuse F2 = 6.3 A mT (dia. 5x20) for fan Fuse F3 = 10 A mT (dia. 5x20) for heater
Fuse F4 = 10 A mT (dia. 5x20) for heater
Glass drying vessel, 6 litre Art. No. 72.783.0001
Drying vessel of stainless steel, 6 litre Art. No. 72.783.0002
Glass drying vessel, 3 x 0.3 litre Art. No. 72.002.0005 Spare glass drying vessel, 0.3 litre Art. No. 02.045.0020 Spare filter bag for glass drying vessel, 0.3 litre Art. No. 02.186.0004
Lid with filter element for 72.002.0005 Art. No. 72.107.0001 1 spare filter element Art. No. 03.186.0024
Quick clamping lid with filter element, incl. 10 spare filters Art. No. 72.643.0001 10 spare filters for 72.643.0001 Art. No. 72.143.0001
Quick clamping lid with filter bag Art. No. 72.107.0002 Spare filter bag with steel ring Art. No. 02.186.0015
Adaptor for sieve drying for Retsch analytical sieves 8“/203 mm Art. No. 72.001.0005
Accessories
2/25/2013 Retsch GmbH 22 0002
Page 23
Wearing parts
10 filter mats Art. No. 72.143.0003
Vulkollan seal in quick clamping lid Art. No. 05.111.0077
O-Ring on steel or glass vessel 185x4 Art. No. 05.114.0044
These operating instructions do not include instructions for repairs. For your own safety repairs should be carried out only by Retsch GmbH, an authorised agent or service technicians.
F1 PT100 temperature sensor is defective or not connected.
Connect or replace the temperature sensor.
F2 Overload temperature Bimetal monitor on the heating tripped.
Reduce the set temperature at the TG200 by at least 5°C.
F3 Fault in blower Blower has stopped or the speed is too low.
Increase the speed by at least 5 or replace blower.
Fault codes
Copyright
Passing on or duplication of this documentation, or use and passing on of the contents, is allowed only with the express permission of Retsch GmbH. Persons infringing are liable to pay damages.
Alterations
Technical data subject to change without notice.
2/25/2013 Retsch GmbH 23 0002
Page 24
Page 25
Authorized person for the compilation of technical documents:
J. Bunke (technical documentation)
The following records are held by Retsch GmbH in the form of Technical Documentation:
Detailed records of engineering development, construction plans, study (analysis) of the measures required for conformity assurance, analysis of the residual risks involved and operating instructions in due form according to the approved regulations for preparation of user information data.
The CE-conformity of the Retsch Rapid Dryer Type TG 200 is assured herewith.
In case of a modification to the machine not previously agreed with us as well as the use of not licensed spare parts and accessories this certificate will lose its validity.
Retsch GmbH Haan, January 2010 Dr. Stefan Mähler
Manager technical services
Retsch GmbH Ret sch-Alle e 1- 5 42781 Haan G erma ny w w w . r e t sc h .c o m
Rapid Dryer
TG 200
FB-EW-805-052 (E) Änderungsstand C 01.2012
CERTIFICATE OF CE-CONFORMITY
Translation
Certificate of CE-Conformity according to:
EC Mechanical Engineering Directive 2006/42/EC
Applied harmonized standards, in particular: DIN EN ISO 12100 Security of machines
EC Directive Electromagnetic Compatibility 2004/108/EC Applied standards, in particular: EN 61236 Electrical equipment for measurement, control and laboratory use in
conjunction with EN 55011 and EN 61000
Additional applied standards, in particular
DIN EN 61010 Safety prescriptions concerning measuring-, operating-, controlling- and
laboratory equipment
Page 26
Page 27
Page 28
Copyright
® Copyright by Retsch GmbH Haan, Retsch-Allee 1-5 D-42781 Haan Federal Republic of Germany
Loading...