Open Loop Circulation ............................................................................................................................ 18
Refrigeration Control Connections (Refrigerating/Heating Circulators only) .......................................... 19
Electrical Power ...................................................................................................................................... 19
Heat Only Circulators and Open Bath Systems ................................................................................. 20
Communication ....................................................................................................................................... 21
USB Communication .......................................................................................................................... 22
Power ...................................................................................................................................................... 23
Safety Set Temperature.......................................................................................................................... 23
Menu Structure ....................................................................................................................................... 27
Main Operational Displays (Home Screens) .......................................................................................... 28
Status Bar ........................................................................................................................................... 28
Home Screens .................................................................................................................................... 29
Adjusting the Temperature Set Point ...................................................................................................... 30
General Operational Settings ................................................................................................................. 32
Unit ..................................................................................................................................................... 33
Language ............................................................................................................................................ 33
Lock Out ............................................................................................................................................. 33
Auto Restart ........................................................................................................................................ 33
Time and Date Settings .......................................................................................................................... 34
Time .................................................................................................................................................... 34
Date .................................................................................................................................................... 34
High Limit / Low Limit Temperatures .................................................................................................. 35
High Alarm / Low Alarm Temperatures .............................................................................................. 35
Control .................................................................................................................................................... 36
Probe Control ..................................................................................................................................... 36
Auto Cool ............................................................................................................................................ 36
Fluid Type ........................................................................................................................................... 37
USB .................................................................................................................................................... 39
Restoring All Factory Default Values ...................................................................................................... 57
System Restoration .................................................................................................................................. 57
Changing Your Circulator's Viewing Angle ............................................................................................ 58
Inert Gas Purge ......................................................................................................................................... 58
Tap Water Cooling .................................................................................................................................... 58
Technical Information ............................................................................................................................... 72
Thank you for choosing this Circulating Bath with Advanced Programmable Temperature Controller. It is
intended for the precise temperature control of suitable liquids in a reservoir. Extremely easy to use and
maintain, it combines design innovation with highly intuitive operation to deliver convenient and versatile
liquid temperature control for a wide range of applications.
WARNING: PolyScience Circulating Baths are not intended for directly controlling the temperature of
foods, pharmaceuticals, medicines, or other objects which may be ingested by or injected in humans
or animals. Any such objects must be isolated from contact with the bath fluid and bath surfaces.
Here are some of the features that make your Circulating Bath so user-friendly:
Intuitive touch screen operation
OpenMode time/temperature programming (no restriction on the number of programs or
steps)
Selection of seven different temperature displays, including time-temperature graphing
Powerful variable-speed pressure/suction pump with external circulation capability
180° viewing radius (Swivel 180™ rotating control head)
DuraTop™ heat and chemical resistant top plate
LidDock™ self-storing reservoir cover (integrated baths only)
Built-in temperature protection
Suitable for use with Class III flammable bath fluids per DIN 12876-1
It will take you very little time to get your new Circulating Bath installed and running. This Operator’s
Manual is designed to guide you quickly through the process. We recommend that you read it thoroughly
before you begin.
Circulating Baths with Advanced Programmable Temperature Controller
Model Type
Reservoir
Capacity
°C °F
AP07R-20 Refrigerating/Heating 7 liter -20° to 200°C -4° to 392°F
AP07R-40 Refrigerating/Heating 7 liter -40° to 200°C -40° to 392°F
AP7LR-20 Refrigerating/Heating 7 liter -20° to 200°C -4° to 392°F
AP15R-30 Refrigerating/Heating 15 liter -30° to 200°C -22° to 392°F
AP15R-40 Refrigerating/Heating 15 liter -40° to 200°C -40° to 392°F
AP28R-30 Refrigerating/Heating 28 liter -30° to 200°C -22° to 392°F
AP45R-20 Refrigerating/Heating 45 liter -25° to 135°C
(1)
-13° to 275°F
(1)
(1)
(1)
AP07H200 Heating Only 7 liter Ambient +10° to 200°C Ambient +20° to 392°F
AP15H200 Heating Only 15 liter Ambient +10° to 200°C Ambient +20° to 392°F
AP20H200 Heating Only 20 liter Ambient +10° to 200°C Ambient +20° to 392°F
AP28H200 Heating Only 28 liter Ambient +10° to 200°C Ambient +20° to 392°F
AP06S150 Stainless Steel Open Tank 6 liter Ambient +10° to 150°C
AP10S150 Stainless Steel Open Tank 10 liter Ambient +10° to 150°C
AP28S150 Stainless Steel Open Tank 28 liter Ambient +10° to 150°C
AP08P100 Polycarbonate Open Tank 8 liter Ambient +10° to 85°C
AP11P100 Polycarbonate Open Tank 11 liter Ambient +10° to 85°C
AP14P100 Polycarbonate Open Tank 14 liter Ambient +10° to 85°C
AP17P100 Polycarbonate Open Tank 17 liter Ambient +10° to 85°C
AP23P100 Polycarbonate Open Tank 23 liter Ambient +10° to 85°C
AP28P100 Polycarbonate Open Tank 28 liter Ambient +10° to 85°C
AP29VB3S Polycarbonate Viscosity 29 liter Ambient +10° to 85°C
AP29VB5R Polycarbonate Viscosity 29 liter Ambient +10° to 85°C
(2)
Ambient +20° to 302°F
(2)
Ambient +20° to 302°F
(2)
Ambient +20° to 302°F
(3)
Ambient +20 to 185°F
(3)
Ambient +20 to 185°F
(3)
Ambient +20 to 185°F
(3)
Ambient +20 to 185°F
(3)
Ambient +20 to 185°F
(3)
Ambient +20 to 185°F
(3)
Ambient +20 to 185°F
(3)
Ambient +20 to 185°F
(2)
(2)
(2)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
(3)
1. Maximum operating temperature at which ±0.005°C temperature stability can be maintained; Advanced Programmable Controller is capable of
higher temperatures.
2. Maximum operating temperature for stainless steel tank. Advanced Programmable Controller is capable of higher temperatures.
3. Maximum operating temperature for polycarbonate tank. Advanced Programmable Controller is capable of higher temperatures.
When installed, operated, and maintained according to the directions in this manual and common safety
procedures, your Circulating Bath should provide safe and reliable temperature control. Please ensure
that all individuals involved in the installation, operation, or maintenance of this Circulating Bath read this
manual thoroughly prior to working with the unit.
This symbol alerts you to a wide range of potential dangers.
This symbol advises you of danger from electricity or electric shock.
This symbol indicates that a hot surface may be present.
This symbol marks information that is particularly important.
/
This symbol indicates alternating current.
These symbols on the Power Switch / Circuit Breaker indicate that they place the main power supply
ON / OFF.
This symbol on the Power Key indicates that it places the unit in a standby mode. It DOES NOT fully
disconnect the unit from the power supply.
Read all instructions pertaining to safety, set-up, and operation.
Proper operation and maintenance is the user’s responsibility.
To prevent injury to personnel and/or damage to property, always follow your workplace’s safety
procedures when operating this equipment. You should also comply with the following safety
recommendations:
WARNING:
This Circulating Bath is suitable for use with Class III flammable fluids per DIN 12876-1. A fire
hazard may be present.
Be aware of the chemical hazards that may be associated with the bath fluid used. Observe all
safety warnings for the fluids used as well as those contained in the material safety data sheet.
Explosive gas mixtures may accumulate if used with insufficient ventilation. Use this Circulating
Bath in a well ventilated area or beneath a suitable fume hood only.
Use only recommended bath fluids; see Technical Information in the rear of this manual for
recommended fluids.
Use only non-acid bath fluids.
WARNING: When using Class III flammable fluids per DIN 12876-1, the user must attach the
following warning labels to the front of the unit so that they are well visible:
Warning Label
W09
Colors: Yellow/black
Mandatory Label
M018
Colors: Blue/white
or
Semi S1-0701
Table A1-2 #9
Colors: Blue/white
WARNING:
Always connect the power cord on this Circulator to a grounded (3-prong) power outlet. Make
certain that the outlet is the same voltage and frequency as your unit.
Never operate the Circulator with a damaged power cord.
Always turn the Circulator OFF and disconnect mains power before performing any maintenance or
service.
WARNING:
Never operate the Circulator without bath fluid in the reservoir. Periodically check the reservoir to
ensure that the liquid depth is within acceptable levels. Always refill the reservoir using the same
bath fluid type that is already in the reservoir. Bath oil must not contain any water contaminants
and should be preheated to the actual bath temperature before adding as there is an explosion
hazard at high temperatures.
Always drain all fluid from the reservoir before moving or lifting your Circulator. Be sure to follow
your organization’s procedures and practices regarding the safe lifting and relocation of heavy
objects.
Danger Area.
Attention! Observe instructions
(operating manual, safety data sheet)
Carefully read the user information prior to
beginning operation.
Scope: EU
Carefully read the user information prior to
beginning operation.
Always allow the bath fluid to cool to ambient temperature before draining.
The reservoir cover, top deck, and/or external pump connections may become hot with continuous
use. Exercise caution when touching these parts.
Always keep within the 85°C maximum operating temperature limit if using a polycarbonate tank.
WARNING: It is the user’s responsibility to properly decontaminate the unit in the event hazardous
materials are spilled on exterior or interior surfaces. Consult manufacturer if there is any doubt
regarding the compatibility of decontamination or cleaning agents.
Regulatory Compliance & Testing
Canada USA (60 Hz units)
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 1:
General Requirements (CAN/CSA C22.2 No. 61010-1-12)
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 1:
General Requirements (UL 61010-1:2012)
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 2-010:
Particular Requirements for Laboratory Equipment for the Heating of Materials (CSA 61010-2-010:15)
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 2-010:
Particular Requirements for Laboratory Equipment for the Heating of Materials (UL 61010-2-010:2015)
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 2-051:
Particular Requirements for Laboratory Equipment for Mixing and Stirring (CSA 61010-2-051:15)
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 2-051:
Particular Requirements for Laboratory Equipment for Mixing and Stirring (UL 61010-2-051:2015)
Your Circulator was packed in a special carton or cartons. You should keep the packaging, along with all
packing materials, until the unit has been installed and you are certain it is working properly.
CAUTION: Remove any loose packing material that may have fallen into the heater/pump housing
during shipping. Before powering up, check that nothing remains around the heater or Circulator
pump.
We recommend that you begin using your Circulator immediately to confirm proper operation, since
beyond one week you may be eligible for warranty repair only (rather than replacement). You’ll find
complete warranty information in the back of this manual.
WARNING: Keep unit upright when moving. Be sure to follow your company’s procedures and
practices regarding the safe lifting and relocation of heavy objects.
Contents
The items included with your Circulator will vary depending on which model Circulating Bath you
purchased.
Refrigerating / Heating Bath Heating Only Bath
Resource Disk with
Operator’s Manual
Reservoir Cover(s) ● ●
3-ft / 0.91 m IEC to IEC
Power Cord
6-ft / 1.82 m IEC to Mains
Power Cord
Refrigeration Control Cable ●
Fittings
Cooling Coil N/A Integral
Bath Lid Sleeves
Certificate of Compliance
Flash Drive ● ●
Pt100 External Temperature
(3)
Probe
Quick-Start Guide ● ●
1. 60Hz and 50Hz models
2. 50Hz models only
3. Included only with Calibration Baths (AP15RCAL)
(3)
● ●
●
● ●
(2)
(1)
(1)
(1)
1/4 in. NPT to 3/16 in. barbed adapter
1/4 in. NPT to 1/4 in. barbed adapter
1/4 in. NPT to 3/8 in. barbed adapter
Your Circulating Bath with Advanced Programmable Temperature Controller is designed to be simple to
set-up and install. The only tools required are a Philips-head screwdriver and a container for adding water
or other suitable fluid to the bath reservoir.
General Site Requirements
Locate your Circulator on a level surface free from drafts and direct sunlight. Do not place it where there
are corrosive fumes, excessive moisture, high room temperatures, or in excessively dusty areas.
Refrigerating / Heating Circulators must be 10.2 cm / 4 inches or more away from walls or vertical
surfaces so that airflow is not restricted.
Avoid voltage drops by using properly grounded power outlets wired with 14 gauge or larger diameter
wire and if possible, be close to the power distribution panel. The use of extension cords is not
recommended; this will reduce the potential for problems caused by low line voltage.
Maximum Fluid Level =
1 inch / 2.54 cm below
underside of top deck
Minimum Fluid Level =
hes / 7.6 cm
below underside of top
deck
Adding Liquid to the Bath Reservoir
WARNING: When using Class III flammable fluids per DIN 12876-1, the user must attach the
following warning labels to the front of the unit so that they are well visible:
Warning Label
W09
Colors: Yellow/black
Mandatory Label
M018
Colors: Blue/white
or
Semi S1-0701
Table A1-2 #9
Colors: Blue/white
WARNING: Read the safety data sheet for the bath fluid being used carefully before filling reservoir.
WARNING: See Technical Information in the rear of this manual for a list of compatible liquids.
WARNING: If the proper fluid level is not maintained, the heater coil may become exposed and
possibly damaged (fluid level too low) or the bath may overflow (fluid level too high).
Danger Area.
Attention! Observe instructions
(operating manual, safety data sheet)
Carefully read the user information prior to
beginning operation.
Scope: EU
Carefully read the user information prior to
beginning operation.
Scope: NAFTA
The liquid in the reservoir should be maintained at a depth between 1 inch / 2.54 cm and 3.0 inches / 7.6
cm below the underside of the bath’s top deck. Upon start up, it may be necessary to add fluid to the bath
to compensate for the fluid required for external circulation. Likewise, be sure to compensate for fluid
displacement when placing samples or other materials in the Circulator’s reservoir.
3.0 inc
WARNING: Always drain all fluid from the reservoir before moving or lifting your Circulator. Be sure
to follow your organization’s procedures and practices regarding the safe lifting and relocation of
heavy objects.
WARNING: To avoid the potential for burns, allow the Circulator to cool completely before cleaning
WARNING: When connecting tubing to an external application, it is the user’s responsibility to make
sure that the tubing and fittings connected to the Circulator are suitable for the fluid being used and
the temperature range of operation.
CAUTION: The Circulator’s bypass tubing is secured to the fluid inlet and outlet connections by high
temperature nylon hose clamps, which can be removed by carefully cutting them with diagonal
cutters.
CAUTION: Secure the tubing to the inlet and outlet fittings using hose clamps with a minimum ID of
7/8 inch (22 mm). Do not operate the unit without hose clamps.
WARNING: If the Circulating Bath will not be used for external circulation, the inlet and outlet ports
should remain connected using the Buna N bypass tubing provided with the unit.
The pump inlet and outlet ports are female ¼ inch NPT connections that permit use of barbed tubing
adapters or hard plumbing fittings. ½ inch (13mm) ID tubing may also be slid over these connections and
held in place with a hose clamp.
If the pump inlet and outlet are not used for external circulation, the Bypass Tubing provided with the unit
should be left in place in order to optimize fluid mixing within the reservoir.
The nylon barbed tubing adapter fittings supplied with the unit are intended for applications from
®
-40° to 93°C. For applications above 93°C, brass, stainless steel, or Teflon
fittings are recommended. ¼
inch NPT to M16 stainless steel male adapter fittings are provided with all 50Hz models.
NOTE: The use of quick-connect fittings is not recommended as they typically restrict flow rate.
External Closed Loop Circulation
Connect the pump inlet and outlet to the external apparatus. To maintain adequate flow, avoid restrictions
in the tubing. When connecting the Circulator to more than two closed loops, the use of a manifold made
of “Y” adapters to divide the fluid into multiple banks is recommended. After setting up multiple closed
loops, check for adequate flow at the return manifold of each loop and check that the bath fluid is at an
adequate level. A booster pump may be added to closed loops without damaging the Circulator’s pump.
The temperature control stability of a closed loop system is better at the external apparatus than in the
Circulator reservoir (provided the control point of the apparatus represents a constant load and is well
insulated). For example, if you circulate fluid through a viscometer at 50°C, the temperature variation
observed in the Circulator reservoir may be ±0.1°C while the temperature variation in the viscometer may
be only ±0.05°C.
Although temperature stability is generally better at the external apparatus control point, depending on the
length of tubing used and the efficiency of the insulation, the actual temperature reading at the external
apparatus may be slightly different than the temperature reading at the Circulator reservoir.
The duplex (pressure/suction) pump permits circulation to and from an external open bath. To prevent
siphoning when the Circulating Bath is turned off, position both baths so that the two fluid levels are at the
same elevation.
Connect the pump inlet and outlet to the
external bath using tubing of the same
diameter and length. The same size fittings
should also be used on both the inlet (suction)
and outlet (pressure). This helps ensure a
balanced flow. A restricting valve or pinch clip
should be installed in the pressure (outlet)
tubing and adjusted to match the return
suction (inlet) flow rate. Cut the external end
of the suction tube into a “V” shape so that the
tube will not seal itself against the wall of the
external tank. Both the pressure and suction
tubing should be securely fastened to the
external tank to prevent movement during use.
When using flexible tubing, the suction tubing
must have a wall thickness that will not collapse under vacuum, particularly when going around bends.
Circulating Bath Height Regulation — Position the ends of the pressure and suction tubes at the desired
maximum fluid level in the external bath and fill the bath to that level. Fill the Circulating Bath to a height
one inch (25mm) below the top of the reservoir. Start the pump and adjust the restricting valve/pinch clip
on the outlet tubing until the liquid height in both baths remains constant. Add fluid to the baths as needed
to compensate for the fluid in the inlet and outlet lines.
Refrigeration Control Connections (Refrigerating/Heating Circulators only)
Electrical Power
WARNING: The Circulator’s power cord must be connected to a properly grounded electrical
receptacle. Make certain that this electrical outlet is the same voltage and frequency as your
Circulator. The correct voltage and frequency for your Circulator are indicated on the identification
label on the back of the Controller.
CAUTION: The use of an extension cord is not recommended. If one is necessary, it must be
properly grounded and capable of handling the total wattage of the unit. The extension cord must not
cause more than a 10% drop in voltage to the unit.
Refrigerating / Heating Circulators
Attach the 3-ft / 0.91 m power cord to the IEC electrical connectors on the Temperature Controller (male)
and the Refrigeration Power Module (female).
Attach the 6-ft / 1.8 m power cord to the IEC electrical connection on the Refrigeration Power Module and
then plug the male connector into the Mains electrical outlet.
Place the Power Switch / Circuit Breaker on the Refrigeration Power Module in the ON position. An hour
glass will appear on the Temperature Controller’s display while the Circulator completes an initialization
sequence. Once completed, “Standby” will appear on the display; the Power Key will also light.
Attach the 6-ft / 1.8 m power cord to the IEC electrical connection on the Temperature Controller and then
plug the male connector into the Mains electrical outlet.
Place the Power Switch / Circuit Breaker on the Temperature Controller in the ON position. The display
on the Controller will light and “Standby” will appear on the display; the Power Key will also light.
The Advanced Programmable Controller features a variety of connectivity options. Following are some
typical ways you can use them to monitor and control the operation of your Circulator.
Two USB ports (A and B) are provided on the rear of the Temperature Controller for use with USB flash
drives to log temperature data and store and/or transfer time/temperature programs. See Basic Operation, Communications and Data Logging for more information.
Ethernet
An Ethernet port is provided on the back of the Temperature Controller to enable you to connect your
Circulator to a computer network.
RS232 / RS485 Serial Communication
CAUTION: Always turn electrical power to the Circulator OFF before making a connection to the
serial (DB9) port.
Your Circulator features RS232 / RS485 serial communication for remote data logging and control
capability. A DB9 connector is provided on the rear of the Temperature Controller for this purpose. See
Basic Operation, Communications and Data Logging for set up information.
The serial interface should be connected to a serial communication port on a remote PC using an
appropriate cable. Information on the RS232 / RS485 commands and communication protocol can be
found in the Technical Information section of this manual.
External (P2) Temperature Probe
Your Circulator is capable of controlling temperature based on either the temperature of the internal bath
or that of an external vessel or device. The connection for the optional external temperature probe is on
the rear of the Temperature Controller. The Temperature Controller automatically detects the external
temperature probe when it is connected. See Replacement Parts & Accessories for other available
lengths and part numbers.
IMPORTANT NOTE FOR CALIBRATION BATHS: The PT100 external temperature probe should be
inserted in the calibration reservoir.
Pin Out Diagrams — External (P2) Temperature Probe Connection
MAXIMUM OPERATING TEMPERATURE @ 200 C, CLASS A 0.003850 OHMS/DEGREES C.
Press . “Initializing” will appear briefly on the display, the Circulator will begin running, and the default
Main Run screen will appear.
Safety Set Temperature
NOTE: The Safety Set Temperature is displayed and set in whichever temperature unit is currently
selected. The default is °C. To set the temperature using °F, see General Operational Settings, Unit.
This is a “Do Not Exceed” temperature setting for your Circulator and is the temperature at which the
heater will be turned OFF should the liquid level in the bath drop too low or the heater malfunctions. It is
normally set about 5° higher than the desired operating temperature. Setting the Safety Set Temperature
is a simple 4-step procedure.
WARNING: The Safety Thermostat is user-adjustable from approximately 40° to 240°C / 104° to 464°F.
Do not force the indicator dial beyond the stops at either end of the dial’s range.
1. Press to access the Main menu.
2. Touch the Safety icon to access the Safety sub-menu.
3. Using a small, Philips head screwdriver, rotate the
Safety Set Thermostat on the rear of the
Temperature Controller until the desired Safety Set
Temperature is displayed (clockwise to increase;
counter-clockwise to decrease).
NOTE: The following examples are intended to illustrate how commonly used touch screen icons
function when displayed. A specific icon may or may not be displayed on a sub-menu page. The
function/operation of icons not shown here are described in the sections associated with the screens
on which they appear.
Help regarding sub-
– Touch to access
CAUTION: The menu and sub-menu screens time-out and revert to the Main Operational Display
(Home Screen) after approximately 30 seconds without any touch screen interaction. Any changes
made will take effect automatically.
Display
Lock Out
Auto Restart
Personalize
Display Filter
SHC
Safety Set
Safety Probe
High Limit
Low Limit
High Alarm
Low Alarm
Ethernet
RS232
USB
RS485 Address
Probe Control
Auto Cool
Fluid Type
P2 – P1
PID
External Cooling Control
Send Internal Data USB/Web
Reset memory and reboot
Timer
Time
Date
Schedule
5% to 100%
°C / °F
On / Off
Available display languages
Temperature display resolution: .XX / .XXX
Yes / No
Yes / No
Off / PNG
Display averaging: 1 to 7 seconds
Specific heat capacity unit: BTU/(LB.°F) / kJ(kg/°C)
Displays current safety set temperature
Displays current safety probe temperature
202°C maximum / 25°C minimum
-72°C minimum / 20°C maximum
202°C maximum / 25°C minimum
-72°C minimum / 20°C maximum
Ethernet connection
1200 to 115,200 baud
Logging On / Off
Addressable
Create, edit, delete, run, save, and transfer time/temperature
programs. No restriction on number of programs or steps.
P1 (internal) / P2 (external)
Set temperature at which refrigeration activates: 2° to 152°C
Optimizes operation for a specific fluid type
Maximum allowable temperature differential between P1 and P2
User / Factory
Operating parameters for optional external cooling valve
Outputs diagnostic file
Resets unit to factory defaults
0 to 999 minutes
Time and time format
Year, Month, Day
Set time and date to auto-start a program
Time remaining until fluid replacement
Reset / display number of days pump has been operating
Reset / display number of days compressor has been operating
Calibration table entry
Number of days unit has been operating
Version
Version
Version
Actual temperature
(P1 = Internal Control; P2 = External Control
Time
Lit when a program is scheduled to run
Main Operational Displays (Home Screens)
Your Circulator features seven Main Operational Displays (Home screens). You can see the available
choices and select a specific Home screen by pressing the
Status Bar
A status bar displaying important operating information is displayed at the top of every Home and Menu
screen. The information included on the status bar is determined by the type of screen on which it
appears.
Actual bath temperature
Shown to either two or three
decimal points
Set point temperature
Shown to two decimal points
Temperature probe controlling
temperature
Internal
External
Temperature Trend graph
Scales from 4
days
Scale currently in use for
temperature trend graph. Press
to change.
P1 Internal temperature
reading
P2 External temperature
reading or status
Set point temperature
Temperature trend line keys:
Black = P1 temperature
Green = Set point temperature
Red: = High and Low Alarm settings
Press highlighted button to add associated line to
temperature trend graph
Temperature probe monitoring
temperature
ternal
External
Home Screens
Primary Home Screens
The background for this style Home screen may be either blue or black.
NOTE: On primary home screens, the actual and set point temperatures are displayed to one decimal
point resolution only, regardless of the resolution selected in the Settings / Display sub-menu.
Temperature Detail Home Screens
fluid
P1 =
P2 =
fluid
P1 = In
P2 =
The background for this style Home screen may be either blue or black.
Temperature Trend Home Screen
minutes to 10
The background for this style Home screen is blue only.
Press this key when any
Home screen or the
Main Menu screen is
displayed
Personalized Home Screen
Adjusting the Temperature Set Point
This is the temperature at which the fluid in your Circulating Bath will be maintained. It may be set to one
one-hundredth of a degree over a range of -50.00° to +200.00°C / -58.00° to +392.000°F. The factory
default set point is +20.00°C / +68.00°C.
NOTE: Although temperature can be displayed with three decimal point resolution, set point
temperatures may only be set to two decimal points.
To optimize operational flexibility, set point temperature changes can be initiated in several ways.
Controller Keypad
NOTE: To change the temperature set point while operating in the Temperature Trend Home screen,
you must press the SET key on the Controller’s front panel. Temperature set point cannot be changed
while a program is running.
Inputs numerical value
Toggles minus sign
ON and OFF
Clears last number entered
Accepts change
Accepts change
Cancels change
Cancels change
CAUTION: The menu and sub-menu screens time-out and revert to the Main Operational Display
(Home Screen) if approximately 30 seconds pass without any touch screen interaction. Any changes
made will take effect automatically.
General Operational Settings
The Settings sub-menu is used to select the Temperature Controller’s general operational characteristics.
Press
to access the Main menu.
Touch the Settings icon to access the Settings sub-menus.
Pump Speed — Sets pump speed. The current setting is displayed between the < (decrease) and >
(increase) icons and is adjustable from 5% to 100%. Touch
Unit — Sets the unit in which temperature will be displayed. The current setting is shown in the nonhighlighted (blue) field. To change, touch Unit icon or the ° icon in the highlighted (white) field. Touch
accept the new temperature unit,
Sound — Turns sound ON and OFF. The current setting is shown in the non-highlighted (blue) field. To
change, touch the Sound icon or the icon in the highlighted (white) field. Touch
Language — Sets the language in which all menus will be displayed.
The current selection is shown to the right of the Language icon. To
change, touch either the left or right highlighted (white) Language field.
A screen showing the available languages will appear:
to cancel.
to accept the new value, χ to cancel.
to accept, χ to cancel.
to
The current setting is shown in the non-highlighted (blue) field. To
change, touch the icon associated with the desired language. Touch
to return to the Settings sub-menu.
Touch ∨ to advance to page 2 in the Settings sub-menu.
Display — Sets the number of decimal points in which actual and set point temperature will be displayed.
The current setting is shown in the non-highlighted (blue) field. To change, touch the Display icon or the
icon in the highlighted (white) field. Touch
NOTE: Temperature is shown to one decimal point only on the Primary (largest) display, regardless of
the display setting. See Main Operational Displays (Home Screens) for more information.
Lock Out — Allows you to block temperature set point and other changes. While this feature is activated,
the unit will remain running at the current settings. See Basic Operation, Enabling/Disabling the Local Lock Out for information on using this feature.
Auto Restart — This determines how the unit will begin operating after a disruption to electrical power.
When YES is selected, the Circulator will begin running automatically when power is restored. When NO
is selected, the Circulator will power up in the Standby mode. A message will also be displayed to alert
you to the power outage. The current setting is shown in the non-highlighted (blue) field. To change,
touch the Auto Restart icon or the icon in the highlighted (white) field. Touch
to cancel.
to accept, χ to cancel;
to accept the new setting,
WARNING: The unit may start automatically after a disruption in electrical power.
Personalize — Allows the personalization of one of the Home screens with an image or text. See
Creating and Installing a Personalized Home Screen for information on adding text/images.
Touch ∨ to advance to page 3 in the Settings sub-menu.
Display Filter —This setting determines the averaging time used when displaying temperature
information. Higher settings will generally result in less fluctuation, while lower settings will more
accurately reflect real-time temperature probe data. Use the < (decrease) and > (increase) icons to
change the current setting; touch
SHC (Specific Heat Capacity)— Sets the unit (BTU or kJ) in which Specific Heat Capacity will be
displayed. Works in conjunction with the Fluid Type setting in the Control sub-menu. To change, touch
the SHC icon or the icon in the highlighted (white) field. Touch
Time and Date Settings
Touch the Time icon on the Main Menu; the Time sub-menu will be displayed.
to accept the new value, χ to cancel.
to accept the new setting, χ to cancel.
Time — Sets the time in either 12 or 24 hour format. To change, touch the Time icon or the actual
displayed time. Depending on the actual time and current time format, one of the following screens will
appear:
12 hour format – AM 12 hour format – PM 24 hour format
To change the time format, touch the 12H or 24H. Touch √to accept the new setting, χ to cancel.
To change the time, touch the up/down arrow icons associated with the hours or minutes. You can also
change the hour or minutes setting by touching the number itself; this will bring up a numeric keypad on
which the correct hour or minute can be entered. Touch
Date — Sets the date in Year/Month/Day format. To change, touch the Date icon or the actual displayed
date; the following screen will appear:
See Controller Setup for
information on changing
this value
Displays the current
safety probe reading.
Highest allowable set point
temperature
Lowest allowable set point
temperature
Temperature above which
unit will automatically shut
down
Temperature below which
unit will automatically shut
own
Current setting; touch
number to access
numeric keypad
To change the current setting, press the appropriate Year, Month, or Day icon and then touch < or >. You
can also change the Day by simply pressing the appropriate date on the calendar. Touch
new setting,
to cancel.
NOTE: See Basic Operation, Timer and Advanced Operation, Event Scheduling for information on the
Timer and Schedule functions.
Safety Settings
Touch the Safety icon on the Main Menu. The following screen will appear:
Touch ∨ to advance to the next page in the Safety sub-menu.
to accept the
d
High Limit / Low Limit Temperatures — These values establish limits on how high or low the
temperature set point may be set. To change, touch the up/down arrow icons. You can also change the
limit value by touching the number itself; this will bring up a numeric keypad on which the desired limit
temperature can be entered. Touch
to accept the new setting, χ to cancel.
High Alarm / Low Alarm Temperatures — These values establish the maximum allowable temperature
range for the unit. If the fluid temperature goes outside of this range, the system will shut down. To
change, touch the up/down arrow icons. You can also change limit value by touching the number itself;
this will bring up a numeric keypad on which the desired limit temperature can be entered. Touch
Touch the Control icon on the Main Menu. Page 1 of the Control sub-menu will appear:
Probe Control — This establishes whether temperature control will be based on the fluid temperature
within the Circulator’s reservoir or the fluid temperature at an external device. Touch the P1 icon to select
internal control; touch the P2 icon to select external control. Touch
cancel.
to accept the new setting, χ to
NOTE: P2 can only be selected when a remote temperature probe is connected to the Temperature
Controller.
Auto Cool — This control setting functions only on refrigerated circulators. The setting determines the
bath temperature at which refrigeration will be activated, permits more precise control when operating at
high temperatures as well as more rapid cool downs. For most applications, a set point that is 15°C above
room temperature is recommended. The Auto-Cool control range is from +2°C to 152°C.
To change, touch the up/down arrow icons. You can also change the auto cool temperature by touching
the number itself; this will bring up a numeric keypad on which the desired value can be entered. Touch
to accept the new setting, χ to cancel.
Cool Command™ Refrigeration — -40°C 7 liter Refrigerating/Heating Circulators and 15 liter and larger
Refrigerating/Heating Circulators feature the Cool Command™ modulating refrigeration control system.
Cool Command allows the refrigeration system to turn on at a fluid temperature up to 152°C when the
temperature set point is changed to or below the Auto-Refrigeration set point (152°C maximum). As a
result, bath fluid cools more quickly.
Conventional Refrigeration — -20°C 7 liter Refrigerating/Heating Circulators use a conventional
refrigeration system. The refrigeration system will turn on when the bath fluid temperature and set point
are below the Auto-Refrigeration set point (70°C maximum).
Restores user-set PID
settings to original
Factory values
Outputs diagnostic
data file
Fluid Type — This sets the specific heat capacity (SHC) for the bath fluid being used to achieve optimal
temperature control. Touch the Fluid Type or Current Bath Fluid icon to access a list of fluids; select
custom from the list to manually enter the SHC. Touch
flu
I P2 – P1 I — This setting is for use when external temperature control (P2) is being used. It establishes
the maximum allowable differential between the external and internal measured temperatures and is
intended to prevent uncontrolled heating or cooling should the external temperature probe be dislodged
or fail. To change, touch the up/down arrow icons. You can also change displayed value by touching the
number itself; this will bring up a numeric keypad on which the desired differential temperature can be
entered. Touch
to accept the new setting, χ to cancel.
to accept the new setting, χ to cancel.
Touch ∨ to advance to page 2 in the Control sub-menu.
PID — This determines whether Factory or User-set PID settings will be used for temperature control. To
manually enter PID values, touch the User icon and then use the increase and decrease icons to adjust
the Proportional (P), Integral (I), and Derivative (D) settings to the desired values. Touch
Allows manual entry of IP
address via numeric
keypad
Selects dynamic
entry of IP address
Turns TCP Control password
protection ON and OFF
Sets TCP password via alphanumeric keypad
Selects manual
entry of IP address
Displays dynamically
assigned IP address
CAUTION: The Advanced Programmable Temperature Controller’s factory PID settings have been
adjusted to achieve optimum temperature control. User-set PID values should only be used by
individuals with a thorough knowledge and understanding of Proportional/Integral/Derivative control.
External Cooling Control — This feature is only functional on units fitted with the optional External
Cooling solenoid valve. See External Cooling Control in the Technical Information section of this manual
for more information.
Send Internal Data USB/WEB — This sends a diagnostic file to either a connected flash drive or web
monitoring page. See Diagnostic Data Logging in the Routine Maintenance & Troubleshooting section of
this manual for more information.
Reset Memory and Reboot — Restores Temperature Controller to all Factory default settings and
values. See Restoring All Factory Default Values.
Communications and Data Logging
Touch the Connect icon on the Main Menu. The following screen will appear:
Ethernet — The Advanced Programmable Controller may be connected directly to a laptop or desktop
computer via its Ethernet connection or indirectly via your facility’s wired or wireless network. Either type
of connection enables you to control and/or monitor the operation of your circulating bath in real time
using an Internet browser by entering its IP address. The Controller’s active serial communications
commands are used to retrieve and/or change operational information. The Ethernet selection in the
Connect sub-menu allows you to configure the IP address and subnet mask. When touched, the following
screen is displayed:
When DHCP is selected, the following screen appears. It displays the dynamically assigned IP address.
NOTE: See the Technical Information section for information regarding Ethernet configuration. See
the Advanced Operation section for information on monitoring/controlling your Circulator using an
internet browser.
RS232 — This is used to set the baud rate for RS232 bi-directional communications and, if desired, a
communications protocol that emulates other common circulating baths. The increase and decrease
icons are used to adjust the baud rate; you can also increase the baud rate by touching the displayed
numerical value. Touch
To select an RS232 communications protocol that emulates another circulator, touch the RS232 icon; the
following screen will appear:
Touch the icon that represents the represents Read (or Get) Temperature command for the device.
Touch
to accept the new settings, χ to cancel.
USB — The Advanced Programmable Temperature Controller has both USB A and USB B ports. The
USB A port is used for data logging; data is stored in a CSV file that can be read by spreadsheet
programs such as Microsoft Excel
touch the Logging icon to turn logging ON. Touch
to accept the new settings, χ to cancel.
®
. To begin data logging, plug a flash drive into the USB A port and then
to accept, χ to cancel.
NOTE: See the Technical Information section for complete information regarding data logging file
formats and content.
The USB B port can be used to monitor or control your Circulator using a personal computer. See the
Technical Information section of this manual for more information.
RS485 — This is used to set the address for RS485 communications. The Temperature Controller
supports cable lengths up to 1000 ft / 304.8 m. Use the increase and decrease icons to enter the desired
RS485 address. You can also enter the address by touching the RS485 icon; this will bring up a numeric
Touch the Time icon on the Main Menu. The Time sub-menu will be displayed.
The Timer may be set from 0 to 999 minutes in one minute increments. Use the increase and decrease
icons to change the displayed time or touch the Minutes icon to access a numeric keypad on which the
desired interval can be entered.
To start the timer, touch the Timer icon. A clock icon and digital counter counting down the time (in
minutes and seconds) will be displayed in the right corner of the status bar. Once it reaches 0:00, digital
counter will turn red and begin counting up so you know how much time has elapsed since the end of the
timed event.
To stop or reset the timer, touch the Timer icon.
NOTE: See Time and Date Settings for information on selecting the time format (12 or 24 hour) and
changing the time and date settings; see Advanced Operation for information on Event Scheduling.
Touch the Settings icon on the Main Menu. The Settings sub-menu will appear; Lock Out appears on
page 2 of this sub-menu.
To enable Local Lock Out, touch YES. A password entry screen will appear:
Enter any desired password from 1 to 5 digits in length and then touch the OK key. This both enables the
Local Lockout and sets the password needed to disable the Local Lock Out.
When the Local Lock Out is enabled, a password entry screen will appear if the operator attempts to
change the set point or access the Main Menu.
To disable the Local Lock Out, enter the password used to enable the lock out and touch OK. Access to
the Main Menu and set point adjustment functions will now be permitted.
NOTE: The password used to enable the Local Lock Out is stored in memory only until it is used to
disable the lock out. To re-enable the Local Lock Out, a new password (either the same password or a
new one) must be entered via the Settings, Lock Out sub-menu.
Press to access the Main menu and then touch the Program icon. The Program sub-menu will
appear:
Initial Program Step — To set the parameters for the first step in your time/temperature program, touch
a PX_(Empty) icon and then Edit. A screen for the first program step will appear:
Touch the From / To icon. The screen for entering the Step Start Temperature, Step End Temperature,
Step Duration (in minutes), and External Cooling will appear:
Step Start Temperature – This is the bath temperature at the beginning of the displayed
Step. It is only settable for the initial step in the program.
Step End Temperature – This is the bath temperature desired at the end of the Step. It can be
higher, lower, or the same as the Step Start Temperature depending on whether this is a ramp or
soak step.
Step Duration – This is the amount of time desired to complete the Step. Enter 0 if you wish to
ramp to the Step End Temperature as quickly as possible. Enter a higher value if you want to
allow a more prolonged time to reach the Step End Temperature or if this is a soak step.
External Cooling -- This feature is only functional on units fitted with the optional External
Cooling solenoid valve and affects the temperature ramp rate. The higher the % setting, the
greater the flow of cooling water through the cooling coil. Valve opening is based on 5 second
increments. If the External Cooling percentage is set at 20% and the Step Duration to 1 minute,
the valve will be open for one second every five seconds spread evenly over the entire minute
(i.e., open for 12 seconds). The cooling icon on the Status Bar will be lit whenever the External
Cooling valve is open.
Enter Step Start and Step End temperatures, the Step Duration, and the External Cooling % by touching
the associated icon and then using the keypad to enter the desired value. Touch
values,
to cancel.
IMPORTANT: The initial Step Start temperature establishes the bath temperature that must be
achieved before the Program will run. If the actual bath temperature is above or below the initial Step
Start temperature, cooling (if the circulator is refrigerated) or heating will be applied to bring the bath to
the Step Start temperature. Once that temperature has been achieved, the Program will begin
running.
Additional Program Steps — Touch ⊕ to add another program step. The “From” temperature value for
each new step will default to the “End” temperature value from the previous step. Make changes to the
Step End and Step Duration values as indicated previously.
to accept the entered
NOTE: A Step Start temperature can only be entered for the first step in a program. All subsequent
steps use the Step End value from the previous step as the Step Start value.
Naming and Setting Program Run Parameters — Touch √ after you have entered all the steps you
want in the program.
and
A screen will appear that allows you to give your program a unique identifying name. There is an 8
character limit on a program name.
After you have named the program, you can set additional parameters for running the program
Priority – Programs may be run using either a Time- or Temperature-based priority. If achieving
successive temperature set points is critical, Temperature should be selected as the priority. If
completing a program in a fixed amount of time is essential, Time should be selected as the
priority.
Loop – This is the number of times the program will run through the programmed steps; I.e.,
automatically begins running Step 1 once the final step in the program has been completed.
Soak Temperature – This is the temperature the bath should be maintained at once the final
step has been completed.
Selects the Program to preview
Previews selected Program
Program Preview screen
Previewing a Program
The Preview function enables you to review a recently created time/temperature program or verify that
the program you want to run is the proper profile. This function is accessed by pressing
Program icon on the Main Menu, and then selecting the program you wish to preview.
The Preview screen includes the following information:
Initial Temperature — This is the start temperature for Step 1.
Soak Temperature — This is the temperature at which the bath will be maintained upon completion
of the program.
touching the
,
Loop Duration — This is the time (in minutes) it will take to complete one loop of the program.
Steps per Loop — This is the number of steps in the program.
Loop Number — This is the number of loops in the program
Fluid Temperature vs Time — This is a graphic representation of one program loop and illustrates
ramp and soak times associated with various program steps.
Press to access the Main menu and then touch the Program icon. The Program sub-menu will
appear:
Select the Program you wish to run and then touch the Start icon to begin running.
IMPORTANT: If the actual bath temperature is above or below the initial Step Start temperature,
cooling (if the circulator is refrigerated) or heating will be applied to bring the bath to the Step Start
temperature. Once that temperature has been achieved, the Program will begin running.
NOTE: A program can also be scheduled to run automatically at a predetermined time and date. See
Event Scheduling for more information.
change time
(4 minutes to 10 days)
The Program Run screen includes the following:
Temperature Information — (P1) the actual bath temperature, (P2) fluid temperature as measured
by the remote temperature sensor if connected and active, and (Set) the set point for the current
temperature step.
Program Name — The name of the program that is running.
Program Information — The priority selected for the program (time or temperature), current and
total number of loops, the current and total number of steps in the loop, the final soak temperature.
Temperature Graph — Actual (P1) bath temperature is always displayed. Touching the colored
icons adds the following information to the graph: P2 temperature (blue); set point (green); high and
low alarm levels (red).
NOTE: When Alarm level information is displayed, the temperature scale on the graph expands
to accommodate those temperature extremes and may cause set point and other temperature
information to be displayed less clearly.
Time Scale — The time scale covered by the graph may be changed by simply touching an area
within the graph. Temperature data can be displayed for periods ranging from the most recent 4
minutes to 10 days.
To temporarily stop a program, touch the Pause icon; the bath temperature will hold at the current
temperature until restarted. Press the Start icon to resume. To permanently stop a program, press the
Stop icon.
NOTE: A “paused” program can be restarted from the point it was paused (the Pause icon changes to
a Start icon). A “stopped” program cannot be restarted.
Editing a Program
Press to access the Main menu and then touch the Program icon. The Program sub-menu will
appear:
Select the Program you wish to edit and touch the Edit icon. The program screen for that program will
appear, with the final program Step displayed.
Adding a Step —To add a Step to the program, touch
a new Step will be added after the final Step in
the program. Enter the Step End temperature and Step Duration by touching the associated icon and
entering the desired values using the keypad. Touch
NOTE: A Step can only be added to the end of a program.
to accept the entered values, χ to cancel.
Deleting a Step — To delete the final program Step, touch the minus (-) icon. Touch √to confirm the
deletion,
to cancel.
NOTE: A Step can only be deleted from the end of a program.
Editing Step Parameters— To edit the Step End temperature, Step Duration, or External Cooling
percentage value for a particular Step, touch the icon for that Step. The parameters for that Step will
appear:
Select the parameter you wish to change and use the numeric keypad to enter the new value. Touch √to
accept,
Editing Program Parameters — To change a program parameter, such as the priority or number of
loops, touch the right arrow icon at the bottom right of the program steps screen. The program
parameters screen will appear:
to cancel.
IMPORTANT: When the Step End temperature is changed, the Step Start temperature for the
subsequent Step is automatically updated. However, the Step End temperature for the subsequent
Step does not change. Depending on the change made, it may be necessary to change the
temperature settings for all following Steps.
You may change any or all program parameters, including the name of the program.Touch √to accept
the changes,
to cancel.
Saving and Uploading Programs
This function allows you to save time/temperature programs to a flash drive or other external data storage
device. The saved file can then be used to restore the original program or uploaded to other Advanced
Programmable Temperature Controllers.
Saving a Program — Press
Program sub-menu will appear:
to access the Main menu and then touch the Program icon. The
Touch any of the program icons; the program list screen will appear.
Select the program you wish to save and touch the USB icon. A message will briefly appear indicating
that the file is being transferred to the USB followed by the file transfer screen.
Select a file to transfer and the touch the right arrow icon. Once the file transfer is complete, the file name
will appear in the list box on the right of the screen.
Uploading a Program — Press
Program sub-menu will appear:
to access the Main menu and then touch the Program icon. The
Touch any of the program icons; the program list screen will appear.
Select a location for the program you wish to upload and touch the USB icon. A message will briefly
appear indicating that the file is being transferred from the USB followed by the file transfer screen.
Select the destination for the upload, select the program you wish to upload, and then touch the left arrow
icon.
Once the file transfer is complete, the program name will appear on the program list.
Touch to change year, month,
or day using up and down
arrow icons
Event Scheduling
IMPORTANT: You must place the Circulator in Standby in order for a scheduled event to run.
This function allows you to schedule a time/temperature program to begin running automatically at a
predetermined time and date. The Circulator must be Off (in Standby); it will turn On automatically and
begin running the program at the scheduled time.
To schedule such an event, press
to access the Main menu and then touch the Time icon. The Time
sub-menu will appear:
Touch the Schedule icon. The scheduling screen will appear:
Selecting a Program — Touch the icon associated with the program you wish to schedule.
Entering the Start Date — Touch the Date icon. A calendar screen will appear.
Enter the Start date for the program: year, month, and date you want the program to start on. Touch √to
accept the scheduled Start date,
Entering the Start Time — Touch the Time icon. A time entry screen will appear:
Enter the Start time for program. Touch √to accept the scheduled Start time, χ to cancel. A screen will
appear showing the date and time at which the program is scheduled to run.
Deleting a Scheduled Event — To delete a scheduled program, select that program and then touch the
No Scheduled Program icon. Scheduling information for that event will be cleared.
Scheduled Event Reminder – When a program has been scheduled to run at a future date and time, the
Program/Event icon on the Status Bar illuminates.
When power is turned OFF, a reminder appears on the Standby screen.
Creating and Installing a Personalized Home Screen
The Advanced Programmable Controller features a special home page that you can easily personalize
with a special message, favorite photo, or other image. Following is an example of a personalized home
screen:
Creating the Image — The personalized home screen requires a 291 x 211 pixel Portable Network
Graphic (PNG) file. The file should be saved to a flash drive and named: PSC.png (be sure to use all
uppercase letters for PSC).
Installing the Image — Place the flash drive containing the PSC.png file in the USB A port. Press
access the Main menu and then touch the Settings icon. The Settings sub-menu will appear:
to
Touch ∨ to advance to page 2 of the Settings sub-menu.
Touch the PNG icon to turn personalization ON and the touch √.The Main Menu screen will appear. Your
image will now be displayed on the personalized home screen.
NOTE: Wait at least 30 seconds after turning personalization on before removing the flash drive from
the USB port.
Setting / value field
Read commands: shows current setting or value
Set commands: enter desired setting or value
Circulator Monitoring and Control Using an Internet Browser
NOTE: The Advanced Programmable Controller’s Ethernet capabilities may be used with any of the
following web browsers: Internet Explorer 9.0 or later, Firefox, Chrome, Safari, or Opera.
NOTE: A user-set password is required if you want to control (change set point, etc.) your Circulator
using an internet browser. A password is not required if all you want to do is view operation. See TCP Control Configuration in this section for information on password setup.
The Advanced Programmable Controller may be connected directly to a laptop or desktop computer via
its Ethernet connection or indirectly via your facility’s wired or wireless network. Either type of connection
enables you to control and/or monitor the operation of your circulating bath in real time using an Internet
browser by entering its IP address. See Ethernet Configuration in the Technical Information section for
information on IP address and TCP control configuration.
To view Circulator status using an internet browser, enter its IP address in the browser address bar. A
screen will appear with the following information and interactive fields:
Displays current calibration
settings for selected
temperature probe
Begins calibration
procedure for selected
temperature probe
Enters Calibration
password:
1234
Accepts password
entered
Calibration
CAUTION: The maximum calibration offset for any given temperature point is ±2.0°C. When
calibrating at 2 or more temperature points, there must be at least a 50°C span between the lowest
calibration temperature and the highest calibration temperature. Calibrations are always performed in
degrees C.
The Advanced Programmable Temperature Controller features 10-point calibration capability for both the
internal and optional external temperature probes. To access the Calibration function, touch the Service
icon on the Main Menu and then advance to page 2 of the Service sub-menu.
Touch the Calibration key. The following screen will appear:
Select the temperature probe to calibrate, enter the number of calibration points (1 to 10), and then touch
the Calibrate key. A password entry screen will appear:
Enter the Calibration password (1234) and touch OK. The following screen will appear:
CAUTION: All time/temperature programs, calibrations, and other user-set values are erased when
Factory default values are restored. Be sure to save any programs you wish to keep to a flash drive
and record any settings that you have changed.
Touch the Control icon on the Main Menu and then advance to page 2 of the Control sub-menu:
Touch the Reset Memory and Reboot key. A password entry screen will appear:
Enter the Restore Defaults password (1234) and touch OK. The Temperature Controller will power down
and then reboot in Standby.
System Restoration
CAUTION: All firmware updates made will be lost when a system restoration is performed. All
operational settings will also be returned to the original factory defaults. Be sure to save any programs
you wish to keep to a flash drive and record any settings that you have changed.
Should you need to restore the Advanced Programmable Controller to its original factory configuration
(firmware as well as all default operational settings), you may do so using its built-in system restoration
utility. System restoration is performed as follows:
1. Turn power to the Controller OFF by pressing the Power Key on the front of the Controller.
2. Place the Power Switch/Circuit Breaker on the rear of the unit in the OFF position.
3. Return the Power Switch/Circuit Breaker to the ON position while pressing and holding the Home
Key. Continue holding the Home Key until the touch screen calibration screen appears.
4. Follow the calibration instructions on the screen. Once you complete the touch screen calibration
procedure, the hour glass image will briefly reappear and followed by the Standby screen.
5. Press the Power Key to return to normal operation.
Your Circulator is equipped with Swivel 180™, an innovative feature which permits viewing of the
temperature display from anywhere over a 180° arc.
NOTE: There are positive stops at 45° intervals; however, the viewing angle may be set anywhere
within a 180° arc.
To change the viewing angle, slide the release latch to the right and rotate the Temperature Controller to
the desired angle. The latch release will automatically return to the locked position at every 45° positive
stop.
Inert Gas Purge
A 0.125 in. / 3 mm port on the rear of the Temperature Controller is provided to allow you to blanket the
surface of the liquid in the bath reservoir with nitrogen or another inert gas to help prevent condensation
and dilution of the bath fluid.
Tap Water Cooling
Tap water cooling allows for more rapid bath cool down from high temperatures and/or more precise
operation at temperatures near ambient.
Heat only Circulating Baths feature an integrated cooling coil as standard equipment. The tap water
connections are made on the rear of the unit. Two 0.25 inch / 64 mm female NPT fittings are provided for
these connections.
Polycarbonate Viscosity Baths feature a cooling coil with two 0.375 inch / 9.5 mm OD straight barb fittings
at which the fluid inlet and outlet connections can be made. Either connection may serve as the fluid inlet
or outlet. Be sure to secure the tubing with the appropriate size hose clamps.
An optional cooling coil is available for use on stainless steel and polycarbonate Open Bath systems. This
cooling coil has two 0.375 inch / 9.5 mm OD straight barb fittings at which the fluid inlet and outlet
connections can be made. Either connection may serve as the fluid inlet or outlet. Be sure to secure the
tubing with the appropriate size hose clamps.
WARNING: The fluid outlet must be connected and flow to a suitable drain or vessel located at a
level below that of the inlet.
Refrigerating/Heating and Heat Only Circulating Baths feature the LidDock® system to eliminate mess
when adding fluid or samples to the reservoir. Specially positioned notches in the inner lip of the top deck
allow you to stand the reservoir cover up upright over the bath opening, allowing condensate to flow back
into the bath.
h
NOTE: 75 liter Refrigerating/Heating Baths and Calibration Baths are not equipped with LidDock®.
WARNING: Always turn your Circulator OFF and disconnect it from the electrical power outlet before
performing any maintenance or service.
WARNING: To avoid the potential for burns, allow the Circulator to cool completely before cleaning
or performing any maintenance.
WARNING: Always drain all fluid from the reservoir before moving or lifting your Circulator. Be sure
to follow your organization’s procedures and practices regarding the safe lifting and relocation of
heavy objects.
Service Sub-Menu
This sub-menu contains important information regarding the operation of your Circulator, such as the
number of days the unit and key systems have been in service as well as the number of days until air
filter and fluid maintenance is required.
Press
Page 1 Page 2
Air Filter Service Interval — This allows you to view the number of days remaining until air filter cleaning
is required and set a service interval for this maintenance (10 to 180 days). This function is active only on
Refrigerating/Heating Circulators.
Touch the + and – icons to increase and decrease the service interval; Reset returns the time remaining
to 180 days.
Fluid Service Interval — This allows you to view the number of days remaining until bath fluid
replacement is required and set a service interval for this maintenance (10 to 180 days).
Touch the + and – icons to increase and decrease the service interval; Reset returns the time remaining
to 180 days.
Clear Pump Operation — This allows you to view the length of time the pump has been operating.
Touch Clear Pump operation to reset it to 0.
to access the Main menu. Touch the Service icon to access the Service sub-menu.
Clear Cooling Operation — This allows you to view the length of time the cooling system has been
operating. Touch Clear Cooling operation to reset it to 0.
Calibration — This allows you to calibrate the Temperature Controller. See the Calibration section of this
manual for more information.
Installation Operation — The length of time the Circulator has been operating.
Display Module Firmware — The version of the currently installed display module firmware.
Power Module Firmware — The version of the currently installed power module firmware.
Cooling Module Firmware — The version of the currently installed cooling module firmware.
Diagnostic Data Logging
This feature enables you to log key operational information that can be used by a service technician to
evaluate performance and diagnose problems. Logged data represents the Circulator’s performance at a
moment in time. The name of the file created is: TFTINFO.xml.
Data may be logged to a flash drive inserted in the USB A port.
1. Place a flash drive in the temperature controller’s USB A port.
2. Go to the second page of the Control sub-menu and select “Send Internal Data USB/WEB”.
3. Remove the flash drive from the USB A port.
Maintaining Clear Bath Water
Optimum temperature and moisture conditions for algae growth exist when using water as a bath fluid. To
prevent algae contamination and minimize the frequency of draining the reservoir, an algaecide such as
Polyclean Bath Algaecide (004-300040) should be used.
WARNING: Always drain all fluid from the reservoir before moving or lifting your Circulator. Be sure
to follow your organization’s procedures and practices regarding the safe lifting and relocation of
heavy objects.
WARNING: Bath fluids should be stored and disposed of according to applicable laws and
regulations.
Refrigerated / Heating and Heat only Circulating Baths
Refrigerated / Heating and Heat only Circulating Baths are equipped with a drain valve and port located
either beneath the front access panel or on the right hand side of the unit.
To drain fluid from the bath, attach a short length of suitable 0.45 inch / 11.5 mm ID tubing to the drain
port and secure it using a hose clamp with a minimum ID of 0.7 inch / 18 mm. Open the drain valve using
a flat blade screwdriver. When closing the valve, do not over tighten.
WARNING: Be sure to close the drain valve before refilling the bath reservoir. Do not over tighten.
Checking the Over-Temperature / Low Liquid Level Safety Systems
Your Circulator incorporates over-temperature and low liquid level protection according to IEC 61010-2-
010. For optimum safety, these systems should be checked at least every six months for proper
operation. These checks must be performed with the unit running.
Over-Temperature Protection
1. Press to access the Main menu.
2. Touch the Safety icon to access the Safety sub-menu.
3. Using a small, Philips head screwdriver, rotate the Safety
Set Thermostat on the rear of the Temperature Controller
until the unit shuts down. The Safety Set temperature at this
point should be the same as the actual bath temperature.
4. Return the Safety Set temperature to the desired overtemperature value.
5. Press the
Low Liquid-Level Protection
1. Set the temperature set point to ambient and allow the Circulator to stabilize at that temperature.
2. Increase the temperature set point to about 5°C above ambient and slowly drain fluid from the bath.
3. Continue draining fluid until the unit shuts down. The fluid level at this point should be approximately
3.0 inch / 7.6 cm below the underside of the Circulator’s top deck.
4. Replace the bath fluid and return to normal operation.
WARNING: It is the user’s responsibility to properly decontaminate the unit in the event hazardous
materials are spilled on exterior or interior surfaces. Consult the manufacturer if there is any doubt
regarding the compatibility of decontamination or cleaning agents.
Temperature Controller
Turn the Temperature Controller OFF by pressing and unplug power cord from the electrical outlet.
Wipe the housing with a clean cloth dampened with a mild detergent and water or mild all-purpose
cleaner.
CAUTION: Do not spray cleaning liquids directly onto the Temperature Controller or allow them to
enter the Controller’s vents. Do not use abrasives as these could scratch the housing or the digital
display.
Bath Reservoir
Bath Reservoir and Wetted Components — A concentrated bath cleaner (polyclean Bath Cleaner, part
number 004-300050) is available to remove mineral deposits from the stainless steel reservoir and the
Temperature Controller’s wetted parts. The cleaner should be added to the bath reservoir at the
prescribed dosage and circulated at 60°C / 140°F until the scale is removed.
CAUTION: Do not use steel wool to clean your Circulator’s bath reservoir.
External Surfaces
Only mild detergents and water or an approved cleaner should be used on the top deck and other
external surfaces of your Circulator. Do not allow cleaning liquids or sprays to enter the vents on the rear
of the Temperature Controller.
Pump Impeller
In the unlikely event that debris becomes lodged in the pump impeller, a soft brush can be used to
remove any lodged particles. If necessary, soak in a solution of distilled water and polyclean Bath Cleaner
to soften before brushing.
CAUTION: Do not use hard utensils or abrasive pads to remove trapped debris.
Condenser, Air Vents, and Reusable Filter
(Refrigerating / Heating Circulators only)
To keep the refrigeration system operating at optimum cooling capacity, the
condenser, removable air filter, and all air vents (front, side, back) should be
kept free of dust and dirt. Be sure to check them on a regular basis and clean
as required.
The reusable filter is easily accessed from the front of the unit by simply
removing the access panel. Use a mild detergent and water solution to wash off any accumulated dust
and dirt. Rinse and dry thoroughly before reinstalling.
Temperature Controller Removal and Re-Installation
Removal
The Temperature Controller on your Circulating Bath is designed to be easily removed from the top deck
without the use of special tools. It is removed as follows:
1. Place the tip of a small flat blade screwdriver under the retaining ring locking tab and pry up
gently.
2. Rotate the Temperature Controller clockwise until it stops (about 0.75 inch / 1.9 cm).
3. Lift the Controller straight up and out of the opening of the Circulator’s top deck.
The top deck of your Circulator incorporates four pins to facilitate positioning of the Temperature
Controller when it is being reinstalled. These pins correspond to keyhole slots on the interior of the
Circulator’s retaining ring.
1. With the retaining ring locking tab oriented above one of the indents on the top deck, slowly lower
the Temperature Controller into the top deck opening until it is resting on top of the positioning
pins.
2. Gently rotate the Temperature Controller until it drops down on the positioning pins.
3. Rotate the Temperature Controller counterclockwise until the Locking Tab engages the indent on
The Advanced Programmable Controller has a built-in feature that allows you to calibrate the touch
screen so that it responds more accurately to your touch. This calibration is performed as follows:
1. Turn power to the Controller OFF by pressing the Power Key on the front of the Controller.
2. Place the Power Switch/Circuit Breaker on the rear of the unit in the OFF position.
3. Return the Power Switch/Circuit Breaker to the ON position. An hour glass image will appear on
the display.
4. When the hour glass image fades away, press and hold the Home Key until the calibration screen
appears.
5. Follow the calibration instructions on the screen. Once you complete the touch screen calibration
procedure, the hour glass image will briefly reappear and followed by the Standby screen.
6. Press the Power Key to return to normal operation.
Display Module Firmware Updates
CAUTION: The current version of firmware as well as all time/temperature programs, calibrations,
and other user-set values are erased when new Display Module Firmware is installed. Be sure to
save any programs you wish to keep to a flash drive and record any settings that you have changed.
The Advanced Programmable Controller provides you with the ability to update its firmware should the
need arise. A firmware update is performed as follows:
1. Load the updated firmware (the file name will be: *.CMD where * = version number) in the root
directory of a flash drive. The flash drive should not contain any other versions of the firmware.
2. Turn power to the Controller OFF by pressing the Power Key on the front of the Controller.
3. Place the Power Switch/Circuit Breaker on the rear of the unit in the OFF position.
4. Insert the flash drive in the USB A port on the back of the Controller.
5. Press and hold the SET key while placing the Power Switch/Circuit Breaker to the ON position.
An hour glass image will appear on the display.
6. Continue holding the SET key until the display changes from black to white. Release the SET key
when the message “Checking Image…This process takes several minutes, please wait” appears.
7. When “Standby” appears on the display, place the Power Switch/Circuit Breaker in the OFF
position.
8. Return the Power Switch/Circuit Breaker to the ON position.
9. When “Standby” appears on the display, turn the Controller ON by pressing the Power Key on the
front panel.
10. Press the Menu Key to access the Main Menu, select Service, and then verify that the Display
Module Firmware version displayed on page 2 of the Service sub-menu matches the one you just
loaded.
11. Return to the Main Menu, select Control, and then select Reset Memory and Reboot on page 2 of
the Control sub-menu.
12. Enter the Restore Defaults password (1234) and touch OK. The Temperature Controller will
power down and then reboot in standby.
NOTE: See System Restoration for information on restoring the Temperature Controller’s original
version of firmware.
Warnings alert you to conditions such as the disconnection of the external temperature sensor or that the
desired set point is above or below the user-set limits. The warning message appears briefly on the status
bar when the condition is detected. It also appears when power is turned Off and then back On.
Warning Message Cause Corrective Action
The set point is above the high limit Desired set point temperature is
above High Limit setting
The set point is below the low limit Desired set point temperature is
below Low Limit setting
Probe 2 has been disconnected External temperature probe
disconnected
The set point is above the safety set
point
Service air filter and reset Air filter cleaning maintenance
The fluid should be replaced Bath fluid replacement
The fluid is above the high alarm Bath fluid temperature above High
The fluid is below the low alarm Bath fluid temperature is below
Desired set point temperature is
above the safety set temperature
reminder
maintenance reminder
Alarm setting
Low Alarm setting
Lower set point temperature or
increase High Limit setting
Raise set point temperature or
decrease Low Limit setting
Reconnect external temperature
probe or select internal (P1) control.
Lower set point or increase safety
set temperature
Clean air filter and reset service
reminder
Replace bath fluid and reset service
reminder
Allow bath fluid to cool or increase
High Alarm setting
Allow bath fluid to warm or
decrease Low High Alarm setting
Faults alert you to conditions which may compromise safety or component failures which affect
performance. When a fault condition is detected, power to the heater, condenser, and pump are
disconnected and a fault message appears continuously on the display. The fault message includes the
type of fault, cause, and corrective action.
Fault Messages Cause Corrective Action
COOLING MODULE
DISCONNECTED
EXTERNAL PROBE FAILURE
Note: This fault will only appear if
External (P2) Control is active.
HIGH ALARM
HIGH SAFETY ALARM
INTERNAL FAILURE An internal failure may have
INTERNAL PROBE FAILURE The internal temperature probe
LOW ALARM The fluid temperature is lower than
LOW LIQUID LEVEL.
LOW LIQUID LEVEL DETECTED Low liquid level has been detected
The refrigeration control module,
detected when the unit was last
powered, is no longer present.
The external temperature probe
may have failed.
The fluid temperature exceeds the
High Alarm setting. Fluid is too
warm.
The fluid temperature exceeds the
Safety Set.
occurred.
may have failed.
the Low Alarm. Fluid is too cold.
The float switch has activated.
through either activation of the float
switch or the detection of an overtemperature condition.
Reconnect the refrigeration control
cable and re-power unit.
Check that the external
temperature probe is plugged in
correctly. Inspect probe for damage
such as frayed or exposed wires. If
problem persists, please contact
your supplier.
Decrease fluid temperature or
increase High Alarm setting. If
problem persists, please contact
your supplier.
Check fluid level. Check fluid
temperature and set point. Ensure
that Safety Set is higher than the
fluid set point. If problem persists,
please contact your supplier.
Check your set up. If problem
persists, please contact your
supplier.
If problem persists, please contact
your supplier.
Increase temperature set point or
decrease Low Alarm setting. . If
problem persists, please contact
your supplier.
Check the fluid level in the bath,
and add more fluid if required.
Recheck your fluid level, and
recheck your alarm settings.
1. Maximum operating temperature at which ±0.005°C temperature stability can be maintained; Advanced Programmable Controller is capable of
higher temperatures.
2. Maximum operating temperature for stainless steel tank; Advanced Programmable Controller is capable of higher tem peratures.
3. Maximum operating temperature for polycarbonate tank; Advanced P rogrammable Controller is capable of higher temperatures.
Ambient +10° to 85°C
Ambient +20° to 185°F
Ambient +10° to 85°C
Ambient +20° to 185°F
Ambient +10° to 85°C
Ambient +20° to 185°F
Ambient +10° to 85°C
Ambient +20° to 185°F
Ambient +10° to 85°C
Ambient +20° to 185°F
-22° to 392°F
(3)
(3)
(3)
(3)
(3)
120V, 60Hz,
10A
120V, 60Hz,
10A
120V, 60Hz,
10A
120V, 60Hz,
10A
120V, 60Hz,
10A
13A
240V, 50Hz,
10A
240V, 50Hz,
10A
240V, 50Hz,
10A
240V, 50Hz,
10A
240V, 50Hz,
10A
13A
Environmental Conditions Indoor use only
Maximum Altitude: 2000 meter
Operating Ambient: 5° to 35°C (41° to 95°F)
Relative Humidity: 80%, non-condensing
Installation Category: II
Pollution Degree: 2
Ingress Protection: IP 31
Climate Class: SN
Software Class: B
Output Waveform: Sinusoidal
Depending on your needs, a variety of fluids can be used with your Circulator. No matter what bath fluid is
selected, it must be chemically compatible with the reservoir and the materials in your Circulator. It must
also be suitable for the desired temperature range.
WARNING: When using Class III flammable fluids per DIN 12876-1, the user must attach the
following warning labels to the front of the unit so that they are well visible:
Warning Label
W09
Colors: Yellow/black
Mandatory Label
M018
Colors: Blue/white
Or
Semi S1-0701
Table A1-2 #9
Colors: Blue/white
WARNING: Always use fluids that satisfy safety, health, and equipment compatibility requirements.
Be aware of the chemical hazards that may be associated with the bath fluid used. Observe all safety
warnings for the fluids used as well as those contained in the material safety data sheet.
Danger Area.
Attention! Observe instructions
(operating manual, safety data sheet)
Carefully read the user information prior to
beginning operation.
Scope: EU
Carefully read the user information prior to
beginning operation.
Scope: NAFTA
For optimum temperature stability, the fluid’s viscosity should be 50 centistokes or less at its lowest
operating temperature. This permits good fluid circulation and minimizes heating from the pump.
For temperatures from 10°C to 90°C, distilled water is recommended. For temperatures below 10°C, a
mixture of laboratory grade ethylene glycol and water should be used. Do not use deionized water.
The following chart is intended to serve as a guide in selecting a bath fluid for your application. For
optimum temperature stability and low vaporization, be sure to stay within the fluid's normal temperature
range.
CAUTION: Always turn electrical power to the Circulator OFF before making a connection to the
serial (DB9) port.
Serial Connector — A DB9 connector is provided on the back panel of the Controller for RS232/RS485
data communication.
Pin RS232 RS485
1
2 TX A
3 RX
4 DTR
5 GND GND
6 DSR
7 RTS
8 CTS B
9
Serial Communications Protocol — The Controller uses the following serial communications settings:
Data bits — 8
Parity — None
Stop bits — 1
Flow control — None
Baud rate — Selectable (Controller/PC baud rates must match). 57600 is recommended.
Communications Commands — RS232 commands must be entered using the command only; RS485
commands must be entered using @ and the RS485 address (E.g., @001) followed by the actual
command. All commands must be entered in the exact format shown. Do not send a [LF] (line feed) after
the [CR] (carriage return). Be sure to follow character case exactly.
A response followed by an exclamation point (!) indicates that a command was executed correctly. A
question mark (?) indicates that the Controller could not execute the command (either because it was in
an improper format or the values were outside the allowable range). A response must be received from
the Controller before another command can be sent. All responses are terminated with a single [CR].
WARNING: Always turn electrical power to the Circulator OFF before making a connection to the
serial (DB9) port.
The DB9 connector on the rear of the Temperature Controller can also be used to turn the Circulator ON
or OFF via a contact closure or VDC control room signal.
Contact Closure — To use a contact closure to turn the Circulator ON and OFF, connect pin 1 to 5. The
Circulator will be ON when the switch is closed and OFF when the switch is open.
VDC Signal — To use a VDC signal to turn the Circulator ON and OFF, connect to pins 9 (Positive) and 5
(Ground). The Circulator will be ON when a +9 to +24 VDC signal is applied; it will be OFF when the
signal is removed.
USB Data Logging
Data is stored in a CSV file that can be read in spreadsheet programs such as Microsoft Excel®. The data
is output in the following format:
MIN.SEC [V] UNIT SET P1 P2
4 [v0013] C 10 15.19 14.89
4 [v0013] C 10 15.18 14.88
4.22 [v0013] C 10 15.03 14.71
4.23 [v0013] C 10 15.02 14.7
4.24 [v0013] C 10 15 14.69
4.25 [v0013] C 10 15 14.69
4.26 [v0013] C 10 14.99 14.68
4.27 [v0013] C 10 14.98 14.67
MIN.SEC = Minutes and seconds.
V = Firmware version.
UNIT = Selected temperature unit.
SET = Temperature set point
P1 = Fluid temperature as measured by the Internal probe
P2 = Fluid temperature as measured by the External probe (if connected)
NOTE: The time stamp on the first line of data represents the time when data began being logged. The
time stamp will indicate 0.01 if data logging was enabled after a flash drive or computer was connected
to the USB port. It will show a later time if data logging was enabled before the storage device was
connected. In the example shown above, the flash drive was connected to the USB port 4 minutes after
data logging was enabled.
The Advanced Programmable Temperature Controller can be monitored and controlled using a personal
computer connected to its USB B port.
Initial Setup
1. Plug the USB A connector of a USB A to USB B cable into an open USB port on the personal
computer and plug the USB B connector into the corresponding port on the Temperature
Controller.
2. The computer operating system will automatically detect a new device and ask to install the
drivers for the device. The drivers are located on the Resource Disk that came with your
Temperature Controller.
3. Place the Resource Disk in the computer’s CD drive and install the drivers.
4. Determine the identity of the communications port your computer assigned to the Temperature
Controller.
Monitoring and Control
When all connections have been made and drivers installed, you can monitor and control the
Temperature Controller using a terminal program and the active serial communications commands.
Ethernet Configuration
NOTE: The Advanced Programmable Controller’s Ethernet capabilities may be used with any of the
following web browsers: Internet Explorer 9.0 or later, Firefox, Chrome, Safari, or Opera.
The Advanced Programmable Controller may be connected directly to a laptop or desktop computer via
its Ethernet connection or indirectly via your facility’s wired or wireless network. Either type of connection
enables you to control and/or monitor the operation of your circulating bath in real time using an Internet
browser by entering its IP address. The Controller’s active serial communications commands are used to
retrieve and/or change operational information.
Direct Computer to Controller Configuration
NOTE: If your computer is already connected to a wired network, you will not be able to directly
connect it to your circulating bath’s Temperature Controller.
1. Open the utility that displays your computer’s available network connections.
2. Select an available wired connection.
3. Enter the following properties for the selected connection:
A. Internet protocol: TCP/IP
B. IP address: any three sets of identical numbers (e.g., 111.111.111) followed by a single
digit (e.g., 111.111.111.5).
C. Subnet mask: 255.255.255.0
4. Enter the IP address into your Controller as outlined in Static IP Addressing, below. Please note
that the first three series of numbers in IP address on the Controller must match those entered in
Step 3B, above. The last number does not have to match.
The Advanced Programmable Controller supports both dynamic (DHCP) and static IP configuration. The
default value is Static. With either type of IP configuration, it may be necessary to work with your IT
department to ensure that both the network and your Controller are configured properly.
1. Press
2. Touch Connect to access the Connect sub-menu.
to access the Main Menu.
3. Touch Ethernet to access the IP address screen. Depending on the current configuration, one
of the following screens will appear:
Static IP address screen Dynamic IP address screen
Static IP Addressing — If your Controller will be connected directly to a computer or your network uses
static IP addressing, you must manually enter the IP address that will be used into your Controller. This is
done as follows:
1. Select Static and touch √.
2. Select an IP address field, enter the correct IP address, and touch
IP address fields.
Touch to create a
password allowing control
using an internet browser
Dynamic IP Addressing — If your network uses DHCP (Dynamic Host Configuration Protocol), all you
need to do is set your Controller for DHCP. The correct address will appear automatically when you
connect the Controller to the network.
TCP Control Configuration
In order to make operational changes (set point, high/low limits and alarms, etc.) through an internet
browser, the TCP Control function must be ON and a password set on the Temperature Controller’s IP
settings screen. You must enter this password in the browser screen’s Login field before you are allowed
to change the Temperature Controller’s operational settings.
TCP control is turned ON and a password created as follows:
1. Select Ethernet from the Connect sub-menu. The IP settings screen will appear:
2. Select Password, create a password (8 characters maximum), and touch √.
CAUTION: External Cooling Control settings are always displayed and set in °C.
NOTE: The Circulator must be equipped with the optional external cooling control valve in order to use
this function. Refer to the instructions included with the optional External Cooling Control Valve for
installation information.
External cooling control provides you with a means of quickly cooling the bath fluid after a period of
operation at an elevated temperature. When properly configured, it also prevents the bath from cooling to
a temperature below that which is intended.
General Operation
External cooling control allows you to initiate a controlled cool down of the bath fluid after operation at an
elevated temperature by automatically and incrementally closing the valve as the bath temperature
decreases.
1. When the bath temperature is above the Maximum Set value, the external cooling valve is fully
open (100%).
2. When bath temperature is between the Maximum Set and Minimum Set values, the external
cooling valve is partially open (user-adjustable from 10% to 90%).
3. When the bath temperature is below the Minimum Set value, the external cooling valve is fully
closed (0%).
Configuring External Cooling Control
1. Select External Cooling Control from the second page of the Control sub-menu. The External
Cooling Control setup screen will appear.
2. Enter the desired Maximum Set, Minimum Set, and Valve % values for the external cooling valve.
Select a value to change by touching the corresponding button. Enter the desired value using the
numeric keypad that appears and then touch
Maximum Set — At bath temperatures above this value, the valve will be open 100%. The
Maximum Set value can be set from 40° to 80°C.
Minimum Set — At bath temperature below this value, the valve will be fully closed. The
Minimum Set value can be set from 20° to 35°C.
to accept the new value.
Valve % — This is the valve setting when bath temperature is between the Maximum Set and
Minimum Set values. The Value % can be set from 10% to 90% open.
3. Touch √ to accept the new external cooling control settings.
Operating External Cooling Control
CAUTION: Be sure to deactivate external cooling control once bath temperature has cooled to the
desired temperature. Failure to do so may result in excessive energy consumption and/or make it
difficult to maintain a bath temperature that is higher than the Minimum Set value.
To initiate a cool down using external cooling control, select External Cooling Control from the second
page of the Control sub-menu, touch Start and then
. The external cooling control valve will open (fully
or partially, depending on bath temperature and settings). The external cooling control valve will
automatically close once the bath temperature reaches the Minimum Set value.
When the bath has cooled to the desired temperature, external cooling control can be deactivated by
selecting External Cooling Control from the Control sub-menu, touching Stop, and then
.
Equipment Disposal (WEEE Directive)
or
This equipment is marked with the crossed out wheeled bin symbol to indicate it is covered by the Waste
Electrical and Electronic Equipment (WEEE) Directive and is not to be disposed of as unsorted municipal
waste. Any products marked with this symbol must be collected separately, according to the
regulatory guidelines in your area.
It is your responsibility to correctly dispose of this equipment at lifecycle-end by handing it over to an
authorized facility for separate collection and recycling. It is also your responsibility to decontaminate the
equipment in case of biological, chemical and/or radiological contamination, so as to protect the persons
involved in the disposal and recycling of the equipment from health hazards. By doing so, you will help to
conserve natural and environmental resources and you will ensure that your equipment is recycled in a
manner that protects human health.
Requirements for waste collection, reuse, recycling, and recovery programs vary by regulatory authority
at your location. Contact your local responsible body (e.g., your laboratory manager) or authorized
representative for information regarding applicable disposal regulations.
If you have followed the troubleshooting steps outlined previously and your Circulator still fails to operate
properly, contact the supplier from whom the unit was purchased. Have the following information
available for the customer service person:
Model, Serial Number, and Voltage (from back panel label)
Date of purchase and purchase order number
Supplier’s order number or invoice number
A summary of the problem
Warranty
The manufacturer agrees to correct for the original user of the product, either by repair (using new or
refurbished parts), or at the manufacturer’s election, by replacement (with a new or refurbished product),
any defects in material or workmanship which develop during the warranty period. The standard warranty
is twenty-four (24) months after delivery of the product. In the event of replacement, the replacement unit
will be warranted for the remainder of the original warranty period or ninety (90) days, whichever is
longer. For purposes of this limited warranty, “refurbished” means a product or part that has been
returned to its original specifications. In the event of a defect, these are your exclusive remedies.
If the product should require service, contact the manufacturer’s/supplier’s office for instructions. When
return of the product is necessary, a return authorization number is assigned and the product should be
shipped, transportation charges pre-paid, in either its original packaging or packaging affording an equal
degree of protection to the indicated service center. To insure prompt handling, the return authorization
number must be placed on the outside of the package. A detailed explanation of the defect should be
enclosed with the item.
The warranty shall not apply if the defect or malfunction was caused by accident, neglect, unreasonable
use, improper service, acts of God, modification by any party other than PolyScience, or other causes not
arising out of defects in material or workmanship.
EXCLUSION OF IMPLIED WARRANTIES. THERE ARE NO WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING, BUT NOT LIMITED TO, THOSE OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE WHICH EXTEND BEYOND THE DESCRIPTION AND PERIOD AS STATED
IN THE OPERATOR’S MANUAL INCLUDED WITH EACH PRODUCT.
LIMITATION ON DAMAGES. THE MANUFACTURER’S SOLE OBLIGATION UNDER THE WARRANTY
IS LIMITED TO THE REPAIR OR REPLACEMENT OF A DEFECTIVE PRODUCT AND POLYSCIENCE
SHALL NOT, IN ANY EVENT, BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES
OF ANY KIND RESULTING FROM USE OR POSSESSION OF THIS PRODUCT.
Some states do not allow: (A) limitations on how long an implied warranty lasts; or (B) the exclusion or
limitation of incidental or consequential damages, so the above limitations or exclusions may not apply to
you. This warranty gives you specific legal rights and you may have other rights that vary from state to
state.