Philips Skylight User manual

Skylight Camera
Site Planning
9347-0111 Rev C
Sec 1: General Requirements ....1-1
Sec 2: Room Layouts .................2-1
Sec 3: Anchor Details .................3-1
Sec 4: Networking.......................4-1
Sec 5: Shipping Information........5-1
Sec 6: Floor Loading...................6-1
Philips Medical Systems 540 Alder Dr. Milpitas, CA 95035
Warranty Disclaimer
Philips provides this document without warranty of any kind, either implied or expressed, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Limitation Of Liability
Philips has taken care to ensure the accuracy of this document. However, Philips assumes no liability for errors or omissions and reserves the right to make changes without further notice to any products herein to improve reliability, function, or design. Philips may make improvements or changes in the product(s) or program(s) described in this document at any time.
About This Document
This document provides site planning information for customers planning to purchase a Skylight camera and their facility engineers, structural engineers, site planners and architects.
CAD Drawings
Drawings in a DWG format for the Skylight system are downloadable from the following website for architects planning room layouts:
http://apps1.medical.philips.com/documents
Copyright Notice
© August 15, 2003 Koninklijke Philips Electronics N.V. All rights reserved.
i
Section 1
General Requirements
General Information
Figure 1-1 on page 1-3 illustrates the Skylight components described below. The Skylight system consists of an overhead frame supported by four 4" diameter
steel posts. The system images patients as they lie on a patient table inside the frame. A carriage translates along the overhead frame
and supports two image-collecting detectors. The detectors can move up and down, in and
out, and rotate around their own axis. This allows the detectors to image patients from above, below, or from the sides of a patient.
Prior to imaging a patient, nuclear medicine technologists attach a pair of “collimators” on the detectors. Collimators are 21" x 26" (53cm x 66cm) sheets of lead. Several different types of collimators exist; collimators weigh between 110 and 260 lbs (50 and 118 kg).
When a pair of collimators are not on the detectors, technologists will store the collimators in a floor-mounted, collimator exchanger on one side of the Skylight frame.
The system also includes two “towers” containing electronic components: the PC tower and the power tower.
The PC tower contains image processing and networking equipment. The power tower contains an uninterruptible power supply (UPS) and other power
supplies.
Collimator
Detector
In/Out
Up/ Down
The PC tower, power tower, collimator exchanger and vanity covers fill one side of the Skylight frame.
1-1
Section 1: General Requirements
Clinical Operation
Although not shown in Figure 1-1, a desktop processing station (a PegBlade) must be located outside the frame, but preferably in the same room. From the PegBlade, physicians view and interpret images produced by the system.
Patients to be imaged by the Skylight system receive a small intravenous injection of a gamma-ray-emitting radionuclide. The collimators contain a pattern of gamma-ray admitting holes perpendicular to the imaging surface of the detectors. The detectors sense the radiation that passes through holes in the collimators.
The detectors transmit electronic data to image processors in the PC tower. The image processors produce images for physician interpretation on the PegBlade processing station.
The Skylight system performs these types of clinical studies:
Emission Computed Tomography (ECT) -- the carrriage remains in one position while the detectors rotate around the patients torso. Image processors produce a set of three dimensional tomographic images. The images show radionuclide concentration in parallel slices of the patient’s internal organs.
Total Body -- the carrriage translates along the length of the patient (usually head to toe). One detector scans above the patient table; the other scans below patient table. The image processors produce two images (anterior and posterior) showing radionuclide concentration in the patients skeleton and other tissue.
Static (Spot) or Dynamic -- the carriage and detectors remain fixed positions near the patient. The system produces either a single image (static), or a sequence of images (dynamic) showing the flow of radionuclide within the patient over time.
1-2
Frame
Frame Support Post (1 of 4)
Carriage Motion
Carriage
Section 1: Room Requirements
Power Tower
Exchanger Support Post
Detector #1
Patient Table
Note: Covers around front posts and filling all open spaces above Power Tower, PC Tower and Exchanger are not shown.
Detector #2 (see detail)
Collimator
Detector
Collimator Exchanger
Vanity Cover
PC Tower
In/Out
Up/ Down
Detector #2 Detail
(Detector #1 is similar)
Figure 1-1. Skylight Camera System.
1-3
Section 1: General Requirements
Collimator Exchanger
Technologists store collimators in vertical drawers of the collimator exchanger. To transfer a collimator from an exchanger drawer to a detector, an operator will
swing the drawer 90° into the room, translate the carriage to the exchanger, and move a detector next to a drawer as Figure 1-2 illustrates.
All Skylight systems come with a two-drawer collimator exchanger. That exchanger will hold four pairs of collimators.
Customers may purchase a second, one-drawer collimator exchanger. The one­drawer exchanger will hold two more pairs of collimators. As Figure 1-3 shows, the one-drawer collimator exchanger resides on the same wall as the two-drawer exchanger.
Skylight Models
The Skylight system comes in one of four models:
a 9 foot Skylight,
a 9 foot California (seismic) Skylight,
an 8 foot Skylight,
an 8 foot California (seismic) Skylight.
The California (seismic) models have thicker-walled steel posts and three top braces above the Skylight frame.
Vibration Specifications
Nuclear medicine cameras do not have floor vibration specifications. This is because:
image collection durations are long (10 - 300 seconds, or more) and floor vibration durations are much smaller,
vibrations are typically sinusoidal and, therefore, tend to cancel out, and
the patient table and detector/gantry assemblies are both floor mounted and
tend to vibrate together.
1-4
Section 1: General Requirements
Collimator exchanger drawer swung 90° around exchanger support post to transfer collimators from or to Detectors
Detector #2 in position to transfer a collimator to or from the collimator exchanger
Figure 1-2. Collimator Exchange Operation.
Exchanger Support Post (two drawer)
Two-Drawer Collimator Exchanger
One-Drawer Collimator Exchanger (optional)
Exchanger Support Post (one drawer)
Figure 1-3. One-Drawer and Two-Drawer Collimator Exchangers
1-5
Section 1: General Requirements
Power Requirements
The power requirements for a Skylight camera and PegBlade are as follows:
Skylight Camera PegBlade
Input Voltage: 208 - 240 V - dedicated 120 V (International) 208 - 240 V - dedicated 220 V
Peak/Quiescent Current: 15 A / 4.8 A 5 A / NA (International) 15 A / 4.8 A 2.5 A / NA
Phase: Single Phase Single Phase Receptacle: Hardwired NEMA 5-15 Note: The Skylight camera includes a UPS and provides power to the detectors,
carriage, patient table, towers, acquisition terminal and collimator exchanger. An optional cardiac gate can plug into a 115 volt receptacle on the patient table.
Air Conditioning Requirements
The system produces these heat loads:
Skylight System PegBlade PegBlade Monitor
5,653 BTU/hr 854 BTU/hr 544 BTU/hr 396 Cal/sec 60 Cal/sec 38 Cal/sec 1,656 watts 250 watts 159 watts
The camera room HVAC system must maintain the temperature between 60°- 75° F (16° - 24° C) with less than 10° F (6° C) variation per hour. Humidity must be between 20% - 75%.
These requirements must be met on a 24 hours per day, 7 days per week basis.
Magnetic Field Limitations
To avoid image quality and video monitor problems, the Skylight system must not be in an area with a magnetic field greater than 1 Gauss.
Floor Levelness
The floor under the four Skylight frame posts, the collimator exchanger post(s) and the patient table must all be at the same level ±1".
1-6
Room Layout
Room Size Requirements
Section 2
9 Ft Skylight W x D x H _ W x D x H _ Minimum Size 12'-6" x 14' x 9' 12'-6" x 14' x 8'
381cm x 427cm x 274cm 381cm x 427cm x 244cm Recommended Size 16'-7" x 16'-10" x 9' 16'-7" x 16'-10" x 8'
505cm x 513cm x 274cm 505cm x 513cm x 244cm Figure. 2-1 illustrates a minimum room size. In such rooms, the site must provide a
separate, nearby room for the PegBlade processing station. Figures 2-2 and 2-3 illustrates two recommended room sizes. Such rooms facilitate
patient access and technologist/physician convenience. To keep frame support posts close to walls, Philips installers can extend one of the frame support posts from 20" to 36" in either the X or Y direction from the normal post position (Figure 2-2 & 2-3).
8 Ft Skylight
Equipment Sizes (9 ft and 8 ft Skylights)
W x D x H Wt
Skylight System1 11'-10" x 13'-6" see Fig 2-4 & 2-5 see pg 6-3 361cm x 411cm
Patient Table (minimum height) 2'-5" x 9'-6" x 1'-9" 320 lbs
74cm x 290cm x 53 145 kg
PegBlade CPU 18" x 18" x 5" 27 lb (placed on customer’s desktop) 46cm x 46cm x 13cm 12 kg
PegBlade monitor 20" x 20" x 20" 69 lb (placed on customer’s desktop) 51cm x 51cm x 51cm 32 kg
Codonics Printer (option) 17" x 21" x 12" 55 lbs (placed on customer’s desktop) 43cm x 53cm x 30cm 25 kg
Important: Unless the floor is a reinforced concrete slab poured on grade, a licensed structural engineer must
_______
1
At installation, Philips must lay a 11'-9.3" x 13'-5.4" rectangular metal drill template on the floor (see Section3). There must be no objects at floor level to interfere with this template.
evaluate floor loading capacity. See Section 6.
2-1
Section 2: Room Layout
Minimum (12'-6" x 14') Room Restrictions:
• Anchor plate for frame support posts placed 1" from wall on side of Two-Drawer Collimator Exchanger (instead of more desireable 3" placement),
• Acquisition Terminal placed outside of room,
• Processing Station placed in separate, nearby room,
• Display Panel and Boom tied in one position (instead of allowing desireable free rotation of boom),
• Doors cannot swing into room.
12'-6" (381cm)
From Main Electrical Panel
1" (3cm)*
4" (10 cm)*
Collimator Exchanger:
Drawer #1 & Drawer #2
J
Power Tower
11'-10" (361cm)
Two Drawer:
6'-9" (206 cm)
PC Tower
Preferred side
of table for
patient access
4" (10 cm)*
Anchor Plate for Frame Support Post (4 places)
Opening must be curtain, sliding door or pocket door which does not open into Skylight frame.
C
L
Frame
14' (427cm)
13'-6" (411cm)
Patient Table
6'-9" (206 cm)
5'-11" (180 cm)
5"
(13 cm)*
* For a 9' Seismic Skylight, all anchor plates must be at least 8" from all walls. For an 8' all anchor plates must be at least 6" from all walls.
J
Electrical Junction Box for hardwiring to 10' Skylight cord
Circuit Breaker, 208-240 VAC, 30 Amp, single phase
120 VAC, 15 Amp, NEMA 5-15
4" (10 cm)*
Seismic Skylight,
Touch Screen & Boom (tied in one position)
5'-11" (180 cm)
C
L
Frame
Preferred side
of table for
operator
Acquisition Terminal
Frame
Carriage
Scale: 1/4" = 1'
4" (10 cm)*
Processing Station: Customer provided desktop; Philips provided PegBlade CPU, monitor & keyboard
Twin RJ-45 Ethernet Connector Box Telephone line for modem
Figure 2-1. Skylight in Minimum Size Room and PegBlade in Separate Room.
2-2
Scale: 1/4" = 1'
From Main Electrical Panel
5'-11" (180 cm)
J
Drawer #1 Drawer #2
Section 2: Room Layout
16'-7" (505cm)
3" (8 cm)
5'-11" (180 cm)
6'-9" (206 cm)
C
L
Frame
6'-9" (206 cm)
4" (10 cm)
Power Tower
Movement of Exchanger Drawers
Movement of Touch Screen and Boom
Patient Table
Acquisition Terminal
Two-Drawer
Collimator Exchanger
Carriage
6'-2" (188 cm)
C
Frame
L
12' -1" (368 cm)
PC Tower
Preferred side of table for patient access
Preferred side
of table for
operator
Frame
Processing Station:
Customer provided desktop. Philips provided PegBlade, monitor/keyboard.
Frame "X" Extension (20" to 36" post-centerline to post-centerline)
"Moved" Frame Support Post
48"
(122 cm)
49"
(124 cm)
16'-10" (513cm)
16'-6" (503cm) max
9'-9" (297 cm) max
1" (3 cm)
J
Electrical Junction Box for hardwiring to 10' Skylight cord Circuit Breaker, 208-240 VAC, 30 Amp, Single Phase
115 VAC, 15 Amp, NEMA 5-15
Twin RJ-45 Ethernet Connector Box
Telephone line for modem
Figure 2-2. Skylight in Recommended Room #1.
2-3
Scale: 1/4" = 1'
Section 2: Room Layout
From Main Electrical Panel
6'-9" (206 cm)
6'-9" (206 cm)
5'-11" (180 cm)
J
Drawer #1 Drawer #2
Power Tower
Movement of Exchanger Drawers
Movement of Touch Screen and Boom
Patient Table
15'-2" (462cm)
Frame
C
L
5'-11" (180 cm)
Two-Drawer
Collimator Exchanger
Carriage
Preferred side of table for patient access
PC Tower
Preferred side
of table for
operator
Frame
"Moved" Frame Support Post
3" (8 cm)
C
L
Frame
18'-1" (551cm)
13'-9" (420cm)
7'-0" (215cm)
Processing Station:
Customer provided desktop. Philips provided PegUltra, monitor/keyboard.
Acquisition Terminal
8'-10" (270 cm) max
14' -9" (450cm) max
4" (10 cm)
J
Electrical Junction Box for hardwiring to 10' Skylight cord
Circuit Breaker, 208-240 VAC, 30 Amp, Single Phase 115 VAC, 15 Amp, NEMA 5-15
Twin RJ-45 Ethernet Connector Box Telephone line for modem
Figure 2-3. Skylight in Recommended Room #2.
Frame "Y" Extension (20" to 36" post-center to post-center)
4'-1"
(124 cm)
4'
(122 cm)
1" (3 cm)
2-4
Finished Ceiling
Frame Top Frame Bottom
Notes: Post height is adjustable ±1.11" to allow for uneven finish floor.
Some carriage components extend 1.5" above frame.
Section 2: Room Layout
Worst case clearance = 108" - (103.86" + 1.11" + 1.50") = 108" - 106.47" = 1.53"
Frame
Gusset
Power Tower
Collimator Exchanger
8'-7.86" ±1.11"
9' minimum
8' ±1.11"
Support Post
Patient Table
Detector
Finished Floor
Right Isometric
View
Detector
Section AA
Elevation (Side) View
Scale: 1/4" = 1'
A A
Plan (Top) View
Figure 2-4. Elevation Details, 9 ft Skylight
(non-California)
2-5
Finished Ceiling
Frame Top Frame Bottom
Section 2: Room Layout
Worst case clearance = 96" - (91.86" + 1.11" + 1.50") = 96" - 94.47" = 1.53"
Notes: Post height is adjustable ±1.11" to allow for uneven finish floor.
Some carriage components extend 1.5" above frame.
Finished Floor
Right Isometric
View
7'-7.86" ±1.11"
8' minimum
7' ±1.11"
Support Post
Patient Table
Frame
Power Tower
Detector
Detector
Section AA
Elevation (Side) View
Scale: 1/4" = 1'
Gusset
Collimator Exchanger
A A
Plan (Top) View
Figure 2-5. Elevation Details, 8 ft Skylight
(non-California)
2-6
Finished Ceiling Frame Top
Frame Bottom
Notes: Post height is adjustable ±1.11" to allow for uneven finish floor.
Section 2: Room Layout
Worst case clearance = 108" - (106.74" + 1.11") = 108" - 107.85" = 0.15"
Frame
Gusset
Power Tower
Collimator Exchanger
8'-10.74" ±1.11"
9' minimum
8' ±1.11"
Support Post
Patient Table
Seismic Top Brace (1 of 3)
Detector
Finished Floor
Right Isometric
View
Detector
Section AA
Elevation (Side) View
Scale: 1/4" = 1'
49"32"32"49"
A A
Seismic Top Brace (1 of 3)
Plan (Top) View
Figure 2-6. Elevation Details, 9 ft Skylight
(California Seismic)
2-7
Loading...
+ 33 hidden pages