FIELD CHANGE ORDER Service
Issued by : Released by : Publication No. :
e
r
APPLIES TO:
Tube assembly combinations RO / SRO and SRM tubes with tube housings ROT 350, ROT 351, ROT 500, ROT 501, ROT-GS 501, ROT 504, ROT-GS 505, ROT 507 with the following type number s :
9874 001 16122 9874 006 11122 9874 094 00362 9890 000 03951 9890 000 63901 9874 002 05122 9874 006 22122 9874 094 00462 9890 000 03961 9890 000 63911 9874 002 07122 9874 006 23112 9874 094 00482 9890 000 63181 9890 000 63931 9874 002 08112 9874 007 01122 9890 000 03362 9890 000 63661 9890 000 63941 9874 004 21122 9874 008 46122 9890 000 03432 9890 000 63671 9890 000 65122 9874 004 23122 9874 008 47122 9890 000 03442 9890 000 63721 9890 000 85001 9874 004 24112 9874 008 60122 9890 000 03562 9890 000 63731 9890 000 85181 9874 005 08122 9874 008 63132 9890 000 03621 9890 000 63751 9890 000 85251 9874 005 09122 9874 008 64132 9890 000 03821 9890 000 63761 9874 005 16122 9874 008 66122 9890 000 03841 9890 000 63841
PMS Hamburg DMC GmbH D. Morgenstern 4512 980 60031
e
c
i
v
r
e
s
e
m
m
o
c
n
d
a
t
i
signed
n
o
ROT 350, ROT 351, ROT 500, ROT 501,
ROT-GS 501, ROT 504,
ROT-GS 505, ROT 507
FCO74000002 March, 2003 740
Tube assembly
TITLE:
Loose fixing screws on the supporting bridge of the tube housing of the tube assembly
LIST OF PAGES & DRAWINGS:
1…4 (03.0)
FCO74000002 1/4
ALL RIGHTS RESERVED
Copyright © 2003 Philips Medical Sy stems
INTRODUCTION: Symptom :
Problems with the radiation geometry during operation of the tube assembly. The centering of the central beam does not coincide with the geometry of the imaging system (film or, respectively, image intensifier).
Cause :
The tube assembly is centered with respect to the system via the supporting bridge of the tube housing. The required stability to keep the tube assembly in position with respect to the central beam is provided by especially specified fixing screws (1). See sketch below:
1
2
Complaints from the field and inspections of returned tube assemblies have revealed that sometimes the fixing screws of the supporting bridge of the tube housing were loose. It was also found out that unspecified lengths of screws were used in the field which caused damage of the tube housing.
Remedy :
Check of whether the fixing screws are tight and, if necessary, retightening of the fixing screws at least once a year. This work can be carried out during regular maintenance or any other service action. To fix the retaining rings of the supporting bridge of the tube housing use six cheese-head screws
Code number screw M5 x 12: 2522 007 15036 Code number spring washer M5: 2512 767 06004
The supporting bridge of the tube housing is fixed at the system side with four screws M6 x _ _ Observe the inserted screw length (7.3 mm).
1
2
(2) with hexagon socket and conical spring washers.
(1) and spring washers. The length of the screws depends on the system.
2/4 FCO74000002