1.6.3 Electrical Data .................................................................................................................................10
1.6.5 Environmental Conditions for Transport and Storage ......................................................................11
1.6.6 Performance Data ...........................................................................................................................12
1.7 Compliance information ......................................................................................................................13
1.8 Conditions of Acceptability ..................................................................................................................14
1.9 Abbreviations and definitions ..............................................................................................................14
1.10 Manual history .................................................................................................................................15
4.1 General ...............................................................................................................................................28
4.2 Replace (complete) BLD on the stand.................................................................................................28
4.13 Replace Micro-switch including Bracket...........................................................................................37
4.14 Replace Fan ....................................................................................................................................38
NICOL X-Ray Beam Limiting Device Introduction and Technical Data
Table 6: CAN Interface
LA2X1
LA2X11
Mnemonic Description
1 - reserved
2 CAN-L CAN-L bus line (dominant low)
3 0VCAN 0V of CAN supply voltage
4 - reserved
5 - reserved
6 0VCAN 0V of CAN supply voltage
7 CAN-H CAN-H bus line (dominant high)
8 RSTCAN Reset-line CAN
9 +12VCAN +12V CAN supply voltage
Introduction and Technical Data NICOL X-Ray Beam Limiting Device
1.6.6 P
ERFORMANCE DATA
Rectangular shutters
• two pairs of independent, symmetrically moving main shutters, 3mm thick lead (mechanically
coupled to backup shutters and near focus shutters)
• max. aperture : 2 x 14.4° (image receptor area 510mm at SID=100cm)
minimum field size : < 2 x 2 mm (at SID=100cm)
•
Iris
• 4 circular shaped iris segments, lead 3mm thick, in combination with the rectangular shutters
• max iris segment aperture : 18.8° (image receptor area 680mm at SID=100cm)
• max circular aperture : 14.4° (image receptor area 510mm at SID=100cm)
• minimum field size : 90mm (at SID=100cm)
Spectral Filter
• Remote controlled, discontinuous rotatable filter disk, containing four selectable filter positions.
For filter material and thickness, see Table 8: Filter material/position assignment.
• Filter switch time : < 350 ms (for adjacent filters)
< 700 ms (full range)
Table 8: Filter material/position assignment
9896 010 2216.
9896 010 2224.
9896 010 2223.
material Aleq
Filter
position
9896 010
2217.
material Aleq
@75kV
Position 0 no filter - no filter Position 1 1 mm Al + 0.1 mm Cu 4.0 mm 1 mm Al + 0.1 mm Cu 4.0 mm
Position 2 1 mm Al + 0.2 mm Cu 6.5 mm 1 mm Al + 0.2 mm Cu 6.5 mm
Position 3
No filter(s)
2 mm Al 2.0 mm 1 mm Al + 0.3 mm Cu 11.0 mm
Light
• Light source : 12V/100W halogen lamp
• Light intensity : > 193 lux (at SID=100 cm)
• X-Ray/Light Field accuracy : < 2% of SID in two directions
(including all X-Ray tube assembly tolerances)
< 2.5 mm (at SID = 100 cm for BLD only)
• Contrast ratio : > 4
• Maximum duty cycle : 80%, max. on time 8 min.
• Lamp on period : 5 - 60 s (SW programmable, accuracy ±0.1s)
• Lamp switch off
2
temp. : 75°C
Ruler
• Range : 30 - 205 cm
• Accuracy : 1.4 cm (at SID=70 cm)
: 4 cm (at SID=205 cm)
• Resolution : 1 mm
@75kV
2
When the BLD has reached a temperature of 75°C, the lamp is switched off automatically (SW-
controlled). It is no longer possible to switch on the lamp, until the temperature is <75°C.
• Inherent filtration : 0.22m Al equivalent @ 75kV (spectral filter NOT included)
• Max. X-Ray tube voltage : 150 kV
• Max. X-Ray tube power : 500W (for 9896 010 2216., for any filter)
350W (for 9896 010 2217., 9896 010 2223.
and 9896 010 2224., any filter)
• Cold reset/warm reset : <30 s/<5 s
1.7 COMPLIANCE INFORMATION
Philips products comply with relevant international and national standards and laws. Information on
compliance will be supplied on request by your local PMS representative, or by:
Philips Medical Systems
PO Box 10.000
5680 DA Best
The Netherlands
Facsimile: +31 40 276 2205
Philips products comply with relevant international and national law and standards on EMC (ElectroMagnetic Compatibility) for this type of equipment when used as intended.
Such laws and standards define both the permissible electromagnetic emission levels from equipment
and its required immunity to electromagnetic interference from external sources.
• IEC 60601-1 "Medical electrical equipment. Part 1: General requirements for safety" (second edition 1988, including
amendment nr. 1 (1991), amendment nr. 2 (1995))
• IEC 60601-1-2 "Medical electrical equipment. Part 1: General requirements for safety. 2. Collateral Standard:
Electromagnetic compatibility - Requirements and tests" (second edition, 2001)
• IEC 60601-1-3 "Medical electrical equipment. Part 1: General requirements for safety. 3. Collateral Standard: General
requirements for radiation protection in diagnostic X-ray equipment" (first edition, 1994-07)
• IEC 60601-1-4 "Medical electrical equipment. Part 1: General requirements for safety. 4. Collateral Standard:
Programmable electrical medical systems" (first edition, 1996-05)
• UL 2601-1 "Medical electrical equipment, Part 1: General requirements for safety" (second edition, October 1997)
• CAN/CSA -C22.2 No. 601.1-M90 "Medical Electrical Equipment. Part 1: General requirements for safety" (November
1990, including Supplement C22.2 No. 601.1S1-94 and Amendment nr. 2 (1998))
• 21CFR, Subchapter J
•
IEC 60601-2-28 "Medical electrical equipment. Part 2: Particular requirements for the safety of X-ray source assemblies
and X-ray tube assemblies for medical diagnosis" (first edition, 03-1993)
Introduction and Technical Data NICOL X-Ray Beam Limiting Device
ONDITIONS OF ACCEPTABILITY
1.8 C
The BLD is compatible with:
• X-ray tube assemblies, which meet the requirements as specified in the table below.
Leakage Radiation
≤ 0.43 mGy/h (≤ 50mR/h)
at conditions 150kV, 3mA and 100cm
distance.
Values of loading factors concerning leakage radiation
For RAD
<500 µSv/h
at conditions 150kV, 500W, 12000
mAs/h, 100cm distance.
For DuoD/DuoD+iris/OmniD <500 µSv/h
at conditions 150kV, 350W, 8400
mAs/h, 100cm distance
Inherent Filtration
≥2.5 mm Aleq
at conditions 150kV, 3mA
Distance Focus - Coupling Flange Interface Plane
64mm ± 2mm
• X-ray systems, which meet the requirements as specified in the table below.
Misalignment of the edges of the of visually defined Field with
≤ 1% of SID
the respective edges of the X-ray Field along either the length
or the width of the visually defined field.
Alignment of the centre of the radiographic X-ray Field with the
≤ 1.5% of SID
center of the image receptor
1.9 ABBREVIATIONS AND DEFINITIONS
Abbreviation Explanation
AEP Area Exposure Product
Aleq Aluminum equivalency
BLD Beam Limiting Device
DuoD Duo Diagnost
FRU Field Replaceable Unit
FSE Field Service Engineer
OmniD Omni Diagnost
PMS Philips Medical Systems
P&P Plug & Play
RAD Radiography
RF Radiography / Fluoroscopy
SID Source Image Distance
V2 Version 2
NICOL X-Ray Beam Limiting Device Introduction and Technical Data
1.10 M
ANUAL HISTORY
Original Text : English
Date Version NameReason of changes
19th June 1999 99.0 P.W.M. Sijbers First Version
9th September 1999 99.1 P.W.M. Sijbers
31st August 2000 00.0 P.W.M. Sijbers
31st October 2001 01.0 P.W.M. Sijbers
3rd December 2003 03.0 P.W.M. Sijbers
17th March 2004 04.0 P.W.M. Sijbers
• NICOL OmniD (9896 010 2224.) added.
• Chapter Conditions of Acceptability added.
• Document updated due to changed PEI-
numbers
• Chapter “1.6.1 Compatibility” extended for
new tubes and Nicol V2
• Chapter “2.2.1 Mounting the Coupling
Flange Nicol V2” added
This chapter contains general instructions about the installation of the BLD. For specific information at
system level, refer to the relevant system reference manual.
2.1.1 TOOLS & TEST EQUIPMENT
• Standard Service Toolkit
• BLD alignment tool, code number 4522 980 31521, available from Service Logistics Best.
2.1.2 S
The BLD is delivered as part of a pre-installed system or in a single package, always containing:
2.2 I
Whenever replacing a BLD, no centering is required, provided the coupling flange is left on the X-Ray
UPPLIED ITEMS
• BLD (including aesthetic cover)
• Coupling flange assembly (to mount and adjust the BLD on the X-Ray tube housing)
• Four countersunk screws (M6x25mm) (used for mounting the shipping bracket to the coupling
flange, also are intended to mount the coupling flange to the X-Ray tube housing)
• Service Manual (No Operator Manual supplied)
NSTALLATION INSTRUCTIONS
• Remove the items from the packing and check them against paragraph 2.1.2 Supplied Items,
for completeness.
This chapter describes the mounting procedure in case a pre-aligned X-ray Tube Assembly is used.
• Unscrew the locking plate screw (item 5, Figure 3, page 19) from each of the two locking plates
(item 1, Figure 3) and push them outwards as far as possible.
• Remove the shipping bracket (item 2, Figure 3) together with the coupling flange (item 3, Figure
3) from the BLD
• Unscrew the shipping bracket from the coupling flange (countersunk screws M6 x 25 mm must
be reused to mount the coupling flange on the tube housing.)
NOTE
Check the screw required length (25 mm) in the applicable tube housing service documentation.
At least 5 mm must be screwed into the tube housing.
________________
• Place the coupling flange (item 1,
Figure 1) on the X-Ray tube
housing (item 2, Figure 1) with
factory mounted alignment ring
(item 3, Figure 1)
• Positioning the reference pin (item
4, Figure 1) to the notch (item 5,
Figure 1) of the alignment ring.
Secure it with the four countersunk
• Unscrew the locking plate screw (item 5, Figure 3, page 19) from each of the two locking plates
(item 1, Figure 3) and push them outwards as far as possible.
• Remove the shipping bracket (item 2, Figure 3) together with the coupling flange (item 3, Figure
3) from the BLD
• Unscrew the shipping bracket from the coupling flange (countersunk screws M6 x 25 mm must
be reused to mount the coupling flange on the tube housing.)
NOTE
Check the screw required length (25 mm) in the applicable tube housing service documentation.
At least 5 mm must be screwed into the tube housing.
________________
• Place the coupling flange (item 1, Figure 2) on the X-Ray tube housing (item 2, Figure 2),
positioning the 0° notch (item 5, Figure 2) to the rear of the X-Ray tube housing, and secure it
with the four countersunk screws M6 x 25 mm, item 3, Figure 2. (see note above)
Figure 2: Mounting the coupling flange assembly
2.2.3 CENTERING THE BLD WITH RESPECT TO THE X-RAY TUBE
All BLD functions have been accurately centered in the factory, relative to the central axis; any further
centering has to be done by adjusting the coupling flange assembly of the BLD with respect to the XRay focus. The specified alignment tool must be used.
The centering procedure is described in paragraph 6.1.4, page 40.
This procedure must be carried out:
• After replacement of NICOL DISC 4522 300 2417x
• After replacement of the X-Ray tube, configurations A and D of Table 3.
It is advised to start/perform faultfinding based on error messages as described in the relevant system
documentation, see system reference manual. Only if this is not possible, or does not lead to a
solution for the problem, this chapter may be of help. The faultfinding directly points to some basic
FRU’s (PCB’s and BLD) and a few FRU’s which are cheap and easy to replace.
3.3 INDICATORS
LED indicators are present on PCB’s LA2 and LA3.
The explanation of the LED indicators is given in Table 10 and Table 11. The location of the LED’s on
the PCB’s is given in Z3-2 to Z3-3.
3.3.1 LA2: P
Number Color Description Faultfinding and solutions
LA2H1 green +24V present In case H1 is OFF:
LA2H2 green +5V present In case H2 is OFF and H1 is ON:
LA2H3 green +5V CAN
LA2H4 green +12V lamp
OWER SUPPLY CONVERTER
Table 10: Indicators LA2
• check system power supply
• check cable connection from system to LA2X2
• replace LA2
• replace BLD
• check +24V (LA2MP1)
• remove LA3, when H2 is lit now, check voltage, and if
1. Replace the DSC buttons, items 1, Figure 4, page 20, by pulling them outwards.
2. Assemble in reverse order.
4.6 REPLACE DSC ASSEMBLY
The DSC-A assembly is situated at the lamp
whereas the
side
DSC-B assembly
at the ruler side. (On the flat-cable of the DSC
assembly is indicated DSC-A respectively DSCB). It is not necessary to remove the BLD from
the X-Ray tube housing assembly for
replacement of the DSC assemblies.
CAUTION
When the lamp has been used, the lamp and its
surrounding material might be (“very”) hot.
________________
is situated
Figure 9: Removal iris motor assembly
Replace DSC-A assembly
1. Remove the Aesthetic Cover, see paragraph
2.2.5, page 20.
2. Be sure the power is switched OFF.
3. Remove the Lamp Side Plate, see paragraph
4.14, page 38.
4. Loosen the 2 screws, item 2 and 3, Figure 10,
page 29.
1. Follow the instructions 1 - 5 of paragraph 4.2.
2. Remove micro-switch connector LA3X5.
3. Remove the 4 sunken screws (items 2, Figure 13) from the LA3 Cover-plate (item 1, Figure 13) and
remove the LA3 Cover-plate
4. Remove 2 screws and washers (items 4) from LA3.
5. Remove the LA3, item 3, by lifting the board (connectors LA3X1 respectively LA3X2 from LA1X3
respectively LA3X4)
6. Exchange PROM LA3D17, see 4.8.1, page 32.
7. Reassemble in reverse order. Be sure that the board connectors (item 6) are positioned exactly in
the correct position onto the backpanel connectors (item 5).
8. Check correct positioning of the connectors LA3X1 respectively LA3X2 onto LA1X3 respectively
LA3X4
Figure 13: Replace LA3 SIFLCO
4.8.1 REPLACE PROM LA3D17
1. Follow steps 1-4 of paragraph 4.8
2. Remove PROM LA3D17 from socket, see Z3-3, by using an appropriate IC extractor
3. Place correct PROM into socket and check correct position, see Z3-3.
4. Reassemble in reverse order, steps 5-6 of paragraph 4.8
For safe handling of PCB’s and PROM’s we refer to ESD Precautions as described in Section 4:
Planned Maintenance Instruction Modules (4522 983 24611 or 4535 440 04161)
Figure 15 page 34 gives in more detail, the connection of the cable of the electronic ruler. In case of
connection/disconnection of the electronic ruler cable, following steps have to be taken:
1. Remove ruler side plate, see chapter 2.2.5 page 20.
2. Remove connector LA1X5 (item 5 Figure 15) and cable (item 2 Figure 15). Cable must be
removed by pulling connector LA5X1 (item 1 Figure 15) through the hole (item 4 Figure 15) in
LA1.
3. Replace cable in reverse order
4. Mount ruler side plate, see chapter 2.2.5 page 20, while
5. Guiding ruler cable (item 3) together with power supply and CAN cable(s) through cable guidance.
1. Remove the aesthetic cover, paragraph 2.2.5, page 20.
2. Switch Power OFF.
3. Disconnect connector LAX5.
4. Remove the micro-switch (items 6 till 10, Figure 18 page 36).
5. Replace the micro-switch (item 8, Figure 18 page 36), reusing items 5, 6, 7, 9 and 10 and reconnect LAX5. Be sure the micro-switch is positioned maximally outwards before tightening the
screws.
6. Switch Power ON
7. Check proper functioning of the micro-switch.
8. Reassemble the aesthetic cover.
9. Check proper functioning of the micro-switch, with aesthetic cover reassembled.
WARNING
Maximum torque for tightening the screws items 6, Figure 18 page 36 = 11Ncm
________________
4.13 REPLACE MICRO-SWITCH INCLUDING BRACKET
1. Remove the aesthetic cover, paragraph 2.2.5, page 20.
2. Switch Power OFF.
3. Disconnect connector LAX5.
4. Remove the micro-switch including the bracket (items 5 till 10, Figure 18 page 36).
5. Replace the micro-switch including the bracket and re-connect LAX5. Be sure the micro-switch is
positioned maximally outwards before tightening the screws.
6. Switch Power ON
7. Check proper functioning of the micro-switch.
8. Reassemble the aesthetic cover.
9. Check proper functioning of the micro-switch, with aesthetic cover reassembled.
This chapter describes the jumper settings for each board. The settings of other BLD programmable
functions, that must be changed with a software service tool are NOT described here, see relevant
Systems reference manual.
5.2 HARDWARE PROGRAMMING
5.2.1 LA1: B
Jumper Position Description Delivery Position
W1 1 - 2
5.2.2 LA2: P
ACKPANEL
2 - 3
OWER SUPPLY CONVERTER
Table 14: Programming LA1
BLD configured as frontal collimator (BLD1)
BLD configured as lateral collimator (BLD2)
Refer to Z3-1 for the position on the PCB
& CAN I
1 - 2
NTERFACE BOARD
(PSC_CAN)
Table 15: Programming LA2
Jumper Position Description Delivery Position
W1 1 - 2
none
W2
BLD programmed as CAN end node
BLD NOT an end node
Refer to Z3-2 for the position on the PCB
jumper NOT mounted
jumper
never
mounted!
5.3 SOFTWARE PROGRAMMING
All software programming is done at system level via the CAN interface. For this purpose, we refer to the
applicable system reference manual(s). Some examples of software programming at system level are:
• manual shutter speed
• shutter move limits (= radiation angle)
• BLD orientation (with respect to X-Ray tube housing)
• After replacement of the X-Ray tube, configurations A and D of Table 3.
•
ENTERING THE
6.1 C
for replacements of NICOL and NICOL V2
NOT
BLD
WITH THE
X-R
AY TUBE
6.1.1 I
NTRODUCTION
For perfect attachment of the BLD to the X-Ray tube housing, it is necessary that the BLD axis coincides
with the median perpendicular line that passes through the center, between the two tube foci. Use BLD
alignment tool, see Figure 32, page 50. Ideally, the resulted picture consists of two concentric circles. In
reality there will be different pictures for small and large focus, see
Figure 22
, page 41.
6.1.2 TOOLS REQUIRED
The BLD alignment tool, code number 4522 980 31521
6.1.3 M
OUNTING THE
BLD A
The BLD alignment tool can be mounted:
• if the coupling flange assembly is not mounted, first perform this procedure as described in
paragraph 2.2.2, page 18.
• if the coupling flange assembly is mounted, similar procedure as for BLD mounting, see
paragraph 0, page 19.
6.1.4 C
HECKING THE CENTERING FOR
To check the BLD centering relative to the X-Ray tube, proceed as follows:
1. Prepare a film cassette (with screen) oriented in the plane perpendicular to the central ray, with the
edges of the film parallel to the longitudinal axis of the tube.
2. Set the following data: 65kV, 30mAs, SID=100mm (or more)
3. Make an exposure with small focus (SF)
Develop the film and determine:
• the center MA of the projection of ring A (see Figure 20)
• the center MB of the projection of ring B (see Figure 20)
• Draw the system coordinates on the film in the A-direction and B-
direction with reference to the film edges. (the positive directions
can be derived from the mark at the right-hand top of the film)
4. Measure the distances P and Q on the film
5. Make an exposure with
Develop the film and determine:
• the center MA of the projection of ring A (see Figure 21)
the center MB of the projection of ring B (see Figure 21)
•
• Draw the system coordinates on the film in the A-direction and B-
direction with reference to the film edges. (the positive directions
can be derived from the mark at the right-hand top of the film)
6. Measure the distances R and S on the film
7. The coupling flange is correctly adjusted when the film image of the
small focus is a mirror image with respect to that of the large focus (see
Figure 22) In this case the following applies:
B-direct ion
-a+a
S
large focus
+b
R
MA
MB
mark
|P| = |-R| ± 0.3mm
|Q| = |-S| ± 0.3mm
8. Further correction is not recommended because the adjustment can
become worse due to measuring errors during the correction
A-direct ion
-b
procedure.
Figure 21: LF alignment
If the above tolerances have
B-direction
been fulfilled, remove the BLD
small focus
+b
mark
large focus
+b
mark
alignment tool and proceed to
BLD mounting as described in
paragraph 0, page 19. If not,
proceed to paragraph 6.1.5.
-a+a
MA
P
MB
Q
-a+a
R
MA
S
MB
P = - R
A-direction
-b
Q = - S
-b
Figure 22: Requirement for alignment
6.1.5 ADJUSTING THE CENTRING
To adjust the BLD centering with the X-Ray tube, proceed as follows:
1. Measure the distances A, B, C and D (see Figure 23) and enter them as default values in Table 17
below.
2. Determine: Ta = (P + R)/2
Ub = (Q + S)/2
Va = Ta x Z
Wb = Ub x Z
with the factor Z derived from the following table, depending on the selected SID:
Preconditions for the Field Indication Plate adjustment:
• The X–ray field and light field should be correctly aligned.
• The output window of the BLD must be parallel to the receptor plane.
• The SID should be as large as possible (100 cm or more).
• Check whether the field indication plate is mounted correctly onto the adjusting strip, item 2
Figure 25
Procedure (refer to Figure 25, Page 44)
1. Close the vertical shutters (shutters A) manually, by using the corresponding DSC-button, to an
opening at receptor or table, of approximately 30 mm. (lamp automatically switches on)
2. Open the horizontal shutters (shutters B) completely (use corresponding DSC-button).
3. Loosen the two screws (items 3, Figure 25) and move the adjusting strip (item 2) and the Field
Indication Plate (item 1, Figure 25) so that the center of the Field Indication Plate cross-lines is
parallel and within 2.5 mm of the light field center line.
4. Close the horizontal shutters to an opening at receptor or table of approximately 30 mm.
5. Open the vertical shutters (shutters A) completely.
6. Move the adjusting strip (item 2, Figure 25) and Field Indication Plate (item 1, Figure 25) so that the
center of the cross of the Field Indication Plate is parallel and within 2.5 mm of the centerline of the
light field.
7. Secure the adjusting strip with the two screws (items 3, Figure 25).
• if the check of the alignment of the X-Ray Field to the Light Field (chapter 2.2.6, page 22) failed
For adjustment of the size of the Light Field with respect to
the X-Ray Field, the following procedure has to be used.
1. Remove aesthetic covers, see chapter 2.2.5, page 20.
2. Remove lamp side plate, see chapter 2.2.5, page 20.
3. Loosen the 2 screws, item 1, Figure 26.
4. IF Light Field > X-Ray Field THEN turn screw, item 2,
Figure 26, clockwise to reduce the Light Field.
5. IF Light Field < X-Ray Field THEN turn screw, item 2,
Figure 26 counter clockwise to enlarge the Light Field.
6. After completing the adjustment, tighten the 2 screws,
item 1, Figure 26.
7. The adjustment must be checked according the
procedure as described in chapter 2.2.6, page 22
6.3.2 LAMP ADJUSTMENT AFTER REPLACING
Figure 26: X-Ray Field/Light Field alignment
LAMP
This procedure only has to be carried out:
• If after replacement of the lamp, the Check the
alignment of the X-Ray Field to the Light Field (chapter
2.2.6, page 22) failed because the Light Field with
respect to the X-Ray Field is shifted in x direction.
(Definition see Figure 29, page 46)
The following procedure is to be used:
1. After replacing the lamp, the alignment of the X-Ray Field with
respect to the Light Field must be checked according the
procedure as described in chapter 2.2.6, page 22
2. If not OK proceed with the following steps.
3. Remove the Field Indication Plate, item 1, Figure 25, page 44
4. Place a film (density 3) or paper on the BLD output screen.
5. Switch on the lamp
6. Look perpendicular onto the mirror and check visually, that the
filaments of the lamp are correctly positioned (A = B), see Figure
27 page 45.
7. If not OK, reposition lamp (Don’t touch lamp with your fingers,
also lamp is getting very hot!)
Figure 27: Lamp Adjustment
8. Finally, the alignment of the X-Ray Field with respect to the Light
Field must be checked again according the procedure as described in chapter 2.2.6, page 22. If not
correct, repeat procedure as described in chapter 2.2.6, page 22, starting from step 6.
• If after replacement of the lamp, the Check the alignment of the X-Ray Field to the Light Field
(chapter 2.2.6, page 22) failed because the Light Field with respect to the X-Ray Field is shifted
in y direction. (Definition see Figure 29, page 46)
Swap Lamp:
1. Allow the lamp to cool down
2. Remove the lamp, do NOT touch the lamp with your fingers.
3. Reinstall the lamp, with the lamp pins switched
4. Perform Lamp Adjustment After Replacing Lamp chapter
6.3.2 page 45 .
5. Check the alignment of the X-Ray Field to the Light Field
(chapter 2.2.6, page 22)
There are NO electrical adjustments required for the BLD.
6.5 SOFTWARE ADJUSTMENTS
All software adjustments with respect to the BLD are described in the relevant system reference manual.
The BLD has NO direct service entry; this is provided by the system and described in the system
reference manual. The following software adjustments with respect to the BLD are applicable: