Philips HeartStart MRx User manual (2005)

M3535A/M3536A Instructions for Use
HeartStart MRx
1Table of Contents
1 Introduction 3
Overview 3 Intended Use 5
Defibrillation Therapy 5 Synchronized Cardioversion Therapy 5 Noninvasive Pacing Therapy (Optional) 5 12-Lead ECG Analysis (Optional) 6 Noninvasive Blood Pressure (NBP) Monitoring (Optional) 6 CO2 Monitoring (Optional) 6 Pulse Oximetry (SpO2) Monitoring (Optional) 6
Indications for Use 7 Safety Considerations 8 Documentation and Training 8
2 Getting Acquainted 9
Basic Orientation 10
Front Panel 10 Side Panels 11 Top Pa n e l 12 Back Panel 13
M3538A Lithium Ion Battery 14
Battery Capacity 14 Battery Life 14
Operating Modes 14
Password Security 15
Display Views 15
General Status 16 Wave Se ct or s 17 Parameter Blocks 18 Soft Key Labels 18 Display Menus 19 Message Windows 19 High Contrast Display 20
Controls 20
Therapy Knob 20 General Function Buttons 21 Defibrillation Controls 22 Soft Keys 22 Indicators 23
i
Responding to Alarms 24 Entering Patient Information 25 Continued Use 25 Return to Owner 26
3 Setting Up 27
Attaching the Carrying Case and Accessory Pouches 27 Storing Accessories 29 Connecting the ECG Cable 31 Connecting the SpO2 Cable 32 Connecting the NBP Interconnect Tubing 33 Connecting the CO2 Nasal FilterLine Connecting the Therapy Cable 35 Installing Paper 36
50mm Printer 36 75mm Printer 37
Installing Batteries 38
Charging Batteries 38 Battery Safety 38
Installing the AC Power Module 39 Installing the Data Card 40
®
34
4 ECG and Arrhythmia Monitoring 41
Overview 41 Monitoring View 42 Preparing to Monitor ECG 43 Electrode Placement 45 Lead Selection 47
Lead Choices 47 Selecting the Lead 48
Arrhythmia Monitoring 48
Aberrantly-Conducted Beats 49 Intermittent Bundle Branch Block 49
Heart Rate and Arrhythmia Alarms 50
Arrhythmia Alarm Latching 50 INOP Messages 51
Setting Alarms 52
Changing Heart Rate or Vtach Alarm Limits 52 Enabling/Disabling Heart Rate and Arrhythmia Alarms 52
Responding to HR and Arrhythmia Alarms 52 Displaying an Annotated ECG 53 Arrhythmia Learning/Relearning 54 Troubleshooting 54
ii
5 Semi-Automated External Defibrillation 55
AED View 56 Preparation 57 Using AED Mode 59
Step 1 - Turn the Therapy Knob to AED 59 Step 2 - Follow the Screen and Voice Prompts 60 Step 3 - Press the Orange Shock Button, if Prompted 62
Troubleshooting 63
6 Manual Defibrillation and Cardioversion 65
Overview 65 Code View 66 Preparing for Defibrillation 67
Using Multifunction Electrode Pads 67 Using External Paddles 68 Using Pediatric Paddles 69 Using Internal Paddles 69
Defibrillating (asynchronously) 70 Performing Synchronized Cardioversion 72
Preparing for Synchronized Cardioversion 72 Delivering a Synchronized Shock 73 Delivering Additional Synchronized Shocks 74 Disabling the Sync Function 74
Troubleshooting 74
7 Noninvasive Pacing 75
Overview 75 Alarms 76 Pacing View 77 Demand Mode Versus Fixed Mode 78 Preparing for Pacing 79 Demand Mode Pacing 80 Fixed Mode Pacing 81 Defibrillating During Pacing 82 Troubleshooting 82
iii
8 Pulse Oximetry 83
Overview 83 Understanding Pulse Oximetry 84 Selecting a Sensor 85 Applying the Sensor 87 Monitoring SpO
Pleth Wave 89
2
88
Setting SpO2 Alarms 90
Changing the SpO2 Alarm Limits 90 SpO2 Desat Alarm 90 Enabling/Disabling the SpO2 Alarms 90
Setting Pulse Rate Alarms 91
Enabling/Disabling the Pulse Rate Alarms 91 Changing the Pulse Rate Alarm Limits 91
Disabling the SpO2 Monitoring Function 92 Caring for Sensors 92 Troubleshooting 92
9 Noninvasive Blood Pressure 93
Overview 93 Preparing to Measure NBP 94 Measuring NBP 96
Changing the NBP Schedule 96
Alarms 97
Changing NBP Alarms 97 Enabling/Disabling NBP Alarms 97
Troubleshooting 97
10 Monitoring Carbon Dioxide 99
Overview 99 Preparing to Measure EtCO
Selecting the Accessories 100
Setting Up Microstream EtCO2 Measurements 101
Using the Nasal FilterLine 101 Using the FilterLine and Airway Adapter 101
Measuring EtCO
2
Setting Up the EtCO2 and AwRR Alarms 102
Changing the EtCO2 Alarm Limits 103 Enabling/Disabling the EtCO2 Alarms 103 Changing the AwRR Alarm Limits 103 Changing the Apnea Time Alarm Limit 104 Enabling/Disabling AwRR Alarms 104
Disabling the EtCO2 Monitoring Function 104 Troubleshooting 104
iv
2
100
102
11 12-Lead ECG 105
Overview 105 Preview Screen 106 Preparation 106 Acquiring the 12-Lead ECG 108 12-Lead Report 109
Accessing Stored Reports 109
Improving Signal Quality 111
Adjusting Wave Size 111
12-Lead Filters 112 Troubleshooting 112
12 12-Lead ECG Transmission 113
Overview 113 Setting Up for Bluetooth Transmission 115
Adding a Bluetooth Device 115 Pairing a Bluetooth Device with the HeartStart MRx 115 Changing Bluetooth Profiles 116 Additional Bluetooth Device Information 116
Transmitting to a Configured Site using Bluetooth 117 Setting Up for RS 232 Transmission 118 Transmitting to a Configured Site Using RS 232 120 Transmitting to a Manually Entered Fax Number 121 Transmission Status 122 Transmitting Stored 12-Lead Reports 123 Cancelling Transmission 123 Troubleshooting 124
13 Configuration 125
Overview 125 Accessing the Configuration Menu 125 Setting the Date and Time 126 Modifying Settings 126 Saving Configuration Settings to a Data Card 127 Loading Configuration Settings from a Data Card 127 Restoring the Default Settings 127 Printing Configuration Settings 127 Configurable Parameters 128
v
14 Data Management 141
Overview 141 Copying from Internal Memory 142 Viewing and Erasing the Data Card 142 Printing the Event Summary 143 Printing the 12-Lead ECG Report 143 Events Recorded 144 Marking Events 148 Printing Events 149
15 Maintenance 151
Overview 151 Automated Tests 152
Automated Test Summary 153
Ready For Use Indicator 155 Shift Checklist and Operational Check 156
Weekly Shock Test 156
Operational Check (Versions Prior to B.05) 157
Performing the Operational Check 157 Operational Check Report for Versions Prior to B.05 162 Operational Check Summary 165
Operational Check (Versions B.05 and Greater) 166
Performing the Operational Check 167 Operational Check Report for Versions B.05 or Greater 173 Operational Check Summary 177
Battery Maintenance 178
Battery Capacity 178 Battery Life 178 Charging Batteries 179 Battery Calibration 179 Storing Batteries 180 Discarding Batteries 181
Cleaning Instructions 182
Monitor/Defibrillator 182 Printer Printhead 182 Paddles, Therapy Cable 183 ECG Cable 183 Carrying Case 184 NBP Cuff 184 SpO2 Sensor and Cable 184
Disposing of the HeartStart MRx 185
Disposing of Empty Calibration Gas Cylinders 185
Supplies and Accessories 186
vi
16 Troubleshooting 193
Symptoms 194 Audio Tones and Alarm Indications 210 Calling for Service 211
17 Specifications and Safety 213
Specifications 213
General 213 Defibrillator 213 ECG and Arrhythmia Monitoring 216 Display 218 Battery 218 Thermal Array Printer 219 Noninvasive Pacing 219 SpO2 Pulse Oximetry 220 NBP 221 EtCO2 221 AwRR 222 Calibration Gas for CO2 Measurement System 223 12-Lead ECG 223 Patient Data Storage 223 Environmental (M3535A) 223 Environmental (M3536A) 225 Symbol Definitions 227
Clinical Performance Summary - Defibrillation 230
Methods 230 Results 230 Conclusion 230
Clinical Performance Summary - Cardioversion 231
Methods 231 Results 231 Conclusion 232
Clinical Performance Summary - Internal Defibrillation 233
Overview 233 Methods 233 Results 233 Conclusion 233
vii
Safety Considerations 234
General 234 Defibrillation 236 Battery 237 Supplies and Accessories 238
Electromagnetic Compatibility 238
Reducing Electromagnetic Interference 238 Restrictions for Use 239 Emissions and Immunity 239 Guidance and Manufacturer’s Declaration 239
Index 247
viii
Notice
About This Edition
Edition 4
Printed in the USA
Publication number M3535-91900
To determine the product level version to which the Instructions for Use are applicable, refer to the version level appearing on the back cover of this book or on the label of the User Documentation CD-ROM that accompanied this device. This information is subject to change without notice.
Philips shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Edition History
Edition Print Date
1 December, 2003
2 June, 2004
3 October, 2004
4 October, 2005
Copyright
Medical Device Directive
The HeartStart MRx complies with the requirements of the Medical Device Directive 93/42/EEC and carries the
mark accordingly.
0123
Manufacturer
Philips Medical Systems 3000 Minuteman Road Andover, MA 01810
Authorized EU-representative
Philips Medizin Systeme Böblingen GmbH Hewlett Packard Str. 2 71034 Böblingen Germany
Canada EMC:ICES-001
For the Declaration of Conformity Statement, please see the Philips Medical web site at http:// incenter.medical.philips.com/PMSPublic. Scroll over the Quality and Regulatory Tab located in the upper left corner of the window. Click to select Regulatory by Modality. Then click to select Defibrillators and select the entry for Declaration of Conformity (DoC)
War ning
.
Copyright © 2005
Koninklijke Philips Electronics N.V.
All rights are reserved. Permission is granted to copy and distribute this document for your organization’s internal educational use. Reproduction and/or distribution outside your organization in whole or in part is prohibited without the prior written consent of the copyright holder.
SMART Biphasic is a registered trademark of Philips.
Use of supplies or accessories other than those recommended by Philips may compromise product performance.
THIS PRODUCT IS NOT INTENDED FOR HOME USE.
IN THE U.S., FEDERAL LAW RESTRICTS THIS DEVICE TO SALE ON OR BY THE ORDER OF A PHYSICIAN.
Radio frequency (RF) interference from nearby transmitting devices may degrade the performance of the HeartStart MRx. Electromagnetic compatibility with surrounding devices should be assessed prior to using the monitor/defibrillator.
1
These Instructions for Use contain the following conventions:
WARNING Warning statements describe conditions or actions that can result in personal injury or loss of life.
CAUTION Caution statements describe conditions or actions that can result in damage to the equipment or loss of
data.
NOTE Notes contain additional information on usage.
"Voice" represents voice prompt messages
Text represents messages that appear on the display
[Soft key] represents soft key labels that appear on the display above the
button to which they correspond.
2
Thank you for choosing the HeartStart MRx monitor/defibrillator. Philips Medical Systems welcomes you to its family of resuscitation devices.
The HeartStart MRx is designed to meet your monitoring and resuscitation needs by providing advanced, multi-parameter monitoring functions and a full range of defibrillation therapies. This guide provides instructions for the safe and proper operation of the device, as well as set-up, configuration, and maintenance information.
Be sure to familiarize yourself with the features and operation of the HeartStart MRx prior to its use.

Overview

The HeartStart MRx is a lightweight, portable, monitor/defibrillator. It provides four modes of operation: Monitor, Manual Defib, AED, and Pacer (optional).
1

1Introduction

In Monitor Mode you can monitor up to four ECG waveforms, acquired through a 3-, 5-, or 10-lead ECG set. Optional monitoring of pulse oximetry (SpO carbon dioxide (EtCO display and alarms are available to alert you to changes in the patient’s condition.
Monitor Mode also provides an optional 12-Lead ECG function, enabling you to preview, acquire, store, and print 12-lead ECG reports, with or without analysis/interpretation.
Manual Defib Mode offers simple, 3-step defibrillation. You analyze the patient’s ECG and, if appropriate: 1) select an energy setting, 2) charge, and 3) deliver the shock. Defibrillation may be performed using paddles or multifunction electrode pads. Manual Defib Mode also allows you to perform synchronized cardioversion and internal defibrillation. If desired, use of Manual Defib Mode may be password protected (for Version A.02 and later devices).
In AED Mode, the HeartStart MRx analyzes the patient’s ECG and determines whether a shock is advised. Voice prompts guide you through the 3-step defibrillation process, providing easy-to-follow instructions and patient information. Voice prompts are reinforced by messages that appear on the display.
Both Manual Defib and AED Mode incorporate the Philips’ low energy SMART Biphasic waveform for defibrillation. The Q-CPR™ option, available in both Manual Defib and AED Modes, offers real­time, measurement and corrective feedback on the rate, depth, and duration of CPR compressions, as well as the frequency and volume of ventilations. It also provides notification of lack of CPR activity. For more information about Q-CPR refer to the Instructions for Use Addendum for Q-CPR.
NOTE Q-CPR™ is a trademark of Laerdal Medical.
) are also available. Measurements from these parameters are presented on the
2
), noninvasive blood Pressure (NBP), and
2
3
1 Introduction Overview
Optional Pacer Mode offers noninvasive transcutaneous pacing therapy. Pace pulses are delivered through multifunction electrode pads, using a monophasic waveform. If desired, use of Pacer Mode may be password protected (for Version A.02 and later devices).
The HeartStart MRx is powered by rechargeable lithium ion batteries. Available battery power is easily determined by viewing the convenient battery power indicators located on the device display or by checking the gauge on the battery itself. Additionally, an external AC or DC Power Module may be applied as a secondary power source and for continual battery charging.
The HeartStart MRx performs Automated Tests on a regular basis. The results of these tests are reported to the Ready For Use (RFU) indicator. Prominently displayed, the RFU indicator communicates the status of your device, letting you know it is operating correctly, needs attention, or is unable to deliver therapy. In addition, performing the specified Operational Check ensures that the HeartStart MRx is functioning properly.
The HeartStart MRx automatically stores critical event data in its internal memory, such as Event Summaries and 12-Lead Reports. The HeartStart MRx also enables you to store data and event information on an optional data card for downloading to Philips’ data management solution, HeartStart Event Review Pro.
The HeartStart MRx is highly configurable to better meet the needs of diverse users. Be sure to familiarize yourself with your device’s configuration before using the HeartStart MRx. See “Configuration” on page 125.
4
Intended Use 1 Introduction

Intended Use

The HeartStart MRx is for use in hospital and pre-hospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support or defibrillation.
When operating as a semi-automatic external defibrillator in AED Mode, the HeartStart MRx is suitable for use by medical personnel trained in basic life support that includes the use of an AED.
When operating in Monitor, Manual Defib or Pacer Mode, the HeartStart MRx is suitable for use by healthcare professionals trained in advanced cardiac life support.

Defibrillation Therapy

Defibrillation therapy is the definitive method for termination of a variety of potentially fatal arrhythmias. The HeartStart MRx provides this therapy through the application of a brief biphasic pulse of electricity to the cardiac muscle. This electrical energy is transferred through attached paddles or disposable multifunction electrode pads applied to the patient’s bare chest.
NOTE Successful resuscitation is dependent on many variables specific to the patient’s physiological state and
the circumstances surrounding the patient event. Failure to have a successful patient outcome is not a reliable indicator of monitor/defibrillator performance. The presence or absence of a muscular response to the transfer of energy during electrical therapy is not a reliable indicator of energy delivery or device performance.
Precautions for Manual Defibrillation Therapy
Defibrillating asystole can inhibit the recovery of natural pacemakers in the heart and completely eliminate any chance of recovery. Asystole should not be routinely shocked.
Precautions for AED Therapy
The AED algorithm is not designed to handle erratic spiking problems caused by a properly or improperly functioning pacemaker. In patients with cardiac pacemakers, the HeartStart MRx may have reduced sensitivity and not detect all shockable rhythms.
NOTE The HeartStart MRx AED mode is not intended for children under 8 years of age. For children older
than 8 years, the American Heart Association recommends that standard operating procedures for AEDs be followed. See the American Heart Association Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Dallas, Texas; AHA; 2000.

Synchronized Cardioversion Therapy

The HeartStart MRx provides synchronized cardioversion therapy by delivering a biphasic, electrical stimulus to the heart immediately following an R-wave detected in the ECG measurement. The SMART Biphasic waveform utilized in the HeartStart MRx has undergone clinical testing demonstrating its effectiveness for cardioversion of atrial fibrillation.

Noninvasive Pacing Therapy (Optional)

The HeartStart MRx provides noninvasive transcutaneous pacing by delivering a monophasic, electrical stimulus to the heart. This stimulus is intended to cause cardiac depolarization and myocardial contraction. The medical care provider selects the stimulus output and rate settings. The energy is delivered through multifunction electrode pads applied to the patient’s bare chest.
5
1 Introduction Intended Use

12-Lead ECG Analysis (Optional)

The HeartStart MRx 12-lead ECG function uses the Philips 12-Lead Algorithm to analyze acquired 12-lead ECG signals from adult and pediatric patients. The algorithm provides an analysis of amplitudes, durations, and morphologies of the ECG waveforms and the associated rhythm, based on the age and sex of the patient. Measurements and interpretive statements are offered to the clinician on an advisory basis only. This information is to be used in conjunction with the clinician’s knowledge of the patient, the results of the physical examination, the ECG waveforms, and other clinical findings.

Noninvasive Blood Pressure (NBP) Monitoring (Optional)

Oscillometric devices measure the amplitude of pressure changes in the occluding cuff as the cuff is deflated from above the systolic pressure. The amplitude suddenly increases as the pulse breaks through the occlusion in the artery. The pressure at which this occurs is very close to the systolic pressure. As the cuff pressure is decreased further, the pulsations increase in amplitude, reach a maximum (which approximates to the mean pressure), and then diminish rapidly. The index of diastolic pressure is taken where this rapid transition begins.
Studies have shown that, especially in critical cases (arrhythmia, vasoconstriction, hypertension, shock), oscillometric devices are more accurate and consistent than devices using other noninvasive measuring techniques.
The readings obtained from automatic oscillometric devices almost always deviate from those produced by the auscultatory technique. It is of particular importance to understand these deviations in environments such as Emergency and Recovery Departments where comparisons with intra-arterial measurements are seldom available.

CO2 Monitoring (Optional)

Capnometry is the measurement and numerical display of end-tidal CO2 (EtCO2) or the maximum expired CO the concentration or partial pressure of expired CO The capnograph provides information not only regarding pulmonary function, but also indirect cardiac function, ventilator function and perfusion.
concentration during a respiratory cycle. The capnograph is a graphical representation of
2
during a respiratory cycle in a “waveform” format.
2

Pulse Oximetry (SpO2) Monitoring (Optional)

A pulse oximeter is a noninvasive device that indicates the oxygen saturation (SpO2) of arterial blood. This measurement is obtained through a probe that directs red and near infrared light through arterial beds. Hemoglobin absorbs these lights differently when it is bound with oxygen. Pulse oximetry measures this difference and translates the measurement into a saturation percentage that is displayed as an SpO
reading.
2
6
Indications for Use 1 Introduction

Indications for Use

The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation.
The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
The SMART Biphasic waveform utilized in the HeartStart MRx has previously undergone clinical testing in adults. These trials support the waveform’s effectiveness for defibrillation of ventricular tachyarrhythmias at 150J. There are currently no clinical studies related to the use of the SMART Biphasic waveform in pediatric applications.

Manual Defibrillation

Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients who are pulseless and unresponsive. Synchronous defibrillation is indicated for termination of atrial fibrillation.
Asynchronous defibrillation therapy is contraindicated in patients that exhibit one or any combination of the following:
• Responsiveness
• Spontaneous breathing
•Palpable pulse

AED Therapy

An AED is to be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are:
• Unresponsive
•Not breathing
•Pulseless
An AED is not to be used on patients that exhibit one or any combination of the following:
• Responsiveness
• Spontaneous breathing
•Palpable pulse

Noninvasive Pacing Therapy

The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.
Noninvasive pacing is contraindicated in the treatment of ventricular fibrillation. Noninvasive pacing in the presence of severe hypothermia may be contraindicated.
7
1 Introduction Safety Considerations

12-Lead ECG

The 12-Lead ECG function is to be used where the clinician decides to evaluate the electrocardiogram of adult and pediatric patients as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to rule out causes for symptoms.

Noninvasive Blood Pressure

The NBP option is intended for use to determine the noninvasive measurement of arterial blood pressure for adult and pediatric patients.
NBP monitoring is contraindicated in neonatal patients or infants whose upper arm circumference is less than 13 cm.
End-tidal CO
2
The EtCO2 option is intended for noninvasive monitoring of exhaled carbon dioxide (EtCO2) and to provide a respiration rate for adult, pediatric, and neonatal patients.

Pulse Oximetry

The SpO2 option is intended for use when it is beneficial to assess oxygen saturation level for adult and pediatric patients.

Safety Considerations

General warnings and cautions that apply to use of the HeartStart MRx are provided in “Specifications and Safety” on page 213. Additional warnings and cautions specific to a particular feature are provided in the appropriate section of this guide.
WARNING Electric shock hazards exist internally. Do not attempt to open the device. Refer servicing to qualified
personnel.

Documentation and Training

Available documentation and training for the HeartStart MRx includes:
• HeartStart MRx Instructions for Use
• HeartStart MRx Quick Reference Cards
• HeartStart MRx Battery Application Note
• HeartStart MRx Web-based User Training (Located at:
www.medical.philips.com/goto/mrxtraining. Enter training access password: meetMRx.)
• HeartStart MRx User Training Video
Other Application Notes can be found on the Philips website at: www.medical.philips.com/goto/productdocumentation.
8

2Getting Acquainted

The HeartStart MRx is designed with your needs in mind. Controls, indicators, and menus are carefully organized to facilitate ease of use, and display information is tailored to the current task.
This chapter will acquaint you with the HeartStart MRx operational modes, display views, controls, and indicators. It also provides general information on use of the device.
2
NOTE If your HeartStart MRx does not have optional SpO
or 12-lead ECG transmission functionality, disregard these controls and the related information described in this chapter.
, CO2, NBP, Pacing, 12-lead ECG acquisition,
2
9
2 Getting Acquainted Basic Orientation

Basic Orientation

HeartStart MRx controls, indicators, and connections are carefully organized to facilitate ease of use.

Front Panel

The front panel contains operational controls and indicators, as shown below.
Figure 1 Basic Orientation (Front)
Label Recess
Mark Event button
Lead Select button
Display
External Power Indicator
Synchronized Cardioversion (Sync) Button
c
n
y
S
u
u
n
n
a
a
M
M
l
l
a
a
20
M
30
15
P
e
e
D
D
1-10
acer
onitor
A D
120
b
b
150
i
i
100
f
f
70
50
n
O
O
n
O
ff
AED
Ready For Use (RFU) Indicator
Therapy Knob
dult ose
170
200
S
elect
E
nergy
1
C
harge
CHARGE button
SHOCK button
2
S
hock
3
Printer (50 mm)
Printer Door
Printer Door
Latch
Print button
Speaker
Alarm Pause button
Event Summary button
Soft keys (4 total)
Navigation buttons
Menu Select button
The front panel also contains the printer and speaker.
Additional controls and indicators are on the paddles (if used) and batteries.
NOTE A palette of colored decals is included with your HeartStart MRx. These colored decals may be applied
to the label recesses located on the device handle to aid in identification. Use an indelible marker to print identification information on the decal.
10
Basic Orientation 2 Getting Acquainted

Side Panels

The left side of the HeartStart MRx has ports for monitoring cables, including ECG, pulse oximetry
). The ECG port may be used to
2
Therapy Connector
CO2 Inlet Port
CO2 Outlet Port
(SpO2), noninvasive blood pressure (NBP), and carbon dioxide (CO connect a 3-, 5-, or 10-lead patient cable. The ECG Out jack may be used to connect to an external monitor. There is also a telephone jack (for future use).
The right side of the HeartStart MRx has a therapy port for paddles (external or internal) or multifunction electrode pads. It also has a slot for a data card to transfer patient information.
Figure 2 Basic Orientation (Right/Left Sides)
2
O
C
r
c
i
M
1
2
m
a
e
r
t
s
o
G
C
E
NBP Port
ECG Port
SpO2 Port
Data Card
ECG Out (Sync) Jack
RJ11 Telephone
G
C
E
Jack
11
2 Getting Acquainted Basic Orientation

Top Panel

The top of the HeartStart MRx has a handle and basic operating instructions. If optional external paddles are present, they reside on the top panel as shown.
Figure 3 Basic Orientation (Top - with Optional Paddles)
12
Basic Orientation 2 Getting Acquainted
L
A

Back Panel

The back panel of the HeartStart MRx has two compartments for lithium ion batteries. Compartment B may instead be used to connect an AC power module. Between the battery compartments is a DC Power Input port.
The back panel also has an RS 232 serial port for 12-lead ECG transmission. The LAN port is for factory use only.
Figure 4 Basic Orientation (Back)
Battery/AC Compartment B
Battery Compartment
AN Connection
Battery
RS 232 Serial Port
AC Power Module
DC Power Input
13
2 Getting Acquainted M3538A Lithium Ion Battery

M3538A Lithium Ion Battery

The HeartStart MRx uses the M3538A Lithium Ion Battery. The battery has a fuel gauge with 5 LED indicators, each representing a charge of approximately 20% of capacity. Press the fuel gauge button to illuminate the fuel gauge.
NOTE A battery should be used as the primary power source, with AC/DC as a secondary source, if desired. If
an AC/DC power module is used as the only power source, the HeartStart MRx will take longer to charge to the desired energy level.

Battery Capacity

A new, fully-charged M3538A battery, operating at room temperature 25oC(77oF), provides approximately 5 hours of monitoring, with ECG, SpO measured every 15 minutes, or at least 50 full-energy discharges.

Battery Life

Battery life depends on the frequency and duration of use. When properly cared for, the M3538A Lithium Ion battery has a useful life of approximately 2 years. To optimize performance, a fully (or nearly fully) discharged battery should be charged as soon as possible.
and CO2 monitored continuously and NBP
2

Operating Modes

The HeartStart MRx has four clinical modes of operation, each with a customized display view. The modes are as follows:
Table 1 Operating Modes and Views
Mode of Operation Display View Description
Monitor Mode Monitoring View, or
12-Lead View
AED Mode AED View Used to analyze ECG and if necessary,
Manual Defib Mode Code View Used to perform asynchronous and
Pacer Mode Pacing View Used to perform demand or fixed mode
NOTE Upon returning to a clinical mode from a non-clinical mode such as Configuration or Data
Management, all settings are re-set to the device’s default values.
Used to monitor ECG, take an optional 12-lead ECG, and monitor optional parameters such as SpO
perform semi-automatic external defibrillation.
synchronous defibrillation (cardioversion).
pacing.
, EtCO2 and NBP.
2
14
Display Views 2 Getting Acquainted

Password Security

Access to Manual Defib Mode and Pacer Mode may be password protected (for Version A.02 and later devices) if enabled in Configuration. If enabled, you will be prompted to enter the password upon moving the Therapy Knob to either the Pacer position or an energy selection. The password is entered by using the Navigation buttons to select the password numbers and then selecting Done to complete the entry. The Charge button and the [Start Pacing] soft key remain inactive until the
C
harge
2
password is entered.
WARNING Use of the Manual Therapy Security password requires the clinician to know and remember the
password, as defined in Configuration. Failure to enter the correct password will prevent the delivery of manual defibrillation or pacing therapy. Prior to selecting this Configuration option, review this potential risk with your Risk Manager.

Display Views

The HeartStart MRx display layout is segmented as shown in Figure 5.
Figure 5 Display Layout
General Status Area
Mark Event
Inops Area
ECG/HR alarms (R or Y)
HRHR
000000
Wav e Sector 1
Wav e Sector 2
Wav e Sector 3
Wav e Sector 4
Soft Keys
ECG Wave 1ECG Wave 1
Wave 2Wave 2
SpO2SpO2 etCO2
100
Wave 3
Wave 4
Softkey#1Softkey#2Softkey#3Softkey
100
90
120120
6060
Pulse
000
02 Mar 2003 10:52Adult Paced
HH:MM:SS
NBP q 30 10:40
000/000(000)
AwRR
000
#4
50
000
30
120
50
SYS
160
90
rhythm statusrhythm status
30
8
Main Menu
Volume
Waves
Measurements/Alarms
Patient Info
Other
Exit
Parameter Block 1
Parameter Block 2
Menu Area
The content of these areas varies with the display view and the function being performed.
15
2 Getting Acquainted Display Views
y

General Status

The general status area of the display contains:
• Mark Event button label
• Patient information
Some modes of operation allow for the entry of patient information via a menu choice. If no information is entered, the patient category is defaulted to Adult, unless configured otherwise, and the pacing status is set to Non-Paced, unless the Paced status is set to Paced for an internally paced patient, or the HeartStart MRx is pacing the patient. In Pacer Mode, Paced status is not displayed.
• Date and time
• Battery power indicators
Battery icons, labeled “A” and “B”, correspond with the battery compartments located at the back of the HeartStart MRx. Each battery icon displays the current available battery power, ranging from hollow (fully discharged) to full (fully charged), as shown in Figure 6. If the AC Power Module is inserted in Compartment B, the no battery icon is displayed.
Figure 6 Battery Charge Level Indicators
AA
No Battery Empty Battery 25% Capacity 50% Capacity 75% Capacity 100% Capacit
A
AAA
• Inop statements
Inop statements appear in the top left of the display if equipment problems are detected.
• ECG/HR alarm status
Alarm messages communicate arrhythmia alarms, as well as overall alarm status (alarms off, alarms paused).
•Event Timer
An Event Timer communicates the elapsed time since the device was turned on.
16
Display Views 2 Getting Acquainted

Wave Sectors

The HeartStart MRx is configured to populate each wave sector with a predetermined waveform, when powered on in Monitor, Manual Defib, or Pacer Mode. Figure 7 shows the default Waves configuration choices. See “Configuration” on page 125 for more information.
A dashed line in a wave sector or an empty wave sector indicates that the waveform source is not connected to the HeartStart MRx. Wave sectors may contain a variety of information, as appropriate to the parameter, view, and task. Additionally, ECG wave sectors contain a calibration bar.
Figure 7 Default Waves Configuration
Wave Sector 1
Wave Sector 1 will only contain an ECG waveform. This is the waveform used by the arrhythmia, heart rate derivation, and AED analysis algorithms. This waveform may be acquired through the therapy port for pads/paddles or the monitoring port for 3-, 5-, or 10-lead electrodes. If the configured source is not connected to the HeartStart MRx when turned on, the first valid ECG source is displayed in Wave Sector 1. Once the configured source is available, it automatically populates Wave Sector 1. Should the configured source then become invalid, a Leads Off condition is displayed. The HeartStart MRx does not revert to the initial source of ECG.
NOTE If Pads are configured as the primary ECG source for Wave Sector 1, the ECG patient cable must be
connected to the MRx and to the monitoring electrodes on the patient in order to change the ECG source to a Leads selection.
NOTE When monitoring using a 3-lead ECG set, the MRx displays only 1 ECG lead at a time.
17
2 Getting Acquainted Display Views
Wave Sectors 2-4
Wave Sectors 2 through 4 are automatically populated when parameter sources (cables/tubing) are connected to the HeartStart MRx. If the parameter source is the configured choice of a particular wave sector, it is displayed in that wave sector. If you connect a parameter source that is not configured to be displayed, it is displayed in the first empty wave sector. If you subsequently connect the configured parameter source, it does not replace the current parameter, instead it populates the first available empty wave sector.
Changing Displayed Waveforms
Wave Sector 1 has a dedicated Lead Select button to change the displayed lead/source. Waveforms displayed in other wave sectors may be changed for the current patient through the display menu. See “Display Menus” on page 19.

Parameter Blocks

Measurements for displayed waveforms and monitored parameters are provided in the parameter blocks. Parameter Block 1 always contains the heart rate and HR alarm settings. The NBP schedule, measurements, and alarm settings are also displayed in Parameter Block 1. Parameter Block 2 contains measurements and alarm settings for SpO measurement is labeled and displays the current value and the high and low alarm limit settings or the
(alarms off) icon. A “-?-” is displayed until a valid measurement can be obtained.
, Pulse, EtCO2 and Airway Respiration Rate (AwRR). Each
2
and EtCO2 measurements are activated when the associated parameter cable/tubing is
SpO
2
connected. When the SpO and requests approval to turn off the measurement.
Parameter Alarm Messages are displayed in the space above each numeric value, replacing the parameter label.

Soft Key Labels

The four soft key labels correspond to the soft key buttons located immediately below. These labels change, as appropriate, according to the current display view and function. Soft key labels appearing as grey text indicate that the soft key is inactive.
and EtCO2 cable/tubing is disconnected, a prompt message informs you
2
18
Display Views 2 Getting Acquainted

Display Menus

Menus with controls and options specific to each function of the HeartStart MRx are easily accessible through the Menu Select and Navigation buttons located on the front panel. Menus are used to adjust volume, select waveforms for display, set alarms, schedule measurements, enter patient information, perform the Operational Check, generate reports, and a variety of other tasks. Menus and submenus are organized to allow you to conveniently make selections and enter information.
To display a menu, press the Menu Select button. Then use the up or down Navigation buttons to scroll through the available choices until the desired selection is highlighted. To activate the selection, press the Menu Select button. Press Exit to close the menu without activating a selection.
Figure 8 Sample Menu
HR/Arrhythmia
Relearn Rhythm
Alarms Off
HR Limits
VTACH Limits

Message Windows

Periodically, message windows appear on the display to provide additional status information, alert you to an error or a potential problem, or direct you to take action. Remain alert to these messages. If a response is required, as shown in Figure 9, use the Navigation and Menu buttons to highlight and select the appropriate answer.
Figure 9 Sample Message Window
Exit
Configuration Not Saved - Exit Anyway?
Yes
No
19
2 Getting Acquainted Controls

High Contrast Display

To optimize visibility of the MRx display when used in bright sunlight, Version A.02 and later devices provide a High Contrast feature which may be enabled. In this view, the MRx display appears using a yellow background with all other screen elements appearing in black or shades of gray. High Contrast is enabled in Manual Defib, Pacer and Monitor Modes by pressing the Menu Select button and selecting High Contrast On from the Main Menu.
NOTE The High Contrast feature does not display colors configured as red or blue, therefore, be sure your
device is configured correctly with the appropriate parameter color settings. See Table 16 on page 132 in Configuration.

Controls

The Therapy Knob is used to turn the HeartStart MRx on in the desired mode of operation. Operating controls are organized by function, with general function buttons located along the left and bottom sides of the display, defibrillation controls to the right of the display, and soft keys immediately below the display. (See Figure 1.)

Therapy Knob

The Therapy Knob serves as the power switch for the HeartStart MRx. It can be set to:
•Off
AED - to enable AED Mode for automated external defibrillation.
Monitor - to enable Monitor Mode for 3- or 5-lead ECG monitoring, 12-lead ECG acquisition, or
monitoring of optional parameters such as SpO2, CO2, and NBP.
Pacer - to enable Pacer Mode (optional) for demand or fixed mode pacing.
Manual Defib - to enable Manual Defib Mode for asynchronous or synchronous defibrillation
(cardioversion) at the selected energy setting.
In Manual Defib Mode the defibrillation energy settings are labeled as 1-9, 10, 15, 20, 30, 50, 70, 100, 120, 150, 170, and 200 Joules. If your unit is equipped with the optional pacing function, the energy settings are labeled as 1-10, 15, 20, 30, 50, 70, 100, 120, 150, 170, and 200 Joules.
20
Loading...
+ 231 hidden pages