Philips HeartStart MRx User manual

MONITOR/DEFIBRILLATOR
CONTROLS, CONNECTIONS
AND INDICATORS
CONTROLS, CONNECTIONS AND INDICATORS
Mark Event
Lead Select
Alarm Pause
Event Summary
CO2 Inlet
CO
2
Outlet
Temp
ECG Out (Sync) Jack
External Power
Soft Keys
Navigation
Soft Keys
1
2
O
C
2
E
E
C
C
G
G
SYNC
c n
y
S
n
n
a
a
M
M
Invasive Pressure 1
Invasive Pressure 2
b
b
i
i
100
f
f
e
e
70
D
D
l
l
a
a
50
u
u
30
20
15
1-10
O
P
acer
M
onitor
Menu Select
NBP
ECG
A
dult
D
ose
120
150
170
200
n
O
n
O
ff
AED
SpO2
Ready For Use (RFU)
S
elect
Therapy Knob
E
nergy
1
C
harge
2
Charge
S
hock
3
Shock
Speaker
Battery/AC Compartment B
Printer
Printer Door Latch
Print
LAN Connection
RS 232 Serial Port
Data Card
AC Power Module
Therapy Connector
Bed Rail Hook Mount
Battery Compartment A
Battery
DC Power Input
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
12-LEAD ECG
12-LEAD ECG
Electrode Placement for 12-Lead ECG
Acquiring the 12-Lead ECG
1. Turn the Therapy Knob to Monitor.
2. Press the 12-Lead soft key.
3. Check for good signal quality.
4. Press the Start Acquire soft key. If prompted, enter patient age and sex.
5. Keep the patient still during the 10-second acquisition period.
Accessing Stored 12-Lead Reports
1. Press the Menu Select button.
2. Use the Navigation buttons to select Reports. Reports for the current patient are listed by date and time.
3. Use the Navigation buttons to select a report and press the Menu Select button.
4. Select Print, Copy, or Delete, and press the Menu Select button.
5. To select another report, repeat steps 1 through 4.
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
ALARMS
USING ALARMS
Responding to Alarms
1. Check the patient.
2. Identify the alarm(s) indicated.
3. Silence the alarm(s) by pressing the Menu Select button or one of the Navigation buttons.
4. Address the alarm condition by selecting one of the following options: Alarms Off - Turn the monitoring parameter's alarms off, displaying next to the parameter value. Acknowledge - Recognize that the alarm condition is present, without further action. As the default option, Acknowledge always appears when an alarm occurs. New Limits - Adjust parameter alarm limits accordingly.
Note: For only Heart Rate/Arrhythmia alarms, your options are either Acknowledge or Acknowledge and New Limits, depending on the alarm.
Checking/Adjusting Alarm Limits
Alarm limits are preset for your device based on its configuration and the patient type. When alarms are on, alarm limits are visible next to the measurement's numeric value.
To adjust an alarm limit setting:
1. Press the Menu Select button.
2. Select Measurements/Alarms and press the Menu Select button.
3. Select the desired measurement from the menu and press the Menu Select button.
4. Select (measurement) Limits from the menu and press the Menu Select button.
5. Adjust the limit and press the Menu Select button.
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
MONITORING
CARBON DIOXIDE
MONITORING CO
Using the Nasal FilterLine
1. Attach the FilterLine tubing to the CO2 Inlet port.
2. Check that both nostrils are clear.
3. Position the nasal FilterLine on the face by inserting the tips into the nostrils.
4. Pass the FilterLine tubing over the ears, then slide the sleeve up the tubing toward the neck to a comfortable fit under the chin.
5. If using dual purpose FilterLine tubing, connect the green tubing to the oxygen source.
6. Check the positioning of the FilterLine regularly to ensure proper function.
Using the FilterLine and Airway Adapter
1. Attach the FilterLine tubing to the CO2 Inlet port.
2. Connect the wide end of the airway adapter to the endotracheal tube.
3. Connect the narrow end of the airway adapter to the ventilator tubing or manual resuscitator.
Measuring EtCO2 and AwRR
The CO2 waveform is automatically displayed in the configured Wave Sector when the FilterLine is connected to the CO2 Inlet port. The measurement values for End-Tidal CO2 (EtCO2) and Airway Respiration Rate (AwRR) are displayed in Parameter Block 2 as shown below.
2
SpO2SpO
2
100
%
100
90
EtCO2
34
mmHg
50 30
AwRR
18
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
rpm
30
8
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
MONITORING
MONITORING
ECG
ECG
MONITORING ECG
Selecting the Lead for ECG Monitoring
It is important to select a suitable lead for monitoring so that a QRS complex can be accurately detected.
Guidelines:
- QRS tall and narrow
+
I
III
II
aVR
+
I
aVL
III
aVF
II
- R-wave above or below baseline (but not bi-phasic)
- T-wave smaller than 1/3 R-wave height
- P-wave smaller than 1/5 R-wave height
+
3-Lead Placement
+
5-Lead Placement
+
+
- For non-transcutaneously paced patients, the pace pulse should be large enough to be detected with minimal repolarization and not wider than the normal QRS complex.
To select a lead for the primary wave sector:
1. Press the Lead Select button.
2. Cycle through lead choices.
To select a lead for an additional wave sector:
1. Press the Menu Select button.
2. Select Waves and press the Menu Select button.
3. Select the wave sector and press the Menu Select button.
4. Select the desired lead source and press the Menu Select button.
5. If needed, adjust the ECG wave size (gain) and press the Menu Select button. Note: Arrhythmia analysis occurs at x1 gain (no matter what the display gain is).
Changing Pacing Status
When monitoring patients with non-transcutaneous pacemakers, it is important to set the paced status correctly. (See MRx Instructions for Use for additional information.)
To change pacing status:
1. Press the Menu Select button.
2. Select Patient Info and press the Menu Select button.
3. Select Paced and press the Menu Select button.
4. Select Ye s/No and press the Menu Select button.
453564042471
*453564042471*
MONITOR/DEFIBRILLATOR
Edition 1
*1*
MONITOR/DEFIBRILLATOR
MONITORING
INVASIVE PRESSURES
MONITORING INVASIVE PRESSURES
Changing a Pressure
1. Press the Menu Select button.
2. Using the Navigation buttons, select the Measurement/Alarms menu and press the Menu Select button.
3. Select Press 1 (or Press 2) and press the Menu Select button.
4. Select Label and press the Menu Select button.
5. Select the appropriate label from the list and press the Menu Select button.
6. Select the size of the scale, the Alarm Source type and high/low Alarm Limits.
Pressure Label Options
ABP - Arterial Blood Pressure ART - Arterial Blood Pressure Ao - Aortic Pressure CVP - Central Venous Pressure ICP - Intracranial Pressure* LAP - Left Atrial Pressure PAP - Pulmonary Artery Pressure RAP - Right Atrial Pressure P1 - Non-specific Pressure P2 - Non-specific Pressure Label (Channel 1) Label (Channel 2) * Cerebral Perfusion Pressure (CPP) is automatically displayed with ICP when one pressure is already set to ABP, ART or Ao and the other pressure is set to ICP.
Zeroing
1. Place the transducer at the appropriate level for the measurement site.
2. Close the transducer stopcock to the patient and vent the transducer to atmospheric pressure. Then pick one of the following methods: Using Menu Select button Using a soft key in Monitor Mode
3. Press the Menu Select 3. Press the button under [Zero button. Pressure] on the display.
4. Using the Navigation button, 4. Using the Navigation buttons, select Measurement/Alarms, select the pressure(s) to zero press the Menu select button from the Zero menu and press
5. Select the pressure label to the Menu Select button. zero. 6-Highlight Zero and press the Menu Select button.
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
MONITORING
NONINVASIVE BLOOD
PRESSURE
MONITORING NBP
Preparing
1. Select the appropriately sized cuff.
2. Attach the cuff to the NBP tubing.
3. Insert the NBP tubing into the NBP port.
4. Apply the blood pressure cuff to the patient's arm or leg. a. Ensure the cuff is completely deflated. b. Wrap the cuff around the limb. c. Check that the edge of the cuff falls within the size range markings.
5. Place the limb at the same level as the patient's heart.
Changing the NBP Schedule
1. Press the Menu Select button.
2. Using the Navigation buttons, select the Measurement/Alarms menu and press the Menu Select button.
3. Select NBP and press the Menu Select button.
4. Select NBP Schedule and press the Menu Select button.
5. Select the desired interval or Manual, and press the Menu Select button. Interval Options: 1, 2.5, 5, 10, 15, 30, 60 and 120 minutes. A Manual NBP measurement may be taken at any time by pressing the Start NBP soft key.
Measuring NBP
To measure NBP, press the Start NBP soft key. Cuff pressure is displayed as the cuff inflates and deflates. When the measurement completes, the NBP values are displayed.
HRHR
82 82
NB
120120
6060
PmmHgM
120/80 (100)
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
anua
SYS 10:
160
90
l
40
MONITOR/DEFIBRILLATOR
OPERATIONAL
CHECK
OPERATIONAL CHECK
Running the Operational Check
1. Insert a charged battery into the HeartStart MRx.
2. Turn the Therapy Knob to 150J.
3. Press the Menu Select button.
4. Using the Navigation buttons, select Other and press the Menu Select button.
5. Select Operational Check and press the Menu Select button.
6. Select Run Op Check and press the Menu Select button.
7. Press the Menu Select button to acknowledge the message Leaving Normal Operating Mode.
8. Carefully read and respond to the prompts.
At the completion of the Operational Check, a report is printed that lists the test results and prompts you to visually inspect the device and cables, and to inventory all accessories and supplies.
Note: For device versions greater than or equal to B.05 with Pacing, run Op Check with a pads cable and test external paddles using the Weekly Shock Test.
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
NONINVASIVE PACING
(TRANSCUTANEOUS PACING)
NONINVASIVE PA C I N G
P
ace
d
R
hythm
(TRANSCUTANEOUS PACING)
Demand Mode Pacing (Requires ECG leads and pads)
1. Turn the Therapy Knob to Pacer.
2. Press the Lead Select button to select the desired lead.
3. Verify white R-wave markers appear above or on the ECG waveform.
4. Press the Pacer Rate soft key, select the desired number of paced pulses per minute, and press the Menu Select button.
5. If needed, press the Pacer Output soft key, use the Navigation buttons to adjust the initial pacer output, and press the Menu Select button.
6. Press the Start Pacing soft key.
7. Verify white pacing markers appear in front of every QRS.
8. If cardiac capture is not obtained, press the Pacer Output soft key, increase the output until capture occurs, decrease the output to the lowest level that still maintains capture, and press the Menu Select button.
9. Verify the presence of a peripheral pulse.
IIII
Fixed Mode Pacing (Can be done with pads only)
1. Turn the Therapy Knob to Pacer.
2. Set the pacer mode to Fixed.
- Select Pacer Mode from the Main Menu.
- Select Fixed and press the Menu Select button.
3. If using leads, press the Lead Select button to select the desired lead.
4. Press the Pacer Rate soft key, select the desired number of paced pulses per minute, and press the Menu Select button.
5. If needed, press the Pacer Output soft key, use the Navigation buttons to adjust the initial pacer output, and press the Menu Select button.
6. Press the Start Pacing soft key.
7. Verify cardiac capture. If it is not obtained, press the Pacer Output soft key, increase the output until capture occurs, decrease the output to the lowest level that still maintains capture, and press the Menu Select button.
8. Verify the presence of a peripheral pulse.
NOTE: If monitoring SpO2 when pacing, activate the Pulse alarms to assess for peripheral perfusion through the SpO2 transducer.
Demand Mode
P
ace
d
R
hythm
Capture
453564042471
*453564042471*
Edition 1
*1*
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
TM
Q-CPR
Q-CPR
Prior to Use
0-8
1. Verify the Pads/CPR cable is connected to the MRx Therapy port.
2. Verify the Compression Sensor is connected to the Pads/CPR cable.
3. Verify a new Compression Sensor Adhesive Pad is applied to the Compression Sensor.
Preparing for Q-CPR
1. Connect the multifunction electrode pads to the Pads/CPR cable.
2. Apply the multifunction pads to the patient in the anterior/anterior position.
3. Position the Compression Sensor on the patient as shown on the green liner (i.e., on the lower half of the sternum).
Q-CPR in Manual Defib Mode
- Q-CPR starts automatically when you turn the Therapy Knob to 150J or perform a chest compression with the Compression Sensor.
- To manually activate Q-CPR, press the Start CPR soft key.
Q-CPR in AED Mode
- Q-CPR starts automatically when you perform a chest compression with the Compression Sensor during the CPR Pause period. If necessary, press the Pause for CPR soft key to activate the Pause period.
- AED Mode provides only corrective text and voice prompt feedback.
Q-CPR™ is a trademark of Laerdal Medical.
MONITOR/DEFIBRILLATOR
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
READY FOR USE INDICATOR
READY FOR USE
INDICATOR
The Ready For Use (RFU) indicator is located on the upper, right corner of the device. It indicates the status of the therapy delivery functions of the monitor/ defibrillator using the following definitions:
A blinking black hourglass symbol indicates the shock, pacing, and ECG functions of the device are ready for use. Sufficient battery power is available for device operation.
A blinking red “X” and a periodic audio chirp indicate no battery is present or a low battery condition. If a battery is inserted and charging, the audio chirp is not present.
A solid red “X” and a periodic audio chirp indicate a failure has been detected that may prevent the delivery of a shock, pacing, or ECG acquisition. An error message will display at power-on.
A solid red “X” without periodic audio chirps indicates either there is no power available, or a device failure has occurred. If, after power is supplied, the indicator reverts to the blinking black hourglass symbol, the device is once again ready for use.
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
MONITORING
SpO2
MONITORING SpO
2
Applying the Sensor
Follow the manufacturer's directions for applying and using the sensor, making sure to observe any warnings and cautions. For best results: Make sure the sensor is dry. If the patient is moving, secure the sensor cable loosely to the patient. Make sure the transducer is not too tight. Keep power cables away from the sensor cable and connections. Avoid placing the sensor in an environment with bright lights. If necessary, cover the sensor with opaque material. Avoid placing the sensor on an extremity with poor perfusion.
Monitoring SpO
2
1. Connect the appropriate sensor cable to the HeartStart MRx.
2. Apply the sensor to the patient.
3. If the HeartStart MRx is not turned on, turn the Therapy Knob to Monitor.
The pleth wave appears after a few seconds, followed by the oxygen saturation value. As the patient's oxygen saturation changes, the SpO2 value is continuously updated and the associated waveform is displayed. See the Alarms Quick Card for information on adjusting alarms.
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
MONITORING
TEMPERATURE
MONITORING TEMPERATURE
To Monitor a Temperature
1. Connect the temperature cable to the HeartStart MRx.
2. Select the correct temperature label for your measurement.
3. Check that the current device settings (including alarm settings) are appropriate for the patient.
4. Apply the temperature probe to the patient.
Selecting a Temperature Label
1. Press the Menu Select button.
2. Using the Navigation buttons, select the Measurement/Alarms menu and press the Menu Select button.
3. Select the temperature label option currently assigned to your measurement (default is Temp) and press the Menu Select button.
4. Select Label and press the Menu Select button.
5. Select the appropriate label from the list provided and press the Menu Select button.
Temperature Label Options
Tesoph - esophageal temperature Trect - rectal temperature Tskin - skin temperature Temp - non-specific temperature label Tcore - core temperature Tnaso - nasopharyngeal temperature Tvesic - urinary bladder Tart - arterial temperature Tven - venous temperature
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
MONITOR/DEFIBRILLATOR
VITAL SIGNS
TRENDING
VITAL SIGNS TRENDING
Viewing Trending Data
1. Place the HeartStart MRx into Monitor Mode.
2. Press the Menu Select button.
3. Using the Navigation buttons, select Trends and press the Menu Select button. The Vital Signs Trending Report appears in the bottom two wave sectors.
Scrolling
LEFT/RIGHT: Use the << or >> soft keys to scroll left and right (backward/ forward) in the report screen. The soft key will be inactive if there is no more data to be viewed in that direction. UP/DOWN: If there are more vital signs than can be shown on the screen, use the Navigation buttons next to the Menu Select button to scroll up and down with the vertical scroll bar on the display. (Make sure there are no menu items active at the time.)
Reading Trending Data
- When trending is initially displayed, the latest (most recent) trending data will appear in the far right column; older data flows to the left.
- The display will auto-update as new data becomes available as long as the latest data is displayed on the screen.
- A -?- indicates that the parameter has invalid information.
- A ? just before a numeric value indicates questionable data.
- An empty space indicates unavailable data.
Printing the Vital Signs Trending Report
You can print the Vital Signs Trending Report two ways:
1. Press the soft key under the Print Trends label. A report for the displayed period and interval is printed.
2. Press the Summary button and select Trends and then Trends Interval. A report for the entire incident period is printed.
453564042471 Edition 1 September 2006
*453564042471*
MONITOR/DEFIBRILLATOR
*1*
MONITOR/DEFIBRILLATOR
WEEKLY SHOCK TEST
WEEKLY SHOCK TEST
How to perform the Weekly Shock Test
Test each type of patient therapy cable used (multifunction electrode pads or paddles). If you are using paddles, make sure the paddles are secure in their pockets and that the Patient Contact Indicator (PCI) LEDs located on the sternum paddles are not lit. If the LEDs light, adjust the paddles in their pockets. If the LEDs continue to light, clean both adult and pediatric paddle electrode surfaces. If you are using multifunction electrode pads, attach a test load to the end of the patient Therapy Cable.
1. Turn the Therapy Knob to 150J
2. Press the Charge Button. NOTE: If it becomes necessary to disarm the defibrillator, press [DISARM]
3. If using:
- Pads, press the Shock button on the HeartStart MRx to deliver a shock into the test load.
- External paddles, simultaneously press the shock buttons located on the paddles to deliver a shock into the test load.
4. The strip prints immediately if configured to do so. If the strip does not print immediately, press the Print button.
5. Confirm on the printed strip that the energy delivered to the test load is 150J +/- 23J (127J to 173J). If not, take the device out of use and call for service.
453564042471 Edition 1 September 2006
*453564042471* *1*
MONITOR/DEFIBRILLATOR
Loading...