The Philips cm200 TEM uses a field emission gun (FEG) which provides very
high resolution but requires more care than a standard TEM filament. It is very
important that a high vacuum is maintained in the gun chamber at all times.
The gun chamber is sealed off from the rest of the column using the gun valve.
The gun valve must be closed whenever a specimen is being exchanged. If you
have a problem with the operation of the instrument or if you need to leave the
room, please close the gun valve by turning the control knob slowly clockwise
to the closed position.
The computer screen:
The cm200 contains many manual controls, such as focus and magnification.
However, many functions are controlled via a computer interface, using the
“softkeys” at the edges of the screen. To select a particular function press the
softkey next to the appropriate function title. To leave a page or function, press
the READY button at the lower right hand side of the screen. In most cases, this
will take you back to the HR-TEM BRIGHT FIELD page.
The functions described here are those required for most users. The other
functions must not be used unless you have received appropriate training.
DO NOT PRESS A SOFTKEY IF YOU DON’T KNOW WHAT IT DOES.
INT ZOOM
INT LIMIT
-
-
-
-
-
-
HR-TEM BRIGHT FIELD
LM
HT
spot 1
focusstep1
defocus -4.30um
plate auto
meter XXX s
exp no
stoc
X: -141.61 um
Y: -78.72 um
Z: 135.69 um
: 0.00 d
B: 0.00 d
1000x
200k
350nm
E7773
50
MODES
PARAMETERS
UTOCON-1
COMPUSTAGE
RSET DEFOC
MEASURING
VACUUM
TEM CAMERA
HR-TEM BRIGHT FIELD page
¾ This page allows access to most of the functions of the microscope.
¾ The page is made up of three columns. The central column shows the
magnification, spotsize, focus step, film exposure information, film
stock levels, and stage positions.
¾ The right-hand column allows access to other pages in the program
via the softkeys. The most commonly required are COMPUSTAGE,
VACUUM STATUS and MODE SELECTION (MODES).
5
cm200 user guide 2009
COMPUSTAGE page
A
A
k
V
¾ The
COMPUSTAGE allows stage position to be monitored and
controlled. Stage positions can be stored and recalled. The central
column on this page contains the same information as on the HRTEM BRIGHT FIELD page.
¾ To store a stage position, select a number in the right-hand column
on the monitor using a softkey (when selected, the number is
highlighted). Press the STORE softkey to save the position.
¾ To return to a saved position, select the appropriate number and
press the RECALL softkey on the right-hand side of the page. The
stage will automatically move to that position.
¾ To clear a saved stage position, select the appropriate number and
press the CLEAR softkey on the left-hand side of the page.
¾ To clear all saved stage positions, press the CLEAR ALL softkey twice
(located on the left-hand side of the page).
¾ The COMPUSTAGE page also contains functions for setting eucentric
height (A-WOBBLER) and for returning the specimen to zero tilt
(RESET AB). The RESET AB softkey must always be pressed before a
specimen is removed.
COMPUSTAGE REGISTER CONTROL
COMPUCTRL
Z DISPL = 0
Z DISPLAY
USER REAL
CLEAR
CLEAR ALL
XY RECALL-
RESET AB
-WOBBLER
LM
HT
spot 1
focusstep
plate auto
meter XXX s
exp no
stoc
X: -141.61 um
Y: -78.72 um
Z: 135.69 um
: 0.00 d
B: 0.00 d
1000x
200k
350nm
-defocus 44.30um
E7773
50
STORE
1
2
3
RECALL
6
cm200 user guide 2009
VACUUM STATUS page
A
¾ The
VACUUM STATUS page shows a diagram of the vacuum system.
The status of the vacuum can be read at the top of the screen. Below
the diagram of the vacuum system is a series of pressure readings,
including the IGP (ion getter pump) reading.
P1: 32 P3:34
P2: 49
¾ Before the gun valve is opened, the IGP reading must be less than 27
and the VACUUM STATUS must be ‘READY’.
MODE SELECTION page
¾ The modes page allows the user to select the TEM mode to be used.
Usually this is HR-TEM. The CONFIGURATION page can be
accessed from this page.
CONFIGURATION page
¾ This page contains information on the state of the microscope
(STANDBY or OPERATE) and on the extraction voltage. As the filament is
turned up, the change in extraction voltage (“Actual”) can be
monitored on this page.
IGP:5
STANDBY
EXTR LIMIT
CONFIGURATION
FEG CONTROL
State: STANDBY
Time: 0 min
: 4.5 kV
ctual: 2.29 kV
--
DISPLAY
FIL OFF
7
cm200 user guide 2009
2. CHECK cm200 STATUS
1. Log in to the EMU Booking System on the PC.
2. Check that the vacuum is OK by confirming that both the UHV and HV
indicators on the far right side of the panel are illuminated. If not, check with EMU staff.
3. Go to the VACUUM STATUS page (from the HR-TEM page, press the
VACUUM softkey). Check that the vacuum status is READY and that the IGP
is less than 27. If not, check with EMU staff.
4. Return to the HR-TEM page (press READY).
5. Check that the High Tension indicator on the right side of the panel is lit. If not, check with EMU staff.
6. Fill the liquid nitrogen dewar and place it under the cold finger. You must
wear safety glasses and protective gloves when handling liquid nitrogen.
7. Ensure that condenser aperture 4 and objective aperture 7 are selected and
that the diffraction aperture is out (swing shift to the right).
1. Press the MODES softkey and then the CONFIGURATION softkey to go to
the CONFIGURATION page.
2. Turn the filament knob on the right-hand side of the
display screen slowly clockwise (three or four clicks at a
time).
Note: If the voltage is increased too quickly, the kV value
will freeze and you will need to seek assistance from
EMU staff.
Monitor the extraction voltage on the screen. When the
voltage approaches its limit (~4.5 kV), you will hear a
beep. Make sure you reach the maximum possible
voltage by turning the knob a few more clicks. The
voltage will usually reach 4.48 kV.
3. Press the READY button and then the HR-TEM softkey to return to the HRTEM BRIGHT FIELD page.
High tension
On
Off
Filament
3
cm200 user guide 2009
9
4. LOAD THE SPECIMEN
There are two specimen holders available:
1) A single-tilt holder for conventional imaging and EDS analysis.
2) A low-background double tilt holder for crystallographic studies and
specialised EDS analysis (see appendix a).
When you begin your cm200 session the single-tilt holder will normally be in
the microscope.
Remove the single-tilt holder:
1. CHECK THAT THE GUN VALVE IS CLOSED. If not, close it by turning the
switch slowly clockwise.
2. Pull the specimen holder out until it cannot
be pulled any further (~8cm).
3. Rotate the holder clockwise as far as it will
go (~120º).
4. Pull the holder gently out of the airlock.
Take care not to knock the specimen holder against the airlock walls as it is
released.
5. Place the specimen holder in the support. Do not touch the brass part of the
specimen holder with your fingers.
6. Lift the specimen securing clamp using the special tool.
7. Load the specimen into the recessed ring, then gently lower the securing
clamp using the special tool.
8. Rotate the specimen holder slightly and gently tap it to ensure that the
specimen stays in place.
1
2
3
Insert the single-tilt holder:
1. CHECK THAT THE GUN VALVE IS CLOSED. If not close it by turning the
switch slowly clockwise.
2. Check the IGP reading by going to the VACUUM STATUS page. The IGP
reading should be less than 27. If not, check with EMU staff.
cm200 user guide 2009
10
3. Insert the rod gently into the specimen airlock. Orient the rod so that the
large key peg is at the 11 o’clock position. Push the rod into the airlock until
you feel a stop and the red light on the airlock illuminates (this indicates that
the airlock is being pumped). After the pump has started it is often useful to
gently wiggle the rod, which allows it to be inserted another few
millimetres.
4. Wait until the red light has turned off.
5. Rotate the rod anti-clockwise (~120º) keeping a firm grip on the holder. The
vacuum will ‘grab’ the rod when it gets to the 6 o’clock position.
6. Guide the rod as it enters the airlock. The key peg should slide into the slot
on the airlock. Monitor any change in vacuum by noting the IGP value. (The
IGP value may rise above 27 during rod insertion – it must be 27 before
operation).
cm200 user guide 2009
11
Loading...
+ 25 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.