Pfeiffer Hicube ECO User Manual

OPERATING INSTRUCTIONS
EN
HICUBE ECO
Turbo pumping stations
Translation of the Original
Dear Customer,
Thank you for choosing a Pfeiffer Vacuum product. Your new turbo pumping station is designed to support you by its performance, its perfect operation and without interfering your individual application. The name Pfeiffer Vacuum stands for high-quality vacuum technology, a comprehensive and complete range of top-quality products and first-class service. With this expertise, we have acquired a multitude of skills contributing to an efficient and secure implementation of our product.
Knowing that our product must not interfere with your actual work, we are convinced that our product offers you the solution that supports you in the effective and trouble-free execution of your individual application.
Please read these operating instructions before putting your product into operation for the first time. If you have any questions or suggestions, please feel free to contact info@pfeiffer- vacuum.de.
Further operating instructions from Pfeiffer Vacuum can be found in the Download Center on our website.
Disclaimer of liability
These operating instructions describe all models and variants of your product. Note that your product may not be equipped with all features described in this document. Pfeiffer Vacuum constantly adapts its products to the latest state of the art without prior notice. Please take into account that online operating instructions can deviate from the printed operating instructions supplied with your product.
Furthermore, Pfeiffer Vacuum assumes no responsibility or liability for damage resulting from the use of the product that contradicts its proper use or is explicitly defined as foreseeable misuse.
Copyright
This document is the intellectual property of Pfeiffer Vacuum and all contents of this document are protected by copyright. They may not be copied, altered, reproduced or published without the prior written permission of Pfeiffer Vacuum.
2/60

Table of contents

1 About this manual 7
1.1 Validity 7
1.2 Target group 7
1.3 Conventions 7
2 Safety 10
2.1 General safety instructions 10
2.2 Safety instructions 10
2.3 Safety precautions 13
2.4 Limits of use of the product 13
2.5 Proper use 14
2.6 Foreseeable improper use 14
Table of contents
1.1.1 Applicable documents 7
1.1.2 Variants 7
1.3.1 Instructions in the text 7
1.3.2 Pictographs 8
1.3.3 Stickers on the product 8
1.3.4 Abbreviations 9
3 Product description 15
3.1 Identifying the product 15
3.2 Scope of delivery 16
3.3 Function 16
3.3.1 Control interface 16
3.3.2 Drive 17
3.3.3 Cooling 17
4 Transportation and Storage 18
4.1 Transporting the pumping station 18
4.2 Transport protection 18
4.3 Storing the pumping station 19
5 Installation 20
5.1 Preparing for set-up 20
5.2 Setting up the pumping station 20
5.3 Anchoring pumping station 20
5.4 Connecting the high vacuum side 21
5.4.1 Designing the counter flange 21
5.4.2 Using accessories for the high vacuum connection 22
5.4.3 Connecting the turbopump externally 23
5.4.4 Installation of ISO-KF flange 23
5.4.5 Attaching ISO-K flange onto ISO-K 24
5.4.6 Installation of ISO-K flange onto ISO-F 24
5.4.7 Fastening CF flange to CF-F 26
5.5 Connecting the exhaust side 27
5.6 Modifying the DCU for remote operation 28
5.7 Connecting accessories for the pumping station 29
5.8 Removing transport lock 29
5.9 Grounding the pumping station 30
5.10 Establishing the mains connection 30
6 Operation 32
6.1 Commissioning 32
6.2 Switching on the turbo pumping station 33
6.3 Normal operation 33
6.4 Operation with gas ballast (applies to MVP 015-2 only) 33
6.5 Operation monitoring 35
3/60
Table of contents
6.5.1 Operating mode display via LED 35
6.5.2 Temperature monitoring 35
6.6 Switching off and venting 35
6.6.1 Shutting down the turbo pumping station 35
6.6.2 Venting 36
7 Maintenance 37
7.1 General maintenance information 37
7.2 Maintenance intervals and responsibilities 37
7.3 Replacing mains fuses 37
7.4 Removing components for maintenance 38
7.4.1 Removing the electrical connections 39
7.4.2 Dismantling the turbopump from the pumping station 39
8 Decommissioning 41
8.1 Shutting down for longer periods 41
8.2 Recommissioning 41
8.3 Disposing of the vacuum pump 41
9 Malfunctions 42
9.1 General 42
9.2 Troubleshooting 42
9.3 Error codes 43
10 Service solutions from Pfeiffer Vacuum 47
11 Accessories 49
12 Technical data and dimensions 50
12.1 General 50
12.2 Technical data 50
12.3 Dimensioned drawings 55
Declaration of Conformity 58
4/60

List of tables

Tbl. 1: Stickers on the product...............................................................................................8
Tbl. 2: Abbreviations used in the document.......................................................................... 9
Tbl. 3: Limits of use for turbopump stations.........................................................................14
Tbl. 4: Pumping station component combination options....................................................16
Tbl. 5: Requirements for on-site high vacuum connection.................................................. 22
Tbl. 6: Changing pumping speed with splinter shield or protective screen..........................22
Tbl. 7: Factory setting of key parameters on delivery..........................................................33
Tbl. 8: Description of the key functions of the DCU.............................................................33
Tbl. 9: LED display and meaning at the DCU......................................................................35
Tbl. 10: Factory settings for delayed venting of turbopumps.................................................36
Tbl. 11: Preset accessory connections..................................................................................38
Tbl. 12: Troubleshooting........................................................................................................43
Tbl. 13: Warning and error messages when using the DCU................................................. 43
Tbl. 14: Error and warning messages for turbopump electronic drive unit............................ 45
Tbl. 15: Error and warning messages for MVP 015 diaphragm pump electronic drive unit...46
Tbl. 16: Conversion table: Pressure units..............................................................................50
Tbl. 17: Conversion table: Units for gas throughput.............................................................. 50
Tbl. 18: Technical data HiCube 30 Eco, DN 40 ISO-KF........................................................51
Tbl. 19: Technical data HiCube 30 Eco, DN 63 CF-F............................................................51
Tbl. 20: Technical data HiCube 30 Eco, DN 63 ISO-K.......................................................... 52
Tbl. 21: Technical data HiCube 80 Eco, DN 40 ISO-KF........................................................52
Tbl. 22: Technical data HiCube 80 Eco, DN 63 CF-F............................................................53
Tbl. 23: Technical data HiCube 80 Eco, DN 63 ISO-K.......................................................... 53
Tbl. 24: Technical data HiCube 300 Eco, DN 100 CF-F........................................................54
Tbl. 25: Technical data HiCube 300 Eco, DN 100 ISO-K...................................................... 55
List of tables
5/60

List of figures

List of figures
Fig. 1: Position of the labels on the product.......................................................................... 9
Fig. 2: Overview: Turbo and backing pump combinations...................................................15
Fig. 3: Example combinations product overview................................................................. 16
Fig. 4: DCU control panel.................................................................................................... 17
Fig. 5: Transporting the turbo pumping station manually.................................................... 18
Fig. 6: Secure HiCube 300 Eco with lashing straps............................................................ 21
Fig. 7: Flange connection ISO-KF to ISO-KF...................................................................... 23
Fig. 8: Flange connection ISO-K to ISO-F, bracket screws.................................................24
Fig. 9: Flange connection CF-F, hexagon head screw and through hole............................26
Fig. 10: Flange connection CF-F, stud screw and tapped hole............................................. 26
Fig. 11: Flange connection CF-F, stud screw and through hole............................................27
Fig. 12: Example backing pump exhaust connection............................................................ 27
Fig. 13: Remove DCU from pumping station.........................................................................28
Fig. 14: DCU connection as remote control...........................................................................29
Fig. 15: Backing pump transport lock.................................................................................... 30
Fig. 16: Example: Connecting the grounding cable...............................................................30
Fig. 17: Power supply connector and master switch on HiCube Eco.................................... 31
Fig. 18: Example: DCU display for second parameter set.....................................................34
Fig. 19: Replace main fuses..................................................................................................38
Fig. 20: Electrical connections...............................................................................................39
Fig. 21: Dismantling the turbopump.......................................................................................39
Fig. 22: Dimensions of HiCube 30 Eco, DN 40 ISO-KF........................................................ 55
Fig. 23: Dimensions of HiCube 30 Eco, DN 63 CF-F............................................................ 56
Fig. 24: Dimensions of HiCube 30 Eco, DN 63 ISO-K...........................................................56
Fig. 25: Dimensions of HiCube 80 Eco, DN 40 ISO-KF........................................................ 56
Fig. 26: Dimensions of HiCube 80 Eco, DN 63 CF-F............................................................ 56
Fig. 27: Dimensions of HiCube 80 Eco, DN 63 ISO-K...........................................................57
Fig. 28: Dimensions of HiCube 300 Eco, DN 100 CF-F........................................................ 57
Fig. 29: Dimensions of HiCube 300 Eco, DN 100 ISO-K.......................................................57
6/60

1 About this manual

IMPORTANT
Read carefully before use.
Keep the manual for future consultation.

1.1 Validity

These operating instructions are for customers of Pfeiffer Vacuum. They describe the function of the designated product and provide the most important information for safe usage of the product. The de­scriptions comply with applicable directives. All information provided in these operating instructions refer to the current development status of the product. The documentation remains valid as long as the cus­tomer does not modify the product in any way.

1.1.1 Applicable documents

HiCube Eco Operating instructions
Declaration of conformity A component of these instructions
Operating instructions for the individual components see product description
About this manual

1.1.2 Variants

This manual applies to HiCube Eco ling turbopump stations:
HiCube 30 Eco
HiCube 80 Eco
HiCube 300 Eco

1.2 Target group

This operating instructions are aimed at all persons performing the following activities on the product:
transport,
setup (installation),
usage and operation,
decommissioning,
maintenance and cleaning,
storage or disposal.
The work described in this document is only permitted to be performed by persons with the appropriate technical qualifications (expert personnel) or who have received the relevant training from Pfeiffer Vac­uum.

1.3 Conventions

1.3.1 Instructions in the text

Usage instructions in the document follow a general structure that is complete in itself. The required ac­tion is indicated by an individual step or multi-part action steps.
Individual action step
A horizontal, solid triangle indicates the only step in an action.
This is an individual action step.
Sequence of multi-part action steps
The numerical list indicates an action with multiple necessary steps.
1.
Step 1
2.
Step 2
3.
...
7/60
C
US
TÜV Rheinland
Fuse
S/N
Mod.
HiCube 80 Eco PM S74 100 00
- - - - - - - -
P/N
Input Mass
D-35614 Asslar
Made in Germany 2018/06
T 2 A H 250 VAC 100-240 V +/-10 % 2.00 A 50/60 Hz
12.5 kg
VOR INBETRIEBNAHME TRANSPORTSICHERUNG LÖSEN BEFORE USE UNFASTEN THE TRANSPORT PROTECTION
warranty seal
About this manual

1.3.2 Pictographs

Pictographs used in the document indicate useful information.
Note
Tip

1.3.3 Stickers on the product

This section describes all the stickers on the product along with their meaning.
Rating plate
The rating plate is located on the back of the pumping station housing.
Transport protection
This sticker indicates that the backing pump transport lock must be released prior to commissioning.
Tbl. 1: Stickers on the product
Closure seal
The product is sealed ex-factory. Damaging or removing a clo­sure seal results in loss of the warranty.
Protection class
The sticker describes protection class 1 for the product. The positioning indicates the position for the ground connection.
Electrical voltage warning
The sticker warns of the risk of electric shock when working with the housing open.
Mandatory: disconnect mains plug
The sticker indicates that the mains plug is to be disconnected from the device prior to installation and maintenance work.
8/60
1 2,3 4
5
6
7
About this manual
Fig. 1: Position of the labels on the product
1 Transport lock sticker 5 Transport lock sticker
2 Closure seal (vacuum pump) 6 Rating plate
3 Ground connection sticker (vacuum pump) 7 Mandatory disconnect mains plug sign
4 Electrical voltage (power supply pack) warning sign

1.3.4 Abbreviations

Abbreviation Meaning in the document
CF flange: Metal-sealed connector conforming to ISO 3669
d Measurement of the diameter (in mm)
DCU Display Control Unit (display and control unit from Pfeiffer Vacuum).
DN Nominal diameter as size description
f Measurement of the rotation speed of a vacuum pump (frequency, in rpm or Hz)
HV High vacuum flange, high vacuum side
ISO flange: Connector conforming to ISO 1609 and ISO 2861
LC Display: Liquid crystal display (LCD)
LED Light-emitting diode
MVP Diaphragm vacuum pump
PE Earthed conductor (protective earth)
[P:xxx] Control parameters of the electronic drive unit. Printed in bold as a three-digit num-
T Temperature (in °C)
TC Turbopump electronic drive unit (turbo controller)
TPS Power supply (turbo power supply)
X3 15-pole D-Sub connecting socket on the electronic drive unit of the turbopump
ber in square brackets. Frequently displayed in combination with a short designation.
Example: [P:312] Software version
Tbl. 2: Abbreviations used in the document
9/60
Safety

2 Safety

2.1 General safety instructions

This document includes the following four risk levels and one information level.
DANGER
Imminent danger
Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
Instructions on avoiding the hazardous situation
WARNING
Possibly imminent danger
Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
Instructions on avoiding the hazardous situation
CAUTION
Possibly imminent danger
Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
Instructions on avoiding the hazardous situation
Danger of property damage
Notice is used to address practices not related to physical injury.
Instructions on avoiding property damage

2.2 Safety instructions

All safety instructions in this document are based on the results of the risk assessment carried out in accordance with Machinery Directive 2006/42/EC Annex I and EN ISO 12100 Section 5. Where applica­ble, all life cycle phases of the product were taken into account.
Risks during transport
WARNING
Danger of serious injury due to falling objects
Due to falling objects there is a risk of injuries to limbs through to broken bones.
Take particular care and pay special attention when transporting products manually.
Do not stack the products.
Wear protective equipment, e.g. safety shoes.
NOTICE
Notes, tips or examples indicate important information on the product or on this document.
10/60
Risks during installation
WARNING
Danger to life from electric shock in the event of a fault
In the event of a fault, devices connected to the mains may be live. There is a danger to life from electric shock when making contact with live components.
Always keep the mains connection freely accessible so you can disconnect it at any time.
Safety
WARNING
Danger to life from electric shock due to improperly performed installation
The device uses voltage that is dangerous on contact as the electrical power supply. Potentially fatal situations arise due to unsafe or incorrectly installation when reaching into the device.
Ensure that the system is safely integrated into an emergency off safety circuit.
Do not carry out any unauthorized modifications or changes to the device.
WARNING
Risk of cuts on moving, sharp-edged parts when reaching into the open high vacuum flange
With the high vacuum flange open, access to sharp-edged parts is possible. A manual rotation of the rotor increases the danger situation. There is the risk of cuts, up to the separation of body parts (e.g. fingertips). There is a risk of hair and loose clothing being drawn in. Objects falling in destroy the tur­bopump during subsequent operation.
Only remove the original protective covers immediately prior to connecting the high vacuum flange.
Do not reach into the high vacuum connection.
Wear protective gloves during installation.
Do not start the turbopump with open vacuum connections.
Always carry out the mechanical installation before electrical connection.
Prevent access to the high vacuum connection of the turbopump from the operator side (e.g. open vacuum chamber).
WARNING
Risk of injury caused by the turbopump breaking away with the vibration compensator in the event of a malfunction
Sudden jamming of the rotor generates high destructive torques in accordance with ISO 27892. When using a vibration compensator, this will probably lead to the turbopump being sheared off in use. The energy that this would release could throw the entire turbopump or shattered pieces from its interior through the surrounding space. Potentially dangerous gases can escape. There is a risk of very serious injuries, including death, and extensive property damage.
Take suitable safety precautions on-site for the compensation of the occurring torques.
Before installing a vibration compensator, you must first of all contact Pfeiffer Vacuum.
CAUTION
Risk of injury from bursting due to high pressure in the exhaust line
Faulty or insufficient exhaust lines cause hazardous situations, e.g. increase in exhaust pressure. There is a risk of bursting. It is not possible to rule out the risk of injuries due to broken pieces flying around, high escaping pressure and damage to the equipment.
Lay the exhaust line without shut-off units.
Observe the permissible pressures and pressure differentials of the product.
Check the exhaust line regularly for correct function.
Risks during operation
WARNING
Danger of death from poisoning due to toxic gases being expelled without an exhaust line
Exhaust gases and vapors are released from the turbo pumping station unhindered during normal us­age. In the case of processes with toxic media, there is a risk of injury and danger of death due to poisoning.
Note the corresponding regulations for handling toxic substances.
Toxic process gases should be safely conveyed away via an exhaust line.
11/60
Safety
CAUTION
Risk of injury from bursting due to high pressure in the exhaust line
Faulty or insufficient exhaust lines cause hazardous situations, e.g. increase in exhaust pressure. There is a risk of bursting. It is not possible to rule out the risk of injuries due to broken pieces flying around, high escaping pressure and damage to the equipment.
Lay the exhaust line without shut-off units.
Observe the permissible pressures and pressure differentials of the product.
Check the exhaust line regularly for correct function.
Risks during maintenance
WARNING
Danger to life from electric shock during maintenance and service work
The device is only completely de-energized when the mains plug has been disconnected and the tur­bopump is at a standstill. There is a danger to life from electric shock when making contact with live components.
Before performing all work, switch off the main switch.
Wait until the turbopump comes to a standstill (rotation speed =0).
Remove the mains plug from the device.
Secure the device against unintentional restarting.
WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work, there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic sub­stances causes environmental damage.
Take suitable safety precautions and prevent health hazards or environmental pollution by toxic process media.
Decontaminate affected parts before carrying out maintenance work.
Wear protective equipment.
Risks during disposal
WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work, there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic sub­stances causes environmental damage.
Take suitable safety precautions and prevent health hazards or environmental pollution by toxic process media.
Decontaminate affected parts before carrying out maintenance work.
Wear protective equipment.
Risks in the event of malfunctions
12/60
WARNING
Danger to life from the turbopump breaking away in the event of a fault
Sudden blocking of the rotor generates high destructive torques in accordance with ISO 27892. If the turbopump is not properly secured, it will tear off. The energy released in this way can propel the en­tire turbopump or shattered pieces from its interior through the surrounding area. Potentially danger­ous gases may escape. There is a risk of serious injury, potentially even fatal, and significant equip­ment damage.
Follow the installation instructions for this turbopump.
Observe the requirements regarding stability and design of the counter flange.
Use only original accessories or fixing material approved by Pfeiffer Vacuum for the installation.
WARNING
Risk of injury caused by the turbopump breaking away with the vibration compensator in the event of a malfunction
Sudden jamming of the rotor generates high destructive torques in accordance with ISO 27892. When using a vibration compensator, this will probably lead to the turbopump being sheared off in use. The energy that this would release could throw the entire turbopump or shattered pieces from its interior through the surrounding space. Potentially dangerous gases can escape. There is a risk of very serious injuries, including death, and extensive property damage.
Take suitable safety precautions on-site for the compensation of the occurring torques.
Before installing a vibration compensator, you must first of all contact Pfeiffer Vacuum.

2.3 Safety precautions

Safety
Duty to provide information on potential dangers
The product holder or user is obliged to make all operating personnel aware of dan­gers posed by this product.
Every person who is involved in the installation, operation or maintenance of the product must read, understand, and adhere to the safety-related parts of this docu­ment.
Infringement of conformity due to modifications to the product
The Declaration of Conformity from the manufacturer is no longer valid if the opera­tor changes the original product or installs additional equipment.
Following installation into a system, the operator is required to check and re­evaluate as necessary the conformity of the overall system in the context of the relevant european Directives before commissioning that system.
General safety precautions when handling the product
Observe all applicable safety and accident prevention regulations.
Check that all safety measures are observed at regular intervals.
Do not expose body parts to the vacuum.
Always ensure a secure connection to the earthed conductor (PE).
Never disconnect plug connections during operation.
Observe the above shutdown procedures.
Before working on the high vacuum connection, wait until the rotor has stopped completely (rota­tion speed f = 0).
Never put the device into operation with the high vacuum connection open.
Keep lines and cables away from hot surfaces (> 70°C).
Never fill or operate the unit with cleaning agents or cleaning agent residues.
Do not carry out your own conversions or modifications on the unit.
Observe the unit protection class prior to installation or operation in other environments.

2.4 Limits of use of the product

Parameter Limit value
Installation location weatherproof (internal space)
Air pressure 750 hPa to 1060 hPa
Installation altitude max. 2000 m
Rel. air humidity max. 80%, at T < 31 °C
up to max. 50% at T < 40 °C
Protection class I
Excess voltage category II
Permissible protection class IP20
13/60
Safety
Parameter Limit value
Degree of contamination 2
Ambient temperature 5°C to 35°C with air cooling
5°C to 40°C with water cooling
Permissible surrounding magnetic field depending on turbopump used
Maximum irradiated thermal output depending on turbopump used
Maximum permissible rotor temperature of the turbopump 90 °C
Tbl. 3: Limits of use for turbopump stations
Notes on ambient conditions
The specified permissible ambient temperatures apply to operation of the turbo­pump at maximum permissible backing pressure or at maximum gas throughput, depending on the cooling type. The turbopump is intrinsically safe thanks to redun­dant temperature monitoring.
The reduction in backing pressure or gas throughput permits operation of the turbopump at higher ambient temperatures as well.
If the maximum permissible operating temperature of the turbopump is exceed­ed, the electronic drive unit first reduces the drive output and switches it off where necessary.

2.5 Proper use

The turbo pumping station is only used to generate the vacuum.
The turbo pumping station is only used to extract dry and inert gases.
The turbo pumping station is only intended for use in enclosed interior spaces.
The turbo pumping station is intended for use as a table-top unit.

2.6 Foreseeable improper use

Improper use of the product invalidates all warranty and liability claims. Any use that is counter to the purpose of the product, whether intentional or unintentional, is regarded as misuse, in particular:
Establishing the voltage supply without correct installation
Installation with non-specified fastening material
Pumping explosive media
Pumping of corrosive media
Pumping of condensing vapors
Pumping of fluids
Pumping of dust
Operation with impermissible high gas throughput
Operation with impermissible high fore-vacuum pressure
Operation with excessively high irradiated heat output
Operation in impermissible high magnetic fields
Operation in an incorrect gas mode
Venting with impermissible high venting rates
Use for pressure generation
Use in areas with ionizing radiation
Operation in explosion-hazard areas
Use in systems in which sporadic loads and vibrations or periodic forces act on the device
The causing of hazardous operating conditions by a presetting on the electronic drive unit that is contrary to the process
Use of accessories or spare parts that are not listed in these instructions
14/60

3 Product description

HiPace 300 HiPace 80 HiPace 30
MVP 030-3MVP 015-4MVP 015-2

3.1 Identifying the product

To ensure clear identification of the product when communicating with Pfeiffer Vacuum, always keep all of the information on the rating plate to hand.
Learn about certifications through test seals on the product or at www.tuvdotcom.com with com­pany ID no. 000021320.
Product description
Fig. 2: Overview: Turbo and backing pump combinations
Feature HiCube Eco Operating in-
HV flange size DN 40/DN 63 DN 100
HiPace 30 PT 0510 BN
Turbopump
Electronic drive unit
Diaphragm Pump
Current supply TPS onboard,
Control unit DCU 002 DCU 002 DCU 002 PT 0250 BN
Air cooling yes yes yes PT 0500 BN
Venting valve optional optional optional PT 0228 BN
Housing heater 1)not available Optional, with CF
TC 110 TC 110 TC 110 PT 0204 BN
MVP 0152
MVP 0154
MVP 0303
24 V DC
HiPace 80 PT 0208 BN
MVP 0152
MVP 0154
MVP 0303
TPS onboard, 24 V DC
flange only
structions
HiPace 300 PT 0202 BN
HiPace 300 H PT 0509 BN
(MVP 0152)
MVP 0154
MVP 0303
TPS onboard, 24 V DC
Optional, with CF flange only
PU 0070 BN
PU 0070 BN
PU 0065 BN
PT 0233 BN
1) Water cooling required
15/60
1 2
3
4
5
6
7
89
10
11
12
13
14
15
Product description
Tbl. 4: Pumping station component combination options

3.2 Scope of delivery

HiCube Eco turbo pumping station
Protective cover for the high vacuum flange
Extension cable M12 on M12, 3 m
Power supply cable, country-specific
Operating instructions for the turbo pumping station and individual components

3.3 Function

Turbo pumping stations are fully automatic pump units ready for connection. A turbo pumping station consists of a portable or mobile vacuum pump unit, with a turbopump and a specially matched backing pump.

3.3.1 Control interface

Fig. 3: Example combinations product overview
1 Housing 9 DCU 002 display and control unit
2 HiPace 80 turbopump, DN 63 ISO-K 10 Attachment hooks (HiCube 300 Eco only)
3 Electronic drive unit TC 110 11 Transport protection
4 Connection cable 12 MVP 030-3 diaphragm pump backing pump
5 Ground terminal 13 Recessed handle
6 MVP 015-2 diaphragm pump backing pump 14 HiPace 300 turbopump, DN 100 CF-F
7 Master switch 15 Air cooling
8 Mains supply plug
The integrated DCU 002 display and control unit serves to control and monitor the turbo pumping sta­tion. The DCU is optionally used as a remote control when removing the housing and using an exten­sion cable.
16/60
DCU
1
2
3
4

3.3.2 Drive

Fig. 4: DCU control panel
1 LC display, illuminated 3 LED operating display
2 Status icons 4 Control keys
Electronic drive unit of the turbopump
Electronic drive unit of the backing pump
Product description

3.3.3 Cooling

Air cooling
Water cooling (optional)
At excessively high temperatures, the electronic drive unit automatically reduces the drive power.
17/60
1
3
2
Transportation and Storage

4 Transportation and Storage

4.1 Transporting the pumping station

WARNING
Danger of serious injury due to falling objects
Due to falling objects there is a risk of injuries to limbs through to broken bones.
Take particular care and pay special attention when transporting products manually.
Do not stack the products.
Wear protective equipment, e.g. safety shoes.
Packing
We recommend keeping the transport packaging and original protective cover.
Instructions for safe transport
Observe the weight specified on packaging.
Use a transport aid if necessary (trolley, lift truck).
Transport the product in its original packaging.
Always move the product upright, and over the flattest possible surface.
Always place the product on an adequately sized, level surface.
Fig. 5: Transporting the turbo pumping station manually
1 Protective cover 3 Footprint
2 Recessed handle
Instructions for safe transport without packaging
For transport without packaging, HiCube Eco turbo pumping stations are equipped with recessed han­dles on the side of the housing frame.
1.
Observe weight specified on the rating plate.
2.
Lift the turbo pumping station with your hands on the two recessed handles.
3.
Always transport the turbo pumping station upright, with its standing surface downwards.
4.
Always place turbo pumping station securely on an adequately sized, level surface.

4.2 Transport protection

Backing pumps in HiCube line turbo pumping stations are secured against damage during transport.
Handling the transport lock
1.
Release the bucking pump transport lock, only directly prior to commissioning at the installation location.
2.
18/60
Observe the installation instructions (see chapter “Removing transport lock”, page 29).

4.3 Storing the pumping station

Packing
We recommend storing the product in its original packaging.
1.
Close all flange openings with the original protective caps.
2.
Close all other connections (e.g. exhaust) with the corresponding protective caps.
3.
Only store the turbo pumping station within the specified limits.
4.
In rooms with humid or aggressive atmospheres: Hermetically seal the turbo pumping station, to­gether with a drying agent, in a plastic bag.
Transportation and Storage
19/60
Installation

5 Installation

The installation of the turbopump and its fastening is of outstanding importance. The rotor of the turbo­pump revolves at very high speed. In practice it is not possible to exclude the risk of the rotor touching the stator (e.g. due to the penetration of foreign bodies into the high vacuum connection). The kinetic energy released acts on the housing and on the anchoring of the turbopump within fractions of a sec­ond.
Comprehensive tests and calculations conforming to ISO 27892 confirm the safety of the turbopump both against crashes (destruction of the rotor blade) and against bursting (breakage of the rotor shaft). The experimental and theoretical results are expressed in safety measures and recommendations for the correct and safe fastening of the turbopump.

5.1 Preparing for set-up

WARNING
Danger to life from electric shock in the event of a fault
In the event of a fault, devices connected to the mains may be live. There is a danger to life from electric shock when making contact with live components.
Always keep the mains connection freely accessible so you can disconnect it at any time.
General notes for the installation of vacuum components
Choose an installation location that permits access to the product and to supply lines at all times.
Observe the ambient conditions given for the area of use.
Provide the highest possible level of cleanliness during assembly.
Ensure that flange components during installation are grease-free, dust-free and dry.

5.2 Setting up the pumping station

WARNING
Risk of cuts on moving, sharp-edged parts when reaching into the open high vacuum flange
With the high vacuum flange open, access to sharp-edged parts is possible. A manual rotation of the rotor increases the danger situation. There is the risk of cuts, up to the separation of body parts (e.g. fingertips). There is a risk of hair and loose clothing being drawn in. Objects falling in destroy the tur­bopump during subsequent operation.
Only remove the original protective covers immediately prior to connecting the high vacuum flange.
Do not reach into the high vacuum connection.
Wear protective gloves during installation.
Do not start the turbopump with open vacuum connections.
Always carry out the mechanical installation before electrical connection.
Prevent access to the high vacuum connection of the turbopump from the operator side (e.g. open vacuum chamber).
Procedure
1.
Maintain a minimum distance of 50 cm to lateral boundaries or adjacent devices.
2.
Maintain a minimum distance of 10 cm to exposed table edges.
3.
Make sure that there are sufficient cooling options for the pumping station.
4.
Install suitable shielding if the surrounding magnetic fields exceed the permissible levels.
5.
If you use housing heating and water cooling, make sure that the temperature at the vacuum chamber connection flange does not exceed 120 °C.
6.
Install suitable shielding so that the irradiated thermal output does not exceed the permissible val­ues when high temperatures occur due to the process.

5.3 Anchoring pumping station

(applies to HiCube 300 Eco only)
20/60
1 2
Installation
In the case of a sudden jam of the turbopump rotor, the resulting torque must be taken up via the pump­ing station frame by the fixture provided by customer. To secure the HiCube 300 Eco pumping station and connected systems, it is essential to anchor the pumping station.
Fig. 6: Secure HiCube 300 Eco with lashing straps
1 Lashing strap 2 Mounting hooks
Required aids
2 x lashing straps of adequate length and loading capacity; approximately 3 cm wide
On-site attachment points to retain the lashing straps
Secure HiCube 300 Eco against twisting
1.
Always attach the HiCube 300 Eco pumping station to the two mounting hooks with lashing straps.
2.
Affix the lashing straps perpendicularly to the mounting hooks opening, i.e., against the turbo­pump rotation direction.
3.
Make sure that the belts are fitted and tensioned correctly. Make sure that the belt loop is in the hollow of the mounting hook.
4.
Attach the lashing straps at the customer site so that they are able to withstand loads of up to 2000 N per attachment point.

5.4 Connecting the high vacuum side

5.4.1 Designing the counter flange

If the rotor is suddenly blocked, the torques arising from the system and the high vacuum flange must be absorbed. The installation elements for turbo pumps are special designs by Pfeiffer Vacuum.
NOTICE
Risk of damage due to incorrect counter flange design
Unevenness on the operator-side counter flange results in stresses in the vacuum pump housing, even when properly attached. This can produce leakage or negative changes in running characteris­tics.
Adhere to the shape tolerances for the counter flange.
Observe the maximum flatness deviations over the entire surface.
Superstructural parts
Installing vertical superstructural parts to the high vacuum connection is the respon­sibility of the operating company. The loading capacity of the high vacuum flange is specific for the turbopump used. The total weight of superstructural parts must not exceed the maximum values specified.
21/60
Installation
Turbo pumping sta­tion
HiCube 30 Eco 500 Nm 50 kg ± 0.05 mm
Applied tor-
2)
que
Super­structural
3)
parts
Flatness Tensile
strength
170 N/mm
4)
2
depth
2.5 x d 3.0 mT 1.2 W
Screw in
5)
Magnetic
6)
field
Heat in­put
HiCube 80 Eco 620 Nm 3.3 mT 0.9 W
HiCube 300 Eco 2000 Nm 5.5 mT 2.4 W
270 N/mm
2
1.5 x d
Tbl. 5: Requirements for on-site high vacuum connection
Only use the approved mounting kits of Pfeiffer Vacuum for the high vacuum connection of the turbopump.

5.4.2 Using accessories for the high vacuum connection

Installation and operation of accessories
Pfeiffer Vacuum offers a series of special, compatible accessories for its products.
Information and ordering options for approved accessories can be found on­line.
The following accessories are not included in the scope of supply.
Using a splinter shield or protective screen
Pfeiffer Vacuum centering rings with splinter shield or protective screen in the high vacuum flange pro­tect the turbopump against foreign bodies from the vacuum chamber. The volume flow rate of the pump reduces according to the flow conductance values and the size of the high vacuum flange.
7)
Flange size
H
2
He N
2
Ar
Splinter shield DN 40 6 9 17 18
Splinter shield DN 63 3 6 15 16
Protective screen DN 63 1 1 4 4
Splinter shield DN 100 5 7 24 24
Protective screen DN 100 2 2 10 8
Tbl. 6: Changing pumping speed with splinter shield or protective screen
Reduced pumping speed in % for the gas type
With ISO flanges, use centering rings with a protective screen or splinter shield.
With CF flanges, always insert the protective screen or splinter shield into the high vacuum flange with the clamping lugs pointing towards the rotor.
Using vibration dampers
Pfeiffer Vacuum vibration dampers are suitable for use on plants sensitive to vibration.
Maximum torque applied in the event of a burst: The theoretically calculated torque in the
2) event of a burst (rotor shaft breakage) according to ISO 27892 was not reached in any experi­mental test.
3) Permissible total weight of superstructural parts on the pumping station
4) Tensile strength of counter flange material in all operating conditions, according to screw in depth
5) Screw in depth of fixing screws in counter flange
6) Maximum permissible magnetic field surrounding pumping station
7) Maximum permissible heat absorption in pumping station
22/60
WARNING
Risk of injury caused by the turbopump breaking away with the vibration compensator in the event of a malfunction
Sudden jamming of the rotor generates high destructive torques in accordance with ISO 27892. When using a vibration compensator, this will probably lead to the turbopump being sheared off in use. The energy that this would release could throw the entire turbopump or shattered pieces from its interior through the surrounding space. Potentially dangerous gases can escape. There is a risk of very serious injuries, including death, and extensive property damage.
1.
Take suitable safety precautions on-site for the compensation of the occurring torques.
2.
Before installing a vibration compensator, you must first of all contact Pfeiffer Vacuum.
1.
Only install a vibration damper with a vertical passage.
2.
Take account of the flow resistance.
3.
Secure the turbo pump to the high vacuum flange in addition.
4.
Observe the fastening of the ISO-K flanges (see chapter “Attaching ISO-K flange onto ISO-
K”, page 24) and (see chapter “Installation of ISO-K flange onto ISO-F”, page 24).

5.4.3 Connecting the turbopump externally

Depending on the configuration, the turbopump can be operated independently from the pumping sta­tion.
Installation
Dismantling the turbopump
Please see the descriptions for removing a turbopump from the pumping station
(see chapter “Removing components for maintenance”, page 38).
Procedure
1.
Follow the installation instructions for the turbopump in the associated operating instructions for the individual components.
2.
Use cable kit PM 071 477 -T (3 m) from the Pfeiffer Vacuum accessories range.
Other lengths available on request.
3.
Extend the fore-vacuum line and the control cable.

5.4.4 Installation of ISO-KF flange

ISO flange connections
For the connection of flanges in ISO-KF or ISO-K design, twisting may occur in the event of sudden blockage of the rotor, despite correct installation.
Leak-tightness of the flange connection, however, is not jeopardized in this re­gard.
Required tools
Allen key
Calibrated torque wrench (Tightening factor ≤ 1.6)
Fig. 7: Flange connection ISO-KF to ISO-KF
Using of clamping ring
1.
Only use the approved mounting kits from Pfeiffer Vacuum for the connection.
2.
Fasten the high vacuum connection on the turbopump and the clamping ring to the counter flange.
3.
Utilize all prescribed components for the turbopump.
4.
Tighten the clamping ring screws evenly.
23/60
Installation
Tightening torque: 3,7 Nm.

5.4.5 Attaching ISO-K flange onto ISO-K

ISO flange connections
For the connection of flanges in ISO-KF or ISO-K design, twisting may occur in the event of sudden blockage of the rotor, despite correct installation.
Leak-tightness of the flange connection, however, is not jeopardized in this re­gard.
Required tools
Wrench, WAF 15
Calibrated torque wrench (Tightening factor ≤ 1.6)
Fig. 8: Flange connection ISO-K to ISO-F, bracket screws
Connection with bracket screw
1.
For the connection of the turbopump, use only the approved mounting kits from Pfeiffer Vacuum.
2.
Connect the flange with the components of the mounting kit according to the figure.
3.
Use for all prescribed components for the turbopump.
4.
Tighten the bracket screws cross-wise in 3 steps.
Tightening torque: 5, 15, 25 ± 2 Nm

5.4.6 Installation of ISO-K flange onto ISO-F

The connection types for the installation of ISO-K with ISO-F flange are:
"Claw and tapped hole"
"Hexagon head set screw and tapped hole"
"Stud screw and tapped hole"
"Stud screw and through-bore"
ISO flange connections
For the connection of flanges in ISO-KF or ISO-K design, twisting may occur in the event of sudden blockage of the rotor, despite correct installation.
Leak-tightness of the flange connection, however, is not jeopardized in this re­gard.
Connection of claw and tapped hole
24/60
1.
Only use the approved mounting kits from Pfeiffer Vacuum for the connection.
2.
Connect the flange according to the diagram using the components in the mounting kit.
3.
Utilize all prescribed components for the turbopump.
4.
Tighten the claws in 3 stages in opposite pairs.
Tightening torque: 5, 10, 16 ± 1 Nm
Connection of hexagon head set screw and tapped hole
1.5 × d
2.5 × d
1.
Only use the approved mounting kits from Pfeiffer Vacuum for the connection.
2.
Place the collar flange over the high vacuum flange on the turbopump.
3.
Insert the snap ring into the side groove on the high vacuum flange of the turbopump.
4.
Fasten the turbopump onto the counter flange in accordance with the diagram using the collar flange, snap ring and centering ring.
5.
Utilize all prescribed components for the turbopump.
6.
Screw the hexagon head set screws into the tapped holes.
Observe the minimum tensile strength of the flange material and the screw-in depth.
7.
Tighten the hexagon head set screws in 3 stages in opposite pairs.
Tightening torque: 5, 10, 16 ± 1 Nm
Connection of stud screw and tapped hole
Installation
1.
Only use the approved mounting kits from Pfeiffer Vacuum for the connection.
2.
Screw the required number of stud screws with the shorter screw-in end into the holes in the counter flange.
Observe the minimum tensile strength of the flange material and the screw-in depth.
3.
Place the collar flange over the high vacuum flange on the turbopump.
4.
Insert the snap ring into the side groove on the high vacuum flange of the turbopump.
5.
Fasten the turbopump onto the counter flange in accordance with the diagram using the collar flange, snap ring and centering ring.
6.
Utilize all prescribed components for the turbopump.
7.
Tighten the nuts in 3 stages in opposite pairs.
Tightening torque: 5, 10, 16 ± 1 Nm
Connection of stud screw and through-bore
1.
Only use the approved mounting kits from Pfeiffer Vacuum for the connection.
2.
Place the collar flange over the high vacuum flange on the turbopump.
3.
Insert the snap ring into the side groove on the high vacuum flange of the turbopump.
4.
Fasten the turbopump onto the counter flange in accordance with the diagram using the collar flange, snap ring and centering ring.
5.
Utilize all prescribed components for the turbopump.
6.
Tighten the nuts in 3 stages in opposite pairs.
7.
Tightening torque: 5, 10, 16 ± 1 Nm
25/60
Installation

5.4.7 Fastening CF flange to CF-F

The connection types for CF installation with CF flange are:
"Hexagon head screw and through hole"
"Stud screw with tapped hole"
"Stud screw with through hole"
Leaks may occur due to the incorrect installation of CF flanges
Inadequate cleanliness when handling CF flanges and copper gaskets results in leaks and may cause process damage.
Always wear suitable gloves before touching or fitting any components.
Fit seals only if dry and free of grease.
Take care of damaged surfaces and cut edges.
Replace the damaged components.
Required tools
Hexagonal wrench (13 WAF)
Calibrated torque wrench (tightening factor ≤ 1.6)
NOTICE
Fig. 9: Flange connection CF-F, hexagon head screw and through hole
Connection of the hexagon head screw and through holes
1.
For the connection of the turbopump, use only the approved mounting kits from Pfeiffer Vacuum.
2.
If used: Insert the protective screen or splinter shield with clamping lugs downwards in the turbo­pump high vacuum flange.
3.
Place the seal exactly in the hollow.
4.
Connect the flange with the components of the mounting kit according to the figure.
5.
Tighten the screw couplings all the way around.
Tightening torque: 22 ± 2 Nm
6.
Then check the torque, since flowing of the sealing material may make it necessary to re-tighten the screws.
Fig. 10: Flange connection CF-F, stud screw and tapped hole
Connection of the stud screw and tapped hole
1.
For the connection of the turbopump, use only the approved mounting kits from Pfeiffer Vacuum.
2.
Screw in the required number of stud screws with the shorter end in the holes on the counter flange.
3.
If used: Insert the protective screen or splinter shield with clamping lugs downwards in the turbo­pump high vacuum flange.
4.
Place the seal exactly in the hollow.
5.
Connect the flange with the components of the mounting kit according to the figure.
6.
Tighten the screw couplings all the way around.
26/60
1
2 (G1/8”)
Tightening torque: 22 ± 2 Nm
7.
Then check the torque, since flowing of the sealing material may make it necessary to re-tighten the screws.
Fig. 11: Flange connection CF-F, stud screw and through hole
Connection of the stud screw and through hole
1.
For the connection of the turbopump, use only the approved mounting kits from Pfeiffer Vacuum.
2.
If used: Insert the protective screen or splinter shield with clamping lugs downwards in the turbo­pump high vacuum flange.
3.
Place the seal exactly in the hollow.
4.
Connect the flange with the components of the mounting kit according to the figure.
5.
Tighten the screw couplings all the way around.
Tightening torque: 22 ± 2 Nm
6.
Then check the torque, since flowing of the sealing material may make it necessary to re-tighten the screws.
Installation

5.5 Connecting the exhaust side

WARNING
Danger of death from poisoning due to toxic gases being expelled without an exhaust line
Exhaust gases and vapors are released from the turbo pumping station unhindered during normal us­age. In the case of processes with toxic media, there is a risk of injury and danger of death due to poisoning.
Note the corresponding regulations for handling toxic substances.
Toxic process gases should be safely conveyed away via an exhaust line.
CAUTION
Risk of injury from bursting due to high pressure in the exhaust line
Faulty or insufficient exhaust lines cause hazardous situations, e.g. increase in exhaust pressure. There is a risk of bursting. It is not possible to rule out the risk of injuries due to broken pieces flying around, high escaping pressure and damage to the equipment.
Lay the exhaust line without shut-off units.
Observe the permissible pressures and pressure differentials of the product.
Check the exhaust line regularly for correct function.
Fig. 12: Example backing pump exhaust connection
1 Diaphragm pump 2 Exhaust silencer (included in the delivery)
27/60
1
2
3
Installation
Procedure
1.
Choose a minimum exhaust line cross section equal to the size of the exhaust connection.
2.
Unscrew the exhaust silencer from the backing pump.
3.
Route the piping downwards from the pump, to prevent condensate return.
Condensate separator
We recommend installing a condensate separator, with condensate drain at the lowest point of the exhaust line.

5.6 Modifying the DCU for remote operation

If required, the display and control unit can be used as a remote control, after modification.
Before carrying out any work, safely disconnect device from mains
1.
Shut down the device.
2.
Wait until all components are at a complete standstill.
3.
Switch off the master switch.
4.
Disconnect the mains cable from the device.
Fig. 13: Remove DCU from pumping station
1 Collar screw with attachment nipple 3 Control cable with M12 connecting plug
2 DCU 002 display and control unit
Remove DCU
1.
Release the 4 collar screws from the corners of the DCU screen.
2.
Safely store the attachment nipple.
3.
Remove DCU from pumping station housing.
Note control cable length.
4.
Unscrew M12 control cable connecting plug from DCU.
Cable included in shipment
Other cable lengths are available on request from the Pfeiffer Vacuum accessories.
28/60
2
1
PM 061 270 -X
RS-485
Fig. 14: DCU connection as remote control
1 Extension cable 2 Control cable with M12 connecting plug
Connect DCU as remote control
1.
Screw interface cable supplied in shipment into DCU RS-485 connection.
2.
Use the PM 061 270 -X coupling.
3.
Connect the interface cable to the coupling, and to the control cable from the pumping station.
Installation

5.7 Connecting accessories for the pumping station

Installation and operation of accessories
Pfeiffer Vacuum offers a series of special, compatible accessories for its products.
Information and ordering options for approved accessories can be found on­line.
The following accessories are not included in the scope of supply.
Accessory connection for TC 110 electronic drive unit
The use of Pfeiffer Vacuum accessories via the TC 110 electronic drive unit is only possible using the corresponding connection cable and/or adapter on the X3 multi-functional connection.
Configuration of the required accessory output via RS-485 using Pfeiffer Vac­uum display and control units or PC.
Detailed instructions can also be found in the "TC | 110 electronic drive unit" operating instructions.
Factory settings
All turbopump accessory connections are preconfigured at the factory. Any confu­sion will cause malfunctions or operational failure.
1.
Please note the initial product configuration.
2.
Note the cable connections and parameter settings, before making changes.
Use additional accessories
1.
Note the present configuration of existing connections and control cables.
2.
Use the Pfeiffer Vacuum display and control unit DCU 002.

5.8 Removing transport lock

All HiCube Eco line turbo pumping stations are equipped with a transport lock for the backing pumps. The backing pump is affixed diagonally to the pumping station baseplate, with cheesehead screws and nuts, on 2 spring-loaded rails.
29/60
1
Installation
Fig. 15: Backing pump transport lock
1 Transport lock hexagon nut
Required tool
Hexagonal wrench (13 WAF)
Procedure
1.
Rotate the transport lock hexagon nut upwards to the stop.
2.
Also release the transport lock on the opposite side.

5.9 Grounding the pumping station

Fig. 16: Example: Connecting the grounding cable
Required aids
Screw M4 × 8
Shake proof washer M4, if required
Suitable grounding cable with M4 cable lug
Procedure
1.
Use a suitable grounding cable to divert applicative interferences.
2.
Route the connection in accordance with locally applicable provisions.
3.
Use the turbopump ground terminal (M4 female thread).

5.10 Establishing the mains connection

WARNING
Danger to life from electric shock due to improperly performed installation
The device uses voltage that is dangerous on contact as the electrical power supply. Potentially fatal situations arise due to unsafe or incorrectly installation when reaching into the device.
Ensure that the system is safely integrated into an emergency off safety circuit.
Do not carry out any unauthorized modifications or changes to the device.
30/60
Installation
WARNING
Danger to life from electric shock in the event of a fault
In the event of a fault, devices connected to the mains may be live. There is a danger to life from electric shock when making contact with live components.
Always keep the mains connection freely accessible so you can disconnect it at any time.
Fig. 17: Power supply connector and master switch on HiCube Eco
1 Mains plug 2 Country-specific power supply cable (in shipment)
Establish mains power supply for HiCube Eco
1.
Make sure that the master switch is off before connecting.
2.
Always ensure a secure connection to the earthed conductor (PE).
3.
Plug the mains cable supplied in the shipment into the connecting plug on the side of the pumping station housing.
4.
Connect the mains cable to the mains power supply on the customer-side.
31/60
Operation

6 Operation

6.1 Commissioning

Important settings and function-related variables are factory-programmed into the vacuum pumps elec­tronic drive units as parameters. Each parameter has a three-digit number and a description. Parame­ters can be used via Pfeiffer Vacuum display and control panels or externally via RS-485, using the Pfeiffer Vacuum protocol.
For both parameter sets, the DCU acts as the user interface. HiCube Eco controls two independent pa­rameter sets, for the turbopump and backing pump. The backing pump parameters are shown hierarchi-
cally in the display, and marked with a symbol.
Parameter Name Designation Option
[P:001] Heating Heating no heating
with heating 1 1
[P:012] EnableVent Enable vents 1 1 1
[P:019] Cfg DO2 Output DO2 config-
uration
[P:023] MotorPump Motor pump 1 1 1
[P:024] Cfg DO1 Output DO1 config-
uration
[P:025] OpMode Backing pump op-
erating mode
[P:035] Cfg Acc A1 Connection A1 ac-
cessories configu­ration
[P:036] Cfg Acc B1 Connection B1 ac-
cessories configu­ration
[P:708] PwrSVal Default power con-
sumption
[P:710] Swoff BKP Backing pump off
threshold in interval mode
[P:711] SwOn BKP Backing pump on
threshold in interval mode
[P:794] Param set Parameter set 1 = advanced 1 = advanced 1 = advanced
[P:795] Servicelin Service row view 309 = Actual
[P: 030]
[P: 717]
ValveMode Gas ballast configu-
ration
StdbySVal Default rotation
speed in standby
Air cooling 4 = Fan (tem-
Water cool­ing
for MVP 015-2
for MVP 015-4
for MVP 030 100 % 100 % 100 %
for MVP 015 0 = auto 0 = auto 0 = auto
for MVP 015 66 % 66 % 66 %
30 Eco 80 Eco 300 Eco
not available
22 = Backing pump standby
15 = Pump state 15 = Pump
1 = Interval mode
perature control­led)
2 = heating 2 = heating 2 = heating
1 = Venting valve, normally closed
90 % 90 % 90 %
100 % 100 % 100 %
25 W 25 W 45 W | MVP 015
40 W 40 W 55 W | MVP 015
speed
HiCube pumping station
0 0
22 = Backing pump standby
state
1 = Interval mode
4 = Fan (tem­perature con­trolled)
1 = Venting valve, normally closed
309 = Actual speed
22 = Backing pump standby
15 = Pump state
1 = Interval mode
4 = Fan (temper­ature controlled)
1 = Venting valve, normally closed
52 W | MVP 030
62 W | MVP 030
309 = Actual speed
32/60
Tbl. 7: Factory setting of key parameters on delivery
NOTICE
Turbopump destruction due to gases with too high molecular masses
The pumping of gases with impermissible high molecular masses leads to the destruction of the tur­bopump.
Make sure that the gas mode is set correctly by [P:027] in the electronic drive unit.
Consult Pfeiffer Vacuum before you use gases with higher molecular masses (> 80).
Commissioning the turbo pumping station
1.
Make sure that the transport lock has been released.
2.
Where applicable, ensure cooling water supply and flow rate.
3.
Where applicable, ensure sealing gas supply and flow rate.
4.
Provide the current supply for the product.

6.2 Switching on the turbo pumping station

Switching on the current supply
Switch off the power supply at the main switch.
After being switched on, the DCU runs a self-test and a check of the connected instruments. The dura­tion of the self-test is approx. 20 seconds and is visualized on the display with a progress bar.
Putting the pumping station into operation
Switch on the turbo pumping station.
Operation

6.3 Normal operation

Operating the DCU
When working with the DCU 002 Pfeiffer Vacuum display and control unit, please see the associ­ated operating instructions supplied in the shipment.
Key Parameter | use Description
Tbl. 8: Description of the key functions of the DCU
corresponds to
[010] = 0 or 1
[308] --> [309] Scroll forwards through the parameter set
[309] --> [308] Scroll back through the parameter set
Press simultaneously Editing mode: Allows the setting of parameter options
Press simultaneously
again
Pumping station ON/OFF: All components are put into/out of operation according to their configuration
Error acknowledgement (Reset): Resets active malfunc­tion messages in case the cause is eliminated.
The arrow - - - > designates the option selection in the dis­play
Selection mode: Accepts the selection made ("change confirmed" is displayed)

6.4 Operation with gas ballast (applies to MVP 015-2 only)

Steam or moisture from pumped media condensing in the backing pump will impact throughput. Inject­ing gas ballast improves (reduces) condensate quantity. The pumping station reaches the specified final vacuum more quickly.
33/60
Operation
The MVP 015-4 and MVP 030-3 backing pumps have a sufficiently high water vapor tolerance, and are not equipped with gas ballast valve.
The MVP 015-2 backing pump is equipped with an electronic gas ballast valve.
The parameters for automatic gas ballast valve operation of the are factory-set for normal applica­tion.
Gas ballast
The HiCube Eco turbo pumping station, with MVP 015-2 diaphragm pump as back­ing pump, is equipped with an automatic gas ballast function. If however run-up time delays still occur, as a result of condensate in the fore-vacuum area, we rec­ommend opening the gas ballast valve manually.
Manual gas ballast valve operation in this case involves control via the DCU function keys.
Fig. 18: Example: DCU display for second parameter set
Open gas ballast valve manually
1.
Open advanced parameter set, if necessary (see chapter “Normal operation”, page 33).
2. Shut down the turbo pumping station.
3. Scroll through the entire first parameter set, and the second parameter set directly following
it, to parameter [030]: ValveMode.
4.
Press both keys simultaneously: this opens Edit mode.
5. Select option "2" = open.
6.
Press both keys simultaneously: this confirms your selection, and exits Edit mode.
Inject gas ballast
1. Switch on the turbo pumping station.
2.
Operate the turbo pumping station with gas ballast valve open.
Operating time: approximately 15 to 30 minutes.
Set gas ballast function to automatic
1. Shut down the turbo pumping station.
2. Scroll through the entire first parameter set, and the second parameter set directly following
34/60
it, to parameter [030]: ValveMode.
3.
Press both keys simultaneously: this opens Edit mode.
4. Select option "0" = auto.
5.
Press both keys simultaneously: this confirms your selection, and exits Edit mode.
6. Switch on the turbo pumping station.

6.5 Operation monitoring

6.5.1 Operating mode display via LED

The LEDs on the front panel of the DCU display basic operating statuses.
LED Symbol LED status Display Meaning
Off without current
Green
Red
On, flashing "Pumping station OFF", rotation speed ≤ 60 rpm
On, inverse flashing "Pumping station ON", set rotation speed not reached
On, constant "Pumping station ON", set rotation speed reached
On, flashing "Pumping station OFF", rotation speed > 60 rpm
Off no error, no warning
On, flashing Warning
On, constant Defect
Operation
Tbl. 9: LED display and meaning at the DCU

6.5.2 Temperature monitoring

If threshold values are exceeded, output signals from temperature sensors allow the vacuum pumps to be brought to a safe condition. Depending on pump type, temperature threshold values for warnings and malfunction messages are saved unchanged in the electronic drive unit. For information purposes, various status queries are prepared in the parameter set.
The drive power of the turbopump is reduced in case of impermissible motor temperature or non­permissibly high housing temperature. This can cause the motor to fall below the set rotation speed switchpoint and so result in switching off of the turbopump.
At a non-permissibly high temperature (> 75 °C) of the backing pump, the motor rotation speed is reduced to nominal rotation speed to prevent overheating. After cooling down (< 72° C) the back­ing pump runs at set rotation speed again.

6.6 Switching off and venting

We recommend
Vent the turbopump after shutdown. By doing so, you prevent particles flowing back into the vacuum system from the fore-vacuum area.

6.6.1 Shutting down the turbo pumping station

Procedure
Shut down the turbo pumping station.
The components shut down according to their configuration. The pumping station remains ready for operation.
Shut off the entire current supply at the master switch.
35/60
Operation

6.6.2 Venting

NOTICE
Damage to the turbopump due to non-permissibly fast pressure rise during venting
Non-permissibly high pressure rise rates place a significant load on the rotor and the magnetic bear­ing of the turbopump. During venting very small volumes in the vacuum chamber or the turbopump, there is a risk of uncontrollable pressure rises. This causes mechanical damage to the turbopump, including potential failure.
Observe the prescribed maximum pressure rise speed of 15 hPa/s.
Avoid manual and uncontrolled venting of very low volumes.
Where necessary, use a venting valve from the Pfeiffer Vacuum range of accessories.
Manual venting
Manual venting describes the standard process for venting the turbo pumping station.
1.
Ensure that the vacuum system is shut down.
2.
Open the black venting screw on the turbopump by one revolution maximum.
3.
Wait for pressure equalization to atmospheric pressure in the vacuum system.
4.
Close the venting screw again.
Use a Pfeiffer Vacuum venting valve
The Pfeiffer Vacuum venting valve is an optional accessory for installation on the turbopump.
The venting valve is normally closed. Control is via the turbopump electronic drive unit, and configura­tion of parameters [P:012] and [P:030]. In the event of a power failure, the turbopump continues to de­liver sufficient energy during its run-down period to initiate a proper venting process. When power is re­stored, the venting process is interrupted.
Switch off the turbopump.
The venting process starts automatically.
Venting speed [P:720] Venting duration [P:721] Venting duration in the event of a power
failure
50 % of rated speed 3600 s 3600 s
Tbl. 10: Factory settings for delayed venting of turbopumps
General information for fast venting
We recommend fast venting of larger volumes in 4 steps.
1.
Use a Pfeiffer Vacuum venting valve for the turbopump, or match the valve cross-section to the size of the recipient and maximum venting rate.
2.
Vent the vacuum system with a maximum rate of pressure rise of 15 hPa/s for a duration of 20 seconds.
3.
Then vent the system with a second venting valve of any size, for example directly at the vacuum chamber.
4.
Wait for pressure equalization to atmospheric pressure in the vacuum system.
36/60

7 Maintenance

7.1 General maintenance information

WARNING
Danger to life from electric shock during maintenance and service work
The device is only completely de-energized when the mains plug has been disconnected and the tur­bopump is at a standstill. There is a danger to life from electric shock when making contact with live components.
Before performing all work, switch off the main switch.
Wait until the turbopump comes to a standstill (rotation speed =0).
Remove the mains plug from the device.
Secure the device against unintentional restarting.
WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work, there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic sub­stances causes environmental damage.
Take suitable safety precautions and prevent health hazards or environmental pollution by toxic process media.
Decontaminate affected parts before carrying out maintenance work.
Wear protective equipment.
Maintenance

7.2 Maintenance intervals and responsibilities

Instructions for carrying out maintenance
1.
Carry out the required maintenance on the pumping station components, in accordance with the instructions in the individual operating instructions.
2.
Clean the turbopump station exterior with a lint-free cloth and a little isopropanol.
3.
Consult Pfeiffer Vacuum Service about shorter maintenance intervals for extreme loads or impure processes.
4.
For all other cleaning, maintenance or repair work, contact the appropriate Pfeiffer Vacuum Serv­ice location.

7.3 Replacing mains fuses

WARNING
Danger to life from electric shock during maintenance and service work
The device is only completely de-energized when the mains plug has been disconnected and the tur­bopump is at a standstill. There is a danger to life from electric shock when making contact with live components.
Before performing all work, switch off the main switch.
Wait until the turbopump comes to a standstill (rotation speed =0).
Remove the mains plug from the device.
Secure the device against unintentional restarting.
Device or system problem
Generally, a faulty fuse is an indication of a problem in the device or system. If the new fuse is again faulty after replacement, contact your nearest Pfeiffer Vacuum Service Center.
37/60
1
2
3
4
Maintenance
Fig. 19: Replace main fuses
1 Master switch 3 Fuses
2 Fuse holder 4 Mains plug
Replacing faulty fuses
Spare parts required
2 x fuses, 5 x 20 (T 2 A H 250 V AC)
1.
Shut down the pumping station.
2.
Disconnect mains cable from the mains plug.
3.
Open the fuse holder above the mains plug.
4.
Replace faulty fuses.
5.
Close fuse holder.
6.
Recommission the pumping station.

7.4 Removing components for maintenance

Customer maintenance on pumping station components may require removal from the pumping station frame.
NOTICE
Malfunction due to changing the connection configuration
The pumping station connections are pre-configured at the factory. If the control lines on the connec­tor are mixed up, this causes the pumping station to malfunction or fail.
When removing components, note their original layout for subsequent re-assembly.
Make a note of the accessories configuration and important setting values from the DCU before disassembly of the pumping station or components.
Connection to X3 Preset accessory
Acc. A1 Air cooling
Acc. B1 Venting valve
DO1 Backing pump (MVP 30-3 only)
Tbl. 11: Preset accessory connections
38/60

7.4.1 Removing the electrical connections

4
5
1
2
3
4
3
2
5
1
Before carrying out any work, safely disconnect device from mains
1.
Shut down the device.
2.
Wait until all components are at a complete standstill.
3.
Switch off the master switch.
4.
Disconnect the mains cable from the device.
Maintenance
Fig. 20: Electrical connections
1 Multifunction connection "X3" 4 Mains cable
2 Connection cable connecting plug 5 Master switch
3 Control cable accessory
Procedure
1.
Disconnect the connection plug from the multi-function connection on the turbopump electronic drive unit.
2.
Disconnect and remove the accessory control lines from the connecting plug.

7.4.2 Dismantling the turbopump from the pumping station

Fig. 21: Dismantling the turbopump
1 Turbopump, complete 4 Fore-vacuum line
2 Fixing screw with washer 5 Mounting plate
3 Fore-vacuum connection
39/60
Maintenance
Required tools
Allen key
Calibrated torque wrench (tightening factor ≤ 1.6)
Procedure
1.
Loosen and remove the fore-vacuum line from the turbopump.
2.
Make sure not to kink or damage the fore-vacuum hose.
3.
Unscrew all 4 fixing screws and washers from the mounting plate.
Fixing screw tightening torque for subsequent reassembly: 10 Nm.
4.
Remove the entire turbopump and mounting plate from frame.
Service turbopump
The opening in the mounting plate makes it easy for customers to perform mainte­nance work on the turbo pump (e.g. change the lubricant reservoir).
40/60

8 Decommissioning

8.1 Shutting down for longer periods

Procedure
1.
Remove the pumping station from the vacuum system if necessary.
2.
Change the operating fluid reservoir of the turbopump if necessary.
3.
Close all flange openings with the original protective caps.
4.
Store the pumping station indoors only, at temperatures between -10 °C and +40 °C.
5.
In rooms with humid or aggressive atmospheres: Hermetically seal the pumping station together with a drying agent in a plastic bag.

8.2 Recommissioning

Risk of damage to the turbopump as a result of operating fluid aging after recommissioning
The shelf life of the operating fluid of the turbopump is limited. Aging of the operating fluid may lead to the failure of the ball bearing and cause damage to the turbopump.
Pay attention to when the operating fluid must be changed:
after maximum 2 years without operation,
after maximum 4 years combined operation and downtimes.
Observe the maintenance instructions and inform Pfeiffer Vacuum Service.
Decommissioning
NOTICE
Procedure for recommissioning the pumping station
1.
Check the pumping station for impurities and moisture.
2.
Clean the turbopump station exterior with a lint-free cloth and a little isopropanol.
3.
If necessary, arrange for Pfeiffer Vacuum Service to fully clean the pumping station.
4.
If necessary, arrange for Pfeiffer Vacuum Service to fully clean the pumping station components.
5.
Install the pumping station according to these instructions (see chapter “Installation”, page 20).
6.
Recommission the pumping station according to these instructions (see chapter “Commission-
ing”, page 32).

8.3 Disposing of the vacuum pump

WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work, there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic sub­stances causes environmental damage.
Take suitable safety precautions and prevent health hazards or environmental pollution by toxic process media.
Decontaminate affected parts before carrying out maintenance work.
Wear protective equipment.
Procedure
Dispose of all substances safely according to local regulations.
41/60
Malfunctions

9 Malfunctions

9.1 General

WARNING
Danger to life from the turbopump breaking away in the event of a fault
Sudden blocking of the rotor generates high destructive torques in accordance with ISO 27892. If the turbopump is not properly secured, it will tear off. The energy released in this way can propel the en­tire turbopump or shattered pieces from its interior through the surrounding area. Potentially danger­ous gases may escape. There is a risk of serious injury, potentially even fatal, and significant equip­ment damage.
Follow the installation instructions for this turbopump.
Observe the requirements regarding stability and design of the counter flange.
Use only original accessories or fixing material approved by Pfeiffer Vacuum for the installation.
WARNING
Risk of injury caused by the turbopump breaking away with the vibration compensator in the event of a malfunction
Sudden jamming of the rotor generates high destructive torques in accordance with ISO 27892. When using a vibration compensator, this will probably lead to the turbopump being sheared off in use. The energy that this would release could throw the entire turbopump or shattered pieces from its interior through the surrounding space. Potentially dangerous gases can escape. There is a risk of very serious injuries, including death, and extensive property damage.
Take suitable safety precautions on-site for the compensation of the occurring torques.
Before installing a vibration compensator, you must first of all contact Pfeiffer Vacuum.
Malfunctions on the pumping station are usually caused by faults on individual components. LEDs on the turbopump and backing pump indicate the operation statuses for the respective device.
The DCU display and control unit indicates basic pumping station operating statuses (see chap-
ter “Operating mode display via LED”, page 35).
The DCU indicates system malfunctions by outputting an error code to the display.

9.2 Troubleshooting

Problem Possible causes Remedy
Pumping station will not start up. None of the LEDs on the DCU illumi­nate.
Current supply interrupted
Incorrect oper­ating voltage
No operating voltage present
Electronic drive unit defective
1.
Check supply lines to pumping station.
2.
Check the 24 V DC output voltage at the “DC out” connection of the built-in power supply pack.
3.
Check the plug contacts on the power supply pack.
4.
Check connection cable between power sup­ply pack, backing pump, and turbopump.
1.
Note rating plate specifications.
2.
Supply the correct operating voltage.
1.
Connect operating voltage
1.
Replace electronic drive unit according to tur­bopump operating instructions.
2.
Contact Pfeiffer Vacuum Service.
42/60
Problem Possible causes Remedy
Pumping station not reaching final pressure.
Tbl. 12: Troubleshooting

9.3 Error codes

Error (** Error E––– **) always result in the shutdown of the connected peripheral devices.
Warnings (* Warning F––– *) are only displayed and do not result in the shutdown of components.
Handling of error messages on the HiCube Eco
1.
2.
Condensate in the backing pump
Gas ballast valve open
First eliminate the cause of the error.
Reset the error message by pressing the key.
1.
1.
Malfunctions
MVP 015-2 only: open the backing pump gas ballast valve (see chapter “Operation with gas
ballast (applies to MVP 015-2 on­ly)”, page 33).
MVP 015-2 only: close the backing pump gas ballast valve (see chapter “Operation with gas
ballast (applies to MVP 015-2 on­ly)”, page 33).
Display in DCU Problem Possible causes Remedy
* Warning F110 * Pressure gauge
** Error E040 ** Hardware error
** Error E042 ** Hardware error
** Error E043 ** Hardware error
** Error E090 ** Internal device
error
** Error E698 ** Communication
error
Tbl. 13: Warning and error messages when using the DCU
Display Problem Possible causes Remedy
E001 Excessive speed
E002 Overvoltage
E006 Run-up fault
E007 Operating fluid low
Pressure gauge faulty
Connection to the pressure gauge disconnected during operation
external RAM faulty
EPROM checksum incorrect
2
EPROM write error
RAM not large enough
DCU is connected to incorrect electronic drive unit
Electronic drive unit is not re­sponding
Incorrect power supply pack used
Run-up time threshold set too low
Gas flow in recipient through leaks or open valves
Still below speed-control switch point run-up time expires
Operating fluid low
Check the cable connection
Carry out a restart with pressure gauge connected
Replace the pressure gauge com­pletely
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Connect the DCU to the correct electronic drive unit
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service. Acknowledge only at rotation speed f = 0
Check the power supply pack type Check the mains voltage
Adjust the run-up time to the process condi­tions Check vacuum chamber for leakage and closed valves Adjust the rotation speed switch point
Check the operating fluid Acknowledge only at rotation speed f = 0
43/60
Malfunctions
Display Problem Possible causes Remedy
E008 Connection, electronic drive
unit - pump faulty
E010 Internal device error
E021 Electronic drive unit does
not detect the pump
E043 Internal configuration error
E044 Excess temperature, elec-
tronics
E045 Excess temperature, motor
E046 Internal initialization error
E091 Internal device error
E092 Unknown connection panel
E093 Motor temperature evalua-
tion faulty
E094 Electronics temperature
evaluation faulty
E098 Internal communication er-
ror
E107 Final stage group error
E108 Rotation speed measure-
ment faulty
E109 Software not released
E110 Operating fluid evaluation
faulty
E111 Operating fluid pump com-
munication error
E112 Operating fluid pump group
error
E114 Final stage temperature
evaluation faulty
E117 Excess temperature, pump
lower part
E118 Excess temperature, final
stage
E119 Excess temperature, bear-
ing
E143 Operating fluid pump ex-
cess temperature
E777 Nominal rotation speed not
confirmed
F001 TMS heat-up time expired
Connection to the pump faulty
Faulty cooling
Faulty cooling
Faulty cooling
Faulty cooling
Faulty cooling
Faulty cooling
Nominal rotation speed not confirmed after re­placing the electronic drive unit
Internal timer for heat­up monitoring exceeded
Check connections Acknowledge only at rotation speed f = 0
Contact Pfeiffer Vacuum Service. Acknowledge only at rotation speed f = 0
Contact Pfeiffer Vacuum Service. Acknowledge only at rotation speed f = 0
Contact Pfeiffer Vacuum Service.
Improve the cooling Check the operating conditions
Improve the cooling Check the operating conditions
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service. Acknowledge only at rotation speed f = 0
Contact Pfeiffer Vacuum Service. Acknowledge only at rotation speed f = 0
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service. Acknowledge only at rotation speed f = 0
Contact Pfeiffer Vacuum Service. Acknowledge only at rotation speed f = 0
Contact Pfeiffer Vacuum Service. Acknowledge only at rotation speed f = 0
Contact Pfeiffer Vacuum Service.
Improve the cooling Check the operating conditions
Improve the cooling Check the operating conditions
Improve the cooling Check the operating conditions
Improve the cooling Check the operating conditions Acknowledge only at rotation speed f = 0
Confirm the nominal rotation speed with [P:
777]
Acknowledge only at rotation speed f = 0
Check the operating conditions
44/60
Display Problem Possible causes Remedy
F003 TMS heating circuit temper-
ature sensor
F007 Undervoltage/power failure
F018 Operating supremacy con-
flict
F021 Blocking signal invalid
F034 Sealing gas flow low
F045 Motor high temperature
F076 Electronics high tempera-
ture
F097 Invalid pump information
F098 Incomplete pump informa-
tion
F100 Rotation speed raised to
minimum value
F115 Pump lower part tempera-
ture evaluation faulty
F116 Bearing temperature evalu-
ation faulty
F117 Pump lower part high tem-
perature
F118 Final stage high tempera-
ture
F119 Bearing high temperature
F143 High operating fluid pump
temperature
F168 High delay
TMS temperature not in the permissible range between +5 °C and 85 °C
Mains failure
Pumping station switch­ed on with [P:010] while E74 input "start/stop" is off (opened)
Signal of the blocking signal monitoring out­side the valid range
Signal of the sealing gas monitoring valid but below the set threshold
[P:791]
Faulty cooling
Faulty cooling
Data of the pump faulty
Connection to the pump faulty
Permissible specifica­tions for rotation speed operation or stand-by in­correct
Faulty cooling
Faulty cooling
Faulty cooling
Faulty cooling
Rate of pressure rise too high, venting rate too high
Check the operating conditions
Contact Pfeiffer Vacuum Service.
Check the mains supply
Switch on the pumping station via E74
Switch the pumping station off via [P:010].
Check the connections of the sealing gas monitoring
Check the parameter options of the accesso­ry outputs
Check and improve the sealing gas supply
Check the operating conditions
Improve the cooling
Check the operating conditions
Improve the cooling
Check the operating conditions
Acknowledging restores factory setting
Contact Pfeiffer Vacuum Service.
Check [P:707] or [P:717]
Obtain the valid rotation speed range from the technical data of the Turbopump
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Improve the cooling
Check the operating conditions
Contact Pfeiffer Vacuum Service.
Check the operating conditions
Improve the cooling
Check the operating conditions
Improve the cooling
Check the operating conditions
Check and adapt pump-specific venting rate
Malfunctions
Tbl. 14: Error and warning messages for turbopump electronic drive unit
Display Problem Possible causes Remedy
E042 Inconsistent software Checksum errors
E091 Unknown hardware
E098 Internal communication error between interface
board and drive
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
45/60
Malfunctions
Display Problem Possible causes Remedy
E117 Excess temperature pump
E173 Pump overflow
E174 Pump blocked
Tbl. 15: Error and warning messages for MVP 015 diaphragm pump electronic drive unit
Faulty cooling
Improve the cooling Check the operating conditions
Contact Pfeiffer Vacuum Service.
Contact Pfeiffer Vacuum Service.
46/60
ANFORDERUNG SERVICE
ERKLÄRUNG KONTAMINIERUNG
Service solutions from Pfeiffer Vacuum

10 Service solutions from Pfeiffer Vacuum

We offer first class service
Long vacuum component service life, coupled with low downtimes, are clear expectations that you have of us. We satisfy your needs with capable products and outstanding service.
We are consistently striving to perfect our core competence, service for vacuum components. And our service is far from over once you’ve purchased a product from Pfeiffer Vacuum. It often enough really just begins then. In proven Pfeiffer Vacuum quality, of course.
Our professional sales engineers and service technicians stand ready to provide hands-on support to you worldwide. Pfeiffer Vacuum offers a complete portfolio of service offerings, ranging from genuine
spare parts right through to service agreements.
Take advantage of Pfeiffer Vacuum Service
Whether for preventative on-site service from our field service, fast replacement with as-new replace­ment products or repair in a Service Center close to you; you have various options for upholding your equipment availability. Detailed information and addresses can be found on our website in the Pfeiff-
er Vacuum Service section.
Advice on the optimum solution is available from your Pfeiffer Vacuum contact partner.
For quick and smooth handling of the service process, we recommend the following steps:
1. Download the current form templates.
Declaration of Service RequestService RequestDeclaration of Contamination
a. Dismantle all accessories and keep them (all external mounted parts as valve, inlet
screen, etc.).
b. Drain the operating fluid/lubricant as necessary.
c. Drain the cooling medium as necessary.
2. Fill out the service request and the declaration of contamination.
3. Send the forms via email, fax or post to your local Service Center.
4. You will receive a response from Pfeiffer Vacuum.
Sending of contaminated products
No units will be accepted if they are contaminated with micro-biological, explosive or radioactive sub­stances. If products are contaminated or if the declaration of contamination is missing, Pfeiffer Vacuum will contact the customer before starting maintenance. In addition, depending on the product and the level of contamination additional decontamination costs may be required.
47/60
ERKLÄRUNG KONTAMINIERUNG
Service solutions from Pfeiffer Vacuum
5. Prepare the product for transport in accordance with the details in the dec­laration of contamination.
a) Neutralize the product with nitrogen or dry air.
b) Close all openings with airtight blank flanges.
c) Seal the product in appropriate protective film.
d) Only pack the product in suitable, stable transport containers.
e) Observe the applicable transport conditions.
6 Affix the declaration of contamination to the outside of the packaging.
7 Then send your product to your local Service Center.
8 You will receive a confirmation message/a quotation from Pfeiffer Vacuum.
For all service orders, our General Terms and Conditions of Sales and Supply and General Terms and
Conditions of Repair and Maintenance apply to vacuum equipment and components.
48/60

11 Accessories

Accessories
We recommend
Please refer to the accessories list for the individual components in their respective operating instructions or online at pfeiffer-vacuum.de.
49/60
Technical data and dimensions

12 Technical data and dimensions

12.1 General

This section describes the basis for the technical data of Pfeiffer Vacuum turbopumps.
Technical data
Maximum values refer exclusively to the input as a single load.
Specifications according to PNEUROP committee PN5
ISO 27892 2010:"Vacuum technology - Turbomolecular pumps - Measurement of rapid shutdown torque"
ISO 21360 2012: “Vacuum technology - Standard methods for measuring vacuum-pump perform­ance - Part 1: General description”
ISO 21360 2018: "Vacuum technology - Standard methods for measuring vacuum-pump perform­ance - Part 4: Turbomolecular vacuum pumps"
Ultimate pressure with test dome after 48 h bake out duration
Gas throughput with water cooling; backing pump = rotary vane pump (10 m3/h)
Cooling water consumption at maximum gas throughput, cooling water temperature 25 °C
Integral leakage rate with 100 % helium concentration, 10 s measurement duration
Sound pressure level at distance to vacuum pump = 1 m
mbar bar Pa hPa kPa Torr | mm Hg
mbar 1
bar 1000 1
Pa 0.01
hPa 1
1 · 10
1 · 10
1 · 10
-3
-5
-3
100 1 0.1 0.75
5
1 · 10
1 0.01
1000 100 750
-3
1 · 10
7.5 · 10
100 1 0.1 0.75
kPa 10 0.01 1000 10 1 7.5
Torr | mm Hg 1.33
1.33 · 10
-3
1 Pa = 1 N/m
133.32 1.33 0.133 1
2
Tbl. 16: Conversion table: Pressure units
mbar l/s
Pa m3/s
sccm Torr l/s
atm cm3/s
mbar l/s 1 0.1 59.2 0.75 0.987
Pa m3/s
sccm
10 1 592 7.5 9.87
1.69 · 10
-2
1.69 · 10
-3
1
1.27 · 10
-2
1.67 · 10
Torr l/s 1.33 0.133 78.9 1 1.32
atm cm3/s
1.01 0.101 59.8 0.76 1
Tbl. 17: Conversion table: Units for gas throughput
-3
-2

12.2 Technical data

Classification HiCube 30 Eco HiCube 30 Eco HiCube 30 Eco
Flange (in) DN 40 ISO-KF DN 40 ISO-KF DN 40 ISO-KF
Flange (out) Silencer, G 1/8" Silencer, G 1/8" Silencer, G 1/8"
Turbopump HiPace 30 HiPace 30 HiPace 30
Backing pump MVP 015-2 MVP 015-4 MVP 030-3
50/60
Technical data and dimensions
Ultimate pressure
Pumping speed for N
2
< 1 · 10-7 hPa < 1 · 10-7 hPa < 1 · 10-7 hPa
22 l/s 22 l/s 22 l/s
Pumping speed backing pump at 50 Hz1 m³/h 0.75 m³/h 1.8 m³/h
Pump-down time for vacuum chamber
56 s 84 s 15 s
size 1 l
Pump-down time for vacuum chamber
560 s 836 s 145 s
size 10 l
Pump-down time for vacuum chamber
5595 s 8358 s 1451 s
size 100 l
Mains requirement: voltage 100 – 240 V 100 – 240 V 100 – 240 V
Mains requirement: frequency (range) 50/60 Hz 50/60 Hz 50/60 Hz
Cooling method, standard Air Air Air
Relative humidity of air 5 – 85 %, not condens-
ing
5 – 85 %, not condens­ing
5 – 85 %, not condens­ing
Weight 11.7 kg 13.4 kg 13.3 kg
Tbl. 18: Technical data HiCube 30 Eco, DN 40 ISO-KF
Classification HiCube 30 Eco HiCube 30 Eco HiCube 30 Eco
Flange (in) DN 63 CF-F DN 63 CF-F DN 63 CF-F
Flange (out) Silencer, G 1/8" Silencer, G 1/8" Silencer, G 1/8"
Turbopump HiPace 30 HiPace 30 HiPace 30
Backing pump MVP 015-2 MVP 015-4 MVP 030-3
Ultimate pressure
Pumping speed for N
2
< 1 · 10-8 hPa < 1 · 10-8 hPa < 1 · 10-8 hPa
32 l/s 32 l/s 32 l/s
Pumping speed backing pump at 50 Hz 1 m³/h 0.75 m³/h 1.8 m³/h
Pump-down time for vacuum chamber
56 s 84 s 15 s
size 1 l
Pump-down time for vacuum chamber
560 s 836 s 145 s
size 10 l
Pump-down time for vacuum chamber
5595 s 8358 s 1451 s
size 100 l
Mains requirement: voltage 100 – 240 V 100 – 240 V 100 – 240 V
Mains requirement: frequency (range) 50/60 Hz 50/60 Hz 50/60 Hz
Cooling method, standard Air Air Air
Relative humidity of air 5 – 85 %, not condens-
ing
5 – 85 %, not condens­ing
5 – 85 %, not condens­ing
Weight 12.9 kg 14.6 kg 14.5 kg
Tbl. 19: Technical data HiCube 30 Eco, DN 63 CF-F
Classification HiCube 30 Eco HiCube 30 Eco HiCube 30 Eco
Flange (in) DN 63 ISO-K DN 63 ISO-K DN 63 ISO-K
Flange (out) Silencer, G 1/8" Silencer, G 1/8" Silencer, G 1/8"
Turbopump HiPace 30 HiPace 30 HiPace 30
Backing pump MVP 015-2 MVP 015-4 MVP 030-3
51/60
Technical data and dimensions
Ultimate pressure
Pumping speed for N
2
< 1 · 10-7 hPa < 1 · 10-7 hPa < 1 · 10-7 hPa
32 l/s 32 l/s 32 l/s
Pumping speed backing pump at 50 Hz1 m³/h 0.75 m³/h 1.8 m³/h
Pump-down time for vacuum chamber
56 s 84 s 15 s
size 1 l
Pump-down time for vacuum chamber
560 s 836 s 145 s
size 10 l
Pump-down time for vacuum chamber
5595 s 8358 s 1451 s
size 100 l
Mains requirement: voltage 100 – 240 V 100 – 240 V 100 – 240 V
Mains requirement: frequency (range) 50/60 Hz 50/60 Hz 50/60 Hz
Cooling method, standard Air Air Air
Relative humidity of air 5 – 85 %, not condens-
ing
5 – 85 %, not condens­ing
5 – 85 %, not condens­ing
Weight 11.7 kg 13.4 kg 13.3 kg
Tbl. 20: Technical data HiCube 30 Eco, DN 63 ISO-K
Classification HiCube 80 Eco HiCube 80 Eco HiCube 80 Eco
Flange (in) DN 40 ISO-KF DN 40 ISO-KF DN 40 ISO-KF
Flange (out) Silencer, G 1/8" Silencer, G 1/8" Silencer, G 1/8"
Turbopump HiPace 80 HiPace 80 HiPace 80
Backing pump MVP 015-2 MVP 015-4 MVP 030-3
Ultimate pressure
Pumping speed for N
2
< 1 · 10-7 hPa < 1 · 10-7 hPa < 1 · 10-7 hPa
35 l/s 35 l/s 35 l/s
Pumping speed backing pump at 50 Hz1 m³/h 0.75 m³/h 1.8 m³/h
Pump-down time for vacuum chamber
56 s 83 s 14 s
size 1 l
Pump-down time for vacuum chamber
557 s 834 s 143 s
size 10 l
Pump-down time for vacuum chamber
5572 s 8335 s 1428 s
size 100 l
Mains requirement: voltage 100 – 240 V 100 – 240 V 100 – 240 V
Mains requirement: frequency (range) 50/60 Hz 50/60 Hz 50/60 Hz
Cooling method, standard Air Air Air
Relative humidity of air 5 – 85 %, not condens-
ing
5 – 85 %, not condens­ing
5 – 85 %, not condens­ing
Weight 12.1 kg 13.8 kg 13.7 kg
Tbl. 21: Technical data HiCube 80 Eco, DN 40 ISO-KF
Classification HiCube 80 Eco HiCube 80 Eco HiCube 80 Eco
Flange (in) DN 63 CF-F DN 63 CF-F DN 63 CF-F
Flange (out) Silencer, G 1/8" Silencer, G 1/8" Silencer, G 1/8"
Turbopump HiPace 80 HiPace 80 HiPace 80
Backing pump MVP 015-2 MVP 015-4 MVP 030-3
52/60
Technical data and dimensions
Ultimate pressure
Pumping speed for N
2
< 1 · 10-8 hPa < 1 · 10-8 hPa < 1 · 10-8 hPa
67 l/s 67 l/s 67 l/s
Pumping speed backing pump at 50 Hz 1 m³/h 0.75 m³/h 1.8 m³/h
Pump-down time for vacuum chamber
56 s 83 s 14 s
size 1 l
Pump-down time for vacuum chamber
557 s 834 s 143 s
size 10 l
Pump-down time for vacuum chamber
5572 s 8335 s 1428 s
size 100 l
Mains requirement: voltage 100 – 240 V 100 – 240 V 100 – 240 V
Mains requirement: frequency (range) 50/60 Hz 50/60 Hz 50/60 Hz
Cooling method, standard Air Air Air
Relative humidity of air 5 – 85 %, not condens-
ing
5 – 85 %, not condens­ing
5 – 85 %, not condens­ing
Weight 13.5 kg 15.2 kg 15.1 kg
Tbl. 22: Technical data HiCube 80 Eco, DN 63 CF-F
Classification HiCube 80 Eco HiCube 80 Eco HiCube 80 Eco
Flange (in) DN 63 ISO-K DN 63 ISO-K DN 63 ISO-K
Flange (out) Silencer, G 1/8" Silencer, G 1/8" Silencer, G 1/8"
Turbopump HiPace 80 HiPace 80 HiPace 80
Backing pump MVP 015-2 MVP 015-4 MVP 030-3
Ultimate pressure
Pumping speed for N
2
< 1 · 10-7 hPa < 1 · 10-7 hPa < 1 · 10-7 hPa
67 l/s 67 l/s 67 l/s
Pumping speed backing pump at 50 Hz1 m³/h 0.75 m³/h 1.8 m³/h
Pump-down time for vacuum chamber
56 s 83 s 14 s
size 1 l
Pump-down time for vacuum chamber
557 s 834 s 143 s
size 10 l
Pump-down time for vacuum chamber
5572 s 8335 s 1428 s
size 100 l
Mains requirement: voltage 100 – 240 V 100 – 240 V 100 – 240 V
Mains requirement: frequency (range) 50/60 Hz 50/60 Hz 50/60 Hz
Cooling method, standard Air Air Air
Relative humidity of air 5 – 85 %, not condens-
ing
5 – 85 %, not condens­ing
5 – 85 %, not condens­ing
Weight 12.1 kg 13.8 kg 13.7 kg
Tbl. 23: Technical data HiCube 80 Eco, DN 63 ISO-K
Classification HiCube 300
Eco
HiCube 300 Eco
HiCube 300 Eco
HiCube 300 Eco
HiCube 300 Eco
HiCube 300 Eco
Flange (in) DN 100 CF-F DN 100 CF-F DN 100 CF-F DN 100 CF-F DN 100 CF-F DN 100 CF-F
Flange (out) Silencer,
G 1/8"
Silencer, G 1/8"
Silencer, G 1/8"
Silencer, G 1/8"
Silencer, G 1/8"
Silencer, G 1/8"
Turbopump HiPace 300 HiPace 300 HiPace 300 HiPace 300 H HiPace 300 H HiPace 300 H
53/60
Technical data and dimensions
Backing pump MVP 015-2 MVP 015-4 MVP 030-3 MVP 015-2 MVP 015-4 MVP 030-3
Ultimate pressure
< 1 · 10-8 hPa < 1 · 10-8 hPa < 1 · 10-8 hPa < 1 · 10
-10
< 1 · 10
hPa
Pumping speed for N
2
Pumping speed
260 l/s 260 l/s 260 l/s 260 l/s 260 l/s 260 l/s
1 m³/h 0.75 m³/h 1.8 m³/h 1 m³/h 0.75 m³/h 1.8 m³/h backing pump at 50 Hz
Pump-down time for
55 s 83 s 14 s 55 s 83 s 14 s vacuum chamber size 1 l
Pump-down time for
553 s 830 s 138 s 553 s 830 s 138 s vacuum chamber size 10 l
Pump-down time for
5532 s 8295 s 1382 s 5532 s 8295 s 1382 s vacuum chamber size 100 l
Mains requirement:
100 – 240 V 100 – 240 V 100 – 240 V 100 – 240 V 100 – 240 V 100 – 240 V voltage
Mains requirement:
50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz frequency (range)
Cooling method,
Air Air Air Air Air Air standard
Relative humidity of air
5 – 85 %, not
condensing
5 – 85 %, not condensing
5 – 85 %, not condensing
5 – 85 %, not condensing
5 – 85 %, not condensing
Weight 17.9 kg 19.6 kg 19.5 kg 17.9 kg 19.6 kg 19.5 kg
-10
hPa < 1 · 10
hPa
5 – 85 %, not condensing
-10
Tbl. 24: Technical data HiCube 300 Eco, DN 100 CF-F
Classification HiCube 300
Eco
HiCube 300 Eco
HiCube 300 Eco
HiCube 300 Eco
HiCube 300 Eco
HiCube 300 Eco
Flange (in) DN 100 ISO-K DN 100 ISO-K DN 100 ISO-K DN 100 ISO-K DN 100 ISO-K DN 100 ISO-K
Flange (out) Silencer,
G 1/8"
Silencer, G 1/8"
Silencer, G 1/8"
Silencer, G 1/8"
Silencer, G 1/8"
Silencer, G 1/8"
Turbopump HiPace 300 HiPace 300 HiPace 300 HiPace 300 H HiPace 300 H HiPace 300 H
Backing
MVP 015-2 MVP 015-4 MVP 030-3 MVP 015-2 MVP 015-4 MVP 030-3
pump
Ultimate
< 1 · 10-7 hPa < 1 · 10-7 hPa < 1 · 10-7 hPa < 1 · 10
-10
hPa < 1 · 10
-10
hPa < 1 · 10
-10
hPa
pressure
Pumping speed for N
Pumping
260 l/s 260 l/s 260 l/s 260 l/s 260 l/s 260 l/s
2
1 m³/h 0.75 m³/h 1.8 m³/h 1 m³/h 0.75 m³/h 1.8 m³/h speed back­ing pump at 50 Hz
Pump-down
55 s 83 s 14 s 55 s 83 s 14 s time for vac­uum cham­ber size 1 l
54/60
Technical data and dimensions
Pump-down time for vac­uum cham­ber size 10 l
Pump-down time for vac­uum cham­ber size 100 l
Mains re­quirement: voltage
Mains re­quirement: frequency (range)
Cooling method, standard
Relative hu­midity of air
Weight 15.9 kg 17.6 kg 17.5 kg 15.9 kg 17.6 kg 17.5 kg
553 s 830 s 138 s 553 s 830 s 138 s
5532 s 8295 s 1382 s 5532 s 8295 s 1382 s
100 – 240 V 100 – 240 V 100 – 240 V 100 – 240 V 100 – 240 V 100 – 240 V
50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz
Air Air Air Air Air Air
5 – 85 %, not
condensing
Tbl. 25: Technical data HiCube 300 Eco, DN 100 ISO-K
5 – 85 %, not condensing
5 – 85 %, not condensing
5 – 85 %, not condensing
5 – 85 %, not condensing
5 – 85 %, not condensing

12.3 Dimensioned drawings

All dimensions given in mm
Fig. 22: Dimensions of HiCube 30 Eco, DN 40 ISO-KF
55/60
Technical data and dimensions
Fig. 23: Dimensions of HiCube 30 Eco, DN 63 CF-F
Fig. 24: Dimensions of HiCube 30 Eco, DN 63 ISO-K
Fig. 25: Dimensions of HiCube 80 Eco, DN 40 ISO-KF
Fig. 26: Dimensions of HiCube 80 Eco, DN 63 CF-F
56/60
Fig. 27: Dimensions of HiCube 80 Eco, DN 63 ISO-K
Technical data and dimensions
Fig. 28: Dimensions of HiCube 300 Eco, DN 100 CF-F
Fig. 29: Dimensions of HiCube 300 Eco, DN 100 ISO-K
57/60

Declaration of Conformity

We hereby declare that the product cited below satisfies all relevant provisions of the follow­ing EU Directives:
Machinery 006/42/EC (Appendix II, No. 1 A)
Electro-magnetic Compatibility 2014/30/EU
Restrictions on the Use of Specific Hazardous Substances 2011/65/EU
Authority for compiling the technical documents rests with Mr. Helmut Bernhardt, Pfeiff­er Vacuum GmbH, Berliner Straße 43, 35614 Aßlar.
Turbo pumping station
HiCube Eco
Harmonized standards and national standards and specifications which have been applied:
DIN EN ISO 12100:2011
DIN EN 1012-2:2011
DIN EN 61000-3-2:2014
DIN EN 61000-3-3:2014
DIN EN 61010-1:2010
DIN EN 61326-1:2013
DIN EN 62061:2013
Signature:
(Dr. Ulrich von Hülsen)
Managing director
Pfeiffer Vacuum GmbH
Berliner Straße 43
35614 Asslar
Germany
Asslar, 2016-09-10
59/60
ed. D - Date 1901 - P/N:PT0553BEN
*PT0553*
Loading...