Perma Pure FC-Series User Manual

PRINCIPLE OF OPERATION
®
tube and transferred to
FC™- Series
Humidifier
User Manual
Nafion® is a registered trademark of DuPont. FC™ is a tradmark of Perma Pure LLC.
Bulletin# 207
PERMA PURE
8 Executive Drive Toms River, N.J. 08755 Phone: 800-337-3762
732-244-0010 Fax: 732-244-8140 e-mail: info@permapure.com Web Site: www.permapure.com
FC HUMIDIFIER SPECIFICATIONS
SETUP 1: GAS TO GAS
SETUP 2: WATER-TO-GAS
B
A
Model Number A B Ports
FC125-240-5PP 8.25" 4.75" 1/2" FC125-240-7PP 10.25" 6.75" 1/2" FC125-240-10PP 13.25" 9.75" 1/2"
FC150-480-7PP 11.50" 6.75" 1" FC150-480-10PP 14.50" 9.75" 1"
FC200-780-7PP 11.25" 6.50" 1" FC200-780-10PP 14.25" 9.50" 1"
FC300-1660-7ABS 12.10" 5.60" 1.5" FC300-1660-10ABS 15.10" 8.60" 1.5" FC300-1660-15ABS 20.10" 13.60" 1.5"
FC400-2500-10PP 16.63" 8.63" 1" or 2" FC400-2500-12PP 18.63" 10.63" 1" or 2" FC400-2500-15PP 21.63" 13.63" 1" or 2" FC400-2500-20PP 26.63" 18.63" 1" or 2"
FC600-7000-8PP 15.5 7.25 2" FC600-7000-14PP 21.5 13.25 2"
In Gas-to-Gas setup, constant depoint wet gas needs to be continuously flowing at same flowrate as sample. DI water is recommended.
GAS-TO-GAS PERFORMANCE
Model Number
FC150-480-7PP 35 84 112 FC150-480-10PP 50 120 160
FC200-780-7PP 57 137 182 FC200-780-10PP 80 195 260
FC300-1660-7ABS 121 291 387 FC300-1660-10ABS 173 415 553 FC300-1660-10HP 173 415 553 FC300-1660-15ABS 259 623 830
FC400-2500-10ABS 260 625 833 FC400-2500-12ABS 312 750 1000 FC400-2500-15ABS 390 938 1250 FC400-2500-20ABS 520 1250 1667
FC600-7000-8PP 583 1400 1860 FC600-7000-14PP 1020 2450 3267
* Units are expressed as flow rate of air (slpm) in relation to approach dewpoint temperature. Approach temperature is defined as difference between wet gas entering humidifier and humidified stream outlet. For example, for FC-125-240-10PP at 25 lpm, wet gas enters humidifier at 70 deg.C at Approached Tem­perature of 4 deg., humidified stream outlet would be
66 deg. C.
Approach Temperature*
4 Deg. 6 Deg. 9 Deg.
In Water-to-Gas setup, heated water needs to be continuously circulated. Flow with greater pressure needs to be flowing inside tubes to prevent tubing collapse. DI water should be circulated at 4 LPM for 50 LPM of dry gas flow.
WATER-TO-GAS PERFORMANCE
Performance can be controlled by increasing heat or by decreasing sample flow. Chart below shows performance of various models.
Water to Gas Humidification using FC-Style Humidifiers
Outlet DP (75C water) Outlet DP (60C water) Outlet DP (40C water) Outlet DP (80C water)
80 75
70 65 60 55
50
Dew Point (C)
45 40 35 30
200-10" 232 464 696 929 1,161 1,393 1,625 1,857 2,089
200-7" 163 325 488 650 813 975 1,138 1,300 1,463
150-10" 143 286 429 571 714 857 1000 1143 1286
150-7" 100 200 300 400 500 600 700 800 900
125-10" 71 143 214 286 357 429 500 571 643
125-7" 50 100 150 200 250 300 350 400 450 125-5" 36 71 107 143 179 214 250 286 321