PerkinElmer CLARUS 680 GC, CLARUS 600 GC User Manual

Page 1
CLARUS 600/680 GC
User’s Guide
GAS CHROMATOGRAPHY
Page 2
Page 3
Clarus 600/680 GC
User’s Guide
Page 4
ii
Release History
Part Number Release Publication Date
09936780 C February 2010
Any comments about the documentation for this product should be addressed to:
User Assistance PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, Connecticut 06484-4794 U.S.A.
Or emailed to: info@perkinelmer.com
Notices
The information contained in this document is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no warranty of any kind with regard to this document, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this publication may be reproduced in any form whatsoever or translated into any language without the prior, written permission of PerkinElmer, Inc.
Copyright © 2010 PerkinElmer, Inc.
Trademarks
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are protected by law.
PerkinElmer is a registered trademark of PerkinElmer, Inc.
Page 5
iii
Table of Contents
Introduction .................................................................................................................... 1-1
About This Manual .......................................................................................................... 1-3
Conventions Used in this Manual .................................................................................... 1-4
Notes, cautions and warnings ................................................................................... 1-4
Warning Signs on the Instrument..................................................................................... 1-5
Screen Abbreviations ....................................................................................................... 1-6
Glossary of Clarus GC Terms .......................................................................................... 1-7
Autosampler Terms .......................................................................................................... 1-7
Instrument-Specific Terms ............................................................................................... 1-8
Glossary of Chromatographic Terms ............................................................................. 1-10
Touch Screen Navigation .............................................................................................. 2-1
Chapter Overview ............................................................................................................ 2-3
Touch Screen ................................................................................................................... 2-4
Tabs........................................................................................................................... 2-6
Non-Configured Devices .......................................................................................... 2-6
Non-installed Devices ............................................................................................... 2-7
Run Button ................................................................................................................ 2-8
Stop Menu ............................................................................................................... 2-11
Signal Button .......................................................................................................... 2-11
Heated Zones .......................................................................................................... 2-13
Navigation Bar ............................................................................................................... 2-16
Injectors ......................................................................................................................... 2-17
Program Buttons ..................................................................................................... 2-18
Display Text ............................................................................................................ 2-18
PSS .......................................................................................................................... 2-18
Packed (Pkd) ........................................................................................................... 2-24
POC......................................................................................................................... 2-28
Capillary (Cap) Injector .......................................................................................... 2-29
Oven ............................................................................................................................... 2-30
Aux Zone ....................................................................................................................... 2-31
Detectors ........................................................................................................................ 2-32
FID .......................................................................................................................... 2-32
TCD ........................................................................................................................ 2-36
ECD ........................................................................................................................ 2-37
PID .......................................................................................................................... 2-39
NPD ........................................................................................................................ 2-40
FPD ......................................................................................................................... 2-41
Timed Events ................................................................................................................. 2-42
Page 6
iv
Events Tab .............................................................................................................. 2-42
Timed Events Table ................................................................................................ 2-43
Using the Active Method ............................................................................................... 3-1
Overview .......................................................................................................................... 3-3
Editing the Active Method ............................................................................................... 3-4
Getting Ready to Edit the Active Method ........................................................................ 3-5
Setting Up Oven and Inlet Temperature Programs .......................................................... 3-6
Configuring the Equilibration Time (Init) ................................................................ 3-7
Setting a Coolant Timeout Value .............................................................................. 3-7
Setting a Coolant Cut-In Temperature ...................................................................... 3-8
Setting Up an Isothermal Program ............................................................................ 3-8
Setting Up a One-Ramp Temperature Program ...................................................... 3-10
Holding an Oven Temperature ............................................................................... 3-11
Setting Injector Temperatures ........................................................................................ 3-14
Displaying Injector Temperature Screens ............................................................... 3-14
Turning Injector Heaters On or Off ........................................................................ 3-16
Changing Injector Temperatures ............................................................................ 3-16
Oven Temperature Programming Mode ................................................................. 3-17
Inlet Temperature Programming Mode .................................................................. 3-18
Setting Up a One-Ramp Injector Program .............................................................. 3-19
Viewing/Setting Carrier Gas Pressure/Flow/Velocity ................................................... 3-21
Setting Carrier Gas Pressure ................................................................................... 3-23
Setting Carrier Gas Flow using PPC ....................................................................... 3-25
Viewing/Setting Carrier Gas Velocity .................................................................... 3-27
Setting the Septum Purge Mode and Offset ................................................................... 3-29
Setting Up Valves .......................................................................................................... 3-37
Setting the Split Mode Using PPC .......................................................................... 3-37
Set up the Splitless Mode for a CAP or PSS Injector .................................................... 3-41
Splitless Injection with Pressure Pulse (Flow Pulse, Velocity Pulse) Technique ... 3-41
Contolling the Autosampler ........................................................................................... 3-51
Autosampler Trays and Carousel ............................................................................ 3-51
Autosampler Touch Screen ..................................................................................... 3-53
Start Button ............................................................................................................. 3-54
Stop Button ................................
............................................................................. 3-55
About Waste and Wash Vials ................................................................................. 3-62
Multi-Program Mode .............................................................................................. 3-62
About Sample Vials ................................................................................................ 3-63
Using a TCD with the Autosampler ....................................................................... 3-63
Available Syringes .................................................................................................. 3-64
Setting Up the Detectors ................................................................................................ 4-1
Overview .......................................................................................................................... 4-3
Page 7
v
FID Overview .................................................................................................................. 4-5
Setting Up an FID ..................................................................................................... 4-6
Automatically Ignite the Flame .............................................................................. 4-11
Manually Ignite the Flame ...................................................................................... 4-13
Using the FID in Corrosive Environments ............................................................. 4-15
Setting up an ECD ......................................................................................................... 4-17
United States Government Regulations for ECDs .................................................. 4-17
Special Instructions for ECD Cell Purchasers ........................................................ 4-18
United Kingdom Regulations ................................................................................. 4-20
ECD Overview ........................................................................................................ 4-20
ECD Set Up ............................................................................................................ 4-22
Setting up a TCD ........................................................................................................... 4-28
TCD Setup .............................................................................................................. 4-29
TCD Setup Summary .............................................................................................. 4-29
Using a TCD with the Autosampler or GSV .......................................................... 4-35
Setting up a PID ............................................................................................................. 4-36
PID Overview ......................................................................................................... 4-36
PID Setup ................................................................................................................ 4-37
Setting Up a NPD........................................................................................................... 4-43
NPD Overview ........................................................................................................ 4-44
Setting Up a NPD ................................................................................................... 4-45
NPD Bead Activation Procedure ............................................................................ 4-48
Setting up an FPD .......................................................................................................... 4-55
FPD Overview ........................................................................................................ 4-55
Setting Up the FPD ................................................................................................. 4-57
FPD Setup Summary .............................................................................................. 4-57
Configure the Linearizer and Signal Filter ............................................................. 4-57
Set the FPD Temperature ........................................................................................ 4-61
Detector Temperatures ................................................................................................... 4-72
Setting the Range ........................................................................................................... 4-74
Setting the FID, PID, or NPD Range ...................................................................... 4-74
Changing the FID, PID, or NPD Range During a Run .................................................. 4-75
Setting the TCD Current (Filament Current) ................................
................................. 4-78
Changing TCD Polarity During a Run ........................................................................... 4-80
Routing Detector Output ................................................................................................ 4-83
Rerouting Output .................................................................................................... 4-83
Attenuating Detector Signals Going to a Recorder ........................................................ 4-87
Optimizing a Detector Signal to an Integrator for Maximum Sample Dynamic Range 4-89
Matching Detector Signals with Your Integrator ........................................................... 4-90
Autozeroing a Detector .................................................................................................. 4-93
Using the Method Editor ............................................................................................... 5-1
Page 8
vi
Method Editor .................................................................................................................. 5-3
Method Edit Mode ........................................................................................................... 5-5
Oven Settings ............................................................................................................ 5-7
Managing Methods .......................................................................................................... 5-9
File Menu .................................................................................................................. 5-9
Using the Tools Menu .................................................................................................... 6-1
Tools Menu ...................................................................................................................... 6-3
Method Editor .................................................................................................................. 6-4
Event Log ......................................................................................................................... 6-5
Maintenance ..................................................................................................................... 6-6
Configuration ................................................................................................................. 6-10
Injectors (Deconfiguring and Configuring Injectors) ............................................. 6-10
Oven ............................................................................................................................... 6-18
Detectors ........................................................................................................................ 6-20
Run/AutoSampler ................................................................................................... 6-26
Setup ....................................................................................................................... 6-28
Pneumatics .............................................................................................................. 6-33
Connecting the PPC Modules ........................................................................................ 6-35
Reference Settings .................................................................................................. 6-48
Relays ..................................................................................................................... 6-48
Zeroing and Calibrating the PPC Modules ............................................................. 6-49
Zeroing the PPC Gas Modules................................................................................ 6-49
Setting the Reference Temperature and Pressure ................................................... 6-53
Calibrating the Carrier Gas Modules ...................................................................... 6-54
Calibrating the Detector Gas Modules .................................................................... 6-57
Calibrating the Oven Temperature ................................................................................. 6-60
Calibrate the Reference Thermometer .................................................................... 6-60
Place the Thermometer Probe in the Oven ............................................................. 6-61
Equilibrate the Oven Temperature .......................................................................... 6-62
Enter the Required Offset Value ............................................................................. 6-65
Remove the Thermometer Probe ............................................................................ 6-65
Utilities ........................................................................................................................... 6-66
Column Flow Calculator ................................
......................................................... 6-67
Split Flow Calculations Screen ............................................................................... 6-71
Column Length Calculator...................................................................................... 6-73
Background Compensation ..................................................................................... 6-74
Stopwatch ............................................................................................................... 6-76
Program (Max Oven Program Rates) ............................................................................. 6-77
Release… ....................................................................................................................... 6-78
Logout… ........................................................................................................................ 6-79
About ............................................................................................................................. 6-80
Page 9
vii
Setting up a Typical Analysis ........................................................................................ 7-1
System Status Screen ....................................................................................................... 7-4
Injector Settings ............................................................................................................... 7-6
Packed Column (PKD) Injector ................................................................................ 7-6
Programmed Split/Splitless (PSS) Injector ............................................................... 7-8
Detector Settings ............................................................................................................ 7-11
Flame Ionization Detector (FID) ............................................................................ 7-11
Detector .......................................................................................................................... 7-16
Thermal Conductivity Detector (TCD) ................................................................... 7-16
Autosampler ................................................................................................................... 7-22
Index
Page 10
viii
Page 11
Introduction
1
Page 12
Introduction
1-2
Page 13
Clarus 600/680 GC User’s Guide
1-3
About This Manual
This guide is divided into following chapters:
Chapter 1 Introduction
This chapter contains a brief introduction to the Clarus GC, the conventions and warnings used in the manual.
Chapter 2 Touch Screen Navigation
This chapter contains information that introduces you to the look and feel of the touch screen. It shows you how to turn on the GC, check the status, and perform fundamental operations such as changing parameters, selecting options, using the tools menus, etc.
Chapter 3 Active Method
This chapter contains step by step explanation of the active method. Detailed information on setting up the autosampler is contained in this chapter.
Chapter 4 Setting Up the Detectors
This chapter contains detailed information on setting up the detectors.
Chapter 5 Using the Method Editor
This chapter describes how to modify, save and use the method editor.
Chapter 6 Using the Tools Menu
The chapter contains detailed information on the tools menu section of the touch screen.
Chapter 7 Setting Up a Typical Analysis
This chapter describes how to run a typical analysis on the Clarus GC using the most common configuration.
Page 14
Introduction
1-4
Conventions Used in this Manual
Normal text is used to provide information and instructions.
Bold text refers to text that is displayed on the screen.
All eight-digit numbers are PerkinElmer part numbers unless stated otherwise.
Notes, cautions and warnings
Three terms, in the following standard formats, are also used to highlight special circumstances and warnings.
NOTE: A note indicates additional, significant information that is provided with some
procedures.
CAUTION
We use the term CAUTION to inform you about situations that could result in serious damage to the instrument or other equipment. Details about these circumstances are in a box like this one.
WARNING
We use the term WARNING to inform you about situations that could result in personal injury to yourself or other persons. Details about these circumstances are in a box like this one.
Page 15
Clarus 600/680 GC User’s Guide
1-5
Warning Signs on the Instrument
For detailed safety information see the Safety Practices chapter in the Clarus 600/680 GC Customer Hardware and Service Guide (P/N 09936781).
Caution, hot surface.
Caution, risk of electric shock.
Caution, laser radiation hazard.
Caution (refer to accompanying documents).
Page 16
Introduction
1-6
Screen Abbreviations
AUX – Auxiliary zone
Cap – capillary split/splitless injector
Cmptr – computer
Ctrl – syringe control parameters
ECD – Electron Capture Detector
FID – Flame Ionization Detector
FPD – Flame Photometric Detector
Gen – generate
Inj/Vial – injections per vial
Int – integrator
kPa – kilopascals
NPD – Nitrogen Phosphorus Detector
PID – Photoionization Detector
Pkd – packed injector
POC – Programmed On-Column Injector
PPC – programmed pneumatic control
Pre – # of preinjection syringe washes
Psi or psig – pounds per square inch (gauge)
PSS – Programmed Split/Splitless Capillary Injector
TCD – Thermal Conductivity Detector
Page 17
Clarus 600/680 GC User’s Guide
1-7
Glossary of Clarus GC Terms
The glossary of Clarus GC terms are divided into two types:
Autosampler Terms
Instrument-Specific Terms
Autosampler Terms
Term
Description
Washes Washing the syringe.
Pre-Wash The number of prewashes of sample to prime the syringe
(no pumping).
Pre-Inject The number of pre-injection syringe washes with a solvent.
Post-Inject The number of post injection syringe washes with a
solvent.
Pumps The number of times the syringe draws up sample and
evacuates it before acquiring the volume. This is done to eliminate bubbles.
Mode The style of injection.
Fast Fast speed of the syringe during sample injection. This is
used to eliminate discrimination in the needle.
Normal Normal speed of the syringe during sample injection.
Slow Slow speed of the syringe during sample injection in order
to inject directly into a wide-bore capillary column, in hot injection port.
Page 18
Introduction
1-8
Instrument-Specific Terms
Term
Description
Attenuation The amplification of the detector output signal.
Background compensation
A routine that automatically subtracts a stored calibrated baseline profile from the signal generated during a GC run.
Ballistically Changing the oven temperature as quickly as possible to
reach a set point.
Soft Cooling Soft-cooling is a technique whereby, the cooling rate of the
GC oven is initially limited by controlling (throttling) the intake of the ambient (cold) air used to cool the oven.
Detector background
The detector output signal when no components are being eluted.
Equilibration The delay time after the method set points have been
reached before the system becomes READY.
Flameout Detects if the flame is on to prevent injection into a
nonfunctioning system and subsequent loss of sample.
Isothermal method
A method in which the oven temperature remains constant throughout a GC run.
Method A collection of parameters that control the GC.
Negative-time event
A timed event that you set to occur before the instrument becomes READY.
Page 19
Clarus 600/680 GC User’s Guide
1-9
Term Description
Parameter An independent variable used to specify a condition to be
met.
PPC
Programmable Pneumatic Control
Pre-run The time after equilibration during which negative-time events
are executed.
PreVent
PPC version only. Sample management system.
Range For a Flame Ionization Detector, range means sensitivity of
the detector output signal.
Ready Indicates that all method parameters have reached their set
points and that you can start your analysis.
Run The time from sample injection to the end of the oven
temperature program.
Sleep mode The GC can be set to a predefined method for gas savings.
Temperature program
PPC version only. A program that specifies the rate of oven temperature increase from one plateau to the next, and the time the oven temperature is held constant at a plateau.
Timed Events Events that take place before or during a GC run as specified
in a timed events table.
Zone A heated area in the GC oven, injector, or detector.
Page 20
Introduction
1-10
Glossary of Chromatographic Terms*
Adsorption – A process that occurs at the surface of a liquid or solid as a result of the attractive forces between the adsorbent and the solute. These forces may be physical or weakly chemical.
Analysis – The complete investigation of a sample by gas chromatographic separation including identification of the sample components and quantitative measurements.
Anode – The positively charged electrode in any electrical circuit to which charged particles and ions are attracted.
Band Broadening – A process that occurs in the GC whereby the peak width for a component increases the longer the component travels through the column.
Baseline – The detector signal to a recorder or integrator when only the carrier gas is passing through the detector.
Baseline drift – Any regular change occurring in the baseline signal from the detector, usually resulting from column temperature and/or gas flow changes.
Blank run – A run without the sample being injected.
Bleed – The evaporation of the stationary phase from a column.
Capillary column (wall coated open tubular column) – A small-internal-diameter
column whose inside wall is coated with a liquid phase.
Carrier gas – The mobile phase of the separation system. An inert gas which transports the sample from the injector through the column to the detector. This gas is usually helium, hydrogen, or nitrogen.
Column conditioning – A process for producing a stable column by heating the column with carrier gas flowing to remove volatile impurities from the stationary phase.
Detectors – Hardware that responds to sample components producing an electrical signal that can be measured to quantitate the amount of each component present.
Flow programming – The stored PPC program is interpreted as real column mass flow settings or capillary column flow settings from which the inlet
*
Reference: Denney, R.C. A Dictionary of Chromatography.
Page 21
Clarus 600/680 GC User’s Guide
1-11
pressure is calculated. This is dependent on the capillary control mode and installed pressure or split control option.
Flow rate – The mass flow of carrier gas or detector gas in milliliters per minute.
Ghost peaks – Peaks that are not due to sample components, for example, peaks
produced by carrier gas impurities, septum, or components from previous analyses.
Injection port – The hardware through which the sample is introduced to the column by injection.
Linearity – Quantitatively all detectors will produce a linear response with respect to solute concentration over a defined range, for example, the Linear Range.
Liquid phase – The material in the column that causes the components to separate because of partitioning of the components between the mobile phase (carrier gas) and the stationary phase (liquid phase).
Lowest limit of detection – The smallest amount of sample that can be detected by the detector being used. Usually defined as any signal that is as great as two times the noise level. Also referred to as Minimal Detectable Quantity (MDQ).
Mobile phase – The gas which carries the solute (sample) along and over the column material. This carrier gas is inert and usually helium, nitrogen, or hydrogen.
Noise – Background signal fluctuations arising from a detector response. This response is the result of the column installed, carrier gas purity, electronic components, etc. The response of any detector is defined by the signal-to-noise ratio.
Partition Coefficient – The differential solubility of a substance in two different phases. In the case of gas–liquid chromatography, the sample components reach an equilibrium between the gas phase (mobile) and the liquid phase (stationary). Each component has a different partition coefficient thus causing separation in the column.
Pressure programming – Pressure control through an independent four-step, three-ramp program for each carrier gas channel.
Resolution – The degree of separation between two peaks.
Retention time – The time interval from the point of injection to the appearance
of the peak maximum, of a component’s signal.
Page 22
Introduction
1-12
Septum – Silicone rubber material placed in the injection port through which the injection is made. When the needle is withdrawn, the silicone rubber reseals, thus not allowing any sample or carrier gas to escape.
Stationary phase – The liquid or solid adsorbent portion of the column that retains components passing through the GC column.
Subambient – Below ambient temperature. The oven cools down below ambient temperature with the use of either liquid nitrogen or liquid carbon dioxide.
Syringes – Precision dispensing devices used to deliver sample to the GC. Liquid and gas syringes are available.
Tailing – When a peak is not symmetrical or Gaussian shaped but the back end is broadened, it is said to be tailing.
Temperature programming – A technique commonly used to increase the rate of elution of the components. After the sample is injected into the oven at a specific temperature, the temperature program increases the oven temperature to the prescribed temperature at a defined rate (in
°C/min).
Unretained peak – A component that is not retained by the column. The time taken for an unretained sample to pass through the column is the same time as the time taken for the carrier gas to pass through.
Velocity programming – Programmed linear velocity where the PPC controller interprets stored PPC programs as the average capillary column linear velocity settings.
Page 23
Touch Screen
Navigation
2
Page 24
Touch Screen Navigation
2-2
Page 25
Clarus 600/680 GC User’s Guide
2-3
Chapter Overview
You operate the Clarus GC using the touch screen on the front of the GC. The touch screen is the interface between you and the instrument, enabling you to control the Clarus GC in order for you to perform your analyses easily and conveniently. The touch screen contains active areas that you touch to perform the required action.
You may find it more convenient to use a stylus or you may like to touch the active areas of the touch screen. Do not use sharp, pointed objects.
A light touch is all that is needed. You do not need to press forcefully on an active area.
Please note that when you touch an active area, various processes are started in the instrument. Some of these processes may take several seconds. Be patient and wait for the process to complete before touching another active area on the touch screen.
This chapter gives a brief overview on touch screen definitions and navigation. For details on using the active method, method editor and tools menu see chapters 3, 4 and 5.
Page 26
Touch Screen Navigation
2-4
Touch Screen
The main display area on the touch screen is referred to as screens. You use various screens to set up your Clarus GC to perform the analyses. These screens contain:
Entry fields which allow you to make entries,
Buttons that you touch to start or stop actions or to display a dialog,
Option Buttons that you select drop down menu options from a list,
Boxes that you use to switch functions on or off and
Radio Buttons that allow you to choose one of two or more options.
The Clarus GC is controlled by a collection of operating parameters called the Active Method. You can prepare and save up to five methods and make any one of them the Active Method. However, the fifth method is reserved for TotalChrom and may be overwritten.
Page 27
Clarus 600/680 GC User’s Guide
2-5
The following Status Screen shows the different sections of the user interface.
The Status Screen displays icons that provide quick access to major areas of the system. The injector and detector buttons show graphic representations of the devices for each channel. If a heated auxiliary zone is configured, the Aux button with a thermometer appears active on the bottom of the screen. The Aux zone button with the carrier gas icon indicates that the PPC carrier gas zone is active. The icon buttons that represent the heated zones (injectors, detectors, oven, and Aux if configured) include a light to indicate the ready/not ready status. A red blinking light indicates not ready and a steady green light indicates ready status.
Click here to view the Signal. If
a method is not running the
The Title bar displays the name of the active method.
The Run button provides
access to the Autosampler and Manual Inject settings.
Channel B appears above
Channel A to emulate the order on top of the GC.
The Icon buttons provide quick
access to all configured areas of the instrument. An indicator light shows the ready/not ready status of each heated zone.
The status bar displays the overall ready/not ready status as well as text messages and the real time clock/date
The bottom bar displays the
Tools pop up menu. When the system is running the Stop button also appears.
The Start button appears so you can start or stop a run.
Page 28
Touch Screen Navigation
2-6
Tabs
Some of the active method screens include tabs to allow you to switch views associated with a specific channel.
Non-Configured Devices
If a device is installed but not configured, or configured as None, both the graphic and light are grayed out.
Touching the button prompts the following dialog box.
Page 29
Clarus 600/680 GC User’s Guide
2-7
The following screen capture shows the drop down list on the injector Configuration screen used to set a device to None.
Non-installed Devices
If a device is not installed, a grayed out deactivated button with no indicator light appears in the designated position.
Touching the button prompts the following dialog box:
Page 30
Touch Screen Navigation
2-8
Run Button
The Run button displays the Manual injection and Autosampler settings if the device is enabled. The bottom bar on these screens includes the Start button. When the system is running, the Start button changes to the Stop button. Since the Start button is only available on the Autosampler and Manual inject screens, you must initiate a run from these screens. However, when the system is running, the Stop button appears on all screens allowing you to Stop the run from anywhere in the application. If the system is Paused, the Resume button appears on all screens.
Autosampler Enabled
When an Autosampler is enabled, the Run button defaults to that screen. The following screens show the enable check box on the configuration screen and the associated Autosampler screen in the Active Method.
Page 31
Clarus 600/680 GC User’s Guide
2-9
When the GC is ready, touch the Start button to begin the run. If you touch Start before the system is ready, the run will start when the system reaches the ready status.
When the Autosampler is disabled, touching the Run button displays the Manual inject screen. When the instrument is ready, touch the Start button to begin the analysis. The countdown enable check box is selected by default. You can stop the countdown by pressing the Cancel button.
Manual Injection
If no Autosampler is installed or configured, the Run button displays the Manual injection screen.
The enable check box is selected by default. Touching the start button initiates the countdown starting from 5 seconds. When the countdown reaches 0, an audible beep prompts you to inject the sample. The GC will then start to program the oven and collect data. If the enable check box is deselected, the timer and tone are disabled. The system starts as soon as the Start Button is pressed and you make an injection simultaneously.
Page 32
Touch Screen Navigation
2-10
Enabled Disabled
When the Stop button is pressed, an abbreviated Stop Menu appears.
Page 33
Clarus 600/680 GC User’s Guide
2-11
Stop Menu
Stop Immediately – Touching this button immediately stops the Active Method and closes the Stop Menu dialog..
Go to a temp, program step – Displays the Go to Temp dialog for you to select the step you wish to go to at this time. Then press OK to activate the selected step.
Cancel – Cancels this Stop Menu dialog and returns you to the Active Method.
Signal Button
The Signal icon displays the graphic signal for the active method. If the system is not running, a baseline appears on the graph.
The Signal screen displays the selected detector signal over a fixed time period (2 minutes) with you selecting the scaling factor using the up and down arrow buttons. The raw signal is unprocessed and can appear slightly larger than the processed signal. The processed signal has gone through attenuation and offset filters.
Raw signal
Processed signal
Page 34
Touch Screen Navigation
2-12
The name of the channel appears on the top left of the screen. Detector controls include Ignite and Autozero.
Display controls include scaling and zeroing.
Ignite Button
Pressing the Ignite button starts the ignition sequence for a PPC-enabled FID. During ignition the plot and signal level are active.
For information on setting up manual pneumatics for the different detectors see the various detector sections in chapter 4 “Setting Up the Detectors” and sections in Chapter 3 on “Using the Active Method” and Viewing or Adjusting Carrier Gas Pressure Using Manual Pneumatics.
AutoZero Button
Autozeroes the selected detector.
Signal Plot
The plot graph displays the signal from the selected detector in “real time.” The window is 2 minutes wide. The signal commences at the y-axis and is plotted out to start + 1.75 minutes. The signal then scrolls back so the start is now +1 minute, the last 0.75 minutes are displayed from there, and the signal then scrolls out again.
The signal is plotted continually before, during and after a run.
Signal Display and Scaling
The signal display at the top of the screen displays the signal level of the detector in mV.
The default signal scale is 0-1024mV. The up/down arrows will adjust the display between 1 and 1024mV by halving or doubling the current factor
Page 35
Clarus 600/680 GC User’s Guide
2-13
Zero Display Button
The Zero Display button offsets the current signal level to 5% f.s.d. on the display.
Heated Zones
The Status screen displays graphic icons to represent the devices installed in the system.
Each type of injector and detector is represented with a customized graphic as well as a text label. Heated zone buttons display a status light indicator in the lower right corner. The indicator lights for individual zones follow the same convention as the system status light that appears in the Status bar. Steady green indicates that the zone is ready. A red blinking light indicates that the zone is not ready.
Injector and Detector Buttons
The screen is organized with Channel B devices above Channel A to reflect the position of the channels on top of the GC. Injector and detector buttons display
Page 36
Touch Screen Navigation
2-14
the active method settings for each device. Touching a device icon displays the selected screen as well as a row of tabs to allow easy access to the injector, oven, detector, and events for that channel. Touching the alternate channel icon on the top navigation bar displays the active screen and associated tabs for the other channel if it is installed and configured. The active method screens include navigation icons on the title bar that allow you to access the same elements available from the Status screen.
Page 37
Clarus 600/680 GC User’s Guide
2-15
Oven Button
Clicking the oven button provides access to the oven program settings and graph. You can set 1 to 10 program steps and up to 9 ramp rates. Use the up and down scroll buttons to the right of the oven entries to view program steps 5 – 10.
If both channels are running and the Oven is accessed from the Status screen, it defaults to Channel A. The oven program screen displays the same tabs and navigation icons as the injector and detector active method screens.
Aux Zone Buttons
If a heated Aux Zone is configured, pressing the Aux button displays a screen that follows the same format as the injector screens and shows the current temperature.
Page 38
Touch Screen Navigation
2-16
Navigation Bar
The navigation bar provides icons to allow users to seamlessly toggle between channels as well as access the same main areas of the system available from the status screen.
The following diagram shows the navigation bar of an Active Method with the corresponding icons on the status screen.
Status Screen Active Method Navigation Bar
Page 39
Clarus 600/680 GC User’s Guide
2-17
Injectors
The injector elements that appear on a particular injector screen depend on injector types as well as selected configuration options. Some injector types offer a variety of options while others are more limited. The labels in the following screen capture identify the main components of an Active Method injector screen. This example shows a screen for a PSS injector with PPC.
Injector Tab for Channel A-PSSI selected
Active Method Name
Temp Ready/Not
Ready Light
Carrier Gas
Ready/NotReady Light
Carrier gas
pressure, flow, or velocity
Split Flow
Ready/Not Ready
Split Flow actual
value
System Status – Oven
Program Step
Time/Date Display
Split flow set
value
Carrier Gas
program button
Temp program
button
Temp Ready/Not
Ready Light
Channel B Parameters
Temp Set Point
Page 40
Touch Screen Navigation
2-18
Program Buttons
The program button appears next to the temperature and Carrier Gas fields for a programmable PSS or POC injector. If the program button associated with a value is active, the entry field is grayed-out. You must go to the program screen to modify the settings.
If the programmable injector is not configured to program the setting, the program button is grayed out. In most cases, fields with a deactivated program button are editable, allowing you to modify the setting directly.
Display Text
The text displayed next to the set point entry fields show the current values. If the injector configuration allows you to display the value in other units, such as velocity or flow, the display text shows the actual values for all of the available options.
PSS
This section describes the main components of the PSS active method screen and the relationship between program and configuration options.
Page 41
Clarus 600/680 GC User’s Guide
2-19
Configuration Options
The following screens show the configuration choices and how they correspond to the entry fields in the active method.
Configuration Screen Active Method
Temperature Program
You can set injector (inlet) temperature programs by selecting, the program
button
on the active method screen is active.
Mode
Description
Inlet
In the inlet programmed mode, the injector is first heated to the injector temperature that is entered in the method. The injector then follows the injector temperature program that you defined.
Oven
The oven programmed mode is the default mode of operation. In this mode, the injector setpoint equals the oven setpoint plus 5 °C.
Page 42
Touch Screen Navigation
2-20
The temperature graph displays the set point value. The label below the graph displays the current step. The text to the left of the graph displays the actual temperature.
To modify the temperature program during a run, touch the program button. You can modify the temperature, time, or rate. However, if the program is currently in that step, the changes will be ignored.
Isothermal Program
If the program is isothermal, the program button is grayed out and the top-level field is active. Changes entered in the field take effect immediately.
Page 43
Clarus 600/680 GC User’s Guide
2-21
Carrier Gas Program
The Carrier Gas options for PSS include flow, velocity, or pressure. If the option for Capillary Control is selected, you can program the carrier gas according to one of the three values. The following example shows that Pressure was selected as shown in the following Configuration screen:
The corresponding actual values for the other units appear as read-only text on the screen as shown in the following Active Method screen.
Page 44
Touch Screen Navigation
2-22
.
Touching the Program button next to the field displays the associated Carrier Pressure Program screen for the entry field value.
Page 45
Clarus 600/680 GC User’s Guide
2-23
Capillary Control Off
The following screens are examples when Capillary Control is off
Split Flow
The PPC system adds the split flow offset to the method's split flow rate setpoint to obtain the Total split flow rate delivered to the inlet system. The total value is equal to the setpoint flow rate (the flow at the column entrance) plus the offset flow rate. The inlet system flow controller delivers the Total flow rate, not the setpoint flow rate.
Page 46
Touch Screen Navigation
2-24
Packed (Pkd)
The packed column injector has a fixed (non-programmable) temperature field. The carrier gas can be programmed by Pressure or flow. The following procedure shows how to navigate through the screens to configure PPC for a packed column injector.
1. Select Configuration from the
Tools menu
2. Touch the Pneumatics button.
.
Page 47
Clarus 600/680 GC User’s Guide
2-25
3. Touch the PPC Connect button.
4. Touch the OK button.
5. Touch Carrier B to Configure
your Packed column injector for PPC.
Page 48
Touch Screen Navigation
2-26
6. Select Pres Read from the drop-
down menu. Then touch the OK button.
7. Touch Close to close the Connect
PPC Device screen.
8. Touch OK on the Configuration
PPC screen.
9. Touch the B Injector button on the
Configuration screen.
Page 49
Clarus 600/680 GC User’s Guide
2-27
10. The Configure Injector B screen
appears.
11. Select Pressure or Flow from the
drop-down menu.
12. To exit out of Configuration,
touch the OK button then touch the Close button.
13. Display the active method and
view the B-Pkd injector parameters as shown on the right.
Page 50
Touch Screen Navigation
2-28
POC
The packed on co lumn injector can be programmed by Inlet (POCI) or Oven (POCO). The POC injector is configured and controlled according to the same parameters as the PSS injector.
Packed On Column Inlet
Packed On Column Oven
Page 51
Clarus 600/680 GC User’s Guide
2-29
Capillary (Cap) Injector
1. From the Configuration menu,
select Cap from the drop-down list.
2. Exit from Configuration and
display the active method screen.
Page 52
Touch Screen Navigation
2-30
Oven
The Oven Temperature Program provides fields and a graph that reflect the current temperature program as shown in the screen on the left below. Pressing the Go to button invokes a dialog box that allows users to raise the temperature to a value associated with another step. The last step of the current program is selected by default. Radio buttons for steps that exceed the number of steps in the current program are deactivated.
In the example below, step 2 is selected. As soon as the user touches the OK button the change is implemented. The graph in the capture on the right shows the immediate temperature rise and the departure from the program.
Graph and status bar show the oven is ready to run a program.
Touching the Go to button displays the Go to Temp dialog box to select a step to go to. In this example step 2 is selected. Press the OK button to immediately raise the oven temperature to the value specified in step 4.
The blue line in the graph reflects the departure from the program.
Page 53
Clarus 600/680 GC User’s Guide
2-31
Aux Zone
If the GC is configured with an auxiliary zone, the button appears on the status screen. Touching the Aux button displays a dialog box with the setpoint and actual temperature.
Page 54
Touch Screen Navigation
2-32
Detectors
Detector screens provide you with a way to set and control a number of parameters during a run, such as temperature, range, and attenuation. Other active elements include a Heater Off button to turn off the detector and an AutoZero check box. Selecting the AutoZero check box prompts the system to zero the signal before the run and also activates the Signal button on the Signal screen. The mV value next to the small signal icon that displays the current signal. The read-only text displays details about configuration settings. The following screen shows the major components that apply to all detector screens. The subsections that describe each detector type includes details about settings that are unique to that particular detector.
FID
The FID (Flame Ionization Detector) uses a hydrogen and air flame to generate ions from the combustion of an organic compound. The detector collects and measures the concentration of these ions to determine the amount of compound
Dynamic text
displays the current signal
Selecting the AutoZero checkbox zeros the signal before the run
Touch the
Ignite button to light the FID
flame
Temperature
setpoint.
Actual
temperature
Indicator lights
show the gas flow status
Selecting this
checkbox turns the
detector heater
Page 55
Clarus 600/680 GC User’s Guide
2-33
in the sample. The FID active method screen varies depending on whether the detector is configured with manual or programmable pneumatics. This section describes the screen components for each type of configuration.
Touching the Ignite button allows you to light the FID flame.
Auto-Ignition
The PPC controlled FID has flameout detection and an auto-ignite feature. Flameout detection occurs 0.5 minutes into the PRE-RUN time and it measures the user-entered baseline threshold level in mV. If the threshold level is exceeded, the flame is considered lit, otherwise the Clarus GC detects that the flame is out and auto-ignite automatically lights the flame before the start of a run.
With the FID temperature set, the air and hydrogen flows set, and auto-ignition activated by setting a positive mV value for the flame offset, the FID attempts auto-ignition at -0.5 min. before the system becomes ready. At this Pre-run time the status bar flashes “
Flame Ignition” as the air flow is shut off,
hydrogen purges the FID for about 10 seconds, and the ignitor energizes for ignition followed by a return of the air flow to the detector to sustain the flame.
Page 56
Touch Screen Navigation
2-34
After the flame ignites, the FID output level is monitored to confirm ignition before producing a READY signal. If ignition is not confirmed, the ignition process repeats up to two more times if needed. If ignition is not achieved after three attempts, the flame gases are set to off and a dialog box appears displays the following message:
Flame 1 or 2 failed to ignite.”
Touch the OK button. The status is made NOT READY and “FID 1 or 2 flameout” appears. If auto-ignition is attempted with the FID gases set to zero flow, the message “No ignition. Flame gases are off” appears on the screen. To proceed, go the PPC screen and touch Reset to the Configuration screen and reset the air and hydrogen to their 450 and 45 mL/min flow rates.
Auto-Ignite
The auto-ignition process only occurs before the start of each run just prior to the instrument becoming READY. However, the manual flame ignition process, performed by touching the Ignite button, can be initiated at any time.
Auto-ignite proceeds in the following sequence:
1. When auto-ignite begins the Clarus GC equilibrates.
2. Once the two minutes of equilibration expire, PRE-RUN begins and the
AutoSystem detects that the flame is not lit and indicates this by alternately flashing Flame? and Method 5 in the status bar.
3. After determining that the flame is not lit, the AutoSystem switches to a not
ready state and attempts to automatically ignite the flame. Flame Ignition 1 appears in the status bar.
4. The air flow is shut-off and hydrogen is purged through the FID for
10 seconds.
5. The ignitor energizes, the flame ignites (indicated by a pop sound), and air flow
is returned to sustain the flame. The FID output level is monitored to confirm that the flame is lit. If a lighted flame is not confirmed, the AutoSystem repeats the ignition process up to two more times.
Page 57
Clarus 600/680 GC User’s Guide
2-35
Configuration Settings
The following screens show the Active Method screen and the corresponding settings in the Configuration screen.
Active Method Configuration Screen
Page 58
Touch Screen Navigation
2-36
TCD
The Thermal Conductivity Detector (TCD) is a dual-channel detector which measures the difference in thermal conductivity between carrier gas flowing through a reference channel and carrier gas plus sample flowing through an analytical channel.
Four tungsten-rhenium filaments are connected in a bridge circuit, two in each channel. The TCD is set up with the same amount of carrier gas flowing through both channels. The presence of a sample component in the carrier gas flowing through the analytical channel changes the resistance of the filaments in that channel. This unbalances the bridge and produces a signal proportional to the amount of sample.
The following screens show the Active Method screen and the corresponding settings in the Configuration screen.
Active Method Configuration Screen
Page 59
Clarus 600/680 GC User’s Guide
2-37
ECD
The Electron Capture Detector (ECD) is the most sensitive detector available for the analysis of electrophilic compounds such as chlorinated hydrocarbons found in pesticide residues. It is concentration-sensitive and nondestructive.
The ECD cell contains a nickel foil in a cylinder through which the carrier gas flows. The coating of the foil's inner surface contains the radioactive isotope Nickel-63 and has a nominal sensitivity of 15 mCi. To optimize detector response, make-up gas is combined with column effluent.
Beta particles emitted from the isotope ionize the carrier gas. The resultant ions and electrons travel to the collector anode assembly under the influence of a pulsed polarizing voltage applied between the source and the collector. The pulse frequency is varied to maintain a constant average current. The presence of an electron absorbing species in the detector decreases the current as the absorbed electrons form ions, which travel more slowly than electrons. The pulse frequency of the polarizing supply is automatically controlled to maintain a constant current and is used to form the detector output signal.
The ECD works with nitrogen carrier or makeup gas or a carrier or makeup gas mixture of 95% argon/5% methane. When using capillary columns at low flows (5 mL/min or less), helium or hydrogen may be used as the carrier gas provided nitrogen or argon/methane is used as the make-up gas. A cross section view of an ECD is shown below.
The 63 Ni radioactive foil coating and requires careful handling to ensure that it does not create a safety hazard. In order to prevent a Thermal Runaway risk, the GC shuts down if any heated zone exceeds 470º C. If this occurs, the following message appears:
Warning
Instrument Shutdown
Thermal Runaway xxxxxx zone
Call your PerkinElmer service representative
Page 60
Touch Screen Navigation
2-38
Instrument shutdown also occurs if there is a PRT (Platinum Resistance Thermometer) or MPU (Micro Processor Unit) failure. In this case, the following message appears:
Instrument Shutdown
xxxx zone PRT error
Call your PerkinElmer service representative
The ECD works with nitrogen carrier or makeup gas or a carrier or makeup gas mixture of 95% argon/5% methane. When using capillary columns at low flows (5 mL/min or less), helium or hydrogen may be used as the carrier gas provided nitrogen or argon/methane is used as the make-up gas.
The following screens show the Active Method screen and the corresponding settings in the Configuration screen.
Active Method Configuration Screen
Page 61
Clarus 600/680 GC User’s Guide
2-39
PID
The Photoionization Detector (PID) is designed to allow the effluent from a chromatographic column to be ionized by ultraviolet light (provided the ionization potential of the effluent is less than that of the UV source). The current produced by the ion flow is measured by the detector and is proportional to the concentration of the ionized material.
The following screens show the Active Method screen and the corresponding settings in the Configuration screen.
Active Method Configuration Screen
Page 62
Touch Screen Navigation
2-40
NPD
The nitrogen phosphorus detector (NPD) is a highly specific thermionic detector for organically bound nitrogen and phosphorus.
The detector operates by electrically heating a glass bead, that contains an alkali metal, to the point where it emits electrons. A mixture of hydrogen gas and air flows around the bead to produce a hydrogen plasma. Stable intermediates are formed in the hydrogen plasma which then capture the electrons emitted from the bead to produce ions. A polarizing field directs these ions to the collector electrode, thereby creating a current which is amplified. Sensitivity is affected primarily by the air flow, and selectivity is affected primarily by the hydrogen flow.
The following screens show the Active Method screen and the corresponding settings in the Configuration screen.
Active Method Configuration Screen
Page 63
Clarus 600/680 GC User’s Guide
2-41
FPD
In the Flame Photometric Detector (FPD), the column effluent is mixed with hydrogen and then burned in air. Light emitted from the flame passes through a lens, a filter, and to a photomultiplier tube, which generates an electrical signal. The Flame Photometric Detector is used for detecting sulfur, phosphorus, or tin compounds, which produce chemiluminescent reactions with emissions at wavelengths characteristic of the S2, Sn, and HPO species.
The following screens show the Active Method screen and the corresponding settings in the Configuration screen.
Active Method Configuration Screen
Page 64
Touch Screen Navigation
2-42
Timed Events
Events Tab
The Events tab displays the timed events table and buttons, valves, and Aux gas controls.
Element Enabling
Timed events table Always
Aux gas button PPC configuration
Relays Always (see below for details)
Insert button Always
Edit button Always
Delete button Always
Clear button Always
Page 65
Clarus 600/680 GC User’s Guide
2-43
Timed Events Table
The timed events table displays the time, name, and value of each event. Timed events may be added and removed while the GC is running. The next event to be executed is automatically selected. Up to 32 events can be specified. The configured hardware determines which events are available. Events that are not applicable should not be available. The following table lists the available events.
The time range is -99 to 999 minutes. Events occurring after the temperature program is complete are ignored.
Add
The Add button displays settings that allow you to specify an event to add to the table.
Page 66
Touch Screen Navigation
2-44
Error Messages
Edit
Touching this button displays the active method page for editing timed events. If the GC is running, then edited events at times before the run time at which the edited list is accepted do not occur until the next run.
Page 67
Clarus 600/680 GC User’s Guide
2-45
Delete
Clear
Touching the Clear button deletes the contents of the entire event table. A dialog box confirms the action.
Pressing Yes clears the table. Touching the No button closes the dialog box and retains the contents of the Events table.
If the table is empty, the following Error message appears:
Page 68
Touch Screen Navigation
2-46
Value
The value box function depends on the event. For some events, such as Carrier X, a numeric value is required. For others, such as TCD polarity, a choice between two options is required.
Page 69
Clarus 600/680 GC User’s Guide
2-47
The PreVent/ProTect/MS Vent valves are never operated during a run and are not required in the Timed Events list.
No timed event should have the same time as another timed event.
Time resolution is 0.01 minutes.
Timed Events may be entered and removed in any sequence.
Pressing the Clear button prompts a dialog box:
”This action will clear the entire table. Do you want to continue? “Yes” “No”
Timed Events are re-ordered chronologically after each entry. The
capacity is 32 entries.
Page 70
Touch Screen Navigation
2-48
Relays
They appear as set in the configuration screen.
User can turn on the initial and/or current settings by touching the check box.
If GC is Ready, changing the initial setting will change the current setting.
The initial settings are stored in the method and are only changed if the user modifies them.
The current settings are changed via Timed Events.
Any settings may be changed during a run.
Page 71
Clarus 600/680 GC User’s Guide
2-49
Aux Zones
Heated Aux Zone
If the GC is configured with an auxiliary zone, the activated button appears on the status screen.
Touching the Aux button displays a dialog box with the setpoint and actual temperature.
To modify the temperature, press the minus or plus button. Touch OK to implement the value and close the dialog box.
Pneumatic Auxiliary Zones
Up to four pneumatic zones can be set for PPC carrier gas control. When one or more of the aux zones is configured, the following button appears on the status screen.
The same button also appears on the Events screen.
Touching the button displays the following screen:
Page 72
Touch Screen Navigation
2-50
Aux Zones that have been set in PPC configuration display the configured unit type and include a status indicator light. The initial values can be modified via timed events. The current value is shown next to the initial value field.
Page 73
Using the
Active Method
3
Page 74
Using the Active Method
3-2
Page 75
Clarus 600/680 User’s Guide
3-3
Overview
The Clarus GC is controlled by a collection of operating parameters called the Active Method. You can prepare and save up to five methods and make any one of them the Active Method. However, the fifth method is reserved for TotalChrom (external computer controlled software) and may be overwritten.
The operating parameters include the an oven temperature program; injector temperature or injector temperature program; detector temperature or detector temperature program; carrier gas pressure set for pressure flow or velocity; valve settings (for example, the operation of a gas sampling valve or split/splitless operation of capillary column); a timed events table, and a variety of parameters for controlling data output.
Active Method parameters are always saved when you turn the Clarus GC off, and automatically reloaded when you turn it back on.
Page 76
Using the Active Method
3-4
Editing the Active Method
If you have read and used the chapters in this manual sequentially, as recommended, by now you have probably installed and conditioned one or two columns, and set up one or two detectors. Also, you were probably not aware at the time that you modified certain Active Method parameters. Since the conditions used then were not necessarily those required by the analyses you want to perform, you should now edit the Active Method to set your operating parameters.
Examples of how to perform the tasks listed below are described in the sections which follow. You should, of course, substitute your values for ours.
Setting up oven temperature programs
Setting injector temperatures/programs (PSS and POC)
Setting detector temperatures
Viewing/setting carrier-gas pressure/programs (PPC)
Viewing/setting carrier-gas flow/programs (PPC)
Setting up valves
Setting FID, PID, NPD, or FPD range
Setting TCD filament current
Routing detector output
Attenuating detector signals
Autozeroing a detector
Building a timed events table
Page 77
Clarus 600/680 User’s Guide
3-5
Getting Ready to Edit the Active Method
Close the oven door.
NOTE: If a column has been installed, make sure you have carrier gas flow through the
column before closing the oven door and elevating the oven temperature.
The Status screen icon buttons provide access to the active method parameters.
NOTE: When you change an Active Method parameter, the change goes into effect immediately.
For example, if the system is READY and you change an injector temperature setpoint, the status screen displays NOT READY until the new setpoint is reached.
Page 78
Using the Active Method
3-6
Setting Up Oven and Inlet Temperature Programs
You can set up ramp programs (they are Init (1), 2, 3, through 10) using the Clarus GC software.
1. From the System Status screen touch the oven button
to display the oven
program screen.
2. Once the Oven program screen appears.
3. Select a field by touching it and use the up and down arrow or keypad
buttons to enter a temperature, equilibration time and rate. Enter a
value of 0 in the Rate field to make that the last step. You will then see the word END appear in the Rate field all the fields that follow are deactivated. You must touch END and then touch the up arrow button to activate the next row. The following figure shows a typical temperature step program.
Page 79
Clarus 600/680 User’s Guide
3-7
Figure 3-1. Typical temperature program steps.
Before you actually set up an oven temperature program, you should decide on the Equilibration Time you need and change the default value if necessary (see below).
Configuring the Equilibration Time (Init)
When you change the oven temperature setpoint, the system heats or cools the oven to the new setpoint, then equilibrates the column(s) for the specified Equilibration Time before the instrument becomes READY.
NOTE: A minimum Equilibration Time of < 1 minute is recommended for temperature programs.
Longer equilibration times are recommended for initial oven temperatures of < 50°C. An equilibration time of 0.0 minutes can be used for isothermal methods.
Setting a Coolant Timeout Value
The purpose of the coolant timeout value is to conserve coolant. If you select a coolant timeout value when you configure the Clarus GC, the coolant solenoid shuts off if a GC run is not started within this time period after the oven reaches the initial set temperature. Once the coolant shuts off, the GC remains not ready and the following message will appear, Coolant Timed Out: Touch OK to Continue.
NOTE: The coolant timeout must be greater than the equilibration time.
When you press OK, the coolant solenoid will then operate again and the oven will cool to the initial temperature.
Page 80
Using the Active Method
3-8
NOTE: When using a POC or PSS injector with the subambient accessory, the injector
temperature must not be set below the oven temperature. You can minimize coolant consumption if using the oven subambient with the PSS or POC installed but not being used, by deconfiguring the PSS or POC.
Setting a Coolant Cut-In Temperature
The coolant cut-in temperature is the temperature at which the coolant starts to cool the oven. The default cut-in temperature is 60 °C. The allowable cut-in temperature range is 30 °C - 110 °C. We recommend that YOU NOTE THE AMBIENT TEMPERARTURE AND SET the cut-in temperature 20 °C above your current ambient temperature.
For example: A room temperature of 20 °C uses a default temperature of 40 °C, whereas a room temperature of 25 °C uses a 45 °C cut-in temperature. Examples for setting up isothermal, 1, and 2 ramp programs follow.
NOTE: Higher cut-in temperatures cause more coolant to be consumed.
Setting Up an Isothermal Program
Setting up an isothermal program requires entering the following information in the sequence shown:
Initial oven temperature (Temp 1)
The total run time for the initial temperature (Time 1)
A temperature rate increase of 0 (Rate 1).
Page 81
Clarus 600/680 User’s Guide
3-9
To set up an isothermal program:
1. From the Oven tab set the initial temperature 1 to 75 ºC using the up and down
arrow or keypad
buttons.
2. Enter a temperature setpoint appropriate for your analysis, for example, 100 °C.
NOTE: The oven status changes to Not Ready. The changing oven temperature is displayed
under the oven icon as the oven heats up to the new setpoint. When the oven reaches the new setpoint, the system equilibrates for the Equilibration time (Init) specified in Oven Configuration screen and then becomes Ready.
For an isothermal run, Init. (TIME 1) is equivalent to the total run time. At the end of the time (TIME 1), the system changes to Not Ready, it equilibrates, and then goes Ready.
To change the hold time:
Go to the Oven tab and enter a Time appropriate for your analysis, for example,
25 min.
NOTE: Since a constant oven temperature is maintained for isothermal operation, the rate of
temperature increase must be set to. If the Active Method you are editing is already isothermal, the RATE parameter will have been properly set at 0 and is indicated on the screen by the word "End”.
Page 82
Using the Active Method
3-10
NOTE: You can set the rate to 0 but whenever a RATE is set to “0,” any subsequent program
steps, if present, are deleted from the Active Method.
Setting Up a One-Ramp Temperature Program
Setting up a one-ramp program requires entering the following information in the sequence shown:
Initial oven temperature (Init 1)
Hold Time for the initial temperature (Init 1)
Rate of temperature increase from the initial to the final oven temperature (Init Rate 1)
Final oven temperature (Init 2)
Isothermal time for the final temperature (Init 2)
Temperature increase of 0 °C/min which signifies the end of the RATE 2 step.
The following example shows how to convert an isothermal program into a one-ramp program using the previous isothermal method oven temperature parameter as a starting point.
Parameter
Value
Init Temp 50 °C
Init Time
5 min
Init Rate
2 °C/min
Temp 2
70 °C
Time 2 10 min
Rate 2 End (0 °C/min)
To set up a one-ramp program:
1. From the Oven tab, for the initial temp enter 50.
2. Enter 5 for initial time.
Page 83
Clarus 600/680 User’s Guide
3-11
3. Enter a ramp initial rate 1 of 2 °C/min.
4. For Temp 2, enter a temperature of 70 °C.
5. Enter a time of 10 min.
Setting Up a Multiramp Temperature Program
The following example shows how to expand the above one-ramp program to a two-ramp program. The parameter values used in the example are shown below:
Parameter
Value
Rate 2
4 °C/min
Temp 3
100 °C
Time 3 5 min
Rate 3 0 °C/min
1. From the oven tab enter a rate of 4.
2. Enter a temperature of 100.
3. Enter a time of 5.
NOTE: You can enter RATE 3, TEMP 4, and TIME 4 to set up a third ramp.
Holding an Oven Temperature
While running a temperature program, you can hold the oven temperature by setting it to the temperature of the unexecuted steps.
To hold an oven temperature during a temperature program:
1. In the System Status screen touch the oven button
.
Page 84
Using the Active Method
3-12
2. After you touch oven button the Oven tab screen appears. Next, touch the go to
temp button
to display the Go To Temp (temperature) screen.
2. From this window touch the button for the temperature level that you want to hold.
Touch OK. Modify the temperature on the Oven tab screen using the up and down arrow or keypad
buttons. When completed return to the system status
screen.
Page 85
Clarus 600/680 User’s Guide
3-13
The oven heats ballistically to the temperature value for oven Temp 2 and holds, instead of executing the time associated with oven Temp 2.
Figure 3-2. Oven Temperature Program for Rate 1.
To hold the oven program at a specific time during or before a run, use an oven HOLD timed event. See the section on Timed Events.
Page 86
Using the Active Method
3-14
Setting Injector Temperatures
Packed, Capillary (Split/Splitless), Programmed On-Column, and Programmed Split/Splitless
The examples in this section were derived by editing the Active Method with a capillary injector installed and configured in position A (front), and a packed injector installed and configured in position B (rear).
This section contains examples of how to:
Display injector temperature screens
Change injector temperatures
Turn injector heaters off
Turn injector heaters on
Set oven programming mode
Set inlet programming mode
Displaying Injector Temperature Screens
From the System Status screen touch the injector button this will
display the injector tab page.
Page 87
Clarus 600/680 User’s Guide
3-15
The injector information displays:
Page 88
Using the Active Method
3-16
Turning Injector Heaters On or Off
To turn the Heater Off touch the check box.
As the injector cools off, the injector temperature is displayed on the above screen.
Changing Injector Temperatures
1. To turn the heaters on and modify temperatures, make sure the Heater Off check
box does not display a check mark.
2. Touch the Temp field and using the up and down arrow or keypad
buttons increase the value to 255 ºC
.
Page 89
Clarus 600/680 User’s Guide
3-17
Oven Temperature Programming Mode
In this mode the POC or PSS injector is configured as Oven. The injector temperature is set to the oven temperature plus 5 °C. This is the default mode and the easiest mode to
operate in. For example, if the oven program is as follows:
Parameter
Value
Temp 1
50 °C
Time 1 2 min
Rate 1 5 °C/min
Temp 2
240 °C
Time 2
10 min
The initial injector temperature will be 55 °C. When the GC run begins, the injector will hold this temperature for 2 minutes, then it will ramp up to 245
°C at 5 °C/min and hold
for 10 min.
Figure 3-3. PSS oven mode temperature program.
NOTE: Opening the oven door turns off the oven and the injector heater. Also, if you reset the
oven temperature during the run to the next level, the injector temperature will follow.
Page 90
Using the Active Method
3-18
Inlet Temperature Programming Mode
The inlet temperature programming mode is used when the POC or PSS injector is configured as Inlet. In this mode, the injector follows the injector temperature program that you entered in the method.
The initial default values are: Initial (Temp 1) is 50 °C, Injection Time is 999 min, and Inj Rate is 0. With non­programmed injectors, an entry of 0 turns the heated zone off. With programmed injectors, a temperature of 0 is possible. Therefore, when 0 is entered for the initial injector setpoint, the following screen appears:
1. From the selected injector tab screen make sure that the injector heater is off.
2. Set the initial temperature to 50
°C.
Page 91
Clarus 600/680 User’s Guide
3-19
Setting Up a One-Ramp Injector Program
To set up a one-ramp injector program, enter the following parameters:
Initial injector temperature (Init Temp)
Isothermal time for the initial temperature (Init Time)
Rate of temperature increase from the initial to the final injector temperature (Init
Rate)
Final injector temperature (Temp 2)
Isothermal time for the final temperature (Time 2)
Temperature increase of 0
°C/min, which signifies the end of the Rate 2 step.
The following example shows how to build a one-ramp temperature program for your POC or PSS injector. The parameter values used in the example are shown below:
Parameter
Value
Init Temp 60 °C
Init Time 0 min
Init Rate
100 °C/min
Inj Temp 2
250 °C
Inj Time 2 15 min
Inj Rate 2
End (0°C/min)
To set up a one-ramp injector program:
1. From injector tab screen touch the temp program button
.
Page 92
Using the Active Method
3-20
3. The temp program screen appears for the selected injector. Use the up and down
arrow buttons or numeric keypad button to program the temperature, time and rate. Refer to the table on the previous page for some examples for values to set up a one ramp injector program.
4. When you have completed setting up the program touch Close button then the
system icon
on the navigation bar to return to the System Status screen.
NOTE: You can enter Rate 2, Temp 3, and Time 3 to set up a second ramp. An entry of 999 for
Rate 1 or 2 will cause the injector to ballistically heat up.
NOTE: If the Injector Time is set with the default value of 999, the injector temperature will hold
at the current temperature until the end of the oven program. Otherwise, the injector temperature will hold for the time you have entered, or until the run ends.
NOTE: If you have the subambient option, the POC and PSS injectors are linked to the oven
subambient option; therefore, you cannot operate the injectors below the oven subambient temperature.
Page 93
Clarus 600/680 User’s Guide
3-21
Viewing/Setting Carrier Gas Pressure/Flow/Velocity
NOTE: For a description of programmable pneumatics control (PPC), refer to “PPC
Fundamentals” In the Hardware Guide.
PPC enables you to select and set the carrier gas Pressure, Flow, or Velocity. In all three modes, you can operate at a constant pressure, flow, or velocity or at a programmed pressure, flow, or velocity. You can also make a pressure pulse, flow pulse, or velocity pulse injection. In this case the initial pressure, flow, or velocity is quite high at the time of injection. This minimizes the expansion in the injector liner as the liquid sample is vaporized and increases the transfer time onto the column. This is recommended for splitless operation of wide boiling point samples. In this case, you would actually program the pressure, flow, or velocity down to the normal column pressure, flow, or velocity for the remainder of the chromatographic run.
The following terms are used in PPC analysis:
Zero removes the flow controller from the GC method and sets the flow to 0.
Flow enables the flow controller in the GC method and sets its initial flow to
30% of the nominal maximum flow rate (according to the installed flow-range element and carrier gas type). The operator has the option to enter a different flow rate and enter a flow program. When in the flow mode, the PPC system also monitors the column inlet pressure. If the inlet pressure falls outside a selectable range, a leak-check fault occurs.
Pressure controls the column inlet pressure and reads the flow rate from the flow sensor in the flow controller. The initial pressure in the method defaults to 5 psig. The operator has the option to enter a different initial pressure and optionally enter a pressure program.
Carrier Gas Mass-Flow Controller Operation
Once the carrier gas operation mode has been selected (see the next section “Setting Carrier Gas Pressure”), you can also modify the default setpoints in the active GC method, edit the carrier-gas setpoints in stored methods, or generate a new method and enter carrier-gas setpoints. The following parameters control the carrier gas.
Page 94
Using the Active Method
3-22
Initial Carrier-Gas Setpoint
The carrier gas mass-flow controller receives set points of flow or pressure depending on the configured carrier gas mode. The program calculates the initial set point from the active GC method. The initial set point is established upon setting up a method, ending a run, or resetting the GC. The flow rate is held constant at the initial value if no additional carrier-flow entries are made in the GC method. The controller receives pressure set points when operating in the pressure control mode.
Carrier-Gas Program
The GC method accommodates up to three additional carrier-gas program steps each consisting of a hold time, ramp rate, and plateau set point. Sequential set points may be greater than or less than the preceding set point. The PPC system programs the carrier­gas controller with a positive or negative ramp as required.
At the end of the GC oven program, the carrier gas resets to the initial set point value in the method even if a carrier-gas program is in progress. The carrier gas also resets if the end of the carrier-gas program occurs before the end of the oven program.
Carrier-Gas Timed Events
You can also use timed events to produce step changes in the set point. Carrier-gas events can occur before or after a run starts. A carrier-gas event supersedes the method set point value at the time the event occurs. However, if the carrier-gas program in the method advances to a new step after a carrier-gas event, the new program step or ramp takes control.
Leak Checking
The Clarus GC monitors the inlet pressure with a carrier gas mass-flow controller in use. After selecting the carrier-gas Flow mode in configuration, you can enter the maximum inlet pressure value. If the inlet pressure drops below or rises above the set limit for more than one minute, a warning message appears.
Page 95
Clarus 600/680 User’s Guide
3-23
Setting Carrier Gas Pressure
To view or set the carrier pressure:
1. From the System Status screen touch the corresponding Injector button (A or B)
for which you want to set the carrier pressure. The injector screen appears.
In this example, the Clarus GC is configured with the PSS capillary mode on for pressure (psi). If the option for capillary control has been selected during configuration, you can program the carrier gas according to one of three values: pressure, velocity, or flow. The carrier gas field on the injector tab screen displays the set point value. The label below the field displays the current step. The actual values for all the available configuration options are displayed to the left as read only text on the screen.
2. Touch the program button
for Carrier Gas.
The Carrier Pressure Program window appears:
Page 96
Using the Active Method
3-24
3. Touch the field, ranging from the Initial field to the step 4 field, and use up and
down arrow or keypad
buttons to enter a pressure, hold time and
rate.
NOTE: The rate for the range is 0 (end) to 99. A rate of 999 is a ballistic setting
(maximum).
CAUTION
The ballistic mode of programming is not recommended. Setting a program rate will be more reproducible.
4. Touch Close to return to the injector screen.
Page 97
Clarus 600/680 User’s Guide
3-25
Setting Carrier Gas Flow using PPC
To view or set the carrier flow:
1. From the System Status screen touch the Injector button (for example, Pkd or
Cap) corresponding to the injector you wish to set the carrier gas flow. The tab for the selected injector appears:
In this example, the Clarus GC is configured with the PSS capillary mode on for flow. If the option for capillary control has been selected during configuration, you can program the carrier gas according to one of three values: pressure, velocity, or flow. The carrier gas field on the injector tab screen displays the set point value. The label below the field displays the current step. The actual values for all the available configuration options are displayed to the left as read only text on the screen.
2. Touch the program button
for Carrier Gas.
Page 98
Using the Active Method
3-26
The Carrier Pressure Program window appears:
3. Touch to highlight each individual field, ranging from the Initial field to the
step 4 field, and use up and down arrow or keypad
buttons to enter
a pressure, HOLD time and HOLD rate.
NOTE: The rate for the range is 0 (end) to 99. A rate of 999 is a go to setting.
CAUTION
The go to mode of programming is not recommended. Setting a program rate will be more reproducible.
4. Touch Close.
Page 99
Clarus 600/680 User’s Guide
3-27
Viewing/Setting Carrier Gas Velocity
You can view/set carrier gas velocity using PPC.
To view or set the carrier gas velocity:
1. From the System Status screen touch corresponding Injector button for which
you wish to set the carrier gas velocity. The page for the selected injector appears:
In this example, the Clarus GC is configured with the capillary mode on for velocity (as shown by vel.-init displayed under the carrier gas field). If the option for capillary control has been selected during configuration, you can program the carrier gas according to one of three values: pressure, velocity, or flow. The Carrier Gas field on the injector tab screen displays the set point value. The label below the field displays the current step. The actual values for all the available configuration options is displayed to the left as read only text on the screen.
2. Touch the program button
for Carrier Gas.
Page 100
Using the Active Method
3-28
The Carrier Velocity Program window appears:
3. Touch each individual field, ranging from the Initial field to the step 4 field, and
use up and down arrow or keypad
buttons to enter a velocity,
equilibration time and rate.
NOTE: The rate for the range is 0 (end) to 99. A rate of 999 is a go to setting
CAUTION
The go to mode of programming is not recommended. Setting a program rate will be more reproducible.
4. Touch Close.
Loading...