3.3 Gas Supply Safety Devices
3.3.1 Gas Supply Cut-off Device
A gas cut-off device, triggered by low oxygen
supply pressure, cuts the supply of nitrous
oxide, and carbon dioxide (if fitted).
The cut-off operates when the oxygen
pressure falls to 186 ±14 kPa (27 ±2 psig).
Gas supplies are reinstated only when the
oxygen supply pressure rises above 227
±14 kPa (33 ±2 psig).
3.3.2 Oxygen Supply Failure Warning
Whistle
A whistle gives an audible warning when
there is a reduction of oxygen supply
pressure.
Operated solely by the remaining oxygen in
the machine system, the warning whistle is
prolonged by an oxygen reservoir built into
the gas circuit, allowing a minimum warning
whistle of 7 seconds duration.
The whistle will start to sound when the
pressure falls to 200 ±21kPa (29 ±3 psig),
and will continue to sound until the pressure
falls to approximately 70 kPa (10 psig).
Oxygen consumption of the whistle is
approximately 2 L/min when sounding and
nil at other times.
WARNING
Gas supply failure systems will not
necessarily operate as indicated during any
procedures that are outside the scope of the
indications for use of the machine (see
section 2).
Machine set to deliver Oxygen only through
Auxiliary Outlets :
The oxygen failure alarm is designed to operate
during normal use of the machine, i.e. when
providing controlled concentrations and flows of
gases to a patient breathing system.
Do not use the machine solely to provide large
flows of oxygen, via the anaesthetic machine
auxiliary outlets, to external devices which may
not be equipped with a supply failure alarm.
3.3.3 Fresh Gas Pressure Relief Valve
The valve is designed to prevent fresh gas
being delivered to the breathing system at
pressures exceeding 41 kPa (6 psi).
This valve also protects machine
components against excessive pressure in
the event of a total blockage of the CGO.
3.3.4 Oxygen Supply Visual Indicator
The indicator (1) is mounted on the front of
the machine and is operated from the
oxygen supply and shows GREEN when the
supply is at working pressure, and RED if the
pressure falls.
WARNING
Gas supply failure systems within the anaesthetic
machine will not necessarily operate as indicated
during any procedures that are outside the scope
of the indications for use of the machine (see
section 2).
Machine set to deliver Air only :
When the machine is operated with an Air flow
only, note that there may still be retained oxygen
in the system, and that the visual O2 alarm will
continue to indicate green, even though oxygen is
not being delivered.
3.3.5 Mechanical AHD
A mechanical link between the oxygen
control valve and a needle valve in the
nitrous oxide flow ensures that the machine
delivers a fresh gas mixture with a minimum
30% (±3%) oxygen, irrespective of the flow
of nitrous oxide set by the anaesthetist.
With the nitrous oxide control valve fully
open, the oxygen and nitrous oxide flows are
then both controlled by the oxygen control
valve.
See section 3.4 for a full description
3.3.6 Low Pressure Gas Tubing
Diameter-indexed tubing is used for the low
pressure gas system - see section 4.
7
DESCRIPTION
1