(A) Machine set to deliver Air only
When the machine is operated with an Air
flow only, note that there may still be
retained oxygen in the system, and that the
visual O2 alarm will continue to indicate
green, even though oxygen is not being
delivered.
(B) Machine set to deliver Oxygen only
through Auxiliary Outlets
The oxygen failure alarm is designed to
operate during normal use of the machine,
i.e. when providing controlled
concentrations and flows of gases to a
patient breathing system, as described in
Section 2 (Purpose).
Do not use the machine solely to provide
large flows of oxygen, via the anaesthetic
machine auxiliary outlets, to external
devices which may not be equipped with a
supply failure alarm.
12. The machine must not be fitted with
more than four operator accessible
mains socket outlets. There is a risk of
an excessive leakage current.
13. The use of antistatic or electrically
conductive breathing hoses is not
recommended when using high
frequency electrical surgery equipment
(e.g. : Diathermy). Burns may be
caused.
14. Before any electrically powered machine
is used clinically for the first time,
check that the hospital engineering
department has carried out an earth
continuity test.
15. Before using any additional electrical
equipment powered by the auxiliary
sockets on the machine, check that the
additional equipment is correctly wired
and is earthed through its plug.
A missing or defective protective earth
conductor may increase earth leakage
currents to the patient to values
exceeding the allowable limits,
resulting in ventricular fibrillation, or
interference with the pumping action of
the heart.
16. Additional equipment placed on the top
shelf must be securely attached.
Take care when moving a fully loaded
machine, particularly when negotiating
ramps.
Check that hoses or power leads are not
trailing on the floor.
3
WARNINGS AND CAUTIONS
17. Prima SP2 series anaesthetic machines
are not MRI compatible.
CAUTIONS
1. Flowmeter needle valves are designed to
seal with light torque and may be damaged
if tightened excessively.
Do not force the control knob past either the
fully open or fully closed positions.
2. Open cylinder valves slowly to avoid
damage to pressure reducing valves.
Ensure that cylinder valves are at least one
full turn open when in use.
3. Under no circumstances should anaesthetic
agents be used for cleaning purposes.
4. After use, always disconnect the machine
from the piped gas supply and/or close the
gas cylinder valves.
5. Mechanical AHD system - The oxygen flow
control is restricted to prevent the needle
valve from fully closing. This ensures a
minimum basal flow of oxygen.
DO NOT attempt to close the flow to zero.
Do not overtighten.
6. Compressed gas supplies must be clean
and dry.
7. When the auxiliary gas outlets are in use on
a machine with cylinder supply only, or if
the pipeline supply is not in use, check flow
rate requirements, and ensure that
adequate back-up cylinders are available.
8. The requirements of IEC 60601-1-1 apply to
any device connected to the auxillary
sockets. Users must be aware of the risks of
increased leakage currents when
equipment is connected to the auxillary
power sockets
9. In the event of malfunction of any device
powered by the auxiliary sockets, check if
the circuit breaker has tripped.
10. On machines with pipeline connections,
note that a malfunction of the central gas
supply within your facility may cause
immediate cessation of gas delivery and
total anaesthesia system failure.