Penlon Prima SP User manual

Prima SP Anaesthetic Machine Range
User Manual
Quality and Assurance in Anaesthesia
THE IMPORTANCE OF PATIENT MONITORING
WARNING
Anaesthetic systems have the capability to deliver mixtures of gases and vapours to the patient which could cause injury or death unless controlled by a qualified anaesthetist.
There can be considerable variation in the effect of anaesthetic drugs on indi­vidual patients so that the setting and observation of control levels on the anaesthesia systems does not in itself ensure total patient safety.
Anaesthesia system monitors and patient monitors are very desirable aids for the anaesthetist but are not true clinical moni­tors as the condition of the patient is also dependent on his respiration and the
functioning of his cardio-vascular system.
IT IS ESSENTIAL THAT THESE ELEMENTS ARE MONITORED FREQUENTLY AND REGULARLY AND THAT ANY OBSERVA­TIONS ARE GIVEN PRECEDENCE OVER MACHINE CONTROL PARAMETERS IN JUDGING THE STATE OF A CLINICAL
PROCEDURE.
IMPORTANT
Servicing and Repairs
In order to ensure the full operational life of this anaesthetic machine, servicing by a Penlon-trained engineer should be undertaken periodically.
The machine must be serviced to the schedule detailed in section 8. Details of these operations are given in this Service Manual, available only for Penlon trained engineers.
For any enquiry regarding the servicing or repair of this machine, contact the nearest accredited Penlon agent:
or communicate directly with:
Service Department Penlon Limited Radley Road Abingdon OX14 3PH UK
Tel: +44 (0) 1235 547063 Fax: +44 (0) 1235 547062
E-mail: service@penlon.co.uk
Always give as much of the following information as possible:
1. Type of equipment
2. Product name
3. Serial number
4. Approximate date of purchase
5. Apparent fault
(i)
FOREWORD
This manual has been produced to provide authorised personnel with information on the function, routine performance, maintenance checks and repair procedures applicable to the Prima SP anaesthetic machine range.
Information contained in this manual is correct at the date of publication. The policy of Penlon Limited is one of continued improvement to its products. Because of this policy, Penlon Limited reserves the right to make any changes which may affect instructions in this manual, without giving prior notice.
Personnel must make themselves familiar with the contents of this manual and the machine’s function before using the apparatus.
Copyright © Penlon Limited, 2002. All rights reserved.
(ii)
CONTENTS
Page No.
USER RESPONSIBILITY 01
1. WARNINGS AND CAUTIONS 02
2. PURPOSE 05
3. DESCRIPTION 06
3.1 Framework and General Construction 06
3.2 Gas Circuit NON-AHD MACHINES 07
3.3 Gas Circuit Schematic 07
3.4 Gas Supply Safety Devices 10
3.5 Mechanical AHD 11
3.6 Pressure Gauges 12
3.7 Flowmeters and Controls 13
3.8 Vaporizers 14
3.9 Common Gas Outlet (CGO) Block 15
3.10 Electrical Power Supply 16
3.11 Third/Fourth Gas Options 17
3.12 Auxiliary Gas Outlets 17
3.13 Oxygen Monitor 18
3.14 A100 Absorber 23
3.15 Prima SP Anaesthetic System - MRI Compatibility 23
4. SPECIFICATION 24
4.1 Physical Dimensions 24
4.2 Gas Supplies 25
4.3 Flowmeters 25
4.4 Gas Pressures 26
4.5 Auxiliary Gas Outlets 27
4.6 Oxygen Failure Warning Devices 27
4.7 Oxygen Flush 27
4.8 Mechanical AHD Systems 27
4.9 Environmental 28
4.10 Third and Fourth Gas Options 28
4.11 Electrical Supply 28
4.12 Oxygen Monitor 29
5. PRE-USE CHECKS 31
5.1 Pre-use Check List 31
5.2 Pre-use Checks (Non-AHD Machines) 33
5.3 Pre-use Checks (Machines with Mechanical AHD) 35
5.4 Leak Rate Check 37
5.5 Electrical Supply 37
5.6 Patient Breathing System 38
5.7 Oxygen Flush 42
5.8 Anaesthetic Gas Scavenge System (AGSS) 42
5.9 Alarm System Testing 43
5.10 Vaporizers 44
5.11 Ventilator 44
5.12 Oxygen Monitor 45
(iii)
CONTENTS
6. FUNCTION TEST 49
6.1 Introduction 49
6.2 Machine Frame 50
6.3 Electrical Safety Tests 50
6.4 Pipeline Gas Supply and Non-Return Valve 51
6.5 Cylinder Gas Supply and Pressure Reducing Valves 52
6.6 Flowmeter Unit 54
6.7 Gas Safety Devices 56
6.8 Vaporizers and Back Bar Manifold Assembly 57
6.9 Reduced Pressure Gas Circuit Leakage 58
6.10 Emergency Oxygen Flush Control Test 58
7. FAULT FINDING 59
7.1 Introduction 59
7.2 Fault Diagnosis Table 60
8. SERVICE SCHEDULE 69
8.1 Service Schedule 69
8.2 Six Month Service Checks 69
8.3 Twelve Month Service Checks 72
8.4 Twenty-four Month Service Checks 75
8.5 Seventy-two Month Service Checks 78
9. SERVICING PROCEDURES 82
9.1 Servicing 82
9.2 Ancillary Equipment 82
9.3 Cleaning and Sterilisation 84
9.4 Gas System Components 85
9.5 Internal Gas Pipework 85
9.7 Front Cover - Remove and Refit 88
9.7 Flowmeter Cover 89
9.8 Gear Linkage - Reset 90
9.9 Control Valve Capsule Removal/Replacement 92
9.10 Oxygen Reservoir - Removal/Replacement 93
9.11 Secondary Regulators 94
9.12 Gas Block - Remove and Refit 95
9.13 Selectatec Compatible Manifold Block 96
9.14 Gas Delivery Switch 97
9.15 Air/N2O Interlock Switch 98
9.16 Oxygen Failure Visual Indicator 99
9.17 CGO - Clippard Valve Replacement 100
9.18 Flowmeter Tubes - Removal 101
9.19 Flowmeter Assembly - Removal 102
9.20 Flowmeter Lighting - Removal 103
9.21 Oxygen Monitor - Battery Replacement 104
9.22 Oxygen Monitor Sensor 105
9.23 A100 Circle System Absorber 106
10. SPARE PARTS LIST 107
11. APPENDIX 130 Care of Back-up Battery 130 Product Classification and Labelling Terminology 131 Wiring Diagram - Lighting 132 Wiring Diagram - General 133
(iv)
USER RESPONSIBILITY
This anaesthetic machine has been built to conform with the specification and operating procedures stated in this manual and/or accompanying labels and notices when checked, assembled, operated, maintained and serviced in accordance with these instructions.
To ensure the safety of this device it must be checked and serviced to at least the minimum standards laid out in this manual. A defective, or suspected defective, product must not under any circumstances be used.
The user must accept responsibility for any malfunction which results from non­compliance with the servicing requirements detailed in this manual.
Additionally, the user must accept responsibility for any malfunction which may result from misuse of any kind or non­compliance with other requirements detailed in this manual.
Statements in this manual preceded by the following words are of special significance:
WARNING means there is a
possibility of injury to yourself or others.
CAUTION means there is a possibility
of damage to the apparatus
or other property.
NOTE indicates points of
particular interest for more efficient and convenient
operation.
Always take particular notice of the warnings, cautions and notes provided throughout this manual.
Worn, broken, distorted, contaminated or missing components must be replaced immediately. Should such a repair become
necessary it is recommended that a request for service advice be made to the nearest Penlon accredited agent.
This device and any of its constituent parts must be repaired only in accordance with written instructions issued by Penlon Limited and must not be altered or modified
USA and Canadian Federal Law restricts the sale and use of this device to, or on the order of, a licensed practitioner.
1
1. WARNINGS AND CAUTIONS
The following WARNINGS and CAUTIONS must be read and
understood before using this anaesthetic apparatus.
WARNINGS
1. This apparatus is designed for use
only with non flammable anaesthetic agents. It must not be used with or in close proximity to flammable anaesthetic agents, due to a possible fire or explosion hazard.
2. Exterior panels must not be
removed by unauthorised personnel and the apparatus must not be operated with such panels missing. On machines with an electrical power supply, there is a possible electric shock hazard.
during test procedures as well as during use with a patient.
6. Prima SP machines must only be used with Sigma Delta vaporizers (or other vaporizers approved by Penlon Limited) installed by means of the Cagemount or Selectatec system. Free-standing vaporizers may be accidentally tipped, resulting in excessive and uncalibrated volumes of anaesthetic drug entering the breathing system
Do not install or connect any vaporizers of any description between the Common Gas Outlet (CGO) and the breathing system unless they are specifically designed for such use. (If this is done, the oxygen flush flow will pass through the vaporizer and may result in gross overdosage when the flush valve is operated.)
3. No oil, grease or other flammable lubricant or sealant must be used on any part of the machine in close proximity to medical gas distribution components. There is a risk of fire or explosion.
4. When attaching cylinders of medical gases ensure that the machine yoke and cylinder faces are dust free and clean and that the sealing washer provided is in position between the cylinder valve and the yoke. Tighten the yoke securely before opening the cylinder valve. Dust and dirt presents a fire hazard in the presence of high pressure gas. Leakage of high pressure gas can cause serious injury.
5. Anaesthesia apparatus must be connected to an anaesthetic gas scavenging system (AGSS) to dispose of waste gas and prevent possible health hazards to operating room staff. This requirement must be observed
7. The breathing system which conveys gases from the anaesthetic machine to the patient and disposes of expired gases is a vital part of the anaesthetic delivery system. Because breathing systems require frequent cleaning and disinfection they are not a permanent part of the anaesthetic machine and therefore cannot be directly under the control of the anaesthetic machine manufacturer. However, we strongly recommend that only breathing systems which have been approved and authorised by Penlon for use with the Prima SP range should be employed. This is particularly important when mechanical ventilation is employed.
8. When mechanical ventilation is employed the patient breathing system must be connected directly to an over-pressure relief valve to prevent the possibility of barotrauma.
2
WARNINGS AND CAUTIONS
9. Always perform a pre-use check of the machine, including vaporizers, ventilator, circle absorber and monitors before clinical use. Follow the pre-use checklist (see section 5) as a minimum requirement. Many clinical accidents occur because of a failure to check for correct function.
10. The machine must not be used if any of the alarm, monitoring or protection system devices are not functioning correctly.
11. The machine must not be fitted with more than four operator accessible mains socket outlets. There is a risk of an excessive leakage current.
12. The use of antistatic or electrically conductive breathing hoses is not recommended when using high frequency electrical surgery equipment (e.g. : Diathermy). Burns may be caused.
14. Before any electrically powered machine is used clinically for the first time, check that the hospital engineering department has carried out an earth continuity test.
15. Before using any additional electrical equipment powered by the auxiliary sockets on the machine, check that the additional equipment is correctly wired and is earthed through its plug. A missing or defective protective earth conductor may increase earth leakage currents to the patient to values exceeding the allowable limits, resulting in ventricular fibrillation, or interference with the pumping action of the heart.
Take care when moving a fully loaded machine, particularly when negotiating ramps. Check that hoses or power leads are not trailing on the floor.
CAUTIONS
1. Flowmeter needle valves are designed to seal with light torque and may be damaged if tightened excessively. Take particular care with the carbon dioxide flowmeter control (if fitted); do not force the control knob past either the fully open or fully closed positions.
2. Open cylinder valves slowly to avoid damage to pressure reducing valves. Ensure that cylinder valves are at least one full turn open when in use.
3. Under no circumstances should anaesthetic agents be used for cleaning purposes.
4. After use, always disconnect the machine from the piped gas supply and/or close the gas cylinder valves.
5. Mechanical AHD system - The oxygen flow control is restricted to prevent the needle valve from fully closing. This ensures a minimum basal flow of oxygen. DO NOT attempt to close the flow to zero. Do not overtighten.
6. Compressed gas supplies must be clean and dry.
16. Additional equipment placed on the top shelf must be securely attached.
3
WARNINGS AND CAUTIONS- O2 Monitor
Oxygen Monitor
WARNINGS
1. We recommend calibration of the oxygen monitor every time the system is turned on, as a safety precaution.
2. Do not attempt to open the fuel cell. The sensor contains a small quantity of electrolyte, classified as a harmful irritant which is potentially hazardous.
3. ALWAYS check the integrity of the sensor assembly before use. See section 3.4.
4. Once exhausted, the sensor must be disposed of according to hospital, local, state and federal regulations.
5. The sensor measures oxygen partial pressure, and its output will rise and fall due to pressure change. An increase in pressure of 10% at he sensor inlet will produce a 10% increase in sensor output.
Using the oxygen monitor
6. When the battery voltage has fall­en to the minimum safe level, the oxygen monitor will automatically shut down to avoid permanent damage to the battery.
7. If the internal battery is fully dis­charged, the oxygen monitor will not function in the event of mains power failures. The battery must be recharged before the oxygen monitor is used clinically, otherwise back-up time can not be guaranteed.
See section 3.13.
CAUTIONS
1. Do not sterilise the oxygen sensor or control unit components. These components are not compatible with sterilisation techniques and damage may result.
2. Do not autoclave or expose the sensor to high temperatures.
3. If the sensor shows signs of being affected by condensation, dry the sensor with soft tissue. Do not use heat to dry the sensor.
NOTES
1. The O2 SENSOR FAULT alarm indicates that one of the following conditions has occurred. a) Internal electrical fault b) Software/electronics fault c) Oxygen sensor fault.
2. The concentration read-out may, in certain conditions of excess pressure, show a value above 100%. To accommodate these conditions it is possible to set the high alarm value up to 105% (see section 5).
3. To maintain maximum sensor life, always remove the unit from the
breathing circuit after use.
4
2. PURPOSE
The Prima anaesthesia workstation range is intended to provide controlled concentrations and flows of anaesthesia gases into a patient breathing system, from where the anaesthesia ventilator and breathing circuit will then deliver this fresh gas mixture to the patient Use in conjunction with anaesthetic vaporizers, breathing hoses and patient connection fittings which comply with the relevant ISO standard or equivalent.
Depending upon the patient circuit selected, the machines can be used in open, semi­open, semi-closed or closed circuit configurations.
The range has been designed to give a wide choice of configurations and accessories, as follows:
Machine size
Prima SP101 Induction machine with a single vaporizer on the backbar manifold, and up to three gas cylinders.
Anti-hypoxic Device (AHD)
Machines can be specified with a mechanical AHD system, designed to minimise the risk of a hypoxic mixture reaching the patient, see section 3.
Oxygen monitor (optional)
The Oxygen Monitor is intended to continu­ously measure and display the concentration of oxygen in breathing gas mixtures used in anaesthesia, and is intended for adult, pae­diatric and neonatal patients. The oxygen monitor is a module within an anaesthesia machine. The oxygen monitor is intended for use by health care providers, i.e. Physicians, Nurses and Technicians for use with patients during general anaesthesia.
Prima SP102 Medium width machine with capacity for two vaporizers on the backbar manifold, and up to four gas cylinders.
Gas supplies
SP101 Up to three gases
SP102 Up to four gases (oxygen,
nitrous oxide, carbon dioxide, and air, with pin­index cylinder yokes (see above), and provision for up to three pipeline supply inlets.
Vaporizer mounting systems
Back bar manifold for Selectatec Compatible, or Cagemount type vaporizers.
5
3. DESCRIPTION
3.1 Framework and General Construction
Frame
The machine has a cast base, extruded aluminium uprights, and aluminium panels.
Mobility
Trolley models have four castors, with a brake on each of the front castors. The castors are five inches diameter with anti­static rubber. A footrest is mounted on the front of the machine. To aid manoeuverability, two side handles are provided.
Mounting posts and brackets
A ‘T-slot’ mounting system is built into each side upright, to allow the use of pole-mount brackets, V-brackets, and ventilator
mounting brackets. The pole mount upright (see illustration) can be used to mount a complete AV-series Ventilator, or bellows unit only. V brackets can be used to mount a gas scavenging system, suction units, and accessories.
Draw units and work surfaces
The machine can be fitted with a base drawer unit (as illustrated) plus two additional smaller drawers. The work surface has has raised edges to prevent instruments, vials etc. from rolling off.
The Prima SP 102 illustrated is equipped with a standard full-width top shelf unit suitable for a large monitor, and a pull-out writing tablet mounted under the work surface. Options available include a CPU tray mounted above the drawer unit.
Prima SP 102 with standard top shelf and base drawer unit
Pole mount system fitted to frame upright
6
DESCRIPTION
3.2 Gas Circuit
Gas Supplies
For each size machine, a variety of cylinder and pipeline combinations can be added to the basic specification of oxygen and nitrous oxide cylinder and pipeline supply.
For example, the Prima SP102 can be sup­plied with an extra two gas cylinders (choose from one additional oxygen, one additional nitrous oxide, one carbon dioxide, one air), and one extra pipeline supply - Air.
Note a) Kits are available for fitment to
existing machines - see section 9 (Ordering Information).
b) Carbon dioxide is not available
on US specification machines.
Cylinder Yokes
The cylinder yokes are rear mounted and conform with ISO standards for pin-index fitting. To ensure that only cylinders of the appropriate gas may be installed the yokes are designed so that the retaining latch cannot be closed unless the index pins are
fully engaged.
Pipeline Inlets
Machines can be fitted with up to three pipeline gas inlets mounted on the rear of the machine.
Pipeline supply hoses are connected by non-interchangeable, threaded unions.
Filters
To prevent dirt entering the gas system, cylinder yokes and pipeline inlets are fitted
with filters.
pressure gauge tapping for direct mounting of a pressure gauge, and a non-return valve to prevent back flow of gas.
In addition, gas blocks for cylinder supplies have a diaphragm pressure regulator to reduce the pressure of the compressed gas supply, and a pressure relief valve, factory set to prevent any pressure build up under the diaphragm should any leakage develop across the reducing valve seat.
Secondary Pressure Regulator
For oxygen, nitrous oxide, and air, a second stage regulator reduces the pressure supplied to the flowmeter controls (see
section 4.4).
The fitment of a secondary regulator for oxygen and nitrous oxide enhances the performance of the mechanical AHD system fitted to some models.
Secondary regulation of the air supply is utilised to allow connection to high pressure air pipeline supplies.
Carbon Dioxide Flow Restrictor
On machines with a carbon dioxide supply, an integral, factory set, flow valve is fitted to restrict the flow of carbon dioxide to 500 ml/min.
3.3 Gas Circuit Schematics
Gas circuit schematics for: a) Non-AHD machines b) Mechanical AHD machines are shown on the following pages, and both show a four-gas machine. All available gas supply options are shown.
Gas Inlet Block
Each individual gas supply, from a cylinder or pipeline, is routed through a separate gas block.
Each gas block has an integral high
Note that carbon dioxide is not available for US specification machines.
7
Prima SP Non-AHD Machine
UK specification four-gas machine Note: US spec. machines are not fitted with CO
Air
N2O
facility
2
CO
O
2
2
Pneumatic pressure source
Filter
Pressure gauge
Pressure regulator
Pressure relief valve
Pneumatic on/off switch
Gas cut-off valve (normally open)
Gas cut-off valve (normally closed)
Gas cut-off valve (open/closed)
Reservoir
Audible alarm
Flow control valve (variable, but note that valve 1 is not user-adjustable)
1
Flowmeter
Vaporizer
Oxygen flush valve
Non-return valve
Power take-off point (or test point)
Gas reference connection
Visual indicator
8
Prima SP Mechanical AHD Machine
UK specification four-gas machine Note: US spec. machines are not fitted with CO
Air
N2O
facility
2
CO
O
2
1
2
Pneumatic pressure source
Filter
Pressure gauge
Gas cut-off valve (normally open)
Gas cut-off valve (normally closed)
Gas cut-off valve (open/closed)
Flowmeter
Vaporizer
Oxygen flush valve
Non-return valve
Pressure regulator
Pressure relief valve
Pneumatic on/off switch
Reservoir
Audible alarm
Flow control valve (variable, but note that valve 1 is not user-adjustable)
9
Power take-off point (or test point) Gas reference connection
Mechanical connection
Visual indicator
DESCRIPTION
3.4 Gas Supply Safety Devices
3.4.1 Gas Supply Cut-off Device
A gas cut-off device, triggered by low oxygen supply pressure, cuts the supply of nitrous oxide, and carbon dioxide (if fitted).
The cut-off operates when the oxygen pressure falls below 172 kPa (25 psi).
Gas supplies are reinstated only when the oxygen supply pressure rises above 227 kPa (33 psi).
3.4.2 Oxygen Supply Failure
Warning Whistle
A whistle gives an audible warning when there is a reduction of oxygen supply pressure. Operated solely by the remaining oxygen in the machine system, the warning whistle is prolonged by an oxygen reservoir built into the gas circuit, allowing a minimum warning whistle of 7 seconds duration.
The whistle will start to sound when the pressure falls to approximately 172 kPa (25 psi), and will continue to sound until the pressure falls to approximately 70 kPa (10 psi).
3.4.4 Air/N
(Machines with Air option)
The user can switch between Air and Nitrous Oxide (A).
NOTE a) The machine will NOT deliver a mixture of Air and nitrous oxide.
b) On machines with Mechanical AHD, the
2/N2O linkage
O continues to operate.
2O Interlock
B
A
3.4.5 Oxygen Supply Visual Indicator
The indicator (B) is operated from the oxygen supply and shows GREEN when the supply is at working pressure, and RED if the pressure falls.
3.4.6 CO2 Flow Restriction
The maximum flow of carbon dioxide is restricted to 500 ml/min by a pre-set flow control valve. This valve is not user-adjustable. Note that carbon dioxide is not available for US specification machines.
Oxygen consumption of the whistle is approximately 2 L/min when sounding and nil at other times.
3.4.3 Fresh Gas Pressure Relief Valve
A pressure relief valve is mounted between the vaporizer back bar and the common gas outlet (CGO) on the inside face of the machine right hand upright.
It is designed to prevent fresh gas being delivered to the breathing system at pressures exceeding 39 kPa (5.4 psi). This valve also protects machine components against excessive pressure in the event of a total blockage of the CGO.
3.4.7 Mechanical AHD
A mechanical link between the oxygen control valve and a needle valve in the nitrous oxide flow ensures that the machine delivers a fresh gas mixture with a minimum oxygen concentration of 30% ±3%, irrespective of the flow of nitrous oxide set by the anaesthetist. With the nitrous oxide control valve fully open, the oxygen and nitrous oxide flows are then both controlled by the oxygen control valve.
See section 3.5 for a full description
3.4.8 Low Pressure Gas Tubing
Diameter-indexed tubing is used for the low pressure gas system - see section 4.
10
DESCRIPTION
3.5 Mechanical AHD (Anti
Hypoxic Device)
3.5.1 Introduction
The Mechanical AHD is housed within the flowmeter module and comprises a gear linkage between the oxygen control valve and a needle valve in the nitrous oxide flow.
The system is designed to control the relative flow rates of oxygen and nitrous oxide. A predetermined minimum oxygen concentration of 30% ±3% in the oxygen / nitrous oxide mixture is maintained over the flow range to ensure that a hypoxic mixture is not supplied from the anaesthetic machine.
3.5.2 Gas Delivery Switch
The Gas Delivery Switch (A) operates on the oxygen supply and must be in the ‘On’ position for normal operation of the anaesthetic machine. The switch consequently controls the supply of all gases provided with a gas cut-off triggered by a predetermined pressure level within the oxygen supply (see section 3.4.1).
A whistle (oxygen failure warning whistle) will sound briefly whenever the gas delivery switch is turned on or off. Note that the whistle functions normally if the oxygen supply fails (see section 3.4.2).
The switch also controls the electrical supply to the optional Flowmeter Lighting unit and
2 Monitor (see section 3.13.2).
O
A
B
NOTE Machines with an Air supply option are fitted with an Air/N2O Interlock switch (B). The machine will NOT deliver a mixture of
Air and nitrous oxide - see section 3.4.4) . This switch must be in the N allow a flow of Nitrous Oxide.
The nitrous oxide control knob on the flowmeter module operates a secondary needle valve in the nitrous oxide flow. It is positioned downstream of the primary valve and therefore is used only to restrict the flow already set by the primary valve, which itself has been determined by the position of the oxygen control knob.
Therefore for any oxygen flow set by the user, the mixture delivered will still contain a minimum 30% ±3% oxygen even with the nitrous oxide control knob fully open. As the nitrous oxide knob is progressively closed, the oxygen content of the mixture increases to 100%.
2O position to
3.5.3 Gear Linkage and Nitrous Oxide Control Valves
A gear linkage connects the oxygen control knob on the flowmeter module and a needle valve in the nitrous oxide flow. This linkage limits the flow of nitrous oxide relative to the flow of oxygen set by the user.
Note that this needle valve acts as the primary nitrous oxide valve, and is actuated only by movement of the oxygen control.
3.5.4 Oxygen Basal Flow
To allow the system to function correctly, an oxygen basal flow is continuously supplied (see section 4.8). This basal flow can only be turned on and off by using the Gas Delivery Switch (A).
CAUTION The oxygen control is restricted to prevent the needle valve from fully closing. This ensures a minimum oxygen basal flow. DO NOT attempt to close the flow to zero. Do not overtighten the knob.
11
DESCRIPTION
3.6 Pressure Gauges
Pressure gauges (50 mm diameter) are located on the front panel below the flowmeter bank.
The gauges for the third and fourth gases (if fitted) are positioned between oxygen and nitrous oxide. Unused gauge positions are blanked out.
All pressure gauges are colour coded and labelled for the gases whose pressures they are indicating.
Cylinder contents are marked CYLINDER and pipeline pressure gauges are marked PIPELINE.
The gauges are calibrated in kPa x 100.
12
DESCRIPTION
3.7 Flowmeters and
Controls
3.7.1 All models
The flowmeters, mounted behind the perspex cover on the left hand side of the machine, are length-indexed to prevent inadvertent, incorrect installation.
All floats indicate flow rate in line with the upper surface as shown below.
Read flow at this level
Each flow control valve is positioned directly underneath the flow tube assembly to which it corresponds, and the control knob is colour-coded for the gas which it controls.
supply of oxygen and must be in the ON position for normal operation of the machine.
Flow control of each gas is achieved by a needle valve comprising a polished stainless steel needle mounted concentrically in a common manifold block. To minimise wear and material pick-up the needle seat is manufactured from silver. The flow control knob is turned counter-clockwise to increase the gas flow.
CAUTION
Needle valves are designed to seal with light torque and may be damaged if tightened excessively. DO NOT USE EXCESSIVE FORCE.
3.7.2 Optional Dual Cascade Flow Tubes
The flow of gas through dual cascade system flow tubes always flows through the low-flow tube first. The high-flow tube will not show any flow until more than 1 L/min is set.
At flows above 1 L/min, the high-flow tube reading indicates the rate of flow for that gas.
The oxygen flow control knob is made physically distinguishable from the other flow controls for identification by touch in accordance with ISO standards.
When fitted, air and carbon dioxide flowmeters are always installed in the inner positions on the flowmeter assembly. These positions are blanked out if air or carbon dioxide are not specified for the machine.
NOTE Machines with an Air supply option are fitted with an Air/N machine will NOT deliver a mixture of Air
and nitrous oxide - see section 3.4.4) . This switch must be in the N allow a flow of Nitrous Oxide.
The gas delivery switch, positioned to the right of the flowmeter bank, controls the
2O Interlock switch (The
2O position to
3.7.3 Carbon Dioxide Flow Restriction
The maximum flow of carbon dioxide (if fitted) is restricted to 500 ml/min.
13
DESCRIPTION
3.8 Vaporizers
CAUTION
Read the instruction manual supplied with the vaporizer before clinical use.
3.8.1 Vaporizer Mounting Systems
Vaporizers for the administration of volatile anaesthetic agents can be fitted to customer’s requirements as follows:
(a) Up to two Penlon Sigma Delta
Selectatec compatible vaporizers (interlock or non-interlock), mounted on a Selectatec compatible backbar.
(b) One or two Penlon Sigma Delta
Cagemount vaporizers mounted on a Modura rail (check that relevant national standards for your country allow fitment of more than one cagemount type vaporizer).
WARNING
Vaporizers must always be securely mounted, and never used free-standing. Unmounted vaporizers may be accidentally tipped resulting in uncalibrated and excessive volumes of liquid anaesthetic drug entering the breathing system.
Vaporizers of any description must not be installed or connected between the Common Gas Outlet (CGO) and the Breathing System, unless they are specifically designed for such use. (If this is done, the oxygen flush flow will pass through the vaporizer, and severe overdosage may result).
3.8.2 Selectatec Compatible Vaporizers
Selectatec compatible vaporizers, e.g. the Sigma Delta with the Selectatec compatibility block, may be mounted on a universal back bar manifold built onto the Prima range as an option.
Single and two-station manifolds are available, with each station fitted with two valve capsule assemblies for vaporizer
connector block attachment.
When a vaporizer is installed on a station the valves on that station open automatically to allow gas flow into and out of the vaporizer. Removal of the vaporizer from the station closes the valves on that station.
Selectatec Compatible vaporizer interlock systems are described in the literature
supplied with the vaporizer.
3.8.3 Cagemount Vaporizer
Vaporizers fitted with cagemount tapers have the male taper (inlet port) on the left, and the female taper on the right (viewing the front of the vaporizer).
It is recommended that detachable cagemount connectors are retained with a safety clip (catalogue number 52275) to prevent inadvertent disconnection.
14
DESCRIPTION
3.9 Common Gas Outlet (CGO) Block
The CGO block is mounted on the rail on the front of the machine, and can be moved along the rail. Slacken the securing screw under the block and carefully slide the block along the rail to the required position. Tighten the screw to hold the block in place.
The fresh gas outlet is located on the front face of the block, with 22 mm male taper and concentric 15 mm female taper. The male taper incorporates the Penlon Safelock system designed to prevent accidental disconnection of the breathing system.
A high mounting position for the CGO is available as an option for induction machines.
Oxygen Flush
An emergency oxygen flush valve button is mounted on the top front of the CGO block
and is marked ‘O Depressing the button provides a delivery of between 35-75 litres/min of oxygen into the common gas outlet. Releasing the button allows the spring­loaded valve to return to its normal position.
2 FLUSH’.
15
DESCRIPTION
3.10 Electrical Power Supply (if specified)
3.10.4 AV900 or AV800 Ventilator (if fitted) Power Supply
3.10.1 Mains Power Supply
Power is fed to the machine via the mains lead, to power an auxiliary output panel, and
optional flowmeter bank light.
NOTE
a) It is the user’s responsibility to ensure that the total sum of leakage currents from additional equipment plugged into the auxiliary sockets plus the leakage current from the machine does not exceed the values specified in any relevant national standards that may apply in the country where the machine is in use. b) Each socket is protected with a 5 Amp fuse.
3.10.2 Auxiliary Power Supply Sockets (if fitted)
An optional mains electricity auxiliary panel with three or four sockets can be specified, and fitted to the rear of the machine.
A
The mains lead for an AV-series ventilator can be plugged into one of the auxiliary power sockets on the rear of the machine.
Should the electrical power supply to the ventilator fail, the ventilator has a battery back up system to power the ventilator for 60 minutes, if the battery has been maintained in a fully charged condition.
Reference must be made to the separate user instruction manual supplied with the ventilator
Charging of the back-up battery takes place automatically when the ventilator mains lead is connected to a ‘live’ mains supply. The OFF indicator on the ventilator front control panel will show a yellow light during charging.
NOTE
The stated battery back-up period will only be available if the battery is kept fully charged. After the back-up power supply has been run down, the ventilator will not function until the battery is in a fully charged state. A fourteen hours recharge will be necessary to bring the battery to full charge.
The supply to the sockets is controlled by an ON/OFF switch (A).
3.10.3 Flowbank Lighting (optional)
The lighting system is controlled by the main ON/OFF switch (A).
3.10.5 Monitor and Other Accessories (if fitted)
The mains lead (or adaptor) for a monitor system or other accessories requiring an electrical supply can be plugged into one of the auxiliary sockets on the rear of the machine. See 3.13.17.
16
DESCRIPTION
3.11 Third/Fourth Gas Options
3.11.1 Air
When air is requested as the third/fourth gas, the machine specification is modified as stated in 4.10.1.
NOTE Machines with an Air supply option are fitted with an Air/N will NOT deliver a mixture of Air and nitrous oxide - see section 3.4.4). This switch must be in the N2O position to allow a flow of Nitrous Oxide.
3.11.2 Carbon Dioxide
When carbon dioxide is requested as the third/fourth gas, the machine specification is modified as stated in 4.10.2.
2O Interlock switch (The machine
Gas supply cut-off devices operate on carbon dioxide in all machines.
An integral, factory-set flow valve is fitted to restrict the flow of carbon dioxide to 500 ml/min.
Note that carbon dioxide is not available on US specification machines.
3.12 Auxiliary Gas Outlets
Oxygen
Auxiliary oxygen outlets are mounted on the right hand side frame upright.
Air
On machines with an air cylinder/pipeline supply, an auxiliary air outlet is fitted above the
oxygen outlets.
17
DESCRIPTION - O2 Monitor
3.13 Oxygen Monitor (Optional)
The oxygen monitor continuously measures and indicates the concentration of oxygen in the breathing system, and triggers an alarm when the concentration varies from the set levels.
CAUTION
If your machine is fitted with an AV900 Ventilator with a built-in Oxygen Monitor, please refer to the AV900 user manual for instructions on setting up and operation.
3.13.1 System Description
The Oxygen Monitor uses a fast-responding, oxygen­specific, self powered sensor that achieves 90% of final value in less than 10 seconds.
Sensor life:
approximately 1 500 000 O
20oC (minimum one year in most normal applications).
2 percent hours at
An external probe is supplied with a 2 m (6 ft) extendable cable and diverter fitting for a standard 15 mm Tee adaptor. The probe has a safety lock.
The system has user-adjustable high-level and low­level alarms with visual and audible indication of alarm conditions.
Easy-to-read, seven segment LED display for high­set, low-set, and oxygen concentration readings.
The monitor is controlled by the machine gas system master On/Off switch (A).
A back-up battery provides a minimum of 60 minutes operation in the event of mains failure.
The battery is charged when the machine is connected to the mains supply.
3.13.2 System On/Off Switch
The switch controls gas delivery from the anaesthetic machine, and electrical power to the oxygen monitor.
The switch must be in the On position to use the oxygen monitor and anaesthetic machine.
A
When switched to On, the monitor will always default to previous settings.
18
DESCRIPTION - O2 Monitor
5
3
2
10
9
1
6
7
8
4
Oxygen monitor control panel
1. O2 Concentration display 6. High set key
2. Low alarm set display 7. Low set key
3. High alarm set display 8. Calibrate key
4. Alarm LEDs 9. Low alarm set / Calibration control
5. Alarm mute key 10. High alarm set control
3.13.3 Displays
Oxygen Percentage Readout
The display provides direct readout of oxygen concentrations in the range of 0-100%. If the oxygen concentration exceeds 100%, the display will flash.
Low Alarm Set Readout
The indicated value represents the oxygen percentage at which the low alarm will be activated. The low alarm set value is limited within 18-99%. To set the low oxygen concentration alarm, see section 3.3.2.
High Alarm Set Readout
The indicated value represents the oxygen percentage at which the high alarm will be activated. The high alarm set value is limited within 19-105% (Note that in certain conditions of excess pressure, the readout may show a value above 100%.) To set the high oxygen concentration alarm, see section 3.3.1.
19
DESCRIPTION - O2 Monitor
3.13.4 Alarm Conditions
HIGH O2 ALARM
The high O oxygen concentration is 1% above the setting. In this alarm condition, a red HIGH O ALARM LED will flash at a 0.5 second rate, accompanied by a high priority sound. To cancel this alarm, the high alarm setting must be equal to, or above the oxygen concentration.
LOW O
The low alarm is triggered when the oxygen concentration is 1% below the setting. In this alarm condition, a red LOW O ALARM LED will flash at a 0.5 second rate, accompanied by a high priority sound. To cancel this alarm, the low alarm setting must be equal to, or below the oxygen concentration.
2 MONITOR INOP (inoperative)
O
This alarm indicates when the oxygen monitor is in malfunction condition. It could be triggered by electronic components failure or software malfunction. In this alarm condition, a red O ALARM LED will flash at a 0.5 second rate, accompanied by a high priority sound. If this mode occurs you can reset the system by pressing the ALARM MUTE and LOW ALARM SET buttons simultaneously for 3 seconds.
2 alarm is triggered when the
2 ALARM
2 MONITOR
will fall very quickly to zero within two to three weeks of normal usage. See section 5 for sensor replacement.
NOTE
always remove from breathing circuit after use.
2
LOW BATTERY
To maintain maximum sensor life,
The low battery alarm indicates that the battery is disconnected or the battery voltage has dropped below acceptable limits. If the monitor is in use under battery power, and the battery voltage is less than 11.5 volts, a low priority alarm is triggered, to indicate that the battery has less than 20
2
minutes life left. If the battery voltage falls to less than 10.8 volts a flashing medium priority alarm is triggered to indicate there is less than 5 minutes power left in the battery. To cancel this alarm, mains power must be On.
NOTE
Penlon Service Centre, or Penlon Customer Service Department in the UK.
At the end of the final 5 minute warning period, the oxygen monitor will shut down, to prevent damage to the battery.
If this condition persists, contact your
MAINS FAILURE
This alarm indicates mains power failure or cut-off. The yellow MAINS FAILURE LED will illuminate, and a low priority sound will be triggered.
2 SENSOR FAULT
O
The alarm is triggered a) when either the oxygen sensor is disconnected or approaching the end of its life. b) if the the oxygen concentration exceeds 110%. In the alarm condition, a red O
2 SENSOR
ALARM LED will flash at a 0.5 second rate, accompanied by a high priority sound. To cancel this alarm, check the sensor connection or replace the sensor.
2 SENSOR LOW
O
This alarm indicates the sensor has approached the end of its life. The yellow O
2 SENSOR LOW LED will light
up, and a low priority sound will be triggered. The sensor must be replaced as the output
3.13.5 Alarm Mute
In an alarm condition, pressing the ALARM MUTE button will deactivate the alarm sounder but the alarm LED will continue to flash. The yellow MUTE ALARM LED will illuminate, accompanied with a SINGLE ‘beep’ sound.
The alarm mute can not be operated: a) until the mute time is over, or the alarm condition has been rectified. b) when the oxygen concentration drops below 18%.
In high priority and medium alarm conditions the alarm mute deactivates the sounder for 30 seconds and 120 seconds respectively.
20
DESCRIPTION - O2 Monitor
3.13.6 The MOX-3 Oxygen Sensor
The MOX-3 oxygen sensor offers quick response, linear output over the entire 0-100% oxygen range, and long service life.
The MOX-3 is a self-powered galvanic cell that generates current, and the current is proportional to the rate of oxygen consumption.
The cell has a highly stable output over its operating life. Significant output loss is only shown at the very end of its life. Typical sensor drift rates are less than 1% per month when the sensor is exposed to gas in typical applications. Sensor lifetime is governed by the mass of lead available to react with the oxygen and its rate of consumption. High oxygen partial pressure and high temperature will increase the sensor output current, thus shortening the operation life. Nevertheless, average sensor life will still exceed one year. At the point where all lead has been consumed, the output will fall very quickly to zero over a period of two to three weeks.
NOTE To maintain maximum sensor life, always remove from breathing circuit after use.
Typical Sensor Installation on Machine CGO
The sensor assembly consists of an external probe and 2 m (6 ft) cable and diverter on a standard 15 mm Tee adaptor. On machines with an A100 Absorber, the probe can also be mounted on the dome of the inspiratory valve (see sec­tion 5).
21
DESCRIPTION - O2 Monitor
3.13.7 Power Supply
Mains Power Supply
Power is fed to the machine via the mains lead to a switching mode power supply, and the oxygen monitor will be powered up by 14.2 v.
The monitor is controlled by the machine gas system master On/Off switch (see 3.13.2).
Back-up Battery
Should the electrical power supply to the machine fail, the emergency battery supply for the unit comes into action automatically.
This is indicated by the illuminated yellow ‘MAINS FAILURE’ LED alarm, accompanied with a ‘beep’ sound.
The battery is maintained in a fully charged state during normal use (i.e. the machine connected to the mains power supply). A fully charged battery will power the unit for a minimum of 60 minutes.
Low Priority Battery Low Alarm
When the battery is discharged, and the mains power supply is not restored, the ‘BATTERY LOW’ LED alarm will illuminate, accompanied with a low prior­ity alarm sound.
Medium Priority Battery Low Alarm
When the battery is further discharged, and the mains power supply is not restored, the ‘BATTERY LOW’ LED will flash at a two second rate and a medium priority audible warning will be given when the minimum safe level of voltage is reached.
Low Battery Shut Down
WARNING When the battery voltage has fallen to the minimum safe level, the oxy­gen monitor will automatically shut down to avoid permanent damage to the battery.
Recharging the Battery
Charging of the back-up battery takes place automatically when the mains power supply is on, irrespective of the position of the machine gas system On/Off switch position.
NOTE The stated battery back-up period will only be available if the battery is kept fully charged. After the back-up power supply has been run down, the oxy­gen monitor will not function until the battery is in a fully charged state. An eight hours recharge will be necessary to bring the battery to full charge.
22
DESCRIPTION
3.14 A100 Absorber
In-board Mounting
The canister and valve block assemblies are mounted under the work surface and can be detached separately for cleaning (see section 7).
3.15 Prima SP Anaesthetic System - MRI Compatibility
The following Prima SP system components are MRI compatible:
Prima SP basic* machine *A 'basic' machine includes any variant of
Back Bar Flowmeter Bank Drawers Monitor Shelves Additional Gases
A100 Absorber (In-board and pole-mounted)
Nuffield 200 Ventilator
Delta Vaporizer
The following components are currently not MRI compatible:
Oxygen Monitor Flowmeter lighting Electrical power outlets AV-series ventilator
NOTE a) MRI Compatible Plastic Laryngoscopes -
see section 3.9 in the Price List.
b) The IDP Pressure Failure Alarm is MRI
compatible when used with appropriate bat­teries - see section 1.5.3 in the Price List.
23
4. SPECIFICATION
4.1 Physical Dimensions
Overall size: Height x Depth x Width (mm)
SP 101 1475 x 620 x 450 SP 102 1475 x 620 x 570
Work surface height: 890 mm Work surface size: 690/570/450 x 370 mm
Loading: 30 kg (66 lb) - evenly distributed.
Writing tablet: 300 x 220 mm
Loading: 10 kg (66 lb) - evenly distributed.
Top shelf: 690/570/450 x 370 mm
Loading: 30 kg (66 lb) - evenly distributed.
Mid shelf: 690/570/450 x 370 mm
Loading: 25 kg (55 lb) - evenly distributed
Drawers: 150 x 325 x 500 mm.
Loading: 10 kg (22 lb) evenly distributed
Castors: Front pair braked
Absorber posts: Swivels on a 254 mm radius around the left
hand side frame upright:
Height adjustment 300 to 760 mm Loading 30 kg (66 lb)
Ventilator bellows post Bushed to accept 25.4 mm (1 inch) or 22
mm (7/8 inch) poles.
Loading 30 kg (66 lb)
Gas scavenging fixing Dovetail clip on frame upright
Loading 30 kg (66 lb)
Common gas outlet: 22 mm male taper with coaxial 15 mm
female taper connections Safelock fitting
Weight:
SP 101 70 kg (154 lb) SP 102 75 kg (165 lb)
24
Loading...
+ 110 hidden pages