Penlon AV-S User manual

AV-S Ventilator
User Manual
Partnership for Life
www.ardusmedical.com
IMPORTANT
(i)
In order to ensure the full operational life of this ventilator, servicing by an engineer trained by the manufacturer should be undertaken periodically.
The ventilator must be serviced to the following schedule:
(a) Six monthly service - inspection and
function testing.
(b) Annual / two year / four year service -
inspection and function testing, and component replacement.
Details of these operations are given in the Service Manual for the AV-S, available only for engineers trained by the manufacturer.
For any enquiry regarding the servicing or repair of this product, contact Penlon Inc.
Penlon Inc. 11515 K-Tel Drive Minnetonka MN 55434
Always give as much of the following information as possible:
1. Type of equipment
2. Product name
3. Serial number
4. Approximate date of purchase
5. Apparent fault
This manual has been produced to provide authorised personnel with information on the function, routine performance and maintenance checks applicable to the AV-S Anaesthesia Ventilator.
Information contained in this manual is correct at the date of publication. The policy of the manufacturer is one of continued improvement to its products. Because of this policy, the manufacturer reserves the right to make any changes which may affect instructions in this manual, without giving prior notice.
Personnel must make themselves familiar with the contents of this manual and the machine’s function before using the apparatus.
FOREWORD
(ii)
The Importance of Patient Monitoring
WARNING Anaesthetic systems have the capability to deliver mixtures of gases and vapours to the patient which could cause injury or death unless controlled by a qualified anaesthetist.
There can be considerable variation in the effect of anaesthetic drugs on individual patients so that the setting and observation of control levels on the anaesthesia systems does not in itself ensure total patient safety. Anaesthesia system monitors and patient monitors are very desirable aids for the anaesthetist but are not true clinical monitors as the condition of the patient is also dependent on his respiration and the functioning of his cardio-vascular system.
IT IS ESSENTIAL THAT THESE ELEMENTS ARE MONITORED FREQUENTLY AND REGULARLY AND THAT ANY OBSERVATIONS ARE GIVEN PRECEDENCE OVER MACHINE CONTROL PARAMETERS IN JUDGING THE STATE OF A CLINICAL PROCEDURE.
Before using any monitoring system or device, the user must check that it conforms to the relevant standard, as listed in the table below.
Parameter / Device Relevant Standard
Pressure Measuring ISO 8835-2 Pressure Limitation Device EN 60601-2-13:2006 - 51.101.1 Exhaled Volume Monitor EN 60601-2-13:2006 - 51.101.4 Breathing System Integrity Alarm System EN 60601-2-13:2006 - 51.101.5 Continuing Pressure Alarm EN 60601-2-13:2006 - 51.101.6 Oxygen Monitor ISO 7767 Carbon Dioxide Monitor ISO 9918 Breathing Circuit ISO 8835-2 Agent Monitor ISO 11196 Gas Scavenging ISO 8835-3
For information on installing and connection of any of these systems or devices, please refer to the relevant manufacturer’s instructions.
Page No.
USER RESPONSIBILITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1. WARNINGS AND CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2. PURPOSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3. DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Ventilation cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Pneumatic system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4 Electrical system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . .. 14
3.5 Control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.5.1 Touchscreen operation and navigator wheel / push-button . . . . . . . . . . . . 15
3.5.2 Additional screen functions and displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.5.3 Start-up screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . 17
3.5.4 Selecting functions and parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5.5 User-adjustable parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5.6 Parameter display identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.6 Operational capability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.7 Output compensation functions . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . 20
3.7.1 Fresh gas compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.7.2 Fresh gas mixture compensation - models with spirometry . . . . . . . . . . . . 20
3.7.3 Compliance compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.7.4 Altitude compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.8 Interface with anaesthetic machine and A200SP absorber . . . . . . . . . . . . 22
3.9 Ventilation Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.9.1 Standby Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.9.2 Volume Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.9.3 Pressure Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.9.4 Spontaneous Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 28
3.9.5 Advanced Spontaneous Breathing Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.9.5.1 SIMV (Synchronised Intermittent Mandatory Ventilation) . . . . . . . . . . . . . . 30
3.9.5.2 SMMV (Synchronised Mandatory Minute Ventilation) . . . . . . . . . . . . . . . 31
3.9.5.3 PSV (Pressure Supported Ventilation) . . . . . . . . . . . . . . . . . .. . . . . . . 32
3.9.5.4 PEEP ( Positive End Expiratory Pressure) . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.10 On-screen Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.11 Spirometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.12 Display Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
3.13 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3.14 Oxygen Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
3.14.1 System operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
3.14.2 The Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
3.14.3 Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.14.4 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
3.14.5 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
3.14.6 Alarm mute. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . 40
4. SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Oxygen Monitor . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
CONTENTS
(iii)
5. PRE-OPERATION PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.1 Ventilator Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.1.1 Mounting the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.1.2 Electrical Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1.3 Ventilator gas supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1.4 Breathing system schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.1.5 Bellows drive gas connection . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.1.6 Anaesthetic gas scavenging system . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . 50
5.1.7 Remote screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.1.8 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.1.9 Breathing system connection and filters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.1.10 Spirometer connections and Zero Flow Calibration . . . . . . . . . . . . . . . . . . . . . 52
5.1.11 Pressure Monitor Connections . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . 54
5.1.12 Leak test / Compliance value calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.1.13 Bellows assembly - adult and paediatric . . . . . . . .. . . . . . . . . . . . . . . . . . . . 57
5.2 Start-up screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.3 Pre-use Checklists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.1 Daily Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.1.1 Alarm System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.1.2 Ventilator Internal Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.1.3 Function Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.2 Weekly Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.4 Oxygen Monitor Set-up . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . 61
5.4.1 Installation . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.4.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.4.3 Sensor Low Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.4.4 Setting the High and Low O
2 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
6. MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.1 Service Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.2.1 Outside Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.2.2 Bellows Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.2.3 Spirometer Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.2.4 Oxygen Monitor Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.2.5 Control Unit Patient Connector Block . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
6.3 Sterilisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
6.4 Oxygen Monitor Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
7. APPENDIX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
1. Back-up battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
2. Menu system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
3. Ventilator spirometry system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
4. Ventilator disposal after use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
5. Approved accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
(iv)
This anaesthesia ventilator has been built to conform with the specification and operating procedures stated in this manual and/or accompanying labels and notices when checked, assembled, operated, maintained and serviced in accordance with these instructions.
To ensure the safety of this device it must be checked and serviced to at least the minimum standards laid out in this manual. A defective, or suspected defective, product must not under any circumstances be used.
The user must accept responsibility for any malfunction which results from non­compliance with the servicing requirements detailed in this manual.
Additionally, the user must accept responsibility for any malfunction which may result from misuse of any kind or non­compliance with other requirements detailed in this manual.
Worn, broken, distorted, contaminated or missing components must be replaced immediately. Should such a repair become
necessary it is recommended that a request for service advice be made to Penlon Inc.
This device and any of its constituent parts must be repaired only in accordance with written instructions issued by the manufacturer and must not be altered or
modified in any way without the written approval of the manufacturer. The user of this equipment shall have the sole responsibility for any malfunction which results from improper use, maintenance, repair, damage or alteration by anyone other than the manufacturer or Penlon Inc.
USA and Canada: Federal Law restricts the sale and use of this device to, or on the order of, a licensed practitioner.
Statements in this manual preceded by the following words are of special significance:
WARNING means there is a
possibility of injury to the user or others.
CAUTION means there is a possibility
of damage to the apparatus
or other property.
NOTE indicates points of
particular interest for more efficient and convenient
operation.
Always take particular notice of the warnings, cautions and notes provided throughout this manual.
USER RESPONSIBILITY
1
1. WARNINGS AND CAUTIONS
The following WARNINGS and CAUTIONS must be read and understood before using this ventilator.
WARNINGS
General Information
1. Personnel must make themselves familiar with
the contents of this manual and the machine’s function before using the ventilator.
Before Using the Ventilator
2. Before the AV-S ventilator is used clinically for
the first time a Calibration Check and Output Check must be successfully completed.
Calibration and output checks must be carried out by a Penlon-trained technician, following the procedure in Appendix 6 in the AV-S Service Manual.
3. Before the ventilator is used clinically for the
first time, verify that the hospital engineering department has carried out an earth continuity test. If the integrity of the protective conductor is in doubt, the ventilator must not be used.
4. Excessive electronic noise caused by other
poorly regulated devices, such as an electrocautery unit, may adversely interfere with the proper functioning of the ventilator.
To avoid this problem, do not connect the ventilator’s power cord into the same electrical wall outlet or adaptor strip into which an electrocautery unit is connected.
5. If used with a mains extension cord, the unit
may be subject to electro-magnetic interference.
6. The driving gas supply must be clean and dry to
prevent ventilator malfunction.
7. This ventilator is designed to be driven by
oxygen or medical air only. The drive gas is set during manufacture and the ventilator is calibrated for that gas. Before the ventilator is used clinically for the first time, the commissioning engineer must confirm that the air/oxygen selection is set correctly for the drive gas that is to be used. The use of any other gas will cause inaccurate operation and may damage the ventilator, resulting in potential injury to the patient.
8. The driving gas is discharged through the opening in the back of the ventilator control unit. The discharged gas may contaminate the environment, and should therefore be extracted using a gas scavenging system.
9. The bellows can only support approximately 1 kPa (10 cmH
2O) differential positive pressure,
above which it may be dislodged from the mounting ring, resulting in dangerous malfunction of the ventilator.
Do not connect a positive end expiratory pressure (PEEP) valve or other restrictive device to the exhaust port on the bellows base.
This would increase the pressure inside the bellows and the bellows could detach from the base, causing serious malfunction.
10. Breathing System
The breathing system which conveys gases from the anaesthetic machine to the patient, and disposes of expired gases, must conform to the requirements of ISO 8835-2. Because breathing systems require frequent cleaning and disinfection they are not a permanent part of the anaesthetic ventilator and therefore cannot be directly under the control of the anaesthetic ventilator manufacturer. However, we strongly recommend that only breathing systems which have been approved and authorised by the manufacturer for use with AV-S should be employed.
Do not use conductive breathing system hoses.
When mechanical ventilation is employed the patient breathing system must be connected directly to a pressure relief valve to prevent the possibility of barotrauma.
11. The spirometer sensors are mounted within the A200SP absorber. Do not fit a spirometer sensor to any other location. The device will not measure exhaled volumes in any other position.
12. The operation of each alarm function should be verified daily.
Periodically check the alarms at clinically suitable intervals. If the audible alarm or the visual indicator of any alarm function fails to activate during any alarm condition or fails to reset after the alarm has been cleared, refer the unit to an authorised service technician.
2
13. Before using the ventilator check that all connections are correct, and verify that there are no leaks.
Patient circuit disconnects are a hazard to the patient. Extreme care should be taken to prevent such occurrences.
It is recommended that Safelock fittings are used throughout the breathing circuit.
14. Check that the cable between the control unit and remote display screen unit is connected before use. Always use a cable type recommended by the manufacturer.
Using the Ventilator
15. The AV-S ventilator is not intended for use in intensive care applications.
16. This apparatus must not be used with, or in close proximity to, flammable anaesthetic agents. There is a possible fire or explosion hazard.
17. Anaesthesia apparatus must be connected to an anaesthetic gas scavenging system (AGSS) to dispose of waste gas and prevent possible health hazards to operating room staff. This requirement must be observed during test procedures as well as during use with a patient.
The scavenging transfer and receiver system must conform to ISO 8835-3. Any problem arising from an improperly functioning scavenging system is solely the user’s responsibility. Do not use a scavenging system that restricts drive gas flow when negative pressure is exerted on it.
18. When the ventilator is connected to a patient, it is recommended that a qualified practitioner is in attendance at all times to react to an alarm or other indication of a problem.
19. In compliance with good anaesthesia practice, an alternative means of ventilation must be available whenever the ventilator is in use.
20. It is recommended that the patient oxygen concentration should be monitored continuously.
WARNINGS AND CAUTIONS
3
21. If the drive gas supply pressure drops below a nominal 241 kPa (35 psi), the LOW DRIVE GAS SUPPLY alarm will activate both audibly and visually. Patient minute volume may be reduced due to lowered flow rates
22. An audible alarm indicates an anomalous condition and should never go unheeded.
23. The characteristics of the breathing circuit connected between the ventilator and the patient can modify or change patient ventilation. To assist the maintenance of the delivered patient tidal volume, the ventilator control system software includes: A) a compliance compensation algorithm, B) a fresh gas compensation algorithm.
However, patient ventilation must be monitored independently from the ventilator. It is the responsibility of the user to monitor patient ventilation.
24. Care must be taken to ensure that the flow sensors are connected correctly to the inspiratory and expiratory ports of the absorber.
25. The Vent Inop (ventilator inoperative) alarm indicates that one of the following conditions has occurred:
a) The drive gas solenoid has failed.
b) The flow control valve has failed.
c) Internal electronic fault.
d) Internal electrical fault.
e) Software error. Note that if a ventilator error is detected,
‘Ventilator Inoperative’ will be displayed on the front control panel display.
26. The High and Low Airway Pressure Alarms are important for patient care. It is important that the sensor is properly located in the expiratory limb of the circuit ­refer to section 5.1.11.
27. The patient must be continuously attended and
monitored when Advanced Breathing Modes are in use.
User Maintenance
28. User maintenance is restricted to cleaning the outside surfaces of the ventilator, see section 6. Other procedures detailed in this manual must be carried out by trained technicians.
Service and repair operations must only be carried out by an engineer trained by the manufacturer. The warranty for this product is void if the product is not maintained in accordance with the service schedule detailed in section 6.1, and the procedures published in the Service Manual for this product.
Control Unit
29. Opening the control unit by unauthorised personnel automatically voids all warranties and specifications.
Prevention of tampering with the control unit is exclusively the user’s responsibility. If the control unit seal is broken, the manufacturer assumes no liability for any malfunction or failure of the ventilator.
30. For continued protection against fire hazards, any replacement fuses must be the identical type and rating as the original components. Replacement must be carried out by trained technician. See section 4 for fuse rating.
31. If the internal battery is fully discharged, the ventilator will not function in the event of mains power failure. The battery must be recharged before the ventilator is used clinically, otherwise backup cannot be guaranteed. See Appendix for battery maintenance. See also CAUTION No. 7.
Used or defective batteries must be disposed of according to hospital, local, state, and federal regulations.
32. No oil, grease or other flammable lubricant or sealant must be used on any part of the ventilator in close proximity to medical gas distribution components. There is a risk of fire or explosion.
33. Exterior panels must not be removed by unauthorised personnel and the apparatus must not be operated with such panels missing. There is a possible electric shock hazard.
Bellows Assembly
34. The valve seat on the patient gas exhalation diaphragm valve in the base of the bellows assembly must be cleaned regularly. Note that the bellows assembly is built into the A200SP Absorber - please refer to User Manual for this product. Failure to keep the valve seat clean could result in the diaphragm sticking, thus preventing exhalation.
Great care must be taken not to damage the precision surface of the valve seat on the patient gas exhalation diaphragm valve in the base of the bellows assembly.
Never use any hard object or abrasive detergent to clean the valve seat; use only a soft cloth. If the valve seat is damaged, the valve will leak and may cause serious ventilator malfunction.
WARNINGS AND CAUTIONS
4
CAUTIONS
1. Do not sterilise the ventilator control unit. The patient block assembly must be removed from the control unit before sterilisation ( see section
6.2.5). All other internal components are not compatible with sterilisation techniques and damage may result.
2. For ventilator components which require sterilisation, peak sterilisation temperatures should not exceed 134oC (275oF) to prevent possible damage. (See section 6).
3. Care must be taken not to let any liquid run into the control unit; serious damage may result.
4. The exhalation valve located in the bellows base assembly and the paediatric bellows adaptor must be cleaned and sterilised separately. Note that the bellows assembly is built into the A200SP Absorber
- please refer to User Manual for this product.
5. Always check for correct fitment, and carry out a full function test before clinical use, if the bellows has been removed and refitted for any reason. Note that the bellows assembly is built into the A200SP Absorber - please refer to User Manual for this product.
6. Always check for correct fitment, and carry out a full function test before clinical use, if the bellows has been removed and refitted for any reason. See section 6.
7. Damage will occur to the battery if it is allowed to remain in a discharged state. Check the battery frequently if the ventilator is in storage (see Appendix 1).
8. Fresh gas compensation is disabled if :
a) The spirometry system is turned OFF through the menu system, or b) The spirometry system is not functioning correctly.
9. Fresh gas mixture compensation is disabled if :
a) The spirometry system is turned OFF through the menu system, or b) The spirometry system is not functioning correctly.
c) The O2 monitor is switched OFF.
10. Circuit compliance is not activated until Fresh Gas Compensation is switched OFF.
NOTES
1. The term ‘cycle’ is used to designate the transition to the exhalation phase.
2. The term ‘trigger’ is used to indicate the transition to the inhalation phase.
WARNINGS AND CAUTIONS
5
Oxygen Monitor
Note that the sensor for the oxygen monitor is built into the A200SP Absorber - for additional information, please refer to the A200SP User Manual.
WARNINGS
1. We recommend a calibration check of the oxygen monitor every time the system is turned on, as a safety precaution.
2. Do not attempt to open the fuel cell. The sensor contains small quantities of : a) electrolyte, classified as a harmful irritant which is potentially hazardous, and b) lead.
Used or defective cells must be disposed of according to hospital, local, state, and federal regulations.
3. ALWAYS check the integrity of the sensor assembly before use.
4. Once exhausted, the sensor must be disposed of according to hospital, local, state and federal regulations.
5. The sensor measures oxygen partial pressure, and its output will rise and fall due to pressure change. An increase in pressure of 10% at the sensor inlet will produce a 10% increase in sensor output.
6. The oxygen sensor is not suitable for sterilisation. If contamination is suspected, fit a new sensor (see section 6.4) and dispose of the contaminated unit according to hospital, local, state and federal regulations.
6
CAUTIONS
1. Do not sterilise any oxygen monitor component.
2. Do not autoclave or expose the sensor to high temperatures.
3. If the sensor shows signs of being affected by condensation, dry the sensor with soft tissue. Do not use heat to dry the sensor.
NOTES
1. The O2 SENSOR FAULT alarm indicates that one of the following conditions has occurred. a) Internal electrical fault b) Software/electronics fault c) Oxygen sensor fault.
2. The concentration read-out may, in certain conditions of excess pressure, show a value above 100%. To accommodate these conditions it is possible to set the high alarm value up to 105% (see section 5).
3. To maintain maximum sensor life: i) always switch off the anaesthetic machine after use, to ensure that the basal flow ceases. ii) disconnect the breathing circuit after use.
4. The accuracy of flow and volume measurements may be reduced if the oxygen monitor is not in use.
5. Fresh gas mixture compensation is disabled if the
oxygen monitor is switched OFF.
WARNINGS AND CAUTIONS - Oxygen Monitor
The AV-S Ventilator is a pneumatically driven, software controlled, multi-mode ventilator, designed for mechanical ventilation of adult and paediatric patients under general anaesthesia. In addition, in spontaneous mode, it can be used to monitor spontaneously breathing patients It is designed for use in closed-circuit anaesthesia.
Indications for use of the device:
The AV-S Ventilator is intended to provide continuous mechanical ventilatory support during anaesthesia. The ventilator is a restricted medical device intended for use by qualified trained personnel under the direction of a physician. Specifically the ventilator is applicable for adult and paediatric patients. The ventilator is intended for use by health care providers, i.e. Physicians and Nurses with patients during general anaesthesia. The AV-S ventilator is not intended for use in intensive care applications.
Oxygen Monitor
The Oxygen Monitor is intended to continuously measure and display the concentration of oxygen in breathing gas mixtures used in anaesthesia, and is intended for adult and paediatric patients. The oxygen monitor is an integral part of the ventilator. The oxygen monitor is intended for use by health care providers, i.e. Physicians and Nurses for use with patients during general anaesthesia.
2. PURPOSE
7
3.1 General Description
The AV-S Ventilator is a pneumatically driven, software controlled, multi-mode ventilator.
The ventilator is time-cycled, volume/pressure controlled, and pressure limited.
The ventilator has compliance compensation and fresh gas compensation. User-selectable gas mixture compensation is a standard feature, plus a user-selectable variable inspiratory pause and sigh option.
Ventilation Modes
Volume Mode - continuous mandatory ventilation Pressure Mode - pressure controlled ventilation Spontaneous, with advanced patient support -
SIMV, SMMV, PSV, PEEP
Patient Monitoring
Airway pressure, measured from the expiratory limb of the breathing circuit.
Tidal Volume and Minute Volume measurement is provided by a dual spirometry system
An integral oxygen monitor system measures oxygen concentration in the breathing circuit inspiratory limb.
The print function provides a permanent record of function activity for up to eight hours during a procedure, or can be used to record waveforms.
Screen
210 mm (8.4 inch) high definition, colour TFT screen, with single/dual waveform display. Mounting: Remote, arm-mounted as illustrated (1) or optional combined control unit / screen (see section 5.1.1).
Bellows unit
The bellows unit (2) is built into the A200SP absorber. A paediatric bellows assembly is available as an option
Drive gas supply
The drive gas supply can be oxygen or air. The supply must be at 310 to 689 kPa (45 to 100 psi). Note that the drive gas is specified by the customer, and set during manufacture. Conversion from one drive gas to another must only be carried out by an authorised service engineer trained by the manufacturer.
3. DESCRIPTION
8
1
2
Spontaneous Mode Patient Support
SIMV - Synchronised Intermittent Mandatory Ventilation
SMMV - Synchronised Mandatory Minute Ventilation
PSV - Pressure Supported Ventilation
PEEP - Positive End Expiratory Pressure
9
Control Unit Rear Panel
Interface and Parameter inputs
5. A200SP Absorber Bag/Vent switch interface, and Spirometer connector
6. Anaesthetic machine interface connector - (primary on/off switch)
7. Pressure Monitor Port
8. Input socket - Oxygen monitor sensor
Data and Printer Ports
9. Data Output
10. Output to remote display
11. Ethernet
12. USB
13. VGA
14. Printer port
15. RS232 (manufacturer’s use only)
NOTE USB port is for access only by engineers trained by the manufacturer. All other data ports are read only. For further information, please contact your distributor’s service department, or the manufacturer.
Gas Connections
1. Ventilator drive gas inlet
- connect to anaesthetic machine
auxiliary gas outlet
2. Bellows Drive Gas Output
- connect to bellows via A200SP
absorber - see section 5.1.5
3. Outlet - Exhaust Valve
- connect to scavenge system - see section 5.1.6
Electrical Connection
4. Electrical mains input and fuse unit
DESCRIPTION
2
7
3
13 14
15
12
1110
9
8
6
5
4
1
3.2 Ventilation Cycle
This section provides a simplified description of the ventilation cycle.
1. Inspiratory Phase
The drive gas proportional valve (1) in the control unit opens. Drive gas is delivered to the bellows housing (2). The patient proportional valve (3) opens, and gas flows through the bleed valve. The back pressure ensures that the exhaust valve (4) is kept closed. Drive gas pressure builds up above the bellows (5), which starts to move down. The diaphragm (6) in the bellows assembly base is held closed, and patient gas is forced out of the bellows base (7) into the breathing system.
2. Beginning of Expiratory Phase
The drive gas proportional valve (1) closes. The patient proportional valve (3) closes. The exhaust valve (4) opens. Patient gas returns to the bellows (5). As the bellows rises, redundant drive gas is pushed out through the exhaust valve.
DESCRIPTION
10
1
4
4
6
7
5
5
2
3
3
1
3
1
4
DESCRIPTION
3. End of Expiratory Phase
With the bellows at the top of its housing fresh gas continues to flow. To prevent a high pressure build up the exhalation diaphragm (6) lifts and allows gas to exit through the exhaust valve (4).
4. PEEP Positive End Expiratory Pressure (user selectable)
The patient proportional valve (3) applies PEEP pressure plus 20 cmH2O to the exhaust valve, which remains closed at this stage. As fresh gas flows in the patient circuit, any pressure increase above PEEP pressure in the bellows (5) will cause gas to bleed past the exhaust valve (4). If there is a fall in pressure in the breathing circuit, the continuous flow from the drive gas proportional valve (1) helps maintain the set PEEP pressure.
11
6
4
5
5
12
DESCRIPTION
A
Pneumatic Flow
Diagram
C
1817
5
8
14
9
12
6
16
13
15
11
7
1
2
10
4
3
B
0 - 80 cmH2O
100 cmH2O
0 - 90 cmH2O
241 kPa (35 psi)
3 to 7 bar
3.3 Pneumatic System
Refer to the pneumatic system diagram on the previous page.
A) Gas inlet manifold block
The AV-S Ventilator is designed to operate on a 310 - 689 kPa (45 -100 psi) drive gas supply (oxygen or air - to customer’s requirement).
1. Drive Gas Inlet Connector
The gas source is connected to the DRIVE GAS SUPPLY fitting on the rear of the ventilator control unit. The gas supply should be capable of a flow rate of 80 L/min while maintaining a minimum pressure in excess of 310 kPa (45 psi).
2. Filter The drive gas is filtered with a 40-micron Input Gas Filter which protects the pneumatic components from incoming particulate matter.
3. The Low Supply Pressure Detector The pressure switch is set at a predetermined level to detect a loss or reduction of the input gas source pressure. When the pressure falls below 235 kPa (35 psi ± 1 psi), the LOW SUPPLY PRESSURE indicator will be displayed and the high priority audible alarm will activate.
4. Input Pressure Regulator Regulates the input drive gas to 260 kPa ± 21 kPa (38 psi ± 3 psi).
5. Cut-off Valve The valve isolates the gas supply : a) when the ventilator is switched off b) when a fault condition occurs.
6. Airway Pressure Sensor Connected to expiratory limb of breathing circuit.
B) Pneumatic Control Manifold Block
7. Drive Gas Proportional Valve
8. Drive Gas Flow Sensor
9. Drive Gas Pressure Sensor
10. Low Pressure Regulator
11. Patient Proportional Valve
12. PEEP pressure sensor
13. Restrictor The restrictor allows a flow of up to 2 L/min (<2 L/min bleeding)
C) Exhaust Manifold Block
14. Check Valve
15. Diaphragm Valve
16. Pressure Relief valve - Set to 100 cmH
2O
17. Exhaust Port ( to AGSS)
18. Bellows drive gas outlet (to bellows assembly)
DESCRIPTION
13
3.4 Electrical System
Mains Supply
The mains supply inlet is designed for connection to the following mains voltage supplies:
100 to 120 VAC, 50 to 60 Hz
200 to 240 VAC, 50 to 60 Hz
Note that the ventilator adjusts automatically to the supply voltage range.
The connector is a standard IEC type.
Back-up Battery
In the event of mains electrical failure, the back­up battery cuts in automatically.
Standard battery:
A fully charged battery will power the ventilator for approximately 30 minutes (depending on ventilator settings).
High-power battery (option):
A fully charged battery will power the ventilator for approximately one hour (depending on ventilator settings).
See Appendix 1 for battery care procedures.
DESCRIPTION
14
DESCRIPTION
3.5 Control Panel
3.5.1 Touchscreen and Navigator Wheel / Push Button
1. On/Off control Switch On: Short internal test sequence
Switch Off: Power down sequence with progress indicator
2. Status indicators for electrical power (mains/battery supply)
Yellow indicator - illuminated whenever power is applied to the unit and internal battery is being charged
Green indicator - illuminates when the unit is switched on
3. Menu switch The menu function provides access to user and service pages, including alarm settings.
4. Alarm mute switch 30 second or 120 second alarm silence, depending on alarm status. Note also that some alarms are not mutable (see 3.13).
5. Navigator Wheel and Press Button Turn the wheel to select a function or parameter, or to alter the value of an active parameter (see 3.5.4 and 3.5.5).
Press to confirm the setting.
6. Active Tabs Touch the screen at the appropriate tab area to activate the required function/parameter (see 3.5.4 and 3.5.5).
15
VT SETmLTPS INT
cmH2O
SET
BPM
SET
I:E
PEEP
cmH2O
LIMIT
cmH2O
VOLUME
SPONT
STANDBY
PSV
SIMV
SMMV
PRESS
AV-S
1
4
3
5
6
2
.
.
IO
o
o
16
DESCRIPTION
VT SETmLTPS INT
cmH2O
SET
BPM
SET
I:E
PEEP
cmH2O
LIMIT
cmH2O
VOLUME
SPONT
STANDBY
PSV
SIMV
SMMV
PRESS
AV-S
.
.
IO
o
o
3.5.2 Additional Screen Functions and Displays
Area A Menu and sub-menu window
Waveform display, plus waveform pause and print symbols
Area B Alarm values window
Area C Gas mixture values window
Oxygen monitor values window
Area D Additional breathing mode information symbols:
(1) Adult mode
(2) Paediatric
mode
(3) Ventilator
mode
(4) Bag mode
(5) Sigh
(6) Inspiratory
pause
Area B
Area C
Area D
Area A
17
DESCRIPTION
3.5.3 Start-up Screens
1. Start-up
At start-up, the introduction screen allows the user to select one of three default settings:
ADULT DEFAULTS
PAEDIATRIC DEFAULTS
SITE DEFAULTS
NOTE a) The user must select one of the above default groups before the ventilator will switch to standby in that default mode
b) SITE DEFAULT is editable in standby mode (see below)
c) Settings can be saved via the service menu to create a new site default
2. Default Settings
Selection
The user can select ADULT, or PAEDIATRIC, or SITE, and view the default parameter settings.
The options will remain, even after the ventilator is turned off.
Site Default Settings
(typical values shown)
Adjust the parameter values from within the Service menu (SITE DEFAULTS)
Press to confirm the new settings for site defaults.
3. Screen Calibration
Adjust the brightness of the screen back lighting.
<Paediatric Defaults< VT Set : 150 mL VM Set : 2.2 Litres T+PS INT : 10 cmH
2O
Set BPM : 15 I : E : 1: 2.0 PEEP : OFF Limit : 38 cmH
2O
Trigger : 1.0 L/min Apnoea Alarm Limit : 15 secs Volume Type: Tidal
CALIBRATE TOUCH SCREEN
SITE DEFAULTS
VIEW
ADULT DEFAULTS
VIEW
PAEDIATRIC DEFAULTS
PRESS TO CONFIRM
<Site Defaults< VT Set : 800 mL VM Set : 6.0 Litres T+PS INT : 10 cmH2O Set BPM : 10 I : E : 1: 2.0 PEEP : OFF Limit : 38 cmH
2O
Trigger : 1.0 L/min Apnoea Alarm Limit : 15 secs Volume Type: Tidal
CALIBRATE TOUCH SCREEN
SITE DEFAULTS
PRESS TO CONFIRM
ADULT DEFAULTS
VIEW
PAEDIATRIC DEFAULTS
VIEW
<Adult Defaults< VT Set : 600 mL VM Set : 6.0 Litres T+PS INT : 10 cmH
2O
Set BPM : 10 I : E : 1: 2.0 PEEP : OFF Limit : 38 cmH
2O
Trigger : 1.0 L/min Apnoea Alarm Limit : 15 secs Volume Type: Tidal
CALIBRATE TOUCH SCREEN
SITE DEFAULTS
VIEW
ADULT DEFAULTS
PRESS TO CONFIRM
PAEDIATRIC DEFAULTS
VIEW
18
DESCRIPTION
3.5.4 Selecting Functions and Parameters
The functions/parameters shown on the screen can be activated as follows:
a) touch the screen at the appropriate tab area, or
b) rotate the navigator wheel and press it when the indicator arrow
is on the required parameter tab
Note that unless Site Defaults are selected (see above) parameters default to factory-set values for Adult or Paediatric patients when the ventilator is switched on from OFF, and no further user selection is made.
3.5.5 User Adjustable Parameters
Variable parameters can be altered by rotating the navigator wheel. When the required value is displayed, press the active tab, or the wheel, to confirm the setting.
Tidal Volume Range 20-1600 ml
Rate 4-100 bpm
I:E Ratio 1:0.3 to 1:8
PEEP 4-20 cmH2O
Can be set to OFF
Pressure Limit
Volume mode: 10-80 cmH2O Pressure mode: 10-50 cmH2O
Alarm limits (user adjustable alarms only - see 3.13)
3.5.6 Parameter Display Identification
1. Active Parameters
Active parameters that can be set for use in the current mode are displayed as:
White Text on Blue
2. Inactive Parameters
Inactive parameters that can be set for any non-current mode are displayed as:
White Text on Blue Label White values on Black
3. Measured Parameters
Yellow values on Black
4. T+PS INIT (target and pressure support initial value)
The initial pressure value can be changed so that when entering either PRESSURE or PSV modes the TARGET value or PSUPP value are pre-selected.
Note that changing either of these limits in their active modes will maintain the value when changing between PSV, PRESSURE, and STANDBY modes.
Loading...
+ 56 hidden pages