Penlon AV-S Service manual (2005)

AV -S Ventilator
Remote Display Module and Interface for use with A200SPAbsorber
Service Manual
Partnership for Life
Draft Copy
July 2005
THE IMPORTANCE OF PATIENT MONITORING
WARNING
Anaesthetic systems have the capability to deliver mixtures of gases and vapours to the patient which could cause injury or death unless controlled by a qualified anaesthetist.
There can be considerable variation in the effect of anaesthetic drugs on individual patients so that the setting and observation of control levels on the anaesthesia systems does not in itself ensure total patient safety.
Anaesthesia system monitors and patient monitors are very desirable aids for the anaesthetist but are not true clinical monitors as the condition of the patient is also dependent on his respiration and the
functioning of his cardio-vascular system.
IT IS ESSENTIAL THAT THESE ELEMENTS ARE MONITORED FREQUENTLY AND REGULARLY AND THAT ANY OBSERVATIONS ARE GIVEN PRECEDENCE OVER MACHINE CONTROL P ARAMETERS IN JUDGING THE STATE OF
A CLINICAL PROCEDURE.
Servicing and Repairs
In order to ensure the full operational life of this ventilator, servicing by a Penlon-trained engineer should be undertaken periodically.
The ventilator must be serviced to the following schedule:
(a) Six monthly service - inspection and
function testing.
(b) Annual / 2 Year / four year service
including component replacement.
Details of these operations are given in this Manual for the AV-S, available only for Penlon trained engineers.
For any enquiry regarding the servicing or repair of this product, contact the nearest accredited Penlon agent:
or communicate directly with: Technical Support
Penlon Limited Abingdon Science Park Barton Lane Abingdon OX14 3PH UK
Tel: 44 (0) 1235 547076 Fax: 44 (0) 1235 547062
E-mail: technicalsupport@penlon.co.uk
Always give as much of the following information as possible:
1. Type of equipment
2. Product name
3. Serial number
4. Approximate date of purchase
5. Apparent fault
IMPORTANT
(i)
This manual has been produced to provide authorised personnel with information on the function, routine performance, service and maintenance applicable to the AV-S Anaesthesia Ventilator.
Information contained in this manual is correct at the date of publication. The policy of Penlon Limited is one of continued improvement to its products. Because of this policy, Penlon Limited reserves the right to make any changes which may affect instructions in this manual, without giving prior notice.
Personnel must make themselves familiar with the contents of this manual and the machine’s function before servicing the apparatus.
Copyright © Penlon Limited, 2005. All rights reserved.
FOREWORD
(ii)
Page No.
USER RESPONSIBILITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1. WARNINGS AND CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2. PURPOSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3. DESCRIPTION
3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Ventilation Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.3.1 System Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.4 Electrical System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.5 Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.5.1 Touchscreen Operation and Navigator wheel / push-button . . . . . . . . . . . . 16
3.5.2 User Adjustable Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.5.3 Operational capability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.5.4 Output Compensation Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.6 Interface with Prima SP and A200SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.7 Ventilation Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.7.1 Standby Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.7.2 Volume Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.7.3 Pressure Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.7.4 Spontaneous Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.7.5 Advanced Spontaneous Breathing Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.7.5.1 SIMV (Synchronised Intermittent Mandatory Ventilation) . . . . . . . . . . . . . 24
3.7.5.2 SMMV (Synchronised Mandatory Minute Ventilation) . . . . . . . . . . . . 25
3.7.5.3 PSV (Pressure Supported Ventilation) . . . . . . . . . . . . . . . . . . . . . . 26
3.7.5.4 PEEP ( Positive End Expiratory Pressure) . . . . . . . . . . . . . . . . . . . . . . 27
3.8 On-screen Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.9 Spirometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.10 Display Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.11 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.12 Oxygen Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.13.1 System Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.12.2 The MOX-3 Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.12.3 Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.12.4 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.12.5 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.12.6 Alarm Mute. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4. SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Oxygen Monitor . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
CONTENTS
(iii)
CONTENTS
(iv)
5. PRE-OPERATION PROCEDURES
5.1 Ventilator Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.1.1 Mounting the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.1.2 Electrical Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.1.3 Ventilator Gas Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.1.4 Breathing System Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.1.5 Bellows Drive Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.1.6 Anaesthetic Gas Scavenging System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.1.7 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.1.8 Breathing System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.1.9 Spirometer Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.1.10 Pressure Monitor Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.1.11 Bellows Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.2 Pre-use Checklists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.2.1 Daily Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.2.2 Function Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.2.3 Weekly Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.3 Oxygen Monitor Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.3.1 Installation . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.3.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.3.3 Sensor Low Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5.3.4 Setting the High and Low O
2 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6. SERVICE PROCEDURES
6.1 Service Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6.2 Control Unit Patient Block Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7. SERVICE SCHEDULE
Service Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8. PARTS LISTS
Preventive Maintenance Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Assemblies 65
9. APPENDIX
1. Back-up Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
2. Menu System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
3. Ventilator Spirometry System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
4. Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Outside Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Bellows and Diaphragm Exhalation Valve . . . . . . . . . . . . . . . . . . . . . 86
Spirometer Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Patient Connector Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Sterilisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Oxygen Monitor - Cleaning and Sterilisation . . . . . . . . . . . . . . . . . . 88
Oxygen Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
This anaesthesia ventilator has been built to conform with the specification and operating procedures stated in this manual and/or accompanying labels and notices when checked, assembled, operated, maintained and serviced in accordance with these instructions.
To ensure the safety of this device it must be checked and serviced to at least the minimum standards laid out in this manual. A defective, or suspected defective, product must not under any circumstances be used.
The user must accept responsibility for any malfunction which results from non­compliance with the servicing requirements detailed in this manual.
Additionally, the user must accept responsibility for any malfunction which may result from misuse of any kind or non­compliance with other requirements detailed in this manual.
Worn, broken, distorted, contaminated or missing components must be replaced immediately. Should such a repair become
necessary it is recommended that a request for service advice be made to the nearest Penlon accredited agent.
This device and any of its constituent parts must be repaired only in accordance with written instructions issued by Penlon Limited and must not be altered or modified
in any way without the written approval of Penlon Limited. The user of this equipment shall have the sole responsibility for any malfunction which results from improper use, maintenance, repair, damage or alteration by anyone other than Penlon or its appointed agents.
USAand Canadian Federal Law restrict s the sale and use of this device to, or on the order of, a licensed practitioner.
Statements in this manual preceded by the following words are of special significance:
WARNING means there is a
possibility of injury to the user or others.
CAUTION means there is a possibility
of damage to the apparatus
or other property.
NOTE indicates points of
particular interest for more efficient and convenient
operation.
Always take particular notice of the warnings, cautions and notes provided throughout this manual.
USER RESPONSIBILITY
1
1. WARNINGS AND CAUTIONS
The following WARNINGS and CAUTIONS must be read and understood before using this ventilator.
WARNINGS
General Information
1. Personnel must make themselves
familiar with the contents of this manual and the machine’s function before using the ventilator.
Before Using the Ventilator
2. Before the ventilator is used clinically
for the first time, verify that the hospital engineering department has carried out an earth continuity test.
3. Excessive electronic noise caused by
other poorly regulated devices, such as an electrocautery unit, may adversely interfere with the proper functioning of the ventilator.
To avoid this problem, do not connect the ventilator’s power cord into the same electrical wall outlet or adaptor strip into which an electrocautery unit is connected.
4. If used with a mains extension cord, the
unit may be subject to electro-magnetic interference.
5. The driving gas supply must be clean
and dry to prevent ventilator malfunction.
6. This ventilator is designed to be driven
by oxygen or medical air only. It is calibrated during manufacture for use with either gas. Before the ventilator is used clinically for the first time, the commissioning engineer must confirm that the internal Air/Oxygen switch is set correctly for the gas that is to be used. The use of any other gas will cause inaccurate operation and may damage the ventilator, resulting in potential injury to the patient.
7. The driving gas is discharged through the opening in the back of the ventilator control unit. The discharged gas may contaminate the environment, and should therefore be extracted using a gas scavenging system.
8. The bellows can only support approximately 1 kPa (10 cmH
2
O) differential positive pressure, above which it may be dislodged from the mounting ring, resulting in dangerous malfunction of the ventilator.
Do not connect a positive end expiratory pressure (PEEP) valve or other restrictive device to the exhaust port on the bellows base.
This would increase the pressure inside the bellows and the bellows could detach from the base, causing serious malfunction.
9. Breathing System The breathing system which conveys
gases from the anaesthetic machine to the patient, and disposes of expired gases, must conform to the requirements of ISO 8835-2. Because breathing systems require frequent cleaning and disinfection they are not a permanent part of the anaesthetic ventilator and therefore cannot be directly under the control of the anaesthetic ventilator manufacturer. However, we strongly recommend that only breathing systems which have been approved and authorised by Penlon for use with AV-S should be employed.
Do not use conductive breathing system hoses.
When mechanical ventilation is employed the patient breathing system must be connected directly to a pressure relief valve to prevent the possibility of barotrauma.
10. Do not connect a spirometer to the exhaust port on the bellows base. The device will not measure exhaled volumes in that position.
2
11. The operation of each alarm function should be verified daily.
Periodically check the alarms at clinically suitable intervals. If the audible alarm or the visual indicator of any alarm function fails to activate during any alarm condition or fails to reset after the alarm has been cleared, refer the unit to an authorised service technician.
12. Before using the ventilator check that all connections are correct, and verify that there are no leaks.
Patient circuit disconnects are a hazard to the patient. Extreme care should be taken to prevent such occurrences.
It is recommended that Penlon Safelock fittings are used throughout the breathing circuit.
Using the Ventilator
13. The AV-S ventilator is not intended for use in intensive care applications.
14. This apparatus must not be used with, or in close proximity to, flammable anaesthetic agents. There is a possible fire or explosion hazard.
15. Anaesthesia apparatus must be connected to an anaesthetic gas scavenging system (AGSS) to dispose of waste gas and prevent possible health hazards to operating room staff. This requirement must be observed during test procedures as well as during use with a patient. The scavenging transfer and receiver system must conform to ISO 8835-3. Any problem arising from an improperly functioning scavenging system is solely the user’s responsibility. Do not use a scavenging system that restricts drive gas flow when negative pressure is exerted on it.
16. When the ventilator is connected to a patient, it is recommended that a qualified practitioner is in attendance at all times to react to an alarm or other indication of a problem.
17. In compliance with good anaesthesia practice, an alternative means of ventilation must be available whenever the ventilator is in use.
18. It is recommended that the patient oxygen concentration should be monitored continuously.
19. If the drive gas supply pressure drops below a nominal 241 kPa (35 psi), the LOW DRIVE GAS SUPPLY alarm will activate both audibly and visually. Patient minute volume may be reduced due to lowered flow rates
20. An audible alarm indicates an anomalous condition and should never go unheeded.
21. The characteristics of the breathing circuit connected between the ventilator and the patient can modify or change patient ventilation. To assist the maintenance of the delivered patient tidal volume, the ventilator control system software includes: A) a compliance compensation algorithm, B) a fresh gas compensation algorithm.
However, patient ventilation must be monitored independently from the ventilator. It is the responsibility of the user to monitor patient ventilation.
22. Care must be taken to ensure that the flow sensors are connected correctly to the inspiratory and expiratory ports of the absorber.
23. The Vent Inop (ventilator inoperative) alarm indicates that one of the following conditions has occurred:
a) The drive gas solenoid has failed. b) The flow control valve has failed. c) Internal electronic fault. d) Internal electrical fault. e) Software error.
Note that if a ventilator error is detected, ‘Ventilator Inoperative’will be displayed on the front control panel display.
WARNINGS AND CAUTIONS
3
24. The High and Low Airway Pressure Alarms are important for patient care. It is important that the sensor is properly located in the expiratory limb of the circuit - refer to section 5.1.10.
25. The patient must be continuously
attended and monitored when Advanced Breathing Modes are in use.
User Maintenance
Control Unit
26. Opening the control unit by unauthorised personnel automatically voids all warranties and specifications.
Prevention of tampering with the control unit is exclusively the user’s responsibility. If the control unit seal is broken, the manufacturer assumes no liability for any malfunction or failure of the ventilator.
27. For continued protection against fire hazards, replace the two fuses only with the identical type and rating of fuse. See section 4 for fuse rating.
28. If the internal battery is fully discharged, the ventilator will not function in the event of mains power failure. The battery must be recharged before the ventilator is used clinically, otherwise backup cannot be guaranteed. See Appendix for battery maintenance. See also CAUTION No. 7.
Used or defective batteries must be disposed of according to hospital, local, state, and federal regulations.
29. No oil, grease or other flammable lubricant or sealant must be used on any part of the ventilator in close proximity to medical gas distribution components. There is a risk of fire or explosion.
30. Exterior panels must not be removed by unauthorised personnel and the apparatus must not be operated with such panels missing. There is a possible electric shock hazard.
31. Check that the cable between the
control unit and remote display screen unit is connected before use. Always use a cable type recommended by the manufacturer.
Bellows Assembly
32. The valve seat on the patient gas exhalation diaphragm valve in the base of the bellows assembly must be cleaned regularly - see section 6.2. Failure to keep the valve seat clean could result in the diaphragm sticking, thus preventing exhalation.
Great care must be taken not to damage the precision surface of the valve seat on the patient gas exhalation diaphragm valve in the base of the bellows assembly.
Never use any hard object or abrasive detergent to clean it; use only a soft cloth. If the valve seat is damaged, the valve will leak and may cause serious ventilator malfunction.
WARNINGS AND CAUTIONS
4
CAUTIONS
1. Do not sterilise the ventilator control unit. The patient block assembly must be removed from the control unit before sterilisation ( see section 6.2.4). All other internal components are not compatible with sterilisation techniques and damage may result.
2. For ventilator components which require sterilisation, peak sterilisation temperatures should not exceed 136oC (275oF) to prevent possible damage. (See section 6).
3. Those parts suitable for ethylene oxide sterilisation should, following sterilisation, be quarantined in a well ventilated area to allow dissipation of residual gas absorbed by the components. Follow the steriliser manufacturer’s recommendations for any special aeration periods required.
4. The exhalation valve located in the bellows base assembly and the paediatric bellows adaptor must be cleaned and sterilised separately. See section 6.
5. Care must be taken not to let any liquid run into the control unit; serious damage may result.
6. Always check for correct fitment, and carry out a full function test before clinical use, if the bellows has been removed and refitted for any reason. See section 6.
7. Damage may occur to the battery if it is allowed to remain in a discharged state. Check the battery frequently if the ventilator is in storage (see Appendix 1).
8. Fresh gas compensation is disabled if :
a) The spirometry system is turned OFF through the menu system, or b) The spirometry system is not functioning correctly.
9. Fresh gas mixture compensation is disabled
if : a) The spirometry system is turned OFF through the menu system, or b) The spirometry system is not functioning correctly.
c) The O2 monitor is switched OFF.
10. Circuit compliance is not activated until Fresh Gas Compensation is switched OFF.
NOTES
1. The term ‘cycle’ is used to designate the transition to the exhalation phase.
2. The term ‘trigger’ is used to indicate the transition to the inhalation phase.
WARNINGS AND CAUTIONS
5
Oxygen Monitor
WARNINGS
1. We recommend calibration of the oxygen monitor every time the system is turned on, as a safety precaution.
2. Do not attempt to open the fuel cell. The sensor contains small quantities of : a) electrolyte, classified as a harmful irritant which is potentially hazardous, and b) lead.
Used or defective cells must be disposed of according to hospital, local, state, and federal regulations.
3. ALWAYS check the integrity of the sensor assembly before use.
4. Once exhausted, the sensor must be disposed of according to hospital, local, state and federal regulations.
5. The sensor measures oxygen partial pressure, and its output will rise and fall due to pressure change. An increase in pressure of 10% at the sensor inlet will produce a 10% increase in sensor output.
CAUTIONS
1. Only use low temperature ethylene oxide sterilisation for the oxygen sensor. The sensor is not compatible with other sterilisation techniques - damage may result. Do not sterilise any other components.
2. Do not autoclave or expose the sensor to high temperatures.
3. If the sensor shows signs of being affected by condensation, dry the sensor with soft tissue. Do not use heat to dry the sensor.
6
NOTES
1. The O2 SENSOR FAULT alarm indicates that one of the following conditions has occurred. a) Internal electrical fault b) Software/electronics fault c) Oxygen sensor fault.
2. The concentration read-out may, in certain conditions of excess pressure, show a value above 100%. To accommodate these conditions it is possible to set the high alarm value up to 105% (see section 5).
3. To maintain maximum sensor life: i) always switch off the anaesthetic machine after use, to ensure that the basal flow ceases. ii) disconnect the breathing circuit after use.
4. The accuracy of flow and volume measurements may be reduced if the oxygen monitor is not in use.
5. Fresh gas mixture compensation is disabled
if the oxygen monitor is switched OFF.
WARNINGS AND CAUTIONS - Oxygen Monitor
The AV-S Ventilator is a software controlled, multi-mode ventilator, designed for mechanical ventilation of adult and paediatric patients under general anaesthesia. In addition, in spontaneous mode, it can be used to monitor spontaneously breathing patients It is designed for use in closed-circuit anaesthesia and also to drive a Mapleson D circuit.
Indications for use of the device:
The AV-S Ventilator is intended to provide continuous mechanical ventilatory support during anaesthesia. The ventilator is a restricted medical device intended for use by qualified trained personnel under the direction of a physician. Specifically the ventilator is applicable for adult and paediatric patients. The ventilator is intended for use by health care providers, i.e. Physicians, Nurses and Technicians with patients during general anaesthesia. The AV-S ventilator is not intended for use in intensive care applications.
Oxygen Monitor
The Oxygen Monitor is intended to continuously measure and display the concentration of oxygen in breathing gas mixtures used in anaesthesia, and is intended for adult and paediatric patients. The oxygen monitor is an integral part of the ventilator. The oxygen monitor is intended for use by health care providers, i.e. Physicians, Nurses and Technicians for use with patients during general anaesthesia.
2. PURPOSE
7
3.1 General Description
The AV-S Ventilator is a pneumatically driven, software controlled, multi-mode ventilator.
The ventilator is a time-cycled, volume/pressure controlled, and pressure limited.
The ventilator has compliance compensation and a user selectable option of an inspiratory pause fixed at 25% of the inspiratory time. In addition, fresh gas compensation and user selectable gas mixture compensation is a standard feature.
Ventilation Modes
Volume Mode - continuous mandatory ventilation Pressure Mode - pressure controlled ventilation Spontaneous, with advanced features ­SIMV, SMMV, PSV PEEP
Patient Monitoring
Airway pressure, measured from the expiratory limb of the breathing circuit.
Tidal volume and Minute Volume measurement is provided by a dual spirometry system
An integral oxygen monitor system measures oxygen concentration in the breathing circuit inspiratory limb.
The print function provides a permanent record of function activity for up to eight hours during a procedure, or can be used to record waveforms.
Screen
210 mm (8.4 inch) high definition, colour TFT screen, with single/dual waveform display.
3. DESCRIPTION
8
AV-S Ventilator
Bellows unit
The bellows unit (1) is built into the A200SP absorber. A paediatric bellows assembly is available as an option
Mounting options
The AV-S integral screen and control unit can be mounted securely on the anaesthetic machine shelf or side bracket.
Drive gas supply
The supply must be at 310 to 689 kPa (45 to 100 psi ). The ventilator drive gas supply can be oxygen or air. Note that the drive gas is specified by the customer prior to delivery. Conversion from one drive gas to another must be carried out by a Penlon-trained service engineer.
DESCRIPTION
9
1
10
Control Unit Rear Panel
Interface and Parameter inputs
5. A200SP Absorber Bag/Vent switch interface Spirometer connector
6. Prima SP Interface connector
7. Pressure Monitor Port
8. Input socket - Oxygen monitor sensor
Data and Printer Ports
9. Data Output
10. Output to remote screen
11. Ethernet
12. USB
13. VGA
14. Printer port
15. RS232
NOTE USB port is for use by Penlon-trained engineers only. All other data ports are read only. For further information, please contact Penlon Technical Support.
Gas Connections
1. Ventilator drive gas inlet
- connect to anaesthetic machine
auxiliary gas outlet
2. Bellows Drive Gas Output
- connect to bellows ( on Prima SP with A200SP absorber, connect to absorber - see section 5.1.5)
3. Outlet - Exhaust Valve
Electrical Connection
4. Electrical mains input and fuse unit
DESCRIPTION
2
7
3
13 14 15
12
1110
9
8
6
5
4
1
3.2 Ventilation Cycle
This section provides a simplified description of the ventilation cycle.
1. Inspiratory Phase
The inspiratory proportional valve (1) in the control unit opens, and bellows drive gas is delivered to the bellows housing (2). The expiratory proportional valve (3) opens and gas flows through the bleed valve. The back pressure ensures that the exhaust valve (4) is kept closed Drive gas pressure builds up above the bellows, which starts to move down. The diaphragm (5) in the bellows assembly base is held closed, and patient gas is forced out of the bellows base (6) into the breathing
system.
2. Beginning of Expiratory Phase
The Inspiratory (1) and Expiratory (3) proportional valves close and the exhaust valve (4) opens. Patient gas returns to the bellows. As the bellows rises, redundant drive gas is pushed out through the exhaust valve.
DESCRIPTION
11
1
4
4
5
6
2
3
3
1
3
1
4
DESCRIPTION
3. End of Expiratory Phase
With the bellows at the top of its housing fresh gas continues to flow. To prevent a high pressure build up the exhalation diaphragm (5) lifts and allows gas to exit through the exhaust valve (4).
4. PEEP Positive End Expiratory Pressure (user selectable)
During PEEP the Exhalation Proportional valve (3) applies PEEP pressure plus 20 cmH2O to the exhaust valve, which remains closed at this stage. As fresh gas flows in the patient circuit, any pressure increase above PEEP pressure in the bellows will cause gas to bleed past the exhaust valve (4). A continuous flow from the Inspiratory proportional valve (1) ensures that any fall in pressure is compensated by driving the bellows as required.
12
5
4
13
DESCRIPTION
A
Pneumatic Flow
Diagram
C
1817
5
8
14
9
12
6
16
13
15
11
7
1
2
10
4
3
B
0 - 80 cmH
2O
100 cmH
2O
0 - 90 cmH2O
241 kPa (35 psi)
3 to 7 bar
3.3 Pneumatic System
3.3.1 System Operation
Refer to the pneumatic system diagram on the previous page.
A) Gas inlet manifold block
The AV-S Ventilator is designed to operate on a 310 - 689 kPa (45 -100 psi) drive gas supply (oxygen or air - to customer’s requirement).
1. DISS Connector
The gas source is connected to the DRIVE GAS SUPPLY fitting on the rear of the ventilator control unit. The gas supply should be capable of a flow rate of 80 L/min while maintaining a minimum pressure in excess of 310 kPa (45 psi).
2. Filter The drive gas is filtered with a 40-micron Input Gas Filter which protects the pneumatic components from incoming particulate matter.
3. The Low Supply Pressure Detector The pressure switch is set at a predetermined level to detect a loss or reduction of the input gas source pressure. When the pressure falls below 235 kPa (35 psi ± 1 psi), the LOW SUPPLY PRESSURE indicator will be displayed and the high priority audible alarm will activate.
4. Input Pressure Regulator Regulates the input drive gas to 260 kPa ± 21 kPa (38 psi ± 3 psi).
5. Cut-off Valve The valve isolates the the gas supply : a) when the ventilator is switched off b) when a fault condition occurs.
6. Airway Pressure Sensor Connected to expiratory limb of breathing circuit.
DESCRIPTION
14
B) Pneumatic Control Manifold Block
7. Inspiratory Proportional Valve
8.
Flow Sensor
9. Drive Gas pressure Sensor
10. Low Pressure Regulator
11. Expiratory Proportional Valve
12. PEEP pressure sensor
13. Restrictor The restrictor allows a flow of up to 2 L/min (<2 L/min bleeding)
C) Exhaust Manifold Block
14. Check Valve
15. Diaphragm Valve
16. Pressure Relief valve Set to 100 cmH
2O
17. Exhaust Port ( to AGSS)
18. Bellows drive gas outlet (to bellows assembly)
3.4 Electrical System
Mains Supply
The mains supply inlet is designed for connection to any mains voltage from 100 to 240 VAC and a frequency of 50 to 60 Hz, without any adjustment. The connector is a standard IEC type.
Back-up Battery
In the event of mains electrical failure, the back­up battery cuts in automatically. A fully charged battery will power the ventilator for approximately 30 minutes.
See Appendix 1 for battery care procedures.
DESCRIPTION
15
VmSET
Litres
3.6
VmMEAS
Litres
3.6
SET
BPM
6
Insp Time
Sec
2
PEEP
cmH2O
5
LIMIT
cmH2O
38
Spont
Mode
Pmean
cmH2O
10
Pmax
cmH2O
24
Standby
Freeze
Waveform
SMMV
%O2 100
33 20
Trigger
cmH2O
-1.0
cm H2O
secs
AV-S
Touchscreen control
1
4
3
5
2
Gas Mixture
O2 + air
.
.
IO
o o
DESCRIPTION
5. Navigator Wheel and Press Button
Turn the wheel to select a function or parameter, or to alter the value of an active parameter.
Press to confirm the setting.
3.5.1.2 Selecting Functions and
Parameters
The functions/parameters shown on the screen can be activated as follows: a) touch the screen at the appropriate tab area. b) rotate the navigator wheel and press it when the indicator arrow is on the required parameter tab
Note that parameters default to factory-set values when the ventilator is switched on and no further user selection is made.
3.5.1.3 User Adjustable Parameters
Variable parameters can be altered by rotating the navigator wheel. When the required value is displayed, press the active tab or
the wheel to confirm the
setting.
3.5 Control Panel
3.5.1 Touchscreen and Navigator Wheel / Push Button
3.5.1.1 Control Panel
1. On/Off control
Switch On: Short internal test sequence Switch Off: 5 second power down sequence with audible tones
2. Status indicators for electrical power
(mains/battery supply)
Yellow indicator
- illuminated whenever power is applied to the unit and internal battery is being charged Green indicator
- illuminates when the unit is switched
on, .
3. Menu switch
The menu function provides access to user and service pages
4. Alarm mute switch
30 second or 120 second Alarm silence, depending on alarm status. Note also that some alarms are not mutable (see 3.11).
16
3.5.2 User Adjustable Parameters
Tidal Volume Range 20 - 1600 ml Rate 4 - 100 bpm I:E Ratio 1:0.3 to 1:8 PEEP 4 - 30 cmH2O
Can be set to OFF
Pressure Limit
Volume mode: 10 - 80 cmH2O Pressure mode: 10 - 70 cmH2O
3.5.3 Operational Capability
Tidal Volume, Rate, and I:E ratio settings are all limited by a maximum inspiratory flow of 75 L/min.
DESCRIPTION
17
0.1
0.2
0.4
0.3
0.6
0.5
1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0 10 20 30 40 50 60 70 80
1:6 1:5 1:4 1:3 1:2 1:1 1:0.3
Rate (bpm)
Tidal Volume (litres) (Vt)
I:E Ratio
The ventilator is capable of operating at the volumes and rates below each I:E ratio curve.
Example
1. Select required volume (Vt) (e.g. 0.8 L)
2. Select rate (e.g. 10 bpm).
3. Select I:E ratio of 1:2. The point X on the graph lies beneath the 1:2 ratio curve, and is therefore within the
ventilator’s capability
.
X
3.5.4 Output Compensation Functions
WARNING The AV-S automatically compensates for fresh gas (spirometry On), fresh gas mixture (spirometry and oxygen monitor On), and altitude. However, the actual tidal volume delivered to the patient may be different to the ventilation parameters set by the user, due to:
A) an extreme compliance condition, B) a substantial system leak, C) patient circuit pressure effects, or D) extreme fresh gas flows
In addition, high fresh gas flows will lead to an increased Vt being delivered to the patient. The patient must
be monitored independently from the ventilator. It is the responsibility of the user to monitor the patient for adequate ventilation.
Fresh Gas Compensation
Adjusts delivered volume up to 60% Alarms if measured volume is 50% different than set volume User adjustable
NOTE Fresh gas compensation is disabled if : a) The spirometry system is turned OFF through the menu system, or b) The spirometry system is not functioning correctly .
Fresh Gas Mixture Compensation
- models with Spirometry
The spirometry system compensates for fresh gas mixture - the user must access the menu system and select the gas mixture that will be used for each clinical procedure.
NOTE Fresh gas mixture compensation is disabled if : a) The spirometry system is turned OFF through the menu system, or b) The spirometry system is not functioning correctly .
If the O2 monitor is switched OFF, a 40%/60% mixture of O2/N2O is assumed.
Altitude Compensation
Monitors ambient pressure Adjusts delivered volume accordingly
DESCRIPTION
18
A
B
3.6 Interface to Prima SP2/3 and A200SP
The AV-S is designed to interface with the Prima SP Anaesthetic Machine and the A200SP Absorber.
3.6.1 Prima SP Interface
The interface cable links the socket (A) on the control panel to a socket on the rear panel of the anaesthetic machine.
a) Turn the Prima Sp Gas Delivery Switch to
ON. The ventilator will power-up.
b) While the Prima SP power is ON, the
Ventilator can be turned OFF and ON, using the ventilator On/Off switch, as described in section 3.5.1.
c) Turn the Prima SP Gas Delivery Switch to
OFF. The ventilator will power-down.
3.6.2 A200SP Absorber Interface
The interface cable links the socket (B) on the control panel to a socket (C) at the rear of the absorber.
a) The A200SP is fitted with fitted with a sensor
that detects the position of the absorber bag/vent control (D). The sensor signal cabling is routed internally to connector (C) and a second cable runs to the the rear of the AV-S control unit.
b) Operation of the Bag/Vent control will trigger
automatic Mode switching on the AV-S ventilator, as follows: i) If the Absorber Bag/Vent control is moved from Vent to Bag, the ventilator will change from Volume Mode, or Pressure Mode, into Spontaneous Mode. ii) Switching the absorber Bag/Vent control from Bag to Vent: The ventilator will reset from Spontaneous Mode to the previously set active mode. iii) If the ventilator is in any mode other than those detailed above, operation of the absorber Bag/Vent control will not affect the ventilator.
NOTE This function can be enabled/disabled through the AV-S on-screen menus (Service Sub­menu, see appendix).
19
DESCRIPTION
D
C
3.7 Ventilation Modes
3.7.1 Standby Mode
Allows parameters to be set. Some patient alarms are active:
High airway pressure (at 80 cmH2O) High/Low O2 Negative pressure
Incorrect Rate/Ratio
DESCRIPTION
20
3.7.2 Volume Mode
The ventilator delivers a mandatory set volume of gas at preset, fixed breath intervals. The Patient is making no respiratory effort.
3.7.2.1 Fresh Gas Compensation
Adjusts delivered volume up to 60% This delivered volume will consist of the volume delivered from the ventilator bellows plus the fresh gas flow from the anaesthetic machine fresh gas supply, minus any compliance loss and minus any leak. This gives a total actual inspired tidal volume. An alarm is triggered if measured volume is 50% different than set volume User adjustable
Altitude Compensation
Monitors ambient pressure Adjusts delivered volume accordingly
3.7.2.2 Operating Functions
Inspiratory Pause function: Creates 25% plateau
Sigh function: When the ventilator is in Volume Cycle mode the "sigh" option is available. When selected, this option provides extra volume for 1 to 4 breaths in 50 (frequency is user selectable). The extra volume will be 50% above the tidal volume set by the user.
Volumes measured if Spirometry function selected
Auto High and Low volume alarms if measured volume different by 50% of set volume User adjustable option
If max pressure limit achieved, ventilator cycles to expiratory phase
3.7.2.3 Volume Type Selection
Use the menu to switch between Tidal Volume and Minute Volume.
NOTE Minute Volume is derived from a rolling average during a 30 second period.
Volume Mode Parameters
Tidal volume 20 – 1600 mL Rate 4 – 100 bpm I:E ratio 1:0.3 – 1:8 PEEP 'Off' or adjustable 4 – 30 cmH2O Inspiratory pressure limit 10 to 80 cmH2O Inspiratory pause 25% (does not affect I:E ratio) Sigh 1.5 x Set Vt is
delivered once, twice, three times or four times every 50 breaths (user selects frequency)
DESCRIPTION
21
3.7.3 Pressure Mode
3.7.3.1 Parameters
The ventilator delivers a volume of gas to achieve a set pressure at fixed breath intervals. The Patient is making no respiratory effort. This is a common mode for the ventilation of small paediatric patients.
Inspiratory pressure 10 - 70 cmH2O Rate 4 – 100 bpm I:E ratio 1:0.3 – 1:8 PEEP 'Off' or adjustable: 4 – 30 cmH2O
Inspiratory decelerating flow controlled by the ventilator according to pressure setting No Inspiratory pause function
3.7.3.2 Pressure Mode Operating Functions
Defaults to 10 cmH2O Maximum Inspiratory Flow to achieve target pressure Sustaining flow maintains circuit pressure Control achieved using exhaust valve
DESCRIPTION
22
3.7.4 Spontaneous Mode
3.7.4.1 Parameters
The ventilator monitors the following patient parameters: Rate
I:E ratio Pressure Tidal volume
Provides waveform displays Inspiratory oxygen is measured
3.7.4.2 Spontaneous mode operating functions
No mechanical ventilation No Inspiratory Pause function
Patient Monitoring (Bag mode and Ventilator mode): Airway pressures FiO2, Vt, Rate I:E ratio, Supply pressures Ventilation conditions
Advanced Spontaneous Breathing modes are selectable from this mode - see below, and section 3.7.5.
3.7.4.3 Advanced Spontaneous Breathing Modes
Support modes available from 'Special Modes' (select from main menu): SIMV SMMV PSV
The A200SPAbsorber Bag/Vent control must be in 'Vent' position for these modes to be selected.
Note that if the system fails to detect an absorber bag/vent switch, a confirm message will be displayed.
DESCRIPTION
23
Loading...
+ 65 hidden pages