Omron Z500-MC15E, Z500-MC10E Operation Manual

Cat. No. Z159-E1-2 Note: Specifications subject to change without notice. Printed in Japan
0103-0.5C
Authorized Distributor:
Profile Measuring System
Z500
Operation Manual
Profile Measuring System Z500 Manual 2: Operation Manual
Cat. No. Z159-E1-2
Cat. No. Z159-E1-2
About this Manual
Manual 1: Setup Manual
Installation
Wiring
Please read the following manuals carefully and be sure you understand the information provided before attempting to install or operate the Z500.
This manual describes the hardware for the Z500
(Multi-Dimensional Controller) and how to install the
components.
Be sure to read this manual first.
Manual 2: Operation Manual
Starting the Z500 and Positioning the Workpiece
Adjusting Images
Setting Measurement Conditions and Executing
Outputting to External Devices
SECTION 2 Basic Operations
This section describes how to start the Z500 and how to display images on the monitor.
SECTION 7 I/O Format
This section provides details on the inputs and outputs used for communications with external devices via terminal blocks or RS-232C.
• Communications settings
• I/O format
• Timing for communications
SECTION 3 Menus for Conversational Menu
• Basic settings required for measurement
• Measurement of consistent workpieces
Menus for Conversational Menu
Settings can be made easily by entering information as requested - just as though you are having a conversation with the Z500.
SECTION 4 Menus for Expert Menu
• Customized settings
• Detailed settings
• Measurement of workpieces placed at different positions
Menus for Expert Menu
More detailed setting, such as position compensation for workpieces placed at different positions, can be made.
Cat. No. Z159-E1-2
Multi-Dimensional Controller
Z500
Operation Manual
Manual
Cat. No. Z000-E1-1
Cat. No. Z158-E1-2
Multi-Dimensional Controller
Setup Manual
Manual
Cat. No. Z000-E1-1
Multi-Dimensional Controller
Z500
OMRON Corporation
Industrial Automation Company
Application Sensors Division Sensing Devices and Components Division H.Q. Shiokoji Horikawa, Shimogyo-ku, Kyoto, 600-8530 Japan Tel:(81)75-344-7068 / Fax:(81)75-344-7107
Regional Headquarters
OMRON EUROPE B.V.
Sensor Business Unit Carl-Benz Str. 4, D-71154 Nufringen, Germany Tel:(49)7032-811-0 / Fax:(49)7032-811-199
OMRON ELECTRONICS LLC
1 East Commerce Drive, Schaumburg, IL 60173 U.S.A. Tel:(1)847-843-7900 / Fax:(1)847-843-8568
OMRON ASIA PACIFIC PTE.LTD.
83 Clemenceau Avenue, #11-01, UE Square, 239920 Singapore Tel:(65)835-3011 / Fax:(65)835-2711
OMRON CHINA CO.,LTD.
BEIJING OFFICE Room 1028, Office Building, Beijing Capital Times Square, No.88 West Chang'an Road, Beijing, 100031 China Tel:(86)10-8391-3005 / Fax:(86)10-8391-3688
Manual
1
INTRODUCTION
SECTION 1 SECTION 2 SECTION 3 SECTION 4 SECTION 5 SECTION 6 SECTION 7 SECTION 8
Multi-Dimensional Controller
Z500
Z500-MC10E/MC15E
Operation Manual
INTRODUCTION 2
SECTION 1
Features
9
SECTION 2
Basic Operations
19
SECTION 3
Menus for Conversational Menu
39
SECTION 4
Menus for Expert Menu
53
SECTION 5
Other Functions
109
SECTION 6
System Settings
143
SECTION 7
I/O Format
153
SECTION 8
Troubleshooting
183
2
INTRODUCTION
Special or Critical Applications
Z500
Operation Manual
INTRODUCTION
Special or Critical Applications
When the Z500 will be used in one of the conditions or applications listed below, allow extra safety mar-
gins in ratings and functions, add extra safety feature such as fail-safe systems, and consult your
OMRON representative.
• Operating conditions or environments which are not described in the manual
• Nuclear power control systems, railroad systems, aviation systems, vehicles, combustion systems,
medical equipment, amusement equipment, or safety equipment
• Other systems, machines, and equipment that may have a serious influence on lives and property and
require extra safety features
Product Availability
Some of the products listed may not be available in some countries. Please contact your nearest
OMRON sales office by referring to the addresses provided at the back of this manual.
Regulations and Standards
The Z500 complies with the international regulations and standards below:
• EC Regulations
EMC Directive: No.89/336/EEC
• EN Standards (European Standards)
EN61326:1997 + A1:1998 + A2:2001 (EMI:Class A)
© OMRON, 2003
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form, or by any means, mechanical, electronic, photocopying, recording, or otherwise, without the prior written permission of OMRON. No patent liability is assumed with respect to the use of the information contained herein. Moreover, because OMRON is constantly striving to improve its high-quality products, the information contained in this manual is subject to change without notice. Every precaution has been taken in the preparation of this manual. Nevertheless, OMRON assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of the information contained in this publication.
Z500
Operation Manual
INTRODUCTION
Table of Contents
3
INTRODUCTION
Table of Contents
INTRODUCTION.................................................................................... 3
Table of Contents......................................................................................................... 3
Precaution on Safety ................................................................................................... 5
Menus............................................................................................................................ 6
Editor's Note................................................................................................................. 8
SECTION 1
Features ............................................................................................... 9
1-1 Features of the Z500............................................................................................ 10
1-2 Operational Flow..................................................................................................14
SECTION 2
Basic Operations................................................................................ 19
2-1 Starting the Z500 and Displaying Images..........................................................20
2-2 Menu Operations ................................................................................................ 23
2-2-1 Input Device ............................................................................................... 23
2-2-2 Screen Displays......................................................................................... 24
2-2-3 Menu Tree ..................................................................................................25
2-2-4 Inputting Values ........................................................................................ 27
2-2-5 Adjustment Menu ...................................................................................... 28
2-3 Screen Types and Display Methods .................................................................. 29
2-3-1 Image Monitor............................................................................................ 30
2-3-2 Profile Monitor ........................................................................................... 33
2-3-3 Digital Monitor ........................................................................................... 35
2-3-4 Trend Monitor ............................................................................................ 36
2-4 Saving Settings and Exiting the Z500................................................................37
SECTION 3
Menus for Conversational Menu....................................................... 39
3-1 Setting Measurement Conditions Using the Conversational Menus............... 40
Step
1 Starting Conversational Menus ................................................................ 40
Step
2 Setting Basic Measurement Conditions ..................................................41
Step
3 Setting Measurement Contents ................................................................ 48
Step
4 Setting Measurement Conditions ............................................................. 49
Step
5 Setting Output Contents............................................................................50
Step
6 Starting Measurements .............................................................................51
3-2 Saving Settings and Exiting the Z500................................................................52
4
INTRODUCTION
Table of Contents
Z500
Operation Manual
INTRODUCTION
SECTION 4
Menus for Expert Menu...................................................................... 53
4-1 Setting Measurement Conditions Using the Expert Menus.............................54
Step
1 Starting Up Expert Menu and Entering Set Mode .................................. 54
Step
2 Adjusting Images .......................................................................................56
Step
3 Position Compensation ............................................................................. 64
Step
4 Measurement Settings...............................................................................70
Step
5 Measurement Conditions .......................................................................... 88
Step
6 Output Settings ........................................................................................100
Step
7 Performing Measurement........................................................................ 107
4-2 Saving Settings and Exiting the Z500..............................................................108
SECTION 5
Other Functions ............................................................................... 109
5-1 Executing Force-zero ........................................................................................110
5-2 Entering Display Mode ...................................................................................... 112
5-3 Changing the Screen Display ........................................................................... 113
5-4 Changing Scenes .............................................................................................. 125
5-5 Entering Tools Mode ......................................................................................... 128
5-6 Testing Measurement Performance (Test) ...................................................... 129
5-7 Checking the Line Beam Position (Surrounding image) ...............................135
5-8 Backing Up Data to a Computer.......................................................................136
SECTION 6
System Settings ............................................................................... 143
6-1 Entering System Mode ...................................................................................... 144
6-2 Setting RS-232C Communications Specifications (Comm) .......................... 145
6-3 Environment Settings........................................................................................146
6-4 Downloading the Sensor Information to the Controller................................148
6-5 Compensating Installation Error (Slant correct).............................................149
6-6 Initializing the Z500 (Initialize) .......................................................................... 151
6-7 Checking the System Version (Version) ........................................................ 152
SECTION 7
I/O Format ......................................................................................... 153
7-1 Terminal Blocks ................................................................................................. 154
7-2 RS-232C.............................................................................................................. 159
7-3 Analog Output.................................................................................................... 180
SECTION 8
Troubleshooting ............................................................................... 183
8-1 Troubleshooting ................................................................................................184
Index .................................................................................................. 187
Z500
Operation Manual
INTRODUCTION
Precaution on Safety
5
INTRODUCTION
Precaution on Safety
OMRON products are manufactured for use according to proper procedures by a qualified operator and only for the purposes described in this manual. The following conventions are used to indicate and classify precautions in this manual. Always heed the information provided with them. Failure to heed precautions can result in injury to people or damage to property.
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury, or property damage.
WARNING Failure to read and understand the information provided in this manual may result in
personal injury or death, damage to the product, or product failure. Please read
each section in its entirety and be sure you understand the information provided in
the section and related sections before attempting any of the procedures or opera-
tions given.
6
INTRODUCTION
Menus
Z500
Operation Manual
INTRODUCTION
Menus
Menus for Conversational Menu
(OUT0 to OUT7)
P.152
P.37
P.149
P.151
P.145
P.49
P.50
P.52
P.48
P.43
P.41
P.70
P.72
P.70
P.86
P.78
P.84
P.87
P.74
P.70
P.70
P.70
P.76
P.40
P.40
P.51
P.144
P.136
P.135
P.129
P.128
P.44
*1
*1
*1
*1
*1
*1
*1
*
1:The same construction
as that for
Height
RS-232C P.105
P.103
P.100
P.100
P.112
*2
*2:The hierarchy under
Display Mode varies depending on the type of screen being displayed.
P.88
P.59
P.63
P.88
P.96
P.90
Scn
(0 to 15)
P.125
Set Start setting
Meas method
Measurement region
Sensitivity
Meas set
Conditions
Output contents
End setting
Height
Step: 2 pts
Step: 3 pts
Edge position
Width
Edge center
Peak/bottom
Define
Change
Calibrate
Output
Delete
Run
Tools
Test
Surrounding
image
Backup
System
Comm
Slant correct
P.148Download
P.146Environment
Initialize
Version
Sensitivity
Details
Conditions
# to avg
Trigger
Details
Analog
Output
Terminals
Display
Adjustment Menu p.28
Save
When
Define
has been selected, however,
setting of
Calibrate
is disabled.
Item and hierarchy that can be set through only the adjustment menu
Z500
Operation Manual
INTRODUCTION
Menus
7
INTRODUCTION
Menus for Expert Menu
(OUT0 to OUT7)
P.37
P.145
P.100
P.64
P.70
P.56
P.88
P.65
P.69
P.56
P.63
P.70
P.58
P.59
P.88
P.96
P.90
RS-232C
P.100
P.105
P.103
P.70
P.72
P.86
P.78
P.84
P.87
P.74
P.70
P.70
P.70
P.76
(0 to15)
P.125
P.54
P.51
P.144
P.136
P.135
P.129
P.128
*1:The same construction
as that for
Height
P.112
*2:The hierarchy under Display Mode
varies depending on the type of screen being displayed.
*1
*1
*1
*1
*1
*1
*1
Scn
Set
Image
Meas method
Region
Sensitivity
Details
Compensatn
Reference
Compensation
method
Meas set
Height
Step: 2 pts
Step: 3 pts
Edge position
Width
Edge center
Peak/bottom
Define
Change
Calibrate
Output
Delete
Conditions
Run
Tools
Test
Surrounding
image
Backup
System
# to avg
Trigger
Details
Output
Analog
Terminals
Comm
Save
Display
*2
Adjustment Menu p.28
When
Define
has been selected, however, setting of
Calibrate
is
disabled.
Item and hierarchy that can be set through only the adjustment menu
P.152
P.149
P.151
Slant correct
P.148Download
P.146Environment
Initialize
Version
8
INTRODUCTION
Editor’s Note
Z500
Operation Manual
INTRODUCTION
Editor’s Note
Visual Aids
The following headings appear in the left column of the manual to help you locate different types of information.
NOTICE
Indicates information required to take full advantage of the functions and performance of the product. Incorrect application methods may result in the loss of damage or damage to the product. Read and follow all precautionary information.
CHECK
Indicates points that are important in using product functions or in application procedures.
REFERENCE
Indicates where to find related information.
TwoSensor
Indicates information required when using 2 sensors.
HELP Indicates information helpful in operation.
Notation
Screen Messages
In this manual, screen message are given in bold/italic. E.g.: System System/Backup means the Menu is hierarchical.
OMRON Product References
All OMRON products are capitalized in this manual. The world "Unit" is also capitalized when it refers to an OMRON product, regardless of whether or not it appears in the proper name of the product.
Z500
Operation Manual
SECTION 1 Features
9
SECTION 1
Features
SECTION 1 explains the features of the Z500 and the flow of operation using the
Conversational and Expert Menus.
1-1 Features of the Z500 10
1-2 Operational Flow 14
10
SECTION 1
1-1 Features of the Z500
Z500
Operation Manual
SECTION 1
Features
1-1 Features of the Z500
1-1-1 Eight Measurement Items for Various Detection Needs
Select the measurement item according to application.
Complicated detection is also possible, by combining multiple measurement
items and with detailed settings.
Simultaneous measurement can be performed at 8 points.
REFERENCE
Refer to page 70.
Height
Step : 2 pts
Step : 3 pts
Edge position
Width
Edge center
Peak/bottom
Define
1-1-2 Simple Setup Using Menus
The Z500 has 2 types of menu. Set the measurement contents according to the
menus displayed on the monitor screen.
Menus for Conversational Menu
Settings can be made easily by entering information as requested - just as
though you are having a conversation with the Z500.
REFERENCE
Refer to page 39.
Menus for Expert Menu
More detailed setting, such as position compensation for workpieces placed at
different positions, can be made.
REFERENCE
Refer to page 53.
1-1-3 A Variety of Output Formats
Not only is analog output available, but data output to external devices can also
be performed via terminal blocks or RS-232C.
REFERENCE
Refer to page 100.
Z500
Operation Manual
SECTION 1
1-1 Features of the Z500
11
SECTION 1
Features
1-1-4 Monitoring While Viewing the Measurement Status
The Z500 has 4 types of monitor. Monitoring suited to the application is possible
at every stage from installation and measurement settings to operation and
maintenance.
REFERENCE
Refer to page 29.
PA S S
mm+0005.18800
Zero's OFF
OUT 0
Thru
1div x: 1.000 [s] y:+10.0000[mm]
9.96ms
S+ / :Display S+ENT:Adjust
Scn 0 Run
PASS
Sen 0
mm
+0005.18800
9.96ms
1div x: 1.000 [s] y:+10.0000[mm]
Thru
S+ / :Display S+ENT:Adjust
Scn 0 Run
PASSmm
+0005.18800
Zero's OFF
9.96ms
OUT 0
BRIGHT
DARK
FAR
NEAR
[ 15] [ 170]
[ .] [ ]
Sen0
LV
PEAK
LV
PEAK
Sen1
S+ / :Display S+ENT:Adjust
Scn 0 Run
Image Monitor
Trend Monitor
Digital Monitor
Profile Monitor
SHIFT
SHIFT
SHIFT
SHIFT
This is the main monitor of the Z500. Three pieces of information; measurement status, sensitivity, and measurement result, can be checked together with their target images.
Chronological change of height distribution of 126 points (profile) can be checked on a shaded 3-D image.
Chronological change of measured values can be checked.
Two or more measurement results can be checked at the same time.
+
+
+
+
SHIFT
SHIFT
SHIFT
SHIFT
++
+
+
9.96ms
0
+0002.65432mm PASS
HI=+0010.00000 LO=-0010.00000
1
+0004.65432mm PASS
HI=+0010.00000 LO=-0010.00000
2
+0000.00000mm ERROR
HI=+0010.00000 LO=-0010.00000
3
-----------mm
HI=+0010.00000 LO=-0010.00000
S+ / :Display S+ENT:Adjust
Scn 0 Run
12
SECTION 1
1-1 Features of the Z500
Z500
Operation Manual
SECTION 1
Features
1-1-5 Flexible Settings for Data Output to External Devices
In addition to simply outputting measurement results, equations can be set to
output to external devices using measurement and judgment results.
REFERENCE
Refer to page 70.
1-1-6 A Wide Variety of Useful Functions
Any Type of Workpiece can be Measured by Using the Image Adjustment Function
Higher measurement accuracy can be assured by using the sensitivity
adjustment function, even if the workpieces to be measured have different
brightness.
REFERENCE
Refer to pages 44 and 59.
Adjustment Menu
Six types of adjustments are available on shortcut keys by displaying the
adjustment menu.
REFERENCE
Refer to page 28.
Checking Surrounding Images of the Area that the Line Beam Strikes
Displays images of the area surrounding the part that the line beam strikes.
This function is useful when the Sensor is mounted inside an installation and the
measurement area cannot be checked directly.
REFERENCE
Refer to page 135.
Z500
Z500
Operation Manual
SECTION 1
1-1 Features of the Z500
13
SECTION 1
Features
Specifying Measurement Timing by Using the Trigger Function
Measurement timing can be specified arbitrarily by using the self-trigger and
external trigger functions. Characteristic quantity can be held while the
measurement is carried out.
(Measurement status at a set timing can be checked on the monitor screen.)
REFERENCE
Refer to page 90.
Smooth Setup Changes
Scenes can be used to change between up to 16 types of measurement setups.
Simply change the scene to change to a different measurement setup.
REFERENCE
Refer to page 125.
14
SECTION 1
1-2 Operational Flow
Z500
Operation Manual
SECTION 1
Features
1-2 Operational Flow
1-2-1 Menus for Conversational Menu
Preparations for Detection
Turn ON the power.
Selecting Measurement Item Appropriate to the Measurement Goal
Setting Detection Conditions and Executing
Confirming Settings
Height Step: 2 pts Step: 3 pts Edge position Width Edge center Define
Refer to page 70.
Displaying Images on the Monitor
Refer to page 20,
2-1
Starting the Z500 and Displaying Images.
Screen Types and Display Methods
Refer to page 29.
Starting Up Conversational Menu and Entering Set Mode
Refer to page 40, See
Step 1 : Starting Conversational Menus.
Setting Basic Measurement Conditions
Refer to page 41, See
Step 2 :
Setting Basic Measurement Conditions.
Setting Measurement Conditions (if necessary)
Refer to page 49, See
Step 4 :
Setting Measurement Conditions.
Refer to page 88 for details.
Selecting and Setting Measurement Items
Refer to page 48, See
Step 3 : Setting Measurement Contents.
Refer to page 70 for details.
Starting Measurements
Refer to page 51, See
Step 6 : Starting Measurements.
Checking the Line Beam Position
Refer to page 135,
5-7Checking the Line
Beam Position (Surrounding image).
Setting Output Contents
Refer to page 50, See
Step 5 : Setting Output Contents.
Refer to page 100 for details.
Performing Test
Refer to page 129,
5-6
Testing Measurement Performance (Test).
Refer to page 39 for information on settings using Conversational Menu.
Refer to this section to learn about the basic operational flow from setting measurement conditions to executing measurements.
CHECK
Z500
Operation Manual
SECTION 1
1-2 Operational Flow
15
SECTION 1
Features
Backing up Data
Refer to page 136,
5-8 Backing Up Data
to a Computer (Backup).
Saving Detection Conditions
Refer to page 52,
3-2
Saving Settings and Exiting the Z500.
Initializing the Z500
Refer to page 151.
Setting Conditions by Model Type
Refer to page 125,
5-4
Changing Scenes.
Setting System Environment Conditions
Refer to page 146,
6-3
Environment Settings.
Communications with External Devices
Refer to page 145,
6-2
Setting RS-232C Communications Specifications (Comm).
Changing and Deleting Settings
Application Setting Operations
Saving Settings
Executing Force-zero
Refer to page 110.
I/O Format
Refer to page 153.
Clearing Measurement Items
Refer to page 87.
Changing the Screen Display
Refer to page 113.
Changing Measurement Contents
Refer to page 86.
When an error message has been displayed on the screen,
Refer to page 184,
8-1Troubleshooting.
Troubleshooting
16
SECTION 1
1-2 Operational Flow
Z500
Operation Manual
SECTION 1
Features
1-2-2 Menus for Expert Menu
Setting Detection Conditions and Executing
Starting Up Expert Menu and Entering Set Mode
Refer to page 54,
Step 1 : Starting Up
Expert Menu and Entering Set Mode.
Adjusting Images
Refer to page 56,
Step 2 : Adjusting Images.
Setting Measurement Conditions
(if necessary)
Refer to page 88,
Step 5 : Measurement Conditions.
Selecting and Setting Measurement Items
Refer to page 70,
Step 4: Measurement Settings.
Starting Measurements
Refer to page 107,
Step 7 : Performing Measurement.
Setting Output Contents
Refer to page 100,
Step 6 : Output Settings.
Setting Position Compensation
(if necessary)
Refer to page 64,
Step 3 : Position Compensation.
Selecting Measurement Item Appropriate to the Measurement Goal
Height Step: 2 pts Step: 3 pts Edge position Width Edge center Define
Refer to page 70.
Preparations for Detection
Turn ON the power.
Displaying Images on the Monitor
Refer to page 20,
2-1
Starting the Z500 and Displaying Images.
Screen Types and Display Methods
Refer to page 29.
Refer to page 54 for information on settings using Expert Menu.
Refer to this section to learn about the basic operational flow from setting measurement conditions to executing measurements.
CHECK
Z500
Operation Manual
SECTION 1
1-2 Operational Flow
17
SECTION 1
Features
Confirming Settings
Checking the Line Beam Position
Refer to page 135,
5-7 Checking the Line Beam Position (Surrounding image).
Changing and Deleting Settings
Clearing Measurement Items
Refer to page 87.
Changing the Screen Display
Refer to page 113.
Changing Measurement Contents
Refer to page 86.
Performing Test
Refer to page 129,
5-6 Testing Measurement Performance (Test).
Backing up Data
Refer to page 136,
5-8 Backing Up Data to a Computer (Backup).
Saving Detection Conditions
Refer to page 108,
4-2 Saving Settings and Exiting the Z500.
Initializing the Z500
Refer to page 151.
Setting Conditions by Model Type
Refer to page 125,
5-4
Changing Scenes.
Setting System Environment Conditions
Refer to page 146,
6-3
Environment Settings.
Communications with External Devices
Refer to page 145,
6-2 Setting RS-232C Communications Specifications (Comm).
Executing Force-zero
Refer to page 110.
I/O Format
Refer to page 153.
Application Setting Operations
Saving Settings
When an error message has been displayed on the screen,
Refer to page 184,
8-1Troubleshooting.
Troubleshooting
18
SECTION 1
1-2 Operational Flow
Z500
Operation Manual
SECTION 1
Features
MEMO
Z500
Operation Manual
SECTION 2 Basic Operations
19
SECTION 2
Basic Operations
SECTION 2 shows basic menu operations of the Z500.
2-1 Starting the Z500 and Displaying Images 20
2-2 Menu Operations 23
2-2-1 Input Device 23
2-2-2 Screen Displays 24
2-2-3 Menu Tree 25
2-2-4 Inputting Values 27
2-2-5 Adjustment Menu 28
2-3 Screen Types and Display Methods 29
2-3-1 Image Monitor 30
2-3-2 Profile Monitor 33
2-3-3 Digital Monitor 35
2-3-4 Trend Monitor 36
2-4 Saving Settings and Exiting the Z500 37
20
SECTION 2
2-1 Starting the Z500 and Displaying Images
Z500
Operation Manual
SECTION 2
Basic Operations
2-1 Starting the Z500 and Displaying Images
This section describes how to start the Z500 and how to display images on the monitor.
Place the workpiece and adjust the Sensor position while monitoring the displayed image.
#1
Starting the Z500
1.
Be sure that the basic Z500 components have been connected correctly.
CHECK
Before connecting components or wiring power supply lines and grounding wires, be sure
to refer to the relevant section in the
Setup Manual
.
REFERENCE
Refer to page 16 in the Setup Manual.
2.
Turn ON the power supply to the monitor.
3.
Turn ON the power supply to the Z500.
The Image Monitor will be displayed.
CHECK
If the Image Monitor is not displayed, use the
SHIFT
+
Right
Keys or
SHIFT
+
Left
Keys
on the console to change the Monitors.
REFERENCE
Refer to page 29, Screen Types and Display Methods.
×× × × × × × × × × × ×
× × ×
×× × × × × ××
× × × × × × × × × ×
× ×
× × × × × × ×
ESC TRIG
SHIFT
ENT
CONSOLE
Z500
Power Supply
Sensor
Monitor
Console
PA S Smm
+0005.18800
Zero's OFF
9.96ms
OUT 0
BRIGHT
DARK
FAR
NEAR
[ 15]
[ 170]
[ .] [ ]
Sen0
LV
PEAK
LV
PEAK
Sen1
S+ / :Display S+ENT:Adjust
Scn 0 Run
Z500
Operation Manual
SECTION 2
2-1 Starting the Z500 and Displaying Images
21
SECTION 2
Basic Operations
#2
Position the Workpiece and Adjust the Image
Position the workpiece to be measured and adjust the displayed image in the
measurement range.
The measurement range varies depending on the Sensor models connected.
REFERENCE
Refer to page 26 in the Setup Manual.
1.
Position the workpiece. In this example, the workpiece shown in the right figure will be described.
14.5mm
(±7.25mm)
17.3mm
(±8.65mm)
20.1mm
(±10.05mm)
+20mm
0mm
-20mm
100mm
+20mm
0mm
-20mm
Example: Z500-SW17
Sensor
Measurement range
Height
Position
22
SECTION 2
2-1 Starting the Z500 and Displaying Images
Z500
Operation Manual
SECTION 2
Basic Operations
A profile graph showing the level differences in the workpiece will be displayed.
2.
Adjust the position of the workpiece if necessary. Adjust the workpiece position if the Profile graph is not appeared in the center of the monitor.
CHECK
If the lines representing level differences have swollen elliptically, the light density is
excessive. If the lines are too thin to see, the light density is insufficient. In either case,
adjust the sensitivity.
REFERENCE
Refer to page 59 for information on how to adjust sensitivity.
PA S Smm
+0005.18800
Zero's OFF
9.96ms
OUT 0
BRIGHT
DARK
FAR
NEAR
[ 15]
[ 170]
[ .] [ ]
Sen0
LV
PEAK
LV
PEAK
Sen1
S+ / :Display S+ENT:Adjust
Scn 0 Run
Profile
Z500
Operation Manual
SECTION 2
2-2 Menu Operations
23
SECTION 2
Basic Operations
2-2 Menu Operations
2-2-1 Input Device
Menu operations are performed on the Console.
Key Function
Escape Key Returns the cursor to the previous menu display or operation.
Trigger Key The effect of the Trigger Key varies with the function.
Enter Key
Executes a function or sets a value. Note: On the F160-KP, also functions as a Cursor Key.
SHIFT Key
Must be pressed in combination with another key to have any effect. Specific functions are assigned to combinations of the
SHIFT
Key and other keys for specific screens.
Up, Down, Left, and Right Keys
The Up and
Down
Keys are used to move the cursor up and down and also to set values. Use the Up Key to increase a value by 1. Use the
Down
Key to decrease a value by 1.
Hold down the Up or
Down
Key to quickly increase or decrease a value. The
Left
and
Right
Keys are used to move the cursor left or
right.
ESC TRIG
SHIFT
Z300-KP
CONSOLE
ENT
ESC
SHIFT
TRIG
ENT
Trigger Key
Enter Key
Up, Down, Left, and Right Keys
Escape Key
SHIFT Key
24
SECTION 2
2-2 Menu Operations
Z500
Operation Manual
SECTION 2
Basic Operations
2-2-2 Screen Displays
The Z500 is operated by selecting functions displayed on the screen. Familiarize yourself with each
function before operating the Z500.
Mode
9.96ms
S+ / :Display S+ENT:Adjust
Scn 0 Run
Scene Number
There are 16 scenes. Setting different measurement conditions for each one allows easy switching between different setups.
Mode (See the table below)
The current operating mode is displayed.
Measurement period (Sampling time)
Key Operations
Displays special key combinations at the bottom of the screen where available. S refers to the SHIFT Key. "S + ENT" indicates that the ENT Key should be pressed while the SHIFT Key is pressed.
Cursor Keys
The cursor is moved to the desired function by pressing the Up, Down, Left, and Right Keys.
It varies depending on the item to be measured.
Display Description
Set Mode to set the inspection conditions.
Run
Performs measurement. The measurement results are output to an external device via terminal block or RS-233C.
Tools Used to save settings and images to a computer as backup.
System
Used to set system conditions for the Z500. Select this mode to switch menus and settings for communications with external devices.
Save
Used to save data to flash memory in the Z500. If new settings have been made, be sure to save the data before quit­ting.
Z500
Operation Manual
SECTION 2
2-2 Menu Operations
25
SECTION 2
Basic Operations
2-2-3 Menu Tree
Z500 menus are hierarchical. The cursor is moved to the required functions to set measurement
conditions. Use the following procedures to move around the menu tree.
1.
Move the cursor to the item to be moved. (Move the cursor to the desired function in Set Mode using the Up and Down Keys.)
2.
Press the ENT Key.
In this example, the cursor will move to the Image Menu.
3.
Repeat (1.) and (2.) to move the cursor to lower levels.
Press the ESC Key once to move to the upper level.
Image
Compensatn Meas set Conditions Output
Scn 0 Set
Image
Compensatn Meas set Conditions Output
Meas method
Region Sensitivity Details
Scn 0 Set
26
SECTION 2
2-2 Menu Operations
Z500
Operation Manual
SECTION 2
Basic Operations
Triangle Mark
Items with an inverted triangle after them have a list of alternatives.The method for selecting the
alternatives is given here.
1.
Move the cursor to the item to be set. In this example, move the cursor to Baud rate.
2.
Press the ENT Key.
The selections will be displayed.
3.
Move the cursor to the desired baud rate using the Up and Down Keys.
4.
Press the ENT Key.
The selections will be registered.
38400bps
8bit
None
1bit
CR
None
End
Scn 0 System
Baud rate Data Length Parity Stop bits Delimiter Flow
: : : : : :
38400bps
8bit
None
1bit
CR
None
End
2400bps 4800bps
9600bps 19200bps
38400bps
: : : : : :
Scn 0 System
Baud rate Data Length Parity Stop bits Delimiter Flow
38400bps
8bit
None
1bit
CR
None
End
Scn 0 System
Baud rate Data Length Parity Stop bits Delimiter Flow
: : : : : :
Z500
Operation Manual
SECTION 2
2-2 Menu Operations
27
SECTION 2
Basic Operations
2-2-4 Inputting Values
The section explains how to input values when setting judgment or measurement conditions.
1.
Move the cursor to the item for which a value is to be changed.
2.
Press the ENT Key.
The cursor will change to a cursor the size of a single digit.
3.
Move the cursor to the digit to be changed using the Left and Right Keys.
4.
Change the value. Use the Up Key to increase the value. Use the Down Key to decrease the value.
5.
Repeat (4.) to change other values.
6.
Press the ENT Key.
The values will be set.
End
: : : : : : : :
Scn 0 Set
[ 0.01000]
[ 0.00000]
None
2pixel
50.0% Clamp
None
5pixel
judgement width Edge width Smoothing Noise Level Error data Profile Peak/btm width
End
: : : : : : : :
Scn 0 Set
None
2pixel
50.0% Clamp
None
5pixel
judgement width Edge width Smoothing Noise Level Error data Profile Peak/btm width
[0000.01000] [ 0.00000]
End
: : : : : : : :
Scn 0 Set
None
2pixel
50.0% Clamp
None
5pixel
judgement width Edge width Smoothing Noise Level Error data Profile Peak/btm width
[0000.02000] [ 0.00000]
End
: : : : : : : :
Scn 0 Set
None
2pixel
50.0% Clamp
None
5pixel
judgement width Edge width Smoothing Noise Level Error data Profile Peak/btm width
[ 0.02000]
[ 0.00000]
28
SECTION 2
2-2 Menu Operations
Z500
Operation Manual
SECTION 2
Basic Operations
2-2-5 Adjustment Menu
Press the SHIFT + ENT Keys in Run Mode. The adjustment menu will be displayed.
The adjustment menu is useful to display the set screens of the following six items without passing
through the menu hierarchies.
CHECK
The adjustment menu can be performed from the Image Monitor, the Digital Monitor, the
Profile Monitor, or the Trend Monitor.
Key Allocation Reference
Trigger Key Display P.112
Enter Key Test P.129
Up Key Sensitivity P.59
Down Key Conditions P.88
Left Key Output P.100
Right Key Meas set P.70
Escape Key None -
Scn 0 Run
9.96ms
ESC:Back S+ / :Switch display
ESC TRG ENT
Display
Test
Sensitvy
Output
Meas set
Conditns
Loading...
+ 161 hidden pages