IntelliSense™ 
Automatic Blood Pressure Monitor 
with Easy Wrap Cuff 
Model HEM-773AC
INSTRUCTION MANUAL

2
TABLE OF CONTENTS
Introduction..................................................................................................3
Notes on Safety............................................................................................4
Know Your Unit ...........................................................................................5
Quick Reference Guide................................................................................6
Initial Set-Up
Battery Installation/Replacement.........................................................7
How to Set Time and Date...................................................................8
How to Use the AC Adapter ................................................................9
Correct Usage
How to Apply the Arm Cuff ........................................................10, 11
How to Take a Reading................................................................12, 13
How to Use the Memory Function ....................................................14
Troubleshooting .........................................................................................15
Caution.......................................................................................................16
Care and Maintenance ...............................................................................17
Warranty.....................................................................................................18
Specifications .............................................................................................19

3
INTRODUCTION
Thank you for purchasing the Omron IntelliSense™ Automatic Blood 
Pressure Monitor with Easy Wrap Cuff, HEM-773AC. Although 
your blood pressure fluctuates throughout the day, no self-adjustment 
is necessary as the monitor intelligently adjusts to your changing 
conditions. The IntelliSense™ monitor uses fuzzy logic intelligence 
to sense both your systolic and diastolic blood pressure values, and the 
result is personalized inflation with every reading. Arm discomfort and 
misreading due to improper cuff inflation are virtually eliminated.
Your new OMRON monitor uses the oscillometric method of blood 
pressure measurement. This means the monitor detects your blood’s 
movement through your brachial artery and converts the movements into 
a digital reading. At the press of a button, the monitor automatically fills 
the cuff with air, starts deflation, then displays your systolic, diastolic, 
and pulse readings on an extra large display panel. 
Please read this instruction manual thoroughly before using 
this unit. For specific information on your own blood pressure, 
CONTACT YOUR PHYSICIAN.
NOTE: 
This device is a digital monitor intended for the use in measuring 
blood pressure and pulse rate in adult population with the arm 
circumference ranging from 9 inches to 17 inches (from 23 cm to 43 cm).

4
NOTES ON SAFETY
• The warning signs and the sample icons shown here are listed for you to use the product safely and correctly 
as well as to prevent the risk and the damage to you and others from happening.
• The icons and meanings are as follow.
War ning sign Contents
Indicates matters in which bodily harm or material damage* may arise as a result 
of incorrect handling.
The battery liquid may leak and damage the main unit. Please observe the following points.
• When you are not going to use the unit for a long period of time (approximately three 
months or more), take out the batteries.
• Replace the worn batteries with new ones immediately.
• Do not use worn and new batteries together.
• Do not insert the batteries with their polarities in wrong direction.
Do not force to bend the arm cuff or the air tube excessively. 
When removing the air tube, pull at the plug. 
Do not apply strong shock or drop the main unit. 
Do not inflate the arm cuff without it being wrapped on the arm.
* Material damage refers to a wide range of damage involving your house, household goods,
domestic animals, and pets.
Self-diagnosis of measured results and treatment are dangerous. Please follow the 
instructions of your doctor.
Do not use the unit on infants or persons who cannot express one’s intention. 
Do not use the unit for any purpose other than measuring blood pressure. 
Do not use a cellular phone near the unit. Improper operation may result.
Do not disassemble, repair, or remodel the main unit or the arm cuff 
of the blood pressure monitor.

5
KNOW YOUR UNIT
Air Tube
Green Mark
Air Plug 
(Connects to Air Jack)
Main Unit:
Components:
Arm Cuff 
(Applicable arm 
cuff for Adults 
9 to 17 in.)
Cover
AC Adapter Jack
Release Button of Cover
START Button
Pulse Rate
User Indicator 
(A or B)
Date/Time Display
Heart Symbol
Memory Call Button
Systolic Blood Pressure
ON/OFF Button
Set Button
Adjust Button
User Selector
Diastolic Blood Pressure
Air Jack
Cuff Storage Container
AC Adapter

6
QUICK REFERENCE GUIDE
1. Avoid eating, smoking, and exercising for at least 30 minutes before 
taking a measurement.
2. Remove tight-fitting clothing from your upper left arm.
3. Sit in a chair with your feet flat on the floor and place your left 
arm on a table so that the cuff is at the same level as your heart.
4. Wrap the arm cuff around your upper left arm, making sure that 
the bottom edge of the cuff is approximately 1/2'' to 1'' above the elbow.
5. Pull the end of the cuff so that the entire cuff is evenly tightened 
around your arm, and the Green Marker on the cuff is above 
the brachial artery.
6. Press the ON/OFF button.
7. After the Heart Symbol ( ) appears on the digital panel, press the 
START button, and remain still until the measurement is complete.
8. When measurement is complete, the monitor displays your blood 
pressure and pulse rate, and automatically deflates the cuff. 
9. Wait 2-3 minutes before taking another blood pressure measurement.
You may require more rest time between readings depending 
on your individual physiological characteristics.

7
BATTERY INSTALLATION/REPLACEMENT
1. Slide the battery cover off 
in the direction of the arrow.
2. Install or replace 4 “AA” size 
batteries so the + (positive) 
and – (negative) polarities 
match the polarities of the 
battery compartment as indicated.
3. Replace the battery cover.
NOTES:
If the Low Battery 
Indicator ( ) appears 
on the display, replace 
all four batteries. 
Long-life alkaline 
batteries are recommended.
Remove the batteries if the monitor will not be used for 
an extended period of time.
If the batteries remain removed longer than 30 seconds, 
the set date and time will be deleted.

8
HOW TO SET TIME AND DATE
1. When the batteries are installed, the 
display will show 12:00 a.m.
2. Press and hold the SET button. 
The year digits (2002) will flash.
3. Press the ADJUST button to advance 
the digit(s) one at a time. 
If you hold down the ADJUST button, 
the digits will advance rapidly.
4. Press SET button when desired number 
is on the display to lock setting.
5. Repeat Step 3-4 for month and date.
6. Repeat Step 3-4 for hour and minutes.

9
HOW TO USE THE AC ADAPTER
To avoid potential damage to the monitor, use only the exclusive Omron 
AC adapter (HEM-ADPT1).
To connect the AC adapter
1. Insert the AC adapter plug into the AC Adapter jack on the 
right side of the main unit.
2. Plug in the AC adapter into a 120V AC outlet (50 - 60 cycles).
To remove the AC adapter
disconnect the adapter plug 
from the AC outlet first and 
then disconnect the cord from 
the monitor’s jack. 
NOTE: 
The monitor is designed not 
to draw power from the 
batteries when the AC 
adapter is in use. 

10
HOW TO APPLY THE 
ARM CUFF      
1. Check if the air plug is securely 
inserted in the main unit.
2. Remove tight-fitting clothing 
from your upper arm.
3. Sit in a chair with your feet flat 
on the floor and place your arm 
on a table so that the cuff will 
be at same level as your heart.
4. Hold the cloth velcro on the cuff 
with your other hand.
5. Turn the palm of your hand upward. 
Apply the cuff on your upper arm.
Cloth fastener
when wrapping 
on your left arm

11
6. Align the Green Marker of the cuff 
over the brachial artery on the 
inside of the arm
The bottom of the cuff should 
be approximately 1/2''-1'' 
above your elbow.
7. Secure the cuff around your arm 
using the cloth velcro.
8. Make sure the cuff is wrapped 
firmly in place.
9. Relax your arm and put the elbow 
on the table.
NOTE: Arm cuff can also be wrapped 
on your right arm. Follow the 
instructions above (numbers 1-9)
Bottom Edge 1/2inch
Green Mark
Cuff Tube
HOW TO APPLY THE 
ARM CUFF      
(Continued)
when wrapping 
on your left arm

12
HOW TO TAKE A READING
1. Press the ON/OFF button.
a) All display symbols appear for
approximately one second. 
b) When the monitor becomes ready 
to measure, the Heart Symbol (  ) 
appears on the display.
2. Press the START button and remain still. 
As the cuff begins to inflate, the monitor 
automatically determines your ideal 
inflation level. Because this monitor 
detects the pulse even during inflation, 
do not move your arm and remain still 
until the entire measurement completes.
• If you want to stop the inflation or 
measurement, press the ON/OFF button. 
The monitor will stop inflating and start 
deflating rapidly, and the monitor 
will turn off.
3. Inflation stops automatically 
and measurement is started.
As the cuff slowly deflates, decreasing 
numbers appear on the display and the 
Heart Symbol ( ) flashes at every 
heartbeat. In rare circumstances, a higher 
inflation may be necessary. In those 
cases, the monitor reinflates the cuff up 
to 30 mmHg higher than initial inflation 
and restarts the measurement.
Inflation Deflation
If you are going to use the unit’s memory function 
to automatically store the measurement values, you 
need to choose user A or B on the User Selector 
before starting a measurement.

Continue to press button
• If your systolic pressure is known 
to be more than 220 mmHg, press and 
hold the START button until the monitor 
inflates 30 to 40 mmHg higher than your 
suspected systolic pressure.
NOTE: 
Do not apply more pressure 
than necessary. 
The monitor will not inflate 
above 280 mmHg.
Instructions for special conditions:
HOW TO TAKE A READING 
(Continued)
4. When the measurement is complete, the arm cuff 
completely deflates and your blood pressure and 
pulse rate are displayed.
NOTE: 
The monitor automatically stores blood 
pressure and pulse rate into the memory.
• When 21 sets of readings are stored 
in memory, the oldest set will be deleted 
to store a new set.
5. Press the ON/OFF button to turn the monitor off.
NOTE: 
If you forget to turn the monitor off, 
it will automatically shut itself off 
after five minutes.

14
HOW TO USE THE MEMORY FUNCTION
This monitor has a memory capable of storing 21 sets of readings each for A 
and B User Selector. Every time you complete the measurement, the monitor 
automatically stores blood pressure and pulse rate.
• Select the user to recall memory by the User Selector.
• Turn on the monitor by pressing ON/OFF button.
• Press the  button to recall stored readings while the Heart 
Symbol ( ) is displayed.
Press these two buttons 
at the same time.
•To delete all the stored readings, press the  button and the 
START button simultaneously.
NOTE: You can not delete the stored readings partially.
Date and Time of stored readings 
will be alternately displayed.

15
If you continue to have difficulty obtaining a reading, call 
OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1-800-634-4350. 
Please tell the representative you have an OMRON Model HEM-773AC.
Error Indicator Cause Correction
Cuff under-inflated.
Movement during 
measurement.
Cuff over-inflated 
(more than 280 mmHg).
Battery voltage 
is excessively low.
Turn monitor off, wait 2-3 minutes
and take another measurement 
(some users may require a longer 
waiting period). Repeat steps listed 
under “How to Take a Reading.” 
If “EE” is displayed again, take 
another measurement by pressing 
and holding the START button as 
shown on instruction page 13 
under “Instructions for 
special conditions.” 
Your cuff may be wrapped 
too tightly around your arm.
Turn the monitor off, carefully 
read and repeat steps listed 
under “How to Take a Reading.” 
Replace all four “AA” batteries 
with new batteries.
Problem caused by abnormal 
memory function.
Contact OMRON HEALTHCARE 
CUSTOMER SERVICE (toll free 
1-800-634-4350).
TROUBLESHOOTING

16
CAUTION
CAUTION:
Changes or modifications not expressly approved by Omron Healthcare, Inc. 
could void the user’s warranty.
NOTE : 
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) 
This product has been tested and found to comply with the limits for a Class B 
digital device, pursuant to part 15 of the FCC rules. These limits are designed 
to provide reasonable protection against harmful interference in a residential 
installation. The product generates, uses, and can radiate radio frequency energy 
and, if not installed and used in accordance with the instructions, may cause 
harmful interference to radio communications. However, there is no guarantee 
that interference will not occur in a particular installation. If the product does 
cause harmful interference to radio or television reception, which can 
be determined by turning the product on and off, the user is encouraged 
to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and receiver.
• Connect the product into an outlet on a circuit different from that to which 
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only) 
This digital apparatus does not exceed the Class B limits for radio noise emissions 
from digital apparatus as set out in the interference-causing equipment standard 
entitled “Digital Apparatus”, ICES-003 of the Canadian Department of Communications.

17
CARE AND MAINTENANCE
To keep your digital blood pressure monitor in the best condition and 
protect the unit from damage follow the directions listed below: 
Clean the monitor with a soft dry cloth. Do not use any abrasive 
or volatile cleaners. Do not attempt to clean the cuff. Never immerse 
the monitor or any components in water. 
Store the monitor in a safe and dry location. Do not subject the 
monitor to extreme hot or cold temperatures, humidity and direct 
sunlight. Avoid subjecting the monitor to strong shocks such as 
dropping on the floor. Remove the batteries if the unit will not be 
used for an extended period of time. 
Use the unit consistent with the instruction provided in this manual.
Never inflate the unit over 280 mmHg or attempt to disassemble the 
monitor. Use only authorized parts and accessories.  

18
WARRANTY
Your HEM-773AC IntelliSense™ Automatic Blood Pressure Monitor 
is warranted to be free from manufacturing defects for a period 
of five years under normal use. The five year warranty excludes 
the arm cuff. The arm cuff is warranted for a one year period.  
This warranty extends only to the original retail purchaser.
Should repair be needed within the warranty period, ship the unit 
prepaid with the proof of purchase date and $5.00 for return 
shipping and insurance to:
Omron Healthcare, Inc. 
Attn: Repair Department 
300 Lakeview Parkway
Vernon Hills, IL 60061 
Be sure to include the model number and your phone number 
on any correspondence.
We will either repair or replace (at our option) free of charge any parts 
necessary to correct defects in the materials or workmanship.
The above warranty is complete and exclusive. The warrantor expressly 
disclaims liability for incidental, special, or consequential damages of any 
nature. (Some states do not allow the exclusion or limitation of incidental 
or consequential damages, so the above warranty may not apply to you.)
Any implied warranties arising by the operation of law shall be limited 
in duration to the term of this warranty. (Some states do not allow 
limitations on how long an implied warranty lasts, so the above 
limitation may not apply to you.)
This warranty gives you specific legal rights, and you may have other 
rights which vary from state to state. As a condition to operation of your 
warranty, the enclosed registration card must be completed and sent 
to us within 10 days from the date of purchase.
FOR CUSTOMER SERVICE CONTACT:
By phone toll free 1-800-634-4350 
By using the feedback form at www.omronhealthcare.com

19
SPECIFICATIONS
AC
Model: HEM-773AC 
Display: LCD Digital Display 
Measurement Range: Pressure: 0 to 280 mmHg
Pulse: 40 to 180/minute
Accuracy/Calibration: Pressure: ±3 mmHg 
Pulse: ±5% of reading
Inflation: Automatic by electric pump 
Deflation: Active electronic control valve 
Rapid Pressure Release: Active electronic control valve 
Pressure Detection: Capacitive pressure sensor 
Measurement Method: Oscillometric method 
Pulse Wave Detection: Capacitive pressure sensor 
Power Source: 4 “AA” batteries or AC adapter  
Battery Life: Approximately 250 uses  
Operating  50°F to 104°F (10°C to 40°C)
Temperatures/Humidity: 30 to 85% RH maximum 
Storage –4°F to 140°F (–20°C to 60°C)
Temperatures/Humidity: 10 to 95% RH maximum 
Main Unit Weight: Approximately 31.8 oz (900 g) not including batteries 
Main Unit Dimensions: Approximately 9" (l) x 7" (w) x 4" (h)
(228 mm x 178 mm x 102 mm)
Cuff Size: Approximately 6" (w) (152 mm)
Cuff Tube 23 5⁄8" (l) (600 mm)
Cuff Circumference: Fits arm circumference 9" to 17" 
(228 mm to 432 mm)
Components: Arm cuff, Instruction manual, 
4 ''AA'' Batteries, AC Adapter
NOTE: These specifications are subject to change without notice.

Manufactured by: OMRON
Distributed by:
OMRON HEALTHCARE, INC.
300 Lakeview Parkway 
Vernon Hills, Illinois 60061 
www.omronhealthcare.com
Copyright © 2003 OMRON HEALTHCARE, INC.
773ACINST Rev A