Omron BP742N Owner's Manual

Page 1
INSTRUCTION MANUAL
5 series
Blood Pressure Monitor
Model BP742N
ESPAÑOL
ENGLISH
Page 2
TABLE OF CONTENTS
Introduction ............................................................................................... 3
Important Safety Information..................................................................... 4
Operating Instructions
Know Your Device..................................................................................... 8
Preparation ............................................................................................. 13
Using the Device..................................................................................... 15
Care and Maintenance
Error Messages and Troubleshooting..................................................... 23
Maintenance and Storage....................................................................... 26
Specifications.......................................................................................... 29
FCC Statement ....................................................................................... 31
Limited Warranty..................................................................................... 32
Guidance and Manufacturer’s Declaration ............................................. 33
Display symbols ..................................................................... 10
Before Taking a Measurement ............................................... 12
Battery Installation.................................................................. 13
Setting the Date and Time...................................................... 14
Applying the Arm Cuff ............................................................ 15
How to Sit Correctly................................................................ 17
Taking a Measurement........................................................... 17
Using the Memory Function ................................................... 20
Error Messages...................................................................... 23
Troubleshooting...................................................................... 25
Maintenance........................................................................... 26
Storage................................................................................... 27
Optional Medical Accessories ................................................ 27
2
Page 3
INTRODUCTION
Thank you for purchasing the OMRON® BP742N Blood Pressure Monitor. Your new blood pressure monitor uses the oscillometric method of blood
essure measurement. This means the monitor detects your blood movement
pr through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
Intended Use
This device is a digital monitor intended for use in measuring blo and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with the measurement result.
Please read this instruction manual thor
Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR PHYSICIAN.
oughly before using the device.
od pressure
3
Page 4
IMPORTANT SAFETY INFORMATION
Warning: Indicates a potentially hazardous situation which, if not
(General Usage)
DO NOT adjust medication based on measurement pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.
This device is not intended to be a diagnostic device. Consult your physician before using the device for any of the following
conditions: common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age, pregnancy, pre-eclampsia, renal diseases. Note that PATIENT motion, trembling, shivering may affect the measurement reading.
Do not use the device on the injured arm or the arm under medical treatment.
Do not apply the arm cuff on the arm while being on an intravenous drip or blood transfusion.
Consult your physician before using the device on the arm with an arterio­venous (A-V) shunt.
Do not use the device with other medical electrical (ME) equipment simultaneously.
Do not use the device in the area the HF surgical equipment, MRI, or CT scanner exists, or in the oxygen rich environment.
The air tube or the AC adapter cable may cause accidental strangulation in infants.
Contains small parts that may cause a choking hazard if swallowed by infants.
avoided, could result in death or serious injury.
results from this blood
4
Page 5
IMPORTANT SAFETY INFORMATION
Caution: Indicates a potentially hazardous situation which, if not
(General Usage)
Always consult your physician. Self-diagnosis of self-treatment are dangerous.
Consult your physician before using the device for any of the following conditions:
• If you have had a mastectomy.
• Do not take measurements more than necessary. It may cause bruising
• People with severe blood flow problems or blood disorders as cuff inflation
Remove the arm cuff if it does not start deflating during the measurement. Do not use this device on infants or persons who cannot express their
intentions. Do not use the device for any purpose other than measuring blood pressure. Use only the approved arm cuff for this device. Use of other arm cuffs may
result in incorrect measurement results. Do not use a mobile phone or other devices that emit electromagnetic fields,
near the device. This may result in incorrect operation of the device. Do not disassemble the monitor or arm cuff. This may cause an inaccurate
reading. Do not use in a location with moisture, or a location where water may splash
on the device. This may damage the device. Do not use the device in a moving vehicle (car, airplane). Read “If your systolic pressure is more than 210 mmHg” of this instruction
manual, if your systolic pressure is known to be more than 210 mmHg. Inflating to a higher pressure than necessary may result in bruising where the arm cuff is applied.
avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
measurement results and
due to blood flow interference.
can cause bruising.
5
Page 6
IMPORTANT SAFETY INFORMATION
(AC Adapter (optional) Usage)
Do not use the AC adapter if the device or the power cord is damaged. Turn of
f the power and unplug the power cord immediately.
Plug the AC adapter into the appropriate voltage outlet. Do not use in a multi-outlet plug.
Never plug in or unplug the power cord from the electric outlet with wet hands.
Fully insert the power plug into the outlet. When disconnecting the power plug from the outlet, do not pull the power
cord. Be sure to pull from the power plug safely. When handling the power cord, take care not to do the following:
Do not damage. Do not break it. Do not tamper with it. Do not forcibly bend or pull. Do not twist. Do not bundle during use.
Do not pinch. Do not place under heavy objects. Wipe the dust off from the power plug. Unplug monitor when not in use. Disconnect the power plug before cleaning. Use only an OMRON AC adapter designed for this device. Use of
unsupported adapters may damage and/or may be hazardous to the device.
(Battery Usage)
Do not insert the batteries with their p Use only 4 “AA” alkaline or manganese batteries with this device. Do not use
other types of batteries. Do not use new and used batteries together. Remove the batteries if the device will not be used for three months or more.
olarities incorrectly aligned.
6
Page 7
IMPORTANT SAFETY INFORMATION
General Precautions
• Do not forcibly crease the arm cuff or the air tube excessively.
• Do not press the air tube while taking a measurement.
• To unplug the air plug, pull on the air plug at the connection with the monitor not the tube itself.
• Do not drop the monitor or subject device to strong shocks or
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not use the device outside the specified environment. It may cause an
ccurate reading.
ina
• Dispose of the device, components and opt applicable local regulations. Unlawful disposal may cause environmental pollution.
ional accessories according to
vibrations.
,
7
Page 8
KNOW YOUR DEVICE
G
H
I
D
A
F E
B
C
DD
J
L
K
J. Arm cuff
(Arm circumference 9” - 17” (22 - 42 cm)) K. Air plug L. Air tube
Contents:
Monitor, arm cuff, battery set, instruction manual, quick start guide
Monitor:
A. Display B. Up/Down buttons C. USER ID selection switch D. Date/Time setting button E. START/STOP button
Arm cuff:
F. Memory button G. Battery compartment H. Air jack I. AC adapter jack
(for optional AC adapter)
8
Page 9
Display:
N
O
P
R
M
S
T
Q
U
V
W
X
Y
KNOW YOUR DEVICE
M. USER ID symbol N. Systolic blood pressure O. Diastolic blood pressure P. Low battery symbol Q. Heartbeat symbol
(Flashes during measurement.)
R. Memory symbol
S. Average value symbol T. Date/Time display U. Movement error symbol V. Irregular heartbeat symbol W. Blood pressure level indicator X. Deflation symbol Y. Pulse display / Memory number
9
Page 10
KNOW YOUR DEVICE
Irregular Heartbeat
Normal Heartbeat
Pulse
Blood pressure
Blood pressure
Short
Long
Pulse
Display symbols
Irregular Heartbeat Symbol ( )
When the monitor detects an irregular rhythm two or more times during the
ement, the irregular heartbeat
measur symbol will appear on the display with the measurement values.
An irregular heartbeat rhythm is defined
hythm that is 25% less or 25% more
as a r than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure.
If the irregular heartbeat symbol displays with your measur recommend you consult your physician. Follow the directions of your physician.
Movement Error Symbol ( )
The movement error symbol is displayed if you move your body during the measurement. Please remove the arm cuff, and wait 2 - 3 minutes. Take another measurement, remain still during measurement.
Average Value Symbol ( )
The average value symbol is displayed when you press and hold the memory button for more than 3 seconds. The most recent average value appears on the display screen.
ement results, we
10
Page 11
KNOW YOUR DEVICE
Blood Pressure Level Indicator
2013 ESH/ESC Guidelines for the management of arterial hypertension
Definitions of hypertension by office
Systolic Blood Pressure  140 mmHg  135 mmHg
Diastolic Blood Pressure  90 mmHg  85 mmHg
These are from statistical values for blood pressure.
and home blood pressure levels
Office Home
11
Page 12
KNOW YOUR DEVICE
Before Taking a Measurement
To help ensure an accurate reading, follow these directions:
1. Avoid bathing, drinking alcohol or caffeine, smoking, 30 minutes before taking a measurement. Rest for a taking the measurement.
2. Stress raises blood pressure. Avoid taking measurements during stressful times.
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your arm.
5. Sit on a chair with your feet flat on the floor. Rest your arm on a table so that
m cuff is at the same level as your heart.
the ar
6. Remain still and do not talk during the measurement.
7. Keep a record of your blood pressure and pu A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time. Try to measure your blood pressure at the same time each day for consistency.
exercising and eating for
t least 15 minutes before
lse readings for your physician.
12
Page 13
PREPARATION
Battery Installation
1. Remove the battery cover.
2. Insert 4 “AA” batteries as
indicated in the battery compartment.
3. Replace the battery cover.
Notes:
• When the low battery symbol ( off, then replace all batteries at the same are recommended.
• The measurement values continue to be stored batteries are replaced.
• The supplied batteries may have a shorter life.
• Dispose of the device, components and option applicable local regulations. Unlawful disposal may cause environmental pollution.
) appears on the display, turn the monitor
time. Long life alkaline batteries
in me
mory even after the
al accessories
according to
13
Page 14
PREPARATION
Minute
Change Confirm
Change Confirm
Change Confirm
Change Confirm
Change Confirm
Year
Month
Day
Hour
: Forward : Back
Setting the Date and Time
Set the monitor to the correct date and time before taking a measurement for the first time.
1. Press the button.
2. Push or to change the year.
Push button to confirm the year and then the month flashes. Repeat the same steps to change the month, day, hour, and minute.
3. Press the START/STOP button to turn the monitor off.
Notes:
• If the batteries have been replaced, the date and time setting will need to be reset.
• If the date and time are not set, “-:--” app
ears during or after measurement.
14
Page 15
USING THE DEVICE
Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm. Do not place the arm cuff over thick clothes.
1. Insert the air plug into the air jack
securely.
2. Wrap the arm cuff firmly in
place around your left upper arm.
The bottom edge of the arm cuff should be 1/2 inch (1 to 2 cm) above the elbow. Air tube is on the inside of your arm and aligned with your middle finger.
1/2 inch (1 - 2 cm)
15
Page 16
USING THE DEVICE
3. Secure closed with the
fabric fastener.
Notes:
When you take a measurement on the right
• arm, the air tube will be at the side of your
. Be careful not to rest your arm on the
elbow
ube.
air t
• The blood pressure can differ between the the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your measurements.
right arm and the left arm, and
16
Page 17
USING THE DEVICE
How to Sit Correctly
To take a measurement, you need to be relaxed and comfortably seated, u
nder comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• Sit with your back and arm being
orted.
supp
• The arm cuff should be placed on your arm
the same level as your heart.
at
Taking a Measurement
Notes:
• To stop the measurement, press the START/STOP button once to deflate the
cuff.
arm
• Remain still and quiet while t
The monitor is designed to take measurements and store the measurement values in the memory for 2 people using USER ID 1 and USER ID 2.
1. Select your User ID (1 or 2).
aking a measurement.
17
Page 18
USING THE DEVICE
START INFLATE DEFLATE COMPLETE
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg. Inflating to a higher pressure than necessary may result in bruising
where the arm cuff is applied.
2. Press the START/STOP button.
The arm cuff will start to inflate automatically.
18
Page 19
USING THE DEVICE
3. Remove the arm cuff.
4. Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes.
Note: Wait 2-3 minutes before taking another measurement. Waiting
between prior to taking a measurement.
measurements allows the arteries to return to the condition
Using the Guest Mode
The monitor stores measurement values for 2 users in the memory. The guest mode can be used to take a single measurement for another user. No measurement values are stored in the memory when the guest mode is selected.
1. Press and hold the START/STOP
button for more than 3 seconds.
The USER ID symbol and the Date/Time display will disappear.
2. Release the START/STOP button when the Date/Time
display turns off.
The arm cuff will start to inflate automatically.
DO NOT adjust medication based on measurement pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.
This device is not intended to be a diagnostic device.
19
results from this blood
Page 20
USING THE DEVICE
Alternating date/time display
Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
Inflating to a higher pressure than necessary may result in bruising where the arm cuff is applied.
Using the Memory Function
The monitor automatically stores the results up to 50 sets for each user (1 and 2). It can also calculate an average value based on the last 3 measurement values taken within 10 minutes.
Notes:
• If there are only 2 measurement values in the memory for that period, the
average will be
• If there is 1 measurement value in the mem
as the average.
• If the memory is full, the monito
• When viewing the measurement value taken without setting the date and time,
“-:--” is
To View the Measurement Values Stored in Memory
1. Select your User ID (1 or 2).
2. Press the button.
The Memory number appears for a second before the pulse rate is displayed.
The newest set is numbered “1”.
based on these 2 values.
ory for that period, this is displayed
r will delete the oldest value.
displayed instead of the date and time.
20
Page 21
USING THE DEVICE
3. Press the or button to view the values stored in
memory.
: To view the older values
: To view the more recent values
To View the Average Value
1. Select your User ID (1 or 2).
2. Press and hold the button for
more than 3 seconds.
Notes:
• If the previous measurement was taken without setting the date and time, the average value is not calculated.
• If there are no measurements results stored in the memory, the screen to the right is displayed.
21
Page 22
USING THE DEVICE
To Delete All the Values Stored in Memory
The values stored in the memory are deleted by USER ID.
1. Select your User ID (1 or 2).
2. Press the button, while the memory symbol appears.
3. While holding the
button down, press the ST
ART/STOP button for
more than 3 seconds.
Note: You cannot partially delete the values stored in the memory.
All values for the user you select will be deleted.
22
Page 23
ERROR MESSAGES AND TROUBLESHOOTING
Error Messages
Error Display Cause Solution
Remove the arm cuff. Wait 2 - 3 minutes
Irregular heartbeats are detected.
and then take another measurement. Repeat the steps in Measurement” appear, contact your physician.
“Taking a
. If this error continues to
Movement during measurement.
The batteries are low.
The batteries are exhausted.
Air plug disconnected.
Arm cuff is applied too loosely.
Air is leaking from
rm cuff.
the a
Carefully read and repeat the steps in “Taking a Measurement”
You should replace the batteries with ne
w ones ahead of time.
Refer to “Battery Installation”.
You should replace the batteries with ne
w ones at once.
Refer to “Battery Installation”.
Insert the plug securely. Refer to “Applying the Arm Cuff”.
Apply the arm cuff tighter. Refer to “Applying the Arm Cuff”.
Replace the arm cuff with a new one. Refer to “Optional Medical Accessories”.
.
23
Page 24
ERROR MESSAGES AND TROUBLESHOOTING
Error Display Cause Solution
Movement during measurement and the arm cuff has not been inflated sufficiently.
The arm cuff was inflated exceeding the maximum allowable pressure, and then deflated automatically when inflating the arm cuff manually.
Movement during measurement.
Repeat measurement. Remain still and do no Refer to
If “E2” appears repeatedly, inflate the arm cuff manually until it is 30 to 40 mmHg above your previous measurement r Refer to
Do not touch the arm cuff and/or bend th measurement. Do not inflate the arm cuff more than necessary. Refer to
Repeat measurement. Remain still and do no Refer to
t talk during measurement.
“Taking a Measurement”
esult.
“Taking a Measurement”
e air tube while taking a
“Taking a Measurement”
t talk during measurement.
“Taking a Measurement”
.
.
.
.
Clothing is interfering with the arm cuff.
Device error. Contact Customer Service.
Remove any clothing interfering with th
e arm cuff.
Refer to “Applying the Arm Cuff”.
24
Page 25
ERROR MESSAGES AND TROUBLESHOOTING
Troubleshooting
Problem Cause and Solution
Replace all batteries with new ones.
No power. No display appears on the monitor.
Measurement values appear too high or too low
.
Check the battery installation for pr
oper placement of the battery polarities. Refer to “Battery Installation”.
Blood pressure varies constantly. Many factors including stress, day, and how you wrap the cuff, may affect your blood pressure. Review the sections “Before Taking a Measurement” and Measurement”
“Taking a
.
time of
25
Page 26
MAINTENANCE AND STORAGE
Maintenance
To protect your device from damage, please observe the following:
• Store the device and the components in a clean, safe location.
• Do not use any abrasive or volatile cleaners.
• Do not wash the device and any components or immerse them in water
• Do not use gasoline, thinners or similar solvents to clean the devic
• Use a soft and dry cloth, or a soft and moistened cloth and neutral soap to clean on the monitor and the arm cuff.
• Changes or modification not ap will void the user warranty. Do not disassemble or attempt to repair the device or components. Consult Customer Service.
.
e.
proved by the manufacturer
26
Page 27
MAINTENANCE AND STORAGE
Storage
1. Unplug the air plug from the air jack.
2. Gently fold the air tube into the arm
cuff.
Note: Do not bend or crease the air tube
excessively.
Do not store the device in the following situations:
• If the device is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapors such as bleach.
• Locations exposed to vibrations, sho
Optional Medical Accessories
Arm cuff AC Adapter
Arm circumference
9” - 17” (22 - 42 cm)
cks or where it will be at an angle.
CD-WR17
(Model: HEM-RML31)
HEM-ADPTW5
27
Page 28
MAINTENANCE AND STORAGE
Using the Optional AC Adapter
1. Insert the AC adapter plug
into the AC adapter jack on the rear side of the monitor.
2. Plug the AC adapter into an
electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first, and then remove the AC adapter plug from the monitor.
28
Page 29
SPECIFICATIONS
Model Display LCD digital display
Measurement range
Accuracy
Inflation Fuzzy-logic controlled by electric pump Deflation Automatic pressure release valve Measurement method Oscillometric method IP classification IP 20
Power source
Battery life Approximately 1000 measurements (using new alkaline batteries) Operating temperature /
humidity Storage temperature /
umidity / air pressure
h
Weight
Dimensions
Cuff circumference 9” to 17” (220 to 420 mm) Memory Up to 50 per user Content s Monitor, arm cuff, battery set, instruction manual, quick start guide
Applied part
Protection against electric shock
BP742N
Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats / min.
Pressure: r3 mmHg or 2% of reading Pulse: r 5
4 “AA” batteries 1.5V or optional AC adapter (INPUT AC100-240V 50/60Hz 0.12A)
50°F to 104°F (10°C to 40°C) / 15 to 90% RH
-4°F to 140°F (-20°C to 60°C) / 10 to 95% RH / 700 to 1060 hPa
Monitor : Approximately 10 5/8 oz. (300 g) not including batteries Arm cuff : Approximately 6 oz. (170 g)
Monitor : Approximately 4 1/4” (w) u 3 1/8” (h) u 5 1/2” (l)
Arm cuff : Approximately 5 3/4” u 23
Internally powered ME equipment (When using only the batteries)
HEM-7131-Z
% of display reading
(107 mm u 79 mm u 14
(145 mm u 594 mm (air tube: 750 mm))
= Type BF
= Class II ME equipment (Optional AC adapter)
1/2” (air tube: 29 1/2”)
1 mm)
29
Page 30
SPECIFICATIONS
Notes:
• These specifications are subject to
• In the clinical validation study, the 5th phase was used on 85 subjects for determination of diastolic blood pressure.
• This device has not been validated for use on pregnant patients.
change without notice.
30
Page 31
FCC STATEMENT
FCC CAUTION Changes or modifications not expressly appr compliance could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which
ceiver is connected.
the re
• Consult the dealer or an experienced rad
oved by the party responsible for
io/TV technician for help.
31
Page 32
LIMITED WARRANTY
Your BP742N Automatic Blood Pressure Monitor, excluding the arm cuff and batteries, is warranted to be free from defects in materials and workmanship appearing within 2 years from the date of purchase, when used in accordance with
he instructions provided with the monitor. The arm cuff is warranted to be free
t from defects in materials and workmanship appearing within 1 year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor. The above warranty extends only to the original retail purchaser.
We will, at our option, replace without charge any monitor or arm cuff covered by the above warranty. Replacement is our only responsibility and your only remedy under the above warranty.
To obtain warranty service contact Customer Service by calling for the address of the inspection center and the return shipping and handling fee.
Enclose the original printed receipt. Include a letter, with your name, address, phone number, and description of the specific problem. Pack the product carefully to prevent damage in transit. Because of possible loss in transit, we recommend insuring the product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS
CONNE ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENT EXPENSES OR DAMAGES.
This warranty provides you with specific le rights that vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE Visit our web site at: www.omronhealthcare.com Call toll free: 1-800-634-4350
AL, CONSEQUENTIAL OR INDIRECT COSTS,
gal rights, and you may have other
1-800-634-4350
32
Page 33
GUIDANCE AND MANUFACTURER’S DECLARATION
OMRON Automatic Blood Pressure Monitor Information for accompanying documents in the scope of IEC60601-1-2:2007 Model: BP742N including AC-adapter
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the
im to prevent unsafe product situations, the IEC60601-1-2 standard has been
a implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.
Medical devices manufactured by OMRON Healthcare conform to this IEC60601-1-2:2007
andard for both immunity and emissions.
st Nevertheless, special precautions need to be observed:
• The use of accessories and cables other t exception of cables sold by OMRON as replacement parts for internal components, may result in increased emission or decreased immunity of the device.
• The medical devices should not be used adjacent In case adjacent or stacked use is necessary verify normal operation in the configuration in which it will be used.
• Refer to further guidance below regarding the EMC environment in which the device
d be used.
shoul
• The MEDICAL ELECTRICAL EQUIPMENT BP742N including AC-adapter needs special pre according to the EMC information provided in this documentations.
• The Essential Performance of the BP742N includ pressure and a pulse rate and using the memory function.
The BP742N including AC-adapter may be interfered with by other equipment, even if that
ther equipment complies with CISPR EMISSION requirements.
o
cautions regarding EMC and needs to be installed and put into service
han those specified by OMRON, with the
to or stacked with other equipment.
, the medical device should be observed to
ing AC-adapter is to measure a blood
33
Page 34
GUIDANCE AND MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP742N including AC-adapter is intended for use in the electromagnetic environment specified below assure that it is used in such environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicke emissions IEC61000-3-3
. The customer or the user of this OMRON BP742N including AC-adapter should
The OMRON BP742N including AC-adapter uses RF
r
Group 1
Class B
Class A
Complies
energy only emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The OMRON BP742N including AC-adapter is suit domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
for its internal function. Therefore, its RF
able for use in all establishments, including
34
Page 35
GUIDANCE AND MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP742N including AC-adapter is intended for use in the electromagnetic environment specified below assure that it is used in such environment.
Immunity test
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply inputlines IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
. The customer or the user of this OMRON BP742N including AC-adapter should
IEC 60601
test leve
±6 kV contact ±8 kV air
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV line(s) to line(s) ±2 kV line(s) to earth
<5 % U
T (>95 % dip
in U
T)
for 0.5 cycle 40 % UT (60 % dip
in U
T)
for 5 cycles 70 % U
T (30 % dip
T)
in U for 25 cycles
<5 % U
T (95 % dip
in U
T)
for 5 sec.
Compliance level
l
±6 kV contact ±8 kV air
±2 kV for power supply lines ±1 kV for input/ output lines
±1 kV line(s) to line(s) ±2 kV line(s) to earth
<5 % U in U for 0.5 cycle
40 % U in U for 5 cycles
70 % U in U for 25 cycles
<5 % U in U for 5 sec.
3 A/m 3 A/m
T (>95 % dip
T)
T (60 % dip
T)
T (30 % dip
T)
T (95 % dip
T)
Electromagnetic
environment - guidance
Floor should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be th
at of a typical commercial and/or hospital environment.
Mains power quality should be that of a typical commercial and/or hospital environment.
Mains power quality should be that of a typical commercial and/or hospital environment. If the user of t
he OMRON BP742N including AC-adapter requires continued operation during power mains interruption, it is recommended that the OMRON BP742N including AC-adapter be powered from an uninterruptible power supply.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical co mmercial or hospital environment.
35
Page 36
GUIDANCE AND MANUFACTURER’S DECLARATION
OMRON BP742N including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP742N including AC-adapter should assure that it is used in such environment.
Immunity test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
Note1: At 80 MHz and 800 MHz, the higher frequency range applies. Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones
and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the OMRON BP742N including AC-adapter is used exceeds the applicable RF compliance level above, the OMRON BP742N including AC-adapter should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the OMRON BP742N including AC-adapter.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Guidance and manufacturer’s declaration - electromagnetic immunity
IEC 60601
st level
te
3 V rms 150 kHz to 80 MHz
3 V/m 80 MHz to
5 GHz
2.
bsorption and reflection from structures, objects, and people.
a
Compli-
ance level
3 V rms
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the OMRON BP742N including AC-adapter including cables, than the recommended separation distance calculated from the equation appropriate to the frequency of the transmitter.
Recommend separation distance
d = 1.2 P
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
ansmitter in watts (W) according to the transmitter
tr manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters as determined by
electromagnetic site survey,
an compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
a
should be less than the
b
36
Page 37
GUIDANCE AND MANUFACTURER’S DECLARATION
Recommended separation distance between portable and mobile RF communications
OMRON BP742N including AC-adapter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this OMRON BP742N including AC-adapter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the OMRON BP742N including AC-adapter as recommended below, according to the maximum output power of the communications equipment.
Output Power of Transmitter
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorpt
equipment and the OMRON BP742N including AC-adapter
Separation distance according to frequency of
in Watt
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
ion and reflection from structures, objects, and people.
150 kHz to 80
MHz
d = 1.2 P
transmitte
80 MHz to 800 MHz
r in meter
d = 1.2 P
800 MHz to 2.5GHz
d = 2.3
P
37
Page 38
Manufactured for: OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN
Distributed by: OMRON HEALTHCARE, INC. 1925 West Field Court Lake Forest, IL 60045 U.S.A.
www.omronhealthcare.com
© 2013 OMRON HEALTHCARE, INC.
Made in Vietnam 2298802-0D
Loading...