13. Guidance and Manufacturer’s Declaration...........................................50
2
Introduction
Thank you for purchasing the OMRON BP7250 Blood Pressure Monitor.
This monitor can be used by multiple users. This blood pressure monitor
uses the oscillometric method of blood pressure measurement. This
means this monitor detects your blood movement through your brachial
artery and converts the movements into a digital reading.
Safety Instructions
This instruction manual provides you with important information about the
OMRON BP7250 Blood Pressure Monitor. To ensure the safe and proper
use of this monitor, READ and UNDERSTAND all of the safety and
operating instructions. If you do not understand these instructions or
have any questions, contact
1-800-634-4350 before attempting to use this monitor. For specific
information about your own blood pressure, consult with your
physician.
Intended Use
This device is a digital monitor intended for use in measuring blood
pressure and pulse rate in adult patient population. The device detects the
appearance of irregular heartbeats during measurement and gives a
warning signal with readings.
Environments of Use: Home
Patient Population: Adult
Receiving and Inspection
Remove this monitor and other components from the packaging and
inspect for damage. If this monitor or any other components is damaged,
DO NOT USE and contact
1-800-634-4350.
Symbols Glossary
For symbol information, visit:
OmronHealthcare.com/symbols-glossary
3
Contraindications
• This monitor is contraindicated for use in ambulatory environments.
• This monitor is contraindicated for use on aircraft.
4
Important Safety Information
Warning
Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
Read the Important Safety Information in this instruction manual
before using this monitor.
Follow this instruction manual thoroughly for your safety.
Keep for future reference. For specific information about your own blood
pressure, CONSULT WITH YOUR PHYSICIAN.
• DO NOT use this monitor on infants, toddlers, children or persons
who cannot express themselves.
• DO NOT adjust medication based on readings from this blood
pressure monitor. Take medication as prescribed by your
physician. ONLY a physician is qualified to diagnose and treat high
blood pressure.
• DO NOT use this monitor on an injured arm or an arm under
medical treatment.
• DO NOT apply the arm cuff on your arm while on an intravenous
drip or blood transfusion.
• DO NOT use this monitor in areas containing high frequency (HF)
surgical equipment, magnetic resonance imaging (MRI) equipment,
computerized tomography (CT) scanners. This may result in
incorrect operation of the monitor and/or cause an inaccurate
reading.
• DO NOT use this monitor in oxygen rich environments or near
flammable gas.
• Consult with your physician before using this monitor if you have
common arrhythmias such as atrial or ventricular premature beats
or atrial fibrillation; arterial sclerosis; poor perfusion; diabetes;
pregnancy; pre-eclampsia or renal disease. NOTE that any of these
conditions in addition to patient motion, trembling, or shivering
may affect the measurement reading.
5
• NEVER diagnose or treat yourself based on your readings. ALWAYS
consult with your physician.
• To help avoid strangulation, keep the air tube and AC adapter cable
away from infants, toddlers or children.
• This product contains small parts that may cause a choking hazard
if swallowed by infants, toddlers or children.
Data Transmission
• This product emits radio frequencies (RF) in the 2.4 GHz band. DO
NOT use this product in locations where RF is restricted, such as
on an aircraft or in hospitals. Turn off the Bluetooth
®
feature in this
monitor and remove batteries and/or unplug the AC adapter when
in RF restricted areas. For further information on potential
restrictions refer to documentation on the Bluetooth usage by the
FCC.
AC Adapter (optional accessory) Handling and Usage
• DO NOT use the AC adapter if this monitor or the AC adapter cable
is damaged. If this monitor or the cable is damaged, turn off the
power and unplug the AC adapter immediately.
• Plug the AC adapter into the appropriate voltage outlet. DO NOT
use in a multi-outlet plug.
• NEVER plug in or unplug the AC adapter from the electric outlet
with wet hands.
• DO NOT disassemble or attempt to repair the AC adapter.
Battery Handling and Usage
• Keep batteries out of the reach of infants, toddlers or children.
6
Caution
Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury to the user or
patient, or cause damage to the equipment or other property.
• Stop using this monitor and consult with your physician if you
experience skin irritation or discomfort.
• Consult with your physician before using this monitor on an arm
where intravascular access or therapy, or an arterio-venous (A-V)
shunt, is present because of temporary interference to blood flow
which could result in injury.
• Consult with your physician before using this monitor if you have
had a mastectomy.
• Consult with your physician before using this monitor if you have
severe blood flow problems or blood disorders as cuff inflation can
cause bruising.
• DO NOT take measurements more often than necessary because
bruising, due to blood flow interference, may occur.
• ONLY inflate the arm cuff when it is applied on your upper arm.
• Remove the arm cuff if it does not start deflating during a
measurement.
• DO NOT use this monitor for any purpose other than measuring
blood pressure.
• During measurement, make sure that no mobile device or any other
electrical device that emits electromagnetic fields is within 12
inches (30 cm) of this monitor. This may result in incorrect
operation of the monitor and/or cause an inaccurate reading.
• DO NOT disassemble or attempt to repair this monitor or other
components. This may cause an inaccurate reading.
• DO NOT use in a location where there is moisture or a risk of water
splashing this monitor. This may damage this monitor.
• DO NOT use this monitor in a moving vehicle such as in a car.
• DO NOT drop or subject this monitor to strong shocks or
vibrations.
• DO NOT use this monitor in places with high or low humidity or
high or low temperatures. Refer to section 10.
7
• During measurement, observe the arm to ensure that the monitor is
not causing prolonged impairment to blood circulation.
• DO NOT use this monitor in high-use environments such as
medical clinics or physician offices.
• DO NOT use this monitor with other medical electrical (ME)
equipment simultaneously. This may result in incorrect operation of
the monitor and/or cause an inaccurate reading.
• Avoid bathing, drinking alcohol or caffeine, smoking, exercising
and eating for at least 30 minutes before taking a measurement.
• Rest for at least 5 minutes before taking a measurement.
• Remove tight-fitting, thick clothing from your arm while taking a
measurement.
• Remain still and DO NOT talk while taking a measurement.
• ONLY use the arm cuff on persons whose arm circumference is
within the specified range of the cuff.
• Ensure that this monitor has acclimated to room temperature
before taking a measurement. Taking a measurement after an
extreme temperature change could lead to an inaccurate reading.
OMRON recommends waiting for approximately 2 hours for the
monitor to warm up or cool down when the monitor is used in an
environment within the temperature specified as operating
conditions after it is stored either at the maximum or at the
minimum storage temperature. For additional information of
operating and storage/transport temperature, refer to section 10.
• DO NOT use this monitor after the durable period has ended. Refer
to section 10.
• DO NOT crease the arm cuff or the air tube excessively.
• DO NOT fold or kink the air tube while taking a measurement. This
may cause an injury by interrupting blood flow.
• To unplug the air plug, pull on the plastic air plug at the base of the
tube, not the tube itself.
• ONLY use the AC adapter, arm cuff, batteries and accessories
specified for this monitor. Use of unsupported AC adapters, arm
cuffs and batteries may damage and/or may be hazardous to this
monitor.
8
• ONLY use the approved arm cuff for this monitor. Use of other arm
cuffs may result in incorrect readings.
• Inflating to a higher pressure than necessary may result in bruising
of the arm where the cuff is applied. NOTE: refer to “If your
systolic pressure is more than 210 mmHg” in section 3 of this
instruction manual for additional information.
Data Transmission
• DO NOT replace batteries or unplug the AC adapter while your
readings are being transferred to your smart device. This may
result in incorrect operation of this monitor and failure to transfer
your blood pressure data.
AC Adapter (optional accessory) Handling and Usage
• Fully insert the AC adapter into the outlet.
• When unplugging the AC adapter from the outlet, be sure to safely
pull from the AC adapter. DO NOT pull from the AC adapter cable.
• When handling the AC adapter cable:
DO NOT damage it.
DO NOT break it.
DO NOT tamper with it.
DO NOT forcibly bend or pull it.
DO NOT twist it.
DO NOT use it if it is gathered in a bundle.
DO NOT pinch it.
DO NOT place it under heavy objects.
• Wipe any dust off of the AC adapter.
• Unplug the AC adapter when not in use.
• Unplug the AC adapter before cleaning this monitor.
9
Battery Handling and Usage
• DO NOT insert batteries with their polarities incorrectly aligned.
• ONLY use 4 “AA” alkaline or manganese batteries with this monitor.
DO NOT use other types of batteries. DO NOT use new and used
batteries together. DO NOT use different brands of batteries
together.
• Remove batteries if this monitor will not be used for a long period
of time.
• If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Consult with your physician immediately.
• If battery fluid should get on your skin, wash your skin immediately
with plenty of clean, lukewarm water. If irritation, injury or pain
persists, consult with your physician.
• DO NOT use batteries after their expiration date.
• Periodically check batteries to ensure they are in good working
condition.
Appears when viewing readings stored in the memory.
Average value symbol
Flashes when viewing the average of the latest 2 or 3
readings taken within the most recent 10 minute timeframe.
Sync symbol
Flashes/appears when your data needs to be transferred
because the stored memory is either almost, or completely
full. Once you pair your monitor with your smart device,
transfer your readings immediately before the monitor
deletes the oldest reading. Up to 60 readings can be stored
in the internal memory of your monitor.
F
G
Date/time display
H
I
Deflation symbol
Appears during cuff deflation.
Heartbeat symbol
Flashes while taking a measurement.
Bluetooth ON symbol
Appears when your readings are being transferred.
Bluetooth OFF symbol
Appears when Bluetooth is disabled.
14
High symbol
Short
Long
Irregular Heartbeat
Normal Heartbeat
J
Appears if your systolic blood pressure is ≥ 130 mmHg or
the diastolic blood pressure is ≥ 80 mmHg. Refer to subsection 1.4 for more information.
Irregular heartbeat symbol
Appears when irregular rhythm is detected 2 or more times
during a measurement. The symbol with the number
appears as below:
Irregular heartbeat is detected 2 times:
Irregular heartbeat is detected 3 times or more:
An irregular heartbeat rhythm is defined as a rhythm that is
K
while your monitor is measuring blood pressure.
25% less or 25% more than the average rhythm detected
or
Irregular heartbeat symbol may be displayed when the
monitor detects arm movements. Refer to section 7 for more
information.
If it continues to appear, we recommend you to consult
with and follow the directions of your physician.
Movement error symbol
Appears with your reading when your body is moving during
L
a measurement. If it appears, remove the arm cuff and wait
for 2 to 3 minutes. Then, reapply the cuff, remain still and
take another measurement.
15
OK symbol
M
N
Pulse display / Memory number
Pulse rate appears after the measurement.
When pressing the button, the memory number will also appear
O
for approximately 1.5 seconds before the pulse rate shows again on
the display.
Flashes when your monitor is connected to your smart
device or readings are transferred successfully.
Low battery symbol
Flashes when batteries are low.
Depleted battery symbol
Appears when batteries are depleted.
16
1.4Classification of BP (Blood Pressure)
Warning
ACC/AHA 2017 Hypertension guidelines define categories of BP in adults
as below.
Categories of BP in Adults*
BP CategorySystolic BPDiastolic BP
Normal < 120 mmHg and < 80 mmHg
Elevated120 - 129 mmHg and < 80 mmHg
Hypertension
Stage 1130 - 139 mmHg or 80 - 89 mmHg
Stage 2≥ 140 mmHg or ≥ 90 mmHg
* Individuals with Systolic BP and Diastolic BP in 2 categories should be
designated to the higher BP category.
BP indicates blood pressure (based on an average of ≥ 2 careful readings
obtained on ≥ 2 occasions).
Source: ACC/AHA 2017 High Blood Pressure Clinical Practice Guideline
• NEVER diagnose or treat yourself based on your readings. ALWAYS
consult with your physician.
17
2. Preparation
Note
2.1Installing Batteries
1. Push down the hook of the
battery cover and pull
downward.
2. Insert 4 “AA” batteries as
indicated in the battery
compartment.
3. Close the battery cover.
As soon as batteries are correctly installed, the year
flashes on the display.
Refer to sub-section 2.2 to set the date and time
manually, or refer to sub-section 4.1 to pair your
monitor to set the date and time automatically.
• When the “” symbol flashes on the display, it is recommended you
replace batteries.
• To replace batteries, turn your monitor off and remove all batteries. Then
replace with 4 new batteries at the same time.
• Replacing batteries will not delete previous readings.
• The supplied batteries may have a shorter life span than new batteries.
• Disposal of used batteries should be carried out in accordance with local
regulations.
18
2.2Setting the Date and Time
Note
Year
Month
Day
Hour
Minute
• Set your monitor to the correct date and time before taking a
measurement for the first time.
• The date and time will automatically be set when the monitor is paired
with “OMRON connect US/CAN” app. Refer to sub-section 4.1.
• As soon as batteries are correctly installed, the year flashes on the
display of your monitor.
1. If nothing is displayed, press the button to display the year.
If the year is already displayed, skip this step.
2. Press the or button to adjust the year.
Press and hold the button to quickly advance the years.
Press and hold the button to quickly go backwards through the
years.
3. Press the button to
confirm the year, then the
month will flash. Repeat
the same steps to adjust
the month, day, hour, and
minute.
19
4. Press the button to store the setting.
Note
Caution
Your monitor automatically turns off 3 seconds after the minutes have
been set in this sequence.
• If you need to reset the date and time, replace batteries or press the
button, then set the date and time as shown in above steps.
• If the date and time are not set, “” flashes during or after
measurement.
2.3Blood Pressure Measurement Tips
To help ensure an accurate reading, follow these directions:
• It is important to take measurements at the same times each day. Taking
measurements in the morning and in the evening is recommended.
• Remember to have a record of your blood pressure and pulse readings
for your physician. A single measurement does not provide an accurate
indication of your true blood pressure. You need to take and record
several readings over a period of time.
• Avoid bathing, drinking alcohol or caffeine, smoking, exercising and
eating for at least 30 minutes before taking a measurement.
• Rest for at least 5 minutes before taking a measurement.
20
2.4Applying the Arm Cuff
Note
Caution
• The following steps are for applying the arm cuff to your left arm. When
you take a measurement on your right arm, follow the cuff wrap
instructions for a right arm use mentioned under “To take a measurement
on your right arm” located at the end of this sub-section.
• The blood pressure can differ between the right arm and the left arm, and
the measured blood pressure readings can be different. OMRON
recommends to always use the same arm for measurement. If the
readings between both arms differ substantially, check with your
physician to determine as to which arm to use for your measurements.
• Remove tight-fitting, thick clothing from your arm while taking a
measurement.
1. Plug the arm cuff into your monitor
by inserting the air plug into the air
jack securely until it clicks.
Click
21
2. Place your hand through the cuff loop. Pull
Note
Bottom edge
0.5 inch (1 to 2 cm)
Bottom edge
the cuff until it reaches your upper left arm.
• The bottom edge of the arm
cuff should be 0.5 inch (1 to 2
cm) above the inside elbow.
The air tube is on the inside of
your arm and aligned with
your middle finger.
3. Make sure that the air tube
is positioned on the inside of your arm and
wrap the cuff securely, so it can not move
around your arm.
To take a measurement on your right arm
When you take a measurement on your right arm, the air tube should run
along the side of your elbow, along the bottom of your arm. Be careful not
to rest your arm on the air tube.
22
2.5Sitting Correctly
To take a measurement, you
need to be relaxed and
comfortably seated in a room at
a comfortable temperature.
Place your arm on the table.
• Sit in a comfortable chair with
your back and arm supported.
• Keep your feet flat and your
legs uncrossed.
• The arm cuff should be placed
on your arm at the same level
as your heart, with the arm
resting comfortably on a table.
23
3. Using Your Monitor without a Smart Device
Note
Caution
3.1Taking a Measurement
You can use your monitor without pairing to a smart device. To pair your
monitor with a smart device, refer to sub-section 4.1.
• To stop the measurement, press the [START/STOP] button once to
deflate the arm cuff.
• DO NOT use this monitor with other medical electrical (ME) equipment
simultaneously. This may result in incorrect operation of the monitor and/
or cause an inaccurate reading.
• Remain still and DO NOT talk while taking a measurement.
1. Press the [START/STOP] button.
All symbols appear on the display before starting the measurement.
24
2. Remain still and do not talk until the entire measurement
Note
INFLATINGSTARTCOMPLETED
process is complete.
The “” symbol flashes at every heartbeat.
After your monitor has detected your blood pressure and pulse rate,
the cuff automatically deflates. Your blood pressure and pulse rate are
displayed.
If either the systolic or the diastolic reading is high (refer to subsection 1.3), the “” symbol appears.
3. Press the [START/STOP] button to turn your monitor off.
• Your monitor automatically turns off after 2 minutes.
• Wait 2 - 3 minutes between measurements. The wait time allows the
arteries to decompress and return to their pre-measurement state.
You may need to increase the wait time depending on your
individual physiological characteristics.
25
Using the Guest Mode
Note
Note
Caution
The guest mode can be used to take a single measurement for another
user.
• No readings are stored in the memory.
1. Press and hold the [START/STOP] button for
more than 3 seconds.
The date/time display will disappear.
2. Release the [START/STOP] button when the date/time display
turns off.
The arm cuff automatically starts to inflate.
If your systolic pressure is more than 210 mmHg
If your systolic pressure is more than 210 mmHg, take a measurement
manually.
After the arm cuff starts to inflate, press and hold the [START/STOP]
button until the monitor inflates 30 to 40 mmHg higher than your expected
systolic pressure.
• Do not inflate above 299 mmHg.
• Inflating to a higher pressure than necessary may result in bruising
where the arm cuff is applied.
26
4. Using Your Monitor with a Smart Device
4.1Pairing Your Monitor with a Smart Device
The date and time on your monitor will automatically be set when you pair
your monitor with your smart device.
Review the list of compatible smart devices at OmronHealthcare.com
1. Enable Bluetooth on your smart device.
2. Download and install the free “OMRON connect US/CAN” app
onto your smart device.
If you already have the “OMRON connect US/CAN” app and have
created your account, open the app and add your new monitor.
3. Open the app and follow the pairing instructions shown on your
smart device.
4. Confirm that your monitor is connected successfully.
When your monitor is connected
successfully to your smart device, the “OK”
symbol appears.
27
5. Press [START/STOP] button to turn your monitor off.
Note
Note
• If the “Err” symbol appears, follow the instructions in
the “OMRON connect US/CAN” app.
• Your monitor automatically turns off after 2 minutes.
• Be aware that OMRON will not be responsible for the
loss of data and/or information in the app.
• “OMRON connect US/CAN” is the only app that we
recommend to use with your monitor to transfer data correctly.
4.2Taking a Measurement
Refer to sub-section 3.1 to start a measurement. Open the app on your
smart device and follow the instructions.
• Make sure that your monitor is within 16 ft. (5m) of your smart device.
Transferring Data Automatically/Manually
• The readings will be transferred to your smart device within 1 hour after
taking a measurement.
• To transfer the readings manually, press the button.
28
4.3Disabling/Enabling Bluetooth
Warning
Note
• This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT
use this product in locations where RF is restricted, such as on an
aircraft or in hospitals. Turn off the Bluetooth feature in this monitor and
remove batteries when in RF restricted areas. For further information on
potential restrictions refer to documentation on the Bluetooth usage by
the FCC.
Bluetooth is enabled as default.
1. When your monitor is off, press and hold the
button for more than 10 seconds.
“oFF” flashes 3 times.
2. Press the [START/STOP] button to turn your monitor off.
• To enable the Bluetooth, press and hold the button for more than
2 seconds. “on” appears on the display.
• Your monitor automatically turns off 2 minutes after the completion.
• When Bluetooth is disabled, the “ ” symbol appears.
29
5. Using Memory Function
Note
Note
Your monitor automatically stores up to 60 readings.
• This monitor can be used by multiple users. When you use memory
function with other users, track (or manage) your own readings
appropriately.
5.1Viewing the Readings Stored in Memory
1. Press the button.
Alternating
date/time
display
The Memory number appears for one second before the pulse rate is
displayed. The most recent reading stored in the memory will be
numbered as “1”.
• If the reading is high (refer to sub-section 1.3), the
“” symbol appears.
• If there are no readings stored in the memory, the
screen to the right is displayed.
• If the memory is full, new readings will replace the
old readings, starting with the oldest reading.
30
2. Press theorbutton repeatedly to scroll through the
Note
previous readings stored in the memory.
: To view the older readings
: To view the more recent readings
5.2Viewing the Average Reading
Your monitor can calculate an average reading based on the latest 2 or 3
readings taken within the most recent 10 minute timeframe.
1. When your monitor is off, press and hold the
button for more than 2 seconds.
The “” symbol flashes.
• If there are only 2 readings in the memory for that
period, the average will be based on these 2
readings.
• If the average reading is high (refer to subsection 1.3), the “” symbol appears.
31
6. Other Settings
Note
6.1Deleting All Readings Stored in Memory
1. Press the button.
After a memory number appears, the latest reading will appear.
2. While holding the button
down, press and hold the
[START/STOP] button for
more than 3 seconds.
The display to the right flashes.
3. While holding the button
down, press and hold the
[START/STOP] button for
more than 3 seconds.
4. Press [START/STOP] button to turn your monitor off.
• All readings will be deleted. You cannot partially delete the readings
stored in the memory.
• Your monitor automatically turns off after 2 minutes.
32
6.2Restoring Your Monitor to the Default Settings
Note
To delete all the information stored in your monitor, follow the instructions
below. Make sure that your monitor is turned off.
1. While holding the
button down, press
and hold the [START/
STOP] button for more
than 3 seconds.
The display to the right
flashes.
2. While holding the
button down, press
and hold the [START/
STOP] button for more
than 3 seconds.
3. Press [START/STOP] button to turn your monitor off.
• Reverting to the default setting of your monitor does not delete the
information in the app.
• Your monitor automatically turns off after 2 minutes.
• You will need to re-pair your monitor when using the monitor again.
Without re-pairing, your readings will not be transferred to your app.
33
7. Error Messages and Troubleshooting
If any of the below problems occur during measurement, check to make
sure that no other electrical device is within 12 inches (30 cm). If the
problem persists, refer to the table below.
Display/ProblemPossible CauseSolution
appears or the arm
cuff does not inflate.
Air plug is not completely
plugged into the monitor.
The arm cuff is not
applied correctly.
Air is leaking from the
arm cuff.
34
Insert the air plug
securely. Refer to subsection 2.4.
Apply the arm cuff
correctly, then take
another measurement.
Refer to sub-section 2.4.
Replace the arm cuff
with a new one. Refer to
section 9.
Display/ProblemPossible CauseSolution
Remain still and do not
talk during a
measurement. If “E2”
appears repeatedly,
inflate the arm cuff
manually until the
systolic pressure is 30 to
40 mmHg above your
previous readings. Refer
to sub-section 3.1.
Inflate the arm cuff
manually until the
systolic pressure is 30 to
40 mmHg above your
previous readings. Refer
to sub-section 3.1.
Do not touch the arm cuff
and/or bend the air tube
while taking a
measurement.
If inflating the arm cuff
manually, refer to the
end of sub-section 3.1.
appears or a
measurement
cannot be complete
after the arm cuff
inflates.
appears
You move or talk during
a measurement and the
arm cuff does not inflate
sufficiently.
The systolic pressure is
above 210 mmHg and a
measurement cannot be
taken.
The arm cuff is inflated
exceeding the maximum
allowable pressure.
appears
You move or talk during
a measurement.
Vibrations disrupt a
measurement.
35
Remain still and do not
talk during a
measurement.
Display/ProblemPossible CauseSolution
Apply the arm cuff
correctly, then take
appears
/
appears
does not flash during
a measurement
appears
appears
The pulse rate is not
detected correctly.
The monitor has
malfunctioned.
The monitor cannot
connect to a smart
device or transmit data
correctly.
another measurement.
Refer to sub-section 2.4.
Remain still and sit
correctly during a
measurement.
If the “ ” symbol
continues to appear, we
recommend you to
consult with your
physician.
Press the [START/
STOP] button again. If
“Er” still appears, contact
customer service. Refer
to section 12.
Follow the instructions
shown in the “OMRON
connect US/CAN” app. If
the “Err” symbol still
appears after checking
the app, contact
customer service. Refer
to section 12.
flashes
The monitor is waiting for
pairing with the smart
device.
36
Refer to sub-section 4.1
for pairing your monitor
with your smart device.
Display/ProblemPossible CauseSolution
flashes
flashes
appears
flashes
The monitor is ready to
transfer your readings to
the smart device.
There are more than 40
readings in memory to
be transferred.
The date and time is not
set.
There are 60 readings in
memory to be
transferred.
Batteries are low.
Open the “OMRON
connect US/CAN” app to
transfer your readings.
Pair or transfer your
readings to the “OMRON
connect US/CAN” app so
you can keep them in
memory in the app, and
this error symbol
disappears.
Replacing all batteries
with 4 new batteries is
recommended. Refer to
sub-section 2.1.
37
Display/ProblemPossible CauseSolution
appears or the
monitor is turned off
unexpectedly during
a measurement.
Nothing appears on
the display of the
monitor.
Readings appear too
high or too low.
Any other
communication
issue occurs.
Any other problems
occur.
Batteries are depleted.
Batteries are depleted.
Battery polarities are not
properly aligned.
Blood pressure varies constantly. Many factors
including stress, time of day, and/or how you apply
the arm cuff, may affect your blood pressure.
Review sub-sections 2.3 - 2.5 and section 3.
Follow the instructions shown in the smart device,
or visit the “Support” section in the “OMRON
connect US/CAN” app for further help. If the
problem still persists, contact customer service.
Refer to section 12.
Press the [START/STOP] button to turn the monitor
off, then press it again to take a measurement. If
the problem continues, remove all batteries and
wait for 30 seconds. Then re-install batteries.
If the problem still persists, contact customer
service. Refer to section 12.
Immediately replace all
batteries with 4 new
batteries. Refer to subsection 2.1.
Check the battery
installation for proper
placement. Refer to subsection 2.1.
38
8. Maintenance
Caution
Caution
8.1Maintenance
To protect your monitor from damage, follow the directions below:
• Changes or modifications not approved by the manufacturer will void
the user warranty.
• DO NOT disassemble or attempt to repair this monitor or other
components. This may cause an inaccurate reading.
8.2Storage
Store your monitor and other components in a clean, safe location.
1. Remove the arm cuff from the monitor.
• To unplug the air plug, pull on the plastic air plug at the base of the tube,
not the tube itself.
39
2. Gently fold the air tube into the arm
Note
cuff.
• Do not bend or crease the air tube
excessively.
Do not store your monitor and other components:
• If your monitor and other components are wet.
• In locations exposed to extreme temperatures, humidity, direct
sunlight, dust or corrosive vapors such as bleach.
• In locations exposed to vibrations or shocks.
8.3Cleaning
• Do not use any abrasive or volatile cleaners.
• Use a soft dry cloth or a soft cloth moistened with mild (neutral) detergent
to clean your monitor and arm cuff and then wipe them with a dry cloth.
• Do not wash or immerse your monitor and arm cuff or other components
in water.
• Do not use gasoline, thinners or similar solvents to clean your monitor
and arm cuff or other components.
8.4Disposal
Dispose of your monitor, other components and optional accessories
according to applicable local regulations.
Unlawful disposal may cause environmental pollution.
40
9. Optional Accessories
Note
Small D-Ring CuffWide-Range D-Ring Cuff
Arm circumference
7" - 9" (17 - 22 cm)
Arm circumference
9" - 17" (22 - 42 cm)
Model:
HEM-CS24-B HEM-CS24-BZ
AC Adapter
Model: HEM-ADPTW5
• Do not throw the air plug away. The air plug can be applicable to the
optional cuff.
HEM-RML31-B HEM-RML31-BZ
For further information, visit our
website at OmronHealthcare.com
41
Model:
Using the AC Adapter (optional accessory)
Note
Caution
• Make sure not to place your monitor in a location where it is difficult to
plug and unplug the AC adapter.
• We recommend keeping batteries in your monitor at all times, even if you
choose to use the AC adapter. If only the AC adapter is used without
keeping batteries in your monitor, you may need to reset the date and
time each time you unplug and plug back the AC adapter. The readings
will not be deleted.
1. Insert the AC adapter plug into
the AC adapter jack on the rear
side of your monitor.
2. Plug the AC adapter into an
electrical outlet.
To unplug the AC adapter, unplug
the AC adapter from the electrical
outlet and then remove the AC adapter plug from the monitor.
• Only use the AC adapter, the arm cuff, batteries and accessories
specified for this monitor. Use of unsupported AC adapters, arm cuffs
and batteries may damage and/or may be hazardous to this monitor.
42
10. Specifications
Model
DisplayLCD digital display
Cuff pressure range0 to 299 mmHg
Blood pressure
measurement range
Pulse measurement
range
Accuracy
InflationAutomatic by electric pump
DeflationAutomatic pressure release valve
Measurement methodOscillometric method
Internally powered ME equipment (When using
only batteries)
Class II ME equipment (Optional AC adapter)
44
Note
• These specifications are subject to change without notice.
• In the clinical validation study, K5 was used on 85 subjects for
determination of diastolic blood pressure.
• This monitor complies with the requirements of ISO 81060-2:2013.
• This monitor has not been validated for use on pregnant patients.
• IP classification is degrees of protection provided by enclosures in
accordance with IEC 60529. This monitor and optional AC adapter are
protected against solid foreign objects of 0.5 inches (12.5 mm) diameter
and greater such as a finger. The optional AC adapter is protected
against vertically falling water drops which may cause issues during a
normal operation.
About a wireless communication interference
This product operates in an unlicensed ISM band at 2.4 GHz. In the event
this product is used near other wireless devices such as microwave and
wireless LAN, which operate on the same frequency band as this product,
there is a possibility that interference may occur.
If interference occurs, stop the operation of the other devices or relocate
this product away from other wireless devices before attempting to use it.
Pictogram on the Product
Types of marks used to indicate the correct positioning
of the artery for the upper arm
Identifies the type of cuff that is compatible with the
blood pressure monitor
45
11. FCC Statement
Note
FCC CAUTION
Changes or modifications not expressly approved by the party responsible
for compliance could void the user's authority to operate the equipment.
This device complies with part 15 of the FCC Rules. Operation is subject
to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
• This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses
and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to
radio communications. However, there is no guarantee that interference
will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is encouraged
to try to correct the interference by one or more of the following
measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
46
This transmitter must not be co-located or operated in conjunction with any
other antenna or transmitter.
This equipment complies with FCC radiation exposure limits set forth for
an uncontrolled environment and meets the FCC radio frequency (RF)
Exposure Guidelines. This equipment has very low levels of RF energy
that are deemed to comply without testing of specific absorption ratio
(SAR).
47
12. Limited Warranty
Your BP7250 Blood Pressure Monitor, excluding the arm cuff and batteries
is warranted to be free from defects in materials and workmanship
appearing within 2 years from the date of purchase, when used in
accordance with the instructions provided with the monitor. The arm cuff is
warranted to be free from defects in materials and workmanship appearing
within 1 year from the date of purchase when the monitor is used in
accordance with the instructions provided with the monitor. The above
warranty extends only to the original retail purchaser.
We will, at our option, replace without charge any monitor or arm cuff
covered by the above warranty. Replacement is our only responsibility and
your only remedy under the above warranty.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the address of the inspection center and the return shipping and
handling fee.
Enclose the original printed receipt. Include a letter, with your name,
address, phone number, and description of the specific problem. Pack the
product carefully to prevent damage in transit. Because of possible loss in
transit, we recommend insuring the product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON
IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY
DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES
AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE
LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS
WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
48
This warranty provides you with specific legal rights, and you may have
other rights that vary by jurisdiction. Because of special local
requirements, some of the above limitations and exclusions may not apply
to you.
FOR CUSTOMER SERVICE
Visit our web site at:OmronHealthcare.com
Call toll free:1-800-634-4350
49
13. Guidance and Manufacturer’s Declaration
OMRON Battery-operated / AC Adapter-operated Blood Pressure Monitor
Information for Accompanying Documents in the Scope of IEC60601-12:2014
Important information regarding Electromagnetic Compatibility
(EMC)
BP7250 manufactured by OMRON HEALTHCARE Co., Ltd. conforms to IEC60601-12:2014 Electromagnetic Compatibility (EMC) standard.
Further documentation in accordance with this EMC standard is available at
OmronHealthcare.com/emc. Refer to the EMC information for BP7250 on the website.
50
The Bluetooth® word mark and logos are registered trademarks owned by the
Bluetooth SIG, Inc. and any use of such marks by OMRON HEALTHCARE Co.,
Ltd. is under license. Other trademarks and trade names are those of their
respective owners.
App Store and App Store logo are service marks of Apple Inc., registered in the
U.S and other countries.
Google Play and the Google Play logo are trademarks of Google LLC.