INTRODUCTION
Thank you for purchasing the OMRON BP710NVA Blood Pressure Monitor.
Your new blood pressure monitor uses the oscillometric method of blood pressure
measurement. This means the monitor detects your blood movement through your
brachial artery and converts the movements into a digital reading. An oscillometric
monitor does not need a stethoscope so the monitor is simple to use.
Intended Use
This device is a digital monitor intended for use in measuring blood pressure and
pulse rate in adult patient population. The device detects the appearance of
irregular heartbeats during measurement and gives a warning signal with the
measurement result.
Please read this instruction manual thoroughly before using the device.
Please keep for future reference. For specific information about your own blood
pressure, CONSULT YOUR PHYSICIAN.
3
Page 4
IMPORTANT SAFETY INFORMATION
Warning: Indicates a potentially hazardous situation which, if not avoided,
(General Usage)
DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your physician. Only a
physician is qualified to diagnose and treat High Blood Pressure.
The monitor is not intended to be a diagnostic device.
Consult your physician before using the device for any of the following
conditions: common arrhythmias such as atrial or ventricular premature beats or
atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age, pregnancy,
pre-eclampsia, renal diseases.
Note that PATIENT motion, trembling, shivering may affect the measurement
reading.
Do not use the device on the injured arm or the arm under medical treatment.
Do not apply the arm cuff on the arm while being on an intravenous drip or blood
transfusion.
Consult your physician before using the device on the arm with an arterio-venous
(A-V) shunt.
Do not use the device with other medical electrical (ME) equipment
simultaneously.
Do not use the device in the area the HF surgical equipment, MRI, or CT scanner
exists, or in the oxygen rich environment.
The air tube may cause accidental strangulation in children, toddlers or infants.
Contains small parts that may cause a choking hazard if swallowed by children,
toddlers or infants.
(Battery Usage)
Keep the battery out of reach of children, toddlers or infants.
could result in death or serious injury.
4
Page 5
IMPORTANT SAFETY INFORMATION
Caution: Indicates a potentially hazardous situation which, if not avoided,
(General Usage)
Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
Consult your physician before using the device if you have had a mastectomy.
Consult your physician before using the device if you have severe blood flow
problems or blood disorders as cuff inflation can cause bruising.
Do not take measurements more often than necessary. It may cause bruising
due to blood flow interference.
Remove the arm cuff if it does not start deflating during the measurement.
Do not use this device on infants or persons who cannot express their intentions.
Do not use the device for any purpose other than measuring blood pressure.
Use only the approved arm cuff for this device. Use of other arm cuffs may result
in incorrect measurement results.
Do not use a mobile phone or other devices that emit electromagnetic fields,
near the device. This may result in incorrect operation of the device.
Do not disassemble the monitor or arm cuff. This may cause an inaccurate
reading.
Do not use in a location with moisture, or a location where water may splash on
the device. This may damage the device.
Do not use the device in a moving vehicle (car, airplane).
Read “If your systolic pressure is more than 210 mmHg” of this instruction
manual, if your systolic pressure is known to be more than 210 mmHg. Inflating
to a higher pressure than necessary may result in bruising where the arm cuff is
applied.
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
5
Page 6
IMPORTANT SAFETY INFORMATION
(Battery Usage)
Do not insert the batteries with their polarities incorrectly aligned.
Use only 4 “AA” alkaline or manganese batteries with this device. Do not use
other types of batteries. Do not use new and used batteries together.
Remove the batteries if the device will not be used for three months or more.
General Precautions
• Do not forcibly crease the arm cuff or the air tube excessively.
• Do not press the air tube while taking a measurement.
• To unplug the air plug, pull on the air plug at the connection with the monitor, not
the tube itself.
• Do not drop the monitor or subject device to strong shocks or vibrations.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not use the device outside the specified environment. It may cause an
inaccurate reading.
• Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental pollution.
6
Page 7
1. KNOW YOUR DEVICE
D
E
A
C
B
F. Arm cuff
(Arm circumference 9" - 17" (22 - 42 cm))
G. Air pl ug
H. Air tube
Contents:
Monitor, arm cuff, battery set, instruction manual, quick start guide
Monitor:
A. Display
B. START/STOP button
C. Memory button
Arm cuff:
D. Battery compartment
E. Air jack
F
G
H
7
Page 8
Display:
1. KNOW YOUR DEVICE
I
J
K
L
N
O
P
M
I. Memory symbol
J. Systolic blood pressure
K. Diastolic blood pressure
L. Low battery symbol
M. Heartbeat symbol
(Flashes during measurement.)
Q
N. Irregular heartbeat symbol
O. Movement error symbol
P. Pulse display / Memory number
Q. Deflation symbol
8
Page 9
1. KNOW YOUR DEVICE
1.1 Display symbols
Irregular Heartbeat Symbol ( )
When the monitor detects an irregular rhythm
two or more times during the measurement,
the irregular heartbeat symbol will appear on
the display with the measurement values.
An irregular heartbeat rhythm is defined as a
rhythm that is 25% less or 25% more than the
average rhythm detected while the monitor is
measuring the systolic and diastolic blood
pressure.
If the irregular heartbeat symbol displays with your measurement results, we
recommend you consult your physician. Follow the directions of your physician.
Blood pressure
Blood pressure
Normal Heartbeat
Pulse
Irregular Heartbeat
Pulse
Short
Long
Movement Error Symbol ( )
The movement error symbol is displayed if you move your body during the
measurement. Please remove the arm cuff, and wait 2 - 3 minutes.
Take another measurement, remain still during measurement.
2013 ESH/ESC Guidelines for the management of arterial
hypertension
Definitions of hypertension by office and home blood pressure levels
OfficeHome
Systolic Blood Pressure 140 mmHg 135 mmHg
Diastolic Blood Pressure 90 mmHg 85 mmHg
These are from statistical values for blood pressure.
9
Page 10
1. KNOW YOUR DEVICE
1.2 Before Taking a Measurement
To help ensure an accurate reading, follow these directions:
1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for
30 minutes before taking a measurement. Rest for at least 15 minutes before
taking the measurement.
5. Sit on a chair with your feet flat on the floor. Rest your arm on a table so that the
arm cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Keep a record of your blood pressure and pulse readings for your physician.
A single measurement does not provide an accurate indication of your true blood
pressure. You need to take and record several readings over a period of time.
Try to measure your blood pressure at the same time each day for consistency.
10
Page 11
2. PREPARATION
2.1 Battery Installation
1. Remove the battery cover.
2. Insert 4 “AA” batteries as indicated
in the battery compartment.
3. Replace the battery cover.
Notes:
• When the low battery symbol () appears on the display, turn the monitor
off, then replace all batteries at the same time. Long life alkaline batteries are
recommended.
• The measurement values continue to be stored in memory even after the
batteries are replaced.
• Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
11
Page 12
3. USING THE DEVICE
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm.
Do not place the arm cuff over thick clothes.
1. Insert the air plug into the air jack
securely.
2. Wrap the arm cuff firmly in place
around your left upper arm.
The bottom edge of the arm cuff should
be 1/2 inch (1 to 2 cm) above the elbow.
Air tube is on the inside of your arm and
aligned with your middle finger.
1/2 inch
(1 - 2 cm)
12
Page 13
3. USING THE DEVICE
3. Secure closed with the
fabric fastener.
Notes:
• When you take a measurement on the right
arm, the air tube will be at the side of your
elbow. Be careful not to rest your arm on the air
tube.
• The blood pressure can differ between the right arm and the left arm, and the
measured blood pressure values can be different. OMRON recommends to
always use the same arm for measurement. If the values between both arms
differ substantially, please check with your physician which arm to use for your
measurements.
13
Page 14
3. USING THE DEVICE
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortable room temperature. Avoid
bathing, drinking alcohol or caffeine,
smoking, exercising or eating 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• Sit with your back and arm being supported.
• The arm cuff should be placed on your arm at
the same level as your heart.
14
Page 15
3. USING THE DEVICE
3.3 Taking a Measurement
Notes:
• To stop the measurement, press the START/STOP button once to deflate the
arm cuff.
• Remain still and quiet while taking a measurement.
15
Page 16
3. USING THE DEVICE
STARTINFLATEDEFLATECOMPLETE
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/STOP
button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
Inflating to a higher pressure than necessary may result in bruising
where the arm cuff is applied.
1. Press the START/STOP button.
The arm cuff will start to inflate automatically.
16
Page 17
3. USING THE DEVICE
2. Remove the arm cuff.
3. Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement result in its memory.
It will automatically turn off after 2 minutes.
Note: Wait 2-3 minutes before taking another measurement. Waiting between
measurements allows the arteries to return to the condition prior to taking
a measurement.
DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your physician. Only a
physician is qualified to diagnose and treat High Blood Pressure.
This device is not intended to be a diagnostic device.
Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
Inflating to a higher pressure than necessary may result in bruising where the
arm cuff is applied.
17
Page 18
3. USING THE DEVICE
3.4 Using the Memory Function
The monitor automatically stores the results up to 14 sets.
Note: If the memory is full, the monitor will delete the oldest value.
To View the Measurement Values Stored in Memory
1. Press the button.
The Memory number appears for a second
before the pulse rate is displayed.
The newest set is numbered “1”.
2. Press the button repeatedly to view the values stored in
memory.
18
Page 19
3. USING THE DEVICE
To Delete All the Values Stored in Memory
1. Press the button, while the memory symbol appears.
2. While holding the button down,
press the START/STOP button for
more than 3 seconds.
Note: You cannot partially delete the values stored in the memory.
19
Page 20
4. ERROR MESSAGES AND TROUBLESHOOTING
4.1 Error Messages
Error DisplayCauseSolution
Remove the arm cuff. Wait 2 - 3
Irregular heartbeats are
detected.
minutes and then take another
measurement. Repeat the steps
in section 3.3. If this error
continues to appear, contact
your physician.
Movement during measurement.
The batteries are low.
The batteries are exhausted.
Air plug disconnected.
Arm cuff is applied too loosely.
Air is leaking from the arm cuff.
20
Carefully read and repeat the
steps in section 3.3.
You should replace the batteries
with new ones ahead of time.
Refer to section 2.1.
You should replace the batteries
with new ones at once.
Refer to section 2.1.
Insert the plug securely.
Refer to section 3.1.
Apply the arm cuff tighter.
Refer to section 3.1.
Replace the arm cuff with a new
one.
Refer to section 5.3.
Page 21
4. ERROR MESSAGES AND TROUBLESHOOTING
Error DisplayCauseSolution
Repeat measurement. Remain
still and do not talk during
measurement.
Movement during measurement
and the arm cuff has not been
inflated sufficiently.
The arm cuff was inflated
exceeding the maximum
allowable pressure, and then
deflated automatically when
inflating the arm cuff manually.
Movement during measurement.
Clothing is interfering with the
arm cuff.
Device error.Contact Customer Service.
Refer to section 3.3.
If “E2” appears repeatedly,
inflate the arm cuff manually
until it is 30 to 40 mmHg above
your previous measurement
result.
Refer to section 3.3.
Do not touch the arm cuff and/or
bend the air tube while taking a
measurement. Do not inflate the
arm cuff more than necessary.
Refer to section 3.3.
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
Remove any clothing interfering
with the arm cuff.
Refer to section 3.1.
21
Page 22
4. ERROR MESSAGES AND TROUBLESHOOTING
4.2 Troubleshooting
ProblemCause and Solution
No power.
No display appears on the monitor.
Measurement values appear too high or
too low.
Replace all batteries with new ones.
Check the battery installation for proper
placement of the battery polarities.
Refer to section 2.1.
Blood pressure varies constantly. Many
factors including stress, time of day, and
how you wrap the cuff, may affect your
blood pressure. Review the section 1.2
and section 3.3.
22
Page 23
5. MAINTENANCE AND STORAGE
5.1 Maintenance
To protect your device from damage, please observe the following:
• Store the device and the components in a clean, safe location.
• Do not use any abrasive or volatile cleaners.
• Do not wash the device and any components or immerse them in
water.
• Do not use gasoline, thinners or similar solvents to clean the
device.
• Use a soft and dry cloth, or a soft and moistened cloth and neutral
soap to clean on the monitor and the arm cuff.
• Changes or modification not approved by the manufacturer will
void the user warranty. Do not disassemble or attempt to repair
the device or components. Consult Customer Service.
23
Page 24
5. MAINTENANCE AND STORAGE
5.2 Storage
1. Unplug the air plug from the air jack.
2. Gently fold the air tube into the arm cuff.
Note: Do not bend or crease the air tube
excessively.
Do not store the device in the following situations:
• If the device is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight, dust or
corrosive vapors such as bleach.
• Locations exposed to vibrations, shocks or where it will be at an angle.
24
Page 25
5. MAINTENANCE AND STORAGE
5.3 Optional Medical Accessories
Arm circumference
7" - 9" (17 - 22 cm)
CD-CS9
(Model: HEM-CS24)
Arm cuff
Arm circumference
9" - 17" (22 - 42 cm)
CD-WR17
(Model: HEM-RML31)
25
Page 26
6. SPECIFICATIONS
Model
DisplayLCD digital display
Measurement rangePressure: 0 to 299 mmHg
AccuracyPressure: r3 mmHg or 2% of reading
InflationFuzzy-logic controlled by electric pump
DeflationAutomatic pressure release valve
Measurement method Oscillometric method
IP classificationIP 20
Power source4 “AA” batteries 1.5V
Battery lifeApproximately 1000 measurements (using new alkaline batteries)
Operating temperature /
humidity / air pressure
Storage temperature /
humidity / air pressure
WeightMonitor : Approximately 8 7/8 oz. (250 g) not including batteries
DimensionsMonitor : Approximately 4" (w) u 3 1/8" (h) u 5 1/8" (l)
Cuff circumference9" to 17" (22 to 42 cm)
MemoryUp to 14
ContentsMonitor, arm cuff, battery set, instruction manual, quick start guide
Applied part
Protection against
electric shock
BP710NVA HEM-7121-ZVA
Pulse: 40 to 180 beats / min.
Pulse: r 5% of display reading
50°F to 104°F (10°C to 40°C) / 15 to 90% RH / 700 to 1060 hPa
-4°F to 140°F (-20°C to 60°C) / 10 to 95% RH / 700 to 1060 hPa
Arm cuff : Approximately 6 oz. (170 g)
Arm cuff : Approximately 5 3/4" u 23 1/2" (air tube: 29 1/2")
= Type BF
Internally powered ME equipment
(103 mm u 80 mm u 129 mm)
(145 mm u 594 mm (air tube: 750 mm))
Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, the 5th phase was used on 85 subjects for
determination of diastolic blood pressure.
• This device has not been validated for use on pregnant patients.
26
Page 27
7. FCC STATEMENT
FCC CAUTION
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
Note:
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference in a residential
installation. This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
27
Page 28
8. LIMITED WARRANTY
Your BP710NVA Blood Pressure Monitor, excluding the arm cuff and batteries, is
warranted to be free from defects in materials and workmanship appearing within
2 years from the date of purchase, when used in accordance with the instructions
provided with the monitor. The arm cuff is warranted to be free from defects in
materials and workmanship appearing within 1 year from the date of purchase when
the monitor is used in accordance with the instructions provided with the monitor.
The above warranty extends only to the original retail purchaser.
We will, at our option, replace without charge any monitor or arm cuff covered by the
above warranty. Replacement is our only responsibility and your only remedy under
the above warranty.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for
the address of the inspection center and the return shipping and handling fee.
Enclose the original printed receipt. Include a letter, with your name, address, phone
number, and description of the specific problem. Pack the product carefully to
prevent damage in transit. Because of possible loss in transit, we recommend
insuring the product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE
IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF
THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL,
INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR
DAMAGES.
28
Page 29
8. LIMITED WARRANTY
This warranty provides you with specific legal rights, and you may have other rights
that vary by jurisdiction. Because of special local requirements, some of the above
limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE
Visit our web site at:www.OmronHealthcare.com
Call toll free:1-800-634-4350
29
Page 30
9. GUIDANCE AND MANUFACTURER’S DECLARATION
OMRON Blood Pressure Monitor (BPM)
Information for accompanying documents in the scope of IEC60601-1-2:2007
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular)
telephones, medical devices in use may be susceptible to electromagnetic interference from
other devices. Electromagnetic interference may result in incorrect operation of the medical
device and create a potentially unsafe situation. Medical devices should also not interfere with
other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to
prevent unsafe product situations, the IEC60601-1-2 standard has been implemented. This
standard defines the levels of immunity to electromagnetic interferences as well as maximum
levels of electromagnetic emissions for medical devices.
Medical devices manufactured by OMRON Healthcare conform to this IEC60601-1-2:2007
standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• The use of accessories and cables other than those specified by OMRON, with the exception
of cables sold by OMRON as replacement parts for internal components, may result in
increased emission or decreased immunity of the device.
• The medical devices should not be used adjacent to or stacked with other equipment.
In case adjacent or stacked use is necessary, the medical device should be observed to
verify normal operation in the configuration in which it will be used.
• Refer to further guidance below regarding the EMC environment in which the device should
be used.
• The MEDICAL ELECTRICAL EQUIPMENT BPM needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC information provided in
this documentations.
• The Essential Performance of the BPM is to measure a blood pressure and a pulse rate and
using the memory function.
The BPM may be interfered with by other equipment, even if that other equipment complies
with CISPR EMISSION requirements.
30
Page 31
9. GUIDANCE AND MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration - electromagnetic emissions
OMRON BPM is intended for use in the electromagnetic environment specified below. The customer
or the user of this OMRON BPM should assure that it is used in such environment.
The OMRON BPM uses RF energy only for its internal
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC61000-3-3
Group 1
Class B
Not Applicable.
Not Applicable.
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
The OMRON BPM is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
31
Page 32
9. GUIDANCE AND MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BPM is intended for use in the electromagnetic environment specified below. The customer
or the user of this OMRON BPM should assure that it is used in such environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply inputlines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
T is the A.C. mains voltage prior to application of the test level.
Note: U
IEC 60601 test
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
T (>95 % dip
<5 % U
T)
in U
for 0.5 cycle
40 % U
T (60 % dip
in U
T)
for 5 cycles
T (30 % dip
70 % U
T)
in U
for 25 cycles
<5 % U
T (>95 % dip
in U
T)
for 5 sec.
3 A/m3 A/m
Compliance level
±6 kV contact
±8 kV air
Not Applicable.Not Applicable.
Not Applicable.Not Applicable.
Not Applicable.Not Applicable.
Electromagnetic environment -
Floor should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
guidance
32
Page 33
9. GUIDANCE AND MANUFACTURER’S DECLARATION
P
P
OMRON BPM is intended for use in the electromagnetic environment specified below. The customer
or the user of this OMRON BPM should assure that it is used in such environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Guidance and manufacturer’s declaration - electromagnetic immunity
IEC 60601
test level
3 V rms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
Compliance
level
Not
Applicable.
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the OMRON
BPM including cables, than the recommended
separation distance calculated from the equation
appropriate to the frequency of the transmitter.
Recommend separation distance
Not Applicable.
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
b
a
33
Page 34
9. GUIDANCE AND MANUFACTURER’S DECLARATION
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones
and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in
which the OMRON BPM is used exceeds the applicable RF compliance level above, the OMRON
BPM should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the OMRON BPM.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
34
Page 35
9. GUIDANCE AND MANUFACTURER’S DECLARATION
P
P
Recommended separation distance between portable and mobile RF communications
OMRON BPM is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of this OMRON BPM can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the OMRON BPM as recommended below, according
to the maximum output power of the communications equipment.
Output Power of Transmitter in
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
Watt
0.01
0.10.380.73
11.22.3
103.87.3
1001223
absorption and reflection from structures, objects, and people.
equipment
and the OMRON BPM
Separation distance according to frequency of transmitter