USB Output Sanitary Fitting Pressure Transducer
Connect Directly to Your PC or Laptop
PX409S Sanitary USB Series
Gage, Absolute, Compound Gage,
Vacuum and Barometric Pressures
10 inH2O to 600 psi (25 mb to 41 bar)
Starts at
PX409S15-015CGUSB,
1.5" flange, ±15 psi,
USB output, $685.
$
585
Fast Delivery!
Stock to 1 Week
on Most Models
Standard
74-03
⻬ Micromachined Silicon Sensor
⻬ 316L SS Wetted Parts
⻬ High 0.08% Accuracy Typ.
(0.14% Max)
⻬ Excellent Long Term Stability
⻬ USB 2.0 and Below Compatible
⻬
Standard USB Cable Termination
⻬ Ruggedized with Secondary
Containment System
⻬ Select Engineering Units
via Software
The PX409S Sanitary USB Series
connects directly to your computer.
The Sanitary Series is 3A
approved and available with
1
1
⁄2 or 2" Tri-GripTM(Tri-Clamp
compatible) fittings to work with
most common applications. Free
software makes data logging and
charting your readings a simple
task. The micromachined silicon
design is ideal for pressure or level
applications in laboratory, test
platforms, or bio/pharmaceutical
applications as well as industrial
applications that require a rugged,
high accuracy transducer. The
micromachined silicon sensor
provides a very stable transducer
with exceptional high accuracy of
®
2 m (6') integral
USB cable connects
directly to your PC.
0.08% typical (0.14% max) and a
broad compensated range of 4 to
60°C (40 to 140°F). The modular
construction allows for fast delivery
of most configurations and fittings.
Delivery is typically stock to 1 week!
Specifications
Accuracy: 0.08% typical best
straight line, 0.14% max (linearity,
hysteresis and repeatability
combined) compound gage models
calibrated in positive direction only
Calibration: 5-point NIST traceable
calibration with zero and span values
calibrated in horizontal direction
Zero Balance: ±0.5% FS typical
±1% max (for ranges ≤2.5 psi ±1%
typical ±2% max)
Span Setting: ±0.5% FS typical
±1% max (for ranges ≤2.5 psi ±1%
typical ±2% max)
Operating Temperature Range:
-15 to 95°C (-5 to 203°F)
Compensated Temperature Range:
Ranges ≤ 5 psi: 4.4 to 60°C
(40 to 140°F)
Ranges > 5 psi: -4 to 85°C
(25 to 185°F)
PX409S20-015GUSB,
2" flange, 0 to 15 psi,
USB output, $585.
Temperature Compensation
Zero and Span Shift Over
Compensated Range:
Ranges ≤2.5 psi:
Zero: 1.00%
Span: 0.70%
Range = 5 psi:
Zero: 0.80%
Span: 0.50%
Ranges > 5psi:
Zero: 0.50%
Span: 0.50%
26B/26HB Barometric Range:
Zero: 0.90%
Span: 0.50%
Minimum Isolation Between Case
and Output Terminations:
100M Ω @ 50 Vdc
Pressure Cycles: 250,000, minimum
Long Term Stability (1-Year):
±0.1% FS typical
Shock: 50 g, 11 mS half sine
shock, (under test)
Vibration: ±20 g (under test)
FREE SOFTWARE INCLUDED!
Each unit includes free software that
converts your PC into a strip chart
recorder or data logger so readings can
be saved and later printed or exported
to a spread sheet file. Also included are
software drivers for some of the most
popular Laboratory programs
SHOP
ONLINE AT
.
order online, visit omega.com/sanitary
Laptop not
included.
To download information and to
1
ø22.25
(0.88)
ø50.39
(1.98)
PX419S15 1½" flange
60.97 (2.40)
5.21 (0.20)
2 m
(6')
2
Bandwidth: DC to 3 Hz typical
A to D Conversion: 14 bit
Response Time: <1 mS
Fill Fluid: NEOBEE M5 food grade
Approved to 3A Sanitary
Specification 74-03
CE Compliant: Meets
industrial emissions and
immunity standard
IEC61326
Environmental
Protection: IP65
Pressure Rating:
1
⁄2 and 2" Tri-Grip
1
TM
Fitting: 600 psi using
suitable clamp
Dimensions: mm (in)
Ø = Diameter
Sanitary product line continues to expand,
visit omega.com/sanitary for details!
Over Pressure: 4 x rated pressure
to maximum rated flange pressure
Wetted Parts: 316L SS
Weight: 285 g (10 oz)
Secondary Containment: 6 x rated
pressure to a maximum of 2000 psi
(not in clamp)
MOST POPULAR
MODELS HIGHLIGHTED!
To Order
Range Range 1.5" Tri-Grip 2.0" Tri-Grip
psi bar Models Models Price
Gage Pressure
0 to 10 inH2O 0 to 25 mb PX409S15-10WGUSB PX409S20-10WGUSB $635
0 to 1 0 to 69 mb PX409S15-001GUSB PX409S20-001GUSB 635
0 to 2.5 0 to 172 mb PX409S15-2.5GUSB PX409S20-2.5GUSB 635
0 to 5 0 to 345 mb PX409S15-005GUSB PX409S20-005GUSB 635
0 to 15 0 to 1 PX409S15-015GUSB PX409S20-015GUSB 585
0 to 30 0 to 2.1 PX409S15-030GUSB PX409S20-030GUSB 585
0 to 50 0 to 3.4 PX409S15-050GUSB PX409S20-050GUSB 585
0 to 100 0 to 6.9 PX409S15-100GUSB PX409S20-100GUSB 585
0 to 150 0 to 10.3 PX409S15-150GUSB PX409S20-150GUSB 585
0 to 250 0 to 17.2 PX409S15-250GUSB PX409S20-250GUSB 585
0 to 500 0 to 34.5 PX409S15-500GUSB PX409S20-500GUSB 585
0 to 600 0 to 41.4 PX409S15-600GUSB — 585
Absolute Pressure
0 to 15 0 to 1 PX409S15-015AUSB PX409S20-015AUSB $610
0 to 30 0 to 2.1 PX409S15-030AUSB PX409S20-030AUSB 610
0 to 50 0 to 3.4 PX409S15-050AUSB PX409S20-050AUSB 610
0 to 100 0 to 6.9 PX409S15-100AUSB PX409S20-100AUSB 610
0 to 150 0 to 10.3 PX409S15-150AUSB PX409S20-150AUSB 610
0 to 250 0 to 17.2 PX409S15-250AUSB PX409S20-250AUSB 610
0 to 500 0 to 34.5 PX409S15-500AUSB PX409S20-500AUSB 610
0 to 600 0 to 41.4 PX409S15-600AUSB — 610
Vacuum Ranges (Negative Gage Pressure)
0 to -10 inH2O 0 to -25 mb PX409S15-10WVUSB PX409S20-10WVUSB $710
0 to -1 psi 0 to -69 mb PX409S15-001VUSB PX409S20-001VUSB 710
0 to -2.5 psi 0 to -172 mb PX409S15-2.5VUSB PX409S20-2.5VUSB 710
0 to -5 psi 0 to -345 mb PX409S15-005VUSB PX409S20-005VUSB 710
0 to -15 psi 0 to -1 PX409S15-015VUSB PX409S20-015VUSB 660
Compound Gage Ranges
±10 inH2O ±25 mb PX409S15-10WCGUSB PX409S20-10WCGUSB $735
±1 psi ±69 mb PX409S15-001CGUSB PX409S20-001CGUSB 735
±2.5 psi ±172 mb PX409S15-2.5CGUSB PX409S20-2.5CGUSB 735
±5 psi ±345 mb PX409S15-005CGUSB PX409S20-005CGUSB 735
±15 psi ±1 PX409S15-015CGUSB PX409S20-015CGUSB 685
Barometric Ranges (Absolute Pressure)
0 to 32 inHg — PX409S15-32BUSB PX409S20-32BUSB $735
16 to 32 inHg — PX409S15-16BUSB PX409S20-16BUSB 735
26 to 32 inHg — PX409S15-26BUSB PX409S20-26BUSB 735
Comes complete with 5-point NIST traceable calibration certificate, CD with set-up, databasing, charting and
driver software.
Ordering Examples: PX409S1.5-015GUSB, 1.5" Tri-Grip fitting, 2 m (6') cable termination with USB
termination, 15 psi gage range, USB Output, $585.
PX409S20-16BUSB, 2" Tri-Grip fitting, 2 m (6') cable termination with USB termination, 16 to 32 inHg
barometric range, USB Output, $735.
(Specify Model Number)
— 0 to 1100 hPa PX409S15-32HBUSB PX409S20-32HBUSB 735
— 550 to 1100 hPa PX409S15-16HBUSB PX409S20-16HBUSB 735
— 880 to 1100 hPa PX409S15-26HBUSB PX409S20-26HBUSB 735