Omega Products FTB691-P Installation Manual

Meter with Computer Display
Digital Pulse Meter
Operations Guide for
FTB690 and FTB690-P
Electronic Water Meters
TABLE OF CONTENTS
English ...........................................................1
Español ..........................................................7
Deutsch .......................................................13
Italiano .........................................................19
Français .......................................................24
E N G L I S H
IMPORTANT NOTICE
Use FTB690 Series meters with water and other chemicals compatible with wetted components (see Specifications Section). Do not use to meter fuel or incompatible chemicals. FTB690 Series meters are available with either a local electronic display, or a conditioned signal output module to provide a digital signal to customer interfacing equipment. FTB690 Series meters measure in gallons or liters. Refer to the Calibration Section for details.
These meters are not legal for trade applica­tions.
FTB690 Series meters are very sensitive to electric noise if operated within one to two inches of some electric motors or other sources of electronic noise.
2. For Soc Fittings use only primer and solvents approved for PVC gluing.
For NPT Fittings wrap all connections with
3 to 4 wraps of thread tape. Make sure the tape does not intrude into the flow path.
3. Attach meter with arrow pointed in the direction of flow.
4. For NPT Fittings - Hand tighten the meter at the housing ends. Do not use a wrench or similar tool to tighten. This can damage the housing.
Conditioned Signal Output Module Wiring
This conditioned signal output module can be wired to provide an open collector signal output or 6-volt square wave output.
When installing an FTB690-P Series meter, the correct K-factor must be entered into the readout device. You can find the K-factor printed on the turbine body. All K-factors on Omega flowmeters are calculated in pulses per gallon (PPG).
Open Collector Signal Output
To achieve an open collector signal output, reference Wiring Diagram 1. The terminal block is located on the back side of the module. The module is factory assembled for open collec­tor signal output. Please provide the (820 ohm minimum) resistor, if not supplied by receiving equipment.
Ten feet (3m) of cable is provided with the module. Trim it to desired length or extend it as necessary. Distances up to 5,000 feet (1,524m) can be achieved for open collector signal output.
INSTALLATION
Install your meter in-line either horizontally or vertically or at the end of the hose adjacent to the nozzle. Installation to metal connections is not recommended. Install as follows:
1. Plan to install turbine with a minimum straight pipe length as follows:
– Upstream from the turbine, allow a mini-
mum straight pipe length of 10 times
the internal diameter of the turbine. – Downstream from the turbine, allow a minimum straight pipe length of 5 times
the internal diameter of the turbine.
Square Wave Output
To achieve square wave output, reference Wiring Diagram 2 and use an Electronic Digital Meter Battery Kit (sold separately) for battery power. The terminal block and battery location are located on the back side of the module. Access as follows:
1. Remove the four Phillips-head screws from the front of the module and lift the module from the turbine.
2. To change terminal block connections, loosen the appropriate screws. Reconnect the wires in the proper positions and tighten the screws.
1
Wiring Diagram 1
+9 to 35 volt DC
Open Collector
Signal
3
2
1
4
Red
White
Black
Common (Ground)
Terminal Block
Customer
Interfacing
Equipment
The terminal block is identified as follows:
Pin #1 6 volt square wave (not used) Pin #2 +9 to 35 volt DC Input Pin #3 Common Ground Pin #4 Open Collector signal Output
3
2
1
4
White
Black
Common (Ground)
Terminal Block
Customer
Interfacing
Equipment
6 volt Square Wave Signal
The terminal block is identified as follows:
Pin #1 6 volt square wave Pin #2 +9 to 35 volt DC Input (not used) Pin #3 Common Ground Pin #4 Open Collector signal Output (not used)
OPEN COLLECTOR SIGNAL OUTPUT
Resistor
(820 ohm min.)
Wiring Diagram 2
2
SQUARE WAVE OUTPUT
3. Install the batteries. Make sure the positive post is in the correct position.
4. Position the module on the turbine housing. To avoid moisture damage, make sure the seal is fully seated. Tighten the four screws on the front of the module.
Ten feet (3m) of cable is provided with the module. Trim the cable to desired length or extend it as necessary.
Verify Meter Accuracy
Before using, check the meter’s accuracy and verify calibration.
1. Make sure there is no air in the system by starting the flow until it runs steadily. Then, stop the flow using a valve or nozzle.
2. Hold down DISPLAY for 3 seconds to zero the meter’s Batch Total. When zeros appear, release the button.
3. Meter an exact known volume into an accurate container. For best results, meter with one continuous full stream.
4. Check the volume against the display or recording equipment. If the amount metered is accurate, field calibration is not necessary. If not, refer to the Calibration Section for further instructions.
OPERATION
Batch and Cumulative Totals
The display maintains two totals. The Cumula­tive Total provides continuous measurement and cannot be manually reset. The Batch Total can be reset to measure flow during a single use. The Cumulative Total is labeled with TOTAL 1 LOCKED indicating that this total is locked and cannot be manually zeroed. Batch Total is labeled with TOTAL 2.
When the Cumulative Total reaches a maximum reading of 999,999, it will automatically reset to zero.
Press the DISPLAY button briefly to switch between the batch, cumulative total, and flowrate.
NOTE: Totalization counts total units without
differentiating between gallons, liters or field calibrated units.
Flowrate Feature
When this feature is activated, the word FLOW­RATE displays to the left on the bottom line.
When FLOWRATE is displayed, the numbers on the middle line reflect the rate of flow, for example, the current gallons per minute (GPM) or liters per minute (LPM).
Display FLOWRATE
To use this feature, press and release DISPLAY until FLOWRATE appears to the left of the bottom line.
Activate the Meter
Turn the meter ON by starting water flow or briefly pressing the DISPLAY button. The meter will display the Batch or Cumulative Total from last use.
Press DISPLAY briefly to display the Batch Total. Hold the DISPLAY button down for 3 seconds to reset the Batch Total to zero.
The meter is programmed to turn off auto­matically if not used for 4 minutes.
Factory and Field Calibration Curves
All calibration information is visible to the user as words in the upper part of the display, above the numeric digits.
All units are configured with a “factory” calibration curve. Both gallons and liters are available (“GAL” or “LTR” will be displayed). Use the CALIBRATE and DISPLAY buttons to switch between gallons and liters. This curve is NOT user adjustable: the word “PRESET” is displayed to show this. (The factory calibration is stored permanently in the computer’s memory.)
The “field” calibration curve may be set by the user, and can be changed or modified at any time using the calibration procedure described below in the Calibration Section. Totals or flowrate derived from the field calibration are visible when the field calibration setting is selected (“CAL B” will be visible on the top line).
3
Selecting a Different Calibration Setting
You can switch between GAL and LTR modes at will without “corrupting” totalizer contents. For example, the computer can totalize 10.00 gallons. If the user switches to LTR mode, the display will immediately change to “37.85” (the same amount in units of liters). GAL / LTR switching also works in FLOWRATE mode.
To select a different calibration setting, first press and hold the CALIBRATE button. Continue to hold it while also pressing and releasing the DISPLAY button. (You may then also release the CALIBRATE button.) The flag indicators in the top line of the display will change to show the newly selected calibration setting. Calibration settings change in this order: GAL, LTR, CAL B, GAL, etc. While fluid is flowing, only the GAL and LTR selections may be made. However, when NO fluid flow is occurring, any setting may be selected.
Field Calibration
Factory calibration settings are custom­programmed into each flowmeter during production, using water at 70°F (21°C). Readings using the standard factory calibration curves may not be accurate in some situations – for example, under extreme temperature conditions. You can field calibrate the meter if you are using fluids other than water.
For improved accuracy under such conditions, the OMEGA flow computer allows for “field” calibration, that is, user entry of custom cali­bration parameters. A “single point” calibration may yield acceptable accuracy in the middle of the flow range, but 5 or more calibration points may yield a higher level of accuracy, especially at the lower end of the flow range. Up to 15 custom calibration points can be entered.
The use of a uniformly dependable, accurate calibration container is highly recommended for the most accurate results. Due to high flowrate, it is strongly recommended that Field Calibration be completed with a combination of volume and weight using fine resolution scales.
Before Beginning Field Calibration
For the most accurate results, dispense at a flowrate which best simulates your actual oper­ating conditions. Avoid “dribbling” more fluid or
repeatedly starting and stopping the flow – these actions will result in less accurate calibra­tions.
Make sure you meet the meter’s minimum flowrate requirements:
FTB690 Series Meters
1/2 inch meter 1 GPM (3.8 LPM) 3/4 inch meter 2 GPM (7.5 LPM) 1 inch meter 5 GPM (18.8 LPM) 1-1/2 inch meter 10 GPM (37.85 LPM) 2 inch meter 20 GPM (75 LPM)
For best results, the meter should be installed and purged of air before field calibration.
Dispense/ Display Field Calibration Procedures
1. Hold down CALIBRATE while pressing and releasing DISPLAY until the field calibration curve appears (“CAL B” message will be displayed). Release both buttons.
2. To calibrate, press and hold the CALI­BRATE button. While continuing to hold CALIBRATE, also press and hold the DISPLAY button. Hold both buttons for about
3 seconds until you see a blinking “dd CAL” message. Once the “dd CAL” message
appears, release both buttons. You are now in field calibration mode.
3. Once the buttons have been released
from Step 2, the display will show the
blinking message “run 01”. If you want to exit the calibration now before dispensing any fluid, go to Step 11.
4. If you want to continue with the calibration, but have not dispensed any fluid yet, make your final preparations to your pumping system, but don’t start pumping yet.
5. Start your pumping system so that fluid flows through the meter. The display will stop blinking and show the “run 01” message. Dispense into a container that allows you to judge the amount of fluid pumped. When you have pumped the desired amount (for example, 10 gallons), stop the fluid flow quickly.
6. Once the flow has stopped, briefly press and release both buttons. At this point the computer display will change to “0000.00” with the left-hand digit blinking.
4
7. Enter the volume (amount) of fluid that you dispensed (for example, if your 10-gallon container is full, enter “10.0” for gallons or “37.85” for liters). To enter numbers, use the CALIBRATE button to change the value of the digit that is blinking and use the DISPLAY button to shift the “blink” to the next digit.
8. Once the correct number is entered, briefly press and release both buttons. The display will now change to a blinking “run 02” message. You have installed the new cal-curve point. You are ready to end calibration (Step
10) or enter another new calibration point (Step 9).
9. To enter another calibration point, go back and repeat Steps 3 through 8. It is possible to set up to 15 cal-curve points, and the “run ##” message will increment each time you repeat the calibration process (run 01, run 02, run 03, etc., up to run 15).
10. To end calibration, press and hold both buttons for about 3 seconds until you see the “CAL End” message. After you release the buttons the computer will resume nor­mal operations with the new cal point(s) active.
11. If you HAVE NOT dispensed any fluid, you can exit calibration without changing the
cal curve. If the message “run 01” is showing
and you have not dispensed any fluid, hold both buttons for about 3 seconds until you see a “CAL End” message. After you
release the buttons, the computer will
resume normal operation and the old curve (if you entered one in the past) is still intact.
CAUTION
Blowing compressed air through the turbine assembly could damage the rotor.
Battery Replacement
The meter is powered by two 3-volt lithium batteries which may be replaced while the meter is installed. When batteries are removed or lose power, the batch and cumulative totals reset to zero but the field and factory calibrations are retained.
If the meter display becomes dim or blank, replace the batteries as follows:
1. Remove the four Phillips-head screws from the face of the meter and lift the faceplate from the turbine.
2. Remove the old batteries and clean any corrosion from the terminals.
3. Install new batteries. Make sure the positive post is in the correct position.
4. When the batteries are replaced, the face­plate will power ON. Check the display to ensure normal functions have resumed before assembling again.
5. Reseat batteries, if necessary, and position the faceplate on the turbine housing.
To avoid moisture damage, make sure the seal is fully seated. Tighten the four screws
on the faceplate.
SPECIFICATIONS
MAINTENANCE
Proper handling and care will extend the life and service of the meter.
Turbine Rotor
The meter is virtually maintenance-free. However, it is important the rotor moves freely. Keep the meter clean and free of contaminants.
If the rotor does not turn freely, apply a penetrat­ing lubricant on the rotor, shaft, and bearings. Remove any debris or deposits from the rotor using a soft brush or small probe. Be careful not to damage the turbine rotor or supports.
Inlet and Outlet:
For Soc Models: FTB691/FTB691-P 1/2" Sch 40, Soc FTB692/FTB692-P 3/4" Sch 40, Soc FTB693/FTB693-P 1" Sch 40, Soc FTB694/FTB694-P 1-1/2" Sch 40, Soc FTB695/FTB695-P 2" Sch 40, Soc
For NPT Models: FTB691-NPT/FTB691-NPT-P 1/2" NPT FTB692-NPT/FTB692-NPT-P 3/4" NPT FTB693-NPT/FTB693-NPT-P 1" NPT FTB694-NPT/FTB694-NPT-P 1-1/2" NPT FTB695-NPT/FTB695-NPT-P 2" NPT
Design Type: Turbine
5
Wetted Components:
Housing: PVC Journal Bearings: Ceramic Shaft: Tungsten Carbide Rotor and Supports: PVDF Retaining Washer: Stainless Steel
Fitting Types: Slip - Schedule 40 Soc or NPT Max. Working Pressure: 150 PSIG @ 73°F
U.S. Measurement
Unit of Measure: Gallon Flow Range:
1/2 inch 1 - 10 GPM 3/4 inch 2 - 20 GPM 1 inch 5 - 50 GPM 1-1/2 inch 10 - 100 GPM 2 inch 20 - 200 GPM
Accuracy: ± 3.0% of reading (Accuracy can be improved with field calibration)
Operating Temperature: +32° to +140° F NOTE: Do not allow fluid to freeze inside meter.
Storage Temperature: –40° to +158° F Weight:
1/2 inch .31 lbs. 3/4 inch .38 lbs. 1 inch .51 lbs. 1-1/2 inch .81 lbs. 2 inch 1.14 lbs.
Dimensions - inches (W x H x L):
1/2 inch 2 x 2.5 x 4.3 3/4 inch 2 x 2.7 x 5.3 1 inch 2 x 3 x 6 1-1/2 inch 2.3 x 3.6 x 7 2 inch 2.8 x 4.1 x 7.5
Metric Measurement
Unit of Measure: Liter Flow Range:
1/2 inch 3.8 - 38 LPM 3/4 inch 7.6 - 76 LPM 1 inch 19 - 190 LPM 1-1/2 inch 38 - 380 LPM 2 inch 76 - 760 LPM
Accuracy: ± 3.0% of reading (Accuracy can be improved with field calibration)
Operating Temperature: 0° to +60° C NOTE: Do not allow fluid to freeze inside meter.
Storage Temperature: –40° to +70° C Weight:
1/2 inch .141 kg 3/4 inch .173 kg 1 inch .232 kg 1-1/2 inch .368 kg 2 inch .518 kg
Dimensions - cm (W x H x L):
1/2 inch 5.1 x 6.4 x 10.9 3/4 inch 5.1 x 6.9 x 13.5 1 inch 5.1 x 7.6 x 15.2 1-1/2 inch 5.8 x 9.1 x 17.8 2 inch 7.1 x 10.4 x 19.1
PARTS
The following replacement parts and acces­sories are available for the FTB690 Series meters:
Part No. Description
FLSC790-BATT Battery Replacement Kit FTB890 O-ring O-Ring FTB691-RK 1/2 inch, Soc Turbine Assembly Kit FTB691-NPT-RK 1/2 inch, NPT, PVC, Turbine Assembly Kit FTB692-RK 3/4 inch, Soc Turbine Assembly Kit FTB692-NPT-RK 3/4 inch, NPT, PVC Turbine Assembly Kit FTB693-RK 1 inch, Soc Turbine Assembly Kit FTB693-NPT-RK 1 inch, NPT, PVC Turbine Assembly Kit FTB694-RK 1-1/2 inch, Soc Turbine Assembly Kit FTB694-NPT-RK 1-1/2 inch, NPT, PVC, Turbine Assembly Kit FTB695-RK 2 inch, Soc Turbine Assembly Kit FTB695-NPT-RK 2 inch, NPT, PVC Turbine Assembly Kit
Computer Kits:
FTB691-CK 1/2 inch, Computer Assy Kit FTB692-CK 3/4 inch, Computer Assy Kit FTB693-CK 1 inch, Computer Assy Kit FTB694-CK 1-1/2 inch, Computer Assy Kit FTB695-CK 2 inch, Computer Assy Kit
6
E S P A Ñ O L
AVISO IMPORTANTE
Utilizar los metros de los Series del FTB690 con agua y otros productos químicos que son compatibles con les componentes que se exponen al líquido (véase la sección de especifica­ciones). No utilizar para medir el incompatible combustible o los productos químicos. Los metros de la serie del FTB690 están disponibles con una visualización electrónica local, o un módulo de salida condicionado de la señal para proporcionar una señal numérica al equipo de interconexión del cliente. Los medidores Series FTB690 miden en galones o litros. Referirse a la sección de la calibración para mayores detalles.
Estos medidores no son legales para las aplicaciones comerciales.
Los medidores de las Series FTB690 son muy sensibles a interferencia electrónica si funcionan a 1 o 2 pulgadas de algunos motores eléctricos o de otras fuentes del uso electrónico.
Instalar su medidor en línea, u horizontalmente, o verticalmente, o en el extremo de la manguera adyacente al inyector. No se recomienda la instalación a las conexiones de metal. Siga estos pasos para instalar:
1. Planee instalar la turbina con una longitud mínima de la pipa recta de esta manera:
– Contra la corriente de la turbina, permita a una longitud mínima de la
– Con la corriente de la turbina, permita una longitud mínima de la pipa recta de 5 veces el diámetro interno de la turbina.
2. Para Las Conexiones Del Soc utilizar solamente los solventes aprobados para pegar PVC.
Para Las Conexiones Del NPT cubrir las con
exiones de pipa con la vueltas de cinta del 3 a 4 veces. Cerciorarse de que la cinta no imponga en la trayectoria del flujo.
3. Unir el metro con la flecha señalada en la dirección del flujo.
INSTALACIÓN
pipa recta de 10 veces el diámetro interno de la turbina.
4. Para Las Conexiones Del NPT utilizar solamente sus manos para apretar las conexiones del metro. No utilizar una llave inglesa o una herramienta similar para apretar. Esto puede dañar la cubierta.
Señal de Salida Condicionada Cableado De Módulo
Este módulo de Señal de salida condicionada se puede conectar para proporcionar una salida de colector abierta o de señal de onda cuadra-da de 6-voltios.
Al instalar un flujómetro de la turbina de la serie del FTB690-P, el factor K correcto se debe incorporar en el dispositivo de la lectura. Usted puede encontrar el factor K impreso en el cuerpo de la turbina. Todos los factores K en flujómetros del Omega se calculan en pulsos por galón (PPG).
Señal de Salida De Colector Abierto
Para alcanzar una señal de salida de colector abierto, refierase por favor al digrama eléctrico
1. El bloque de terminales está situado en
el lado trasero del módulo. El módulo viene montado de fábrica para señal de colector abierta. Por favor proporcionar (el resistor de un minimo de 820 ohmios), si no es provisto por el equipo de recepción.
Diez pies (3m) de cable se proporcionan con el módulo. Ajustar el cable a la longitud deseada o extender el cable cuanto le sea necesario. Se puede alcanzar una señal de salida de colector abierto hasta distancias de 5.000 pies (1,524m).
Salida de corrente de Onda Cuadrada
Para lograr una salida de corriente de onda cuadrada, refierase por favor al digrama eléc­trico 2 y utilize un kit electrónico de bateria del medidor digital (vendido por separado) para la fuente de energia de la bateria. El bloque de terminales y la localización de la bateria están situados en el lado trasero del modulo. Acceda al módulo de la siguiente manera:
1. Quitar los cuatro tornillos de cabeza Phillips del frente del módulo. Levantar el módulo de la turbina.
7
2. Para cambiar las conexiones del bloque de terminales, aflojar los tornillos apro­piados. Volver a conectar los alambres en las posiciones apropiadas y apretar los tornillos.
3. Instalar las baterias. Cerciorarse de que el poste positivo esté en la posición correcta.
4. Colocar el módulo en la cubierta de la turbina. Para evitar daños causados por la humedad, cerciorarse de que el anillo esté asentado completamente. Apretar los cuatro tornillos en el frente del módulo.
Diez pies (3m) de cable se proporcionan con el módulo. Ajustar el cable a la longitud deseada o extender el cable cuanto le sea necesario.
Verificar La Exactitud Del Metro
Antes de usar, comprobar la exactitud del metro y verificar la calibración.
1. Cerciorarse de que no haya aire en el sistema comenzando el flujo hasta que funciona constantemente. Entonces, detenga el flujo usando una válvula o un inyector.
2. Mantenga el botón de la EXHIBICIÓN
durante 3 segundos para llevar a cero el
total de medidor. Cuando aparezcan los ceros, suelte el botón.
3. Con el medidor, mida un volumen exacto
en un envase exacto. Para mejores resul­ tados, medir con una corriente complete y
continua.
4. Comprobar el volumen contra la pantalla o el equipo de grabacion. Si la cantidad medi-
da es exacta, la calibración de campo no
es necesaria. Si no, refeierase a la sección de la calibración.
OPERACIÓN
Hornada y Totales Acumulativos
La pantalla mantiene dos totales. El total acumulativo proporciona la medida continua y no puede ser reajustado manualmente. El total de hornada se puede reajustar para medir el flujo durante una sola vez . El total acumulativo se etiqueta con el TOTAL 1 LOCKED. Esto indica que el total esta bloqueado y no puede ser puesto a cero manualmente. El total de hornada se etiqueta con el TOTAL 2.
Cuando el total acumulativo alcanza una lec­tura máxima de 999.999, se reajustará auto­máticamente a cero.
Presionar el botón de DISPLAY brevemente para cambiar entre la hornada, el total acumu­lativo, y el índice de flujo.
NOTA: Totalization cuenta las unidades totales
sin distinguir entre los galones, los litros
o las unidades calibradas de campo.
Característica Del Índice De Flujo
Cuando se activa esta característica, aparece la palabra FLOWRATE a la izquierda hasta abajo.
Cuando se exhibe el FLOWRATE, los números en la línea media reflejan el régimen, por ejemplo, los galones actuales por minuto (GPM) o los litros por minuto (LPM).
Exhibición Del Índice De Flujo
Para utilizar esta característica, presionar y soltar el botón del DISPLAY hasta que FLOWRATE aparece a la izquierda hasta abajo.
Activar El Metro
Encienda el metro comenzando el flujo del agua o brevemente presionando el botón del DISPLAY. El medidor exhibirá la hornada o el total acumulativo a partir de la última vez que fue utilizado.
Presionar el botón del DISPLAY brevemente para exhibir el total de hornada. Oprima el botón de DISPLAY por 3 segundos para reajustar el total de hornada a cero.
El medidor se apaga automáticamente si no es usado durante 4 minutos.
Curvas De Calibración De La Fábrica y Del Campo
Toda la información de la calibración es visible al usuario como palabras en la parte superior de la exhibición, sobre los dígitos numéricos.
Todas las unidades se configuran con una curva de calibración de la “fábrica”. Los galones y los litros están disponibles. (el “GAL” o el “LTR” será visible). Utilizar los botones del CALIBRATE y del DISPLAY para cambiar entre los galones y los litros. Esta curva de calibración
8
no es ajustable por el usuario. La palabra PRESET se exhibe para demostrar esto. (La calibración de la fábrica se almacena permanentemente en la memoria de computadora.)
La curva de calibración de “campo” se puede fijar por el usuario. La calibración se puede cambiar o modificar en cualquier momento usando los procedimientos de la calibración descritos en la sección de la calibración. Los totales o el índice de flujo derivados de la calibración de campo son visibles cuando se selecciona el ajuste de la calibración de campo (la “CAL B” será visible en la línea superior).
Seleccionar Un Ajuste Diverso De La Calibración
Usted puede cambiar entre los modos del GAL y del LTR a voluntad sin afectar los totales. Por ejemplo, la computadora puede sumar 10,00 galones. Si el usuario cambia al modo del LTR, la exhibición cambiará inmediatamente a “37,85” (la misma cantidad en las unidades de los litros). La conmutación del GAL/LTR también trabaja en el modo del FLOWRATE.
Para seleccionar un ajuste diverso de la calibra­ción, oprima y sostenga el botón de la CALIBRATE. Continuar sosteniendo el botón mientras que también presiona y suelta el botón de DISPLAY. (usted puede entonces también soltar el botón de la CALIBRATE.) Los indica­dores de la bandera de la línea superior de la exhibición cambiarán para demostrar el nuevo ajuste seleccionado de la calibración. Los ajustes de la calibración se cambian en este orden: GAL, LTR, CAL B, GAL, etc. Mientras que está fluyendo el líquido, sólo las selecciones del galón y del litro pueden ser hechas. Sin embargo, cuando no está fluyendo NINGÚN líquido, cualquier selección puede ser hecha.
Calibración Del Campo
Los ajustes de la calibración de la fábrica se programan especificamente en cada flujómedidor durante su producción usando agua a 70°F (21°C). Las lecturas que utilizan las curvas de calibración estándares de la fábrica pueden no ser exactas en algunas situaciones. Por ejemplo, cuando se encuentran bajo condiciones de temperatura extremas. Usted puede campo calibrar el metro si usted está utilizando los líquidos con excepción del agua.
Para la exactitud mejorada bajo tales condi­ciones, la computadora OMEGA de flujo tienen en cuenta la calibración del “campo” (es decir un apunte del usuario dentro de los parámetros de calibración especiales). La calibración de “un solo punto” puede rendir una exactitud aceptable en medio de la gama del flujo. Cinco o más puntos de calibración pueden rendir un nivel más alto de exactitud, especialmente en el extremo inferior de la gama del flujo. Hasta 15 puntos de calibración especiales pueden ser incorporados.
Se recomienda altamente usar un envase confiable, y exacto para la calibración. Para obtener resultados más exactos. debido al alto índice de flujo, se recomienda fuertemente que la calibración de campo esté determinada con una combinación del volumen y del peso usando las escalas de alta resolución.
Antes De Comenzar La Calibración De Campo
Para resultados más exactos, dispense un índice de flujo que simule lo mejor posible sus condiciones de funcionamiento reales. Evite “de gotear” más líquido o en varias ocasiones, o el comenzar y de parar el flujo. Estas acciones darán lcomo resultado calibraciones menos exactas.
Cerciorese de reunir todos los requisitos míni­mos del índice de flujo del medidor:
Medidores de las Series del FTB690
Medidor de 1/2 pulgada de
1 GPM (3,8 LPM)
Medidor de 3/4 pulgada de
2 GPM (7,5 LPM)
Medidor de 1 pulgada de
5 GPM (18,8 LPM)
Medidor de 1-1/2 pulgadas de
10 GPM (37,85 LPM)
Medidor de 2 pulgadas de
20 GPM (75 LPM)
Para mejores resultados, el medidor se debe instalar y purgar del aire antes de la calibración de campo.
9
Loading...
+ 25 hidden pages