Omega Products FTB691A Installation Manual

Meter with Computer Display
Digital Pulse Meter
FTB691A through FTB695A
Electronic Water Meters
TABLE OF CONTENTS
English ...........................................................1
Español ..........................................................7
Deutsch .......................................................12
Italiano .........................................................18
Français .......................................................23
E N G L I S H
IMPORTANT NOTICE
Use FTB690 Series meters with water and other chemicals compatible with wetted components (see Specifications Section). Do not use to meter fuel or incompatible chemicals. FTB690 Series meters are available with either a local electronic display, or a conditioned signal output module to provide a digital signal to customer interfacing equipment. FTB690 Series meters with computer display measure in gallons or liters. Refer to the Calibration Section for details.
These meters are not legal for trade applica­tions.
FTB690 Series meters are very sensitive to electric noise if operated within one to two inches of some electric motors or other sources of electronic noise.
INSTALLATION
Install your meter in-line either horizontally or vertically or at the end of the hose adjacent to the nozzle. Installation to metal connections is not recommended. Install as follows:
1. Plan to install turbine with a minimum straight pipe length as follows:
– Upstream from the turbine, allow a mini-
mum straight pipe length of 10 times
the internal diameter of the turbine. – Downstream from the turbine, allow a minimum straight pipe length of 5 times
the internal diameter of the turbine.
2. For NPT Fittings wrap all connections with 3 to 4 wraps of thread tape. Make sure the tape does not intrude into the flow path.
3. Attach meter with arrow pointed in the direction of flow.
4. For NPT Fittings - Hand tighten the meter at the housing ends. Do not use a wrench or similar tool to tighten. This can damage the housing.
Conditioned Signal Output Module Wiring
This conditioned signal output module can be wired to provide an open collector signal output or 6-volt square wave output.
When installing an FTB690 Series meter, the correct K-factor must be entered into the readout device. You can find the K-factor printed on the turbine body. All K-factors on Omega flowmeters are calculated in pulses per gallon (PPG).
Open Collector Signal Output
To achieve an open collector signal output, reference Wiring Diagram 1. The terminal block is located on the back side of the module. The module is factory assembled for open collec­tor signal output. Please provide the (820 ohm minimum) resistor, if not supplied by receiving equipment.
Ten feet (3 m) of cable is provided with the module. Trim it to desired length or extend it as necessary. Distances up to 5,000 feet (1,524 m) can be achieved for open collector signal output.
Square Wave Output
To achieve square wave output, reference Wir­ing Diagram 2 and use an FLSC790-BATT (sold separately) for battery power. The terminal block and battery location are located on the back side of the module. Access as follows:
1. Remove the four Phillips-head screws from the front of the module and lift the module from the turbine.
2. To change terminal block connections, loosen the appropriate screws. Reconnect the wires in the proper positions and tighten the screws.
1
Wiring Diagram 1
+9 to 35 volt DC
Open Collector
Signal
3
2
1
4
Red
White
Black
Common (Ground)
Terminal Block
Customer
Interfacing
Equipment
The terminal block is identified as follows:
Pin #1 6 volt square wave (not used) Pin #2 +9 to 35 volt DC Input Pin #3 Common Ground Pin #4 Open Collector signal Output
3
2
1
4
White
Black
Common (Ground)
Terminal Block
Customer
Interfacing
Equipment
6 volt Square Wave Signal
The terminal block is identified as follows:
Pin #1 6 volt square wave Pin #2 +9 to 35 volt DC Input (not used) Pin #3 Common Ground Pin #4 Open Collector signal Output (not used)
OPEN COLLECTOR SIGNAL OUTPUT
Resistor
(820 ohm min.)
Wiring Diagram 2
2
SQUARE WAVE OUTPUT
3. Install the batteries. Make sure the positive post is in the correct position.
4. Position the module on the turbine housing. To avoid moisture damage, make sure the seal is fully seated. Tighten the four screws on the front of the module.
Ten feet (3 m) of cable is provided with the module. Trim the cable to desired length or extend it as necessary.
Verify Meter Accuracy
Before using, check the meter’s accuracy and verify calibration.
1. Make sure there is no air in the system by starting the flow until it runs steadily. Then, stop or divert the flow using a valve or nozzle.
2. Hold down DISPLAY for 3 seconds to zero the meter’s Batch Total. When zeros appear, release the button.
4. Check the volume against the display or recording equipment. If the amount metered is accurate, field calibration is not necessary. If not, refer to the Calibration Section for further instructions.
Batch and Cumulative Totals
The computer maintains two totals. The Cumu­lative Total provides continuous measurement and cannot be manually reset. The Batch Total can be reset to measure flow during a single use. The Cumulative Total is labeled TOTAL 1, Batch Total is labeled TOTAL 2 BATCH.
When the Cumulative Total reaches a display reading of 999,999 the computer will highlight an X10 icon. This indicates to the operator that a zero must be added to the 6 digits shown. When the next rollover occurs, the computer will highlight an X100 icon. This indicates to the operator that two zeros must be added to the 6 digits shown.
Press the DISPLAY button briefly to switch between the TOTAL 1, TOTAL 2 BATCH and FLOWRATE. Press DISPLAY briefly to display the TOTAL 2 BATCH. Hold the DISPLAY button for 3 seconds to reset the Batch Total to zero.
When fluid is flowing through the meter, a small propeller icon is highlighted.
Flowrate Feature
To use this feature, press and release DISPLAY button until FLOWRATE icon appears. The factory set time base will be highlighted to the right of FLOWRATE (M = minutes, H = hours, D = days). When FLOWRATE is invoked, the display will be indicating rate of flow.
OPERATION
Computer Display
All operations are reflected in the LCD readout. The large center digits indicate amounts, where smaller words or “icons” located above and below indicate specific information regarding totals, flow, calibration and units of measure.
Activate the Meter
Computer is on continuously and always ready to perform. The computer is powered by field replaceable batteries. When display becomes dim, faded or the low battery message appears (see below), the batteries need to be replaced. Reference the Maintenance Section for details.
Factory and Field Calibration
All calibration information is visible to the user as icons on the top line of the display, above the numeric digits.
All units are configured with a “factory” calibra­tion. Both gallons and liters are available (“GL” or “LT” will be displayed). While holding the CALI­BRATE button, briefly press DISPLAY to toggle between gallons and litres. This factory calibra­tion (indicated with FAC) is permanently program­med into the computer and is not user adjustable.
NOTE: Your computer may have other units of
measure programmed into it. If so, holding the CALIBRATE button and momentarily pressing the DISPLAY button will toggle through all factory set units. Other pos­sible units are: IGL (imperial gallon), QT (quart), CF (cubic feet), CM (cubic meter), BL (42 gal. barrel), CC (cubic centimeter) or OZ (ounce).
3
Switching between different units will not corrupt the Total’s contents. For example, in GL mode, the computer totalizes 10.00 gallons, if the user switches to LT mode, the display will read 37.85 litres (the same volume, different unit).
The “field” calibration may be set by the user, and can be changed or modified at any time using the calibration procedure described below in the Calibration Section. Totals or flowrate derived from the field calibration are invoked when the FAC icon is no longer visible on the top line of the display.
CALIBRATION
Verify Accuracy Before Beginning Field Calibration
For the most accurate results, dispense at a flowrate which best simulates your actual operating conditions. Avoid “dribbling” more fluid or repeatedly starting and stopping the flow. This can result in less accurate calibrations.
Make sure you meet the meter’s minimum flowrate requirements:
FTB690A Series Meters
1/2 inch meter 1 GPM (3.8 LPM) 3/4 inch meter 2 GPM (7.5 LPM) 1 inch meter 5 GPM (18.9 LPM) 1-1/2 inch meter 10 GPM (37.9 LPM) 2 inch meter 20 GPM (75 LPM)
The use of a uniformly dependable, accurate calibration container is recommended for the most accurate results. Due to high flowrate, it is strongly recommended that calibration be completed with a combination of volume and weight using fine resolution scales. For best results, the meter should be installed and purged of air before field calibration.
Field Calibration Procedures (Dispense/Display Method)
1. To field calibrate, press and hold CALI­BRATE and DISPLAY buttons for about 3 seconds until you see FLdCAL. Release both buttons and you will see dd000.0. You are now in the field calibration mode.
2. Dispense a known amount of fluid at a flowrate representative of the application. Any amount between .1 and 999.9 units can be used. Display will count up while fluid is flowing through the meter.
3. The DISPLAY button can then be pushed to select the digit location and the CALI­BRATE button can be pushed to scroll the desired value at the blinking position. Edit the amount shown with the value that was dispensed above. Values from 000.1 to
999.9 can be entered.
4. When satisfied with the value, press both CALIBRATE and DISPLAY buttons simul­taneously. CALEnd will be displayed and unit will go back to normal operation, less the FAC (factory calibration) icon.
5. The meter will now be operating with a cus­tom calibration number unique to the above dispense procedure. No unit of measure (gallon, litre, etc.) icon will be highlighted.
NOTE: To return to factory calibration (FAC),
press and hold both CALIBRATION and DISPLAY buttons for about 3 seconds, until FAcCAL is displayed. Then release buttons. Unit should return to normal operation and FAC icon visible.
NOTE: If the field calibration mode is entered
and NO fluid is dispensed, then upon leav­ing, the computer will use data from the last successful field calibration.
Field Calibration with Computer Display
Field Calibration and Factory Calibration are defined in the Operation Section. Factory calibration settings are programmed into each computer during manufacturing, using water at 70° F (21° C). Readings using the Factory Calibration (FAC) may not be accurate in some situations, for example, under extreme tem­perature conditions, non-standard plumbing configurations or with fluids other than water.
4
MAINTENANCE
SPECIFICATIONS
Proper handling and care will extend the life and service of the meter.
Turbine Rotor
The meter is virtually maintenance-free. However, it is important the rotor moves freely. Keep the meter clean and free of contaminants.
If the rotor does not turn freely, apply a penetrat­ing lubricant on the rotor, shaft, and bearings. Remove any debris or deposits from the rotor using a soft brush or small probe. Be careful not to damage the turbine rotor or supports.
CAUTION
Blowing compressed air through the turbine assembly could damage the rotor.
Battery Replacement
The computer display is powered by two 3-volt lithium batteries which may be replaced while the meter is installed. When batteries are removed or lose power, the batch and cumula­tive totals and the field and factory calibrations are re tained.
If the display becomes dim, blank or the low battery message appears (see below), replace the batteries as follows:
1. Remove the four Phillips-head screws from the face of the meter and lift the faceplate from the turbine.
2. Remove the old batteries and clean any corrosion from the terminals.
3. Install new batteries. Make sure the positive post is in the correct position.
4. When the batteries are replaced, the face­plate will power ON. Check the display to ensure normal functions have resumed before assembling again.
To avoid moisture damage, make sure the seal is fully seated. Tighten the four screws
on the faceplate.
Inlet and Outlet:
For NPT Models: FTB691A-NPT/FTB691A-NPT-P 1/2" NPT FTB692A-NPT/FTB692A-NPT-P 3/4" NPT FTB693A-NPT/FTB693A-NPT-P 1" NPT FTB694A-NPT/FTB694A-NPT-P 1-1/2" NPT FTB695A-NPT/FTB695A-NPT-P 2" NPT
Design Type: Turbine Wetted Components:
Housing: PVC Journal Bearings: Ceramic Shaft: Tungsten Carbide Rotor and Supports: PVDF Retaining Washer: Stainless Steel
Fitting Types: Schedule 80 NPT Max. Working Pressure: 225 PSIG @ 73° F
U.S. Measurement
Unit of Measure: Gallon Flow Range:
1/2 inch 1 - 10 GPM 3/4 inch 2 - 20 GPM 1 inch 5 - 50 GPM 1-1/2 inch 10 - 100 GPM 2 inch 20 - 200 GPM
Accuracy: ± 3.0% of reading (Accuracy can be improved with field calibration)
Operating Temperature: +32° to +140° F NOTE: Do not allow fluid to freeze inside meter.
Storage Temperature: –40° to +158° F Weight: (Includes Computer Display. Conditioned
Signal Module adds .30 lbs.)
1/2 inch .55 lbs. 3/4 inch .67 lbs. 1 inch .84 lbs. 1-1/2 inch 1.38 lbs. 2 inch 1.78 lbs.
Dimensions - inches (W x H x L):
(Includes Computer Display. Conditioned Signal Module adds 1.1 inch to height)
1/2 inch 2.0 x 2.8 x 5.5 3/4 inch 2.0 x 2.9 x 5.5 1 inch 2.0 x 3.3 x 6.2 1-1/2 inch 2.3 x 3.9 x 7.6 2 inch 3.5 x 4.5 x 7.9
5
Metric Measurement
Unit of Measure: Liter Flow Range:
1/2 inch 3.8 - 38 LPM 3/4 inch 7.6 - 76 LPM 1 inch 19 - 190 LPM 1-1/2 inch 38 - 380 LPM 2 inch 76 - 760 LPM
Accuracy: ± 3.0% of reading (Accuracy can be improved with field calibration)
Operating Temperature: 0° to +60° C NOTE: Do not allow fluid to freeze inside meter.
Storage Temperature: –40° to +70° C Weight: (Includes Computer Display. Conditioned
Signal Module adds .14 kg)
1/2 inch .249 kg 3/4 inch .304 kg 1 inch .381 kg 1-1/2 inch .626 kg 2 inch .807 kg
Dimensions - cm (W x H x L): (Includes Com-
puter Display. Conditioned Signal Module adds
2.8 cm to height)
1/2 inch 5.0 x 7.1 x 13.9 3/4 inch 5.0 x 7.3 x 13.9 1 inch 5.0 x 8.3 x 15.7 1-1/2 inch 5.8 x 9.9 x 19.3 2 inch 8.8 x 11.4 x 20.0
PARTS
The following parts and accessories are avail­able for the FTB690 Series meters:
Part No. Description
FLSC790-BATT Battery Replacement Kit FTB890 O-ring O-Ring FTB691A-NPT-RK 1/2 in., NPT, PVC, Turbine Assembly Kit FTB692A-NPT-RK 3/4 in., NPT, PVC Turbine Assembly Kit FTB693A-NPT-RK 1 in., NPT, PVC Turbine Assembly Kit FTB694A-NPT-RK 1-1/2 in., NPT, PVC, Turbine Assembly Kit FTB695A-NPT-RK 2 in., NPT, PVC Turbine Assembly Kit
Computer Kits:
FTB691-CK 1/2 in., Computer Assy Kit FTB692-CK 3/4 in., Computer Assy Kit FTB693-CK 1 in., Computer Assy Kit FTB694-CK 1-1/2 in., Computer Assy Kit FTB695-CK 2 in., Computer Assy Kit
6
E S P A Ñ O L
AVISO IMPORTANTE
Utilizar los metros de los Series del FTB690 con agua y otros productos químicos que son compatibles con les componentes que se exponen al líquido (véase la sección de especifica­ciones). No utilizar para medir el incompatible combustible o los productos químicos. Los metros de la serie del FTB690 están disponibles con una visualización electrónica local, o un módulo de salida condicionado de la señal para proporcionar una señal numérica al equipo de interconexión del cliente. Los medidores Series FTB690 miden en galones o litros. Referirse a la sección de la calibración para mayores pantalla de ordenador detalles.
Estos medidores no son legales para las aplicaciones comerciales.
Los medidores de las Series FTB690 son muy sensibles a interferencia electrónica si funcionan a 1 o 2 pulgadas de algunos motores eléctricos o de otras fuentes del uso electrónico.
Instalar su medidor en línea, u horizontalmente, o verticalmente, o en el extremo de la manguera adyacente al inyector. No se recomienda la instalación a las conexiones de metal. Siga estos pasos para instalar:
1. Planee instalar la turbina con una longitud
mínima de la pipa recta de esta manera: – Contra la corriente de la turbina, permita a una longitud mínima de la
– Con la corriente de la turbina, permita una longitud mínima de la pipa recta de 5 veces el diámetro interno de la turbina.
2. Para las Conexiones del NPT cubrir las con
exiones de pipa con la vueltas de cinta del
3 a 4 veces. Cerciorarse de que la cinta no
imponga en la trayectoria del flujo.
3. Unir el metro con la flecha señalada en la
dirección del flujo.
INSTALACIÓN
pipa recta de 10 veces el diámetro interno de la turbina.
4. Para las Conexiones del NPT utilizar solamente sus manos para apretar las conexiones del metro. No utilizar una llave inglesa o una herramienta similar para apretar. Esto puede dañar la cubierta.
Señal de Salida Condicionada Cableado de Módulo
Este módulo de Señal de salida condicionada se puede conectar para proporcionar una salida de colector abierta o de señal de onda cuadra da de 6-voltios.
Al instalar un flujómetro de la turbina de la serie del FTB690, el factor K correcto se debe incor­porar en el dispositivo de la lectura. Usted puede encontrar el factor K impreso en el cuerpo de la turbina. Todos los factores K en flujómetros del Omega se calculan en pulsos por galón (PPG).
Señal de Salida de Colector Abierto
Para alcanzar una señal de salida de colector abierto, refierase por favor al digrama eléctrico
1. El bloque de terminales está situado en
el lado trasero del módulo. El módulo viene montado de fábrica para señal de colector abierta. Por favor proporcionar (el resistor de un minimo de 820 ohmios), si no es provisto por el equipo de recepción.
Diez pies (3 m) de cable se proporcionan con el módulo. Ajustar el cable a la longitud deseada o extender el cable cuanto le sea necesario. Se puede alcanzar una señal de salida de colector abierto hasta distancias de 5.000 pies (1,524 m).
Salida de corrente de Onda Cuadrada
Para lograr una salida de corriente de onda cuadrada, refierase por favor al digrama eléc­trico 2 y utilize un FLSC790-BATT (vendido por separado) para la fuente de energia de la bateria. El bloque de terminales y la localización de la bateria están situados en el lado trasero del modulo. Acceda al módulo de la siguiente manera:
1. Quitar los cuatro tornillos de cabeza Phillips del frente del módulo. Levantar el módulo de la turbina.
7
2. Para cambiar las conexiones del bloque de terminales, aflojar los tornillos apro­piados. Volver a conectar los alambres en las posiciones apropiadas y apretar los tornillos.
3. Instalar las baterias. Cerciorarse de que el poste positivo esté en la posición correcta.
4. Colocar el módulo en la cubierta de la turbina. Para evitar daños causados por la humedad, cerciorarse de que el anillo esté asentado completamente. Apretar los cuatro tornillos en el frente del módulo.
Diez pies (3 m) de cable se proporcionan con el módulo. Ajustar el cable a la longitud deseada o extender el cable cuanto le sea necesario.
Verificar la Exactitud del Metro
Antes de usar, comprobar la exactitud del metro y verificar la calibración.
1. Cerciorarse de que no haya aire en el sistema comenzando el flujo hasta que funciona constantemente. Entonces, detener o des-
viar el flujo mediante una válvula o la boquilla.
2. Mantenga el botón de la DISPLAY durante 3 segundos para llevar a cero el total de medidor. Cuando aparezcan los ceros, suelte el botón.
3. Con el medidor, mida un volumen exacto en
un envase exacto. Para mejores resultados,
medir con una corriente complete y continua.
4. Comprobar el volumen contra la pantalla o el equipo de grabacion. Si la cantidad medi-
da es exacta, la calibración de campo no
es necesaria. Si no, refeierase a la sección de la calibración.
OPERACIÓN
Ordenador de Pantalla
Todas las operaciones se reflejan en la lectura del LCD. Los dígitos grandes del centro indican cantidades, donde las palabras más pequeñas o “iconos” que se encuentran por encima y por debajo indican la información específica sobre los totales, el flujo, la calibración y unidades de medida.
Para Activar el Contador
El ordenador está encendido continuamente y siempre dispuesto a operar. El equipo es ali­mentado por baterías reemplazables de campo. Cuando la pantalla se vuelve opaco, desteñido o el mensaje de batería baja (vea más abajo), las baterías necesitan ser reemplazadas. Ref­erencia de la sección de mantenimiento para más detalles.
Lote y los Totales Acumulados
El computadora mantiene dos totales. El total acumulativo proporciona la medida continua y no puede ser reajustado manualmente. El total de hornada se puede reajustar para medir el flujo durante una sola vez. El total acumulativo se etiqueta con el TOTAL 1 LOCKED. Esto in­dica que el total esta bloqueado y no puede ser puesto a cero manualmente. l total acumulado es etiquetado con TOTAL 1, lote total se etiqueta TOTAL 2 BATCH.
Cuando el total acumulado llega a una lectura de la pantalla de la computadora 999.999 resaltar un icono de X10. Esto indica al operador que un cero se debe agregar a los 6 dígitos que se muestran. Cuando el traspaso se produzca la próxima vez, el equipo se hará hincapié en un icono X100. Esto indica al operador que dos ceros se debe agregar a los 6 dígitos que se muestran.
Pulse brevemente el botón DISPLAY para cambiar entre el TOTAL 1, TOTAL 2 BATCH y la configuración FLOWRATE. Pulse brevemente DISPLAY para mostrar un total de TOTAL 2 BATCH. Mantenga pulsado el botón DISPLAY durante 3 segundos para restablecer el lote total a cero.
Cuando el líquido fluye a través del medidor, un icono de hélice está resaltado.
Atributo del Indice de Flujo
Para utilizar esta función, pulse el botón DIS­PLAY y la FLOWRATE hasta que el icono apa­rece. La fábrica del conjunto de base de tiempo se pondrá de relieve a la derecha del caudal de agua (M = minutos, h = horas, D = días). Cuando el cuadal de agua fluye, en la pantalla se indica la velocidad de flujo.
8
Loading...
+ 22 hidden pages