Omega Products DPGM409S15-001BA User Manual

DIGITAL PRESSURE GAUGE FOR SANITARY/CLEAN-IN-PLACE APPLICATIONS
HIGH 0.08% ACCURACY
Gauge, Absolute, Sealed Gauge Compound Gauge, Vacuum and Barometric Models
0-25 mb to 0-41 bar
DPGM409S Metric
Sanitary Series
®
PRESSURE
GAUGES
G
Standard
U Easy to Read 25 mm (1")
U High 0.08%
Accuracy Typical
U Hi and Lo Alarms
Standard
U MIN/MAX/AVG Reading U User Selectable Analog
Output 0 to 5 Vdc, 0 to 10 Vdc or 4 to 20 mA
U Data Logging and
Charting Software Included
U Wireless Transmitter
Option Sends Reading to Remote Location
U Tested to Industrial
CE Standards
U Battery Powered with
Long Battery Life
U 5-Point NIST Traceable
Calibration Certificate
Fast Delivery –
Typically Stock to
1-Week!
OMEGA's DPGM409S Sanitary/ CIP digital pressure gauges with metric ranges incorporates a rugged 316 SS enclosure designed for washdown and marine applications with a precision 0.08% accuracy pressure transducer with 316L SS wetted parts and body. The result is an extremely durable digital pressure gauge designed to survive harsh industrial environments. The large 25 mm (1") display is easy to read at a distance up to 11 m (36'). Models are available in gauge, sealed gauge, compound gauge, absolute, vacuum and barometric pressure ranges. Standard features include a highly stable silicon sensor, selectable MIN/MAX and AVG readings, Hi and LO alarms and an externally powered user selectable analog output.
IP65 case.
DPGM409S15-17.5BG shown smaller than actual size.
An optional wireless transmitter feature with software allows the signal to be transmitted to a remote location such as a PC for charting and data logging. Power is from an internal long life battery (included) or an externally supplied dc voltage source. Each unit is supplied with configuration software that allows the user connect to a pc via a standard USB connection (cable included) and select a variety of user configurable features including update rate, dampening, units, backlight and auto off. The analog output can also be independently scaled from the display.
G-1
SPECIFICATIONS
RoHS Compliant CE Compliant: Meets industrial
emissions and immunity standard IEC61326 for industrial locations
Accuracy: ±0.08% Best Straight Line Typ. (Linearity, Hysteresis and Repeatability combined)
Calibration: 5-Pt NIST traceable calibration with zero and span values, calibrated in horizontal direction (compound gauge models calibrated in positive direction only). calibrated in horizontal direction
Zero Balance: ±0.5% FS typical ±1% maximum (for ranges 170 mb ±1% Typical ±2% maximum)
Span Setting: ±0.5% FS typical ±1% maximum (for ranges 170 mb ±1% Typical ±2% maximum)
Storage/Operating Temperature Range: -15 to 115°C (-5 to 240°F)
Compensated Temperature Range:
Ranges 350 mb: 4.4 to 60°C
(40 to 140°F) Ranges >350 mb: -4 to 85°C (25 to 150°F)
Zero and Span shift over Compensated Range: Ranges 170 mb: Zero: 1.00%
Span: 0.70% Range = 350 mb: Zero: 0.80% Span: 0.50% Ranges > 350 mb: Zero: 0.50% Span: 0.50% 550 & 880HB Barometric Range: Zero: 0.90% Span: 0.50%
Minimum Isolation Between Case and Output Terminations:
100M Ω @ 50 Vdc Analog Output: User selectable 0 to
5 Vdc, 0 to 10 Vdc of 4 to 20 mA (external 24 Vdc power required)
Note: Compound gage ranges have a unidirectional analog output equal to their full span; i.e. -1 to +1 bar = 0 to 10 Vdc.
Set Jumper to Select Current or Software selection of 0 to 5 Vdc, Vdc, or 4 to 20 mA
Analog Output Cable: 2.7 m (8.9') with ferrite core included
Analog Output Accuracy: 0.08% typical (0.15% max in RF field @ 10V/m)
Pressure Cycles: 250,000, minimum Long Term Stability (1-Year): ±0.1%
FS Typical
Voltage:
0 to 10
Fill Fluid: NEOBEE M5 Food Grade Approved to 3A Sanitary Specification 74-03
Environmental Protection: NEMA 4X (IP65) depending upon electrical termination
Pressure Rating: 1½" and 2" Tri-Grip™ fitting (compatible the Tri-Clamp
®
):
40 bar using suitable clamp Over Pressure: 4 x rated pressure to
maximum rated flange pressure Secondary Containment: 6 x rated
pressure to a maximum of 138 bar (not in clamp)
Display: 4-digit LCD with user selectable backlighting, (9999 count) 25 mm (1") character height
Computer Interface: UBS connection for set-up, cable included
Sample/Display Range: User selectable from 0.35 to 30 seconds (preset to 1/second)
Power: One 3.6V lithium 8.4 Ah capacity (C cell) Included (two included with wireless option) (replacement battery model no. BATT-C-3V)
Battery Life (Typical): Up to 4-years Enclosure Material: 316 stainless
Steel and ABS center gasket
Enclosure Finish: Electropolished Enclosure Environmental Rating:
Weatherproof, NEMA 4X (IP65) Connection Location: Lower
(head rotates to position display)
Front Panel Functions
Backlight: Manually turns backlight
on/off
Min/Max: Recalls min/max values Zero/Clear: Zeros display or clears
min/max
Software Settings (via USB)
Update Rate: 4 options; 10, 5, 2, 1
times per second
Units: psi, inHg, inH Lock: Allows the lock-out of front
panel functions Alarms: User selectable High and
Low alarm limits (open collector) Backlight: On/Off, 10 sec, 30 sec,
1 min, 5 min Wireless Transmitter (Optional):
channel number, Transmission rate, alarms, sensor offset, chart recording, data logging
Calibrate: Zero and Span Overrange Indication: "_ _ _ _"
O, bar, mbar
2
Wireless Option Specifications
Transmit Sample Rate: User
programmable from 1 sample/2 minute to 1 sample/2 seconds
Radio Frequency (RF) Transceiver Carrier: ISM 2.4 GHz
RF Output: 10 dBm (10 mW) Range of RF Link:
Outdoors (Line of Sight):
Up to 120 m (400')
Indoors/Urban: Up to 40 m (130')
Included Software: Requires Windows
7 (32 bit); XP, Vista Data Transmission to Host: Pressure
reading, ambient temperature reading, RF transmit strength and battery level
Dimensions: Display: 115 D x 89 W x 74 mm H
(4.5 x 3.5 x 2.9")
Sensor: 72 to 89 mm L (2.8 to 3.5")
below display
Weight: 900 g (2 lb) typical depending upon configuration
DPGM409S15-17.5BG,
shown smaller than actual size.
®
G-2
Loading...