Omega Products DPGM409-001BDWU User Manual

DIGITAL DIFFERENTIAL PRESSURE GAUGES wITH METRIC FITTINGS AND RANGES
0.08% ACCURACY wITH 316L SS wETTED PARTS AND BODY
Uni-Directional Differential Pressure Ranges
0-25 mb to 0-70 bar (0-10 inH
0 to 0-1000 psi)
2
DPGM409
PRESSURE
GAUGES
G
Differential Series
Standard
Fast Delivery –
Typically Stock to
2-weeks!
U Wet/Wet Differential
Pressure Measurement
U 5-Point NIST Traceable
Calibration
Backlit LCD Display
U High 0.08% Accuracy U Hi and Lo Alarms
Standard
U MIN/MAX/AVG Reading U User Selectable Analog
Output 0 to 5 Vdc, 0 to 10 Vdc or 4 to 20 mA
U Bar Graph on Display
Magnetic stylus (included).
DPGM409-17.5BDWU,
shown smaller than
actual size.
U Wireless Transmitter
Option Sends Reading to Remote Location and Includes Charting and Data Logging Software
U CE Certified U Battery Powered with
Long Battery Life
U NEMA 4X (IP65)
Enclosure
U Metric Ranges and
Pressure Connections
IP65 case.
Standard integral pressure sensor
OMEGA’s DPGM409 digital differential pressure gauge incorporates a rugged 316 SS enclosure designed for washdown and marine applications with a precision 0.08% accuracy pressure transducer and 316L SS wetted parts and body. The result is an extremely durable digital pressure gauge. You get all the durability and stability of the micromachined silicon sensor in a rugged NEMA 4X (IP65) package.
G-M1
SPECIFICATIONS
CE Compliant: Meets EN1326-1: 2006
for industrial locations with analog output cables <3 m (9.8') and supplied ferrite core
Accuracy (Combined Linearity, Hysteresis and Repeatability):
±0.08% BSL Storage Temperature: -40 to 82°C
(-40 to 180°F) Operating Temperature: -18 to 66°C
(0 to 150°F)
Thermal Effects (Over Operating Range):
Zero Balance:
Ranges 25 mb to 70 mb: ±1% span Ranges >70 mb: ±0.5% span
Span Setting:
Ranges 25 mb to 70 mb: ±1% span Ranges >70 mb: ±0.5% span
Minimum Resistance Between Transducer Body and Any Wire:
100M Ω @ 50 Vdc
Pressure Cycles: 1 million minimum Long Term Stability (1-yr.):
±0.1% FS typical Analog Output: User selectable 0 to
5 Vdc, 0 to 10 Vdc or 4 to 20 mA (external 24 Vdc power supply required)
Notes: 1. Bidirectional ranges have a unidirectional analog output equal to their full span; i.e. -15 to +15 psi = 0 to 10 Vdc.
2. Analog output is independently scalable
Set jumper to Select Current or Software selection of 0 to 5 Vdc, 10 Vdc, or 4 to 20 mA
Analog Output Cable: 2.7 m (9') with ferrite core (included)
Analog Output Accuracy:
0.08% typical (0.15% maximum in RF field @ 10V/m)
Line/Static Pressure: 500 psi maximum applied to both sides simultaneously
Proof Pressure (differential): 25 mb: = 10 times range 70 mb: = 6 times range 170 mb to 50 Bar: = 4 times range 70 Bar Range: = 3 times range
Hi-Side Containment Pressure (Differential):
25 to 350 mb: To 1000 psi 1 to 70 bar: To 3000 psi
Display: 4-digit LCD with user selectable backlighting, (9999 count) 25 mm (1") character height
Sample/Display Range: User selectable from 0.38 to 30 seconds (preset to 1/second)
Voltage:
0 to
New security feature uses magnetic stylus (included) for keypad activation. No accidental button presses. Perfect for dirty hands or if wearing gloves.
Computer Interface: USB connection
for set-up, cable included Power: One 3.6V lithium 8.4 Ah
capacity (C cell) included (two included with wireless option) (replacement battery model no. BATT-C-3V)
Battery Life (Typical): Up to 4-years Enclosure Material: 316 Stainless
Steel and ABS center gasket
Enclosure Finish: Electropolished Enclosure Environmental Rating:
Weatherproof, NEMA 4X (IP65) Sensor Wetted Parts:
316L stainless steel
Pressure Connection: G ¼ male Dimensions:
Display: 115 D x 89 W x 74 mm H
(4.5 x 3.5 x 2.9")
Sensor: 72 to 89 mm L (2.8 to 3.5")
below display, 80 mm (3.1") wide"
Weight: 1 kg (2.2 lb) typical depending upon configuration
Front Panel Functions
Backlight: Manually turns backlight
on/off
Min/Max: Recalls min/max values Zero/Clear: Zeros display or clears
min/max
Software Settings (via USB)
(USB connection is for set-up only) Units: psi, inHg, inH
O, bar, mbar
2
Lock: Allows the lock-out of front panel functions
Alarms: User selectable “High” and “Low” alarm limits (open collector)
Backlight: On/Off, 10 seconds, 30 seconds, 1 minute, 5 minute
Wireless Transmitter (Optional): Channel number, transmission rate, alarms, sensor offset, chart recording, data logging
Calibrate: Zero and Span Overrange Indication: “_ _ _ _”
Wireless Option Specifications
Transmit Sample Rate: User
programmable from 1 sample/2 min to 1 sample/2 sec
Radio Frequency (RF) Transceiver Carrier: ISM 2.4 GHz
RF Output: 10 dBm (10 mW) Range of RF Link: Up to 120 m (400')
outdoors (line of sight); Up to 40 m (130') indoors/urban
Included Software: Requires Windows
Data Transmission to Host: Pressure reading, ambient temperature reading, RF transmit strength and battery level
®
7 (32 bit); XP, Vista
G-M2
Loading...