Omega Products DPGM409-001BA User Manual

DIGITAL PRESSURE GAUGES WITH METRIC FITTINGS AND RANGES
0.08% ACCURACY WITH STAINLESS STEEL WETTED PARTS AND BODY
Gauge Pressure: 0-25 mb to 0-345 bar
Sealed Gauge: 0-7 bar to 0-345 bar Absolute Pressure: 0-350 mb
to 0-345 bar Compound Gauge: ±25 mb to
±1 bar Vacuum (Negative Gauge):
0 to -25 mb to 0 to -1 bar Barometric: 0-1100 hPa to
880-1100 hPa
DPGM409 Series
GAUGES
PRESSURE
G
Standard
U 5-PT NIST Traceable
Calibration Included
U High 0.08% Accuracy U Easy to Read 25 mm (1")
U Hi and Lo Alarms
Standard
U MIN/MAX/AVG Reading U User Selectable Analog
Output 0 to 5 Vdc, 0 to 10 Vdc or 4 to 20 mA
U Included Setup and
Service Software Allows for Fast Maintenance and Calibration via USB Connection
U Wireless Transmitter
Optional
U Wireless Models Include
Data Logging Software for Remote Recording and Charting
U CE Certified U Battery Powered with
Long Battery Life
U Analog Output Standard
Magnetic stylus (included).
DPGM409-070BG, shown
smaller than actual size.
Fast Delivery –
Typically Stock to
2-Weeks!
OMEGA’s DPGM409 digital pressure gauge incorporates a rugged 316 SS enclosure designed for washdown and marine applications with a precision 0.08% accuracy pressure transducer with 316L SS wetted parts and body. The result is an extreamly durable digital pressure gage designed to survive harsh industrial environments. The large 25 mm (1") display is easy to read at a distance up to 11 m (36'). Models are available in Gauge, Sealed Gauge, Compound Gauge, Absolute, Vacuum and Barometric pressure ranges. Standard features include a highly stable silicon sensor, selectable MIN/MAX and AVG readings, Hi and LO alarms and an
NEMA 4X case.
Standard integral pressure sensor
externally powered user selectable analog output. An optional wireless transmitter feature with software allows the signal to be transmitted to a remote location such as a PC for charting and data logging. Power is from an internal long life battery (included) or an eternally supplied dc voltage source. Each unit is supplied with configuration software that allows the user connect to a PC via a standard USB connection (cable included) and select a variety of user configurable features including update rate, units, and backlight.
G-M1
SPECIFICATIONS
RoHS Compliant CE Compliant: Meets EN1326-1: 2006
for industrial locations with analog output cables less than 3 m (9.8') and supplied ferrite core
Accuracy (Combined Linearity, Hysteresis and Repeatability):
±0.08% BSL (0.15% max in 10V/m RF field) (compound gauge calibrated in positive direction only)
Temperature: Storage: -40 to 82°C (-40 to 180°F)
Compensated: -18 to 66°C
(0 to 150°F)
Thermal Effects (Over Compensated Range): Zero Shift: Ranges >350 mb: ±0.3% span
Ranges 350 mb: ±0.5% span
Span Shift: Ranges >350 mb: ±0.3% span
Ranges 350 mb psi: ±0.5% span
Minimum Resistance Between Transducer Body and Any Wire:
100MΩ @ 50 Vdc
Pressure Cycles: 1 million minimum Long Term Stability (1-yr):
±0.1% FS typical Analog Output: User selectable
0 to 5 Vdc, 0 to 10 Vdc or 4 to 20 mA (external 24 Vdc power supply required)
Note: Compound gauge ranges have a unidirectional analog output equal to their full span; i.e. -350 to 350 mb = 0 to 10 Vdc.
Analog Output: Cable: 2.7 m (8.9') with
ferrite core included Accuracy: 0.08% typical (0.15% max in RF field @ 10V/m)
Overpressure: Gauge/sealed gauge 25 mb: 10 times span
70 mb: 6 times span 170 mb to 245 bar: 4 times span 345 bar: 1034 bar max
Absolute/Barometric Ranges: 350 mb abs: 6 times span
Ranges >350 mb to 245 bar abs: 4 times span
345 bar Range: To 1034 bar abs Containment Pressure: Gauge and
Sealed Gauge pressure 25 to 350 mbar: To 70 bar
70 mb to 70 bar: To 210 bar 100 to 345 bar: To 1034 bar
Absolute/Barometric Pressure: 350 mb to 70 bar: to 400 bar
Display: 4-digit LCD with user selectable backlighting, (9999 count) 25 mm (1") character height
Computer Interface: USB connection for set-up, cable included
New security feature uses magnetic stylus (included) for keypad activation. No accidental button presses. Perfect for dirty hands or if wearing gloves.
Update Rate: User selectable from 0.25
to 10 sec. (preset to 1/sec) Power: One 3.6V lithium 4.8 Ah
capacity (C cell) Included (two with wireless option) (replacement battery
BATT-C-3V) Battery Life (Typical): Up to 4-years Enclosure:
Material: 316 Stainless Steel and
ABS center gasket
Finish: Electropolished Environmental Rating:
Weatherproof, NEMA 4X (IP65)
Sensor Wetted Parts: 316L Stainless Steel
Pressure Connection: G
1
4 male
Dimensions: Display: 115 L x 89 W x 74 mm D
(4.5 x 3.5 x 2.9") Sensor: 72 to 89 mm L (2.8 x 3.5") below display
Weight: 900 g (32 oz) Typical depending upon configuration
Front Panel Functions
Backlight: Manually turns backlight on/off Min/Max: Recalls min/max values Zero/Clear: Zeros display or clears min/
max
Software Settings (via USB)
Update Rate: 4 options, 10, 5, 2, 1
times per second Units: psi, inHg, inH
units; must be specified at time of order Lock: Allows the lock-out of
front panel functions
O, bar, mbar, hPa
2
Alarms: User selectable “High” and “Low” alarm limits (open collector)
Analog Output: User Selectable 0 to 5 Vdc, 0 to 10 Vdc, or 4 to 20 mA (requires external 24 Vdc power)
Backlight: On/off, 10 sec, 30 sec, 1 min, 5 min
Wireless Transmitter (Optional): Channel number, transmission rate, alarms, sensor offset, chart recording, data logging
Overrange Indication: “_ _ _ _”
Wireless Option Specifications
Transmit Sample Rate: User
programmable from 1 sample/2 min to 1 sample/2 sec
Radio Frequency (RF) Transceiver Carrier: ISM 2.4 GHz
RF Output: 10 dBm (10 mW) Range of RF Link:
Outdoors (Line of Sight):
Up to 120 m (400')
Indoors/Urban: Up to 40 m (130')
Included Software: Requires Windows
7 (32 bit); XP or Vista Internal Battery: Two 3.6V lithium
4.8 Ah capacity (C-cell) included, (replacement battery BATT-C-3V)
Data Transmission to Host: Pressure reading, ambient temperature reading, RF transmit strength and battery level
®
G-M2
Loading...
+ 1 hidden pages