DIGITAL DIFFERENTIAL PRESSURE
GAUGE VERY HIGH ACCURACY
316L SS WETTED PARTS AND BODY
Uni-Directional Ranges
0-10 inH2O to 0-1000 psi
(0-25 mb to 0-70 bar)
DPG409 Differential
Pressure Series
Standard
U Wet/Wet Differential
Pressure Measurement
U 5-Point NIST Traceable
Calibration Included
U Easy to Read 25 mm (1")
Backlit LCD Display
U High 0.08% Accuracy
U Hi and Lo Alarms
Standard
U MIN/MAX/AVG Reading
U User Selectable Analog
Output 0 to 5 Vdc, 0 to
10 Vdc or 4 to 20 mA
U Wireless Transmitter
Option Sends Reading to
Remote Location and
Includes Charting and
Data Logging Software
U CE Certified
U Battery Powered with
Long Battery Life
U
NEMA 4X (IP65) Enclosure
OMEGA’s DPG409 digital differential
pressure gauge incorporates a rugged
316 SS enclosure designed for
washdown and marine applications
with a precision 0.08% accuracy
pressure transducer and 316L SS
wetted parts and body. The result is
an extremely durable digital pressure
gauge. You get all the durabilty and
stability of the micromachined silicon
sensor in a rugged NEMA 4X
(IP65) package.
SPECIFICATIONS
CE Compliant: Meets
EN1326-1: 2006 for industrial
locations with analog output cables
<3 m (9.8') and supplied ferrite core
Accuracy (Combined Linearity,
Hysteresis and Repeatability):
±0.08% BSL
Storage Temperature: -40 to 82°C
(-40 to 180°F)
Operating Temperature: -18 to 66°C
(0 to 150°F)
Thermal Effects (Over
Operating Range):
Zero Balance:
Ranges 10 inH
Span Setting:
Ranges 10 inH
Minimum Resistance Between
Transducer Body and Any Wire:
100M Ω @ 50 Vdc
Pressure Cycles: 1 million min
Long Term Stability (1-Yr.): ±0.1%
FS typical
Analog Output: User selectable 0 to
5 Vdc, 0 to 10 Vdc or 4 to 20 mA
(external 24 Vdc power supply required)
Note: Analog output is independently
scalable
Set jumper to select current or
Analog Output Cable: 2.7 m (9') with
ferrite core (included)
Analog Output Accuracy: 0.08%
typical (0.15% maximum in RF field @
10V/m
Line/Static Pressure: 500 psi maximum
applied to both sides simultaneouly
Proof Pressure (Differential):
10 inH
Hi-Side Containment Pressure
(Differential):
10 inH
15 to 1000 psi: To 3000 psi
Display: 4-digit LCD with user
selectable backlighting, (9999 count)
25 mm (1') character height
Sample/Display Rate: User selectable
from 0.38 to 30 sec (preset to 1/sec)
Magnetic stylus
included.
Fast Delivery –
Typically Stock to
2 Weeks!
O to 1 psi: ±1% span
Ranges >1 psi: ±0.5% span
Ranges >1 psi: ±0.5% span
Software selection of 0 to 5 Vdc,
0 to 10 Vdc, or 4 to 20 mA
2
1 psi = 6 times range
2.5 to 750 psi = 4 times range
1000 psi Range = 3 times range
O to 5 psi: To 1000 psi
2
2
O to 1 psi: ±1% span
2
O = 10 times range
G-1
voltage.
Computer Interface: USB connection
for set-up, cable included
Power: One 3.6V lithium 8.4 Ah
capacity (C cell) included (two included
with wireless option) (replacement
battery model no. BATT-C-3V)
Battery Life (Typical): Up to 4-years
Enclosure Material: 316 Stainless
Steel and ABS center gasket
Enclosure Finish: Electropolished
Enclosure Environmental Rating:
Weatherproof, NEMA 4X (IP65)
Sensor Wetted Parts: 316L
Stainless Steel
Pressure Connection: ¼-18 NPT male
Dimensions:
Display: 115 D x 89 W x 74 H mm
(4.5 x 3.5 x 2.9")
Sensor: 72 to 89 mm L (2.8 to 3.5")
below display, 80 mm (3.1") wide
Weight: 1 kg (2.2 lb) Typical depending
upon configuration
Front Panel Functions
Backlight: Manually turns backlight
on/off
Min/Max: Recalls min/max values
Zero/Clear: Zeros display or clears
min/max
PRESSURE
GAUGES
G
DPG409-050DWU
shown smaller
than actual size.
Software Settings (via USB)
(USB connection is for set-up only)
Units: psi, inHg, inH2O, bar, mbar
Lock: Allows the lock-out of front
panel functions
Alarms: User selectable “High” and
“Low” alarm limits (open collector)
Backlight: On/Off, 10 sec, 30 sec,
1 min, 5 min
Wireless Transmitter (Optional):
Channel number, transmission rate,
alarms, sensor offset, chart recording,
data logging
Calibrate: Zero and span
Overrange Indication: “_ _ _ _”
To Order Visit omega.com/dpg409_diff for Pricing and Details
WIRELESS OPTION
Transmit Sample Rate: User
programmable from 1 sample/2 min to
1 sample/2 sec
Radio Frequency (RF)
Transceiver Carrier: ISM 2.4 GHz
RF Output: 10 dBm (10 mW)
Range of RF Link: Up to 120 m (400')
outdoors (line of sight); Up to
40 m (130') indoors/urban
Included Software: Requires
Windows® 7 (32 bit); XP, Vista
Data Transmission to Host: Pressure
reading, ambient temperature reading,
RF transmit strength and battery level
RANGE/
RESOLUTION MODEL NO.
psi BAR
1
⁄4 NPT MALE THREAD
WET/WET UNI-DIRECTIONAL DIFFERENTIAL PRESSURE MODELS
UWTC-REC2-D,
48-channel wireless receiver
with analog output and display.
0 to 10.00 inH
0 to 1.000 psi 0 to 70 mbar DPG409-001DWU
0 to 2.500 0 to 170 mbar DPG409-2.5DWU
0 to 5.000 0 to 350 mbar DPG409-005DWU
O 0 to 25 mbar DPG409-10WDWU
2
0 to 15.00 0 to 1 DPG409-015DWU
0 to 30.00 0 to 2 DPG409-030DWU
0 to 50.00 0 to 3.5 DPG409-050DWU
0 to 100.0 0 to 7 DPG409-100DWU
0 to 150.0 0 to 10 DPG409-150DWU
0 to 250.0 0 to 17.5 DPG409-250DWU
0 to 500.0 0 to 35 DPG409-500DWU
0 to 750.0 0 to 50 DPG409-750DWU
0 to 1000 0 to 70 DPG409-1KDWU
WET/WET UNI-DIRECTIONAL DIFFERENTIAL PRESSURE MODELS
WITH WIRELESS TRANSMITTER
Wireless Receivers (For Use with
Wireless Transmitter Option)
MODEL NO. DESCRIPTION
UWTC-REC1
receiver, USB powered
UWTC-REC2-(*) 48-channel wireless
receiver with analog
output
UWTC-REC2-D-(*)
* Specify analog output signal: “V1” for 0 to 5 Vdc;
“V2” for 0 to 10 Vdc, “TC” for Type K thermocouple,
or “MA” for 4 to 20 mA.
Comes complete with operator’s manual, battery, set-up and configuration software (charting and databasing software with common drivers)
USB cable, 2.7 m (9') analog output cable with ferrite core and 5-point NIST traceable calibration certificate.
To order wireless option add suffix “-W” to model number for additional cost.
Ordering Examples: DPG409-015DWU, 0 to 15 psi uni-directional differential pressure gauge with analog output.
DPG409-100DWU-W, 0 to 100 psi uni-directional differential pressure gauge with analog output and wireless transmitter.
48-channel wireless
48-channel wireless
receiver with analog
output and digital display
0 to 10.00 inH
0 to 1.000 psi 0 to 70 mbar DPG409-001DWU-W
0 to 2.500 0 to 170 mbar DPG409-2.5DWU-W
0 to 5.000 0 to 350 mbar DPG409-005DWU-W
0 to 15.00 0 to 1 DPG409-015DWU-W
0 to 30.00 0 to 2 DPG409-030DWU-W
0 to 50.00 0 to 3.5 DPG409-050DWU-W
0 to 100.0 0 to 7 DPG409-100DWU-W
0 to 150.0 0 to 10 DPG409-150DWU-W
0 to 250.0 0 to 17.5 DPG409-250DWU-W
0 to 500.0 0 to 35 DPG409-500DWU-W
0 to 750.0 0 to 50 DPG409-750DWU-W
0 to 1000 0 to 70 DPG409-1KDWU-W
O 0 to 25 mbar DPG409-10WDWU-W
2
G-2