Omega Products DPG4000-SW Installation Manual

omega.com
e-mail: info@omega.com
For latest product manuals:
omegamanual.info
User ’s Guide
M-4934/0710
DPG4000-SW Data Logging Software
Shop online at
USA: One Omega Drive, Box 4047
ISO 9001 Certified Stamford CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378 / 1-800-622-BEST
®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
®
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: En Espan˜ ol: (001) 203-359-7803 e-mail: espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
Servicing Europe:
Benelux: Postbus 8034, 1180 LAAmstelveen, The Netherlands
Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Czech Republic: Frystatska 184, 733 01 Karvina´, Czech Republic
Tel: +420 (0)59 6311899 FAX: +420 (0) 5 9 6 3 1111 4 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
France: 11, rue Jacques Cartier, 78280 Guyancourt, France
Tel: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427 Toll Free in France: 0800 466 342 e-mail: sales@omega.fr
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5BD United Kingdom Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
OMEGAnet®Online Service Internet e-mail
www.omega.com info@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
1. Introduction
This Manual describes the data logging capability and operation of the pressure gauge, plus the operation of the related DPG4000-SW Microsoft Windows application. All DPG4000's with firmware with revision 3.0 or higher are capable of data logging when used with the DPG4000-SW software. This manual complements and extends the user manual received with the pressure gauge. Please refer to the gauge manual for additional information on its operation.
The DPG4000-SW Data Logging Software application is used to configure logging parameters in the gauge, and to retrieve logged data from the gauge for storage and analysis on a personal computer. Before you start check to make sure your data logging kit came with the following items:
1) DPG4000-SW CD
2) RS -232 Cable with stereo style jack (this end plugs into the gauge)
3) RS -232 to USB adapter (with mini-CD) for newer computers that only have USB ports. Refer to the instructions packaged with the USB adapter to load the driver and operate the adapter.
4) Users manual
Refer to figure 1 to see how to connect the RS-232 cable to the Gauge.
Figure 1
2. Description
The gauge can log multiple sets of data, each of which is identified by a sequential run number. Data logging runs are started and stopped using new items inserted into the gauge's CONFIG menu sequence. Logging parameters can be changed from one data logging run to the next, and data logging runs may be as short as a single sample or as
1
long as desired subject to the maximum storage capacity of the gauge. The gauge can log pressure only for each sample, or it can log
pressure and temperature for each sample. This selection applies to all runs and must be made prior to starting the first run.
When logging pressure only, the gauge has a capacity of 8500 data records for storing run identification data and log sample data. Each new data logging run requires 7 records for run identification overhead. One record is required to log each sample, except for the triple value case (average, minimum, maximum) where 3 records are required to log each sample set.
Assuming a single run, the maximum capacity for all but the triple value case is 8493 samples, and the maximum capacity for the triple value case is 2831 sample sets.
When logging pressure and temperature, the gauge has a capacity of 5100 data records for storing run identification data and log sample data. Each new data logging run requires 7 records for run identification overhead. One record is required to log each sample, except for the triple value case (average, minimum, maximum) where 3 records are required to log each sample set.
Assuming a single run, the maximum capacity for all but the triple value case is 5093 samples, and the maximum capacity for the triple value case is 1697 sample sets.
The interval time setting determines the frequency of sample calculation and storage. Eleven discrete interval times are available, ranging from 1 second to 1 hour.
A sample value is calculated at the end of each interval based on all of the pressure measurements made during that interval. The sample value may be the last measurement made in the interval, or it may be the average, minimum, maximum, or median of all of the measurements made during that interval.
A sample value may be stored for every interval during the run, only for those intervals where the sample value exceeds a specified high or low limit, or only for those intervals where the pressure differential during the interval exceeds a maximum span.
The data logging function is disabled by default in all new gauges. The DPG4000-SW application is used to enable and disable the data log function in a gauge.
Two data logging modes are available, Download and Demand. Multiple log runs are allowed in either mode, with the potential number of runs dependent on the number of samples logged in each run and the capacity of the gauge.
When the data logging mode is set to Download, the DPG4000-SW application is used to enter or change the data logging parameters and
2
send them to the gauge. The gauge remains in this mode until either the DPG4000-SW application or the gauge 'CLEAr DATA' function is used clear the logged data in the gauge, at which time the mode is set to Demand. In Download mode none of the parameters may be changed using the gauge menus and the downloaded values are always used for each new log run. The DPG4000-SW application may be used to change the logging parameters between runs without affecting previously logged data stored in memory. The Download mode data logging parameters are:
• Interval time selection:
- 1, 2, 5, 10, 30 seconds
- 1, 2, 5, 10, 30 minutes
- 1 hour
• Maximum number of samples to store per run. This may range from
1 to the full capacity of the gauge.
• Log type selection:
- Continuous. This is the most commonly used log type where a
sample is stored at every interval during the run. The sample value stored is defined by the sample type selection below. The run can be stopped manually, or stops automatically when the log memory is full or the maximum number of samples per run is logged.
- Trip high, where interval samples are stored during the run only
when the sample values exceed a high limit. Sample storage starts when an interval's sample value becomes greater than the trip setpoint, and continues at every interval until a subsequent interval's sample value becomes less than the trip reset. Sample storage then stops until a sample value again becomes greater than the trip setpoint. The sample value stored, and used for comparison, is defined by the sample type selection below. The run can be stopped manually, or stops automatically when the log memory is full or the maximum number of samples per run is logged.
- Trip low, where interval samples are stored during the run only
when the sample values exceed a low limit. Sample storage starts when an interval's sample value becomes less than the trip setpoint, and continues at every interval until a subsequent interval's sample value becomes greater than the trip reset. Sample storage then stops until a sample value again becomes less than the trip setpoint. The sample value stored, and used for comparison, is defined by the sample type selection below. The run can be stopped manually, or stops automatically when the log memory is full or the maximum number of samples per run is logged.
3
- Trip Range, where interval samples are stored during the run
only when the pressure differential during the interval exceeds a maximum span or range of values. Sample storage starts when the difference between an interval's maximum and minimum measurements becomes greater than the trip setpoint, and continues at every interval until a subsequent interval's difference becomes less than the trip reset. Sample storage then stops until an interval's difference again becomes greater than the trip setpoint. The run can be stopped manually, or stops automatically when the log memory is full or the maximum number of samples per run is logged.
• Sample type selection:
- Last measurement made in the sample interval.
- Average of all measurements made in the sample interval.
- Minimum of all measurements made in the sample interval.
- Maximum of all measurements made in the sample interval.
- Median of all the measurements made in the sample interval, the
midpoint between minimum to maximum measurements.
- Average, minimum, and maximum of all measurements made in
the sample interval. This requires 3 records of storage per logged sample set and reduces capacity by a factor of 3.
• Log pressure unit selection. All supported pressure units including
custom.
• Trip setpoint expressed in log unit selection.
• Trip reset expressed in log unit selection.
• Log custom pressure unit factor.
When the data logging mode is set to the Demand mode the user can set the gauge up to data log without the need to first connect it to the computer. The gauge menus are used to enter and change some data logging parameters, while the remaining parameters default to fixed values. The gauge menus may be used to change the logging parameters between runs without affecting previously logged data stored in memory. The demand mode data logging parameters are:
• Interval time selection:
- 1, 2, 5, 10, 30 seconds
- 1, 2, 5, 10, 30 minutes
- 1 hour
• Maximum number of samples to log. This may range from 1 to the
full capacity of the gauge.
• Log type selection fixed at 'Continuous'.
4
• Sample type selection fixed at 'Last measurement made in the
sample interval'.
• Log pressure unit selection. The unit is selected with the existing
configuration function before the log run is started.
• Trip setpoint not used.
• Trip reset not used.
• Log custom pressure unit factor. The factor is selected with the
existing configuration function before the log run is started
The gauge does not have real time clock capability, so the time stamp on all samples in a run is elapsed days, hours, minutes and seconds from run start. The DPG4000-SW application provides a feature for relating each uploaded run start point to a real time base date and time. When this feature is used, all uploaded elapsed time stamps in the run are recalculated accordingly to get true time stamps of the form year, month, day, hour, minute, second.
3. Gauge Operation
3.1 Data Logging Disabled
Until data logging functionality is enabled using the DPG4000-SW application, the gauge operates as described in the user manual received with the gauge. You must connect the gauge to the
DPG4000-SW software to enable the logging function. This is true for both Demand and Download modes.
3.2 Download Mode
Once data logging functionality is enabled, a new item is inserted at the start of the gauge configuration menu tree to direct the user to a completely separate branch which contains all data logging functions. This new menu item shows 'LOG' in the main display and 'DATA' in the secondary display. Press the CONFIG key to proceed to the existing configuration functions, or press the ENTER key to proceed to the new data logging branch. The data logging configuration menu sequence (new branch) is as follows:
1) Start or stop data logging
• Main display: LOG
• Secondary display: FULL if the log memory is full, OFF if not
logging data, else ON if logging data.
• the ▼ and ▲ keys are ignored if the log memory is FULL,
otherwise press either key to toggle between OFF and START if
5
not logging data, or to toggle between ON and STOP if logging data.
• press the CONFIG key to step to the next menu item.
• press the ENTER key to exit configuration if the display is FULL,
ON, or OFF, to stop logging and exit configuration if the display is STOP, or to start logging and exit configuration if the display is START.
2) Display number of free log records
• Main display: number of free records
• Secondary display: FREE
• the and keys are ignored.
• press the CONFIG key to step to the next menu item if not
logging, otherwise it exits configuration.
• press the ENTER key to exit configuration.
3) Erase all logged data and set to Demand mode
• Main display: CLEAr
• Secondary display: DATA
• the and keys are ignored.
• press the ENTER key to step to the next menu item and confirm
erasure.
• press the CONFIG to exit configuration without erasing the data.
4) Confirm erasure of all logged data and set to Demand mode
• Main display: CLEAr
• Secondary display: CNFRM
• the and keys are ignored.
• press the ENTER key to erase the data, set Demand mode, and
exit configuration after briefly displaying CLr to confirm erasure.
• press the CONFIG key to exit configuration without erasing the
data.
While not logging data, the gauge operates as described in the user manual received with the gauge.
When a log run is started, the pressure unit and custom factor are forced to the downloaded selections. They are restored to their original values at the end of the run.
While logging data, the following operating changes occur:
• Pressing the power button causes the log run to be stopped
before power is turned off.
6
• Forced power off on low battery detection causes the log run to
be stopped before power is turned off.
• The zero key is active and an event record (including the change
in the zero offset value) is placed in the log.
• Auto off is disabled. It is restored to its original value at the end
of the run.
• The engineering units, damp, sample rate, tare, and custom
factor items on the existing configuration menu are disabled.
• The units field on the display cycles between the units, the text
LOG, and the text RUN at a rate of once per second. This provides visual feedback that data logging is active, and the number after the text RUN identifies the run number for later reference.
3.3 Demand Mode
The Demand mode allows the user to conduct an impromptu data measurement/log without the need of a computer to configure all of the logging parameters.
Once data logging functionality is enabled, a new item is inserted at the start of the configuration menu tree to direct the user to a completely separate branch which contains all data logging functions. This new menu item shows 'LOG' in the main display and 'DATA' in the secondary display. Press the CONFIG key to proceed to the existing configuration functions, or press the ENTER key to proceed to the new data logging branch. The data logging configuration menu sequence (new branch) is as follows:
1) Start or stop data logging
• Main display: LOG
• Secondary display: FULL if the log memory is full, OFF if not
logging data, else ON if logging data.
• the and keys are ignored if the log memory is FULL,
otherwise press either key to toggle between OFF and START if not logging data, or to toggle between ON and STOP if logging data.
• press the CONFIG key to step to the next menu item.
• press the ENTER key to exit configuration if the display is FULL,
ON, or OFF, to stop logging and exit configuration if the display is STOP, or to start logging and exit configuration if the display is START.
7
Loading...
+ 19 hidden pages