1
/4" NPT
2.88"
3.38"
0.75"
1.65"
Turn at
hex
fitting
only!
0.75"
Two 3 ft long, 4conductor & 6
conductor
22 AWG shielded
cables
SPECIFICATIONS
Ranges & Resolution
30.0 inHg vacuum, ±15.00, 3.00, 5.00, 15.00 psig
30.0, 100.0, 199.9, 300, 500, 1000, 3000, 5000 psig
Absolute reference:15.00, 100.0 psia
Optional Units
Most engineering units such as kPa, atm, bar,
mbar, inHg, mmHg, inH2O, ftH2O, torr, kg/cm2,
cmH
2
O, oz/in
2
Display (type, size, update rate)
Ranges up to 1999: 3
1
/
2 digit LCD,
1
/
2" digit height
3000 & 5000 psi ranges: 4 digit LCD, 0.4" H digits
3 readings per second nominal display update
Accuracy (linearity, hysteresis, repeatability)
±0.25% of full scale or better, ±1 least significant
digit
Temperature Stability(rel. to 77°F or 25°C)
±1% FS for offset & span, 32 to 158°F (0 to 70°C)
typical
±2% FS for offset & span, 32 to 158°F (0 to 70°C)
typical for 3 and 5 psi ranges
Alarm Deadbands (hysteresis)
1% of of full scale
Alarm Outputs
Dual form C (SPDT) relay contacts
Individual Setpoint 1 and Setpoint 2 settings via
top-accessible multiturn potentiometers
HI/LO alarm configuration standard
Relay contacts rated 1A/24VDC, 0.5A/115VAC
maximum, non-inductive
Alarm Indicators
Bicolor (red/green) LEDs on front panel
Alarm Response Time
100 milliseconds typical
Test Function
Front panel TEST button toggles SP1 and SP2
alarm status and 4-20 mA output to test level,
independent of pressure input to allow testing of
system operation.
Retransmission Output
4-20 mA DC true analog output, 50 milliseconds
typical response time.Output drive (compliance)
determined by power source.
Power
DC 9 to 32 VDC, or AC 8 to 24VAC 50-60Hz
1.0 watt max.power consumption
Storage Temp:–40 to 203°F (–40 to 95°C)
Operating Temp:–4 to 185°F (–20 to 85°C)
Compensated Temp:32 to 158°F (0 to 70°C)
Weight (approximate)
Gauge: 9 ounces, shipping weight: 1 pound
Housing
Extruded aluminum case, light gray epo xy powder
coated
Black polycarbonate cover, front and rear gaskets
Electrical Connection
3 ft long, 4-conductor 22AWG cable for power
and 4-20 mA output.
3 ft long, 6-conductor 22AWG alarm relay cable
Pressure/Vacuum Connection & Material
1
/
4
" NPT male.All wetted parts are 316SS
Overpressure & Burst
5000 psig for 3000 psig range
7500 psig for 5000 psig range
All others 2x rated pressure minimum
Burst: 4x rated pressure minimum or 10,000 psi,
whichever is less
Using the 4-20 mA Output
The 4-20 mA output is driven by the transducer rather than the display and
thus is a true analog output.The output is filtered to improve noise immunity
and thus has a response time of about 50 msec.
The power supply (–) lead is tied to the 4-20 mA output ground. Therefore
the DC supply (–) lead should be considered common with regard to the 420 mA output (–) or ground connection.
Be sure to observe the output compliance (voltage drive) capabilities of the
gauge. The compliance, and therefore the maximum loop resistance the
output can drive, is a function of the supply voltage to the gauge. Consult
the graph below for maximum loop resistance vs. supply voltage.Too large a
loop resistance will cause the gauge output to “limit” or saturate before
reaching its full 20 mA output.
Calibration
These gauges are factory calibrated and normally do not require
recalibration before being put into service. If calibration is required, lift
calibration label on the top of unit to access the Zero and Span controls.this
label is designed to be reused many times.
For GAUGE reference units the setting of the Zero control is correct when
the gauge reads zero, with the “–” sign occasionally flashing, when the
pressure port is open to the ambient. Zero calibration must be done before
span calibration is attempted. Since the calibration potentiometers are non-
interactive, gauge reference units may be rezeroed without affecting the
span.
Span calibration should only be done when a pressure reference of known
accuracy is available. The calibration equipment accuracy should be better
than 4 times the accuracy of the gauge. Zero calibration must be done
before span calibration. Record readings at three to five points over the
range of gauge and adjust span control to minimize error and meet
specifications.
Calibration of ABSOLUTE reference units is more difficult and is not
recommended in the field, unless the user has access to equipment
required to calibrate absolute-reference units, such as vacuum generation
and atmospheric pressure measurement equipment.
Gauges can be returned to Omega Engineering for recalibration. NIST
traceability is available.
The DPG1000DAR has internal controls that adjust the “agreement”
between the displayed value and the 4-20 mA loop current. These are
carefully set at the factory and should not normally be adjusted. Should the
user feel this adjustment is necessary, the display zero/span must be
accurately set first. Then, after removing the rear cover, the “loop zero” and
“loop span” controls may be trimmed for 4.00 mA of loop current at the low
end of the range, and 20.00 mA at the high end of the range, respectively.
Accurate pressure generation and measurement, and current measurement
equipment are required to successfully complete this calibration.
OMEGA Engineering Inc./Made in USA
TEST
SP1 SP2
1400
1200
1000
800
600
400
200
0
Max Loop Resistance (Ohms)
8 121620242832
Supply V oltage
Voltage Compliance for 4-20 mA Output
PSIG
DC
AC
Gauge power supply
9-32 VDC
or
8-24 VAC 50/60 Hz
(–)BLACK
(+)RED
(+)WHITE
(–)GREEN
Remote device
Optional
shield
connection
Zero SP2
Span
Test
SP1
See "Electrical Connections"
for relay wiring