OHMMETER INDICATOR/CONTROLLER
1
⁄8 DIN
DP2000-O1 shown
actual size.
DP2000-0 Series
U 1999 or 9999 Count Display Span
U Analog Outputs Available
U Proportional Control Available
The DP2000/9000-O is a panel mount resistance meter
with direct readout in ohms. Resistance ranges are
jumper selectable. It is complete with current excitation
and open resistance detection. With the addition of
analog and control outputs, it can be incorporated as
a system component in ATE and closed-loop control
applications.
POWER AND DISPLAY OPTIONS
Six types of power supplies are available: 120 Vac,
240 Vac, 24 Vac, 5 Vdc, isolated 9 to 32 Vdc and
isolated 26 to 56 Vdc. An LED display is standard, an
LCD display is optional and is recommended for viewing
in bright ambient light. A NEMA (IP65) 4 splash-proof
lens cover is available.
Signal input and power connections are made via a
rear barrier terminal strip. The motherboard rear edge
connector provides access to hold and test, polarity,
clock, and the standard analog output and optional
analog outputs. Decimal point positions can be selected
by jumpers.
ANALOG OUTPUT OPTIONS
A 1 mV/count (±2 V full-scale) or 0.1 mV/count (±1 V
full-scale) analog output is standard and is ideal for
driving a strip-chart recorder. An additional analog
output can be provided by an optional vertical plugin board. Available output signals are 0 to 5 Vdc, 0 to
10 Vdc, 0 to 1 mA (source or sink) and 4 to 20 mA
(source or sink). The top and bottom of each output
range can be scaled to fit a user-selected display span.
CONTROL OUTPUT OPTIONS
Additional outputs can be provided by a horizontal
upper board. Available options include single-setpoint
control with one 10 A relay, dual-setpoint control
with two 10 A relays, 4 to 20 mA proportional control
(source or sink), time-proportional 2 A solid-state relay
control, and isolated, parallel BCD output.
SPECIFICATIONS
Input Configuration
Connection: 2- or 4-wire
Zero Adjustment: ±50 counts
Overvoltage Protection (Differential): 15 Vp
Open-Circuit Voltage (maximum): 12V
Conversion
Input Configuration: Differential, bipolar
Technique: Dual slope, average value
Polarity: Automatic
Signal Integration Period: 100 ms, nominal
Reading Rate: 2.5/s, nominal
Accuracy at 25°C
Overall Accuracy: ±0.05% of reading ±1 count
Span Tempco: ±0.006% of reading/°C
Zero Tempco: ±2.5 m/°C ±0.001% of full scale/°C
Warmup to Rated Accuracy:1 min
D-1
INSTRUMENTATION
D
Display
LED: Red, 14.2 mm (0.56"), 7-segment
LCD: 12.7 mm (0.50"), 7-segment
Power
AC Models: 120, 240, or 24 Vac 10%/-15%,
49 to 440 Hz
DC Models: 9 to 32 Vdc, isolated to 300 Vp;
26 to 56 Vdc, isolated to 300 Vp; 5 Vdc ±5%,
non-isolated
Common Mode
Voltage: 1500 Vp test (354 Vp per IEC spacing)
Rejection (DC to 60 Hz): 120 dB
Environmental
Operating Temperature: 0 to 60°C (32 to 140°F)
Storage Temperature: -40 to 85°C (-40 to 185°F)
Humidity: 95% RH, non-condensing @ 40°C (104°F)
Mechanical
Bezel: 96 W x 48 H x 8 mm D (3.78 x 1.89 x 0.31")
Depth Behind Bezel: 139.8 mm (5.50")
Panel Cutout: 92 W x 45 mm H (3.62 x 1.77")
Weight: 17 oz (480 g)
Case Material: 94 V-0 UL-rated polycarbonate
DC CURRENT INPUTS O SERIES
CODE*
O1
O2
O3
O4
Ordering Example: DP2000-O3, LED 120 Vac, 1mV/count, DC
current input 0 to 1.999 kΩ.
DP2000
RANGE
19.99 Ω – 10 MΩ 4.2 mA
199.9 Ω – 100 MΩ 4.2 mA
1.999 kΩ 999.9 Ω 1 Ω 420 µA
19.99 kΩ 9.999 kΩ 10 Ω 42 µA
DP9000
RANGE
RES
EXC
CURRENT
To Order Visit omega.com/dp2000-o for Pricing and Details
MODEL NO. DESCRIPTION
DP2
DP9
* Refer to chart above for input code options.
Accessory
MODEL NO. DESCRIPTION
DPP-5
Comes complete with operator’s manual.
Ordering Example: DP2000-O1, 3½ digit ohmmeter, red LED display, 120 Vac power, 1mV/count, DC current input 0 to 19.99 Ω.
3½-DIGIT FOR ±1999 COUNT
4-DIGIT FOR ±9999 COUNT
0 0 0 -X A. POWER AND DISPLAY
0
1
2
3
4
5
6
7
8
9
A
B
1
DIN panel punch
⁄
8
0
1
2
3
4
5
6
0
1
2
3
4
5
(*)
-O
-G
LED; 120 Vac (50/60 Hz)
LCD; 120 Vac (50/60 Hz) (DP2000 only)
LED; 240 Vac (50/60 Hz)
LCD; 240 Vac (50/60 Hz) (DP2000 only)
LED; 9 to 32 Vdc, isolated
LCD; 9 to 32 Vdc, isolated (DP2000 only)
LED; 5 Vdc
LCD; 5 Vdc (DP2000 only)
LED; 24 Vac
LCD; 24 Vac (DP2000 only)
LED; 26 to 56 Vdc, isolated
LCD; 26 to 56 Vdc, isolated (DP2000 only)
B. ANALOG OUTPUTS
1 mV/count (DP2) or 0.1 mV/count (DP9)
(supplied on all units)
0 to 5 Vdc
0 to 10 Vdc
0 to 1 mA (internally driven)
4 to 20 mA (internally driven)
4 to 20 mA (externally driven)
4 to 20 mA (isolated)
C. CONTROL OUTPUTS
None
Dual setpoint, 10 A relay (SPDT)
Proportional 4 to 20 mA
Proportional/time proportioning, 2 A relay
Parallel BCD, isolated
Single setpoint, 10 A relay (SPDT)
D. SIGNAL CONDITIONER INPUTS
Ohmmeter
ADDITIONAL OPTIONS
Green LED display
D-2
ACC
±0.05%
of
reading
±1
count