HYSTEROFLOW / HYSTEROBALANCE II
HYSTEROSCOPIC FLUID MANAGEMENT SYSTEM
QUICK REFERENCE GUIDE
Option 1: Technical Assistance
Option 2: Customer Service
Option 3: Repair
Option 4: Field Inventory

2 HYSTEROFLOW QUICK REFEREN C E G U I D E
W A R N I N G
This quick reference guide is only a summary
of the HysteroFlow/HysteroBalance II
Instructions for Use. For detailed operating instructions,
be sure to follow the Instruction Manual that was
included with your device when purchased.

HYSTEROFLOW QUICK REFEREN C E G U I D E 3
C O N N E C T I N F L O W T U B I N G
1. Turn the unit on.
2. After self test, press Menu to check
or change system settings. (Refer
to page 4 for menu settings)
3. Hang fluid irrigation bags on the
upper scale hooks.
4. Close inflow and outflow stopcocks
on the instrument.
5. Connect blue Leur Lok on the inflow
tubing to the inflow port on the
instrument.
6. Press OK next to Operative Mode
and press Start.
7. Carefully insert the pressure chamber
on the inflow tubing into the lower
retainer on the pump, and wrap the
tubing up and around the roller wheel,
locking it into place.
8. Clamp tubing and spike fluid irrigation
bag(s). Open one clamp.