Olympus HysteroFlow, HysteroBalance II Quick Reference Manual

HYSTEROFLOW / HYSTEROBALANCE II
HYSTEROSCOPIC FLUID MANAGEMENT SYSTEM
QUICK REFERENCE GUIDE
Option 1: Technical Assistance Option 2: Customer Service
Option 3: Repair Option 4: Field Inventory
2 HYSTEROFLOW QUICK REFEREN C E G U I D E
W A R N I N G
This quick reference guide is only a summary
of the HysteroFlow/HysteroBalance II
Instructions for Use. For detailed operating instructions,
be sure to follow the Instruction Manual that was
included with your device when purchased.
HYSTEROFLOW QUICK REFEREN C E G U I D E 3
C O N N E C T I N F L O W T U B I N G
1. Turn the unit on.
2. After self test, press Menu to check
or change system settings. (Refer to page 4 for menu settings)
3. Hang fluid irrigation bags on the upper scale hooks.
4. Close inflow and outflow stopcocks on the instrument.
5. Connect blue Leur Lok on the inflow tubing to the inflow port on the instrument.
6. Press OK next to Operative Mode and press Start.
7. Carefully insert the pressure chamber on the inflow tubing into the lower retainer on the pump, and wrap the tubing up and around the roller wheel, locking it into place.
8. Clamp tubing and spike fluid irrigation bag(s). Open one clamp.