
II. NOM
ENCLATURE
The
Model
BHT
consists
of
various
components
and
interchangeable
accessories
as
shown
in
the
photo
below. A
wide variety of combinations,
standard
or
optional,
isavailable
according
to
your
req u i
re
men ts.
Eyepiece
Halogen
lamp
holder
Observation
tube
Microscope
stand
Revolving
nosepiece

@
Explanations
in
detail
Mounting
the
stage
1) Loosen the stage
clamping
screw
O
Ov
rotating counterclockwise.
(Fig.
1)
2)
Insert
the stage
into the
mounting
dove-
tail
of
the
microscope
stand slowly and
lock with
clamping screw.
Mounting
the
revolving
nosepiece
1)
L-oosen
the
nosepiece
clamping
screw
O.
(Fig.2)
2) Aligning
the nosepiece dovetail
slide to
the mounting
block @,
push
in
the
nose-
piece
slowly all the way.
NOTE:
Do not
tilt
or rock
the nose-
piece
while inserting
into
the
mountinq
block.
Mounting
the observation
tube
1) Loosen
the clamping
knob
e
tutty.
pull
spring-loaded
clamping
knob
O
. This
will
cause
the locating
pin
@
to
withdraw.
(Fig.
3) lf
the
pin
does
not,
loosen
the
screw
further
until
the
pin
withoraws.
2)
Witfr clamping
knob
e
pulled
out,
insert
the
circular
dovetail
of
the
observation
tube
into
the ring
dovetail.
3) Tighten
the
clamping
knob.
Fis.
1
@
@
Fig.2
Fig.3

c)
l-
o)
-
P
:l
C)
Summary
of
Putting
dre Microscope
into
Operation
Model
BHT
A. Match
the voltage
selector
switch
to local mains voltage
(page
g).
B.
Switch
on
the light
source
(page
9).
C.
Place
a specimen
slide on
the mechanical
stage
(page
9).
D.
Coarse
focus
with
the 10X
objective
(page
10.
13).
E.
Make
interpupillary distance
and
diopter
adjustments
(page
1l
).
F.
Adjust the condenser
position
(page
12).
G.
Swing in the desired
objective.
H.
Adjust light
intensity.
l. Fine focus.
J.
Adjust
aperture
iris diaphragm
and
field
iris diaphragm
(page
12).
Adjustment
of
lllumination
System
for
Various
objective
Powers
*N.A.
is somewhat
low,
but
still compatible
with
a
100X
objective.
(Cut
off
this
page
at
dotted
line and
put
it
on
the wall
near the microscopefor useasa
reminder
of micro-
scopic
procedure.)
Objective
magnif
ication
Achromatic-
aplanatic
condenser
BH2.AAC
Swing-out
condenser
BH2-SC
Low
power
condenser
BH2-UL-C
Compatible
Compatible
Compatible
Swing in
top
lens
oL:yMPus

D.
CondenserAdjustment
1. Condenser
Centration
1)
Stop
down
the
f
ield
iris
diaphragm
with
knurled
ring
O by
rotating
in
the
direc-
tion
of
the
arrow.
(Fig'
14)
2)
Use
the
condenser
height
adjustment
knob
@
to
move
the
condenser
uP
and
down
until
an
image
of
the
field
dia-
phragm
can
be
seen
clearly
in
the
eye-
oieces.
The
rotation
of the
knob
in
the
direction
of
the
arrow
lowers
the
con-
Field
of
view
lf
the
polygonal
image
of
the
iris
the
field
diaphragm
is
centered.
o
The
field iris
diaphragm
controts
the
diameter
of
the
ray
bundle
impining
on
the
speci-
men
surface
and
therefore,
by
stopping
down
the
fierd
diaphragm
until
it
is
slightly
larger
than
the
field
of
view,
it
can
reduce
stray
light,
which
in
turn
increases
image
defi-
nition
and
contrast.
Fig. 14
denser.
Field
iris
diaphragm
ffi+o
Bring
the
field
iris
diaphragm
image
into
denser
centering
knobs
@.
(Fig.
1a)
the
center
of
the
field
of
view
with
the
two
con-
4)Widenthediameteroftheirisdiaphragmprogressive|y'
diaphragmbecomesinscribedinthefie|ditmeansthat
(Fig.
1
5)
Field
lris
DiaPhragm
Aperture
lris
DiaPhragm
o
rn
order
to
achieve
optimum
objective
performance,
the
opening
of
the
aperture
rrs
diaphragm
should
be
matched
to
the
numerical
aperture
of
the
objective
in
use'
lt
is
often
preferable,
however,
to
stop
down
the
aperture
diaphragm
slightly
more
than
in-
dicated
by
the
objective
N.A.
Thiswill
result
in
better
image
contrast,
increased
depth
of
focus
and
a
flatter
field.
After
completing
focus
adjustment,
re-
move
one
of
the
eyepieces
from
the
ob-
servation
tube
and
look
into
the
empty
eyepiece
tube.
As
You
stoP
down
the
aperture
iris
diaphragm,
the
image
of the
iris
diaphragm
can
be
seen
in
the
object-
ive
pupil.
Adjust
the
opening
of the
diaphragm
to
match
the
N'A.
of
the
ob-
jective
in
use.
lf
the
specimen
is
low
in
contrast,
it
is recommended
to
stop
down
lo
70%
-
8O%o
of the
objective
N.R'
(Fig'
16)
Opening
of the
aperture
diaPhragm
Objective
exit
PuPil
12
Fig.
16

.
Focus Adjustment
1. Tension
of Coarse
Ad
justment
Knobs and
Fine
Adjustment.
Although the tension
of
the
coarse adjust-
ment
knobs has
been already adjusted
for
optimum
performance
by
the
manufacturer,
it is
possible
to
personally
adjust the tension
of
the
coarse
adjustment
for either
heavy or
light
movement
depending
on
the
operator's
preference
by
rotating the tension
adjust-
ment rins
e.
(Fig.
17)
The
ring can
be rotated
by
inserting
a screwdriver into
one of the holes
on
the
periphery
of
the
ring. The
clockwise
rotation
(in
the
direction
of
the arrow) tightens
the
coarse adjust-
ment
knobs. Do not loosen
the ring too
much, because the stage
may drop
or
the
fine
adjustment
knobs may
slip.
NOTE: Do
not
rotate the
right
and
left coarse
adjustment
knobs
in
the
opposite
directions
simultaneously.
lf
the stage
drops and the
specimen
goes
out of
focus,
the tension
adjustment
ring
is
too
loose. Tighten the
ring.
Use
of Rubber
Cap
for
Fine
Adjustment
Knob
Attaching
this
cap over the
fine
adjustment
knob
increases
the
sensitivity
of
the
fine
focus-
ing
motion.
(The
rubber
cap
is
optionally
avail-
able.)
2.
Pre-Focusing Lever
This lever
@
is
provided
to
prevent
possible
contact
between specimen
and
objective
as
well as
to
simplify
coarse
focusing.
(Fig.
1B)
The
lever
is
locked after
coarse
focus
has
been accomplished.
This
prevents
further
upward
travel
of
the stage
by means of the
coarse
adjustment
knobs,
and automatically
provides
a limiting stop lf
the stage is
lowered
and then
raised again. The
pre-focusing
lever
does
not restrict f ine
focusing.
3. Adjustment
of
Stage
Block
Height
ln addition
to the
vertical
movement of the
stage
by means
of
coarse
and
fine
adjust-
ments, the stage
block
height can be changed
for
observation
of
specimens'which
are
thicker
than standard slides,
e.g. chambers,
flasks,
etc. with
much
larger
thickness.
The
stage
block
height can be adjusted
by
loosening
the stage
block locking screw
O
'19
Then,
dislocate
the
After
lowering the
with
the Allen
wrench
provided
and
retightening it
at
the
upper
position.
lower limit stop
pin
beneath the stage
block
into
a
lower tapped
hole.
stage
block,
reclamp
the
stage block
locking
screw
@.
(Fig.
19)
t-
t:
t\..
Er r \
I t----l
ll---ri--- | | |
It---: | | t
l.:
l/
Fig.
17
F
is. 18
13

4.
Focusing Procedure
Use the
field of view eyepieces
for focusing
on
the
film
plane.
Each
field of vieweyepiece
has a
focusing
front
lens and
a reticle with 4
frames,
each
frame
indicating the area covered
by a
specif ic
power
N F
K
photo
eyepeice.
(Fig.
22!-.
The number at each frame
indicates the
magnification of
the
photo
eyepiece.
The
image in the
field
of view eyepiece
and
the
image
on
the
film
plane
are
in
focus
at
the
same time. Several
type
field of view eye-
pieces
are available,
according
to
the
film
size
employed.
Field of
view
eyepiece
35WH Kl OX
PWH
Kl
OX 4X5WH K
1OX
MHWH
Kl OX
Attachment camera
35mm Back 3T+" x 4Y+"
Polaroid
Back
4" x5"
Sheet
Film
or
Polaroid
Film
Holder
16
mm
Bolex
camera
120 Roll
Film
Holder
1) Select the
field
of view eyepiece matching
the
camera back
in
use
and insert it into the
right eyepiece
tube
of
the trinocular
tube,
aligning
locating
groove
and
locating
pin.
2)
While
looking through
the
field
of view eyepiece, rotate
the
eyepiece front
lens
in
screw
mount to
focus
on
the
double cross
lines in
the
field.
For
sharp
focusing
with objectives
4X
or
lower,
the
focusing
magnifier
FT
is recommended.
3) Bring the specimen
detail
to
be
photographed
within the
frame
corresponding to the
power
of
the
NFK eyepiece
in use
and
focus
on
the specimen
with
the
microscope fine
adjustment
knobs.
Make
sure the light
path
selector
knob on
the
observation
tube
is
either
on the white
(V)
or
yellowgreen
(CV)
band.
4) lt
is
recommended to
tighten the tension adjustment ring considerably to
prevent
the
stage
from dropping during
long
exposures.
Fig.22
15

VII. TROUBLESHOOTING
lf
you
are
unable to obtain
table
below as
pointers
for
full
performance
from
your
microscope,
please
consult
with
the
troubleshooting.
Phenomenon
Cause
Remedy
1. Optical
System
a)
With illuminator
switched
on, the
field
of view
is
dark.
Field
iris
diaphragm
is
not
opened
sufficiently.
Open diaphragm
to
proper
diameter.
Condenser
is
lowered too
much.
Adjust condenser height.
Light
path
selector lever
is
pulled
out
to
C
position.
Push
in lever up
to CV
or V
position.
b)
Field
of
view
is
cut off or
i
I lum
inated
irregularly.
Light
path
selector
lever
is
stopped midway.
Ctick
it
into
proper position
according to
your purpose.
Nosepiece
is
not
clicked
into
place.
Slightly
rotate
nosepiece until
it clicks
into
place.
Nosepiece
is
not correctly
mounted.
Insert nosepiece dovetail
into
microscope frame
all
the way,
then lock.
The
power
of objective
used
exceeds
the illumination ca-
pacity
of
condenser.
Choose a
condenser
to meet
your purpose.
Condenser
is not centered.
Center condenser.
Field
iris
diaphragm
is
stopped
down excessively.
Open
diaphragm
to
proper
diameter.
c) Dust
or dirt
is
visible
in
the
f
ield
of view.
Dust,
etc.
on
light exit
lens. Remove
dust, etc.
Clean
front
lenses.
Dust on
condenser
top
lens.
Dirty
specimen.
Dust on eyepiece.
d) Excessive
image
contrast.
Condenser is lowered
too
much.
Adjust
condenser height.
Aperture
iris
diaphragm
is
stopped down excessively.
Open diaphragrn
to
proper
diameter.
17

Phenomenon
Cause Remedy
c) Lamp
goes
off
and
on.
Bulb
f
ilament is
likely
to
burn out.
Replace
bulb.
Loose electric
connections. Check
all connections.
d)
Bulb burns out frequently.
Line voltage
selector switch is
not matched
with local mains
voltage.
Match
selector
switch
to
mains voltage.
Bulb
is
not
standard one.
Use standard bulb.
3.
Coarse and
Fine
Adjustments
a)
Coarse adjustment knob
is
too tight.
Tension adjustment ring
is
tightened too much.
Loosen ring
properly.
User
is trying
to
raise
stage
above
the
focusing
limit
im-
posed
by
the
engaged
pre-
focusing
lever.
Unlock
lever.
b)
Stage drops
or
specimen
goes
out of focus
during
observation
due
to
slipping
fine
adjustment knobs.
Tension adjustment
ring
is
too loose.
Tighten ring
properly.
c)
Stage cannot be
raised
to
the upper limit.
Pre-focusing
lever
is
engaged
in lower than focusing
posi-
tion.
Unlock
lever.
d)
Stage
cannot be lowered
to the
lower
limit.
Stage is
mounted
too
low.
Raise
stage
mount
with
Allen
wrench.
e)
Objective
front
lens hits
specimen before coming
into
focus.
Specimen is
placed
on
stage
upside down.
Reverse
specimen.
4. Observation
Tubes
a) Incomplete binocular
vi-
sion.
I
nterpupillary
d
istance is
not
correctly
adjusted.
Correct the interpupillary disĀtance.
Diopter adjustment
is incom-
plete.
Complete
the
diopter
adjust-
ment.
Right
and left eyepieces
are
not matched.
Use a
pair
of matched eye-
pieces.
User
is
unaccustomed
to
bi-
nocular
vision.
Prior
to
looking
into the bi-
nocular observation
tube,
look
at
a
far
away object.
5.
Stage
a)
lmage easily
goes
out of
focus
when
you
touch
the
stage.
Stage
is
not
correctly
locked.
Clamp stage securely.
b)
Specimen stops
midway
on
the
east-west
traverse.
Specimen
is
not correctly
po-
sitioned.
Adjust specimen
position.