Operating Instructions
PCTSTestr™ 50 Pocket Tester
Applications
|
|
• Drinking water |
• Printing industry |
• Agriculture |
|||
|
• Aquaculture |
• Ecology |
• Swimming pools |
|
• Aquariums and |
• Electroplating |
• Verification of |
|
fish farms |
rinse tanks |
reverse osmosis |
|
• Boiler blow- |
• Food sectors |
system operation |
|
|
||
|
down |
• Hydroponics |
• Water and waste- |
|
|
water treatment |
|
|
• Car washes |
|
|
|
• Labs |
|
|
|
|
|
|
|
|
|
Getting Started
The PCTSTestr 50 Pocket Tester has been factory calibrated and usually works well out of the box. However, after extended periods of non-use, it is best to remove the sensor cap and soak the sensor in warm tap water for 10 minutes or so. Prior to taking the measurements, periodic calibration with certified standards is recommended for best accuracy.
Measurement Parameter Setting
1.Press ON/OFF () to power on the tester.
2.Press MENU/v to enter setup window. Press HOLD/ to select Measure. The display shows pH, Cond, TDS and Salinity.
3.Scroll down by pressing MENU/v to toggle between pH, Cond, TDS and Salinity. Press HOLD/ to select pH.
4.The display shows the selected parameter with a .
menu |
Key Info |
|
|
|
|
|
|
|
hold |
|
|
|
hold |
|
||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
pH |
|
|
|
|
|
|
Scroll |
MENU |
|
|
|
Measure |
|
|
|
|
|
|
|
pH |
|||
|
|
|
|
|
|
|
Cond |
|
|
|||||||||
|
Settings |
|
|
|
|
|||||||||||||
|
|
Accept |
HOLD |
|
|
|
|
|
|
|||||||||
|
|
|
|
|
|
|
|
|
||||||||||
|
|
|
Temp Set |
|
|
TDS |
|
|
|
|||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||
|
|
EXIT |
CAL |
|
Reset |
|
|
Salinity |
|
|
|
|||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
pH Buffer Set Selection
PCTSTestr 50 Pocket Tester features USA (pH 4.01, pH 7.00 and pH 10.01) or NIST (pH 4.01, pH 6.86, and pH 9.18) standards. Select either one to suit your requirements.
1.Press MENU/v to enter setup window. Press HOLD/
to select Settings. The display shows Buffer, TDS Factor and Backlight.
1
Temperature Coefficient
1.Pres MENU/v to enter setup window. Scroll down by pressing MENU/v to select Temp Set.
2.Press HOLD/ to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
3.Scroll down by pressing MENU/v to toggle between Set °C/°F, Temp Cal and Temp Coeff.
4.Press HOLD/ to select Temp Coeff or MENU/v to adjust the Temp Coeff.
5.Press HOLD/ to confirm the Temp Coeff value. The new value is automatically confirmed with a .
menu |
|
Key Info |
|
|
|
|
|
|
hold |
|
|
|
hold |
|
hold |
|
|||
|
|
|
|
|
|
|
Measure |
|
|
Set °C/°F |
|
|
|
|
|
|
|||
|
|
Scroll |
MENU |
|
|
|
|
|
|
Adjust |
|
|
2.00 |
||||||
|
|
|
Settings |
|
|
|
|
|
|
||||||||||
|
|
Accept |
HOLD |
|
|
|
|
Temp Cal |
|
|
Temp Coeff |
|
|
|
|||||
|
|
|
|
|
|
|
|
|
|
|
|||||||||
|
|
|
|
|
Temp Set |
|
|
|
|
|
|
|
|||||||
|
|
|
|
|
|
|
Temp Coeff |
|
|
|
2.00 |
|
|
|
|||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
|
EXIT |
CAL |
|
|
Reset |
|
|
|
|
|
|
|
|
|
||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
pH Calibration
Calibration should be done regularly, recommended once a week. Calibrate up to three points using either the USA or the NIST buffer set standards.
1.Press ON/OFF () to power on the tester if needed.
2.Dip electrode about 2 cm to 3 cm into the pH standard buffer solution.
3.Stir gently and press CAL/ESC to enter calibration mode. The CAL indicator will be displayed. The upper display will show the measured reading based on the last calibration while the lower display will indicate the pH standard buffer solution.
Note: To abort calibration, press CAL/ESC to escape.
4.Allow about 2 minutes for the tester reading to stabilize. The timer icon blinks during this time. Once the reading is stabilized, the timer stops blinking. Automatic confirmation happens when the buffer is found and
the display returned to measurement window with reading calibrated to pH standard buffer solution.
5.Repeat with other buffers if necessary. Rinse electrode before dipping into next buffer.
Note: The calibration mode allows you to perform up to three calibration points. Calibration is automatically confirmed with the buffer identification. No user interaction is required after starting the calibration by pressing CAL/ESC.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
cal |
|
|
CAL |
|
|
|
|
|
CAL |
|
|
|
|
pH |
|
esc |
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
|
Place the tester |
|
|
pH |
|
|
|
|
|
|||||
|
|
|
|
|
|
|
|
pH |
|||||||
|
|
in the buffer to |
|
|
|
7.45 |
|
|
|
7.01 |
|
||||
|
|
|
|
|
|
||||||||||
|
|
start or press |
cal |
|
7.01 |
|
|
|
|
25.5°C |
|||||
|
|
esc |
|
|
|
|
|
||||||||
|
|
to escape. |
|
|
|
|
|
|
|
|
|
|
|
4
1065O100_MAN_35634-35.indd 1
2.Press HOLD/ to select Buffer. Display shows USA and NIST.
3.Press HOLD/ to select USA or scroll down by pressing MENU/v to toggle between the two buffer standards.
4.The display shows the selected buffer standard with a .
menu |
Key Info |
|
|
|
|
|
|
|
hold |
|
|
|
hold |
|
|
hold |
|
||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Scroll |
MENU |
|
|
Measure |
|
|
|
Buffer |
|
|
|
|
|
|
|
|
|
||||
|
|
|
|
|
|
|
USA |
|
|
|
USA |
||||||||||||
|
|
|
|
|
|
TDS Factor |
|
|
|
|
|
|
|||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||||||||
|
|
Accept |
HOLD |
|
|
|
Settings |
|
|
|
|
|
NIST |
|
|
|
|
||||||
|
|
|
Temp Set |
|
|
Backlight |
|
|
|
|
|
|
|||||||||||
|
|
EXIT |
CAL |
|
|
|
|
|
|
|
|
|
|
||||||||||
|
|
|
Reset |
|
|
|
|
|
|
|
|
|
|
|
|
|
|||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
TDS Factor Setting
1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Settings.
2.Press HOLD/ to select Settings. The display shows Buffer, TDS Factor and Backlight.
3.Scroll down by pressing MENU/v to toggle between the Buffer, TDS Factor and Backlight. Press HOLD/ to select the TDS Factor.
4.Press HOLD/ to select the default TDS factory setting or MENU/v to adjust the setting.
5.Press HOLD/ to confirm the selection of the setting. The display shows the selected value (TDS factor) with a .
menu Key Info |
hold |
hold |
hold |
Scroll |
MENU |
|
|
Measure |
|
Buffer |
|
Adjust |
|
0.66 |
|||
Accept |
HOLD |
|
|
Settings |
|
|
|
TDS Factor |
|
|
TDS Factor |
|
|
|
|
|
|
|
|
|
|||||||
|
|
|
|
|
|
|
|
|
|||||
|
|
Temp Set |
|
|
|
|
|
|
|||||
|
|
|
|
|
|
||||||||
|
|
|
Backlight |
0.71 |
|
|
|||||||
EXIT |
CAL |
|
|
|
|
|
|||||||
|
|
Reset |
|
|
|
|
|
|
|
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
Backlight Settings
1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Settings.
2.Press HOLD/ to select Settings. The display shows Buffer, TDS Factor and Backlight.
3.Scroll down by pressing MENU/v to toggle between Buffer, TDS Factor and Backlight. Press HOLD/ to select Backlight.
4.The display shows ON and OFF. Scroll down by pressing MENU/v to toggle between ON and OFF. Backlight ON increases readability in low-light conditions.
5.Press HOLD/ to select the desired backlight option. The display shows the selected backlight option with a .
2
Calibration for Conductivity, TDS, or Salinity
For best results, periodic calibration with an accurate standard is recommended prior to measurement. Use the calibration standard value that is close to your intended sample value. The tester will retain one calibration value in each mode (conductivity, TDS, salinity) when the instrument is powered off. The conductivity value can be calibrated automatically or manually, while the TDS & salinity values require manual calibration. The tester will begin in the measurement mode that was used when it was powered off. See “Measurement Parameter Setting” to change the desired parameter.
Automatic Calibration for Conductivity
1.Remove the cap and press ON/OFF () to power on.
2.Dip the sensor in at least 30 mm of calibration standard.
3.Stir gently and press CAL/ESC to begin the calibration.
4.The display will show CAL followed by the default value. CAL is indicated on the display during calibration mode.
5.If the reading is within the calibration range of the automatically recognized standards; 80 (84 μS/cm), 1410 (1413 μS/cm),
or 12.90 (12.88 mS/cm), the icon is displayed when the automatic calibration standard value has been detected.
6.Press HOLD/ to accept the auto conductivity standard and finish the calibration.
7.Display returns to Measurement window.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
cal |
|
|
CAL |
|
|
|
|
|
CAL |
|
|
|
|
COND |
|
esc |
|
|
|
|
|
|
|
|
|
|
|||||
|
|
|
Place the tester in |
|
|
µS/cm |
|
|
|
|
|
||||
|
|
|
|
|
|
|
|
|
|
µS/cm |
|||||
|
|
|
the calibration STD |
|
|
|
844 |
|
|
|
844 |
|
|||
|
|
|
|
|
|||||||||||
|
|
|
to start or press |
|
1413 |
|
|
|
|
23.5°C |
|||||
|
|
|
cal |
|
|
|
|
|
|
|
|
|
|||
|
|
|
esc to escape. |
|
|
|
|
|
|
|
|
|
Manual Calibration
When the conductivity reading is outside calibration range of the automatic conductivity standards or when TDS or salinity is used, the tester will require manual adjustment.
1.Repeat steps 1 to 4 from “Automatic Calibration for Conductivity”.
2.Press MENU/v to manually adjust the value to the desired reading.
Note: The adjustment will decrease only, however the adjustment will eventually cycle to the highest available value after decreasing by 40% of the initial value.
3.Press HOLD/ to accept and finish the calibration when the desired value is selected.
Note: To abort calibration, press CAL/ESC to escape.
5
menu Key Info |
hold |
hold |
hold |
|
|
|
|
|
Measure |
Buffer |
|
|
|
Backlight |
|
|
|
|||
Scroll |
MENU |
|
|
|
|
|
|
|
OFF |
|||||||
|
|
Settings |
|
|
|
TDS Factor |
|
|
|
ON |
|
|
||||
Accept |
HOLD |
|
|
|
|
|
|
|
|
|
|
|
||||
|
|
Temp Set |
|
|
|
|
|
|
||||||||
|
|
|
|
|
|
|
|
|
|
|||||||
|
|
|
|
|
|
|
|
|
OFF |
|
|
|
||||
EXIT |
CAL |
Backlight |
|
|
|
|
|
|
||||||||
|
|
Reset |
|
|
|
|
|
|
||||||||
|
|
|
|
|
|
|
|
|
|
|||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
Temperature Settings
1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Temp Set. Press HOLD/
to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
2.Press HOLD/ to select Set °C/°F. Scroll down by pressing MENU/v to toggle between °C and °F.
3.Press HOLD/ to select temperature unit. The display shows the selected temperature setting with a .
menu Key Info |
hold |
hold |
hold |
|
|
|
|
Measure |
|
|
|
|
|
|
|
|
|
|
|
Scroll |
MENU |
|
|
|
|
Set °C/°F |
|
|
|
°C |
|
|
°C |
||
|
|
Settings |
|
|
|
|
|
|
|
||||||
Accept |
HOLD |
|
|
|
Temp Cal |
|
|
°F |
|
|
|||||
|
|
|
|
|
|
|
|
|
|||||||
|
|
Temp Set |
|
|
|
|
|
|
|||||||
EXIT |
CAL |
|
|
|
|
Temp Coeff |
|
|
|
|
|||||
|
|
Reset |
|
|
|
|
|
|
|
||||||
|
|
|
|
|
|
|
|
|
|
|
Temperature Calibration
1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Temp Set.
2.Press HOLD/ to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
3.Scroll down by pressing MENU/v to toggle between Set °C/°F, Temp Cal and Temp Coeff. Press HOLD/ to select Temp Cal.
4.The lower display shows the current measured temperature reading based on the last set offset and the upper display shows the current measured temperature reading based on factory default calibration.
5.Dip the tester into a solution of known temperature and allow time for the built-in temperature sensor to stabilize.
6.Press MENU/v to adjust the temperature value or press the HOLD/ to confirm the calibrated value as the new temperature value of the solution.
Note: To exit this program without confirming the calibration, press CAL/ESC.
menu |
Key Info |
|
|
|
|
|
|
|
hold |
|
|
|
hold |
|
hold |
|
|||||
|
|
|
|
|
|
|
|
|
Measure |
|
|
Set °C/°F |
|
|
|
|
|
|
|||
|
|
Scroll |
MENU |
|
|
|
|
|
|
|
Temp Cal |
|
|
2.00 |
|||||||
|
|
|
|
Settings |
|
|
|
|
|
|
|||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
|
|
Accept |
HOLD |
|
|
|
|
|
Temp Cal |
|
|
|
23 4 °C |
|
|
|
|||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||||||
|
|
Temp Set |
|
|
|
|
|
|
|
|
|
|
|||||||||
|
|
EXIT |
CAL |
|
|
|
|
|
|
Temp Coeff |
|
|
23 4 °C |
|
|
|
|||||
|
|
|
|
Reset |
|
|
|
|
|
|
|
||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
3
4.Once the calibration is finished and user has accepted the changes, measurement window will now show the calibrated reading.
Note: The auto conductivity standards are 84 μS/cm, 1413 μS/cm & 12.88 mS/cm.
cal |
CAL |
CAL |
CAL |
COND |
esc |
||||
|
Place the tester in |
|
|
|
|
the calibration STD |
µS/cm |
µS/cm |
µS/cm |
|
to start or press |
844 |
844 |
1413 |
|
cal |
1413 |
820 |
23.5°C |
|
esc to escape. |
Measurement
1.Press ON/OFF () to power on the tester if needed.
2.Dip the electrode in about 2 cm to 3 cm into the test solution. Stir and let the reading stabilize. The timer icon will blink during this time. Once the reading is stabilized,
the timer stops blinking and will appear to indicate the stability of the reading.
CAUTION: Testing dry samples is not accurate |
|
|
and can lead to sensor damage or |
pH |
|
|
pH |
|
breakage. Soils must be wet and |
4.01 |
|
free of particulates that may scratch |
|
25.5°C |
the glass sensor. Excessive force |
|
|
|
|
|
into dry samples can cause glass |
|
|
breakage. |
|
|
3.Note the value or press HOLD/ to freeze the reading. To release the reading, press HOLD/ again.
4.Press ON/OFF () for 3 seconds to turn off tester. If you do not press a button for 8.5 minutes, the tester will automatically shut off to conserve batteries.
User Reset
Reset to the user’s default settings by using the User Reset function. Buffer selection and user temperature calibration are not affected by the user reset function.
1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Reset. Press HOLD/ to select Reset. The display shows User Reset and Factory Reset.
2.Press HOLD/ to select User Reset.
3.The display automatically shows No and Yes. Scroll down by pressing MENU/v to toggle between No and Yes.
4.Press HOLD/ to confirm either No or Yes. The display shows the User Reset option with a .
menu |
Key Info |
|
|
|
|
|
|
|
hold |
|
|
|
hold |
|
|
|
|
hold |
|||
|
|
|
|
|
|
|
|
Measure |
|
|
|
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
|
Scroll |
MENU |
|
|
User Reset |
|
|
|
|
No |
|
|
||||||||
|
|
|
Settings |
|
|
|
|
|
|
|
|
|
|||||||||
|
|
Accept |
HOLD |
|
|
|
|
|
|
|
|
|
|
|
|
|
|||||
|
|
|
|
|
|
|
Factory |
|
|
Yes |
|
||||||||||
|
|
|
Temp Set |
|
|
|
|
|
|
||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||||
|
|
EXIT |
CAL |
|
|
|
|
|
Reset |
|
|
|
|
|
|
||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||||||
|
|
Reset |
|
|
|
|
|
|
|
|
|
|
|||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
6
9/5/2017 3:09:46 PM