Oakton PCTSTestr 50 Instruction Manual

4.5 (2)
Oakton PCTSTestr 50 Instruction Manual

Operating Instructions

PCTSTestr50 Pocket Tester

Applications

 

 

• Drinking water

• Printing industry

• Agriculture

 

• Aquaculture

• Ecology

• Swimming pools

 

• Aquariums and

• Electroplating

• Verification of

 

fish farms

rinse tanks

reverse osmosis

 

• Boiler blow-

• Food sectors

system operation

 

 

 

down

• Hydroponics

• Water and waste-

 

 

water treatment

 

• Car washes

 

 

• Labs

 

 

 

 

 

 

 

Getting Started

The PCTSTestr 50 Pocket Tester has been factory calibrated and usually works well out of the box. However, after extended periods of non-use, it is best to remove the sensor cap and soak the sensor in warm tap water for 10 minutes or so. Prior to taking the measurements, periodic calibration with certified standards is recommended for best accuracy.

Measurement Parameter Setting

1.Press ON/OFF () to power on the tester.

2.Press MENU/v to enter setup window. Press HOLD/to select Measure. The display shows pH, Cond, TDS and Salinity.

3.Scroll down by pressing MENU/v to toggle between pH, Cond, TDS and Salinity. Press HOLD/to select pH.

4.The display shows the selected parameter with a .

menu

Key Info

 

 

 

 

 

 

 

hold

 

 

 

hold

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pH

 

 

 

 

 

Scroll

MENU

 

 

 

Measure

 

 

 

 

 

 

 

pH

 

 

 

 

 

 

 

Cond

 

 

 

Settings

 

 

 

 

 

 

Accept

HOLD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temp Set

 

 

TDS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EXIT

CAL

 

Reset

 

 

Salinity

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pH Buffer Set Selection

PCTSTestr 50 Pocket Tester features USA (pH 4.01, pH 7.00 and pH 10.01) or NIST (pH 4.01, pH 6.86, and pH 9.18) standards. Select either one to suit your requirements.

1.Press MENU/v to enter setup window. Press HOLD/

to select Settings. The display shows Buffer, TDS Factor and Backlight.

1

Temperature Coefficient

1.Pres MENU/v to enter setup window. Scroll down by pressing MENU/v to select Temp Set.

2.Press HOLD/to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.

3.Scroll down by pressing MENU/v to toggle between Set °C/°F, Temp Cal and Temp Coeff.

4.Press HOLD/to select Temp Coeff or MENU/v to adjust the Temp Coeff.

5.Press HOLD/to confirm the Temp Coeff value. The new value is automatically confirmed with a .

menu

 

Key Info

 

 

 

 

 

 

hold

 

 

 

hold

 

hold

 

 

 

 

 

 

 

 

Measure

 

 

Set °C/°F

 

 

 

 

 

 

 

 

Scroll

MENU

 

 

 

 

 

 

Adjust

 

 

2.00

 

 

 

Settings

 

 

 

 

 

 

 

 

Accept

HOLD

 

 

 

 

Temp Cal

 

 

Temp Coeff

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temp Set

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temp Coeff

 

 

 

2.00

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EXIT

CAL

 

 

Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pH Calibration

Calibration should be done regularly, recommended once a week. Calibrate up to three points using either the USA or the NIST buffer set standards.

1.Press ON/OFF () to power on the tester if needed.

2.Dip electrode about 2 cm to 3 cm into the pH standard buffer solution.

3.Stir gently and press CAL/ESC to enter calibration mode. The CAL indicator will be displayed. The upper display will show the measured reading based on the last calibration while the lower display will indicate the pH standard buffer solution.

Note: To abort calibration, press CAL/ESC to escape.

4.Allow about 2 minutes for the tester reading to stabilize. The timer icon blinks during this time. Once the reading is stabilized, the timer stops blinking. Automatic confirmation happens when the buffer is found and

the display returned to measurement window with reading calibrated to pH standard buffer solution.

5.Repeat with other buffers if necessary. Rinse electrode before dipping into next buffer.

Note: The calibration mode allows you to perform up to three calibration points. Calibration is automatically confirmed with the buffer identification. No user interaction is required after starting the calibration by pressing CAL/ESC.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

cal

 

 

CAL

 

 

 

 

 

CAL

 

 

 

 

pH

 

esc

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Place the tester

 

 

pH

 

 

 

 

 

 

 

 

 

 

 

 

 

pH

 

 

in the buffer to

 

 

 

7.45

 

 

 

7.01

 

 

 

 

 

 

 

 

 

start or press

cal

 

7.01

 

 

 

 

25.5°C

 

 

esc

 

 

 

 

 

 

 

to escape.

 

 

 

 

 

 

 

 

 

 

 

4

1065O100_MAN_35634-35.indd 1

2.Press HOLD/to select Buffer. Display shows USA and NIST.

3.Press HOLD/to select USA or scroll down by pressing MENU/v to toggle between the two buffer standards.

4.The display shows the selected buffer standard with a .

menu

Key Info

 

 

 

 

 

 

 

hold

 

 

 

hold

 

 

hold

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Scroll

MENU

 

 

Measure

 

 

 

Buffer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

USA

 

 

 

USA

 

 

 

 

 

 

TDS Factor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accept

HOLD

 

 

 

Settings

 

 

 

 

 

NIST

 

 

 

 

 

 

 

Temp Set

 

 

Backlight

 

 

 

 

 

 

 

 

EXIT

CAL

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TDS Factor Setting

1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Settings.

2.Press HOLD/to select Settings. The display shows Buffer, TDS Factor and Backlight.

3.Scroll down by pressing MENU/v to toggle between the Buffer, TDS Factor and Backlight. Press HOLD/to select the TDS Factor.

4.Press HOLD/to select the default TDS factory setting or MENU/v to adjust the setting.

5.Press HOLD/to confirm the selection of the setting. The display shows the selected value (TDS factor) with a .

menu Key Info

hold

hold

hold

Scroll

MENU

 

 

Measure

 

Buffer

 

Adjust

 

0.66

Accept

HOLD

 

 

Settings

 

 

 

TDS Factor

 

 

TDS Factor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temp Set

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Backlight

0.71

 

 

EXIT

CAL

 

 

 

 

 

 

 

Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Backlight Settings

1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Settings.

2.Press HOLD/to select Settings. The display shows Buffer, TDS Factor and Backlight.

3.Scroll down by pressing MENU/v to toggle between Buffer, TDS Factor and Backlight. Press HOLD/to select Backlight.

4.The display shows ON and OFF. Scroll down by pressing MENU/v to toggle between ON and OFF. Backlight ON increases readability in low-light conditions.

5.Press HOLD/to select the desired backlight option. The display shows the selected backlight option with a .

2

Calibration for Conductivity, TDS, or Salinity

For best results, periodic calibration with an accurate standard is recommended prior to measurement. Use the calibration standard value that is close to your intended sample value. The tester will retain one calibration value in each mode (conductivity, TDS, salinity) when the instrument is powered off. The conductivity value can be calibrated automatically or manually, while the TDS & salinity values require manual calibration. The tester will begin in the measurement mode that was used when it was powered off. See “Measurement Parameter Setting” to change the desired parameter.

Automatic Calibration for Conductivity

1.Remove the cap and press ON/OFF () to power on.

2.Dip the sensor in at least 30 mm of calibration standard.

3.Stir gently and press CAL/ESC to begin the calibration.

4.The display will show CAL followed by the default value. CAL is indicated on the display during calibration mode.

5.If the reading is within the calibration range of the automatically recognized standards; 80 (84 μS/cm), 1410 (1413 μS/cm),

or 12.90 (12.88 mS/cm), the icon is displayed when the automatic calibration standard value has been detected.

6.Press HOLD/to accept the auto conductivity standard and finish the calibration.

7.Display returns to Measurement window.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

cal

 

 

CAL

 

 

 

 

 

CAL

 

 

 

 

COND

 

esc

 

 

 

 

 

 

 

 

 

 

 

 

 

Place the tester in

 

 

µS/cm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

µS/cm

 

 

 

the calibration STD

 

 

 

844

 

 

 

844

 

 

 

 

 

 

 

 

 

to start or press

 

1413

 

 

 

 

23.5°C

 

 

 

cal

 

 

 

 

 

 

 

 

 

 

 

 

esc to escape.

 

 

 

 

 

 

 

 

 

Manual Calibration

When the conductivity reading is outside calibration range of the automatic conductivity standards or when TDS or salinity is used, the tester will require manual adjustment.

1.Repeat steps 1 to 4 from “Automatic Calibration for Conductivity”.

2.Press MENU/v to manually adjust the value to the desired reading.

Note: The adjustment will decrease only, however the adjustment will eventually cycle to the highest available value after decreasing by 40% of the initial value.

3.Press HOLD/to accept and finish the calibration when the desired value is selected.

Note: To abort calibration, press CAL/ESC to escape.

5

menu Key Info

hold

hold

hold

 

 

 

 

 

Measure

Buffer

 

 

 

Backlight

 

 

 

Scroll

MENU

 

 

 

 

 

 

 

OFF

 

 

Settings

 

 

 

TDS Factor

 

 

 

ON

 

 

Accept

HOLD

 

 

 

 

 

 

 

 

 

 

 

 

 

Temp Set

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OFF

 

 

 

EXIT

CAL

Backlight

 

 

 

 

 

 

 

 

Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temperature Settings

1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Temp Set. Press HOLD/

to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.

2.Press HOLD/to select Set °C/°F. Scroll down by pressing MENU/v to toggle between °C and °F.

3.Press HOLD/to select temperature unit. The display shows the selected temperature setting with a .

menu Key Info

hold

hold

hold

 

 

 

 

Measure

 

 

 

 

 

 

 

 

 

 

Scroll

MENU

 

 

 

 

Set °C/°F

 

 

 

°C

 

 

°C

 

 

Settings

 

 

 

 

 

 

 

Accept

HOLD

 

 

 

Temp Cal

 

 

°F

 

 

 

 

 

 

 

 

 

 

 

 

 

Temp Set

 

 

 

 

 

 

EXIT

CAL

 

 

 

 

Temp Coeff

 

 

 

 

 

 

Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temperature Calibration

1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Temp Set.

2.Press HOLD/to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.

3.Scroll down by pressing MENU/v to toggle between Set °C/°F, Temp Cal and Temp Coeff. Press HOLD/to select Temp Cal.

4.The lower display shows the current measured temperature reading based on the last set offset and the upper display shows the current measured temperature reading based on factory default calibration.

5.Dip the tester into a solution of known temperature and allow time for the built-in temperature sensor to stabilize.

6.Press MENU/v to adjust the temperature value or press the HOLD/to confirm the calibrated value as the new temperature value of the solution.

Note: To exit this program without confirming the calibration, press CAL/ESC.

menu

Key Info

 

 

 

 

 

 

 

hold

 

 

 

hold

 

hold

 

 

 

 

 

 

 

 

 

 

Measure

 

 

Set °C/°F

 

 

 

 

 

 

 

 

Scroll

MENU

 

 

 

 

 

 

 

Temp Cal

 

 

2.00

 

 

 

 

Settings

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accept

HOLD

 

 

 

 

 

Temp Cal

 

 

 

23 4 °C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temp Set

 

 

 

 

 

 

 

 

 

 

 

 

EXIT

CAL

 

 

 

 

 

 

Temp Coeff

 

 

23 4 °C

 

 

 

 

 

 

 

Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

4.Once the calibration is finished and user has accepted the changes, measurement window will now show the calibrated reading.

Note: The auto conductivity standards are 84 μS/cm, 1413 μS/cm & 12.88 mS/cm.

cal

CAL

CAL

CAL

COND

esc

 

Place the tester in

 

 

 

 

the calibration STD

µS/cm

µS/cm

µS/cm

 

to start or press

844

844

1413

 

cal

1413

820

23.5°C

 

esc to escape.

Measurement

1.Press ON/OFF () to power on the tester if needed.

2.Dip the electrode in about 2 cm to 3 cm into the test solution. Stir and let the reading stabilize. The timer icon will blink during this time. Once the reading is stabilized,

the timer stops blinking and will appear to indicate the stability of the reading.

CAUTION: Testing dry samples is not accurate

 

 

and can lead to sensor damage or

pH

 

 

pH

breakage. Soils must be wet and

4.01

free of particulates that may scratch

 

25.5°C

the glass sensor. Excessive force

 

 

 

 

into dry samples can cause glass

 

 

breakage.

 

 

3.Note the value or press HOLD/to freeze the reading. To release the reading, press HOLD/again.

4.Press ON/OFF () for 3 seconds to turn off tester. If you do not press a button for 8.5 minutes, the tester will automatically shut off to conserve batteries.

User Reset

Reset to the user’s default settings by using the User Reset function. Buffer selection and user temperature calibration are not affected by the user reset function.

1.Press MENU/v to enter setup window. Scroll down by pressing MENU/v to select Reset. Press HOLD/to select Reset. The display shows User Reset and Factory Reset.

2.Press HOLD/to select User Reset.

3.The display automatically shows No and Yes. Scroll down by pressing MENU/v to toggle between No and Yes.

4.Press HOLD/to confirm either No or Yes. The display shows the User Reset option with a .

menu

Key Info

 

 

 

 

 

 

 

hold

 

 

 

hold

 

 

 

 

hold

 

 

 

 

 

 

 

 

Measure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Scroll

MENU

 

 

User Reset

 

 

 

 

No

 

 

 

 

 

Settings

 

 

 

 

 

 

 

 

 

 

 

Accept

HOLD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Factory

 

 

Yes

 

 

 

 

Temp Set

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EXIT

CAL

 

 

 

 

 

Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6

9/5/2017 3:09:46 PM

Loading...
+ 1 hidden pages