Oakton EcoTestr Salt1 Instruction Manual

Operating Manual
Testr™ Salt1 Pocket Tester
Eco
Keypad Functions
Short press = <2 seconds
Long press = >2 seconds
1. Short press to turn on; long press to turn off.
MEAS
CAL
<
2. When turned off, long press to enter setup mode.
3. In mode setting, short press to change parameter.
1. When turned on, long press to enter calibration mode.
2. In calibration mode, short press to confirm calibration.
3. In mode setting, short press to confirm parameter selection.
Calibration
1. Rinse the probe in distilled water and gently dry with a clean tissue. Short press
2. Long press CAL/ to enter calibration mode; short press
3. Dip the probe into calibration solution (SKU 00653-89). Stir gently, leave it to stand. Wait for the measurement stability icon ( then short press CAL/ to complete the 1-point calibra­tion. Tester returns to measurement mode, and calibra­tion icon “H” appears on bottom left side of display.
/MEAS to exit.
) to appear and stay on the display;
/MEAS to turn on the tester.
Measurement
1. Short press /MEAS to turn on the tester. Rinse probe in distilled water and dry it gently with a clean tissue to remove excess water.
2. Stir the probe in the sample solution gently, leave it to stand. Wait for the stability icon () to remain on, then take a reading.
3. Rinse off the tester thoroughly in distilled water after each test.
Notes:
• This salinity tester measures practical salinity converted
from electrical conductivity (NaCl concentration). It does not measure absolute salinity.
• The unit has already been calibrated after manufacture.
User can use the tester immediately or can test it in the calibration solutions to test its accuracy. If error is large, calibrate the tester before using.
• We recommend replacing the calibration solution after
5 to 10 calibrations to maintain accuracy.
Setting the Parameters
When tester is off, long press /MEAS to enter setup mode. Short press press CAL/ and parameter will flash, then short press
/MEAS to choose desired parameter. Short press CAL/ to confirm parameter selection. Long press to return to measurement mode.
Symbol Menu setting Selection
P1 Select salinity unit ppt – mg/L ppt
P2
P3
Notes:
• Select “Yes” for P3 to restore the calibration to the
theoretical values and parameter setting to original values. When tester’s calibration or measurement performs abnormally, this function can be adopted so the tester returns to factory default setting and then users can conduct calibration or take measurements again.
/MEAS to switch from P1 to P2 to P3. Short
/MEAS
Factory
default
Select temperature
unit
Restore to factory
default
°F – °C °C
No – Yes No
1065O101_MAN_Salt1_35634-12.indd 1 03/22/2018 3:55:21 PM
Symbol
Er 1
Self-diagnosis information
Wrong calibration solution, or measured value is not within the range of the tester.
How to fix
• Check if calibration
solution is correct.
• Check if probe is
damaged.
Wait for the measurement stability icon () to appear and stay,
Er 2
CAL/ is pressed before measurement is stable ( appears).
then press CAL/ .
Specifications
Range 0 to 10.00 ppt (mg/L) Resolution 0.01 ppt Accuracy ±1% full-scale
Salinity*
Calibration points
Automatic temperature compensation
Range 32 to 122°F (0 to 50°C)
Temperature
Resolution 0.1°F/°C Accuracy ±0.9°F (0.5°C)
* Tester measures practical salinity converted from electrical conductivity
(NaCl concentration). It does not measure absolute salinity.
Power: four AAA batteries (included); >400 hours of continuous operation
Low-voltage warning:
battery status icon flashes
Auto power-off: tester automatically turns off after 8 minutes of nonuse
IP rating: IP67 (waterproof), floats on water when sensor cap is on
Dimensions (L x W x H): 7" x 1.5" x 1.25" (17.8 x 4 x 3.1 cm) Weight: 3.8 oz (107 g)
1 point; auto standard
recognition
32 to 122°F
(0 to 50°C)
Ordering InformationSelf-Diagnostic Messages
Model Product description Catalog number
Salt1
EcoTestr pocket
salinity tester
35634-12
Battery Installation
The tester uses four AAA batteries. Please install batteries according to the following steps. Note the correct direction of battery installation: the positive side (+) of every single battery must face up. Incorrect installation of batteries will cause damage to the tester and create a potential hazard.
4 x AAA Batteries
2
1
3
+
+
_
_
+
+
_
_
1. Pull the battery cap up.
2. Slide the battery cap along the direction of arrow.
3. Open the battery cap.
4. Insert the batteries (ALL POSITIVE SIDES FACING UP).
5. Close the battery cap.
6. Slide and lock the battery cap along the direction of arrow.
7. Fit the tester’s cap while making sure to push all the way down. The tester’s waterproof design may be compromised if the cap is not fitted correctly.
5
7
4
6
Warranty
We warrant this instrument to be free from defects in material and workmanship and agrees to repair or replace free of charge, at option of Oakton Instruments, any malfunctioned or damaged product attributable to responsibility of Oakton Instruments, for a period of two years from the delivery (a six-month limited warranty applies to probes). This warranty does not apply to defects resulting from actions such as misuse (violation of the instructions in this manual or operations in the manner not specified in this manual), improper maintenance, and unauthorized repairs. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or probe.
Oakton Instruments reserves the right to update the information in this manual without giving notice in advance.
www.4oakton.com
1065O101_MAN_Salt1_35634-12 March 2018
1065O101_MAN_Salt1_35634-12.indd 2 03/22/2018 3:55:23 PM
Loading...