SenzTx
Electrochemical Oxygen Analyzer
WITH 4-20mA ANALOGUE OUTPUT
Short Form User Manual
Table of Contents
1.1. General Information ........................................................................................................................ 3
1.1.2 Intended Use of the SenzTx ................................................................................................. 3
1.1.3 The Measuring Cell/Sensor Element ........................................................................ 3
1.1.4 Unintended Use of the SenzTx Analyzer ........................................................................... 4
1.1.5 Construction ............................................................................................................................ 4
1.1.6 Calibration of the Measuring Instruments ........................................................................... 4
1.2. Technical Description ..................................................................................................................... 5
1.2.1 Connection .............................................................................................................................. 5
1.2.2 Technical Data ........................................................................................................................ 5
1.2.3 Installation................................................................................................................................ 6
1.3. Maintenance and Calibration of Measuring Cells .................................................................... 8
30/07/2018 Rev.1.2 Page 3 /10
1.1. General Information
1.1.2 Intended Use of the SenzTx
The Analyzer are designed to monitor a sample containing an oxygen content to be measured within the
specified ppm or % ranges.
The Analyzer can be used in combination with the control system (PLC) or as a standalone oxygenÂmeasuring-system.
1.1.3 The Measuring Cell/Sensor Element
This Analyzer uses an electrochemical measuring cell, which is specific to oxygen. There are nearly no
cross sensitivities for other substances. But certain substances may damage the measuring cell or
shorten its lifetime. The measuring cell is like a battery and is usually has an operating life of between
one and two years, depending on Oxygen Hours exposure and or degree of exposure to aggressive
gases/gas mixtures. Furthermore, the measuring cell may dry out in the extremely dry atmospheres
necessitating early sensor replacement.
Note!
Operating the ppm sensor at oxygen levels of >1000 ppm (e.g. in air) does not
damage the sensor element irreversibly, but it should be avoided. If exposed
to air, it will take several hours until the sensor will measure low oxygen levels
correctly in inert gas. Longer exposure to high oxygen concentration will
shorten the lifetime of the sensor.
Risk of damage!
The SenzTx should not be stored in a way that environmental air may
reach the Sensor. This will considerably shorten the lifetime of the
cell and should be avoided.
The package should not be unwrapped until the SenzTx is ready to be
put into service.