Nipro Surdial Trouble shooting manual

4.4 (56)

TROUBLE SHOOTING MANUAL

SINGLE PATIENT DIALYSIS MACHINE

Manufacturer NIPRO CORPORATION

3-9-3, Honjo-Nishi, Kita-Ku, OSAKA 531-8510, JAPAN

PHONE +81-6-6372-2331 / FAX +81-6-6371-7422

Version No.

1.GENERAL DESCRIPTION

When TROUBLE is detected,

check symptom and condition and perform

TROUBLE SHOOTING

to check them for finding trouble cause to determine action to be taken.

Condition, check method, cause and action to be taken for each trouble are described in this Manual.

Perform check after reading Cautions for Safety thoroughly contained in Service Manual.

When power ON of SURDIAL is required to perform check, take care of electric shock.

In accordance with procedure specified in

SERVICE MANUAL,

take action and check the result.

Construction drawings and adjusting procedures of each Assy and component, list of components and the like are contained in Service Manual.

2.ALARM/ERROR INDICATIONS AND DESCRIPTION

NO.

ALARM/ERROR INDICATION

NO.

ALARM/ERROR INDICATION

 

 

 

 

 

 

 

 

E01

RECOVER FROM EMERGENCY STOP

E41

VCa CLOSED CIRCUIT LEAK ERROR

E02

T1 TEMPERATURE SENSOR ERROR

E42

VCb CLOSED CIRCUIT LEAK ERROR

E03

T2 TEMPERATURE SENSOR ERROR

E43

VCb CLOSED CIRCUIT LEAK ERROR

E04

T3 TEMPERATURE SENSOR ERROR

E44

AIR ELIMINATION TANK ERROR

E05

DIALYSATE TEMPERATURE MAX ALARM

E45

BLOWN-OUT FUSE DETECTION

E06

DIALYSATE TEMPARATURE MIN ALARM

E46

WATCH-DOG ERROR

E07

DIALYSATE PRESSURE MAXLIMIT

E47

RAM ERROR

E08

DIALYSATE PRESSURE MINLIMIT

E48

EEPROM ERROR

E09

DAIALYSATE PRESSUREMAX ALARM

E49

EPROM ERROR

E10

DIALYSATE PRESSUREMIN ALARM

E50

RAM INITIALIZE

E11

VENOUS PRESSURE MAX LIMIT

E51

EEPROM INITIALIZE

E12

VENOUS PRESSURE MIN LIMIT

E52

SWITCH ERROR

E13

VENOUS PRESSURE MAX ALARM

E53

I/O PORT ERROR

E14

VENOUS PRESSURE MIN ALARM

E54

CONNECTION ERROR SBY0151-SBY0156

E15

ARTERIAL PRESSURE MAX LIMIT

E55

BP BATTERY ERROR

E16

ARTERIAL PRESSURE MIN LIMIT

E56

COUPLER SW ON ALARM(DIALYSIS)

E17

ARTERIAL PRESSURE MAX ALARM

E57

COUPLER SW ON ALARM(PREP.)

E18

ARTERIAL PRESSURE MIN ALARM

E58

COUPLER SW OFF ALARM

E19

TMP ALARM

E59

UF MAX ALARM

E20

DIALYSATE CONDUCTIVITY MAX ALARM

E60

SYRINGE INFUSION COMPLETION

E21

DIALYSATE CONDUCTIVITY MIN ALARM

E61

COMPLETION OF SYRINGE.P INFUSION

E22

BLOOD LEAK SENSOR ERROR

E62

PERFORMED PRIMING

E23

BLOOD LEAK ALARM

E63

PREFORMED TARGET UF.

E24

AIR BUBBLES ARE DETECTED

E64

NO WATER RINSING

E25

HEATER OVER HEAT ALARM

E65

PREPARE DIALYSATE CONC.

E26

SHORT WATER SUPPLY

E66

CONC CONNECTOR TO RINSING PORT.

E27

SHORT DIALYSATE FLOW

E67

RESET SW

E28

DIALYSATE FLOW STOP

E68

UF DIALYSE SW

E29

SN SWITCH TIME ALARM

E69

BYPASS SW

E30

BLOOD PUMP ERROR

 

ALARM/ERROR DESCRIPTION

E31

BLOOD PUMP STOP ERROR

 

Disabled LCD indication

E32

CLV POSITION ERROR

 

Disabled SW function

E33

VISCOUS PUMP DRIVING ERROR(VPa)

 

Disabled syringe pump operation

E34

VISCOUS PUMP DRIVING ERROR(VPb)

 

Blood pump error

E35

VP DRAWING VOLUME IS AT MAX LIM.

 

Level change in venous drip chamber

E36

VCa CLOSED CIRCUIT LEAK ERROR

 

Air inclusion in dialysate

E37

VCa CLOSED CIRCUIT LEAK ERROR

 

Unstable dialysate flow rate

E38

VCb CLOSED CIRCUIT LEAK ERROR

 

Disabled discharge by pressing drain SW

E39

VCb CLOSED CIRCUIT LEAK ERROR

 

UFR error exceeding allowable upper/lower limits

E40

VCa CLOSED CIRCUIT LEAK ERROR

 

Date and time are initialized.

PREFACE

Thank you for purchase of NIPRO Single Patient Dialysis Machine “SURDIAL”.

Information for serviceman or technician required to perform check and maintenance of SURDIAL is given by this Trouble Shooting Manual. All operations described in this Manual should be performed by only qualified serviceman or technician who should be familiar with replacement of components of microprocessor based electronic devices and should have basic knowledge concerning conventional dialysate line of dialysis machine. The serviceman or technician should be familiar with all operations contained in this Manual by reading it thoroughly to full understand them before starting trouble shooting.

Use this Manual together with Operation and Service Manuals of SURDIAL. Same descriptions contained in them are minimal.

Generally, information required to operate SURDIAL is contained in Operation Manual of SURDIAL though technical description is also contained partially. Information related to maintenance, component replacement and various settings is contained in Service Manual of SURDIAL. In this Manual, information related to method of trouble shooting is contained.

Description concerning the trouble shooting contained in this Manual is protected by copyright owned by the manufacturer. Partial or whole copy, duplication or preparation of version in any foreign language of this Manual should not be allowed without obtaining prior written approval of the manufacturer.

E01 RECOVER FROM EMERGENCY STOP

Detection

 

 

Recovery to normal from emergency machine stop due to power failure caused during execution of sequence

 

 

 

 

 

 

 

 

 

 

 

 

Detection

 

Audible

Dialysate line function

 

BP/SP stop

CLV function

Reset

process

 

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rinse

 

 

 

(3 times)

Execution of sequence

before

 

Preparation

 

 

error detection

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Becoming error free of upper/lower alarm limits of

Reset

 

Automatic recovery to normal

 

Note

 

venous, arterial and dialysate pressures by error

 

 

 

 

 

 

 

 

 

 

detection

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Conditionor cause

Check and action

Recovery from power failure

Not required

Intentional OFF-ON of power breaker

Not required

Intentional OFF-ON of power SW

Not required

 

Check Identify position(s) of leakage or short-circuit of AC230V

 

(AC110V) line.

 

Disconnect components one by one from the line to identify

OFF of breaker due to leakage or short

damaged one(s) by testing or operate components in the line

one by one manually to do so.

circuit of AC230V (AC110V) line

 

 

 

Check correct earth connection of SURDIAL and disconnect

 

components after disconnecting power plug from receptacle.

 

Action: Repair or replace damaged one(s).

Incorrect power plug connection

Action: Connect the plug to receptacle securely.

 

Check: Pull lightly each of harnesses connecting POWER PLUG

 

SWITCH GEAR TERMINAL BLOCK2 POWER SWITCH

 

NOISE FILTER TERMINAL BLOCK1 POWER SUPPLY

 

SUB-BOARD CN412 SUB-BOARD CN413 MAIN BOARD

Incorrect power line connection

CN1 in series to check loosening, incorrect contact and

 

disconnection.

 

Action If the loosening is detected, tighten securely. If the incorrect

 

contact and/or disconnection are/is detected, replace relevant

 

harness(es) with new one(s).

Conditionor cause

 

 

Check and action

 

 

 

Check: Identify

whether

error

cause(s)

is (are)

main board,

 

sub-board and/or DC+5V/±12V switching power supply.

 

Check that

actual

output

voltage is

4.8V

(nominal: 5VDC)

 

when measured at test points on main and sub-board.

 

Action

 

 

 

 

 

Repetition of error of recovery from

When 4.8V is not outputted to the main board and sub-board:

Replace sub-board with new one or the switching power

emergency stop

supply with new one.

 

 

 

 

 

 

 

 

When it is outputted to sub-board, while it is not outputted to

 

main board:

 

 

 

 

 

Replace main board with new one.

 

 

 

When it is outputted to the main board and sub-board:

 

Replace main board with new one.

 

 

E02 T1 TEMPERATURE SENSOR ERROR

Detection

 

T1 temperature exceeding measurable limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

 

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

 

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rinse wait

 

(7 times)

Continuance

 

 

Rinse

 

 

 

 

 

 

Rinse end

 

 

 

 

 

 

Air purge

 

 

 

 

 

 

Dialysis

 

(Continuous)

 

 

 

 

 

wait

 

Dialysate line stop

 

 

 

 

 

Preparation

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rese

Press [RESET] SW.

 

Note

 

 

 

t

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

 

 

 

 

Check: Check whether indications of all temperature sensors and other

 

 

 

sensors such as dialysate pressure sensor are incorrect.

Indication error of temperature and

Check stable output of voltage of 4.9V (nominal: 5Vref) at test

other analog system

 

point on main board.

 

 

 

 

Action: Replace main board with new one.

 

 

 

 

Check: Check no incorrect wiring, loosening, corrosion and/or

 

 

 

disconnection at TM1 T1 terminal of main board.

 

Incorrect [ ] terminal connection

Action: If loosened, tighten securely.

 

 

 

 

 

 

 

 

If incorrectly wired, correct.

 

 

 

 

If corroded or disconnected, replace [T1] with new one.

 

 

 

Check: Remove TM1 T1 terminal from main board

and measure

Internal disconnection or

short circuit

resistance between Red and White to find

short circuit,

disconnection and fluctuation.

 

of [T1]

 

 

 

 

 

 

 

 

 

 

Action: Replace [T1] with new one.

 

Introduction of high temperature water

Action: Overhaul dialysate line to correct potential damage.

from water supply source

 

 

 

 

 

 

Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits

Incorrect

indicated

temperature

of main board using temperature calibrator.

 

calibration of main board

 

If the calibration is impossible, replace main board with new

 

 

 

one.

 

E03 T2 TEMPERATURE SENSOR ERROR

Detection

 

T2 temperature exceeding measurable limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

 

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

 

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rinse wait

 

(7 times)

Continuance

 

 

Rinse

 

 

 

 

 

 

Rinse end

 

 

 

 

 

 

Air purge

 

 

 

 

 

 

Dialysis

 

(Continuous)

 

 

 

 

 

wait

 

Dialysate line stop

 

 

 

 

 

Preparation

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rese

Press [RESET] SW.

 

Note

After Ver 3, measurable extent of upper limit while

t

 

rinse isn’t detectable.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

 

 

 

 

Check: Check whether indications of all temperature sensors and other

 

 

 

sensors such as dialysate pressure sensor are incorrect.

Indication error of temperature and

Check stable output of voltage of 4.9V (nominal: 5Vref) at test

other analog system

 

point on main board.

 

 

 

 

Action: Replace main board with new one.

 

 

 

 

Check: Check no incorrect wiring, loosening, corrosion and/or

 

 

 

disconnection at TM1 T2 terminal of main board.

 

Incorrect [ 2] terminal connection

Action: If loosened, tighten securely.

 

 

 

 

 

 

 

 

If incorrectly wired, correct.

 

 

 

 

If corroded or disconnected, replace [T2] with new one

 

 

 

Check: Remove TM1 T2 terminal from main board

and measure

Internal disconnection or

short circuit

resistance between Red and White to find

short circuit,

disconnection and fluctuation.

 

of [T2]

 

 

 

 

 

 

 

 

 

 

Action: Replace [T2] with new one.

 

 

 

 

Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits

Incorrect

indicated

temperature

of main board using temperature calibrator.

 

calibration of main board

 

If the calibration is impossible, replace main board with new

 

 

 

one.

 

E04 T3 TEMPERATURE SENSOR ERROR

Detection

 

 

T3 temperature exceeding measurable limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

 

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

 

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rinse wait

 

(7 times)

Continuance

 

 

Rinse

 

 

 

 

 

 

 

 

Rinse end

 

 

 

 

 

 

Air purge

 

 

 

 

 

 

 

 

Dialysis

 

 

 

(Continuous)

 

 

 

 

 

wait

 

 

 

Dialysate line stop

 

 

 

 

 

Preparation

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW.

 

Note

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

 

 

 

 

Check: Check whether indications of all temperature sensors and other

 

 

 

sensors such as dialysate pressure sensor are incorrect.

Indication error of temperature and

Check stable output of voltage of 4.9V (nominal: 5Vref) at test

other analog system

 

point on main board.

 

 

 

 

Action: Replace main board with new one.

 

 

 

 

Check: Check no incorrect wiring, loosening, corrosion and/or

 

 

 

disconnection at TM1 T3 terminal of main board.

 

Incorrect [ 3] terminal connection

Action: If loosened, tighten securely.

 

 

 

 

 

 

 

 

If incorrectly wired, correct.

 

 

 

 

If corroded or disconnected, replace [T2] with new one

 

 

 

Check: Remove TM1 T3 terminal from main board

and measure

Internal disconnection or

short circuit

resistance between Red and White to find

short circuit,

disconnection and fluctuation.

 

of [T3]

 

 

 

 

 

 

 

 

 

 

Action: Replace [T3] with new one.

 

 

 

 

Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits

Incorrect

indicated

temperature

of main board using temperature calibrator.

 

calibration of main board

 

If the calibration is impossible, replace main board with new

 

 

 

one.

 

E05 DIALYSATE TEMPERATURE MAX ALARM

Detection

 

Dialysate temperature T3 of upper alarm limit during heater ON

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

 

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

 

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Preparation

 

(Continuous)

Auto bypass

 

 

Prep. end

 

 

 

 

Heater OFF

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press [RESET] SW to reset.

 

 

 

 

 

Reset

 

Auto

reset is caused after release from alarm

Note

 

 

 

 

 

condition for 10 seconds.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Actually higher dialysate temperature

Check: Check whether actual dialysate temperature is higher than

upper alarm limit by touching dialysate supply line with hand

 

 

when the line is connected to dialyzer or by measurement of

 

temperature of dialysate discharged from coupler outlet using

 

thermometer when patient is not connected to the machine.

Failure of main board

Action: Replace main board with new one.

 

Incorrect calibration of indicated

Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits

temperature of main board

of main board using temperature calibrator.

If the calibration is impossible, replace main board with new one.

Higher supply water temperature

Action: Adjust supply water temperature as specified.When heat exchanger is used : 5 to 30When heat exchanger is not used 17 to 30

Lower priming flow rate

Action: Adjust priming flow rate to 700 to 750mL/min.

Actually

acceptable

dialysate

Check: Check whether actual dialysate temperature is acceptable by

temperature

 

touching dialysate supply line with hand when the line is

 

 

 

connected to dialyzer or by measurement of temperature of

 

 

 

dialysate discharged from coupler outlet using thermometer

 

 

 

when patient is not connected to the machine.

 

 

 

Check: Check whether indicated temperature is fluctuated when

Failure of T3 temperature sensor

harness of T3 temperature sensor is shaken.

 

 

 

Check also whether indicated T3 temperature is deviated when

 

 

 

temperature calibrator is connected.

 

 

 

Action: If check result is unacceptable, replace T3 temperature sensor

 

 

 

with new one.

 

 

 

Check: Check whether indicated T3 temperature is deviated when

Incorrect calibration of main board

temperature calibrator is connected.

 

 

 

Action: Calibrate [T3] circuit of main board using temperature

 

 

 

calibrator.

 

 

 

If the calibration is impossible, replace main board with new

 

 

 

one.

E06 DIALYSATE TEMPERATURE MIN ALARM

Detection

 

Dialysate temperature T3 of lower alarm limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

 

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

 

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis

 

 

(3 times)

Auto bypass

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press [RESET] SW to reset.

 

 

 

 

 

Reset

 

Auto reset is caused

after release from alarm

Note

 

 

 

 

 

condition for 10 seconds.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Actually lower dialysate temperature

Failure of main board

Incorrect calibration of indicated temperature of main board

Lower supply water temperature

Failure of heater

Higher priming flow rate

Actually acceptable dialysate temperature

Failure of T3 temperature sensor

Incorrect calibration of main board

Check: Check whether actual dialysate temperature is lower than lower alarm limit by touching dialysate supply line with hand when the line is connected to dialyzer or by measurement of temperature of dialysate discharged from coupler outlet using thermometer when patient is not connected to the machine.

Action: Replace main board with new one.

Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits of main board using temperature calibrator.

If the calibration is impossible, replace main board with new one.

Action: Adjust supply water temperature as specified.When heat exchanger is used : 5 to 30When heat exchanger is not used 17 to 30

Check: Measure resistance between both ends of heater at room temperature after power OFF.

Action: Replace heater if the resistance is out of the below ranges: 230V spec. 38.3 to 44.1Ω

110V spec. 8.7 to 10.0Ω

Action: Adjust priming flow rate to 700 to 750ml/min.

Check: Check whether actual dialysate temperature is acceptable by touching dialysate supply line with hand when the line is connected to dialyzer or by measurement of temperature of dialysate discharged from coupler outlet using thermometer when patient is not connected to the machine.

Check: Check whether indicated temperature is fluctuated when harness of T3 temperature sensor is shaken.

Check also whether indicated T3 temperature is deviated when temperature calibrator is connected.

Action: If check result is unacceptable, replace T3 temperature sensor with new one.

Check: Check whether indicated T3 temperature is deviated when temperature calibrator is connected.

Action: Calibrate [T3] circuit of main board using temperature calibrator.

If the calibration is impossible, replace main board with new one.

E07 DIALYSATE PRESSURE MAX LIMIT

Detection Dialysate pressure of +400mmHg

Detection

Audible

Dialysate line function

 

BP/SP stop

CLV function

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rinse wait

(7 times)

UF stop

 

2 sec.

2 sec.

Rinse

ON of V4 and V10 for 2 sec.

 

 

 

 

Rinse end

 

 

 

 

 

 

Air purge

 

 

 

 

 

 

Dialysis wait

(Continuous)

 

 

 

 

 

 

Preparation

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

Auto reset of BP, SP and CLV

Note

 

 

 

Press [RESET] SW to reset UF stop.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Pressure rise due to failure of [PR1]

Action: Adjustment of [PR1]

In adjustment is impossible, repair or replace [PR1].

 

Leak from [V10]

Action: Repair or replacement of [V10]

Error occurrence at gas purge

 

Occlusion of [R2]

Action: Change [R2] so that dialysate pressure becomes 0 to

 

100mmHg during gas purge.

Excess tightening of [PC2]

Action: Adjust [PC2] so that dialysate pressure becomes 0 to

 

100mmHg during gas purge.

Impossible opening of [V4]

Check: Check whether [V4] functions when it is turned on or off in

 

maintenance mode 4 screen.

 

If not functions, remove connector from [V4] and check whether

 

24V is outputted to the connector when [V4] is ON in

 

maintenance mode 4 screen.

Action: If 24V is not outputted, replace main board with mew one.

If it is outputted, replace [V4] with mew one.

Pressure rise due to folding of blood

Action: Correct folding.

line

 

Check: Disconnect hose from the sensor Assy to expose to

 

atmosphere for checking whether indicated pressure is nearly

 

0mmHg.

Failure of dialysate pressure sensor

 

board or the sensor

Action: If the indicated pressure is not nearly 0mmHg, calibrate the

 

sensor board.

 

If calibration is impossible, replace the board or sensor with new

 

one.

 

 

E07 DIALYSATE PRESSURE MAX LIMIT

Condition or cause

Check and action

Disconnection of harness from the

Check: Pull lead wire (harness) lightly to check whether it is separated

or loosened.

sensor Assy or insufficient contact of

 

connector of it

Action: Repair

 

Drain tube is blocked.

Check Remove drain tube from drain port and check drain liquid is

 

drained from port correctly or not.

No folding/kinking of Drain tube

Action Remove the trouble cause.

Drain tube tip is placed too higher

Action Drain tube tip must be placed above the machine floor by 0 to

position.

60 cm.

Drain tube is too long.

Action Use tube of an inter diameter of 8 mm and a length of 2 m or

 

shorter to connect to the port.

 

 

E08 DIALYSATE PRESSURE MIN LIMIT

Detection Dialysate pressure of -400mmHg

Detection

Audible

Dialysate line function

 

BP/SP stop

CLV function

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rinse wait

(7 times)

UF stop

 

2 ses.

2 sec.

Rinse

ON of V4 and V10 for 2 sec.

 

 

 

 

Rinse end

 

 

 

 

 

 

Air purge

 

 

 

 

 

 

Dialysis wait

(Continuous)

 

 

 

 

 

 

Preparation

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

Auto reset of BP, SP and CLV

Note

 

 

 

Press [RESET] SW to reset UF stop.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

 

Check: Check whether the line is folded or clamped by forceps

Occlusion of dialysate line to dialyzer

visually.

 

 

Action: Correct folding or clamp.

Excess occlusion in dialyzer

Action: Replace dialyzer with new one.

 

Check: Check stop of leak by clamping [V4] line with forceps.

Leak from [V4]

Action: Clean [V4] by rinsing after disassembling or replace it with new

 

 

one.

Opening of [V4] due to malfunction of

Action: Clean air-elimination tank [AS2] by rinsing after disassembling

[FSW]

or replace it with new one.

 

Action: Adjust pressure in closed line.

Occlusion of [RV2]

If adjustment is impossible, Clean [RV2] by rinsing after

 

disassembling or replace it with new one.

Error occurrence during gas purge

Failure of [V10] Check: Check whether [V10] functions when it is turned on or off in maintenance mode 4 screen.

If not functions, disconnect connector from [V10] and turn on it in maintenance mode 4 screen to check whether 24V is outputted to the connector.

Action: If 24V is not outputted, replace main board with new one.

If it is outputted, replace [V10] with new one.

Failure of dialysate pressure sensor board or the sensor

Disconnection of harness from the sensor Assy or insufficient contact of connector of it

Low water supply flow

No water supply

Low water supply pressure

Alarm while gaspurge

On the acid rince without an acid rince option.

Check: Disconnect hose from the sensor Assy to expose to atmosphere for checking whether indicated pressure is nearly 0mmHg.

Action: If the indicated pressure is not nearly 0mmHg, calibrate the sensor board.

If calibration is impossible, replace the board or sensor with new one.

Check: Pull lead wire (harness) lightly to check whether it is separated or loosened.

Action: Repair

Action Increase water supply flow rate more than 750 mL/min. Check/Action See E26 SHORT WATER SUPPLY. Check/Action See E26 SHORT WATER SUPPLY. Check/Action See E26 SHORT WATER SUPPLY. Check/Action See E26 SHORT WATER SUPPLY.

E09 DIALYSATE PRESSURE MAX ALARM

Detection Dialysate pressure of set upper alarm limit

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

 

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

UF stop

 

 

2 sec.

2 sec.

 

Preparation

 

ON of V4 and V10 for 2 sec.

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

Auto reset of BP, SP and CLV

Note

By the alarm actuation, upper/lower alarm limits of

Press [RESET] SW to reset UF stop.

venous pressure become error free.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Simultaneous actuation of this alarm

Check: See E13 VENOUS PRESSURE MAX ALARM.

with venous pressure max alarm

Action: See E13 VENOUS PRESSURE MAX ALARM.

 

Dialysate pressure rise due to UF stop

Action: Remove cause of UF stop.

Dialysate pressure rise due to failure of

Action: Adjust [PR1].

[PR1]

If adjustment is impossible, repair [PR1] or replace it with new

one.

 

Leak from [V10]

Action: Repair [V10] or replace it with new one.

Leak from [V5] and [V7]

Action: Repair [V5] and [V7] or replace them with new ones.

 

 

E10 DIALYSATE PRESSURE MIN ALARM

Detection

 

Dialysate pressure of set lower alarm limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

 

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

 

Reset

process

 

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

UF stop

 

 

2 sec.

2 sec.

 

Preparation

 

 

ON of V4 and V10 for 2 sec.

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Auto reset of BP, SP and CLV

Note

By the alarm actuation, upper/lower alarm limits of

 

Press [RESET] SW to reset UF stop.

venous pressure become error free.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Simultaneous actuation of this alarm with venous pressure min alarm

Check: See E14 Venous pressure min alarm.

Action: See E14 Venous pressure min alarm.

 

Action: Adjust pressure in closed line.

Occlusion of [RV2]

If adjustment is impossible, clean [RV2] by rinsing after

 

disassembling or replace it with new one.

 

Check: Check no dialysate pressure change by clamping [V4] line with

 

forceps.

Leak from [V4]

 

 

Action: In the case of no change, clean [V4] by rinsing after

 

disassembling or replace it with new one.

Opening of [V4] due to failure of [FSW]

Check: See E44 AIR ELIMINATION TANK ERROR

The opening may be simultaneously

 

caused with E44 AIR ELIMINATION

Action: See E44 AIR ELIMINATION TANK ERROR

TANK ERROR

 

Leak from [V6] and [V8]

Action: Repair [V6] and [V8] or replace them with new ones.

Excess occlusion in dialyzer

Action: Replace dialyzer with new one if required.

E11 VENOUS PRESSURE MAX LIMIT

Detection

 

Venous pressure of +400mmHg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

(7 times)

UF stop

 

 

Rinse

 

 

 

 

 

 

 

 

Rinse end

 

 

 

 

 

 

Air purge

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

 

 

 

 

 

 

Preparation

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW to recover.

Note

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Folding or occlusion of blood line

 

between Venous Drip Chamber and

Action: Remove the trouble cause.

patient or occlusion of the line with

 

Robert clamp

 

Thrombus formed at mesh of venous

Action: Remove thrombus or replace blood line with new one.

drip chamber or at shunt connector

 

Error occurrence during single needle

Action: Set upper pressure change limit of SN correctly.

dialysis

 

 

Check: Disconnect blood line from venous pressure inlet port to

 

expose to atmosphere and check indicated venous pressure is

Failure of venous pressure sensor

nearly 0mmHg.

board

 

 

Action: If nearly 0mmHg is not indicated, calibrate the board.

 

If calibration is impossible, replace the board with new one.

E12 VENOUS PRESSURE MIN LIMIT

Detection

 

Venous pressure of -400mmHg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

(7 times)

UF stop

 

 

Rinse

 

 

 

 

 

 

 

 

Rinse end

 

 

 

 

 

 

Air purge

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

 

 

 

 

 

 

Preparation

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW to recover.

Note

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Simultaneous occurrence of this error

Action: Remove cause of dialysate related error.

with dialysate related error

 

Simultaneous occurrence of this error

Action: Remove cause of arterial pressure related error

with arterial pressure related error

 

Press of DRAIN SW without removing

 

coupler of dialyzer under occluding

Note: DRAIN SW operation-miss

venous line with Robert clamp

 

 

Check: Disconnect blood line from venous pressure inlet port to

 

expose to atmosphere and check indicated venous pressure is

Failure of venous pressure sensor

nearly 0mmHg.

board

 

 

Action: If nearly 0mmHg is not indicated, calibrate the board.

 

If calibration is impossible, replace the board with new one.

E13 VENOUS PRESSURE MAX ALARM

Detection

 

Venous pressure of set upper alarm limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

 

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

UF stop

 

 

 

Preparation

 

 

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW to recover.

Note

By the alarm actuation, upper/lower alarm limits of

 

dialysate pressure become error free.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

 

Check: Check shunt occlusion or blood line folding.

Occlusion of venous line.

 

 

Action: Remove the trouble.

 

Check: Check shunt occlusion or blood line folding.

Occlusion of arterial line

 

 

Action: Remove the trouble.

Venous pressure change due to patient

Action: If the change is significant, raise somewhat upper alarm limit of

movement

venous pressure at discretion of doctor.

 

Check: Check whether discharge flow rate, pressure applied to rolling

 

tube and rotor rotation of blood pump are normal.

Significant discharge pressure change

 

of blood pump

Action: Adjust the occlusion or the pump rpm correctly.

 

If adjustment is impossible, replace the pump or sub-board

 

with new one.

Nipro Surdial Trouble shooting manual

E14 VENOUS PRESSURE MIN ALARM

Detection

 

Venous pressure of set lower alarm limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

 

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

UF stop

 

 

 

Preparation

 

 

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW to recover.

Note

By the alarm actuation, upper/lower alarm limits of

 

dialysate pressure become error free.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Blood pressure drop of patient

Action: Stop UF and monitor patient condition.

Correct if set lower alarm limit of venous pressure is incorrect.

Check: Check no introduction of blood in venous pressure sensor line.

Dislodgment of venous line from

 

venous pressure inlet port

Action: When blood is not introduced in the line, connect venous line

 

to the port. If introduced, replace the sensor line with new one.

 

Check: Check shunt occlusion or venous line folding.

Occlusion of venous line

 

 

Action: Remove the trouble.

 

Check: Check shunt occlusion or arterial line folding.

Occlusion of arterial line.

 

 

Action: Remove the trouble.

Venous pressure change due to patient

Action: If the change is significant, lower somewhat lower alarm limit

movement

of venous pressure at discretion of doctor.

 

Check: Check whether discharge flow rate, pressure applied to rolling

 

tube and rotor rotation of blood pump are normal.

Significant discharge pressure change

 

of blood pump

Action: Adjust the occlusion or the pump rpm correctly.

 

If adjustment is impossible, replace the pump or sub-board

 

with new one.

E15 ARTERIAL PRESSURE MAX LIMIT

Detection

 

Arterial pressure of +400mmHg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

( times)

UF stop

 

 

Rinse

 

 

 

 

 

 

 

 

Rinse end

 

 

 

 

 

 

Air purge

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

 

 

 

 

 

 

Preparation

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW to recover.

Note

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Simultaneous actuation of this error

Action: Remove cause of dialysate pressure related error.

with dialysate pressure related error

 

Simultaneous actuation of this error

Action: Remove cause of venous pressure related error.

with venous pressure related error

 

Folding or occlusion of blood line

 

between arterial Drip chamber and

Action: Remove the trouble cause.

dialyzer or occlusion of the line with

 

Robert clamp

 

Folding or occlusion of blood line

 

between venous Drip chamber and

Action: Remove the trouble cause.

patient or occlusion of the line with

 

Robert clamp

 

Thrombus formed at mesh of arterial

Action: Remove thrombus or replace blood line with new one.

drip chamber or

 

 

Check: Disconnect blood line from arterial pressure inlet port to

 

expose to atmosphere and check indicated arterial pressure is

Failure of arterial pressure sensor

nearly 0mmHg.

board

 

 

Action: If nearly 0mmHg is not indicated, calibrate the board.

 

If calibration is impossible, replace the board with new one.

E16 ARTERIAL PRESSURE MIN LIMIT

Detection

 

Arterial pressure of -400mmHg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

( times)

UF stop

 

 

Rinse

 

 

 

 

 

 

 

 

Rinse end

 

 

 

 

 

 

Air purge

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

 

 

 

 

 

 

Preparation

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW to recover.

Note

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

 

Check and action

Simultaneous actuation of this error

Action: Remove cause of dialysate pressure related error.

with dialysate pressure related error

 

Thrombus formed at arterial

shunt

Action: Remove thrombus or replace blood line with new one.

connector

 

 

 

Folding or occlusion of blood line

 

between arterial drip chamber

and

Action: Remove the trouble cause.

patient or occlusion of the line with

 

Robert clamp

 

 

 

 

Check: Disconnect blood line from arterial pressure inlet port to

 

 

expose to atmosphere and check indicated arterial pressure is

Failure of arterial pressure sensor

nearly 0mmHg.

board

 

 

 

 

Action: If nearly 0mmHg is not indicated, calibrate the board.

 

 

If calibration is impossible, replace the board with new one.

E17 ARTERIAL PRESSURE MAX ALARM

Detection

 

Arterial pressure of set upper alarm limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

 

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

UF stop

 

 

 

Preparation

 

 

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW to recover.

Note

By the alarm actuation, set upper/lower alarm limits

 

of dialysate pressure become error free.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

 

Check: Check shunt occlusion and blood line folding.

Occlusion of venous line.

 

 

Action: Correct the trouble.

 

Check: Check shunt occlusion and blood line folding.

Occlusion of arterial line.

 

 

Action: Correct the trouble.

Arterial pressure change due to patient

Action: If the change is significant, raise set upper alarm limit of

movement

arterial pressure somewhat at discretion of doctor.

 

Check: Check whether discharge flow rate, pressure applied to rolling

 

tube and rotor rotation of blood pump are normal.

Significant discharge pressure change

 

of blood pump

Action: Adjust the occlusion or the pump rpm correctly.

 

If adjustment is impossible, replace the pump or sub-board

 

with new one.

E18 ARTERIAL PRESSURE MIN ALARM

Detection

 

Arterial pressure of set lower alarm limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

 

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis wait

(Continuous)

UF stop

 

 

 

Preparation

 

 

 

 

 

 

 

 

Prep. end

 

 

 

 

 

 

 

 

 

 

Dialysis

 

 

 

 

 

 

 

 

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW to recover.

Note

By the alarm actuation, set upper/lower alarm limits

 

of dialysate pressure become error free.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

Check and action

Blood pressure drop of patient

Action: Stop UF and monitor patient condition.

Correct if set lower alarm limit of venous pressure is incorrect.

Check: Check no introduction of blood in arterial pressure sensor line.

Dislodgment of blood line form arterial

pressure inlet port Action: When blood is not introduced in the line, connect arterial line to the port. If introduced, replace the arterial line with new one.

Check: Check shunt occlusion and blood line folding.

Occlusion of venous line.

 

Action: Correct the trouble.

 

Check: Check shunt occlusion and blood line folding.

Occlusion of arterial line.

 

 

Action: Correct the trouble.

Arterial pressure change due to patient

Action: If the change is significant, raise set upper alarm limit of

movement

arterial pressure somewhat at discretion of doctor.

 

Check: Check whether discharge flow rate, pressure applied to rolling

 

tube and rotor rotation of blood pump are normal.

Significant discharge pressure change

Action: Adjust the occlusion or the pump rpm correctly.

of blood pump

If adjustment is impossible, replace the pump or sub-board

 

 

with new one

E19 TMP ALARM

Detection TMP (VP - DP) of set alarm limit

Detection

Audible

Dialysate line function

 

 

BP/SP stop

CLV function

Reset

process

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis

 

(Continuous)

UF stop

 

 

Recovery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset

 

Press [RESET] SW to recover.

Note

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

 

Check and action

Excess occlusion in dialyzer

 

Action: Replace dialyzer with new one at discretion of doctor.

Lower setting of TMP alarm limit

 

Action: Set TMP to appropriate value.

E09

DIALYSATE

PRESSURE

MAX

 

ALARM

 

 

 

E10

DIALYSATE

PRESSURE

MIN

Action: See E09 DIALYSATE PRESSURE MAX ALARM

ALARM

 

 

See E10 DIALYSATE PRESSURE MIN ALARM

E13 VENOUS PRESSURE MAX ALARM

See E13 VENOUS PRESSURE MAX ALARM

Simultaneous actuation of this alarm

See E14 VENOUS PRESSURE MIN ALARM

with E14 VENOUS PRESSURE MIN

 

ALARM

 

 

 

E07 DIALYSATE PRESSURE MAX LIMIT

 

E08 DIALYSATE PRESSURE MIN LIMIT

Action: See E07 DIALYSATE PRESSURE MAX LIMIT.

E11 VENOUS PRESSURE MAX LIMIT

See E08 DIALYSATE PRESSURE MIN LIMIT.

Simultaneous actuation of this alarm

See E11 VENOUS PRESSURE MAX LIMIT.

with

E12 VENOUS PRESSURE

MIN

See E12 VENOUS PRESSURE MIN LIMIT.

LIMIT

 

 

 

 

E20 DIALYSATE CONDUCTIVITY MAX ALARM

Detection

 

Dialysate conductivity of upper alarm limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detection

 

Audible

 

 

Dialysate line function

 

 

BP/SP stop

CLV function

 

Reset

process

 

alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dialysis

 

(Continuous)

 

Auto bypass

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press [RESET]SW.

 

 

 

 

 

 

 

 

 

Reset

Auto reset after release from alarm condition for 10

Note

 

 

 

 

 

sec.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition or cause

 

 

 

 

 

Check and action

 

Abnormal actual conductivity

 

 

Check: Take dialysate sample from [SL] to measure actual

 

 

 

 

 

 

 

 

conductivity by specified method.

 

No adjustment of conductivity

 

 

Action: Set conductivity correctly and calibrate indicated conductivity

 

 

 

 

 

 

 

 

in maintenance mode 5 screen.

 

Failure of main board

 

 

Check: See E33 VISCOUS PUMP DRIVING ERROR(VPa) and

 

 

 

 

 

 

 

 

 

E34 VISCOUS PUMP DRIVING ERROR(VPb).

 

 

 

 

 

 

 

 

 

Action: See E33 VISCOUS PUMP DRIVING ERROR(VPa) and

 

 

 

 

 

 

 

 

 

E34 VISCOUS PUMP DRIVING ERROR(VPb).

 

Failure of [V13] or [V14]

 

 

Action: Repair [V13] or [V14] (V15)) or replace it with new one.

Failure of [V2] or [V3]

 

 

Action Repair [V2] or [V3] or replace it with new one.

 

Higher or lower priming flow rate

 

 

Action: Adjust the flow rate to 700-750mL/min.

 

Failure of [C2]

 

 

Action: Replace [C2] with new one.

 

 

 

Folding,

occlusion or collapsing

of

Action: Correct the trouble.

 

 

 

concentrate line tube

 

 

 

 

 

 

 

 

 

E20 DIALYSATE CONDUCTIVITY MAX ALARM

Condition or cause

 

Check and action

Normal actual conductivity

 

Check: Take dialysate sample from [SL] to measure conductivity by

 

 

specified method.

[CD] electrode contaminated with dirt

Action: Rinse the electrode with acetic acid followed by water rinsing

such as CaCO3 or leak from

sufficiently.

conductivity sensor

 

If the dirt can not be removed, use specified detergent such

 

 

as CARON or SYUNMA to remove it after disassembling

 

 

conductivity sensor.

Incorrect [CD] terminal connection

Check: Check no miss-wiring, loosening, corrosion and

 

 

disconnection of TM1 CD terminal of main board.

 

 

Action: If loosened, tighten the connection securely.

 

 

If miss-wired, correct.

 

 

If corroded or disconnected, replace harness with mew one.

Internal disconnection or short circuit

Check: After disconnecting TM1 CD terminal from main board,

of [CD] harness

 

measure resistance between [CD] and [GND] to check no short

 

 

circuit, disconnection and fluctuation of measured resistance.

 

 

Action: Replace [CD] harness with new one.

Failure of main board

 

Check: Change ambient temperature of main board to check whether

 

 

conductivity and temperature indications are changed.

 

 

Action: If incorrectly indicated, replace main board with new one.

When Na infusion unit (option) is used:

 

Higher or lower concentration

of Na

Action: Use 10% Na concentrate.

concentrate

 

 

 

Use of other chemical

added

Action: Set proper upper alarm limit of conductivity by considering

concentrate

 

 

electrolyte concentration.

 

 

Use of other concentrate

 

Action: Use 10% Na concentrate.

 

 

Loading...
+ 79 hidden pages