Warning for Using this Product ........................................................................................................................................... 2
3 Offset function ........................................................................................................................................................ 8
IIName and Function of Each Part ............................................................................................................................... 11
1Name of Each Part .......................................................................................................................................... 11
1 T-RCP Remote Control Pad................................................................................................................................. 14
2 Startup and Shutdown of the Remote Control Pad............................................................................................... 16
3 Registering Data of the Objectives ....................................................................................................................... 17
For deleting the registered data........................................................................................................................... 19
Important: list of objectives requiring data registration......................................................................................... 19
4 Registering Installation Information of Objectives Mounted on Nosepiece........................................................... 19
2 Registering and Restoring the Offset Amount ...................................................................................................... 23
2.1 Registering the Offset Amount....................................................................................................................... 23
2.2 Restoring the Offset Amount ......................................................................................................................... 24
3 Registering and Restoring the Vertical Position of the Nosepiece........................................................................ 24
3.1 Registering the Vertical Position of the Nosepiece........................................................................................ 24
3.2 Escaping the Objective.................................................................................................................................. 25
3.3 Restoring the Vertical Position of the Nosepiece........................................................................................... 25
3.4 About the Movement of the Objective in Registering and Restoring the Vertical Position of the Nosepiece. 26
VConnecting a PC........................................................................................................................................................ 27
Connecting a PC ................................................................................................................................................. 27
Specification for serial interface communication.................................................................................................. 27
Communication commands................................................................................................................................. 27
Rear view of the PF Controller ............................................................................................................................ 33
1 During System Startup ......................................................................................................................................... 35
2 While Using the Perfect Focus Functions............................................................................................................. 35
VIIICare and Maintenance............................................................................................................................................... 38
2Cleaning This Product ...................................................................................................................................... 38
3Disinfecting This Product ................................................................................................................................. 38
• Before using this system, read warnings and cautions described in “Warning and Caution for Using this Product”
carefully and thoroughly, and be sure to follow them.
• For the devices using with this system (epi-fl attachment, differential interference contrast attachment, various
motorized devices, and so on), read warnings and cautions described in the instruction manuals of these devices
carefully and thoroughly, and be sure to follow them.
Operating Procedures
1 Basic Operation
Power on the microscope.
1
Caution: When the PF controller is turned on, the PF module will be initialized for about 10 seconds
and the information about the objectives on the nosepiece are acquired from the HUB
controller. So, before the PF controller, the microscope must be turned on.
Besides, if the information of the objectives is changed, the PF controller must be turned off
once, and then, turned on again.
Using the T-RCP remote control pad, register the information of the objectives mounted on the microscope into
2
the hub controller.
For detailed information about the registration procedure, see Chapter III, “Before Using this System.”
Turn on the power switch (switch to “I”) located on the right side face of the PF controller to power on the
3
controller.
The [POWER] LED, located on the front face of the PF controller, is lit.
Select the desired objective and put the specimen on the stage.
4
• Put the objective supported by this system into the optical path. Refer to IX. Specifications.
• The objective can be selected by using the remote control pad. For more information, see the instruction
manual of the remote control pad.
• Apply water immersion or oil immersion, if required. For more information, see, the instruction manual of the
microscope.
Focus on the specimen by using the coarse/fine focus knob on the microscope.
5
21
IV Operating Procedures
Press the [AF ON] key on the PF controller.
6
(1) The LED of the [AF ON] key is lit and the focus control
starts.
Caution: Manual operation of the fine focus knob is
disabled during the focus control. Do not
touch the coarse focus knob, however,
because this knob is not disabled during the
focus control.
(2) When the objective comes into the focus control range,
the [FOCUS] LED blinks. When the reference position
(boundary surface) comes into focus, the [FOCUS] LED is
lit.
* The focus control range is defined that the focus control
can be performed, and is predetermined for each
objective.
For blinking of the [DISABLE] LED:
The reference position (boundary surface) is not in the focus
control range if the LED blinks. Adjust the objective position by
moving using the coarse/fine focus knob of the microscope so
that the reference position (boundary surface) comes into the
focus control range. When the reference position (boundary
surface) comes into the focus control range, the focus control
starts.
* The completion of focusing on the boundary surface does not mean that the target position of the
specimen comes into the focus. Perform the step 7 below, the [OFFSET] knob operation, for your
target. In particular, when a dry type objective is used, the initial position may be shifted from the
boundary surface because of variations of cover glass thickness. So, after the [FOCUS] LED is lit,
adjust the offset knob to focus on the target.
Time-out function:
This system has the time-out function to keep the objective from contact with specimen. This function will be
activated and the objective will be stopped if the signal cannot be detected within 5 seconds during the focusing
with the [AF ON] operation. (The objective moves about 400 µm for the 5 seconds.)
When the time-out function is activated, the buzzer beeps, the LED of the [AF-ON] turns off, and the up/down
movement of the objective stops.
Recovery from the time-out function:
The reference position (boundary surface) is out of the range. Push the [OFFSET-RESET] key to restore the
offset amount to its original status. Focus on the specimen manually. And then, push the [AF-ON] key again.
22
IV Operating Procedures
Focus on the target of the specimen by manipulating the
7
[OFFSET] knob on the PF controller and start observation.
• Fine and coarse operations are provided for the function of
the [OFFSET] knob. The [FINE/COARSE] key can be used
to switch between the FINE operation and the COARSE
operation.
• The FINE/COARSE state can be checked with the LED on
the key.
Caution: The offset amount can be operated only when the
focus is being controlled (when the [AF ON] LED
is lit).
2 Registering and Restoring the Offset Amount
The offset amount specified during the observation can be registered into the memory of the PF controller. By pressing
the [FOCUS OFFSET – RECALL] key, the registered offset amount can be restored even after the offset amount is
changed by manipulating the [OFFSET] knob.
• The offset amount setting can be overwritten if required.
• Only one offset amount can be registered for one objective (one objective address). (Six positions are available in
total.)
• The registered offset amounts is stored while the PF controller is powered on.
2.1 Registering the Offset Amount
Adjust the offset amount by manipulating the [OFFSET] knob,
1
when the LED on the [AF ON] key of the PF controller is lit.
2
Press the [FOCUS OFFSET – MEMORY] key for over two
2
seconds.
The LED on the [MEMORY] key is lit and the offset amount is
registered.
1
23
IV Operating Procedures
2.2 Restoring the Offset Amount
Press the [FOCUS OFFSET – RECALL] key.
1
The nosepiece automatically moves to the position for where
the offset is registered.
Caution: Manual operation of the fine focus knob is
disabled during the focus control. Do not touch
the coarse focus knob, however, because this
knob is not disabled during the focus control.
3 Registering and Restoring the Vertical Position of the
Nosepiece
The TE2000-E microscope has a linear encoder inside, and the vertical position (the Z-axis position) of the nosepiece
can be reproduced by using the coordinates of the linear encoder. With this function, the vertical position of the
nosepiece can be registered into the memory of the PF controller for the desired observation condition.
Once the position is registered, pressing the [AF – RECALL] key can return the nosepiece to the registered position or its
vicinity and then the system restarts the focus control even after the nosepiece position was changed by manipulating
the coarse and fine focus knobs. (Note: The nosepiece may not return to its registered absolute position of the vertical
movement. It will start the focus control again from the vicinity of the registered position.)
This function is useful for the case supplying oil to the objectives or replacing the specimen where the objectives must be
escape.
• The vertical position of the nosepiece can be overwritten if required.
• Only one offset amount can be registered for one objective (one objective address). (Six positions are available in
total.)
• The registered vertical position of the nosepiece is stored while the PF controller is powered on.
3.1 Registering the Vertical Position of the Nosepiece
Focus on the target of the specimen.
1
• When the [AF ON] LED is lit:
Focus on the target by rotating the [OFFSET] knob.
• When the [AF ON] LED is not lit:
Focus on the target by rotating the coarse/fine focus knob
on the microscope.
Press the [FOCUS OFFSET – MEMORY] key for over two
2
seconds.
The LED on the [MEMORY] key is lit and the vertical position
of the nosepiece is registered.
24
IV Operating Procedures
3.2 Escaping the Objective
Press the [AF ON] key on the PF controller
1
(when the [AF ON] LED is lit).
The LED of the [AF ON] key is turned off and the focus control
is canceled.
Caution; The focusing control must be canceled before
changing or replacing the objectives.
Lower the nosepiece by manipulating the coarse focus knob, then, perform the required operation.
2
Examples of the operation:
• Replacing specimens
• Changing objectives
• Supplying oil or deionized water (or distilled water) to the objective
• The coarse focus knob is manipulated for other reasons, but the specimen is not moved.
3.3 Restoring the Vertical Position of the Nosepiece
The status must be one of the following two conditions.
1
[AF - MEMORY] LED: ON
[AF – ON] LED: OFF
Press the [AF – RECALL] key of the PF controller.
AF = OFF
or
[AF - MEMORY] LED: ON
[AF - ON] LED: ON
[DISABLE] LED: Blink
And no signal was detected although
the [AF - ON] key was pressed.
2
The buzzer sounds (one short beep). And then, the nosepiece
starts moving with blinking the [DISABLE] LED.
Caution: Manual operation of the fine focus knob is
disabled while restoring the nosepiece vertical
position. Do not touch the coarse focus knob,
however, because this knob is not disabled while
restoring.
Action of the equipment: When the nosepiece returned to the
3
registered vertical position or its vicinity, the [AF ON] LED is lit
and the focus control starts.
25
IV Operating Procedures
If the buzzer sounds (twice short beep or one long beep),
the vertical position of the nosepiece could not be restored. Perform
the followings:
• Manipulate the [AF – MEMORY] operation at the focus position
again.
• Change the height of the stage by manipulating the coarse/fine
focus knob.
* The completion of focusing on the boundary surface does not mean that the target position of the specimen comes
into the focus. Perform the step 7, the [OFFSET] knob operation, in “Basic Operation ” for your target. In particular,
when a dry type objective is used, the initial position may be shifted from the boundary surface because of variations
of cover glass thickness. So, after the [FOCUS] LED is lit, adjust the offset knob to focus on the target.
3.4 About the Movement of the Objective in Registering and
Restoring the Vertical Position of the Nosepiece
(1) Push the [AF-MEMORY] key after focusing.
Glass bottom dish
(1)
(2)
WD/2
(4)
(5)
(6)
(7)
Beep!
The vertical position of the nosepiece is
stored. (The tip of the objective is defined as
“Z1” in this explanation.)
(2) Escape the objective to perform an operation
such as oil exchange or so on.
(3) Push the [AF-RECALL] key.
(4) The nosepiece is driven to the registered
vertical position, Z1, and additionally driven
upward at a distance of the half of the WD.
(The WD is the working distance of the
objective.)
However, if the half of the WD is greater than
750 µm, 750 µm is set in place of the
calculated value.
(5) The focus position (boundary surface) is
searched for downward.
(6) When the focus position is detected, the
searching operation ends and the AF is turned
on.
(7) If the focus position cannot be detected, the
search operation ends in 20 seconds and a
sound beeps.
Z1
26
V
Connecting a PC
The hub controller can be controlled from the IBM PC-AT compatible computer connected to it.
Connecting a PC
Using a cross wired cable with D-SUB (9 pins) connectors, connect the “PC connector” of the hub controller and the
“serial port” of the PC.
Pin number of the PC
connector
1 DCD -
2 RxD Input
3 TxD Output
4 DTR Output
5 SG (GND)
6 DSR Input
7 RTS -
8 CTS -
9 RI -
Signal
Direction of the
signal
Communications cable
PC connector
PC connector: D-SUB (9pin) male
Pins numbered one, seven, eight, and nine
are opened inside the hub controller.
D-SUB (9 pins: female to female) and general cross wired cable (three meters maximum)
Specification for serial interface communication
Interactive signal connection circuit RS-232C (EIA standard compliant)
Baud rate 9600 bps
Data length 8 bit
Start bit 1
Stop bit 1
Parity check None
Communication commands
Contact your nearest Nikon representative.
27
V
Assembling
I
This chapter explains how to assemble this system and how to connect required components.
• Before assembling, read warnings and cautions described in “Warning and Caution for Using this Product”
• To prevent electric shock and fire, always turn off the power switch (switch to “O”) of all devices and unplug the
• Do not let fingers or hands get caught in parts or components.
• Scratches or finger prints on the lens adversely affects the image quality. Pay close attention to assemble the
Warning
carefully and thoroughly, and be sure to follow them.
power cord from the wall outlet.
Caution
system and do not touch the lens.
• Components of this system are precision optical instruments. Handle them carefully, and do not subject them to
any shocks. The precision of the objectives in particular can be adversely affected even by a weak shock.
Required tools
Two 2-mm hexagonal screwdrivers (supplied with the microscope)
One 4-mm hexagonal wrench (supplied with the microscope)
One 2.5-mm hexagonal wrench (supplied with the motorized cassette holder)
28
Block diagram
VI Assembling
T-PFB
perfect focus module *
Filter slider
Stage
Stage-up kit
T-PFC
perfect focus
controller *
Two pillars *
On the opposite side
T-FLC2-E
motorized cassette
holder *
Objective
Remote control
pad
AC
adapter
Motorized
nosepiece
Stage-up kit
Power cord
(for the AC adapter)
Stage-up kit
AC
adapter
Illumination pillar
Epi-fl attachment
Auxiliary filter
slider
Epi-illumination
lamp house
Power supply
Power cord
(for the power supply)
To the power
source for
the episcopic
illumination
External fine focus
knob (optional)
Either connection will do.
Power cord
(for the AC adapter)
”*” denotes the component of this system.
29
VI Assembling
1 Checking Components
Check that the following components are prepared.
1. TE-2000E main body (including the hub controller and remote control pad)
2. TE-2000E stage-up kit
3. T-FLC2-E motorized cassette holder
4. T-PFB PF module (including two pillars)
5. T-PFC PF controller
6. Sensor cable
7. Control cable
8. Serial cable
9. AC adapter (for the PF controller)
10. IR-CUT filter (diameter 25 mm x 1 and 45 mm x 1)
2 Mounting the T-FLC2-E Motorized Cassette Holder
Insert the cassette holder into the attaching groove from over
1
the microscope.
Remove filter block cover. And fix the cassette holder with
2
screws on the left and right side by using the hexagonal
wrench.
Connect the cable of the cassette holder to the connector of the
3
FL block on the hub controller .
30
VI Assembling
3 Mounting the T-PFB PF Module
Mount the two pillars for the PF module onto the motorized
1
cassette holder.
Viewing from the microscope’s front, mount the pillar with a
smaller cut on the right side of the holder facing the cut to the
rear.
On the left side of the holder, mount the pillar with a larger cut
facing to the front.
Fix each pillar using two hexagon socket headed bolts.
Mating the groove to the protrusions of each pillar, place the
2
module on the pillars.
Retain the PF module on the pillars using four hexagon socket
3
headed bolts.
4 Mounting Stage-up Kits
Mount three stage pillars.
1
Mating the positioning protrusions of the pillars to the grooves
of the microscope’s main body and retain each of them using
two hexagon socket headed bolts.
Fit the nosepiece pillar and retain it with two hexagon socket
2
headed bolts.
5 Mounting the Motorized Nosepiece
Mount the main body of the motorized nosepiece on the
1
nosepiece pillar attached in step 4, “Mounting Stage-up Kits.”
1
2
1
1
Retain the nosepiece with two M5-hexagon socket headed
2
bolts.
31
VI Assembling
A
Connect the cable to the “Nosepiece connector” on the hub
3
controller.
Caution: Be sure to insert the objectives after mounting the stage
because they may be damaged if made contact.
6 Mounting the Stage
Place the stage on the stage pillars attached in step 4,
1
“Mounting Stage-up Kits.”
Retain the stage using four M5-hexagon socket headed bolts.
2
7 Mounting IR-cut Filters
This system uses the near infrared light (infrared band) derived from the infrared LED for the focus control. Since the
near infrared light (infrared band) used for the illumination system affects adversely on the focus control, therefore, place
the IR-cut filter on each optical path of the diascopic illumination and epi-fl illumination.
Caution: The Attached IR-cut filter reduces the intensity of ultraviolet rays too. Pay attention to observe with fluorescent
technique using the strong ultraviolet rays.
Attach the IR-cut filter (45 mm in diameter)
1
(1) Attach the filter to the filter slider.
(2) Insert the filter slider into the illumination pillar.
Attach the IR-cut filter (25 mm in diameter)
uxiliary filter slider
2
(1) Attach the filter to the auxiliary filter slider.
frame
IR-cut
filter
32
VI Assembling
(2) Insert the filter slider into the epi-fl attachment.
8 Connecting Cables
Locate the PF controller on the stable place near the microscope, and then, connect the following cables. Be sure to
check that the hub controller and PF controller are both not powered when connecting cables.
Rear view of the PF Controller
(1) DCIN connector
Connects to the DC
power supply from the
AC adapter.
(2) CONTROL (EXP)
connector
Connects to the EXP
connector on the hub
controller with the control
cable.
Cables
1. Control cable
(3) SERIAL (STAGE) connector
Connects to the STAGE
connector on the hub
controller with the serial
communication cable,
(5) AUX. IN connector
Connects the cable of the
T-EFN external fine focus
knob (option). (This
connection is not required
if the T-EFN external fine
focus knob is not used.)
(4) SENSOR connector
Connects to the SENSOR connector
of the PF module with the sensor
cable.
Round type plug: to the EXP connector of the hub
controller
Rectangular type plug: to the control (EXP)
connector of the PF controller
Caution: Be sure to connect the round type plug to the EXP connector of the hub controller. Do not plug to
the AUX.IN connector of the PF controller.
33
VI Assembling
2. Serial communication cable
3. Sensor cable
Rectangular type plug: to the STAGE connector of
the hub controller
Round type plug: to the SERIAL (STAGE)
connector of the PF controller
To the SENSOR connector of the PF controller
To the SENSOR connector of the PF module
Checking Connections
After connecting cables, power on the system and check to see each component.
Turn on the microscope.
1
Caution: When the PF controller is turned on, the PF module
will be initialized for about 10 seconds and the information
about the objectives on the nosepiece are acquired from the
HUB controller. So, before the PF controller, the microscope
must be turned on.
Besides, if the information of the objectives is changed, the PF
controller must be turned off once, and then, turned on again.
Turn on the [POWER] switch (switch to “I”) of the PF controller.
2
Check that the [POWER] LED is lit and the [OBJ.ERR] LED is
not lit.
34
V
II
If the system does not function properly, take appropriate action as described below. If the problem is still not resolved
after referring to this section, please contact your nearest Nikon representative.
Troubleshooting
1 During System Startup
Problems Causes Countermeasures
The buzzer issues beeps
twice in a row.
Both the [OFFSET_UPPER]
LED and the
[OFFSET_LOWER] LED are
lit at the same time.
The
[MICROSCOPE-OBJ.ERR]
LED is blinking.
• The motorized nosepiece is not
mounted on the microscope.
• The objective is not in the click stop
position.
• The SENSOR connector of the PF
module and the SENSOR connector of
the PF controller are not connected
with the SENSOR cable.
• The EXP connector of the hub
controller and the CONTROL (EXP)
connector of the PF controller are not
connected with the CONTROL cable.
• Mount the motorized nosepiece on the
microscope.
• Rotate the objective into the click stop
position.
• Connect the SENSOR connector of the
PF module and the SENSOR
connector of the PF controller with the
SENSOR cable.
• Connect the EXP connector of the hub
controller and the CONTROL (EXP)
connector of the PF controller with the
CONTROL cable.
2 While Using the Perfect Focus Functions
Problems Causes Countermeasures
The [FOCUS] LED is lit at
times while the [AF ON] LED
is not lit.
-
It is normal. The [FOCUS] LED is lit when
the boundary surface is detected. Use it
as a guide of focusing.
The
[MICROSCOPE-OBJ.ERR]
LED is lit.
The
[MICROSCOPE-OBJ.ERR]
LED is lit. And the buzzer
issues beeps twice in a row.
The
[MICROSCOPE-UPPER]
LED is lit.
The
[MICROSCOPE-LOWER]
LED is lit.
• The hub controller has no usable
objective.
• No usable objective is selected.
• The cable connection is wrong.
The objective is not in the click stop
position.
The objective reaches the upper limit of
the vertical control.
The objective reaches the lower limit of
the vertical control.
• Register a usable objective into the hub
controller.
• Select a usable objective. For usable
objectives, see the list of objectives in
the Specification.
• Connect cables correctly referring to
the Chapter VI, “Assembling.”
Rotate the objective into the click stop
position.
Lower the specimen position (lower the
stage).
Raise the specimen position (raise the
stage).
35
VII Troubleshooting
Problems Causes Countermeasures
The [OFFSET_UPPER] LED
is lit.
The [OFFSET_LOWER]
LED is lit.
Pressing the [FOCUS
OFFSET-MEMORY] key
causes nothing (no beep
from the buzzer).
Pressing the [AF-MEMORY]
key causes nothing (no beep
from the buzzer).
Pressing the [AF-RECALL]
key causes nothing (no
bleep from the buzzer).
Pressing the [AF-RECALL]
key causes beeps twice from
the buzzer and the
[AF-RECALL] does not work.
The [AF-RECALL] operation
ends after a long sound of
buzzer.
The offset amount reaches the upper limit
of the adjustable range.
The offset amount reaches the lower limit
of the adjustable range.
• The key was not pressed long enough.
• The [AF-RECALL] operation is in progress.
• Press the key longer than two seconds.
• The [AF-MEMORY] operation has not
been done.
• No usable objective is selected (the
[MICROSCOPE-OBJ.ERR] LED is lit).
The objective is close to the upper or
lower limit of the vertical control.
• The focus position could not be
detected.
• The objective reached the upper or
lower limit of the vertical control while
focusing.
Lower the objective with the [OFFSET]
knob until the LED is turned off.
Raise the objective with the [OFFSET]
knob until the LED is turned off.
• Press the key longer than two seconds.
• The [FOCUS OFFSET-MEMORY] key
is disabled during the [AF-RECALL]
operation.
• The [AF-MEMORY] key is disabled
during the [AF-RECALL] operation.
• Perform the [AF-MEMORY] operation
in advance.
• Select a usable objective.
Change the stage position so that its
vertical position is apart from the upper or
lower limit.
• Perform the [AF-MEMORY] operation
at the focus position again.
• Change the stage position so that its
vertical position is apart from the upper
or lower limit.
• The nosepiece is rotated during the
[AF-RECALL] operation. (The system
operation is stopped for safety if the
nosepiece is rotated during the
[AF-RECALL] operation.)
• The focusing signal is distorted during
the [AF-RECALL] operation. If one of
the following actions is taken, the
focusing signal may be distorted and
the system stops its operation because
of safety:
* Abrupt vertical position change is
done by manipulating the coarse
focus knob of the microscope.
* The specimen is inclined.
* The specimen is removed.
36
VII Troubleshooting
Problems Causes Countermeasures
Pressing the [AF ON] key
causes nothing (no bleep
from the buzzer).
The [DISABLE] LED is
blinking.
About 5 second after starting
the [AF ON] operation, the
buzzer beeps and the
focusing stops.
• No usable objective is selected (the
[MICROSCOPE-OBJ.ERR] LED is lit).
• No specimen is set.
• The objective is far from the boundary
surface of the specimen.
• The shape or material of the specimen
is not supported by this system.
• No specimen is set.
• The objective is far from the boundary
surface of the specimen.
• The shape or material of the specimen
is not supported by this system.
• Select a usable objective.
• Set a specimen and focus on it
manually.
• Adjust the objective closer to the
boundary surface of the glass by
manipulating the [OFFSET] knob.
• This perfect focus system may not
work precisely for specimens covered
by a thick glass plate or having a high
refractive index (n is nearly equal to
1.5). See p.7, 4.1 “Suitable specimen.”
• Set a specimen and focus on it
manually.
• Push the [OFFSET-RESET] key to
restore the offset amount to its original
status.
• This perfect focus system may not
work precisely if the specimen is
covered by a thick glass plate or the
aqueous solution has a high refractive
index (n is nearly equal to 1.5). See
p.7, 4.1 “Suitable specimen.”
37
V
III
Care and Maintenance
1 Lens Cleaning
Do not let dust, fingerprints, and so on get on the lenses. Dirt on the lenses, filters, and so on will adversely affect the
view of the image. If any of the lenses get dirty, clean them as described below:
• Brush away dust with a soft brush, or wipe it away gently with gauze.
• Only if there are fingerprints or grease on a lens, dampen a piece of soft, clean cotton cloth, lens tissue, or gauze with
absolute alcohol (ethyl or methyl alcohol) and wipe.
• Only when removing the immersion oil off from the objective, use petroleum benzine. If you cannot obtain petroleum
benzine, use methyl alcohol. However, because methyl alcohol does not clean as well as petroleum benzine, it will be
necessary to wipe the surfaces repeatedly. (Usually, wiping three or four times is sufficient to clean the lenses.) Wipe
with absolute alcohol (ethyl or methyl alcohol) to finish.
• Never use petroleum benzine to clean the entrance lens at the bottom of the eyepiece tube and prism surface of the
eyepiece tube.
• Absolute alcohol and petroleum benzine are highly flammable. Be careful when handling it, when around open flames,
when turning the power switch on / off, and so on.
• When using absolute alcohol, be sure to follow the instructions provided by the manufacturer.
2 Cleaning This Product
• Nikon recommends that you use a silicon cloth to clean this product.
• For persistent dirt, dampen a piece of gauze with neutral detergent and wipe lightly.
• Note that using organic solvents could result in discoloration of the plastic parts.
3 Disinfecting This Product
• Nikon recommends that you use 70% medical alcohol for normal disinfection of this product.
• If a specimen is spilled, check that the specimen is a safe material or not. In case of spillage of a hazardous
specimen onto this product, follow your standard laboratory procedures.
• Note that using organic solvents could result in discoloration of the plastic parts.
4 Storage
Store this product in a dry place where mold is not likely to form.
Store the objectives and eyepieces in a desiccator or similar container with a drying agent.
Put a vinyl cover over this product to protect them from dust.
Before putting on the vinyl cover, turn off the power switch on the microscope (flip it to the “O” side) and wait until the
lamphouse is cooled down.
5 Periodical Inspections (Expenses Charged)
Nikon recommends inspecting this product periodically (expenses charged) in order to maintain peak performance.
Contact your nearest Nikon representative for details.
38
IX
Microscope model TE2000-E
Target specimen Specimen in liquid (culture cell and so on)
Available container Glass bottom dish (glass thickness: No. 1S)
Glass thickness Standard No. 1S (150 to 180 µm)
Detection method Active type, infrared LED image projection method
Specifications
Detected
boundary
surface
Detector Inline CCD
Detector light source
Offset method Optical offset type: The offset value can be operated with the knob of the controller.
Offset amount
Focusing time 700 msec or less (near the boundary surface)
Focusing accuracy 1/3 of the depth of focus for each objective
Available microscopy
Available objective
Water or oil
immersion type
objective
Dry type objectiveBoundary surface between the glass and the air
Boundary surface between the glass and specimen (culture medium)
Refer to the figure on p.6 in Chapter I. “3. Offset function.”
Refer to the figure on p.6 in Chapter I. “3. Offset function.”
Infrared LED wavelength: 770 nm
illumination power: 0.1 mW or less
Dry type objective: +100 to –10 µm
Water immersion type objective: +50 to –2 µm
Oil immersion type objective: +10 to –0.5 µm
Note: Some objectives are available for greater offset amount than above.
Memory function 1 Offset value registering/calling (for each objective, up to 6 values)
Memory function 2
Dimensions and weight T-PFB:H 46 x W 210 x D 125 mm (without protrusions) 1.5 Kg
Service condition
Designated AC adapter
Nosepiece up/down position registering/calling (for each objective, up to 6
positions)
T-PFC:H 65 x W 150 x D 202 mm (without protrusions) 2.2 Kg
T-FLC2-E:H 56 x W 166 x D 165 mm (partially H 117) 1.3 Kg
Operating environment
Temperature: +20 to +38 °C (for control unit: 0 to +40 °C).
Relative humidity: 60% or less (for control unit: 80% or less) * No condensation.
Altitude: 2000m max.
Pollution degree: Degree 2
Installation category: Category II
Electric shock protection class: Class I (AC adapter)
Indoor use only
Storage environment
Temperature: - 20 to +60 °C
Relative humidity: 90% or less. * No condensation.
Manufacturer: ILAN Electronics Ltd.
Type: F1650K
Input ratings: 100 to 240 VAC, max 1.2 A, 50-60 Hz.
Output ratings: 12 VDC, max 3.5 A
Others: UL listed product, GS approved, and CE satisfied.
Power cord
Safety standards compliance
For using in the area supplied voltage is 100 to 120 V:
UL listed detachable cord set, 3 conductor grounding Type SVT, AWG 18, 3 m
long maximum, rated at 125 V AC minimum.
For using in the area supplied voltage is 220 to 240 V:
Approved cord set according to EU/EN standards, 3 conductor grounding type
HO5VV-F, AWG 18, 3 m long maximum, rated at 250 V AC minimum.
LED product category:
Class 1 LED Product
CE marking
This product meets EU Low Voltage Directive requirements.
This product meets EU EMC Directive requirements. (EN61326)
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference
when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if
not
installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference
in which case the user will be required to correct the interference at his own
expense.
This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la norme NMB-003 du
Canada.
This product meets Australian EMI. (AS/NZS CISPR11 Group 1 Class B)
40
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.