Nikon E200 User Manual

Page 1
En
411.4 (Eyepoint height when PD is 64mm.)
191.3
315
407.4 (Minimum microscope height)
414.6 (When PD is 64mm.)
105
188.5
247.7
206.4
226.8
101.5
381.7 (When PD is 64mm.)
Biological Microscope
Dimensional Diagram
Optical system CFI60 (infinity optical system)
Parfocal distance: 60mm
Magnification 40–1500X for observation
8–500X for 35mm photomicrography
Eyepiece tube E2-TB Binocular tube, E2-TF Trinocular tube
Siedentopf type (Inclination: 30°, Interpupillary distance: 47-75 mm, 360° rotatable) C-TE Ergonomic Binocular Tube (inclination: 10­30º, extension: 40mm)
Eyepiece CFI E 10X (F.O.V.: 20mm), CFI E 15X (F.O.V.: 12mm) Photo lens PLI projection lens: 2X, 2.5X, 4X, 5X Nosepiece Quadruple nosepiece, reversed type Coarse/fine Fine: 0.2mm per rotation, Coarse: 37.7mm per
focusing rotation, Minimum reading: 2 microns on left-side
fine control knob, Coarse motion torque adjustable, Refocusing system incorporated in stage, Stage handle and focusing knob are at equal distance from the operator
Stage Rectangular 216 x 150 mm surface stage mounted
on the main body. Cross travel 78 x 54 mm using low-positioned X/Y coaxial control knob
Objectives CFI E Plan Achromat 4X N.A. 0.10 (W.D. 30mm)
CFI E Plan Achromat 10X N.A. 0.25 (W.D. 0.7mm)
CFI E Plan Achromat 40X N.A. 0.65 (W.D.
0.65mm) CFI E Plan Achromat 100X Oil N.A. 1.25 (W.D.
0.23mm) As an option, CFI Achromat DL and other higher-grade CFI60 objectives can be used
Condenser E2 Abbe condenser N.A. 1.25; leaf-type
aperture diaphragm with position guide markings for respective CFI E Plan objectives Optional condensers: E2 Phase condenser N.A. 1.25; leaf-type aperture diaphragm with position guide markings for respective CFI Achromat DL objectives For microscope without field diaphragm: C-C Achromat Condenser Achromat Swing-out Condenser Darkfield Condenser, dry type C-C Phase Contrast Condenser
Illumination system 6V-20W halogen bulb, 6V-30W halogen bulb Intermediate E200 epi-fluorescence attachment, Teaching
attachment
* head (face-to-face and side-by-side types),
Drawing tube, Eye level riser
*Maximum intermediate space 50mm
E200 Specifications
Printed in Japan (0810-05)T Code No. 2CE-MTDH-8
This brochure is printed on recycled paper made from 40% used material.
Specifications and equipment are subject to change without any notice or obligation on the part of the manufacturer. October 2008 ©2000-08 NIKON CORPORATION
WARNING
TO ENSURE CORRECT USAGE, READ THE CORRESPONDING MANUALS CAREFULLY BEFORE USING YOUR EQUIPMENT.
NIKON CORPORATION
6-3, Nishiohi 1-chome, Shinagawa-ku, Tokyo 140-8601, Japan phone: +81-3-3773-8973 fax: +81-3-3773-8986
http://www.nikon.com/products/instruments/
NIKON INSTRUMENTS INC.
1300 Walt Whitman Road, Melville, N.Y. 11747-3064, U.S.A. phone: +1-631-547-8500; +1-800-52-NIKON (within the U.S.A. only) fax: +1-631-547-0306
http://www.nikoninstruments.com/
NIKON INSTRUMENTS EUROPE B.V.
Laan van Kronenburg 2, 1183 AS Amstelveen, The Netherlands phone: +31-20-44-96-222 fax: +31-20-44-96-298
http://www.nikoninstruments.eu/
NIKON INSTRUMENTS (SHANGHAI) CO., LTD.
CHINA phone: +86-21-5836-0050 fax: +86-21-5836-0030 (Beijing branch) phone: +86-10-5869-2255 fax: +86-10-5869-2277 (Guangzhou branch) phone: +86-20-3882-0552 fax: +86-20-3882-0580
NIKON SINGAPORE PTE LTD
SINGAPORE phone: +65-6559-3618 fax: +65-6559-3668
NIKON MALAYSIA SDN. BHD.
MALAYSIA phone: +60-3-7809-3688 fax: +60-3-7809-3633
NIKON INSTRUMENTS KOREA CO., LTD.
KOREA phone: +82-2-2186-8410 fax: +82-2-555-4415
NIKON CANADA INC.
CANADA phone: +1-905-602-9676 fax: +1-905-602-9953
NIKON FRANCE S.A.S.
FRANCE phone: +33-1-4516-45-16 fax: +33-1-4516-45-55
NIKON GMBH
GERMANY phone: +49-211-941-42-20 fax: +49-211-941-43-22
NIKON INSTRUMENTS S.p.A.
ITALY phone: +39-055-300-96-01 fax: +39-055-30-09-93
NIKON AG
SWITZERLAND phone: +41-43-277-28-67 fax: +41-43-277-28-61
NIKON UK LTD.
UNITED KINGDOM phone: +44-208-247-1717 fax: +44-208-541-4584
NIKON GMBH AUSTRIA
AUSTRIA phone: +43-1-972-6111-0 fax: +43-1-972-6111-40
NIKON BELUX
BELGIUM phone: +32-2-705-56-65 fax: +32-2-726-66-45
Page 2
The first microscope in its class to feature CFI60 infinity optics,
plus an anti-mold design to en sure years of high-quality use
F
or the first time, the CFI60 optical system—
infinity optics of the world's highest level—has
been incorporated into a microscope of this class.
The result is stunningly sharp and clear images at
any magnification. Of course, the Eclipse E200
maintains the same operational ease and rigidity
common to all Nikon microscopes and is compatible
with many accessories for higher-grade Eclipse series
microscopes.
Is mold a problem? Not any more.
By using anti-mold paint and sealing anti-mold agent
inside the microscope at critical places, the Eclipse
E200 is designed for use even in hot and humid
environments.
Moreover, this instrument packs all of these features
into a contemporary design that blends in with the
clean interiors of your hospital and examination
rooms.
Best of all, the Eclipse E200 is affordable. Once you
use it, you'll realize what an outstanding performer it
is, despite its modest price.
Eclipse E200-F (model with field diaphragm)
Eclipse E200
Page 3
Superb infinity optics, plus ergonomic touches
The CFI60 optical system combines Nikon's renowned CF optical design with infinity optics to overcome the limitations of the traditional infinity
design. CFI60 optics provide longer working distances and higher N.A.'s. These new optics deliver startlingly clear images at any magnification because chromatic aberrations and curvature of field are both corrected over the entire field of view when the field number is 20mm. Nikon developed new dedicated CFI E Plan Achromat objectives exclusively for the E200. Also, you can use other higher-grade objectives available for the Eclipse series whenever your laboratory situation calls for it.
Anti-mold design
Mold is a formidable enemy of microscopes. Before you know it, it can begin growing on the interior optical surfaces of the microscope and ruin performance. Using anti-mold paint, plus anti-mold agent sealed at critical places inside the microscope, the Eclipse E200 is designed to resist mold growth. In tests, an anti-mold treated unit was able to resist the growth of mold for three consecutive years at an average temperature of 30° C (86° F) and 80% humidity.
Ergonomic design equals comfortable operation
Comfort is ensured allowing long hours of use, thanks to Nikon’s thoughtful ergonomic design. This is the same design incorporated into Nikon’s other laboratory and research-grade Eclipse series microscopes. For example, the focus knob and the stage handle are located equidistant from the operator, permitting one-handed operation in a natural posture without twisting the shoulders. Because these controls are low positioned, you can operate the microscope while resting your arms comfortably on the desk. Moreover, the low-profile stage makes exchange of specimen slides easy, while the low inclination angle eyepiece tube provides comfortable viewing.
Ergonomic binocular tube
With this option, users can adjust not only the eyepiece tube tilt angle, but the eyepiece length to suit their build, eliminating discomfort and strain during long hours of observation.
* Use of this accessory in combination with other equipment may
produce darker images around the periphery.
* Must be attached directly to the main body.
4 5
One-piece construction from arm to base, a stage design where its up/down mechanism is located in the base, plus a wide footprint of 188.5mm across the back all provide
greater rigidity and resistance to vibrations, contributing to superior images.
optical system
Improved
Eye-level riser
Up to two eye-level risers can be mounted to raise the height of the eyepoint—25mm each for a total of 50mm.
Robust, vibration resistant construction
Comparison of the positions of the stage handle and focus control knob
View angle is adjustable by 10º to 30º.
Eyepiece length can be extended by up to 40mm.
E200 configured with an ergonomic binocular tube
Anti-mold
paint and agent
E200 Conventional models
Page 4
Design ideas that ensure superb opti cal quality and increased ease of use
6 7
Revolving nosepiece
A reversed-type nosepiece creates more space at the front of the stage, making handling of specimen slides fast and easy. In addition, the CFI60 optical design eliminates extra optical elements in the nosepiece for enhanced image sharpness. Another advantage of CFI60 objectives is that their increased objective lengths and longer working distances provide more working space around the nosepiece.
Refocusing stage
Nikon has created a unique innovation. The Refocusing Stage eliminates the need to refocus the image manually, making specimen handling safer and easier. In this unique design, the stage can be instantly dropped by pushing it down to exchange specimens or oil the slide, then returns to the original position as soon as the hand is removed. The wide stage surface can accommodate two slide glasses at the same time. In addition, this stage has an array of features including: – Increased resistance to vibrations due to the design of the in-
base focus mechanism.
– Low-profile design that creates more space around the objective
for increased freedom in specimen handling and easier operation.
– A belt-drive mechanism to eliminate the projection of the rack at
the edge of the stage for better ergonomics and smoother
movement. – Removable specimen holder for fast hand scanning of slides. – Improved XY cross travel, providing the comfort and feel similar
to Nikon’s higher grade Eclipse series microscopes.
Upper limit stopper
When using short-working-distance objectives such as 40X or greater, you can set the upper limit of the stage movement, so that the objective doesn’t hit the specimen slide, protecting both from damage. Thanks to this feature, even novices or operators who need to change slides often can perform their job easily and quickly. The limit height can be set in two levels using a stopper bolt—either at the standard position or 2mm lower. This feature is very useful except when extraordinarily thick specimens are used.
New
New
Eyepiece tube
The Siedentopf-type eyepiece tube is inclined at 30 degrees to ensure comfortable viewing in a natural posture. Designed for use by operators with different builds, this eyepiece tube has a narrow minimum interpupillary distance of 47mm, while the eyepoint height can be raised 34mm when interpupillary distance is 64mm by simply swinging the front part of the eyepiece tube up 180˚. For extremely tall users, eye­level risers are available to customize the microscope.
Eyepieces
The E200’s new eyepieces feature a wider field of view for a microscope of this class and are available in 10X (F.O.V. 20) and 15X (F.O.V. 12) types. These eyepieces also feature built-in diopter adjustment that allows the operator to adjust diopters separately for the right and left easily. In addition, these eyepieces accept measuring reticles that will always be in sharp focus with the specimen. Moreover, they can be locked, preventing theft and eliminating the possibility of damage during transit.
CFI60 objectives
Nikon’s exclusive CFI60 objectives provide numerous benefits: longer working distances, high numerical apertures, flat images over the entire field of view with virtually no curvature of field when the field number is 20mm. To match your laboratory requirements, the E200 provides a wide selection of objectives to choose from. These include the new CFI E Plan Achromat objectives developed for the E200 or other higher-grade Eclipse series objectives.
Improved
CFI E Plan Achromat objectives
Low position High position
E2-TB binocular tube E2-TF trinocular tube
* Eyepiece lens is optional.
F.O.V. 18 (conventional model)F.O.V. 20 (E200)
Wide stage surface with no rack sticking out of the stage
2mm
Cover glass
Slide glass
Upper limit (High)
Upper limit (Low)
Stage surface
Page 5
A wealth of accessories to broaden the range of microscopic applications
8 9
Condenser
Although the stage is low-positioned for comfort, there is ample space around the condenser for easy access. The condenser also features an aperture diaphragm that comes complete with position guide markings for respective E Plan objectives to make operation quick and easy.
Replacing the lamp is safe and easy
Turning the microscope upside down to replace the lamp is no longer necessary. Simply open the lens unit cover to make replacement.
Model with field diaphragm available.
A model with a built-in field diaphragm allows the use of Koehler illumination. It features: – A field lens unit with a field diaphragm
that has position-guide markings for respective objectives.
– Easy and safe lamp replacement
procedure.
Improved
Wide variety of condensers
Abbe, Phase, and other Eclipse series condensers, except the universal type, can be used with the E200.
Phase contrast
Simple phase contrast observation at 10X, 20X and 40X is possible with a single phase annulus slider. The aperture diaphragm automatically opens when the slider is inserted into the condenser. Phase contrast 100X slider and darkfield slider up to 40X objectives are available as options.
Epi-fluorescence
The E200-dedicated epi-fluorescence attachment is available for users who want to begin fluorescence observation. Although affordably priced, this attachment allows comprehensive epi-fluorescence as well as UV excitation observations.
Simple polarizing
This method is ideal for observing amyloid and crystals. To set up, install the polarizer over the field lens and the analyzer.
Field diaphragm with position-guide markings
E200-F
E2 Abbe condenser (left) and E2 phase condenser
Simple polarizing set C
CFI Achromat DL objectives for phase contrast
Filters are optional.
Page 6
System Diagram
10 11
Accessories (cont’d.)
Ergonomic binocular tube
Drawing tube
Allows accurate sketching of the image being observed.
Eye-level riser
Two risers can be inserted to increase the eyepoint height— 25mm per riser for a total of 50mm.
Teaching heads
Face-to-face* and side-by-side teaching heads are available.
* Not recommended for use in conjunction with a photomicrographic system, because it makes the microscope top-heavy.
Object marker
Allows the point of interest within a specimen to be marked with ink.
ND filter for objective lenses (ND3)
When this filter is used in combination with a 4x or 10x objective lens, image brightness can be adjusted to nearly equal to that of a 40x objective lens.
Digital Sight series digital camera system
When the trinocular eyepiece tube is used, a digital camera can be attached.* The DS-Fi1-L2, a color camera head combined with a standalone control unit, enables focusing and viewing of images on the built-in 8.4-in. LCD monitor without
the need for a PC. The optimum imaging conditions that match the observation technique in use can be automatically set by clicking on
the appropriate “scene mode.” Storing optimal images is easy.
*With a 4x objective lens, the image may show
uneven illumination depending on the camera used.
Plastic storage case
The rugged, lightweight storage case is handy for transporting and storing the E200.
Both eyepiece length and tilt angle are adjustable.
* Not recommended for use in conjunction with
the Y-THF Teaching Unit, because it makes the microscope top-heavy.
A
Y-IER Eye-level Riser
B
Intermediate Module
E2-FM Epi-Fl
C-SHG Mercury Illuminator 50W
Intermediate Modules
Attachment
A
Drawing Tube
H-III
Photomicrographic System*
Projection Lens PLI 2x, 2.5x, 4x, 5x
V-T Photo Adapter
Eyepiece TubesIntermediate Module Eyepieces
Digital Sight Series
Digital Camera
C-0.7x DXM
Relay Lens
Y-TV TV Tube
F
C-mount TV
Adapter A
C-mount
Adapter 0.55x
Y-TV 0.55x TV Tube
Digital Sight Series
F
Digital Camera
HH
A
D
E2-TB Binocular Tube E2-TF Trinocular Tube
A
A
Y-THS Teaching Unit B (Side-by-Side)
C-TE Ergonomic Binocular Tube
(100/0, 50/50 or 100/0, 0/100)
Y-THP Pointer
Y-THPS Teaching Head Support
H
C-TEP DSC Port
Y-THF Teaching Unit (Face-to-Face)*
CFI E10x
H
CFI E15x
A
*Not recommended for use in conjunction with a photomicrographic system, because it makes the microscope top-heavy.
C
E2 Simple
ABCD
Analyzer
ND Filter for Objective Lenses
CFI60 Objective Lens CFI60 Objective Lens
Dia. Polarizer
E
ABCD
E2 Simple
Analyzer
Dia. Polarizer
E
ND Filter for Objective Lenses
Condensers
E2 Phase Condenser
E
E2 Abbe Condenser
Blue Filter
Phase Contrast Slider 10x–40x
Phase Contrast Slider 100x
Darkfield Slider
Diffuser Slider
Loading...