We welcome your comments about this manual. Your comments and suggestions help us
improve our manuals. Please circle the number for each of the following statements
corresponding to your evaluation and add comments in the space provided.
Fax or send your completed comment card to:
Fax: +81 (3) 5996-8100
International Div., Sales Promotion Section, Nihon Kohden Corp., 1-31-4, Nishiochiai
Shinjuku-ku, Tokyo 161-8560, Japan
Strongly Agree 1Disagree4
Agree2Strongly Disagree 5
Neutral3
cutting line
This manual is organized.12345
I can find the information I want.12345
The information is accurate.12345
I can understand the instructions.12345
The illustrations are appropriate and helpful.12345
The manual length is appropriate.12345
Comments:
Thank you for your cooperation. We appreciate it very much.
Name:
Occupation/Position:
Hospital/Company:
Address:
Phone:
GENERAL HANDLING PRECAUTIONS
This device is intended for use only by qualified medical personnel.
Use only Nihon Kohden approved products with this device. Use of non-approved
products or in a non-approved manner may affect the performance specifications of
the device. This includes, but is not limited to, batteries, recording paper, pens,
extension cables, electrode leads, input boxes and AC power.
Please read these precautions thoroughly before attempting to operate the instrument.
1.To safely and effectively use the instrument, its operation must be fully understood.
2.When installing or storing the instrument, take the following precautions:
(1) Avoid moisture or contact with water, extreme atmospheric pressure, excessive
humidity and temperatures, poorly ventilated areas, and dust, saline or sulphuric air.
(2) Place the instrument on an even, level floor. Avoid vibration and mechanical shock,
even during transport.
(3) Avoid placing in an area where chemicals are stored or where there is danger of gas
leakage.
(4) The power line source to be applied to the instrument must correspond in frequency
and voltage to product specifications, and have sufficient current capacity.
(5) Choose a room where a proper grounding facility is available.
3.Before Operation
(1) Check that the instrument is in perfect operating order.
(2) Check that the instrument is grounded properly.
(3) Check that all cords are connected properly.
(4) Pay extra attention when the instrument is in combination with other instruments to
avoid misdiagnosis or other problems.
(5) All circuitry used for direct patient connection must be doubly checked.
(6) Check that battery level is acceptable and battery condition is good when using battery-
operated models.
4.During Operation
(1) Both the instrument and the patient must receive continual, careful attention.
(2) Turn power off or remove electrodes and/or transducers when necessary to assure the
patient’s safety.
(3) Avoid direct contact between the instrument housing and the patient.
5.To Shutdown After Use
(1) Turn power off with all controls returned to their original positions.
(2) Remove the cords gently; do not use force to remove them.
Operator's Manual ZS-940PAi
(3) Clean the instrument together with all accessories for their next use.
6.The instrument must receive expert, professional attention for maintenance and
repairs. When the instrument is not functioning properly, it should be clearly
marked to avoid operation while it is out of order.
7.The instrument must not be altered or modified in any way.
8.Maintenance and Inspection:
(1) The instrument and parts must undergo regular maintenance inspection at least every 6
months.
(2) If stored for extended periods without being used, make sure prior to operation that the
instrument is in perfect operating condition.
(3) Technical information such as parts list, descriptions, calibration instructions or other
information is available for qualified user technical personnel upon request from your
Nihon Kohden distributor.
9.When the instrument is used with an electrosurgical instrument, pay careful
attention to the application and/or location of electrodes and/or transducers to avoid
possible burn to the patient.
10. When the instrument is used with a defibrillator, make sure that the instrument is
protected against defibrillator discharge. If not, remove patient cables and/or
transducers from the instrument to avoid possible damage.
iiOperator's Manual ZS-940PA
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and
workmanship for one year from the date of delivery. However, consumable materials such as
recording paper, ink, stylus and battery are excluded from the warranty.
NKC or its authorized agents will repair or replace any products which prove to be defective
during the warranty period, provided these products are used as prescribed by the operating
instructions given in the operator’s and service manuals.
No other party is authorized to make any warranty or assume liability for NKC’s products.
NKC will not recognize any other warranty, either implied or in writing. In addition, service,
technical modification or any other product change performed by someone other than NKC or
its authorized agents without prior consent of NKC may be cause for voiding this warranty.
Defective products or parts must be returned to NKC or its authorized agents, along with an
explanation of the failure. Shipping costs must be pre-paid.
This warranty does not apply to products that have been modified, disassembled, reinstalled or
repaired without Nihon Kohden approval or which have been subjected to neglect or accident,
damage due to accident, fire, lightning, vandalism, water or other casualty, improper installation
or application, or on which the original identification marks have been removed.
In the USA and Canada other warranty policies may apply.
CAUTION
United States law restricts this device to sale by or on the order of a physician.
Equipment Authorization Requirement
Operation of this equipment requires the prior coordination with a frequency coordinator
designated by FCC for the Wireless Medical Telemetry Service.
Operator's Manual ZS-940PAiii
EMC RELATED CAUTION
This equipment and/or system complies with the International Standard
IEC60601-1-2 for electromagnetic compatibility for medical electrical equipment
and/or system. However, an electromagnetic environment that exceeds the
limits or levels stipulated in the IEC60601-1-2, can cause harmful interference to
the equipment and/or system or cause the equipment and/or system to fail to
perform its intended function or degrade its intended performance. Therefore,
during the operation of the equipment and/or system, if there is any undesired
deviation from its intended operational performance, you must avoid, identify
and resolve the adverse electromagnetic effect before continuing to use the
equipment and/or system.
The following describes some common interference sources and remedial
actions:
1.Strong electromagnetic interference from a nearby emitter source such as an
authorized radio station or cellular phone:
Install the equipment and/or system at another location if it is interfered with
by an emitter source such as an authorized radio station. Keep the emitter
source such as cellular phone away from the equipment and/or system.
2.Effect of direct or indirect electrostatic discharge:
Make sure all users and patients in contact with the equipment and/or system
are free from direct or indirect electrostatic energy before using it.
3.Electromagnetic interference with any radio wave receiver such as radio or
television:
If the equipment and/or system interferes with any radio wave receiver, locate
the equipment and/or system as far as possible from the radio wave receiver.
If the above suggested remedial actions do not solve the problem, consult your
Nihon Kohden Corporation subsidiary or distributor for additional suggestions.
ivOperator's Manual ZS-940PA
Conventions Used in this Manual and Instrument
Warnings, Cautions and Notes
Warnings, cautions and notes are used in this manual to alert or signal the reader to specific
information.
WARNING
A warning alerts the user to the possible injury or death associated with the use
or misuse of the instrument.
CAUTION
A caution alerts the user to possible injury or problems with the instrument
associated with its use or misuse such as instrument malfunction, instrument
failure, damage to the instrument, or damage to other property.
NOTE
A note provides specific information, in the form of recommendations,
prerequirements, alternative methods or supplemental information.
Operator's Manual ZS-940PAv
Explanations of the Symbols in this Manual and Instrument
The following symbols found in this manual/instrument bear the respective descriptions as
given.
On Panel
SymbolDescriptionSymbolDescription
Defibrillation proof type BF
applied part
Defibrillation proof type CF
applied part
Attention, consult operator’s
manual
Direction for attaching
battery cover
Inside Battery Case
Direct current
Serial number
Year of manufacture
CSA mark
SymbolDescriptionSymbolDescription
Battery position
On LCD
Attention, consult operator’s
manual
SymbolDescriptionSymbolDescription
Full battery
Replace battery
NIBP cannot be measured
Replace batteryCheck electrode
viOperator's Manual ZS-940PA
Introduction
The ZS-940PA transmitter transmits ECG, respiration, SpO2, NIBP and pulse waveform from a
patient to a Nihon Kohden monitor for continuous monitoring. The transmitter can change
channels when connected to the QI-901PK channel writer. The front LCD displays SpO
NIBP, pulse rate, pulse waveform amplitude, electrode condition mark, battery condition and
NIBP measuring mode and interval.
Read the operator’s manual for the receiving monitor together with this manual before operation.
CAUTION
• Do not use the same channel for different patients. Otherwise, two patients’
data will be lost due to mutual modulation interference, or another patient’s
data may appear on the receiving monitor screen.
• Do not use transmitters of adjacent channels in a hospital. Otherwise, radio
waves from one transmitter affect the receiver of the adjacent channel’s
transmitter and there may be interference.
NOTE
• To prevent interference between channels, assign a channel administrator in
the hospital and only he or she should manage channel assignment.
• Use Nihon Kohden parts and accessories to assure maximum performance
from your instrument.
• For stable signal reception, it is recommended to use a diversity antenna
system on the receiving monitor. Otherwise, spike noise from transient fading
of electric field strength (for example, people moving) may interfere with the
transmitter signal and may be mistaken as an arrhythmia on the receiving
monitor.
• NIBP cannot be measured on a neonate. (ECG, respiration and SpO
monitored on a neonate.)
can be
2
%,
2
Operator's Manual ZS-940PA1
Panel Description
Front Panel
LCD:
Displays measuring data, settings and
other information.
NIBP START/STOP key:
Starts/stops NIBP measurement in
selected mode.
NIBP INTERVAL key:
Selects NIBP measurement mode.
Refer to the WARNING below.
(This symbol is attached to the rear of the
battery case cover.)
Battery case:
Contains three 1.5 V AA batteries.
WARNING
Close the battery case cover during operation.
If the transmitter is used with the battery case cover open, the patient may get
an electrical shock when defibrillation is performed, and electrostatic discharge
by the patient may intermittently interfere with the waveform or data.
2Operator's Manual ZS-940PA
Rear Panel
Lock plate:
Fastens the transmitter to an NIBP cuff.
Refer to the WARNING below.
WARNING
If detergent or liquid spills into the transmitter, stop using it and contact your
Nihon Kohden distributor. If a wet transmitter is used, the patient or anyone in
contact with the transmitter may receive an electric shock or patient leakage
current over the allowed amount may flow.
Operator's Manual ZS-940PA3
Top Panel
NIBP socket:
Connects the cuff hose.
Refer to the WARNING on the
next page.
Refer to the WARNING on the
next page.
L
R
F
Refer to the symbol
page.
Refer to the symbol page.
Bottom Panel
Channel number label:
Indicates the channel number of the
transmitter. Attach the channel number label
to the panel of the monitor.
Refer to the WARNING on the next
page.
ECG/impedance RESP socket:
Connects the electrode lead for
measuring ECG and/or respiration
by the impedance method.
Refer to the symbol page.
SpO
socket:
2
Connects the SpO
2
probe.
4Operator's Manual ZS-940PA
WARNING
• Before performing defibrillation, check that the electrode leads and SpO
probe attached to the patient are properly connected to the transmitter.
Touching the metal parts of disconnected leads and probes causes serious
electrical shock or injury by discharged energy.
• When performing defibrillation, all persons must keep clear of the bed and
must not touch the patient, any equipment connected to the patient or the
metal parts of leads and probes connected to the patient. Failure to follow this
warning may result in serious electrical burn, shock or other injury.
• When performing defibrillation, discharge as far as possible from electrodes
and medicine on the chest of the patient. If there is a possibility that the
defibrillator paddle could touch electrodes and medicine, remove electrodes
and medicine from the patient. If the defibrillator contacts these materials, the
discharged energy may cause serious electrical burn on the patient.
• When using this transmitter with an ESU, the ESU return plate and the
electrodes for monitoring must be firmly attached to the patient. If the return
plate is not attached correctly, it may burn the patient’s skin where the
electrodes are attached. Refer to the instruction manual for the ESU.
2
CAUTION
Do not shake or swing the transmitter holding the leads/cables connected to the
transmitter. The transmitter may come off and injure someone or damage
surrounding instruments.
“CUFF” is displayed with the cuff inflation pressure during
measurement.
4 Check electrode markAppears when an electrode or electrode lead becomes
detached during ECG measurement.
5 Battery replacement markAppears when the batteries are weak. For details, refer to
the “Battery Condition Indication” section.
6 Message display areaDisplays messages.
7 NIBP measurement modeDisplays NIBP measurement mode. When set to auto mode,
the measurement interval is displayed.
8 NIBP interval bar graphIn auto NIBP measurement, the remaining time from the last
measurement to the next measurement is displayed as a bar
graph.
9 Pulse level bar graphDisplays pulse level in 7 steps.
10 %SpO
2
11 PRDisplays pulse rate when NIBP or SpO
Displays SpO2 data.
the SpO
probe is attached to the patient, the real time pulse
2
rate is displayed. When the SpO
patient, the pulse rate at the end of NIBP measurement is
displayed.
7
is measured. When
2
probe is not attached to the
2
6Operator's Manual ZS-940PA
Notes on Parameter Settings
When monitoring NIBP and SpO2, the following settings must be set as indicated in the table to
properly transmit the monitoring data to the receiving monitor. Otherwise, data cannot be
displayed on the receiving monitor even when NIBP and SpO
Some monitors require the software to be upgraded. For details, contact your Nihon Kohden
distributor.
TYPE settingINHIBIT SpO2 DURING NIBP settingSpO2 probe attachment site
ON
A*
1
OFF
*1Always set to “A”.
2
*
When the SpO2 probe is attached to the same limb as the NIBP cuff and the cuff is inflated,
the SpO
value becomes unstable and SpO2 or PR alarm may occur.
2
are measured on the transmitter.
2
Probe can be attached to the
same limb as the cuff
Probe must be attached to the
limb without cuff*
2
Operator's Manual ZS-940PA7
Important Safety Information
General
WARNING
• Never use this transmitter in the presence of any flammable anesthetic gas or
high concentration oxygen atmosphere. Failure to follow this warning may
cause explosion or fire.
• Never use this transmitter in a high-pressure oxygen medical care tank.
Failure to follow this warning may cause explosion or fire.
• Do not take this transmitter into the MRI test room. This transmitter is not
designed to be used during MRI tests.
• If detergent or liquid spills into the transmitter, stop using it and contact your
Nihon Kohden distributor. If a wet transmitter is used, the patient or anyone in
contact with the transmitter may receive an electric shock or patient leakage
current over the allowed amount may flow.
• Before cleaning or disinfecting the transmitter, remove the batteries from the
transmitter.
• The transmitter cannot be sterilized.
Output Signal
WARNING
Do not use the output signal from the receiving monitor as the synchronization
signal for other equipment such as IABP, MRI, echocardiography or defibrillator
because there may be time delay between the monitor and the other equipment
caused by waveform transmission delay and spike noise may interfere on the
output signal and be mistaken as a trigger.
8Operator's Manual ZS-940PA
Preparation
Installing (Replacing) Batteries
WARNING and CAUTION for Battery Handling
WARNING
• Do not dispose of the battery in fire because it may explode.
• Do not use a disassembled or damaged battery. The contents of the battery
are harmful and the battery may catch fire.
• If the contents of the battery contact the skin or clothes, immediately wash it
thoroughly with water.
• Never short-circuit the + and – terminals. The battery may overheat and catch
fire.
• Take care that the patient does not swallow batteries.
CAUTION
• Refer to the battery and battery charger manuals for details on handling the
batteries.
• Do not handle the batteries with wet hands.
• When the transmitter is not in use, remove batteries. When batteries are
installed, battery power is consumed even if measurement is not performed.
Especially, when NiMH batteries remain in the transmitter when the
transmitter is not in use, the battery may become unusable from
overdischarge and leak liquid which will damage the transmitter.
• The battery charger must be used outside the patient environment.
Battery Lifetime
Use three AA type alkaline dry cell batteries. NiMH rechargeable batteries can also be used.
TypeLifetime (Measuring parameters)
ECG, SpO2, NIBPECG, SpO
2
ECG only
NiMH secondary2 days2 days2.5 days
Alkaline primary1 day2.5 days3 days
The above data is when new batteries are used at room temperature, NIBP is measured in auto
mode at 60 minute intervals and SpO
is measured on an index finger of a male patient with
2
weight 60 kg.
Operation time depends on the thickness of SpO
Operator's Manual ZS-940PA9
probe attachment site.
2
NOTE
When using rechargeable NiMH batteries, shallow charging/discharging
shortens battery capacity. For details, refer to the battery operator’s manual.
Installing (Replacing) Batteries
CAUTION
Battery replacement must be performed by medical staff. When replacing
batteries of the transmitter currently used for a patient, disconnect electrode
leads from the transmitter before replacing batteries or do not touch the patient
during replacement.
If electrode leads are attached to the patient and a person replacing batteries touches the patient
during battery replacement, patient leakage current over the allowed amount may flow.
CAUTION
• Replace all batteries at the same time.
• Do not use different types of batteries together.
NOTE
Insert the batteries with the correct polarity (+ and –).
Procedure
1.Remove the battery case cover.
2.Insert three new or fully charged batteries into the
battery case observing the correct polarity.
10Operator's Manual ZS-940PA
3.Close the cover.
NOTE
Remove the batteries before disposing of the
transmitter.
Situations Requiring Battery Replacement
Replace the batteries when any of the following occurs.
• The transmitter LCD displays the “
• The transmitter generates a constant alarm (continuous “peep” sound).
• The transmitter LCD does not display anything when the power is turned on.
• The monitor displays the battery replacement message on the screen.
Battery Condition Indication
The battery condition is indicated as follows.
IndicationConditionReceiving Monitor
Fully charged battery
Batteries are low. Replace
batteries.
Batteries are low. NIBP cannot
be measured. Replace batteries.
No
indication
Dead batteries
” or “”mark.
Batteries are full. There is no indication on
the monitor.
Message requiring battery replacement is
displayed.
No signal can be transmitted to the monitor.
There is no indication on the monitor.
Operator's Manual ZS-940PA11
Turning the Transmitter On/Off
Turning On the Power
Peep
Check Items Before Use
Before turning on the transmitter power, check the following to confirm that the transmitter can
be used in normal and safe condition.
When the batteries are installed correctly, the power is
turned on. A one second “peep” sounds and the startup
screen appears. (There is no “peep” sound when there is
no battery power.)
NOTE
Replace the batteries when the LCD displays
the “
Turning Off the Power
To turn off the power, remove batteries.
” or “”mark.
Appearance
• There are no damaged or dirty parts on the outside of the transmitter (Power switch, LCD,
keys, sockets, battery case cover, battery case, lock plate, etc.).
• The transmitter is completely dry.
• The electrode lead, SpO
• There are no damaged or dirty parts on the disposable SpO
probe and NIBP cuff are not broken.
2
probe, disposable electrodes or
2
disposable NIBP cuff.
Batteries
• The battery polarity is correct.
• The battery case spring is firmly fixed and the battery is not loose.
• The battery case cover is firmly closed.
Channel Setting
• The transmitter channel corresponds to those of the receiving monitor.
• There is no transmitter in the surrounding area with the same channel.
12Operator's Manual ZS-940PA
Check Items After Power On
After turning on the power, check the following.
Power On
• The Power switch is not broken.
• The transmitter generates about a one second “peep” sound and the startup screen appears.
• The transmitter does not generate a continuous “peep” sound.
• The transmitter does not give excessive heat.
• The transmitter LCD displays a “
” mark.
• The transmitter does not interfere with the operation of other medical instruments in use.
Basic Operation
• The “signal loss” message is not displayed on the receiving monitor when the transmitter is
inside the receiving range of the monitor.
• The battery replacement message is not displayed on the monitor.
• The keys on the transmitter function properly.
Check Items After Use
To use the transmitter in safe and optimum condition for next time, check the following.
Before Turning Power Off
• Temporarily changed settings are changed back to the previous settings.
• There was no malfunction on the transmitter.
Storage
• ECG electrode leads, SpO
probe and NIBP cuff are cleaned and disinfected.
2
• When the transmitter gets wet, liquid is wiped off and the transmitter is thoroughly dried.
• There are enough consumables, such as disposable electrodes.
• The transmitter power is turned off.
• The batteries are removed from the transmitter when it will not be used for a long time.
• Dead batteries are disposed of properly.
Operator's Manual ZS-940PA13
Changing the Transmitter Channel
The channel of the transmitter can be changed. The optional QI-901PK Channel Writer is
required.
WARNING
The following action must be taken to properly receive the transmitter signal of
the correct patient on the receiving instrument. Otherwise, there may be signal
loss or signals may mix causing a serious accident, such as monitoring a
different patient.
• Assign a channel administrator in the hospital and only he or she should
manage channel assignment on his or her responsibility.
• The channel administrator must manage the channels in the facility so that
there is no signal interference.
• When the transmitter channel is changed, the channel administrator must
check that the channel on the receiving monitor is also changed and the signal
is properly received.
• The channel administrator must replace the channel number label on the
transmitter with the new one after changing the channel.
NOTE
The software version of the QI-901PK channel writer must be 02-01 or later to
change the channel on the ZS-940PA transmitter.
To check the transmitter channel, refer to “CHANNEL/TYPE” in the “Changing System Setup
Settings” section.
14Operator's Manual ZS-940PA
Changing Parameter Setup Settings
The initial settings on the PARAMETER SETUP screen must be changed before monitoring.
Changing these settings during monitoring interrupts monitoring.
Parameter Setup Setting List
The factory default settings are underlined.
Setting ItemDescriptionSettings
SELECTABLE
INTERVALS
INITIAL
INTERVAL
INITIAL CUFF
PRESS
NIBP MODE
AFTER STAT
START/FINISH
SOUND
OLD NIBP DATA
AFTER
INHIBIT SpO
DURING NIBP
2ND PARAMETERSet SpO2 and PR display order.SpO2, PR
2
Select the NIBP measurement modes for
the mode selection.
Select the initial NIBP measurement
mode at power on.
Select the NIBP cuff inflation pressure.
Select the NIBP measurement mode
after completing STAT measurement.
Turn ON or OFF the sound for NIBP
measurement start/finish.
Select whether to hide or dim the NIBP
data after measurement and how long to
wait after measurement to dim or hide it.
2.While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery.
The SETUP screen appears.
Operator's Manual ZS-940PA15
3.Press the NIBP START/STOP key to enter the PARAMETER SETUP screen.
When the cursor is moved to “EXIT” by pressing the NIBP INTERVAL key and the NIBP
START/STOP key is pressed, the startup screen appears, then the monitoring screen
appears.
4.To select or change a setting, press the NIBP START/STOP key.
To move the cursor, press the NIBP INTERVAL key.
Selects or changes setting
Moves cursor
When the cursor is moved to “RETURN” by pressing the NIBP INTERVAL key and the
NIBP START/STOP key is pressed, the SETUP screen appears.
Changing Settings
SELECTABLE INTERVALS
During monitoring, when the NIBP INTERVAL key is pressed, the measurement mode changes
according to the modes selected in this item. MANUAL mode is already selected for the mode
selection.
1.Press the NIBP INTERVAL key to move the
cursor to the desired mode.
2.Press the NIBP START/STOP key to select or
unselect the mode. Selectable modes are: STAT,
5, 10, 15, 30, 60, 120 and 240 min.
16Operator's Manual ZS-940PA
INITIAL INTERVAL
Select the initial NIBP measurement mode at power on.
1.Press the NIBP INTERVAL key to move the
cursor to “INITIAL INTERVAL”.
2.Press the NIBP START/STOP key to select
the mode. Selectable modes are the modes
selected for “SELECTABLE INTERVALS”
and “STAT” and “MAN.” (MANUAL).
INITIAL CUFF PRESS
Select the NIBP cuff inflation pressure.
1.Press the NIBP INTERVAL key to move the
cursor to “INITIAL CUFF PRESS”.
2.Press the NIBP START/STOP key to select
the inflation pressure from 120, 150, 180, 210
and 240 mmHg.
NIBP MODE AFTER STAT
Select the NIBP measurement mode after completing the STAT measurement.
1.Press the NIBP INTERVAL key to move the
cursor to “NIBP MODE AFTER STAT”.
2.Press the NIBP START/STOP key to select
the mode. The selected mode is automatically
selected for “SELECTABLE INTERVALS”
as well.
Operator's Manual ZS-940PA17
START/FINISH SOUND
Turn on or off the sound for NIBP measurement start and finish.
Start sound
End sound
1.Press the NIBP INTERVAL key to move the
cursor to “START/FINISH SOUND”.
2.Press the NIBP START/STOP key to turn ON
or OFF.
OLD NIBP DATA/AFTER
Select whether to dim or hide the NIBP data after measurement and how long to wait after NIBP
measurement to dim or hide it.
1.Press the NIBP INTERVAL key to move the
cursor to “OLD NIBP DATA/AFTER”.
2.Press the NIBP START/STOP key to select the
setting.
DATA:DIMNIBP data is dimmed after the “AFTER” interval.
HIDE NIBP data is hidden after the “AFTER” interval. “– – –” is displayed on the
screen.
AFTER:Select the interval after NIBP measurement to dim or hide.
DimmedHidden
18Operator's Manual ZS-940PA
INHIBIT SpO2 DURING NIBP
Set whether or not to monitor SpO
during NIBP measurement.
2
When the SpO
OFF, the pulse may become unstable and SpO
set this setting to ON so that SpO
When the SpO
probe is attached to the same limb as the NIBP cuff and this setting is set to
2
or PR alarm may occur. It is recommended to
2
is not measured during NIBP measurement.
2
probe is attached to the other limb from the NIBP cuff, this setting can be set to
2
OFF.
NOTE
• When the “CHANNEL/TYPE” on the SYSTEM SETUP screen is set to “8”, this
“INHIBIT SpO
during NIBP measurement.
• When this “INHIBIT SpO
SpO
during NIBP Measurement” section.
2
ON:Stops SpO
OFF:SpO
2ND PARAMETER
Set the display order of SpO
DURING NIBP” setting becomes invalid and SpO2 is monitored
2
DURING NIBP” is set to OFF, refer to the “Monitoring
2
1.Press the NIBP INTERVAL key to move the
cursor to “INHIBIT SpO
DURING NIBP”.
2
2.Press the NIBP START/STOP key to select
“ON” or “OFF”.
monitoring during NIBP measurement.
2
is monitored during NIBP measurement.
2
and PR.
2
1.Press the NIBP INTERVAL key to move the
cursor to “2ND PARAMETER”.
2.Press the NIBP START/STOP key to select
“SpO2” or “PR”.
When set to SpO
Operator's Manual ZS-940PA19
2
When set to PR
Changing System Setup Settings
NOTE
Changing System Setup settings must be done only by a qualified personnel.
System Setup Setting List
The factory default settings are underlined.
Setting ItemDescriptionSettings
CHANNEL/TYPE
PRESSURE UNITSelect the units for NIBP.mmHg, kPa
LANGUAGESelect the language for screen display.JPN, ENG
SYSTEM
INITIALIZE
Displays the transmitter channel and select the
TYPE.
Initializes all settings to the factory default
settings.
Displaying the SYSTEM SETUP Screen
1.Remove one battery.
2.While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery.
The SETUP screen appears.
A, 8
3.Press the NIBP INTERVAL key to move the cursor to “SYSTEM SETUP”.
4.Press the NIBP START/STOP key to enter the
SYSTEM SETUP screen.
When the cursor is moved to “EXIT” by
pressing the NIBP INTERVAL key and the
NIBP START/STOP key is pressed, the startup
screen appears, then the monitoring screen
appears.
20Operator's Manual ZS-940PA
5.To select or change a setting, press the NIBP START/STOP key.
To move the cursor, press the NIBP INTERVAL key.
Selects or changes setting
Moves cursor
When the cursor is moved to “RETURN” by pressing the NIBP INTERVAL key and the
NIBP START/STOP key is pressed, the SETUP screen appears.
Changing Settings
CHANNEL/TYPE
Select “A” for TYPE. The channel of this transmitter is also displayed.
1.On the SYSTEM SETUP screen, press the NIBP START/STOP key to select “A”.
Channel of this transmitter
NOTE
Always select “A” for TYPE. If “8” is selected,
NIBP data cannot be properly transmitted to
the receiving monitor.
PRESSURE UNIT
Select the unit for NIBP.
1.Press the NIBP INTERVAL key to move the
cursor to “PRESSURE UNIT”.
2.Press the NIBP START/STOP key to select
“mmHg” or “kPa”.
Operator's Manual ZS-940PA21
LANGUAGE
Select the language for screen display.
1.Press the NIBP INTERVAL key to move the
cursor to “LANGUAGE”.
2.Press the NIBP START/STOP key to select the
language.
SYSTEM INITIALIZE
Do the following procedure to initialize the settings to the factory default settings.
1.Press the NIBP INTERVAL key to move the cursor to “SYSTEM INITIALIZE”.
2.Press the NIBP START/STOP key. The “EXECUTE” message appears.
3.Press the NIBP START/STOP key to initialize the settings to the factory default settings.
22Operator's Manual ZS-940PA
Attaching NIBP Cuff, Electrodes and SpO2 Probe
to the Patient
The transmitter can be attached to an arm of the patient or placed on the bedside. The required
length of the electrode leads and SpO
attached to the patient.
Monitoring SpO2 during NIBP Measurement
When the SpO
probe is attached to the same limb as the NIBP cuff, the blood
2
flow decreases during NIBP measurement and pulse wave cannot be detected
and SpO
cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on
2
the PARAMETER SETUP screen is set to ON (factory default setting), SpO
monitoring is paused during NIBP measurement to avoid SpO2 alarm
occurrence. However, when monitoring SpO
careful when reading SpO
probe cable depends on how the transmitter is to be
2
NOTE
on the same limb as NIBP, be
2
values.
2
2
When monitoring SpO
is important, attach the probe to the limb to which the
2
NIBP cuff or catheter is not attached.
Attachment Examples
When transmitter is attached on an
arm
When placing the transmitter on a bedside,
place it on a stable and flat place. If the
transmitter falls off, it may be damaged.
Operator's Manual ZS-940PA23
When transmitter is placed on a bedside
NOTE
Attaching the NIBP Cuff
Selecting the NIBP Cuff
Select the NIBP cuff appropriate for the patient.
NOTE
NIBP cannot be measured on neonates using this transmitter.
Reusable Cuffs
When attaching the transmitter to the patient arm, a special NIBP cuff is required. An optional
YN-990P extension hose (1.5 m) is available to extend the length between the NIBP socket on
the transmitter and NIBP cuff (e.g. when not attaching the transmitter to the patient arm and
placing the transmitter on a bedside).
Reusable cuffModelWidth (cm)Air hose length (cm)
For adult
StandardYP-943P1315
LargeYP-944P1515
Air hose
Width
When not attaching the transmitter to the patient arm, the following cuffs can be used. To use
these cuffs, an optional YN-990P extension hose (1.5 m) is required.
Reusable cuffModelWidth (cm)Air hose length (cm)
For infantsYP-960T5
For
children
For adults
24Operator's Manual ZS-940PA
SmallYP-961T7
StandardYP-962T10
StandardYP-963T13
LargeYP-964T15
Width
Air hose
15
Disposable Cuffs
CAUTION
The disposable cuffs are not sterilized. If necessary, sterilize the disposable
cuffs using glutaraldehyde solution by following the instructions for the
glutaraldehyde.
When not attaching the transmitter to the patient arm, the following disposable cuffs can be used.
To use these cuffs, an optional YN-990P extension hose (1.5 m) is required.
Reusable cuffModelWidth (cm)Air hose length (cm)
For infantsYP-910P6
For childrenYP-912P9
For adults
SmallYP-913P12
StandardYP-914P14
LargeYP-915P16
Width
Air hose
20
Extension Hose
CAUTION
When using an extension hose, check that the extension hose is not bent or
squeezed. Otherwise, the cuff may not inflate or deflate. If the cuff cannot
deflate, it may cause congestion on the patient at the cuff attachment site.
YN-990P extension hose, 150 cm
Operator's Manual ZS-940PA25
Reference for selecting a cuff
The AHA (American Heart Association) recommends that the cuff width be 40% of the
circumference of the upper arm. Refer to the following graph and select the cuff which suits the
patient’s arm.
NOTE
• If a range of arm circumference appropriate for the cuff is prescribed, use a
cuff within that range.
• To obtain accurate measured values, select a wide cuff which can be attached
to the upper arm. Measuring with a very narrow cuff may result in measured
values higher than the actual values.
• The YP-943P NIBP cuff is for standard size adult. Do not use this cuff when it
does not fit the patient.
Cuff Width and Arm Circumference
Reusable Cuffs
Cuff width (cm)
15
10
Children small YP-961T
5
Infants YP-960T
Adults large YP-944P (cuff for transmitter)
YP-964T
Adults standard YP-943P (cuff for transmitter)
YP-963T
Children standard YP-962T
Disposable Cuffs
Cuff width (cm)
0
20
15
10
5
0
102030
Children standard YP-912P
Infants YP-910P
102030
40
Adults large YP-915P
Adults standard YP-914P
Adults small YP-913P
40
50
Arm circumference (cm)
5060
Arm circumference (cm)
60
26Operator's Manual ZS-940PA
Connecting the NIBP Cuff to the Transmitter
When Using YP-943P/944P NIBP Cuff
To attach the YP-943P/944P NIBP cuff to the transmitter, the lock plate is required.
YP-943P/944P NIBP cuff
Lock plate
Front cover
Belt for the strap
Lock plate pocket
Top tab
For attaching the NIBP cuff to the transmitter
Air hose
Front cover open
NOTE
Do not roll up or put weight on the cuff with the lock
plate attached to it. Otherwise, the lock plate may
break.
D ring
Belt
Bottom tab
Operator's Manual ZS-940PA27
1.Remove the lock plate from the
transmitter.
2.Insert the lock plate into the lock plate
pocket on the NIBP cuff.
3.Attach the transmitter to the lock plate
by inserting the tabs on the lock plate
into the slots on the transmitter.
4.Cover the transmitter with the front
cover of the NIBP cuff.
5.Connect the air hose to the NIBP socket
on the transmitter. Turn the cuff
connector joint until it clicks.
2
1
28Operator's Manual ZS-940PA
When Using YP-960T series or YP-910P series NIBP Cuff
To use these NIBP cuffs, an optional YN-990P extension hose (1.5 m) is required.
NOTE
Connect the joints properly. If there is an air leak, NIBP cannot be measured
properly.
1.Connect the NIBP cuff to the extension hose.
2.Connect the other end of the extension hose to
the NIBP socket on the transmitter. Turn the
joint clockwise until it clicks.
To disconnect the cuff from the transmitter,
turn the hose joint counterclockwise.
Attaching the NIBP Cuff to the Patient
WARNING
Be careful when measuring NIBP on a patient with known bleeding disorders or
coagulation. After NIBP measurement, there may be dot hemorrhage, or
circulatory disorder by thrombus where the cuff was attached.
CAUTION
• Do not wrap the cuff on an arm or thigh which is used for injection. NIBP
measurement on an arm or thigh which is used for injection may cause a
reflux of blood and stop injection.
• Do not wrap the cuff too tight. It may cause poor blood circulation and
congestion. If the cuff is wrapped too loosely, the NIBP value may be
increased.
• If the skin gets irritated or redness appears on the skin from the cuff, change
the attachment site or stop using the cuff.
Operator's Manual ZS-940PA29
• NIBP and SpO2 can be measured on the same limb, but the SpO2 monitoring
may not be accurate during NIBP measurement. Be careful when reading the
SpO
values.*
2
• Do not reuse disposable cuffs.
* Monitoring SpO2 during NIBP Measurement
When the SpO
during NIBP measurement and pulse wave cannot be detected and SpO
properly. When “INHIBIT SpO
set to ON (factory default setting), SpO
avoid SpO
NIBP, be careful when reading SpO
probe is attached to the same limb as the NIBP cuff, the blood flow decreases
2
cannot be monitored
2
DURING NIBP” on the PARAMETER SETUP screen is
2
monitoring is paused during NIBP measurement to
2
alarm occurrence. However, when monitoring SpO2 on the same limb as the
2
values.
2
NOTE
• Measuring NIBP at a site other than the upper arm gives different values from
those measured at the upper arm. When making diagnosis based on the NIBP
values, measure NIBP on an upper arm.
• To accurately detect the pulsatile flow of the artery, the cuff should be
wrapped around a bare upper arm.
• Do not use an abnormal cuff. The cuff deteriorates from use and cleaning.
Before use, check the cuff and confirm that there is no flaw, crack or hole in it.
Be careful not to damage the inflation bag. If the inflation bag has a hole or a
flaw, it may burst during use. Dispose of an abnormal cuff and replace it with
a new one.
• Refer to the NIBP cuff manual for details.
Cuff Position
Place the cuffed upper arm (brachium) at the same height as
When placing the
transmitter on a bed, make
sure that the hose is not
bent.
the patient’s heart. If the cuff is not at the same level as the
heart, the weight of the blood affects the blood pressure
reading. The pressure difference per unit height is 0.7
mmHg/cm. The blood pressure reading decreases when the
arm is higher than the heart and increases when lower.
Heart
The best measuring condition is when the patient is lying on
his/her back with arms and legs relaxed. If the cuff position
cannot be on the same level as the heart, the displayed blood
pressure reading must be mathematically adjusted.
30Operator's Manual ZS-940PA
Using the YP-943P/944P NIBP Cuff
Belt
1.Attach the NIBP cuff to the transmitter. Refer to
the “Connecting the NIBP Cuff to the
Transmitter” section.
2.Insert the end of the cuff into the belt and then
through the D ring as shown at left.
D ring
End of cuff
3.Fold back the cuff at the D ring and fasten it
using the velcro tape.
Make sure that the cuff is not attached on a joint.
NOTE
The cuff must not wrap around the elbow.
Operator's Manual ZS-940PA31
Attaching the Strap to the Transmitter
NOTE
• Use the strap to prevent the transmitter from falling.
• Do not attach the clip to hard objects such as thick cloth or zipper. It will break
the clip.
Attach a strap provided with the transmitter to the NIBP cuff and patient clothes.
To open the clip, firmly
pull out the tab in
direction of the arrow.
To adjust the strap length,
push down the tab on the
adjuster and slide.
Belt for the strap on the NIBP cuff
1.Adjust the length of the strap.
2.Clip one end of the strap to the belt for the strap
on the NIBP cuff.
3.Clip the other end of the strap to the patient’s
clothes as shown left.
32Operator's Manual ZS-940PA
Using the YP-960T series Reusable Cuffs or YP-910P series Disposable Cuffs
ARTERY
1.Put the cuff on the upper arm so that the
mark of “ARTERY ” aligns with the artery
of the patient.
2.Wrap the cuff so that “INDEX ” comes
within the “ RANGE ”.
RANGE
If “Index
”, change the cuff size.
” is not within the “ RANGE
Attaching Electrodes
Selecting Electrode Lead
CAUTION
Use Nihon Kohden specified electrodes and electrode leads. With electrodes
and electrode leads other than specified ones, the “CHECK ELECTRODE”
message appears and monitoring may stop.
The following electrode leads can be used on the transmitter (option).
BR-903PA,
3 electrodes,
clip type
Operator's Manual ZS-940PA33
BR-913PA,
3 electrodes,
snap type
BR-902PA,
2 electrodes,
clip type
BR-912PA,
2 electrodes,
snap type
Connecting the Electrode Lead to the Transmitter
Connect the electrode lead to the ECG/RESP socket on the transmitter.
When the transmitter is attached on an arm
CAUTION
Hold the connector of the electrode lead when connecting/disconnecting the
electrode lead. If you disconnect the electrode lead by pulling the lead, it
damages the electrode lead.
Selecting the Electrode Position
Follow the physician’s instructions for electrode placement when available.
For ECG monitoring, electrodes are attached only on the chest to allow patient movement and
obtain continuous stable ECG. Following leads are examples. When also monitoring
respiration, refer to the “Electrode Position for Respiration Monitoring” section.
NOTE
The optimum electrode positions for ECG measurement of a patient are not
always optimum for respiration measurement of the patient. Select positions
suitable for both ECG and respiration measurements, or positions which have
priority for one measurement.
34Operator's Manual ZS-940PA
Three Electrodes
• Lead MII, which is similar to standard lead II, used when ECG measurement has priority
Electrode Position
SymbolLead Color
AHAIECAHAIEC
Left infraclavicular fossaLALBlackYellow
Right infraclavicular fossaRARWhiteRed
Below lowest rib on the left
anterior axillary line
• Lead MI, which is similar to standard
lead I
Change F/LL and L/LA of the lead
LLFRedGreen
• Lead MIII, which is similar to standard lead
III.
Change R/RA and L/LA of the lead MII.
MII.
If the electrode position shown above is not available due to chest surgery, attach the electrodes
to the root of the limbs or below the clavicles for stable ECG monitoring.
Operator's Manual ZS-940PA35
Two Electrodes
With the optional BR-912PA and BR-902PA electrode leads, measurement with two electrodes
is available. L/LA electrode is not used.
This is effective for a neonate or a patient whose body area is small and difficult to attach three
electrodes.
(ex.) Lead MII, which is similar to standard lead II
Electrode Position
Right infraclavicular fossaRARWhiteRed
Below lowest rib on the left
anterior axillary line
Difference between measurement with two electrodes and three electrodes
Measurement with two electrodes is less stable than measurement with three electrodes because
of noise and body movement. Pay attention to this point. If ECG of necessary quality cannot be
obtained, measure with three electrodes.
Attaching Electrodes to the Patient and Connecting the Electrode Leads to
Disposable Electrodes
Prepare the Patient Skin
Shave off excessive body hair.
To reduce skin impedance, clean the electrode site with cream or with a cotton pad moistened
with alcohol. Thoroughly dry the skin with a clean cotton pad.
SymbolLead Color
AHAIECAHAIEC
LLFRedGreen
NOTE
• For a patient with frequent body movement, rub the sites with Skinpure skin
preparation gel. However, do not use Skinpure skin preparation gel on
sensitive skin.
• Do not place electrodes on a wound or on an inflamed, wrinkled or uneven skin
surface.
36Operator's Manual ZS-940PA
Attaching Electrodes to the Patient
CAUTION
Do not reuse disposable products.
NOTE
• To maintain good contact between the electrode and skin, check that the paste
of the disposable electrode is not dry.
• When contact between the disposable electrode and skin becomes poor,
replace electrodes with new ones immediately. Otherwise, contact impedance
between the skin and the electrode increases and the correct ECG cannot be
obtained.
Refer to the electrode operator’s manual for details.
1.Carefully remove the backing paper from the
electrode. Avoid touching the adhesive
surface.
2.Place the electrode on the previously cleaned
skin. Pay attention to the electrode lead color
and symbol.
3.Clip the electrode lead to the electrode.
4.Fasten the electrode lead wire with surgical
tape with an extra length of wire between the
tape and the electrode. This lessens the
movement of electrode leads by body
movement and helps stable monitoring.
Operator's Manual ZS-940PA37
Electrode Position for Respiration Monitoring
Place the R/RA and F/LL electrodes so that the lungs are between the electrodes.
NOTE
The optimum electrode positions for ECG measurement of a patient are not
always optimum for respiration measurement of the patient. Select positions
suitable for both ECG and respiration measurements, or positions which have
priority for one measurement.
Electrode Position Examples
Position 1
In this position, respiration measurement is available; however, there is a difference in amplitude
between different patients.
R or RAF or LL
Right infraclavicular
fossa
Position 2
In this position, the waveform amplitude is usually large and the ECG lead is similar to Lead
MII. This position can be generally recommended.
R or RAF or LL
Right infraclavicular
fossa
Fifth intercostal space on the
left midclavicular line, V4
Fifth intercostal space on the
left midaxillary line, V6
Position 3
In this position, the respiration waveform is optimum, but the ECG lead is unusual.
R or RAF or LL
Right midaxillary at the
horizontal level of V4
38Operator's Manual ZS-940PA
Fifth intercostal space on the
left midaxillary line, V6
Position 4
In this position, the respiration measurement is influenced by the impedance variation of the
abdomen, so the cardiac pulse wave included in the respiration wave is reduced. Note that the
waveform is inverted in phase compared with the chest movement (the waveform goes down
during inspiration). It is difficult to measure the ECG at the same time.
R or RAF or LL
Lowest rib on the right
anterior axillary line
Lowest rib on the left
anterior axillary line
Attaching the SpO2 Probe
Selecting the SpO2 Probe
Select an appropriate probe for the patient.
CAUTION
• Use Nihon Kohden specified SpO2 probe to assure maximum performance
from your instrument.
• Do not use damaged or disassembled probe. It causes incorrect
measurement and may hurt the patient.
Reusable Probes
When using a TL-201T finger probe, choose the appropriate cable length for attachment.
ProbeCable LengthPatientAttachment site
Finger probe TL-201T
0.6 m
Adult or children
20 kg or more
Finger
1.6 m
Attachment tape
Adult or Infant
Finger or toeMulti-site probe TL-220T
3 kg or more
Neonate
Instep and sole
3 kg or less
Operator's Manual ZS-940PA39
Disposable Probes
CAUTION
• Use the disposable probe only for one patient. Never reuse the disposable
probe for another patient because it causes cross infection.
• Disposable probes are not sterilized.
ProbePatientAttachment site
TL-251TAdult
30 kg or more
Finger or toe
TL-252TChild
3 to 40 kg
TL-253TNeonate
3 kg or less
Multi-site Y probe TL-260T
Attachment tape
Low birth weight infant
1 kg or less
Neonate or Child
3 kg or more
Neonate
3 kg or less
Adult
50 kg or more
40 mm
Neonate
3 kg or less
Cable length TL-051S: 80 cm
TL-052S: 160 cm
Child or Adult
15 to 50 kg
Finger or toe
Instep and sole
Instep and
sole
Attachment
tape S
Finger or toeAttachment
tape S
Instep and
sole
Attachment
tape L
FingerTL-051S/052S
Instep and sole
FingerTL-061S/062S
35 mm
Infant
Toe
3 to 15 kg
Cable length TL-061S: 80 cm
TL-062S: 160 cm
40Operator's Manual ZS-940PA
Connecting the SpO
Connect the probe to the SpO
Probe to the Transmitter
2
socket on the transmitter.
2
When the transmitter is attached on an arm
CAUTION
Hold the connector when connecting/disconnecting the probe. If you disconnect
the SpO
probe by pulling the cable, it damages the cable.
2
Attaching the Probe to the Patient
Attach the probe to the patient by referring to the probe’s manual. Make sure that the light
emitter and photo detector of the probe face each other at the attachment site.
WARNING
• When using a TL-201T finger probe, do not fasten it to a finger by wrapping the
probe to the site with some tape. It may cause poor blood circulation,
congestion, pressure necrosis or burn.
• When using probes other than a TL-201T finger probe, do not wrap the tape too
tight. Check the blood circulation condition by observing the skin color and
congestion at the skin peripheral to the probe attachment site. Even for shortterm monitoring, there may be burn or pressure necrosis from poor blood
circulation, especially on neonates or low birth weight infants whose skin is
delicate. Accurate measurement cannot be performed on a site with poor
peripheral circulation.
Operator's Manual ZS-940PA41
• Check the circulation condition by observing the skin color of the measuring
site and the pulse waveform. Change the measuring site every 8 hours for
disposable probes and every 4 hours for reusable probes. The skin
temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and
cause a burn or pressure necrosis. When using the probe on the following
patients, take extreme care and change the measurement site more
frequently according to symptoms and degree.
• A patient with a fever
• A patient with a peripheral circulation insufficiency
• Neonate or low birth weight infant with delicate skin
CAUTION
• If the attachment site is dirty with blood, clean the attachment site before
attaching the probe. If there is nail polish on the attachment site, remove the
polish. Otherwise, the amount of transmitted light decreases and the
measured data may be incorrect or measurement cannot be performed.
• If the skin gets irritated or redness appears on the skin from the probe,
change the attachment site or stop using the probe.
• When the probe is attached on an appropriate site with sufficient circulation,
but the error message confirming the probe attachment repeatedly appears,
the probe may be deteriorated. Replace it with a new one.
• Do not use a probe that is deteriorated by aging. Accurate measurement
cannot be performed.
• When using probes other than a TL-201T finger probe on a neonate, be careful
when removing the adhesive tape from neonatal skin.
• When removing a probe taped to the skin, do not pull the cable. Otherwise the
cable may break.
• When removing the probe from the attachment tape, do not pull the sensor
cable. Otherwise the cable may get damaged.
• Before using the TL-260T multi-site Y probe, be sure to attach the probe to the
sponge attachment tape S or L. Do not use the probe without the sponge
attachment tape attached. It causes incorrect measurement and may damage
the attachment site on the skin.
• When fixing the TL-260T multi-site Y probe with the sponge attachment tape,
confirm that the adhesive part of the tape is not on the skin. The adhesive may
cause oversensitive symptoms on the skin such as redness or itch. If the
adhesive touches the skin, remove it carefully and slowly because neonatal
skin is very delicate.
• Do not use a dirty sponge attachment tape. The measurement value may be
incorrect.
• Refer to the probe instruction manual for details.
42Operator's Manual ZS-940PA
Locking the Keys on the Transmitter
To prevent the patient from pressing the keys on the transmitter during monitoring, you can lock
the NIBP START/STOP and NIBP INTERVAL keys.
1.Press the NIBP INTERVAL key for about 3 seconds.
2.A “pip, pip” sounds and the “LOCK KEYS” message is displayed on the LCD.
Pip, pip
LOCK
KEYS
When the NIBP START/STOP key or NIBP INTERVAL key is pressed while the keys are
locked, the “PRESS INT. KEY 3S TO UNLOCK” message appears.
To unlock the keys:
1.Press the NIBP INTERVAL key for about 3 seconds.
2.A “pip, pip” sounds and the keys are unlocked. The “UNLOCK KEYS” message appears
and the keys are unlocked.
Operator's Manual ZS-940PA43
Monitoring
When preparation is done, monitoring starts.
NIBP Oscillometric Method
NIBP is measured from the change in amplitude pattern of pulsatile oscillation in cuff pressure
as the cuff pressure is reduced from above systolic to below diastolic pressure. The occlusiveoscillometry method uses this to determine the systolic, diastolic and mean arterial pressure.
NIBP Monitoring
Selecting the Initial Cuff Inflation Pressure
The initial cuff inflation pressure can be changed on the PARAMETER SETUP screen. The
default setting is 180 mmHg. To change the setting, refer to the “Changing Parameter Setup
Settings” section.
Selecting the Measurement Mode and Interval
Measurement Modes
There are three measurement modes: manual, auto and STAT. The selected mode or interval is
displayed at the lower right of the screen.
The measurement mode and interval can be changed by pressing the NIBP INTERVAL key.
When the key is pressed, the measurement mode changes according to the modes selected at
“SELECTABLE INTERVALS” on the PARAMETER SETUP screen. MANUAL mode is
already selected for the mode selection.
To select the modes for the mode selection, refer to the “Changing Parameter Setup Settings”
section.
Manual Measurement
In Manual mode, a single NIBP measurement is performed when the NIBP START/STOP key
is pressed.
STAT (Continuous) Measurement
In STAT mode, measurement is continuously repeated for 15 minutes after the NIBP START/
STOP key is pressed.
When the STAT measurement for 15 minutes is completed, the measurement mode automatically
changes to the Manual mode or Auto mode of selected interval depending on the “NIBP MODE
44Operator's Manual ZS-940PA
AFTER STAT” setting on the PARAMETER SETUP screen. The default setting is Manual
mode. Refer to the “Changing Parameter Setup Settings” section.
The STAT measurement completes within 15 minutes. When more than 12 minutes elapse from
the start of measurement, there will be no more measurement performed and the measurement
mode changes to the mode selected for “NIBP MODE AFTER STAT” on the PARAMETER
SETUP screen.
Auto Measurement
In Auto mode, measurement is performed automatically at the preset time intervals.
In Auto mode, a single measurement can be performed by pressing the NIBP START/STOP key
between auto measurements.
Measuring NIBP
WARNING
• Be careful when measuring NIBP on a patient with known bleeding disorders
or coagulation. After NIBP measurement, there may be dot hemorrhage, or
circulatory disorder by thrombus where cuff is attached.
• NIBP measurement may be incorrect in the following cases.
• When using an electrical surgery unit.
• When there is body movement.
• When the pulse wave is small (insufficient peripheral circulation).
• Too many arrhythmia.
• When there is vibration.
• When there is a rapid blood pressure change.
• During CPR.
• When performing NIBP measurements in STAT mode or 5 minute intervals,
periodically remove the cuff from the patient for ventilation. Otherwise, the
skin temperature may increase at the cuff attachment site by 2 or 3°C (4 or
5°F). When measuring a patient with a fever or peripheral circulation
insufficiency, it may cause a burn.
CAUTION
When performing NIBP measurement repeatedly, have a rest between
measurements to recover adequate circulation.
Operator's Manual ZS-940PA45
NOTE
• When measuring patients who are conscious, help the patient to relax.
Measurement may not be accurate if the patient’s arm is tense or if the patient
talks.
• The data for measurement on a leg tends to be higher than measurement on
the arm. When making diagnosis based on the NIBP values, measure NIBP on
an upper arm.
• Do not apply pressure to the cuff or air hose. NIBP may not be measured
correctly because of noise or NIBP measurement may stop due to the NIBP
safety circuit.
• When the transmitter is attached to the patient arm and the NIBP
measurement is performed when moving, tell the patient to relax and keep
quiet. Otherwise, measurement may be stopped or remeasurement is
repeated due to body movement.
• If there is an abnormal noise generated during measurement, stop using the
transmitter and contact your Nihon Kohden distributor.
• Do not measure NIBP of a patient on whom an IABP is being used.
Measurement may be incorrect due to the mixing of the patient’s own pulse
and IABP pulse.
• NIBP cannot be measured on a neonate using this transmitter.
1.Select the measurement mode by pressing the NIBP INTERVAL key.
2.Press the NIBP START/STOP key to perform measurement.
46Operator's Manual ZS-940PA
The cuff is inflated and the inflation pressure is displayed on the screen.
Inflation pressure
In manual mode: Measurement is performed once.
In STAT mode:Measurement is performed repeatedly for 15 minutes.
In auto mode:The first measurement is performed when the NIBP START/STOP key
is pressed. The second measurement is performed when the current time
in the transmitter reaches the selected time interval.
To stop measurement during measurement, press the NIBP START/STOP key again.
In STAT mode, after completing the STAT measurement, the measurement mode changes to
the mode set for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen.
In auto mode, to stop measurement in auto mode, change the mode to manual. To cancel
one measurement, press the NIBP START/STOP key during measurement.
After the measurement is complete, the measured data is displayed on the screen and is
transmitted to the monitor.
When SpO2 is not monitored, the pulse rate at the end of NIBP measurement is displayed.
During auto mode measurement, the measurement mode can be changed. During the interval,
press the NIBP INTERVAL key to change the mode. When “MANUAL” is displayed for more
than one second, the measurement in auto mode is stopped.
A buzzer can be set to sound at the start and end of NIBP measurement. Refer to the “Changing
Parameter Setup Settings” section.
Operator's Manual ZS-940PA47
Data Display After NIBP Measurement
When the time set at “OLD NIBP DATA” on the PARAMETER SETUP screen elapses after
the last measurement, the NIBP data is dimmed or hidden. Whether to dim or hide the old data
can also be selected at “OLD NIBP DATA”. Refer to the “Changing Parameter Setup Settings”
section.
Monitoring SpO
When the SpO
during NIBP measurement and pulse wave cannot be detected and SpO
properly. When “INHIBIT SpO
to ON (factory default setting), SpO
SpO
alarm occurrence. However, when monitoring SpO2 on the same limb as the NIBP, be
2
careful when reading SpO
during NIBP Measurement
2
probe is attached to the same limb as the NIBP cuff, the blood flow decreases
2
cannot be monitored
2
DURING NIBP” on the PARAMETER SETUP screen is set
2
monitoring is paused during NIBP measurement to avoid
2
values.
2
ECG and Respiration Monitoring
When the electrodes are attached and the ECG leads are connected to the electrodes, heart rate,
ECG, respiration rate and respiration waveform appear on the monitor.
WARNING
Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac
monitoring and Diagnostic Equipment*
The bioelectric impedance measurement sensor of a minute ventilation rateadaptive implantable pacemaker may be affected by the transmitter which is
connected to the same patient. If this occurs, the pacemaker may pace at its
maximum rate and the transmitter may give incorrect data to the monitor. If this
occurs, disconnect the electrode leads from the patient or change the setting on
the pacemaker by referring to the pacemaker’s manual. For more details,
contact your pacemaker distributor or Nihon Kohden distributor.
* Minute ventilation is sensed in rate-adaptive pacemakers by a technology known as bioelectric
impedance measurement (BIM). Many medical devices in addition to pacemakers use this
technology. When one of these devices is used on a patient with an active, minute ventilation
rate-adaptive pacemaker, the pacemaker may erroneously interpret the mixture of BIM signals
created in the patient, resulting in an elevated pacing rate.
For more information, see the FDA web site.
http://www.fda.gov/cdrh/safety.html
48Operator's Manual ZS-940PA
WARNING
When using this transmitter with an ESU, the ESU return plate and the
electrodes for monitoring must be firmly attached to the patient. If the return
plate is not attached correctly, it may burn the patient’s skin where the
electrodes are attached. Refer to the instruction manual for the ESU.
CAUTION
Turn off the power of cell telephones, small wireless devices and other devices
which produce strong electromagnetic interference. Otherwise, the waveforms
and measurements are affected by such interference and the displayed data
may be incorrect.
NOTE
• Noise generated from an electrosurgery unit may interfere on an ECG
waveform, but will not damage it.
• If an electric blanket is used and incorrect heart rate is displayed on the
monitor, turn off the pacing spike detection on the monitor.
• Turn the pacing spike detection to ON on the monitor when monitoring a
pacemaker patient. Pacing pulse is detected by the transmitter and
transmitted to the monitor. If the pacing spike detection is turned OFF, QRS
and pacemaker spike may not be distinguished and pacemaker failure may
not be recognized.
Electrode Detachment
The “
message is displayed on the screen of the monitor in the following conditions.
• Electrode is detached from skin.
• Electrode lead is disconnected from the electrode.
• Polarization voltage between the electrode and skin is excessively high.
In these cases, check the cause and if necessary, replace electrodes with new ones.
” mark is displayed on the LCD of the transmitter or the “CHECK ELECTRODE”
CAUTION
When the “ELECTRODE OFF” or “CHECK ELECTRODE” message is displayed
on the receiving monitor, check electrodes and electrode leads and remove the
cause. While the “ELECTRODE OFF” or “CHECK ELECTRODE” message is
being displayed, there is no ECG monitoring and no alarms.
Operator's Manual ZS-940PA49
SpO2 Monitoring
When monitoring starts, SpO2 and pulse waveform are sent to the monitor and SpO2, pulse rate
and pulse level bar graph are displayed on the transmitter LCD.
WARNING
• Measurement may be incorrect in the following cases.
• When the patient’s carboxyhemoglobin or methemoglobin increases
abnormally.
• When dye is injected in the blood.
• When using an electrical surgery unit.
• During CPR.
• When there is body movement.
• When there is vibration.
• When measuring at a site where there are venous pulses.
• When the pulse wave is small (insufficient peripheral circulation).
• When using an IABP (intra-aortic balloon pump).
• Check the circulation condition by observing the skin color of the measuring
site and pulse waveform. Change the measuring site every 8 hours for
disposable probes and every 4 hours for reusable probes. The skin
temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and
cause a burn or pressure necrosis. When using the probe on the following
patients, take extreme care and change the measurement site more
frequently according to symptoms and degree.
• A patient with a fever
• A patient with peripheral circulation insufficiency
• Neonate or low birth weight infant with delicate skin
• When not monitoring SpO
Otherwise, noise may interfere from the probe sensor and incorrect data is
displayed on the screen.
, disconnect the SpO2 cable from the transmitter.
2
CAUTION
• Turn off the power of cell telephones, small wireless devices and other devices
which produce strong electromagnetic interference. Otherwise, the
waveforms and measurements are affected by such interference and the
displayed data may be incorrect.
• Normally external light does not affect monitoring, however, strong light such
as an operating lamp or sunlight may affect monitoring. If affected, cover the
measuring site with a blanket.
• Do not pull or bend the probe cable, and do not run over the probe cable with
50Operator's Manual ZS-940PA
caster feet. Do not immerse the probe cable in detergents or water. Failure to
follow these cautions may cause cable discontinuity, short circuit, skin burn
on the patient and incorrect measurement data. Replace any broken probe
with a new one.
• When the probe is attached on an appropriate site with sufficient circulation
and the error message confirming the probe attachment repeatedly appears,
the probe may be deteriorated. Replace it with a new one.
• When the probe failure message appears on the screen, replace it with a new
one. Otherwise SpO
data may not be accurate.
2
• While a patient is on medication which causes vasodilation, the pulse
waveform may change and in rare cases the SpO
value may not be displayed.
2
NOTE
In order to maintain sufficient blood circulation, keep the measurement site
warm by covering it with a blanket or something similar. Warming the site is
effective, especially for a patient with a small pulse amplitude.
SpO
and PR Display Order
2
You can select the display order for SpO
“Changing Parameter Setup Settings” section.
and PR (pulse rate) on the LCD. Refer to the
2
Monitoring SpO
When the SpO
during NIBP measurement and pulse wave cannot be detected and SpO
properly. When “INHIBIT SpO
to ON (factory default setting), SpO
SpO
alarm occurrence. However, when monitoring SpO2 on the same limb as NIBP, be careful
2
when reading SpO
When monitoring SpO
during NIBP Measurement
2
probe is attached to the same limb as the NIBP cuff, the blood flow decreases
2
cannot be monitored
2
DURING NIBP” on the PARAMETER SETUP screen is set
2
monitoring is paused during NIBP measurement to avoid
2
values.
2
is important, attach the probe to the limb to which the NIBP cuff or
2
catheter is not attached.
When SpO2 monitoring is paused during NIBP measurement, the SpO2 value just before the
start of NIBP measurement and an
When NIBP measurement is not completed after 30 seconds, “– – –” is displayed for the SpO
mark are displayed on the transmitter for 30 seconds.
2
value. The same data also appears on the monitor screen.
Operator's Manual ZS-940PA51
NOTE
• When continuous SpO2 monitoring is necessary, attach the probe to the limb to
which the NIBP cuff is not attached and set “INHIBIT SpO
the PARAMETER SETUP screen to OFF.
• When the probe is attached to the same limb as the NIBP cuff, set the sync
source to a parameter other than SpO
• When monitoring SpO
during STAT NIBP measurement, attach the probe to
2
on the receiving monitor.
2
the limb to which the NIBP cuff is not attached.
DURING NIBP” on
2
52Operator's Manual ZS-940PA
Display and Message List
Battery Indication
IndicationCauseCountermeasure
Fully charged battery
Batteries are low.
No
indication
Batteries are low. NIBP cannot be measured.
Dead batteries
Replace batteries.
ECG
IndicationCauseCountermeasure
Electrode lead is disconnected from the
electrode.
Electrode lead is disconnected from the
transmitter.
Electrode lead discontinuity.
Electrode is not firmly attached to the skin.
Polarization voltage is abnormally high.
Firmly connect the electrode
lead to the electrode.
Firmly connect the electrode
lead to the transmitter.
Replace the electrode lead
with a new one.
Replace the electrode with a
new one.
Operator's Manual ZS-940PA53
SpO
2
MessageCauseCountermeasure
During NIBP
measurement
Detecting
body
movement
SpO2 CHECK
PROBE
SpO2 CHECK
PROBE SITE
SpO2 DETECTING
PULSE
SpO2 monitoring is paused for
NIBP measurement.
Considerable body movement.
The probe is not attached to
the patient properly.
The probe is not attached to
the patient properly.
The probe is disconnected
from the SpO
socket on the
2
transmitter.
The probe is not attached at
the appropriate site.
Wait for NIBP measurement to
finish.
When the message is displayed
frequently, check the patient
condition and, if necessary, change
the attachment site.
Attach the probe to the patient
properly.
Connect the probe to the SpO
socket.
Attach the probe to a site 6 to 14
mm thick.
Probe is expired.Replace the probe with a new one.
Searching for the correct pulse
wave.
Wait until the pulse wave is
detected.
The SpO2 value cannot be
obtained because the
waveform is unstable.
The probe is not attached to
Attach the probe to the patient
properly.
the patient properly.
SpO
measurement site is
SpO2 LIGHT
INTERFERENCE
SpO2 PROBE
FAI LU RE
SpO2 WEAK PULSE
2
under fluorescent light,
surgical light, sunlight, etc.
Probe is expired.Replace the probe with a new one.
Probe is damaged or shortcircuited.
Poor peripheral circulation.
The probe is attached too
tightly and is obstructing the
blood circulation.
Cover the measurement site with a
blanket or cloth.
Replace the probe with a new one.
Check the patient condition and
change the attachment site.
Check the probe attachment
condition and if necessary,
reattach the probe.
2
54Operator's Manual ZS-940PA
NIBP
MessageCauseCountermeasure
The cuff and extension hose are not
NIBP AIR LEAK
NIBP CANNOT
DETECT PULSE
NIBP CUFF
OCCLUSION
NIBP HIGH CUFF
PRESS
NIBP INFLATION
PRESS LOW
NIBP MEAS TIMEOUT
NIBP MODULE
FAI LU RE
NIBP REMEASURING
NIBP SAFETY
CIRCUIT RUNNING
(When this message is
displayed, measurement
cannot be performed for
40 seconds.)
NIBP SYS OUT OF
RANGE
NIBP WEAK PULSE
NIBP ZEROING
properly connected.
The cuff hose (or extension hose) is
not properly connected to the NIBP
socket.
The cuff or extension hose is
damaged.
The patient’s pulse wave is small.Measure by palpation or auscultation.
The cuff is not wrapped on the
patient properly.
Transmitter malfunction.
Enormous pressure was applied by
the pressure of the cuff.
Insufficient cuff inflation pressure.
The measuring time exceeded the
specified time due to arrhythmia,
body movement, vibration or, cuff
or air hose being squeezed.
Module malfunction.
NIBP is being remeasured due to
arrhythmia, body movement,
vibration or, cuff or air hose being
squeezed.
Measurement stopped by the safety
circuit.
The maximum blood pressure
cannot be measured even when the
cuff inflation pressure exceeded 280
mmHg when using adult cuff.
The patient’s pulse wave is too
small.
The cuff is wrapped too loosely.Wrap the cuff properly.
The cuff size is not appropriate.Use the appropriate cuff.
NIBP zero balance is being
adjusted.
Connect them properly.
Replace with a new one.
Wrap the cuff on the patient properly.
Immediately remove the cuff from the
patient and contact your Nihon Kohden
distributor.
Remove the cause.
Wait for the remeasurement to be
performed with increased cuff inflation
pressure.
Remove the cause if the cause is body
movement, vibration or squeezing of
cuff or hose.
Contact your Nihon Kohden
distributor.
If the message still appears after
remeasurement, remove the cause if
the cause is body movement, vibration
or squeezing of cuff or hose.
Check that the hose is not bent or
squeezed.
Wait 40 seconds, then perform
remeasurement. If the message still
appears, contact your Nihon Kohden
distributor.
Measure by palpation or auscultation.
Measure by palpation or auscultation.
Do not touch the cuff during zeroing.
Wait for the message to disappear.
Operator's Manual ZS-940PA55
p
Troubleshooting
If the problem still remains after checking the following, contact your Nihon Kohden distributor.
Transmitter
ProblemCauseCountermeasure
displayed on the
LCD after
turning the
power on.
displayed on the
monitor after
turning the
transmitter
power on.
Signal receiving
condition is
oor.
Batteries are not installed
correctly. The battery
polarity is wrong.
Batteries are completely
discharged.
The channel of the
transmitter and monitor
does not match.
The “TYPE” on the
SYSTEM SETUP screen
is not set to “A”.
Another transmitter of the
same channel is used
nearby.
Signals are mixing.Follow the instructions of your channel
The transmitter is
damaged.
Install the batteries correctly.Nothing is
Replace the batteries with new ones.
Set the correct channel on the monitor.Nothing is
Check that the “TYPE” is set to “A”.
Refer to the “Notes on Parameter
Settings” and “Changing System Setup
Settings” sections.
Turn the transmitter power off. If the
monitor still receives a signal, there is a
high probability that another transmitter
of the same channel is used nearby.
Follow the instruction of your channel
administrator and use another transmitter
of a different channel.
administrator and use another transmitter
of a different channel.
Contact your Nihon Kohden distributor.
56Operator's Manual ZS-940PA
ECG
ProblemCauseCountermeasure
The heart rate is
unstable.
The “CHECK
ELECTRODE”
message appears
on the receiving
monitor.
ECG baseline is
thick.
(Hum is
overlapping)
Respiration
waveform
measurement is
unstable.
Pacing detection setting on the
monitor is not correct.
Turn off the pacing detection setting
on the monitor.
When monitoring a pacemaker
patient, turn on pacing detection.
Electrode lead is disconnected
from the electrode.
Firmly connect the electrode lead to
the electrode.
Electrode lead discontinuityReplace the electrode lead with a new
one.
Electrode is not firmly
Replace the electrode with a new one.
attached to the skin.
Polarization voltage is
abnormally high.
The gel on the electrode is
Use Nihon Kohden specified
electrodes.
Replace the electrode with a new one.
dried out.
The gel on the electrode is
coming off.
Electric blanket is used.Cover the blanket with a shield cover.
Hum filter is set to OFF on the
Set the filter to ON.
monitor
The gel on the electrode is
Replace the electrode with a new one.
dried out.
The gel on the electrode is
coming off.
SpO
2
ProblemCauseCountermeasure
SpO2 data is
unstable and
not reliable.
The probe size is not
appropriate for the patient.
Probe attachment condition
is poor. Probe is partly
Use the appropriate probe for the patient.
Firmly attach the probe according to the
procedure in the probe operator’s manual.
detached from the skin.
External light gets in.
Measurement site is dirty.
Remove dirt and nail polish.
Patient is wearing nail
polish.
Probe is attached to the
same limb that is used for
NIBP measurement.
When the probe and cuff are attached to
the same limb, set “INHIBIT SpO
2
DURING NIBP” setting on the
PARAMETER SETUP screen to ON.
Attach the probe to the opposite limb.
Avoid a site where blood circulation
condition changes greatly.
Operator's Manual ZS-940PA57
NIBP
ProblemCauseCountermeasure
Cuff inflation
pressure is less than
10 mmHg.
The cuff does not
inflate when the
NIBP START/STOP
key is pressed.
Abnormal
measurement results
are displayed.
The cuff is suddenly
deflated during
inflation.
Auto mode
measurement does
not start even when
the time interval has
passed.
The cuff suddenly
inflates.
Cannot connect cuff
to the air hose.
The cuff hose is not connected
to the NIBP socket properly.
The cuff is not wrapped
around the arm or is wrapped
too loosely.
The cuff hose is not connected
to the NIBP socket.
The cuff hose or extension
hose may be folded or
squeezed when the cuff
pressure display on the screen
increases quickly but the
actual cuff does not inflate.
The cuff size is not correct.Select the cuff which fits the
The cuff is not wrapped
around the arm correctly.
NIBP data is not correct
because of body movement.
Vibration on the cuff.
The NIBP START/STOP key
is pressed during inflation.
The NIBP INTERVAL key is
pressed and the measurement
mode is changed.
The measurement mode is set
to auto mode.
Unspecified cuff is used.Use a cuff specified by Nihon
Connect the cuff hose to the socket
properly.
Wrap the cuff around the upper
arm.
Connect the cuff hose to the socket
firmly.
Check the cuff hose and air hose.
patient’s limb circumference.
Wrap the cuff around the upper
arm, not too tightly or too loosely.
Prevent the patient from moving
during measurement.
Check that nothing is touching the
cuff during measurement.
Change the measuring site.
Check the measurement mode and
interval.
Check the time interval. If
necessary, stop measurement.
Kohden.
58Operator's Manual ZS-940PA
ProblemCauseCountermeasure
Cannot measure
NIBP.
Vibration on the cuff.
The cuff hose or extension
Check that nothing is touching the
cuff during measurement.
Remove the cause.
hose is bent or squeezed.
The cuff has worn out.Use a new cuff.
Measuring over a long
occurs.
period of time at short
intervals.
Thrombus occurs.Measuring on a patient
with known bleeding
Increase the measuring interval.Blood congestion
Do not measure NIBP over a long
time.
Do not perform NIBP measurement
on such a patient.
disorders or coagulation.
NIBP data on the
screen is --- or dark.
The time set for “OLD
NIBP DATA” on the
PARAMETER SETUP
When NIBP is measured again, the
data is displayed in normal
brightness.
screen elapsed from the last
measurement.
Three loud pip
sounds indicting
NIBP measurement
The cuff is not deflated
enough to start another
measurement.
Wait 30 seconds and measure again.
cannot be started.
Numeric data is
displayed but NIBP
messages are not
displayed on the
The “TYPE” on the
SYSTEM SETUP screen is
not set to “A”.
Check that the “TYPE” is set to “A”.
Refer to the “Notes on Parameter
Settings” and “Changing System
Setup Settings” sections.
monitor.
Operator's Manual ZS-940PA59
Maintenance
To use the instrument in safe and optimum condition, perform maintenance check once every six
months.
CAUTION
Do not disassemble the transmitter when performing maintenance and
inspection. Do not repair the transmitter. When there is any problem with the
transmitter after maintenance and inspection, contact your Nihon Kohden
distributor.
A maintenance check sheet is provided at the end of this section. Make a copy of this check
sheet before performing maintenance check.
1. External Check
• There are no damaged or dirty parts on the outside of the transmitter.
• The battery case cover is not damaged, the spring is firmly fixed and the battery case cover
can be closed firmly.
• NIBP socket is not damaged.
• Keys are not damaged.
• Electrode leads are not damaged.
• There is no blood or chemicals on the transmitter.
2. Transmitter Channel
Check that the channel of the transmitter and the label match.
1.Check that the channel number label attached to the transmitter is not torn or removed.
Channel label
60Operator's Manual ZS-940PA
2.Remove one battery.
3.While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery.
The SETUP screen appears.
4.Press the NIBP INTERVAL key to move the cursor to “SYSTEM SETUP”.
5.Press the NIBP START/STOP key to enter
the SYSTEM SETUP screen. The channel of
this transmitter is displayed.
Channel
6.Check that the channel displayed on the LCD
matches the label on the transmitter.
Operator's Manual ZS-940PA61
3. LCD Display
Check that there are no dots missing on the LCD.
1.Remove one battery.
2.While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery.
The SETUP screen appears.
3.Press the NIBP INTERVAL key twice to move the cursor to “MANUAL CHECK”.
4.Press the NIBP START/STOP key to
enter the MANUAL CHECK screen.
5.Check that the cursor is on “LCD
TEST” and press the NIBP START/
STOP key.
62Operator's Manual ZS-940PA
6.Every time the NIBP INTERVAL key is pressed, the screen changes as below. Check that
there are no dots missing.
When the NIBP START/STOP key is pressed, the screen returns to the MANUAL
CHECK screen.
4. Switch and Key Operation
Power Switch
Check that the Power switch turns the power on and off.
NIBP START/STOP Key
1.Attach the NIBP cuff to your upper arm.
2.Press the NIBP START/STOP key. Check that the cuff inflates and deflates properly.
3.Press the NIBP START/STOP key again. During inflation, press the NIBP START/STOP
key to check that the cuff deflates properly.
NIBP INTERVAL Key
1.Press the NIBP INTERVAL key and check that the NIBP measuring mode can be changed.
2.Select any interval and press the NIBP START/STOP key to perform auto measurement.
Check that the NIBP is measured at the selected interval.
Operator's Manual ZS-940PA63
5. NIBP Cuff for Attaching Transmitter to Patient Arm
The NIBP cuff is a consumable. Check the following and when necessary, replace it with a new
one.
Appearance
• There are no dirty parts.
• There are no broken stitches on the cuff.
• The label on the cuff is readable.
• The velcro tape on the cuff is not removed and there are no broken stitches.
• The lock plate is not damaged and functions properly.
Inflation bag
• The inflation bag is not torn or damaged.
• There is no water inside the inflation bag.
• The connector on the inflation bag is not damaged.
64Operator's Manual ZS-940PA
Hospital/Organization:
Service Personnel:
Maintenance Check Sheet
Instrument Name:
Instrument Model:
Instrument Serial Number:
Hardware Revision Number:
Software Revision Number:
1. External CheckOKNo
2. Transmitter ChannelOKNo
3. LCD DisplayOKNo
4. Switch and Key OperationOKNo
5. NIBP Cuff for Attaching Transmitter to Patient ArmOKNo
Overall Judgement
OK
Transmitter
ZS-940PA
Can be used but needs maintenance
Maintenance required. Cannot be used.
Operator's Manual ZS-940PA65
Repair Parts Availability Policy
Nihon Kohden Corporation (NKC) shall stock repair parts (parts necessary to maintain the
performance of the instrument) for a period of 8 years from the date of delivery. In that period
NKC or its authorized agents will repair the instrument. This period may be shorter than 8 years
if the board or part necessary for the faulty section is not available.
66Operator's Manual ZS-940PA
Lifetime and Disposal
Disposing of Used Batteries
Battery Lifetime
Replace the batteries when the battery replacement indication appears on the transmitter. When
using rechargeable batteries, recharge them.
TypeLifetime (Measuring parameters)
ECG, SpO2, NIBPECG, SpO
NiMH secondary2 days2 days2.5 days
Alkaline primary1 day2.5 days3 days
The above data is when new batteries are used at room temperature, NIBP is measured in auto
mode at 60 minute intervals and SpO
weight 60 kg.
Operation time depends on the thickness of SpO2 probe attachment site.
Disposal
is measured on an index finger of a male patient with
2
NOTE
Remove the batteries before disposing of the transmitter.
Before disposing of the batteries, check with your local solid waste officials for details in your
area for proper disposal. It may be illegal to dispose of these batteries in the municipal waste
stream.
2
ECG only
Disposing of Electrodes, SpO2 Probes and NIBP Cuffs
Refer to the manual of each item.
Operator's Manual ZS-940PA67
Cleaning, Disinfection and Sterilization
Transmitter and Electrode Leads
CAUTION
• If detergent or liquid spills into the transmitter, stop using it and contact your
Nihon Kohden distributor. If a wet transmitter is used, the patient or anyone in
contact with the transmitter may receive an electric shock or patient leakage
current over the allowed amount may flow.
• Before cleaning or disinfecting the transmitter, remove the batteries from the
transmitter.
• The transmitter cannot be sterilized.
Cleaning
Wipe the transmitter and electrode leads with a soft cloth moistened with disinfecting alcohol or
neutral detergent diluted with water. After cleaning, dry them completely.
Disinfection
CAUTION
• Do not immerse the electrode lead connector in liquid.
• Do not disinfect with hypochlorous acid.
• Use the recommended concentration.
Wipe the outside surface of the transmitter and electrode lead with a non-abrasive cloth
moistened with any of the disinfectants listed below. Use the recommended concentration.
• Never allow liquid to get inside the rubber cuff.
• Do not sterilize or disinfect the cuff with ultraviolet light or ozone.
Cleaning
To clean the cuff, remove the lock plate and carefully pull out the inflation bag from the cloth
cover.
Cloth cover:Wash with neutral detergent and water. Thoroughly dry it. When washing in a
washing machine, put it in a net.
Inflation bag: Wipe with a soft cloth or cotton moistened with disinfecting alcohol.
Thoroughly dry it.
Disinfection
To disinfect the cuff, use glutaraldehyde solution. Use the recommended concentration of the
disinfectant. Refer to the disinfectant manual for details. After disinfection, clean the cuff as
described above.
Operator's Manual ZS-940PA69
Specifications
Measuring Parameters
Measuring waveforms:ECG, Respiration in impedance method, pulse
Measuring numeric data:SpO
Transmitting Data
Waveform data:ECG, respiration, pulse wave
Numeric data:SpO
Status information:Battery replacement, channel ID, type of transmitter, check
Displayed Data
SpO2, NIBP, pulse rate, pulse wave bar graph, check electrode, battery replacement, NIBP
measurement mode and status information
ECG Measurement
Channels:1
Input range:±5 mV or more
DC offset:±500 mV or more
Input impedance:5 MΩ or more (5 Hz)
Pacing pulse detection:ANSI/AAMI EC13
Measuring method: Impedance method
Impedance range: 0 to 2 kΩ or less
SpO
Measurement
2
Display range:Depends on the receiving monitor
Measuring range:0 to 100%, in 1% steps
Minimum display range:1%
Measuring accuracy (When the measuring accuracy of the SpO
±1 (80% ≤ SpO
±2 (50% ≤ SpO
≤ 100%)
2
< 80%)
2
probe is not considered):
2
Less than 50% is not specified.
(When considering the measuring accuracy of the SpO
±2 (80% ≤ SpO
±3 (70% ≤ SpO
probe):
2
≤ 100%)
2
< 80%)
2
Less than 70% is not specified.
70Operator's Manual ZS-940PA
NIBP Measurement
Displayed items:Systolic, diastolic, mean
Cuff pressure display range:0 to 300 mmHg
Measurement modes:Manual, STAT, auto at 5, 10, 15, 30, 60, 120 or 240 minute
interval
Pulse Rate
Measuring range:30 to 200 beats/minute ±8 beats/min (NIBP)
30 to 200 beats/min ±3% ±1 beat/min (SpO
)
2
Transmitter
FCC regulation:FCC part 95 Subpart H
Wireless Medical Telemetry Service (WMTS)
Field strength limits:<200 mV/m (at 3 m)
Undesired emission:below 960 MHz: 200 µV/m (at 3 m)
above 960 MHz: 500 µV/m (at 3 m)
Antenna:Internal
Transmission channel:indicated on the transmitter
Transmission frequency range:608.0125 to 613.9875 MHz
Channel spacing:25 kHz (12.5 kHz when interleave)
Type of emission:F1D
Occupied bandwidth:<8.5 kHz
Effective radiated power:1.0 mW (conducted)
Power Requirements
Operating voltage:3.2 to 4.8 V
Battery type:Three AA type NiMH secondary batteries
Three AA type alkaline dry cell primary batteries
Battery lifetime:
TypeLifetime (Measuring parameters)
ECG, SpO2, NIBPECG, SpO
2
ECG only
NiMH secondary2 days2 days2.5 days
Alkaline primary1 day2.5 days3 days
The above data is when new batteries are used at room temperature, NIBP is
measured in auto mode at 60 minute intervals and SpO
is measured on an index
2
finger of a male patient with weight 60 kg.
Operation time depends on the thickness of SpO
probe attachment site.
2
Dimension and Weight
Dimension:114 W × 103 H × 58 D (mm)
Weight:about 350 g (including batteries)
Operator's Manual ZS-940PA71
Environment
Operating environment
Operating temperature:5 to 40°C, 41 to 104°F
When using NIBP cuff, 10 to 40°C, 50 to 104°F
Operating humidity:30 to 85% (non-condensing)
Operating atmospheric pressure: 70 to 106 kPa
Storage environment
Storage temperature:−20 to 65°C, −4 to 149°F
Storage humidity:10 to 95%
Storage atmospheric pressure:70 to 106 kPa
IEC 60601-2-30 (1999)
According to the type of protection
against electrical shock:INTERNALLY POWERED EQUIPMENT
According to the degree of protection
against electrical shock:
ECG and impedance method respiration: DEFIBRILLATION-PROOF TYPE CF APPLIED
PA RT
SpO
and NIBP:DEFIBRILLATION-PROOF TYPE BF APPLIED
2
PA RT
According to the degree of protection
against harmful ingress of water:IPX0 (Ordinary equipment)
According to the degree of safety of
application in the presence of a
FLAMMABLE ANAESTHETIC
MIXTURE WITH AIR, OR WITH
OXYGEN OR NITROUS OXIDE:Equipment not suitable for use in the presence of
FLAMMABLE ANAESTHETIC MIXTURE WITH
AIR, OR WITH OXYGEN OR NITROUS OXIDE
According to the mode of operation:CONTINUOUS OPERATION
72Operator's Manual ZS-940PA
Standard Accessories
1
2
No.NameModelQ’tySupply Code No.
1NIBP cuff for adult, standardYP-943P1S938B
2Strap---1Y236
The following parts are available for replacement.
34
Lock plate is a
standard accessory of
the YP-943P/944P
NIBP cuff.
No.NameModelQ’tySupply Code No.
3Battery case cover16144-012004
4Lock plate16113-049585
Operator's Manual ZS-940PA73
Options
CAUTION
Use only Nihon Kohden electrodes, electrode leads, SpO2 probes and NIBP cuffs
to assure maximum performance from your instrument.
Transmitter
Channel writer, QI-901PK
ECG/RESP
NameApplicationModelQ’tySupply Code No.
Electrode
lead
3 electrodes,clip type,
lead length 80 cm
3 electrodes, snap type,
lead length 80 cm
2 electrodes,clip type,
lead length 80 cm
2 electrodes,snap type,
lead length 80 cm