Nihon Kohden ZS 940PA User Manual

TRANSMITTER
ZS-940PA
0614-009355
Model: ZS-940PA
Manual code no.: 0614-009355
Reader Comment Card
We welcome your comments about this manual. Your comments and suggestions help us
improve our manuals. Please circle the number for each of the following statements
Fax or send your completed comment card to:
Fax: +81 (3) 5996-8100
International Div., Sales Promotion Section, Nihon Kohden Corp., 1-31-4, Nishiochiai
Shinjuku-ku, Tokyo 161-8560, Japan
Strongly Agree 1 Disagree 4
Agree 2 Strongly Disagree 5
Neutral 3
cutting line
This manual is organized. 12345
I can find the information I want. 12345
The information is accurate. 12345
I can understand the instructions. 12345
The illustrations are appropriate and helpful. 12345
The manual length is appropriate. 12345
Comments:
Thank you for your cooperation. We appreciate it very much.
Name:
Occupation/Position:
Hospital/Company:
Address:
Phone:
GENERAL HANDLING PRECAUTIONS
This device is intended for use only by qualified medical personnel. Use only Nihon Kohden approved products with this device. Use of non-approved products or in a non-approved manner may affect the performance specifications of the device. This includes, but is not limited to, batteries, recording paper, pens, extension cables, electrode leads, input boxes and AC power.
Please read these precautions thoroughly before attempting to operate the instrument.
1. To safely and effectively use the instrument, its operation must be fully understood.
2. When installing or storing the instrument, take the following precautions:
(1) Avoid moisture or contact with water, extreme atmospheric pressure, excessive
humidity and temperatures, poorly ventilated areas, and dust, saline or sulphuric air.
(2) Place the instrument on an even, level floor. Avoid vibration and mechanical shock,
even during transport.
(3) Avoid placing in an area where chemicals are stored or where there is danger of gas
leakage.
(4) The power line source to be applied to the instrument must correspond in frequency
and voltage to product specifications, and have sufficient current capacity.
(5) Choose a room where a proper grounding facility is available.
3. Before Operation
(1) Check that the instrument is in perfect operating order. (2) Check that the instrument is grounded properly. (3) Check that all cords are connected properly. (4) Pay extra attention when the instrument is in combination with other instruments to
avoid misdiagnosis or other problems. (5) All circuitry used for direct patient connection must be doubly checked. (6) Check that battery level is acceptable and battery condition is good when using battery-
operated models.
4. During Operation
(1) Both the instrument and the patient must receive continual, careful attention. (2) Turn power off or remove electrodes and/or transducers when necessary to assure the
patient’s safety. (3) Avoid direct contact between the instrument housing and the patient.
5. To Shutdown After Use
(1) Turn power off with all controls returned to their original positions. (2) Remove the cords gently; do not use force to remove them.
Operator's Manual ZS-940PA i
(3) Clean the instrument together with all accessories for their next use.
6. The instrument must receive expert, professional attention for maintenance and repairs. When the instrument is not functioning properly, it should be clearly marked to avoid operation while it is out of order.
7. The instrument must not be altered or modified in any way.
8. Maintenance and Inspection:
(1) The instrument and parts must undergo regular maintenance inspection at least every 6
months.
(2) If stored for extended periods without being used, make sure prior to operation that the
instrument is in perfect operating condition.
(3) Technical information such as parts list, descriptions, calibration instructions or other
information is available for qualified user technical personnel upon request from your Nihon Kohden distributor.
9. When the instrument is used with an electrosurgical instrument, pay careful attention to the application and/or location of electrodes and/or transducers to avoid possible burn to the patient.
10. When the instrument is used with a defibrillator, make sure that the instrument is protected against defibrillator discharge. If not, remove patient cables and/or transducers from the instrument to avoid possible damage.
ii Operator's Manual ZS-940PA
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and workmanship for one year from the date of delivery. However, consumable materials such as recording paper, ink, stylus and battery are excluded from the warranty.
NKC or its authorized agents will repair or replace any products which prove to be defective during the warranty period, provided these products are used as prescribed by the operating instructions given in the operator’s and service manuals.
No other party is authorized to make any warranty or assume liability for NKC’s products. NKC will not recognize any other warranty, either implied or in writing. In addition, service, technical modification or any other product change performed by someone other than NKC or its authorized agents without prior consent of NKC may be cause for voiding this warranty.
Defective products or parts must be returned to NKC or its authorized agents, along with an explanation of the failure. Shipping costs must be pre-paid.
This warranty does not apply to products that have been modified, disassembled, reinstalled or repaired without Nihon Kohden approval or which have been subjected to neglect or accident, damage due to accident, fire, lightning, vandalism, water or other casualty, improper installation or application, or on which the original identification marks have been removed.
In the USA and Canada other warranty policies may apply.
CAUTION
United States law restricts this device to sale by or on the order of a physician.
Equipment Authorization Requirement
Operation of this equipment requires the prior coordination with a frequency coordinator designated by FCC for the Wireless Medical Telemetry Service.
Operator's Manual ZS-940PA iii
EMC RELATED CAUTION
This equipment and/or system complies with the International Standard IEC60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or system. However, an electromagnetic environment that exceeds the limits or levels stipulated in the IEC60601-1-2, can cause harmful interference to the equipment and/or system or cause the equipment and/or system to fail to perform its intended function or degrade its intended performance. Therefore, during the operation of the equipment and/or system, if there is any undesired deviation from its intended operational performance, you must avoid, identify and resolve the adverse electromagnetic effect before continuing to use the equipment and/or system.
The following describes some common interference sources and remedial actions:
1.Strong electromagnetic interference from a nearby emitter source such as an
authorized radio station or cellular phone: Install the equipment and/or system at another location if it is interfered with by an emitter source such as an authorized radio station. Keep the emitter source such as cellular phone away from the equipment and/or system.
2.Effect of direct or indirect electrostatic discharge:
Make sure all users and patients in contact with the equipment and/or system are free from direct or indirect electrostatic energy before using it.
3.Electromagnetic interference with any radio wave receiver such as radio or
television: If the equipment and/or system interferes with any radio wave receiver, locate the equipment and/or system as far as possible from the radio wave receiver.
If the above suggested remedial actions do not solve the problem, consult your Nihon Kohden Corporation subsidiary or distributor for additional suggestions.
iv Operator's Manual ZS-940PA
Conventions Used in this Manual and Instrument
Warnings, Cautions and Notes
Warnings, cautions and notes are used in this manual to alert or signal the reader to specific information.
WARNING
A warning alerts the user to the possible injury or death associated with the use or misuse of the instrument.
CAUTION
A caution alerts the user to possible injury or problems with the instrument associated with its use or misuse such as instrument malfunction, instrument failure, damage to the instrument, or damage to other property.
NOTE
A note provides specific information, in the form of recommendations, prerequirements, alternative methods or supplemental information.
Operator's Manual ZS-940PA v
Explanations of the Symbols in this Manual and Instrument
The following symbols found in this manual/instrument bear the respective descriptions as given.
On Panel
Symbol Description Symbol Description
Defibrillation proof type BF applied part
Defibrillation proof type CF applied part
Attention, consult operator’s manual
Direction for attaching battery cover
Inside Battery Case
Direct current
Serial number
Year of manufacture
CSA mark
Symbol Description Symbol Description
Battery position
On LCD
Attention, consult operator’s manual
Symbol Description Symbol Description
Full battery
Replace battery NIBP cannot be measured
Replace battery Check electrode
vi Operator's Manual ZS-940PA
Introduction
The ZS-940PA transmitter transmits ECG, respiration, SpO2, NIBP and pulse waveform from a patient to a Nihon Kohden monitor for continuous monitoring. The transmitter can change channels when connected to the QI-901PK channel writer. The front LCD displays SpO NIBP, pulse rate, pulse waveform amplitude, electrode condition mark, battery condition and NIBP measuring mode and interval.
Read the operator’s manual for the receiving monitor together with this manual before operation.
CAUTION
• Do not use the same channel for different patients. Otherwise, two patients’ data will be lost due to mutual modulation interference, or another patient’s data may appear on the receiving monitor screen.
• Do not use transmitters of adjacent channels in a hospital. Otherwise, radio waves from one transmitter affect the receiver of the adjacent channel’s transmitter and there may be interference.
NOTE
• To prevent interference between channels, assign a channel administrator in the hospital and only he or she should manage channel assignment.
• Use Nihon Kohden parts and accessories to assure maximum performance from your instrument.
• For stable signal reception, it is recommended to use a diversity antenna system on the receiving monitor. Otherwise, spike noise from transient fading of electric field strength (for example, people moving) may interfere with the transmitter signal and may be mistaken as an arrhythmia on the receiving monitor.
• NIBP cannot be measured on a neonate. (ECG, respiration and SpO monitored on a neonate.)
can be
2
%,
2
Operator's Manual ZS-940PA 1
Panel Description
Front Panel
LCD:
Displays measuring data, settings and other information.
NIBP START/STOP key:
Starts/stops NIBP measurement in selected mode.
NIBP INTERVAL key:
Selects NIBP measurement mode.
Refer to the WARNING below. (This symbol is attached to the rear of the battery case cover.)
Battery case:
Contains three 1.5 V AA batteries.
WARNING
Close the battery case cover during operation. If the transmitter is used with the battery case cover open, the patient may get an electrical shock when defibrillation is performed, and electrostatic discharge by the patient may intermittently interfere with the waveform or data.
2 Operator's Manual ZS-940PA
Rear Panel
Lock plate:
Fastens the transmitter to an NIBP cuff.
Refer to the WARNING below.
WARNING
If detergent or liquid spills into the transmitter, stop using it and contact your Nihon Kohden distributor. If a wet transmitter is used, the patient or anyone in contact with the transmitter may receive an electric shock or patient leakage current over the allowed amount may flow.
Operator's Manual ZS-940PA 3
Top Panel
NIBP socket:
Connects the cuff hose.
Refer to the WARNING on the next page.
Refer to the WARNING on the next page.
L
R
F
Refer to the symbol page.
Refer to the symbol page.
Bottom Panel
Channel number label:
Indicates the channel number of the transmitter. Attach the channel number label to the panel of the monitor.
Refer to the WARNING on the next page.
ECG/impedance RESP socket:
Connects the electrode lead for measuring ECG and/or respiration by the impedance method.
Refer to the symbol page.
SpO
socket:
2
Connects the SpO
2
probe.
4 Operator's Manual ZS-940PA
WARNING
• Before performing defibrillation, check that the electrode leads and SpO probe attached to the patient are properly connected to the transmitter. Touching the metal parts of disconnected leads and probes causes serious electrical shock or injury by discharged energy.
• When performing defibrillation, all persons must keep clear of the bed and must not touch the patient, any equipment connected to the patient or the metal parts of leads and probes connected to the patient. Failure to follow this warning may result in serious electrical burn, shock or other injury.
• When performing defibrillation, discharge as far as possible from electrodes and medicine on the chest of the patient. If there is a possibility that the defibrillator paddle could touch electrodes and medicine, remove electrodes and medicine from the patient. If the defibrillator contacts these materials, the discharged energy may cause serious electrical burn on the patient.
• When using this transmitter with an ESU, the ESU return plate and the electrodes for monitoring must be firmly attached to the patient. If the return plate is not attached correctly, it may burn the patient’s skin where the electrodes are attached. Refer to the instruction manual for the ESU.
2
CAUTION
Do not shake or swing the transmitter holding the leads/cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments.
Operator's Manual ZS-940PA 5
LCD
10
1234
LOCK KEYS
6
5
11
9
8
No. Name Description
1 NIBP SYS Displays NIBP systolic value. 2 NIBP DIA Displays NIBP diastolic value. 3 NIBP MEAN Displays NIBP mean value.
“CUFF” is displayed with the cuff inflation pressure during measurement.
4 Check electrode mark Appears when an electrode or electrode lead becomes
detached during ECG measurement.
5 Battery replacement mark Appears when the batteries are weak. For details, refer to
the “Battery Condition Indication” section. 6 Message display area Displays messages. 7 NIBP measurement mode Displays NIBP measurement mode. When set to auto mode,
the measurement interval is displayed. 8 NIBP interval bar graph In auto NIBP measurement, the remaining time from the last
measurement to the next measurement is displayed as a bar
graph. 9 Pulse level bar graph Displays pulse level in 7 steps.
10 %SpO
2
11 PR Displays pulse rate when NIBP or SpO
Displays SpO2 data.
the SpO
probe is attached to the patient, the real time pulse
2
rate is displayed. When the SpO
patient, the pulse rate at the end of NIBP measurement is
displayed.
7
is measured. When
2
probe is not attached to the
2
6 Operator's Manual ZS-940PA
Notes on Parameter Settings
When monitoring NIBP and SpO2, the following settings must be set as indicated in the table to properly transmit the monitoring data to the receiving monitor. Otherwise, data cannot be displayed on the receiving monitor even when NIBP and SpO
Some monitors require the software to be upgraded. For details, contact your Nihon Kohden distributor.
TYPE setting INHIBIT SpO2 DURING NIBP setting SpO2 probe attachment site
ON
A*
1
OFF
*1Always set to “A”.
2
*
When the SpO2 probe is attached to the same limb as the NIBP cuff and the cuff is inflated, the SpO
value becomes unstable and SpO2 or PR alarm may occur.
2
are measured on the transmitter.
2
Probe can be attached to the same limb as the cuff Probe must be attached to the limb without cuff*
2
Operator's Manual ZS-940PA 7
Important Safety Information
General
WARNING
• Never use this transmitter in the presence of any flammable anesthetic gas or high concentration oxygen atmosphere. Failure to follow this warning may cause explosion or fire.
• Never use this transmitter in a high-pressure oxygen medical care tank. Failure to follow this warning may cause explosion or fire.
• Do not take this transmitter into the MRI test room. This transmitter is not designed to be used during MRI tests.
• If detergent or liquid spills into the transmitter, stop using it and contact your Nihon Kohden distributor. If a wet transmitter is used, the patient or anyone in contact with the transmitter may receive an electric shock or patient leakage current over the allowed amount may flow.
• Before cleaning or disinfecting the transmitter, remove the batteries from the transmitter.
• The transmitter cannot be sterilized.
Output Signal
WARNING
Do not use the output signal from the receiving monitor as the synchronization signal for other equipment such as IABP, MRI, echocardiography or defibrillator because there may be time delay between the monitor and the other equipment caused by waveform transmission delay and spike noise may interfere on the output signal and be mistaken as a trigger.
8 Operator's Manual ZS-940PA
Preparation
Installing (Replacing) Batteries
WARNING and CAUTION for Battery Handling
WARNING
• Do not dispose of the battery in fire because it may explode.
• Do not use a disassembled or damaged battery. The contents of the battery are harmful and the battery may catch fire.
• If the contents of the battery contact the skin or clothes, immediately wash it thoroughly with water.
• Never short-circuit the + and – terminals. The battery may overheat and catch fire.
• Take care that the patient does not swallow batteries.
CAUTION
• Refer to the battery and battery charger manuals for details on handling the batteries.
• Do not handle the batteries with wet hands.
• When the transmitter is not in use, remove batteries. When batteries are installed, battery power is consumed even if measurement is not performed. Especially, when NiMH batteries remain in the transmitter when the transmitter is not in use, the battery may become unusable from overdischarge and leak liquid which will damage the transmitter.
• The battery charger must be used outside the patient environment.
Battery Lifetime
Use three AA type alkaline dry cell batteries. NiMH rechargeable batteries can also be used.
Type Lifetime (Measuring parameters)
ECG, SpO2, NIBP ECG, SpO
2
ECG only
NiMH secondary 2 days 2 days 2.5 days Alkaline primary 1 day 2.5 days 3 days
The above data is when new batteries are used at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO
is measured on an index finger of a male patient with
2
weight 60 kg. Operation time depends on the thickness of SpO
Operator's Manual ZS-940PA 9
probe attachment site.
2
NOTE
When using rechargeable NiMH batteries, shallow charging/discharging shortens battery capacity. For details, refer to the battery operator’s manual.
Installing (Replacing) Batteries
CAUTION
Battery replacement must be performed by medical staff. When replacing batteries of the transmitter currently used for a patient, disconnect electrode leads from the transmitter before replacing batteries or do not touch the patient during replacement.
If electrode leads are attached to the patient and a person replacing batteries touches the patient during battery replacement, patient leakage current over the allowed amount may flow.
CAUTION
• Replace all batteries at the same time.
• Do not use different types of batteries together.
NOTE
Insert the batteries with the correct polarity (+ and –).
Procedure
1. Remove the battery case cover.
2. Insert three new or fully charged batteries into the battery case observing the correct polarity.
10 Operator's Manual ZS-940PA
3. Close the cover.
NOTE
Remove the batteries before disposing of the transmitter.
Situations Requiring Battery Replacement
Replace the batteries when any of the following occurs.
• The transmitter LCD displays the “
• The transmitter generates a constant alarm (continuous “peep” sound).
• The transmitter LCD does not display anything when the power is turned on.
• The monitor displays the battery replacement message on the screen.
Battery Condition Indication
The battery condition is indicated as follows.
Indication Condition Receiving Monitor
Fully charged battery
Batteries are low. Replace batteries. Batteries are low. NIBP cannot be measured. Replace batteries.
No
indication
Dead batteries
” or “ ”mark.
Batteries are full. There is no indication on the monitor.
Message requiring battery replacement is displayed.
No signal can be transmitted to the monitor. There is no indication on the monitor.
Operator's Manual ZS-940PA 11
Turning the Transmitter On/Off
Turning On the Power
Peep
Check Items Before Use
Before turning on the transmitter power, check the following to confirm that the transmitter can be used in normal and safe condition.
When the batteries are installed correctly, the power is turned on. A one second “peep” sounds and the startup screen appears. (There is no “peep” sound when there is no battery power.)
NOTE
Replace the batteries when the LCD displays the “
Turning Off the Power
To turn off the power, remove batteries.
” or “ ”mark.
Appearance
• There are no damaged or dirty parts on the outside of the transmitter (Power switch, LCD, keys, sockets, battery case cover, battery case, lock plate, etc.).
• The transmitter is completely dry.
• The electrode lead, SpO
• There are no damaged or dirty parts on the disposable SpO
probe and NIBP cuff are not broken.
2
probe, disposable electrodes or
2
disposable NIBP cuff.
Batteries
• The battery polarity is correct.
• The battery case spring is firmly fixed and the battery is not loose.
• The battery case cover is firmly closed.
Channel Setting
• The transmitter channel corresponds to those of the receiving monitor.
• There is no transmitter in the surrounding area with the same channel.
12 Operator's Manual ZS-940PA
Check Items After Power On
After turning on the power, check the following.
Power On
• The Power switch is not broken.
• The transmitter generates about a one second “peep” sound and the startup screen appears.
• The transmitter does not generate a continuous “peep” sound.
• The transmitter does not give excessive heat.
• The transmitter LCD displays a “
” mark.
• The transmitter does not interfere with the operation of other medical instruments in use.
Basic Operation
• The “signal loss” message is not displayed on the receiving monitor when the transmitter is inside the receiving range of the monitor.
• The battery replacement message is not displayed on the monitor.
• The keys on the transmitter function properly.
Check Items After Use
To use the transmitter in safe and optimum condition for next time, check the following.
Before Turning Power Off
• Temporarily changed settings are changed back to the previous settings.
• There was no malfunction on the transmitter.
Storage
• ECG electrode leads, SpO
probe and NIBP cuff are cleaned and disinfected.
2
• When the transmitter gets wet, liquid is wiped off and the transmitter is thoroughly dried.
• There are enough consumables, such as disposable electrodes.
• The transmitter power is turned off.
• The batteries are removed from the transmitter when it will not be used for a long time.
• Dead batteries are disposed of properly.
Operator's Manual ZS-940PA 13
Changing the Transmitter Channel
The channel of the transmitter can be changed. The optional QI-901PK Channel Writer is required.
WARNING
The following action must be taken to properly receive the transmitter signal of the correct patient on the receiving instrument. Otherwise, there may be signal loss or signals may mix causing a serious accident, such as monitoring a different patient.
• Assign a channel administrator in the hospital and only he or she should manage channel assignment on his or her responsibility.
• The channel administrator must manage the channels in the facility so that there is no signal interference.
• When the transmitter channel is changed, the channel administrator must check that the channel on the receiving monitor is also changed and the signal is properly received.
• The channel administrator must replace the channel number label on the transmitter with the new one after changing the channel.
NOTE
The software version of the QI-901PK channel writer must be 02-01 or later to change the channel on the ZS-940PA transmitter.
To check the transmitter channel, refer to “CHANNEL/TYPE” in the “Changing System Setup Settings” section.
14 Operator's Manual ZS-940PA
Changing Parameter Setup Settings
The initial settings on the PARAMETER SETUP screen must be changed before monitoring. Changing these settings during monitoring interrupts monitoring.
Parameter Setup Setting List
The factory default settings are underlined.
Setting Item Description Settings
SELECTABLE INTERVALS
INITIAL INTERVAL
INITIAL CUFF PRESS
NIBP MODE AFTER STAT START/FINISH SOUND
OLD NIBP DATA AFTER
INHIBIT SpO DURING NIBP 2ND PARAMETER Set SpO2 and PR display order. SpO2, PR
2
Select the NIBP measurement modes for the mode selection.
Select the initial NIBP measurement mode at power on.
Select the NIBP cuff inflation pressure.
Select the NIBP measurement mode after completing STAT measurement. Turn ON or OFF the sound for NIBP measurement start/finish. Select whether to hide or dim the NIBP data after measurement and how long to wait after measurement to dim or hide it.
2
Turn SpO NIBP measurement.
monitoring on or off during
STAT, 5, 10, 15, 30, 60, 120, 240 MAN., 5 min, 10 min, 15 min, 30 min, 60 min, 120 min, 240 min 120 mmHg, 150 mmHg, 180 mmHg, 210 mmHg, 240 mmHg MAN., 5 min, 10 min, 15 min, 30 min
ON, OFF/ON, OFF
DATA: HIDE, DIM AFTER: 5 min, 10 min, 30 min
ON, OFF
Displaying the PARAMETER SETUP Screen
1. Remove one battery.
2. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery. The SETUP screen appears.
Operator's Manual ZS-940PA 15
3. Press the NIBP START/STOP key to enter the PARAMETER SETUP screen.
When the cursor is moved to “EXIT” by pressing the NIBP INTERVAL key and the NIBP START/STOP key is pressed, the startup screen appears, then the monitoring screen appears.
4. To select or change a setting, press the NIBP START/STOP key. To move the cursor, press the NIBP INTERVAL key.
Selects or changes setting
Moves cursor
When the cursor is moved to “RETURN” by pressing the NIBP INTERVAL key and the NIBP START/STOP key is pressed, the SETUP screen appears.
Changing Settings
SELECTABLE INTERVALS
During monitoring, when the NIBP INTERVAL key is pressed, the measurement mode changes according to the modes selected in this item. MANUAL mode is already selected for the mode selection.
1. Press the NIBP INTERVAL key to move the cursor to the desired mode.
2. Press the NIBP START/STOP key to select or unselect the mode. Selectable modes are: STAT, 5, 10, 15, 30, 60, 120 and 240 min.
16 Operator's Manual ZS-940PA
INITIAL INTERVAL
Select the initial NIBP measurement mode at power on.
1. Press the NIBP INTERVAL key to move the cursor to “INITIAL INTERVAL”.
2. Press the NIBP START/STOP key to select the mode. Selectable modes are the modes selected for “SELECTABLE INTERVALS” and “STAT” and “MAN.” (MANUAL).
INITIAL CUFF PRESS
Select the NIBP cuff inflation pressure.
1. Press the NIBP INTERVAL key to move the cursor to “INITIAL CUFF PRESS”.
2. Press the NIBP START/STOP key to select the inflation pressure from 120, 150, 180, 210 and 240 mmHg.
NIBP MODE AFTER STAT
Select the NIBP measurement mode after completing the STAT measurement.
1. Press the NIBP INTERVAL key to move the cursor to “NIBP MODE AFTER STAT”.
2. Press the NIBP START/STOP key to select the mode. The selected mode is automatically selected for “SELECTABLE INTERVALS” as well.
Operator's Manual ZS-940PA 17
START/FINISH SOUND
Turn on or off the sound for NIBP measurement start and finish.
Start sound
End sound
1. Press the NIBP INTERVAL key to move the cursor to “START/FINISH SOUND”.
2. Press the NIBP START/STOP key to turn ON or OFF.
OLD NIBP DATA/AFTER
Select whether to dim or hide the NIBP data after measurement and how long to wait after NIBP measurement to dim or hide it.
1. Press the NIBP INTERVAL key to move the cursor to “OLD NIBP DATA/AFTER”.
2. Press the NIBP START/STOP key to select the setting.
DATA: DIM NIBP data is dimmed after the “AFTER” interval.
HIDE NIBP data is hidden after the “AFTER” interval. “– – –” is displayed on the
screen.
AFTER: Select the interval after NIBP measurement to dim or hide.
Dimmed Hidden
18 Operator's Manual ZS-940PA
INHIBIT SpO2 DURING NIBP
Set whether or not to monitor SpO
during NIBP measurement.
2
When the SpO OFF, the pulse may become unstable and SpO set this setting to ON so that SpO
When the SpO
probe is attached to the same limb as the NIBP cuff and this setting is set to
2
or PR alarm may occur. It is recommended to
2
is not measured during NIBP measurement.
2
probe is attached to the other limb from the NIBP cuff, this setting can be set to
2
OFF.
NOTE
• When the “CHANNEL/TYPE” on the SYSTEM SETUP screen is set to “8”, this “INHIBIT SpO during NIBP measurement.
• When this “INHIBIT SpO SpO
during NIBP Measurement” section.
2
ON: Stops SpO OFF: SpO
2ND PARAMETER
Set the display order of SpO
DURING NIBP” setting becomes invalid and SpO2 is monitored
2
DURING NIBP” is set to OFF, refer to the “Monitoring
2
1. Press the NIBP INTERVAL key to move the cursor to “INHIBIT SpO
DURING NIBP”.
2
2. Press the NIBP START/STOP key to select “ON” or “OFF”.
monitoring during NIBP measurement.
2
is monitored during NIBP measurement.
2
and PR.
2
1. Press the NIBP INTERVAL key to move the cursor to “2ND PARAMETER”.
2. Press the NIBP START/STOP key to select “SpO2” or “PR”.
When set to SpO
Operator's Manual ZS-940PA 19
2
When set to PR
Changing System Setup Settings
NOTE
Changing System Setup settings must be done only by a qualified personnel.
System Setup Setting List
The factory default settings are underlined.
Setting Item Description Settings
CHANNEL/TYPE
PRESSURE UNIT Select the units for NIBP. mmHg, kPa LANGUAGE Select the language for screen display. JPN, ENG SYSTEM INITIALIZE
Displays the transmitter channel and select the TYPE.
Initializes all settings to the factory default settings.
Displaying the SYSTEM SETUP Screen
1. Remove one battery.
2. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery. The SETUP screen appears.
A, 8
3. Press the NIBP INTERVAL key to move the cursor to “SYSTEM SETUP”.
4. Press the NIBP START/STOP key to enter the SYSTEM SETUP screen.
When the cursor is moved to “EXIT” by pressing the NIBP INTERVAL key and the NIBP START/STOP key is pressed, the startup screen appears, then the monitoring screen appears.
20 Operator's Manual ZS-940PA
5. To select or change a setting, press the NIBP START/STOP key. To move the cursor, press the NIBP INTERVAL key.
Selects or changes setting
Moves cursor
When the cursor is moved to “RETURN” by pressing the NIBP INTERVAL key and the NIBP START/STOP key is pressed, the SETUP screen appears.
Changing Settings
CHANNEL/TYPE
Select “A” for TYPE. The channel of this transmitter is also displayed.
1. On the SYSTEM SETUP screen, press the NIBP START/STOP key to select “A”.
Channel of this transmitter
NOTE
Always select “A” for TYPE. If “8” is selected, NIBP data cannot be properly transmitted to the receiving monitor.
PRESSURE UNIT
Select the unit for NIBP.
1. Press the NIBP INTERVAL key to move the cursor to “PRESSURE UNIT”.
2. Press the NIBP START/STOP key to select “mmHg” or “kPa”.
Operator's Manual ZS-940PA 21
LANGUAGE
Select the language for screen display.
1. Press the NIBP INTERVAL key to move the cursor to “LANGUAGE”.
2. Press the NIBP START/STOP key to select the language.
SYSTEM INITIALIZE
Do the following procedure to initialize the settings to the factory default settings.
1. Press the NIBP INTERVAL key to move the cursor to “SYSTEM INITIALIZE”.
2. Press the NIBP START/STOP key. The “EXECUTE” message appears.
3. Press the NIBP START/STOP key to initialize the settings to the factory default settings.
22 Operator's Manual ZS-940PA
Attaching NIBP Cuff, Electrodes and SpO2 Probe
to the Patient
The transmitter can be attached to an arm of the patient or placed on the bedside. The required length of the electrode leads and SpO attached to the patient.
Monitoring SpO2 during NIBP Measurement When the SpO
probe is attached to the same limb as the NIBP cuff, the blood
2
flow decreases during NIBP measurement and pulse wave cannot be detected and SpO
cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on
2
the PARAMETER SETUP screen is set to ON (factory default setting), SpO monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO careful when reading SpO
probe cable depends on how the transmitter is to be
2
NOTE
on the same limb as NIBP, be
2
values.
2
2
When monitoring SpO
is important, attach the probe to the limb to which the
2
NIBP cuff or catheter is not attached.
Attachment Examples
When transmitter is attached on an arm
When placing the transmitter on a bedside, place it on a stable and flat place. If the transmitter falls off, it may be damaged.
Operator's Manual ZS-940PA 23
When transmitter is placed on a bedside
NOTE
Attaching the NIBP Cuff
Selecting the NIBP Cuff
Select the NIBP cuff appropriate for the patient.
NOTE
NIBP cannot be measured on neonates using this transmitter.
Reusable Cuffs
When attaching the transmitter to the patient arm, a special NIBP cuff is required. An optional YN-990P extension hose (1.5 m) is available to extend the length between the NIBP socket on the transmitter and NIBP cuff (e.g. when not attaching the transmitter to the patient arm and placing the transmitter on a bedside).
Reusable cuff Model Width (cm) Air hose length (cm)
For adult
Standard YP-943P 13 15
Large YP-944P 15 15
Air hose
Width
When not attaching the transmitter to the patient arm, the following cuffs can be used. To use these cuffs, an optional YN-990P extension hose (1.5 m) is required.
Reusable cuff Model Width (cm) Air hose length (cm)
For infants YP-960T 5
For children
For adults
24 Operator's Manual ZS-940PA
Small YP-961T 7
Standard YP-962T 10
Standard YP-963T 13
Large YP-964T 15
Width
Air hose
15
Disposable Cuffs
CAUTION
The disposable cuffs are not sterilized. If necessary, sterilize the disposable cuffs using glutaraldehyde solution by following the instructions for the glutaraldehyde.
When not attaching the transmitter to the patient arm, the following disposable cuffs can be used. To use these cuffs, an optional YN-990P extension hose (1.5 m) is required.
Reusable cuff Model Width (cm) Air hose length (cm)
For infants YP-910P 6
For children YP-912P 9
For adults
Small YP-913P 12
Standard YP-914P 14
Large YP-915P 16
Width
Air hose
20
Extension Hose
CAUTION
When using an extension hose, check that the extension hose is not bent or squeezed. Otherwise, the cuff may not inflate or deflate. If the cuff cannot deflate, it may cause congestion on the patient at the cuff attachment site.
YN-990P extension hose, 150 cm
Operator's Manual ZS-940PA 25
Reference for selecting a cuff
The AHA (American Heart Association) recommends that the cuff width be 40% of the
circumference of the upper arm. Refer to the following graph and select the cuff which suits the
patient’s arm.
NOTE
• If a range of arm circumference appropriate for the cuff is prescribed, use a cuff within that range.
• To obtain accurate measured values, select a wide cuff which can be attached to the upper arm. Measuring with a very narrow cuff may result in measured values higher than the actual values.
• The YP-943P NIBP cuff is for standard size adult. Do not use this cuff when it does not fit the patient.
Cuff Width and Arm Circumference
Reusable Cuffs
Cuff width (cm)
15
10
Children small YP-961T
5
Infants YP-960T
Adults large YP-944P (cuff for transmitter)
YP-964T
Adults standard YP-943P (cuff for transmitter)
YP-963T
Children standard YP-962T
Disposable Cuffs
Cuff width (cm)
0
20
15
10
5
0
10 20 30
Children standard YP-912P
Infants YP-910P
10 20 30
40
Adults large YP-915P
Adults standard YP-914P
Adults small YP-913P
40
50
Arm circumference (cm)
50 60
Arm circumference (cm)
60
26 Operator's Manual ZS-940PA
Connecting the NIBP Cuff to the Transmitter
When Using YP-943P/944P NIBP Cuff
To attach the YP-943P/944P NIBP cuff to the transmitter, the lock plate is required.
YP-943P/944P NIBP cuff
Lock plate
Front cover
Belt for the strap
Lock plate pocket
Top tab
For attaching the NIBP cuff to the transmitter
Air hose
Front cover open
NOTE
Do not roll up or put weight on the cuff with the lock plate attached to it. Otherwise, the lock plate may break.
D ring
Belt
Bottom tab
Operator's Manual ZS-940PA 27
1. Remove the lock plate from the transmitter.
2. Insert the lock plate into the lock plate pocket on the NIBP cuff.
3. Attach the transmitter to the lock plate by inserting the tabs on the lock plate into the slots on the transmitter.
4. Cover the transmitter with the front cover of the NIBP cuff.
5. Connect the air hose to the NIBP socket on the transmitter. Turn the cuff connector joint until it clicks.
2
1
28 Operator's Manual ZS-940PA
When Using YP-960T series or YP-910P series NIBP Cuff
To use these NIBP cuffs, an optional YN-990P extension hose (1.5 m) is required.
NOTE
Connect the joints properly. If there is an air leak, NIBP cannot be measured properly.
1. Connect the NIBP cuff to the extension hose.
2. Connect the other end of the extension hose to the NIBP socket on the transmitter. Turn the joint clockwise until it clicks.
To disconnect the cuff from the transmitter, turn the hose joint counterclockwise.
Attaching the NIBP Cuff to the Patient
WARNING
Be careful when measuring NIBP on a patient with known bleeding disorders or coagulation. After NIBP measurement, there may be dot hemorrhage, or circulatory disorder by thrombus where the cuff was attached.
CAUTION
• Do not wrap the cuff on an arm or thigh which is used for injection. NIBP measurement on an arm or thigh which is used for injection may cause a reflux of blood and stop injection.
• Do not wrap the cuff too tight. It may cause poor blood circulation and congestion. If the cuff is wrapped too loosely, the NIBP value may be increased.
• If the skin gets irritated or redness appears on the skin from the cuff, change the attachment site or stop using the cuff.
Operator's Manual ZS-940PA 29
• NIBP and SpO2 can be measured on the same limb, but the SpO2 monitoring may not be accurate during NIBP measurement. Be careful when reading the SpO
values.*
2
• Do not reuse disposable cuffs.
* Monitoring SpO2 during NIBP Measurement
When the SpO during NIBP measurement and pulse wave cannot be detected and SpO properly. When “INHIBIT SpO set to ON (factory default setting), SpO avoid SpO NIBP, be careful when reading SpO
probe is attached to the same limb as the NIBP cuff, the blood flow decreases
2
cannot be monitored
2
DURING NIBP” on the PARAMETER SETUP screen is
2
monitoring is paused during NIBP measurement to
2
alarm occurrence. However, when monitoring SpO2 on the same limb as the
2
values.
2
NOTE
• Measuring NIBP at a site other than the upper arm gives different values from those measured at the upper arm. When making diagnosis based on the NIBP values, measure NIBP on an upper arm.
• To accurately detect the pulsatile flow of the artery, the cuff should be wrapped around a bare upper arm.
• Do not use an abnormal cuff. The cuff deteriorates from use and cleaning. Before use, check the cuff and confirm that there is no flaw, crack or hole in it. Be careful not to damage the inflation bag. If the inflation bag has a hole or a flaw, it may burst during use. Dispose of an abnormal cuff and replace it with a new one.
• Refer to the NIBP cuff manual for details.
Cuff Position
Place the cuffed upper arm (brachium) at the same height as
When placing the transmitter on a bed, make sure that the hose is not bent.
the patient’s heart. If the cuff is not at the same level as the heart, the weight of the blood affects the blood pressure reading. The pressure difference per unit height is 0.7 mmHg/cm. The blood pressure reading decreases when the arm is higher than the heart and increases when lower.
Heart
The best measuring condition is when the patient is lying on his/her back with arms and legs relaxed. If the cuff position cannot be on the same level as the heart, the displayed blood pressure reading must be mathematically adjusted.
30 Operator's Manual ZS-940PA
Using the YP-943P/944P NIBP Cuff
Belt
1. Attach the NIBP cuff to the transmitter. Refer to the “Connecting the NIBP Cuff to the Transmitter” section.
2. Insert the end of the cuff into the belt and then through the D ring as shown at left.
D ring
End of cuff
3. Fold back the cuff at the D ring and fasten it using the velcro tape.
Make sure that the cuff is not attached on a joint.
NOTE
The cuff must not wrap around the elbow.
Operator's Manual ZS-940PA 31
Attaching the Strap to the Transmitter
NOTE
• Use the strap to prevent the transmitter from falling.
• Do not attach the clip to hard objects such as thick cloth or zipper. It will break the clip.
Attach a strap provided with the transmitter to the NIBP cuff and patient clothes.
To open the clip, firmly pull out the tab in direction of the arrow.
To adjust the strap length, push down the tab on the adjuster and slide.
Belt for the strap on the NIBP cuff
1. Adjust the length of the strap.
2. Clip one end of the strap to the belt for the strap on the NIBP cuff.
3. Clip the other end of the strap to the patient’s clothes as shown left.
32 Operator's Manual ZS-940PA
Using the YP-960T series Reusable Cuffs or YP-910P series Disposable Cuffs
ARTERY
1. Put the cuff on the upper arm so that the
mark of “ARTERY ” aligns with the artery
of the patient.
2. Wrap the cuff so that “INDEX ” comes
within the “ RANGE ”.
RANGE
If “Index
”, change the cuff size.
” is not within the “ RANGE
Attaching Electrodes
Selecting Electrode Lead
CAUTION
Use Nihon Kohden specified electrodes and electrode leads. With electrodes and electrode leads other than specified ones, the “CHECK ELECTRODE” message appears and monitoring may stop.
The following electrode leads can be used on the transmitter (option).
BR-903PA, 3 electrodes, clip type
Operator's Manual ZS-940PA 33
BR-913PA, 3 electrodes, snap type
BR-902PA, 2 electrodes, clip type
BR-912PA, 2 electrodes, snap type
Connecting the Electrode Lead to the Transmitter
Connect the electrode lead to the ECG/RESP socket on the transmitter.
When the transmitter is attached on an arm
CAUTION
Hold the connector of the electrode lead when connecting/disconnecting the electrode lead. If you disconnect the electrode lead by pulling the lead, it damages the electrode lead.
Selecting the Electrode Position
Follow the physician’s instructions for electrode placement when available. For ECG monitoring, electrodes are attached only on the chest to allow patient movement and obtain continuous stable ECG. Following leads are examples. When also monitoring respiration, refer to the “Electrode Position for Respiration Monitoring” section.
NOTE
The optimum electrode positions for ECG measurement of a patient are not always optimum for respiration measurement of the patient. Select positions suitable for both ECG and respiration measurements, or positions which have priority for one measurement.
34 Operator's Manual ZS-940PA
Three Electrodes
• Lead MII, which is similar to standard lead II, used when ECG measurement has priority
Electrode Position
Symbol Lead Color
AHA IEC AHA IEC
Left infraclavicular fossa LA L Black Yellow Right infraclavicular fossa RA R White Red Below lowest rib on the left anterior axillary line
• Lead MI, which is similar to standard lead I Change F/LL and L/LA of the lead
LL F Red Green
• Lead MIII, which is similar to standard lead III. Change R/RA and L/LA of the lead MII.
MII.
If the electrode position shown above is not available due to chest surgery, attach the electrodes to the root of the limbs or below the clavicles for stable ECG monitoring.
Operator's Manual ZS-940PA 35
Two Electrodes
With the optional BR-912PA and BR-902PA electrode leads, measurement with two electrodes is available. L/LA electrode is not used. This is effective for a neonate or a patient whose body area is small and difficult to attach three electrodes.
(ex.) Lead MII, which is similar to standard lead II
Electrode Position
Right infraclavicular fossa RA R White Red Below lowest rib on the left anterior axillary line
Difference between measurement with two electrodes and three electrodes Measurement with two electrodes is less stable than measurement with three electrodes because of noise and body movement. Pay attention to this point. If ECG of necessary quality cannot be obtained, measure with three electrodes.
Attaching Electrodes to the Patient and Connecting the Electrode Leads to Disposable Electrodes
Prepare the Patient Skin
Shave off excessive body hair. To reduce skin impedance, clean the electrode site with cream or with a cotton pad moistened with alcohol. Thoroughly dry the skin with a clean cotton pad.
Symbol Lead Color
AHA IEC AHA IEC
LL F Red Green
NOTE
• For a patient with frequent body movement, rub the sites with Skinpure skin preparation gel. However, do not use Skinpure skin preparation gel on sensitive skin.
• Do not place electrodes on a wound or on an inflamed, wrinkled or uneven skin surface.
36 Operator's Manual ZS-940PA
Attaching Electrodes to the Patient
CAUTION
Do not reuse disposable products.
NOTE
• To maintain good contact between the electrode and skin, check that the paste of the disposable electrode is not dry.
• When contact between the disposable electrode and skin becomes poor, replace electrodes with new ones immediately. Otherwise, contact impedance between the skin and the electrode increases and the correct ECG cannot be obtained.
Refer to the electrode operator’s manual for details.
1. Carefully remove the backing paper from the electrode. Avoid touching the adhesive surface.
2. Place the electrode on the previously cleaned skin. Pay attention to the electrode lead color and symbol.
3. Clip the electrode lead to the electrode.
4. Fasten the electrode lead wire with surgical tape with an extra length of wire between the tape and the electrode. This lessens the movement of electrode leads by body movement and helps stable monitoring.
Operator's Manual ZS-940PA 37
Electrode Position for Respiration Monitoring
Place the R/RA and F/LL electrodes so that the lungs are between the electrodes.
NOTE
The optimum electrode positions for ECG measurement of a patient are not always optimum for respiration measurement of the patient. Select positions suitable for both ECG and respiration measurements, or positions which have priority for one measurement.
Electrode Position Examples
Position 1 In this position, respiration measurement is available; however, there is a difference in amplitude between different patients.
R or RA F or LL
Right infraclavicular fossa
Position 2 In this position, the waveform amplitude is usually large and the ECG lead is similar to Lead MII. This position can be generally recommended.
R or RA F or LL
Right infraclavicular fossa
Fifth intercostal space on the left midclavicular line, V4
Fifth intercostal space on the left midaxillary line, V6
Position 3 In this position, the respiration waveform is optimum, but the ECG lead is unusual.
R or RA F or LL
Right midaxillary at the horizontal level of V4
38 Operator's Manual ZS-940PA
Fifth intercostal space on the left midaxillary line, V6
Position 4 In this position, the respiration measurement is influenced by the impedance variation of the abdomen, so the cardiac pulse wave included in the respiration wave is reduced. Note that the waveform is inverted in phase compared with the chest movement (the waveform goes down during inspiration). It is difficult to measure the ECG at the same time.
R or RA F or LL
Lowest rib on the right anterior axillary line
Lowest rib on the left anterior axillary line
Attaching the SpO2 Probe
Selecting the SpO2 Probe
Select an appropriate probe for the patient.
CAUTION
• Use Nihon Kohden specified SpO2 probe to assure maximum performance from your instrument.
• Do not use damaged or disassembled probe. It causes incorrect measurement and may hurt the patient.
Reusable Probes
When using a TL-201T finger probe, choose the appropriate cable length for attachment.
Probe Cable Length Patient Attachment site
Finger probe TL-201T
0.6 m
Adult or children 20 kg or more
Finger
1.6 m
Attachment tape
Adult or Infant
Finger or toeMulti-site probe TL-220T
3 kg or more
Neonate
Instep and sole
3 kg or less
Operator's Manual ZS-940PA 39
Disposable Probes
CAUTION
• Use the disposable probe only for one patient. Never reuse the disposable probe for another patient because it causes cross infection.
• Disposable probes are not sterilized.
Probe Patient Attachment site
TL-251T Adult
30 kg or more
Finger or toe
TL-252T Child
3 to 40 kg
TL-253T Neonate
3 kg or less
Multi-site Y probe TL-260T
Attachment tape
Low birth weight infant 1 kg or less Neonate or Child 3 kg or more Neonate 3 kg or less Adult 50 kg or more
40 mm
Neonate 3 kg or less
Cable length TL-051S: 80 cm
TL-052S: 160 cm
Child or Adult 15 to 50 kg
Finger or toe
Instep and sole
Instep and sole
Attachment tape S
Finger or toe Attachment
tape S Instep and sole
Attachment
tape L FingerTL-051S/052S
Instep and sole
FingerTL-061S/062S
35 mm
Infant
Toe
3 to 15 kg
Cable length TL-061S: 80 cm
TL-062S: 160 cm
40 Operator's Manual ZS-940PA
Connecting the SpO
Connect the probe to the SpO
Probe to the Transmitter
2
socket on the transmitter.
2
When the transmitter is attached on an arm
CAUTION
Hold the connector when connecting/disconnecting the probe. If you disconnect the SpO
probe by pulling the cable, it damages the cable.
2
Attaching the Probe to the Patient
Attach the probe to the patient by referring to the probe’s manual. Make sure that the light emitter and photo detector of the probe face each other at the attachment site.
WARNING
• When using a TL-201T finger probe, do not fasten it to a finger by wrapping the probe to the site with some tape. It may cause poor blood circulation, congestion, pressure necrosis or burn.
• When using probes other than a TL-201T finger probe, do not wrap the tape too tight. Check the blood circulation condition by observing the skin color and congestion at the skin peripheral to the probe attachment site. Even for short­term monitoring, there may be burn or pressure necrosis from poor blood circulation, especially on neonates or low birth weight infants whose skin is delicate. Accurate measurement cannot be performed on a site with poor peripheral circulation.
Operator's Manual ZS-940PA 41
• Check the circulation condition by observing the skin color of the measuring site and the pulse waveform. Change the measuring site every 8 hours for disposable probes and every 4 hours for reusable probes. The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree.
• A patient with a fever
• A patient with a peripheral circulation insufficiency
• Neonate or low birth weight infant with delicate skin
CAUTION
• If the attachment site is dirty with blood, clean the attachment site before attaching the probe. If there is nail polish on the attachment site, remove the polish. Otherwise, the amount of transmitted light decreases and the measured data may be incorrect or measurement cannot be performed.
• If the skin gets irritated or redness appears on the skin from the probe, change the attachment site or stop using the probe.
• When the probe is attached on an appropriate site with sufficient circulation, but the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one.
• Do not use a probe that is deteriorated by aging. Accurate measurement cannot be performed.
• When using probes other than a TL-201T finger probe on a neonate, be careful when removing the adhesive tape from neonatal skin.
• When removing a probe taped to the skin, do not pull the cable. Otherwise the cable may break.
• When removing the probe from the attachment tape, do not pull the sensor cable. Otherwise the cable may get damaged.
• Before using the TL-260T multi-site Y probe, be sure to attach the probe to the sponge attachment tape S or L. Do not use the probe without the sponge attachment tape attached. It causes incorrect measurement and may damage the attachment site on the skin.
• When fixing the TL-260T multi-site Y probe with the sponge attachment tape, confirm that the adhesive part of the tape is not on the skin. The adhesive may cause oversensitive symptoms on the skin such as redness or itch. If the adhesive touches the skin, remove it carefully and slowly because neonatal skin is very delicate.
• Do not use a dirty sponge attachment tape. The measurement value may be incorrect.
• Refer to the probe instruction manual for details.
42 Operator's Manual ZS-940PA
Locking the Keys on the Transmitter
To prevent the patient from pressing the keys on the transmitter during monitoring, you can lock the NIBP START/STOP and NIBP INTERVAL keys.
1. Press the NIBP INTERVAL key for about 3 seconds.
2. A “pip, pip” sounds and the “LOCK KEYS” message is displayed on the LCD.
Pip, pip
LOCK KEYS
When the NIBP START/STOP key or NIBP INTERVAL key is pressed while the keys are locked, the “PRESS INT. KEY 3S TO UNLOCK” message appears.
To unlock the keys:
1. Press the NIBP INTERVAL key for about 3 seconds.
2. A “pip, pip” sounds and the keys are unlocked. The “UNLOCK KEYS” message appears and the keys are unlocked.
Operator's Manual ZS-940PA 43
Monitoring
When preparation is done, monitoring starts.
NIBP Oscillometric Method
NIBP is measured from the change in amplitude pattern of pulsatile oscillation in cuff pressure as the cuff pressure is reduced from above systolic to below diastolic pressure. The occlusive­oscillometry method uses this to determine the systolic, diastolic and mean arterial pressure.
NIBP Monitoring
Selecting the Initial Cuff Inflation Pressure
The initial cuff inflation pressure can be changed on the PARAMETER SETUP screen. The default setting is 180 mmHg. To change the setting, refer to the “Changing Parameter Setup Settings” section.
Selecting the Measurement Mode and Interval
Measurement Modes
There are three measurement modes: manual, auto and STAT. The selected mode or interval is displayed at the lower right of the screen.
The measurement mode and interval can be changed by pressing the NIBP INTERVAL key. When the key is pressed, the measurement mode changes according to the modes selected at “SELECTABLE INTERVALS” on the PARAMETER SETUP screen. MANUAL mode is already selected for the mode selection.
To select the modes for the mode selection, refer to the “Changing Parameter Setup Settings” section.
Manual Measurement
In Manual mode, a single NIBP measurement is performed when the NIBP START/STOP key is pressed.
STAT (Continuous) Measurement
In STAT mode, measurement is continuously repeated for 15 minutes after the NIBP START/ STOP key is pressed.
When the STAT measurement for 15 minutes is completed, the measurement mode automatically changes to the Manual mode or Auto mode of selected interval depending on the “NIBP MODE
44 Operator's Manual ZS-940PA
AFTER STAT” setting on the PARAMETER SETUP screen. The default setting is Manual mode. Refer to the “Changing Parameter Setup Settings” section.
The STAT measurement completes within 15 minutes. When more than 12 minutes elapse from the start of measurement, there will be no more measurement performed and the measurement mode changes to the mode selected for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen.
Auto Measurement
In Auto mode, measurement is performed automatically at the preset time intervals.
In Auto mode, a single measurement can be performed by pressing the NIBP START/STOP key between auto measurements.
Measuring NIBP
WARNING
• Be careful when measuring NIBP on a patient with known bleeding disorders
or coagulation. After NIBP measurement, there may be dot hemorrhage, or circulatory disorder by thrombus where cuff is attached.
• NIBP measurement may be incorrect in the following cases.
• When using an electrical surgery unit.
• When there is body movement.
• When the pulse wave is small (insufficient peripheral circulation).
• Too many arrhythmia.
• When there is vibration.
• When there is a rapid blood pressure change.
• During CPR.
• When performing NIBP measurements in STAT mode or 5 minute intervals,
periodically remove the cuff from the patient for ventilation. Otherwise, the skin temperature may increase at the cuff attachment site by 2 or 3°C (4 or 5°F). When measuring a patient with a fever or peripheral circulation insufficiency, it may cause a burn.
CAUTION
When performing NIBP measurement repeatedly, have a rest between measurements to recover adequate circulation.
Operator's Manual ZS-940PA 45
NOTE
• When measuring patients who are conscious, help the patient to relax. Measurement may not be accurate if the patient’s arm is tense or if the patient talks.
• The data for measurement on a leg tends to be higher than measurement on the arm. When making diagnosis based on the NIBP values, measure NIBP on an upper arm.
• Do not apply pressure to the cuff or air hose. NIBP may not be measured correctly because of noise or NIBP measurement may stop due to the NIBP safety circuit.
• When the transmitter is attached to the patient arm and the NIBP measurement is performed when moving, tell the patient to relax and keep quiet. Otherwise, measurement may be stopped or remeasurement is repeated due to body movement.
• If there is an abnormal noise generated during measurement, stop using the transmitter and contact your Nihon Kohden distributor.
• Do not measure NIBP of a patient on whom an IABP is being used. Measurement may be incorrect due to the mixing of the patient’s own pulse and IABP pulse.
• NIBP cannot be measured on a neonate using this transmitter.
1. Select the measurement mode by pressing the NIBP INTERVAL key.
2. Press the NIBP START/STOP key to perform measurement.
46 Operator's Manual ZS-940PA
The cuff is inflated and the inflation pressure is displayed on the screen.
Inflation pressure
In manual mode: Measurement is performed once. In STAT mode: Measurement is performed repeatedly for 15 minutes. In auto mode: The first measurement is performed when the NIBP START/STOP key
is pressed. The second measurement is performed when the current time in the transmitter reaches the selected time interval.
To stop measurement during measurement, press the NIBP START/STOP key again.
In STAT mode, after completing the STAT measurement, the measurement mode changes to the mode set for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen.
In auto mode, to stop measurement in auto mode, change the mode to manual. To cancel one measurement, press the NIBP START/STOP key during measurement.
After the measurement is complete, the measured data is displayed on the screen and is transmitted to the monitor.
When SpO2 is not monitored, the pulse rate at the end of NIBP measurement is displayed.
During auto mode measurement, the measurement mode can be changed. During the interval, press the NIBP INTERVAL key to change the mode. When “MANUAL” is displayed for more than one second, the measurement in auto mode is stopped.
A buzzer can be set to sound at the start and end of NIBP measurement. Refer to the “Changing Parameter Setup Settings” section.
Operator's Manual ZS-940PA 47
Data Display After NIBP Measurement
When the time set at “OLD NIBP DATA” on the PARAMETER SETUP screen elapses after the last measurement, the NIBP data is dimmed or hidden. Whether to dim or hide the old data can also be selected at “OLD NIBP DATA”. Refer to the “Changing Parameter Setup Settings” section.
Monitoring SpO
When the SpO during NIBP measurement and pulse wave cannot be detected and SpO properly. When “INHIBIT SpO to ON (factory default setting), SpO SpO
alarm occurrence. However, when monitoring SpO2 on the same limb as the NIBP, be
2
careful when reading SpO
during NIBP Measurement
2
probe is attached to the same limb as the NIBP cuff, the blood flow decreases
2
cannot be monitored
2
DURING NIBP” on the PARAMETER SETUP screen is set
2
monitoring is paused during NIBP measurement to avoid
2
values.
2
ECG and Respiration Monitoring
When the electrodes are attached and the ECG leads are connected to the electrodes, heart rate, ECG, respiration rate and respiration waveform appear on the monitor.
WARNING
Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac monitoring and Diagnostic Equipment* The bioelectric impedance measurement sensor of a minute ventilation rate­adaptive implantable pacemaker may be affected by the transmitter which is connected to the same patient. If this occurs, the pacemaker may pace at its maximum rate and the transmitter may give incorrect data to the monitor. If this occurs, disconnect the electrode leads from the patient or change the setting on the pacemaker by referring to the pacemaker’s manual. For more details, contact your pacemaker distributor or Nihon Kohden distributor.
* Minute ventilation is sensed in rate-adaptive pacemakers by a technology known as bioelectric
impedance measurement (BIM). Many medical devices in addition to pacemakers use this technology. When one of these devices is used on a patient with an active, minute ventilation rate-adaptive pacemaker, the pacemaker may erroneously interpret the mixture of BIM signals created in the patient, resulting in an elevated pacing rate.
For more information, see the FDA web site. http://www.fda.gov/cdrh/safety.html
48 Operator's Manual ZS-940PA
WARNING
When using this transmitter with an ESU, the ESU return plate and the electrodes for monitoring must be firmly attached to the patient. If the return plate is not attached correctly, it may burn the patient’s skin where the electrodes are attached. Refer to the instruction manual for the ESU.
CAUTION
Turn off the power of cell telephones, small wireless devices and other devices which produce strong electromagnetic interference. Otherwise, the waveforms and measurements are affected by such interference and the displayed data may be incorrect.
NOTE
• Noise generated from an electrosurgery unit may interfere on an ECG waveform, but will not damage it.
• If an electric blanket is used and incorrect heart rate is displayed on the monitor, turn off the pacing spike detection on the monitor.
• Turn the pacing spike detection to ON on the monitor when monitoring a pacemaker patient. Pacing pulse is detected by the transmitter and transmitted to the monitor. If the pacing spike detection is turned OFF, QRS and pacemaker spike may not be distinguished and pacemaker failure may not be recognized.
Electrode Detachment
The “
message is displayed on the screen of the monitor in the following conditions.
• Electrode is detached from skin.
• Electrode lead is disconnected from the electrode.
• Polarization voltage between the electrode and skin is excessively high.
In these cases, check the cause and if necessary, replace electrodes with new ones.
” mark is displayed on the LCD of the transmitter or the “CHECK ELECTRODE”
CAUTION
When the “ELECTRODE OFF” or “CHECK ELECTRODE” message is displayed on the receiving monitor, check electrodes and electrode leads and remove the cause. While the “ELECTRODE OFF” or “CHECK ELECTRODE” message is being displayed, there is no ECG monitoring and no alarms.
Operator's Manual ZS-940PA 49
SpO2 Monitoring
When monitoring starts, SpO2 and pulse waveform are sent to the monitor and SpO2, pulse rate and pulse level bar graph are displayed on the transmitter LCD.
WARNING
• Measurement may be incorrect in the following cases.
• When the patient’s carboxyhemoglobin or methemoglobin increases abnormally.
• When dye is injected in the blood.
• When using an electrical surgery unit.
• During CPR.
• When there is body movement.
• When there is vibration.
• When measuring at a site where there are venous pulses.
• When the pulse wave is small (insufficient peripheral circulation).
• When using an IABP (intra-aortic balloon pump).
• Check the circulation condition by observing the skin color of the measuring site and pulse waveform. Change the measuring site every 8 hours for disposable probes and every 4 hours for reusable probes. The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree.
• A patient with a fever
• A patient with peripheral circulation insufficiency
• Neonate or low birth weight infant with delicate skin
• When not monitoring SpO Otherwise, noise may interfere from the probe sensor and incorrect data is displayed on the screen.
, disconnect the SpO2 cable from the transmitter.
2
CAUTION
• Turn off the power of cell telephones, small wireless devices and other devices which produce strong electromagnetic interference. Otherwise, the waveforms and measurements are affected by such interference and the displayed data may be incorrect.
• Normally external light does not affect monitoring, however, strong light such as an operating lamp or sunlight may affect monitoring. If affected, cover the measuring site with a blanket.
• Do not pull or bend the probe cable, and do not run over the probe cable with
50 Operator's Manual ZS-940PA
caster feet. Do not immerse the probe cable in detergents or water. Failure to follow these cautions may cause cable discontinuity, short circuit, skin burn on the patient and incorrect measurement data. Replace any broken probe with a new one.
• When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one.
• When the probe failure message appears on the screen, replace it with a new one. Otherwise SpO
data may not be accurate.
2
• While a patient is on medication which causes vasodilation, the pulse waveform may change and in rare cases the SpO
value may not be displayed.
2
NOTE
In order to maintain sufficient blood circulation, keep the measurement site warm by covering it with a blanket or something similar. Warming the site is effective, especially for a patient with a small pulse amplitude.
SpO
and PR Display Order
2
You can select the display order for SpO “Changing Parameter Setup Settings” section.
and PR (pulse rate) on the LCD. Refer to the
2
Monitoring SpO
When the SpO during NIBP measurement and pulse wave cannot be detected and SpO properly. When “INHIBIT SpO to ON (factory default setting), SpO SpO
alarm occurrence. However, when monitoring SpO2 on the same limb as NIBP, be careful
2
when reading SpO
When monitoring SpO
during NIBP Measurement
2
probe is attached to the same limb as the NIBP cuff, the blood flow decreases
2
cannot be monitored
2
DURING NIBP” on the PARAMETER SETUP screen is set
2
monitoring is paused during NIBP measurement to avoid
2
values.
2
is important, attach the probe to the limb to which the NIBP cuff or
2
catheter is not attached.
When SpO2 monitoring is paused during NIBP measurement, the SpO2 value just before the start of NIBP measurement and an When NIBP measurement is not completed after 30 seconds, “– – –” is displayed for the SpO
mark are displayed on the transmitter for 30 seconds.
2
value. The same data also appears on the monitor screen.
Operator's Manual ZS-940PA 51
NOTE
• When continuous SpO2 monitoring is necessary, attach the probe to the limb to which the NIBP cuff is not attached and set “INHIBIT SpO the PARAMETER SETUP screen to OFF.
• When the probe is attached to the same limb as the NIBP cuff, set the sync source to a parameter other than SpO
• When monitoring SpO
during STAT NIBP measurement, attach the probe to
2
on the receiving monitor.
2
the limb to which the NIBP cuff is not attached.
DURING NIBP” on
2
52 Operator's Manual ZS-940PA
Display and Message List
Battery Indication
Indication Cause Countermeasure
Fully charged battery
Batteries are low.
No
indication
Batteries are low. NIBP cannot be measured.
Dead batteries
Replace batteries.
ECG
Indication Cause Countermeasure
Electrode lead is disconnected from the electrode. Electrode lead is disconnected from the transmitter.
Electrode lead discontinuity.
Electrode is not firmly attached to the skin.
Polarization voltage is abnormally high.
Firmly connect the electrode lead to the electrode. Firmly connect the electrode lead to the transmitter. Replace the electrode lead with a new one.
Replace the electrode with a new one.
Operator's Manual ZS-940PA 53
SpO
2
Message Cause Countermeasure
During NIBP measurement
Detecting body movement
SpO2 CHECK PROBE
SpO2 CHECK PROBE SITE
SpO2 DETECTING PULSE
SpO2 monitoring is paused for NIBP measurement.
Considerable body movement.
The probe is not attached to the patient properly.
The probe is not attached to the patient properly. The probe is disconnected from the SpO
socket on the
2
transmitter. The probe is not attached at the appropriate site.
Wait for NIBP measurement to finish. When the message is displayed frequently, check the patient condition and, if necessary, change the attachment site. Attach the probe to the patient properly.
Connect the probe to the SpO socket.
Attach the probe to a site 6 to 14 mm thick.
Probe is expired. Replace the probe with a new one.
Searching for the correct pulse wave.
Wait until the pulse wave is
detected. The SpO2 value cannot be obtained because the waveform is unstable. The probe is not attached to
Attach the probe to the patient
properly.
the patient properly. SpO
measurement site is
SpO2 LIGHT INTERFERENCE
SpO2 PROBE FAI LU RE
SpO2 WEAK PULSE
2
under fluorescent light, surgical light, sunlight, etc.
Probe is expired. Replace the probe with a new one.
Probe is damaged or short­circuited.
Poor peripheral circulation.
The probe is attached too tightly and is obstructing the blood circulation.
Cover the measurement site with a
blanket or cloth.
Replace the probe with a new one.
Check the patient condition and
change the attachment site.
Check the probe attachment
condition and if necessary,
reattach the probe.
2
54 Operator's Manual ZS-940PA
NIBP
Message Cause Countermeasure
The cuff and extension hose are not
NIBP AIR LEAK
NIBP CANNOT DETECT PULSE
NIBP CUFF OCCLUSION
NIBP HIGH CUFF PRESS
NIBP INFLATION PRESS LOW
NIBP MEAS TIME­OUT
NIBP MODULE FAI LU RE
NIBP REMEASURING
NIBP SAFETY CIRCUIT RUNNING (When this message is displayed, measurement cannot be performed for 40 seconds.)
NIBP SYS OUT OF RANGE
NIBP WEAK PULSE
NIBP ZEROING
properly connected. The cuff hose (or extension hose) is not properly connected to the NIBP socket. The cuff or extension hose is damaged.
The patient’s pulse wave is small. Measure by palpation or auscultation.
The cuff is not wrapped on the patient properly.
Transmitter malfunction.
Enormous pressure was applied by the pressure of the cuff.
Insufficient cuff inflation pressure.
The measuring time exceeded the specified time due to arrhythmia, body movement, vibration or, cuff or air hose being squeezed.
Module malfunction.
NIBP is being remeasured due to arrhythmia, body movement, vibration or, cuff or air hose being squeezed.
Measurement stopped by the safety circuit.
The maximum blood pressure cannot be measured even when the cuff inflation pressure exceeded 280 mmHg when using adult cuff. The patient’s pulse wave is too small.
The cuff is wrapped too loosely. Wrap the cuff properly.
The cuff size is not appropriate. Use the appropriate cuff.
NIBP zero balance is being adjusted.
Connect them properly.
Replace with a new one.
Wrap the cuff on the patient properly.
Immediately remove the cuff from the patient and contact your Nihon Kohden distributor.
Remove the cause.
Wait for the remeasurement to be performed with increased cuff inflation pressure.
Remove the cause if the cause is body movement, vibration or squeezing of cuff or hose.
Contact your Nihon Kohden distributor. If the message still appears after remeasurement, remove the cause if the cause is body movement, vibration or squeezing of cuff or hose. Check that the hose is not bent or squeezed. Wait 40 seconds, then perform remeasurement. If the message still appears, contact your Nihon Kohden distributor.
Measure by palpation or auscultation.
Measure by palpation or auscultation.
Do not touch the cuff during zeroing. Wait for the message to disappear.
Operator's Manual ZS-940PA 55
p
Troubleshooting
If the problem still remains after checking the following, contact your Nihon Kohden distributor.
Transmitter
Problem Cause Countermeasure
displayed on the LCD after turning the power on.
displayed on the monitor after turning the transmitter power on.
Signal receiving condition is
oor.
Batteries are not installed correctly. The battery polarity is wrong. Batteries are completely discharged. The channel of the transmitter and monitor does not match. The “TYPE” on the SYSTEM SETUP screen is not set to “A”.
Another transmitter of the same channel is used nearby.
Signals are mixing. Follow the instructions of your channel
The transmitter is damaged.
Install the batteries correctly.Nothing is
Replace the batteries with new ones.
Set the correct channel on the monitor.Nothing is
Check that the “TYPE” is set to “A”. Refer to the “Notes on Parameter Settings” and “Changing System Setup Settings” sections. Turn the transmitter power off. If the monitor still receives a signal, there is a high probability that another transmitter of the same channel is used nearby. Follow the instruction of your channel administrator and use another transmitter of a different channel.
administrator and use another transmitter of a different channel. Contact your Nihon Kohden distributor.
56 Operator's Manual ZS-940PA
ECG
Problem Cause Countermeasure
The heart rate is unstable.
The “CHECK ELECTRODE” message appears on the receiving monitor.
ECG baseline is thick. (Hum is overlapping)
Respiration waveform measurement is unstable.
Pacing detection setting on the monitor is not correct.
Turn off the pacing detection setting on the monitor. When monitoring a pacemaker
patient, turn on pacing detection. Electrode lead is disconnected from the electrode.
Firmly connect the electrode lead to
the electrode. Electrode lead discontinuity Replace the electrode lead with a new
one. Electrode is not firmly
Replace the electrode with a new one. attached to the skin. Polarization voltage is abnormally high. The gel on the electrode is
Use Nihon Kohden specified
electrodes.
Replace the electrode with a new one. dried out. The gel on the electrode is coming off. Electric blanket is used. Cover the blanket with a shield cover. Hum filter is set to OFF on the
Set the filter to ON. monitor The gel on the electrode is
Replace the electrode with a new one. dried out. The gel on the electrode is coming off.
SpO
2
Problem Cause Countermeasure
SpO2 data is unstable and not reliable.
The probe size is not appropriate for the patient. Probe attachment condition is poor. Probe is partly
Use the appropriate probe for the patient.
Firmly attach the probe according to the
procedure in the probe operator’s manual. detached from the skin. External light gets in. Measurement site is dirty.
Remove dirt and nail polish. Patient is wearing nail polish. Probe is attached to the same limb that is used for NIBP measurement.
When the probe and cuff are attached to
the same limb, set “INHIBIT SpO
2
DURING NIBP” setting on the
PARAMETER SETUP screen to ON.
Attach the probe to the opposite limb.
Avoid a site where blood circulation
condition changes greatly.
Operator's Manual ZS-940PA 57
NIBP
Problem Cause Countermeasure
Cuff inflation pressure is less than 10 mmHg.
The cuff does not inflate when the NIBP START/STOP key is pressed.
Abnormal measurement results are displayed.
The cuff is suddenly deflated during inflation.
Auto mode measurement does not start even when the time interval has passed. The cuff suddenly inflates.
Cannot connect cuff to the air hose.
The cuff hose is not connected to the NIBP socket properly. The cuff is not wrapped around the arm or is wrapped too loosely. The cuff hose is not connected to the NIBP socket. The cuff hose or extension hose may be folded or squeezed when the cuff pressure display on the screen increases quickly but the actual cuff does not inflate.
The cuff size is not correct. Select the cuff which fits the
The cuff is not wrapped around the arm correctly.
NIBP data is not correct because of body movement.
Vibration on the cuff.
The NIBP START/STOP key is pressed during inflation.
The NIBP INTERVAL key is pressed and the measurement mode is changed.
The measurement mode is set to auto mode.
Unspecified cuff is used. Use a cuff specified by Nihon
Connect the cuff hose to the socket properly. Wrap the cuff around the upper arm.
Connect the cuff hose to the socket firmly. Check the cuff hose and air hose.
patient’s limb circumference. Wrap the cuff around the upper arm, not too tightly or too loosely.
Prevent the patient from moving during measurement. Check that nothing is touching the cuff during measurement. Change the measuring site.
Check the measurement mode and interval.
Check the time interval. If necessary, stop measurement.
Kohden.
58 Operator's Manual ZS-940PA
Problem Cause Countermeasure
Cannot measure NIBP.
Vibration on the cuff.
The cuff hose or extension
Check that nothing is touching the cuff during measurement.
Remove the cause. hose is bent or squeezed. The cuff has worn out. Use a new cuff. Measuring over a long
occurs.
period of time at short intervals.
Thrombus occurs. Measuring on a patient
with known bleeding
Increase the measuring interval.Blood congestion
Do not measure NIBP over a long
time.
Do not perform NIBP measurement
on such a patient. disorders or coagulation.
NIBP data on the screen is --- or dark.
The time set for “OLD NIBP DATA” on the PARAMETER SETUP
When NIBP is measured again, the
data is displayed in normal
brightness. screen elapsed from the last measurement.
Three loud pip sounds indicting NIBP measurement
The cuff is not deflated enough to start another measurement.
Wait 30 seconds and measure again.
cannot be started. Numeric data is displayed but NIBP messages are not displayed on the
The “TYPE” on the SYSTEM SETUP screen is not set to “A”.
Check that the “TYPE” is set to “A”.
Refer to the “Notes on Parameter
Settings” and “Changing System
Setup Settings” sections.
monitor.
Operator's Manual ZS-940PA 59
Maintenance
To use the instrument in safe and optimum condition, perform maintenance check once every six months.
CAUTION
Do not disassemble the transmitter when performing maintenance and inspection. Do not repair the transmitter. When there is any problem with the transmitter after maintenance and inspection, contact your Nihon Kohden distributor.
A maintenance check sheet is provided at the end of this section. Make a copy of this check sheet before performing maintenance check.
1. External Check
• There are no damaged or dirty parts on the outside of the transmitter.
• The battery case cover is not damaged, the spring is firmly fixed and the battery case cover can be closed firmly.
• NIBP socket is not damaged.
• Keys are not damaged.
• Electrode leads are not damaged.
• There is no blood or chemicals on the transmitter.
2. Transmitter Channel
Check that the channel of the transmitter and the label match.
1. Check that the channel number label attached to the transmitter is not torn or removed.
Channel label
60 Operator's Manual ZS-940PA
2. Remove one battery.
3. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery. The SETUP screen appears.
4. Press the NIBP INTERVAL key to move the cursor to “SYSTEM SETUP”.
5. Press the NIBP START/STOP key to enter the SYSTEM SETUP screen. The channel of this transmitter is displayed.
Channel
6. Check that the channel displayed on the LCD matches the label on the transmitter.
Operator's Manual ZS-940PA 61
3. LCD Display
Check that there are no dots missing on the LCD.
1. Remove one battery.
2. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery. The SETUP screen appears.
3. Press the NIBP INTERVAL key twice to move the cursor to “MANUAL CHECK”.
4. Press the NIBP START/STOP key to enter the MANUAL CHECK screen.
5. Check that the cursor is on “LCD TEST” and press the NIBP START/ STOP key.
62 Operator's Manual ZS-940PA
6. Every time the NIBP INTERVAL key is pressed, the screen changes as below. Check that there are no dots missing.
When the NIBP START/STOP key is pressed, the screen returns to the MANUAL CHECK screen.
4. Switch and Key Operation
Power Switch
Check that the Power switch turns the power on and off.
NIBP START/STOP Key
1. Attach the NIBP cuff to your upper arm.
2. Press the NIBP START/STOP key. Check that the cuff inflates and deflates properly.
3. Press the NIBP START/STOP key again. During inflation, press the NIBP START/STOP key to check that the cuff deflates properly.
NIBP INTERVAL Key
1. Press the NIBP INTERVAL key and check that the NIBP measuring mode can be changed.
2. Select any interval and press the NIBP START/STOP key to perform auto measurement. Check that the NIBP is measured at the selected interval.
Operator's Manual ZS-940PA 63
5. NIBP Cuff for Attaching Transmitter to Patient Arm
The NIBP cuff is a consumable. Check the following and when necessary, replace it with a new one.
Appearance
• There are no dirty parts.
• There are no broken stitches on the cuff.
• The label on the cuff is readable.
• The velcro tape on the cuff is not removed and there are no broken stitches.
• The lock plate is not damaged and functions properly.
Inflation bag
• The inflation bag is not torn or damaged.
• There is no water inside the inflation bag.
• The connector on the inflation bag is not damaged.
64 Operator's Manual ZS-940PA
Hospital/Organization:
Service Personnel:
Maintenance Check Sheet
Instrument Name:
Instrument Model:
Instrument Serial Number:
Hardware Revision Number:
Software Revision Number:
1. External Check OK No
2. Transmitter Channel OK No
3. LCD Display OK No
4. Switch and Key Operation OK No
5. NIBP Cuff for Attaching Transmitter to Patient Arm OK No
Overall Judgement
OK
Transmitter
ZS-940PA
Can be used but needs maintenance
Maintenance required. Cannot be used.
Operator's Manual ZS-940PA 65
Repair Parts Availability Policy
Nihon Kohden Corporation (NKC) shall stock repair parts (parts necessary to maintain the performance of the instrument) for a period of 8 years from the date of delivery. In that period NKC or its authorized agents will repair the instrument. This period may be shorter than 8 years if the board or part necessary for the faulty section is not available.
66 Operator's Manual ZS-940PA
Lifetime and Disposal
Disposing of Used Batteries
Battery Lifetime
Replace the batteries when the battery replacement indication appears on the transmitter. When using rechargeable batteries, recharge them.
Type Lifetime (Measuring parameters)
ECG, SpO2, NIBP ECG, SpO
NiMH secondary 2 days 2 days 2.5 days Alkaline primary 1 day 2.5 days 3 days
The above data is when new batteries are used at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO weight 60 kg.
Operation time depends on the thickness of SpO2 probe attachment site.
Disposal
is measured on an index finger of a male patient with
2
NOTE
Remove the batteries before disposing of the transmitter.
Before disposing of the batteries, check with your local solid waste officials for details in your area for proper disposal. It may be illegal to dispose of these batteries in the municipal waste stream.
2
ECG only
Disposing of Electrodes, SpO2 Probes and NIBP Cuffs
Refer to the manual of each item.
Operator's Manual ZS-940PA 67
Cleaning, Disinfection and Sterilization
Transmitter and Electrode Leads
CAUTION
• If detergent or liquid spills into the transmitter, stop using it and contact your Nihon Kohden distributor. If a wet transmitter is used, the patient or anyone in contact with the transmitter may receive an electric shock or patient leakage current over the allowed amount may flow.
• Before cleaning or disinfecting the transmitter, remove the batteries from the transmitter.
• The transmitter cannot be sterilized.
Cleaning
Wipe the transmitter and electrode leads with a soft cloth moistened with disinfecting alcohol or neutral detergent diluted with water. After cleaning, dry them completely.
Disinfection
CAUTION
• Do not immerse the electrode lead connector in liquid.
• Do not disinfect with hypochlorous acid.
• Use the recommended concentration.
Wipe the outside surface of the transmitter and electrode lead with a non-abrasive cloth moistened with any of the disinfectants listed below. Use the recommended concentration.
Disinfectant Concentration (%) Glutaraldehyde solution 2.0 Hydrochloric alkyl diaminoethylglycine 0.5 Benzalkonium chloride 0.2 Benzethonium chloride solution 0.2 Chlorohexidine gluconate solution 0.5
68 Operator's Manual ZS-940PA
SpO2 Probe
Refer to the probe manual.
YP-943P/944P NIBP Cuffs
CAUTION
• Do not autoclave.
• Use only glutaraldehyde solution.
• Never allow liquid to get inside the rubber cuff.
• Do not sterilize or disinfect the cuff with ultraviolet light or ozone.
Cleaning
To clean the cuff, remove the lock plate and carefully pull out the inflation bag from the cloth cover. Cloth cover: Wash with neutral detergent and water. Thoroughly dry it. When washing in a
washing machine, put it in a net.
Inflation bag: Wipe with a soft cloth or cotton moistened with disinfecting alcohol.
Thoroughly dry it.
Disinfection
To disinfect the cuff, use glutaraldehyde solution. Use the recommended concentration of the disinfectant. Refer to the disinfectant manual for details. After disinfection, clean the cuff as described above.
Operator's Manual ZS-940PA 69
Specifications
Measuring Parameters
Measuring waveforms: ECG, Respiration in impedance method, pulse Measuring numeric data: SpO
Transmitting Data
Waveform data: ECG, respiration, pulse wave Numeric data: SpO Status information: Battery replacement, channel ID, type of transmitter, check
Displayed Data
SpO2, NIBP, pulse rate, pulse wave bar graph, check electrode, battery replacement, NIBP measurement mode and status information
ECG Measurement
Channels: 1 Input range: ±5 mV or more DC offset: ±500 mV or more Input impedance: 5 M or more (5 Hz) Pacing pulse detection: ANSI/AAMI EC13
, NIBP, pulse rate
2
and NIBP
2
electrodes, abnormal polarization voltage, pacing data, SpO light interference
Based upon Pacemaker pulse rejection Capability
2
Respiration Measurement
Measuring method: Impedance method Impedance range: 0 to 2 k or less
SpO
Measurement
2
Display range: Depends on the receiving monitor Measuring range: 0 to 100%, in 1% steps Minimum display range: 1%
Measuring accuracy (When the measuring accuracy of the SpO
±1 (80% SpO
±2 (50% SpO
100%)
2
< 80%)
2
probe is not considered):
2
Less than 50% is not specified.
(When considering the measuring accuracy of the SpO
±2 (80% SpO
±3 (70% SpO
probe):
2
100%)
2
< 80%)
2
Less than 70% is not specified.
70 Operator's Manual ZS-940PA
NIBP Measurement
Displayed items: Systolic, diastolic, mean Cuff pressure display range: 0 to 300 mmHg Measurement modes: Manual, STAT, auto at 5, 10, 15, 30, 60, 120 or 240 minute
interval
Pulse Rate
Measuring range: 30 to 200 beats/minute ±8 beats/min (NIBP)
30 to 200 beats/min ±3% ±1 beat/min (SpO
)
2
Transmitter
FCC regulation: FCC part 95 Subpart H
Wireless Medical Telemetry Service (WMTS) Field strength limits: <200 mV/m (at 3 m) Undesired emission: below 960 MHz: 200 µV/m (at 3 m)
above 960 MHz: 500 µV/m (at 3 m) Antenna: Internal Transmission channel: indicated on the transmitter Transmission frequency range: 608.0125 to 613.9875 MHz Channel spacing: 25 kHz (12.5 kHz when interleave) Type of emission: F1D Occupied bandwidth: <8.5 kHz Effective radiated power: 1.0 mW (conducted)
Power Requirements
Operating voltage: 3.2 to 4.8 V Battery type: Three AA type NiMH secondary batteries
Three AA type alkaline dry cell primary batteries Battery lifetime:
Type Lifetime (Measuring parameters)
ECG, SpO2, NIBP ECG, SpO
2
ECG only
NiMH secondary 2 days 2 days 2.5 days Alkaline primary 1 day 2.5 days 3 days
The above data is when new batteries are used at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO
is measured on an index
2
finger of a male patient with weight 60 kg. Operation time depends on the thickness of SpO
probe attachment site.
2
Dimension and Weight
Dimension: 114 W × 103 H × 58 D (mm) Weight: about 350 g (including batteries)
Operator's Manual ZS-940PA 71
Environment
Operating environment
Operating temperature: 5 to 40°C, 41 to 104°F
When using NIBP cuff, 10 to 40°C, 50 to 104°F Operating humidity: 30 to 85% (non-condensing) Operating atmospheric pressure: 70 to 106 kPa
Storage environment
Storage temperature: 20 to 65°C, 4 to 149°F Storage humidity: 10 to 95% Storage atmospheric pressure: 70 to 106 kPa
Electromagnetic Compatibility
IEC 60601-1-2 (1993) - Collateral Standard: Electromagnetic compatibility - Requirement and
tests
Emissions: CISPR11 Group 1, Class B
Safety Standards
Safety standard: CAN/CSA-C22.2 No. 601-1 M90 (1990)
CAN/CSA-C22.2 No. 601-1. 1S1-94 (1994) CAN/CSA-C22.2 No. 601-1. 1B-90 (R2002) IEC 60601-1 (1988) IEC 60601-1 Amendment1 (1991) IEC 60601-1 Amendment2 (1995) IEC 60601-1-2 (1993) IEC 60601-2-27 (1994)
IEC 60601-2-30 (1999) According to the type of protection against electrical shock: INTERNALLY POWERED EQUIPMENT According to the degree of protection against electrical shock:
ECG and impedance method respiration: DEFIBRILLATION-PROOF TYPE CF APPLIED
PA RT
SpO
and NIBP: DEFIBRILLATION-PROOF TYPE BF APPLIED
2
PA RT According to the degree of protection against harmful ingress of water: IPX0 (Ordinary equipment) According to the degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE: Equipment not suitable for use in the presence of
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE
According to the mode of operation: CONTINUOUS OPERATION
72 Operator's Manual ZS-940PA
Standard Accessories
1
2
No. Name Model Q’ty Supply Code No.
1 NIBP cuff for adult, standard YP-943P 1 S938B 2 Strap --- 1 Y236
The following parts are available for replacement.
34
Lock plate is a standard accessory of the YP-943P/944P NIBP cuff.
No. Name Model Q’ty Supply Code No.
3 Battery case cover 1 6144-012004 4 Lock plate 1 6113-049585
Operator's Manual ZS-940PA 73
Options
CAUTION
Use only Nihon Kohden electrodes, electrode leads, SpO2 probes and NIBP cuffs to assure maximum performance from your instrument.
Transmitter
Channel writer, QI-901PK
ECG/RESP
Name Application Model Q’ty Supply Code No.
Electrode lead
3 electrodes,clip type, lead length 80 cm 3 electrodes, snap type, lead length 80 cm 2 electrodes,clip type, lead length 80 cm 2 electrodes,snap type, lead length 80 cm
BR-903PA 1 K911A
BR-913PA 1 K910B
BR-902PA 1 K907B
BR-912PA 1 K908B
74 Operator's Manual ZS-940PA
SpO
m
m
2
Name
Finger probe (reusable)
Cable
length
0.6 m P225H
1.6
Model/
Code No.
TL-201T
Multi-site probe (reusable) TL-220T
Q’ty
1
Supply
Code No.
P225F
P225G SpO2 probe (for adult, disposable) TL-251T P201A SpO2 probe (for child, disposable) TL-252T P201B SpO2 probe (for neonate, disposable)
1.6 m
TL-253T P201C
Multi-site Y probe (for low birth weight infant/child/
TL-260T P205A
5
neonate, disposable) SpO2 probe(for adult/neonate, disposable) SpO2 probe (for child/infant, disposable)
0.8 m TL-051S P228A
1.6 m TL-052S P228B
0.8 m TL-061S P229A
1.6
TL-062S
P229B COTTONY tape 340703 20 P259 Foam tape for TL-051S/052S/ 061S/062S Attachment tape for TL-220T/ 251T/252T/253T
---
---
Attachment tape S for TL-260T P260A Attachment tape L for TL-260T
4 × 25 package
3 × 30 package
24
P260
P263
P260B
NIBP
Name
Cuff for adult, for attaching transmitter to patient arm
Width
(cm)
Standard 13 YP-943P* S938B
Large 15
Air Hose
Length
(cm)
15
Model Q’ty
YP-944P*
Supply
Code
No.
1
S938C
Cuff for infants 5 YP-960T S943A
Cuff for children
Cuff for adult
Small 7 YP-961T S943B Standard 10 YP-962T S943C
15
1 Standard 13 YP-963T S944B Large 15
YP-964T
S944C Disposable cuff for infants 6 YP-910P --­Disposable cuff for children 9 YP-912P ---
Disposable cuff for adults
Small 12 YP-913P --­Standard 14 YP-914P --­Large 16
20
YP-915P
20
---
Extension hose 150 YN-990P 1 S903
* The lock plate is provided with these NIBP cuffs.
Operator's Manual ZS-940PA 75
Transmission Frequencies
USA
BAND
Transmission
frequency
(MHz)
608.0000 --- 608.3750 9030 608.7500 9060
608.0125 9001 608.3875 9031 608.7625 9061
608.0250 9002 608.4000 9032 608.7750 9062
608.0375 9003 608.4125 9033 608.7875 9063
608.0500 9004 608.4250 9034 608.8000 9064
608.0625 9005 608.4375 9035 608.8125 9065
608.0750 9006 608.4500 9036 608.8250 9066
608.0875 9007 608.4625 9037 608.8375 9067
608.1000 9008 608.4750 9038 608.8500 9068
608.1125 9009 608.4875 9039 608.8625 9069
608.1250 9010 608.5000 9040 608.8750 9070
608.1375 9011 608.5125 9041 608.8875 9071
608.1500 9012 608.5250 9042 608.9000 9072
608.1625 9013 608.5375 9043 608.9125 9073
608.1750 9014 608.5500 9044 608.9250 9074
608.1875 9015 608.5625 9045 608.9375 9075
608.2000 9016 608.5750 9046 608.9500 9076
608.2125 9017 608.5875 9047 608.9625 9077
608.2250 9018 608.6000 9048 608.9750 9078
608.2375 9019 608.6125 9049 608.9875 9079
608.2500 9020 608.6250 9050 609.0000 9080
608.2625 9021 608.6375 9051 609.0125 9081
608.2750 9022 608.6500 9052 609.0250 9082
608.2875 9023 608.6625 9053 609.0375 9083
608.3000 9024 608.6750 9054 609.0500 9084
608.3125 9025 608.6875 9055 609.0625 9085
608.3250 9026 608.7000 9056 609.0750 9086
608.3375 9027 608.7125 9057 609.0875 9087
608.3500 9028 608.7250 9058 609.1000 9088
608.3625 9029 608.7375 9059 609.1125 9089
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
12.5kHz step Channel No.
A
76 Operator's Manual ZS-940PA
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
12.5kHz step Channel No.
609.1250 9090 609.5375 9123 609.9500 9156
609.1375 9091 609.5500 9124 609.9625 9157
609.1500 9092 609.5625 9125 609.9750 9158
609.1625 9093 609.5750 9126 609.9875 9159
609.1750 9094 609.5875 9127 610.0000 9160
609.1875 9095 609.6000 9128 610.0125 9161
609.2000 9096 609.6125 9129 610.0250 9162
609.2125 9097 609.6250 9130 610.0375 9163
609.2250 9098 609.6375 9131 610.0500 9164
609.2375 9099 609.6500 9132 610.0625 9165
609.2500 9100 609.6625 9133 610.0750 9166
609.2625 9101 609.6750 9134 610.0875 9167
609.2750 9102 609.6875 9135 610.1000 9168
609.2875 9103 609.7000 9136 610.1125 9169
609.3000 9104 609.7125 9137 610.1250 9170
609.3125 9105 609.7250 9138 610.1375 9171
609.3250 9106 609.7375 9139 610.1500 9172
609.3375 9107 609.7500 9140 610.1625 9173
609.3500 9108 609.7625 9141 610.1750 9174
609.3625 9109 609.7750 9142 610.1875 9175
609.3750 9110 609.7875 9143 610.2000 9176
609.3875 9111 609.8000 9144 610.2125 9177
609.4000 9112 609.8125 9145 610.2250 9178
609.4125 9113 609.8250 9146 610.2375 9179
609.4250 9114 609.8375 9147 610.2500 9180
609.4375 9115 609.8500 9148 610.2625 9181
609.4500 9116 609.8625 9149 610.2750 9182
609.4625 9117 609.8750 9150 610.2875 9183
609.4750 9118 609.8875 9151 610.3000 9184
609.4875 9119 609.9000 9152 610.3125 9185
609.5000 9120 609.9125 9153 610.3250 9186
609.5125 9121 609.9250 9154 610.3375 9187
609.5250 9122 609.9375 9155 610.3500 9188
A
Operator's Manual ZS-940PA 77
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
12.5kHz step Channel No.
610.3625 9189 610.7750 9222 611.1875 9255
610.3750 9190 610.7875 9223 611.2000 9256
610.3875 9191 610.8000 9224 611.2125 9257
610.4000 9192 610.8125 9225 611.2250 9258
610.4125 9193 610.8250 9226 611.2375 9259
610.4250 9194 610.8375 9227 611.2500 9260
610.4375 9195 610.8500 9228 611.2625 9261
610.4500 9196 610.8625 9229 611.2750 9262
610.4625 9197 610.8750 9230 611.2875 9263
610.4750 9198 610.8875 9231 611.3000 9264
610.4875 9199 610.9000 9232 611.3125 9265
610.5000 9200 610.9125 9233 611.3250 9266
610.5125 9201 610.9250 9234 611.3375 9267
610.5250 9202 610.9375 9235 611.3500 9268
610.5375 9203 610.9500 9236 611.3625 9269
610.5500 9204 610.9625 9237 611.3750 9270
610.5625 9205 610.9750 9238 611.3875 9271
610.5750 9206 610.9875 9239 611.4000 9272
610.5875 9207 611.0000 9240 611.4125 9273
610.6000 9208 611.0125 9241 611.4250 9274
610.6125 9209 611.0250 9242 611.4375 9275
610.6250 9210 611.0375 9243 611.4500 9276
610.6375 9211 611.0500 9244 611.4625 9277
610.6500 9212 611.0625 9245 611.4750 9278
610.6625 9213 611.0750 9246 611.4875 9279
610.6750 9214 611.0875 9247 611.5000 9280
610.6875 9215 611.1000 9248 611.5125 9281
610.7000 9216 611.1125 9249 611.5250 9282
610.7125 9217 611.1250 9250 611.5375 9283
610.7250 9218 611.1375 9251 611.5500 9284
610.7375 9219 611.1500 9252 611.5625 9285
610.7500 9220 611.1625 9253 611.5750 9286
610.7625 9221 611.1750 9254 611.5875 9287
A
78 Operator's Manual ZS-940PA
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
12.5kHz step Channel No.
611.6000 9288 612.0125 9321 612.4250 9354
611.6125 9289 612.0250 9322 612.4375 9355
611.6250 9290 612.0375 9323 612.4500 9356
611.6375 9291 612.0500 9324 612.4625 9357
611.6500 9292 612.0625 9325 612.4750 9358
611.6625 9293 612.0750 9326 612.4875 9359
611.6750 9294 612.0875 9327 612.5000 9360
611.6875 9295 612.1000 9328 612.5125 9361
611.7000 9296 612.1125 9329 612.5250 9362
611.7125 9297 612.1250 9330 612.5375 9363
611.7250 9298 612.1375 9331 612.5500 9364
611.7375 9299 612.1500 9332 612.5625 9365
611.7500 9300 612.1625 9333 612.5750 9366
611.7625 9301 612.1750 9334 612.5875 9367
611.7750 9302 612.1875 9335 612.6000 9368
611.7875 9303 612.2000 9336 612.6125 9369
611.8000 9304 612.2125 9337 612.6250 9370
611.8125 9305 612.2250 9338 612.6375 9371
611.8250 9306 612.2375 9339 612.6500 9372
611.8375 9307 612.2500 9340 612.6625 9373
611.8500 9308 612.2625 9341 612.6750 9374
611.8625 9309 612.2750 9342 612.6875 9375
611.8750 9310 612.2875 9343 612.7000 9376
611.8875 9311 612.3000 9344 612.7125 9377
611.9000 9312 612.3125 9345 612.7250 9378
611.9125 9313 612.3250 9346 612.7375 9379
611.9250 9314 612.3375 9347 612.7500 9380
611.9375 9315 612.3500 9348 612.7625 9381
611.9500 9316 612.3625 9349 612.7750 9382
611.9625 9317 612.3750 9350 612.7875 9383
611.9750 9318 612.3875 9351 612.8000 9384
611.9875 9319 612.4000 9352 612.8125 9385
612.0000 9320 612.4125 9353 612.8250 9386
A
Operator's Manual ZS-940PA 79
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
A
12.5kHz step Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
12.5kHz step Channel No.
612.8375 9387 613.2500 9420 613.6625 9453
612.8500 9388 613.2625 9421 613.6750 9454
612.8625 9389 613.2750 9422 613.6875 9455
612.8750 9390 613.2875 9423 613.7000 9456
612.8875 9391 613.3000 9424 613.7125 9457
612.9000 9392 613.3125 9425 613.7250 9458
612.9125 9393 613.3250 9426 613.7375 9459
612.9250 9394 613.3375 9427 613.7500 9460
612.9375 9395 613.3500 9428 613.7625 9461
612.9500 9396 613.3625 9429 613.7750 9462
612.9625 9397 613.3750 9430 613.7875 9463
612.9750 9398 613.3875 9431 613.8000 9464
612.9875 9399 613.4000 9432 613.8125 9465
613.0000 9400 613.4125 9433 613.8250 9466
613.0125 9401 613.4250 9434 613.8375 9467
613.0250 9402 613.4375 9435 613.8500 9468
613.0375 9403 613.4500 9436 613.8625 9469
613.0500 9404 613.4625 9437 613.8750 9470
613.0625 9405 613.4750 9438 613.8875 9471
613.0750 9406 613.4875 9439 613.9000 9472
613.0875 9407 613.5000 9440 613.9125 9473
613.1000 9408 613.5125 9441 613.9250 9474
613.1125 9409 613.5250 9442 613.9375 9475
613.1250 9410 613.5375 9443 613.9500 9476
613.1375 9411 613.5500 9444 613.9625 9477
613.1500 9412 613.5625 9445 613.9750 9478
613.1625 9413 613.5750 9446 613.9875 9479
613.1750 9414 613.5875 9447 614.0000 ---
613.1875 9415 613.6000 9448
613.2000 9416 613.6125 9449
613.2125 9417 613.6250 9450
613.2250 9418 613.6375 9451
613.2375 9419 613.6500 9452
A
80 Operator's Manual ZS-940PA
Loading...