MELAG MELAquick 12+ p, MELAquick 12+ User Manual

Page 1
Dear doctor, we thank you for your confidence demonstrated by the purchase of this MELAG product.
As an owner-run and operated family concern founded in 1951, we have a long history of successful specialization in hygiene products for practice-based use. Our focus on innovation, quality and the highest standards of operational reliability has established MELAG as the world’s leading manufacturer in the instrument treatment and hygiene field.
You, our customer are justified in your demand for the best products, quality and reliability. Providing "competence in hygiene" and "Quality – made in Germany", we guarantee that these demands will be met. Our certified quality
management system is subject to close monitoring: one instrument to this end is our annual multi-day audit conducted in accordance with ISO 13485 and ISO 9001. This guarantees that all MELAG products are manufactured and tested in accordance with strict quality criteria.
The MELAG management and team.
User Manual
Steam sterilizer
MELAquick® 12+ MELAquick
®
12+ p
From software version 5.15
EN
Page 2
User manual MELAquick® 12+/12+ p
Valid MELAquick® 12+/12+ p
From software version 5.15
Responsible for content:
MELAG Medizintechnik oHG
Geneststraße 6 – 10
10829 Berlin
Germany
Email: info@melag.com
www.melag.com
© 2017
Document: BA_quick12+_12+p_EN_v12.docx / Rev.: 12 – 17/2441
Date of update: 2017-12-13
We reserve the right to technical alterations
For doctors, physicians assistants and service personnel
Please read this User Manual before you start operation of the
steam sterilizer.
The instructions include important safety information. Make sure to keep the User Manual near to the steam sterilizer. It represents a component of the product.
Page 3
Foreword
Thank you for deciding to buy this MELAG steam sterilizer.
Device description
The device name "steam sterilizer" is used in this User Manual to designate the steam sterilizer
MELAquick 12+/12+ p.
User Manual
The User Manual contains important safety instructions and information which you need to operate the
steam sterilizer. Read these instructions
carefully and thoroughly in proper sequence.
Avoid dangers
Please read all the safety instructions attentively before using the steam sterilizer.
About these Instructions
Symbol
Meaning
Explanation
Danger
Risk to health
Indicates a dangerous situation which if not avoided
could entail slight to life-threatening injuries.
Attention
Observe without fail
Indicates a dangerous situation which if not avoided
could entail damage to the instruments, the prac tic e equipment or the steam sterilizer.
Important
information
Draws your attention to important information.
Table 1: Meaning of the symbols used in this User Manual
Example of emphasis
Meaning
Explanation
Conductivity measurement
Glossary entry
Words or phrases marked with an arrow are explained
in the
glossary. The glossary is listed alphabetically. It
can be found at the end of this manual.
Universal-Program S
Software citation
Words or phrases appearing on the →display of the
steam sterilizer are marked as software citations.
Chapter 6 – Logging
Cross reference
Reference to another text section within this manual.
Figure 1/(5)
Cross reference
Reference to a detail in a figure – in the example, to
part no. 5 in figure 1.
Table 2: Meaning of the emphasized text within this User Manual
Disposal
Electrical and electronic
equipment
MELAG devices are synonymous for long-term quality. When you
eventually need to decommission your MELAG device, we offer a special disposal service. Simply contact your stockist.
Accessories and
consumption media
Dispose of accessories and consumption media which you no longer
require in the appropriate manner. Comply with all relevant disposal specification in terms of possibly contaminated waste.
Packaging
The packaging protects the device against transport damage. The
packaging materials have been selected for their environmentally-friendly and recycling properties and can be recycled. Returning the packaging to the material flow reduces the amount of waste and saves raw materials. Dispose of all non-required packaging materials at the collection points of the dual system.
Page 4
Symbols on the steam sterilizer
In affixing this CE mark, the manufacturer declares that this medical product fulfils the basic
requirements of the medical device directive. The
four
-digit number confirms that this is monitored by an approved
certification agency.
Manufacturer of the medical device date of manufacture
of the medical device
Serial
number of the medical device by the manufacturer
This User Manual contains important safety informations. Failure to comply of the safety instructions could result in human and
material damage.
Please read this user manual carefully before commissioning the device. The manual includes important safety information. The functionality and value-retention of this sterilizer depends on the care accorded to it. Please store this user manual carefully a
nd in close proximity to your sterilizer. It
represents a component of the product.
The symbol of the struck out waste bin identifies a device that may not be disposed in the domestic waste. The vendor is responsible for appropriate disposal of th
e device - it must be delivered to the vendor to be disposed
of. With the designation of an apparatus with this symbol, the manufacturer furthermore declares that he satisfies all requirements of the law concerning the release, redemption and environmental
ly sound
disposal of electric and electronic appliances.
Indication of the scale of the chamber volume
Opterating temperature
of the device
Operating pressure
of the devic e
Draws your attention to a hot surface. During o peration the twist grip may become hot. Let the steam sterilizer cool down after progr am comple tion .
Page 5
Safety Instructions
To operate the steam sterilizer, pay attention to the following safety instructions
given below and in the individual chapters of this Operating
Manual.
Appropriate use
Do not sterilize any liquids with this steam sterilizer.
Power cable and
mains socket
The power cable may not be replaced ba an inadequate measured
power cable.
Never damage or alter the plug or mains cable.
Never operate the steam sterilizer if the mains cable or power plug are
damaged.
Never unplug by pulling on the power cable. Always take a grip on the
plug.
Setting up, installation,
commissioning
Only have the steam steriliz er set up, installed, and started up by
people authorized by MELAG.
Only operate the steam sterilizer in areas which are not subject to
explosion hazards.
The connections for electrical provision and water supply and discharge
must be set-up by trained personnel.
Documentation media (computer, CF card reader, etc.) must be placed
in such a way that they cannot come into contact with liquids.
Preparation and sterilization
of instruments
Follow the instructions of the instrument manufacturers for preparing
and sterilizing instruments.
Observe the relevant standards and directives for the preparation and
sterilization of instruments, e.g.
RKI [Robert Koch Institute] and
DGSV [German Society for Sterile Supply].
Only use wrapping materials and systems which are suitable for steam
sterilization according to the manufacturer's information.
Program termination
Take heed that when opening the door after terminating a running
program, hot water steam can escape out of the chamber, depending on when the program was terminated.
Depending on the time of the program termination, the load might not
be sterile. Observe the clear instructions shown on the
display of t he steam sterilizer. If necessary, sterilize the affected objects after rewrapping.
Removal of the items to be
sterilized
Never use force to open the door.
Use the basket lifter to remove the insert basket and the package
holder. Never touch the
sterilized
items, the chamber or the door with
unprotected hands. The elements are hot.
After removal, place the insert basket and the package holder on a dry,
disinfected surface or use the holder and the frame as an alternative.
Check the wrapping of the sterilized objects for damage when removing
them from the steam sterilizer. If a wrapping is damaged, wrap the object again and re-sterilize it.
Maintenance and repair
Have the maintenance and repair done only by →authorized persons.
When exchanging spare parts, you may use only the original MELAG
components.
Carrying the steam sterilizer
Use the supplied carrying straps to transport the steam sterilizer and
attach them to the sides e.g. using package tape.
Malfunctions
If repeated error messages occur while operating the steam sterilizer,
turn the device off, unplug the power plug and notify your specialist dealer.
Page 6
Disposal
MELAG devices are synonymous for long-term quality. When you
eventually need to decommission your MELAG device, we offer a special disposal service.
Simply contact your stockist.
Dispose of accessories and consumption media which you no longer
require in the appropriate manner. Comply with all relevant disposal specification in terms of possibly contaminated waste.
Accessories and
consumption media
Packaging
The packaging protects the device against transport damage. The
packaging materials have been selected for their environmentally­friendly and recycling properties and can be recycled. Returning the packaging to the material flow reduces the amount of waste and saves raw materials. Dispose of all non-required packaging materials at the collection points of the dual system.
Page 7
CONTENTS
CONTENTS
Chapter 1 – Performance Specifications ............ 8
Proper use .......................................................... 8
User benefits ....................................................... 9
Overview of sterilization programs .................... 10
Chapter 2 – Device Description ......................... 11
Scope of delivery .............................................. 11
Device views ..................................................... 12
Accessories ....................................................... 13
Effective capacity .............................................. 14
Control panel ..................................................... 14
Supply with feed water ...................................... 15
Chapter 3 – Initial Commissioning .................... 16
Pre-conditions for the set-up, installation and
commissioning .................................................. 16
Set-up location requirements ............................ 16
Establishing connections .................................. 17
Installation examples for connecting a water
treatment unit .................................................... 19
Compressed air connection (only MELAquick
12+) ................................................................... 21
Electrical connection ......................................... 22
Preconditions for commissioning ...................... 22
Record of installation and setting up ................. 22
Switch on the steam sterilizer ........................... 23
Opening the door .............................................. 23
Closing the door ................................................ 24
Chapter 4 – Sterilization ..................................... 25
Create prerequisites .......................................... 25
Prepare items to be sterilized ............................ 25
Loading the steam sterilizer .............................. 26
Selecting the program ....................................... 28
Select automatic pre-heating ............................ 28
Select additional drying ..................................... 29
Starting the program ......................................... 29
Program run ...................................................... 29
Manual program termination ............................. 30
Sterilization phase is finished ............................ 31
Drying phase ..................................................... 31
Program is finished ........................................... 32
Display batch number ....................................... 32
Remove sterilized items .................................... 32
Storage of sterilized items ................................. 33
Chapter 5 – Logging ........................................... 34
Batch documentation ........................................ 34
Setting the date and time .................................. 35
Use log printer as output medium ..................... 35
Employing MELAflash as output mediu m ......... 36
MELAnet Box as output medium ....................... 37
Computer as output medium ............................. 37
Automatic immediate log output ........................ 38
Subsequent log issue ........................................ 39
Display protocol memory................................... 40
Reading logs correctly ...................................... 40
Chapter 6 – Maintenance ................................... 42
Cleaning ............................................................ 42
Avoid formation of spots ................................... 43
Use qualitatively high-grade feed water ............ 44
Exchange the door gasket ................................ 44
Maintenance ..................................................... 44
Chapter 7 – Operating pauses ........................... 45
Sterilization frequency ....................................... 45
Pause times ...................................................... 45
Shut-down ......................................................... 45
Transport........................................................... 45
Recommissioning after relocat ion ..................... 46
Chapter 8 – Function test................................... 47
Automatic function test ...................................... 47
Manual function test .......................................... 47
Checks in daily operation .................................. 47
Helix test body system ...................................... 47
Display water quality ......................................... 48
Chapter 9 – Malfunctions ................................... 49
Warning messages ........................................... 49
Malfunction message ........................................ 49
Before you call .................................................. 49
No display on the screen .................................. 56
Poor drying results ............................................ 56
Open door in case of a power failure ................ 56
Overview of programs ........................................ 57
MAIN MENU ..................................................... 57
Menu: Function ................................................. 58
Technical tables .................................................. 59
Pressure-time charts ......................................... 59
Technical data ..................................................... 62
Glossary .............................................................. 63
Appendix A - Accessories ................................. 65
Page 8
Chapter 1 – Performance Specifications
8
Chapter 1 – Performance
Specifications
In this chapter you learn Under which conditions and for which purpose you should use this steam
sterilizer
What benefit you can achieve by using this steam sterilizer Which sterilization programs you can use
Proper use
Application area
The steam sterilizer is intended for use in every kind of general medical area where no Class B
steam sterilizer
is required for the instruments used
and their type of packaging.
Sterilisation tasks
According to →DIN EN 13060 this steam sterilizer is considered a Class S sterilizer. As a universal
steam sterilizer, it is suited for sterilizing
wrapped
/unwrapped solid instrumen ts, simple hollow items
(Hollow Body B)
and dental transfer instruments1. Large hollow bodies
such as hoses or similar may not be sterilized in the
this steam sterilizer.
Danger of burning!
Observe the following instructions for using the steam sterilizer:
Do not sterilize any liquids with this steam sterilizer. It is not licensed for the sterilization of liquids.
Failure to observe these instructions may lead to delayed boiling, damage to the steam sterilizer and burns.
Attention
Use the steam sterilizer only for the intended purposes outlined in
the related technical documents and only in connection with the devices and components approved by MELAG.
The sterilization of instruments with this ste am ster iliz er, like the foregoing instrument preparati on, is only to be done by competent personnel.
Only use instruments and wrappings which are suitable for steam sterilization according to the manufacturer's information.
Failure to observe these provisions could result in damage and/or can compromise safety.
1
For further information please refer to the certificate of suitability in the appendix.
Page 9
Chapter 1 – Performance Specifications
9
User b
enefits
Universal use
The steam sterilizer sterilizes on the basis of the →fractionated flow method. This method ensures the complete and effective wetting or penetration of the items to be sterilized with saturated steam. The procedure can also be used to safely and quickly sterilize wrapped instruments.
Low time expenditure,
overheating protection
To generate the sterilization steam, the steam ster iliz er uses an
integrated steam generator. The sterilizer chamber is protected from overheating and you can sterilize instruments in a short time.
Automatic pre-heating
The cold chamber is preheated on activation of preheating or can be maintained at a specific temperature between two sterilizations. This shortens the program time periods and reduces the formation of condensate to give improved drying results.
Clean feed water according
to DIN EN 13060, Appendix C
The steam sterilizer works with the feed water one-way system. Th is means that fresh feed water is used for every sterilization run. The quality of the feed water is constantly monitored via integrated conductivity mea
surements. This prevents spots on instruments and
soiling of the
steam sterilizer
, provided the instruments have been careful ly
prepared.
Optimal drying for wrapped
items to be sterilized
The sterilized items are dried with the →pulsating overpressure drying method. This way you achieve optimal drying results for wrapped objects as well.
Optimized total
operating time
The steam sterilizer works with an electronic parameter control. This way the
steam sterilizer optimizes the total operating time of a
program depending on the load.
High safety with
extensive
safety devices
The steam sterilizer verifies pressure and temperature in the sterilization chamber at all times and its door locking system does not permit opening the door when there is overpressure in the chamber. A
process
evaluation system is integrated in the electronics of the
steam sterilizer. It
compares the process parameters such as temperature, time and pressure during a program. It monitors the parameters with respect to their limit values at activation and control, and guarantees a safe and successful sterilization. If one or
more parameters are above or fall below the determined limit
values, the
steam sterilizer outputs a warning or fault messages and, if
necessary, aborts the program. In the event of an automatic program termination, observe the information on the display.
Additional
function check
With the →conductivity measurement, you can verify the water quality of the feed water.
Effective batch
documentation
The steam sterilizer has an internal log memory which can save up to 40 program cycles. All data of the completed programs are automatically stored here. You can read out the internal log memory immediately after the end of the program, or at a later time.
Page 10
Chapter 1 – Performance Specifications
10
Overview of sterilization pr ograms
The results of this chart
indicate which tests were performed at the steam
sterilizer
. The marked fields show compliance with all applicable sections of the
Standard
DIN EN 13060.
Type tests
Universal-Program S
Quick-Program S
Prion-Program S
Program type as per
DIN EN 13060
Type S
Type S
Type S
Dynamic pressure test of the
sterilization chamber
X X X
Empty chamber test
X X X
Solid load
X X X
simple hollow items (hollow
body B)
X X X
Dental transfer instruments1)
X X X
Single wrapping
X
---
X
Drying →solid load
X X X
Sterilization temperature
134 °C
134 °C
134 °C
Sterilization temperature band
134.5 ... 136.5 °C
134.5 ... 136.5 °C
134.5 ... 136.5 °C
Sterilization pressure
2 bar
2 bar
2 bar
Sterilization time
5:30 min
3:30 min
20:30 min
1) For further information please refer to the certificate of suitability in the appendix.
Page 11
Chapter 2 – Device Des cription
11
Chapter 2 – Device Description
In this chapter you learn
Which components are included in a standard scope of delivery Which components the steam sterilizer contains What safety devices the steam sterilizer has How the operating elements are designed and how you should use them How the menus of the user interface are structured What kind of water supply the steam sterilizer needs
Scope of delivery
Standard scope of delivery
MELAquick 12+/12+ p
User Manual
Warranty certificate
Manufacturer's inspection repo r t
Record of installation and setting up
1 Basket lifter
1 Sieve tray
1 Support plate
1 Feed water container with connecting hose and cable
for water level
monitoring
1 Waste water container with connecting hose and cab le f or water
level monitoring
2 Replacement device fuses
Only for MELAquick 12+
Hose for compressed air 5 m
Coupling connector (DN 7.2)
Stopper Set to disable the drying
As option
Universal basket including small part insert
Package Holder for wrapped instruments
Frame
Holding bow
Support with ISO adapters
Support with round holders
Log printer MELAprint 42/44
MELAnet Box
MELAflash CF card printer with CF card and card reader
Page 12
Chapter 2 – Device Des cription
12
Device views
Front side
(2)
(3)
(1)
(4)
(1) Twist grip (2) Operating and display field (3) Mains switch (4)
Suction fitting for drying air with insert (only MELAquick 12+ p)
Rear side
(5)
(12)
(14)
(6)
(7)
(8)
(9)
(10)
(13
)
(11)
(5) Device fuse (
6) Cable c
onnection for water level monitoring in the
feed water container
(
7) Cable c
onnection for water level monitoring in the
waste water container
(
8) Serial interface (RS232)
(
9) Sterile filter and compre ss e d air connec tion (only MELAquick 12+)
(1
0) Mains supply for IEC plug
(
11) Connection for external waste water container
(
12) Slot for emergency release
(
13) Safety valve
(
14) Connection for external feed water container
Fig. 1: Device views
Page 13
Chapter 2 – Device Des cription
13
Top view, open
(1) (2) (3)
Wiping rubber (under sliding
door)
Door
gasket
Sieve
tray
Fig. 2: Top view
Accessories
Universal basket for up to 12 handpieces with semicircular insert for universal use. Instruments up to a length of 20 cm can be sterilized in the basket.
The package holder for wrapped instruments can be removed from the steam sterilizer chamber usi ng t he basket lifter by hooking into the upper holes.
Fig. 3: Universal basket with small parts insert (optional) Fig. 4: Package holder (optional)
For cooling down the instruments the universal basket can be removed from the steam sterilizer and placed on the frame.
The holding bow is hooked into the sideways venting slots of the steam sterilizer and is used to hold the universal basket while cooling down the instruments.
Fig. 5: Frame (optional) Fig. 6: Holder (optional)
Page 14
Chapter 2 – Device Des cription
14
Support with round holders for 7 handpieces
Support with ISO adapters 7 handpieces
Fig. 7: Support with round holders (optional) Fig. 8: Support with ISO adapters (optional)
Effective capacity
Device type
Diameter
Depth
Volume
MELAquick 12+/12+ p
9 cm
20 cm
1.35 Litres
Control panel
The control panel consists of a two-line alphanumerical LED display and four membrane keys.
Function key (-) and (+)
to select, set and display special functions: print, date/time, preheating, total batches, conductivity, acknowledge error, (+)
KEY for
unlocking the door
Program selection button (P)
to select the sterilization programs/ test
programs
and select or set options
(sub-
menus) of the special functions
Start – Stop button (S)
to
start programs, terminate
programs / drying as well as control of the special functions
2-line LED display
for program status display and parameter display
Time (h:min:s)
Chamber pressure (bar) and
(steam-) temperature (°C)
12:27:12
0.02 bar 25
Page 15
Chapter 2 – Device Des cription
15
Initial state
The display switches to the initial state after every switch-on, where the current time and the chamber pressure in bar and the (steam) temperature in °C are displayed.
Supply with feed water
The steam sterilizer requires →demineralised or →distilled →feed water for steam production according to →DIN EN 13060, Appendix C. The feed water supply is provided either by an external feed water container or with
a water treatment unit (e.g. MELA
dem 40/MELAdem 47).
Feed water via the external
feed water container
If you employ the feed water container for the feed water supply, you must from time to time fill it manually. At the given point in time, the
steam sterilizer outputs an appropriate
message. The external feed water container holds 5 litr
es and suffices for approx.
20
sterilizations depending on the selected program.
The maximum water
demand for a program run is 280 ml.
Feed water via the water
treatment system
A water treatment unit is connected to the domestic water supply. It produces the
feed water required for generating the steam for the
steam
sterilizer
.
If you use a water treatment unit, then there is always sufficient feed water available. You no longer have to manually fill the feed water container.
Page 16
Chapter 3 – Initial Commissioning
16
Chapter 3 – Initial Commissioning
In this chapter you learn
Who may set up, install and start up the steam sterilizer Which prerequisites you must create for setting up, installing and starting up the
steam sterilizer
How to turn the steam sterilizer on
Pre
-conditions for the set-up, installation
and commissioning
Only have the steam steriliz er set up, installed, and started up by
people authorized by MELAG.
Danger
According to the currently valid VDE standards, this device is not
suitable for use in explosive areas.
The steam sterilizer is intended for use outside the patient environment. The device should be located a minimum of 1.5 m radius away from the treatment area.
Damages to the steam sterilizer and/or injuries can occur in case this precaution is not observed.
Set
-up location requirements
Location
Set up the steam sterilizer in a dry and dustproof location on a horizontal surface.
The relative humidity should amount to
max. 80% at 31 °C, decreasing in
a linear fashion up to a relative humidity of
max. 50% at 40 °C. We
recommend an ambient temperature of between 5
-40 °C (recommended
max. 25 °C). The load capacity of the installation surface of a steam sterilizer must
amount to
a minimum of 25 kg. The containers for the feed water and
waste water exert max. 5 kg on the installation surface.
Attention
Without fail, keep the specified distance free to the surrounding
surfaces at the sides and above the steam sterilizer.
Non-observance can lead to heat accumulation. This could adversely affect the function of the steam sterilizer and entail the shortening of the components' service life and result in extended program times.
Distance to the
surrounding surfaces
The distance to the surrounding surfaces must be at least 5 centimetres on both sides and 10 centimetres at the
rear side. The steam sterilizer
should be freely accessible from above, so that the
steam sterilizer
can be
filled and good ventilation is guara
nteed.
The setup location must be selected so that the temperature in the area of the sterile filter and along the compressed air hose cannot exceed 40°C.
Page 17
Chapter 3 – Initial Commissioning
17
Space requirements
MELAquick 12+
MELAquick 12+ p
Width
Height Depth
A= 19.5 cm
B= 43 cm C=47 cm D= 29 cm E= 9 cm F= 5 cm G= 10 cm H= 22 cm
A= 20.5 cm
B= 43 cm C=46 cm D= 29 cm E= 9 cm F= 5 cm G= 10 cm H= 22 cm
Additional space for feed
water supply
Other than for the steam sterilizer, the feed water and the waste water container require space.
Space requirement for water
treatment unit
MELAdem 40
The attachment can be next to the steam sterilizer or above or below
the steam sterilizer
MELA
dem 47
Attachment next to the steam steriliz er or above or below the steam
sterilizer (e.g. lower cabinet)
Water treatment unit
Width
Height
Depth
Diameter
MELAdem 40
32 cm
35 cm
15 cm
MELAdem 47 module housing
39 cm
47 cm
15 cm
MELAdem 47 water storage tank
51 cm
24 cm
Establishing connections
The feed water and waste water container should be installed in immediate proximity, below the
steam sterilizer and easily access ibl e.
The length of the connection hoses
and the cables amounts to 1.5 m.
During installation they can be cut to the desired l
ength. If needed, hose
lengths
and cables up to a maximum of 5 m may be ordered and used.
The mains socket must be freely accessible after installation so that the device can be taken from the electricity supply at any time.
Connection of feed water
The feed water connection hose (Ø 6 mm) is inserted to the end into the feed water connector (
Fig. 1/(14)) on the rear side of the steam sterilizer
and se
cured with the union nut. In order to enquiry the water level in the
container, the cable is connected on the
rear side of the steam sterilizer
(
Fig. 1/(6)) as well.
Waste water connection
The waste water connection hose (Ø 8 mm) is inserted to the end into the waste water connector (
Fig. 1/(11)) on the rear side of the steam sterilizer
and secured with the union nut. In order to enquiry the water level in the
Page 18
Chapter 3 – Initial Commissioning
18
container, the cable is connected on the rear side of the steam sterilizer (
Fig. 1(7)) as well.
If the connection for feed and waste water enquiry were mixed up by mistake, the information on the display for container filling and/or emptying will be incorrectly displayed.
PLEASE NOTE
The waste
-water hose must be laid dip-free with continuous descent.
Danger
It is absolutely necessary to insert both hoses to the end into the
couplings.
In case of non-observance there is danger of burns, as hot steam may escape on the rear side and/or air may enter the lines when sucking feed water and impact the functionality of the steam sterilizer.
After the connection fill in tap water (approx. 10 cm high) into the waste water container in order to cool the condensate coil and add commercially available
instrument disinfectant to prevent germ formation in the waste
water container.
Use half of the amount recommended by the
manufacturer in order to preserve the material of the condensate coil. This is required during initial commissioning and after every emptying of the waste water container.
Fig.
9: Connections
Connection hoses Cables for water level monitoring
Waste water container
Feed water container
Requirements placed on
the water treatment unit
All MELAG water treatment units comply with the following requirements:
The water pressure at the output of the water treatment unit amounts
to min. 0.5 and max. 10 bar.
The temperature of the feed water reaches max. 40 °C.
At 0.5 bar, the flow rate reaches min. 0.8 l/min for a short time and
then an average of 0.025 l/min (= 1.5 l/h).
Connection to a water
treatment unit
If the steam sterilizer should receive feed water fully automatically from a water treatment unit, the
steam sterilizer is connected using a
PTFE hose
with exterior diameter Ø 6 mm to
the water treatment unit. It is also
possible to connect the
steam sterilizer
in parallel to a different unit, using
a T
-piece, to an already existing water treatment unit of sufficient
capacity.
Page 19
Chapter 3 – Initial Commissioning
19
Next, you must switch the feed water supply to External in the "Function" menu.
Select the menu Function by simultaneous brief pressing of the
KEYS (+) and (-). The display shows Function: Last batch numb er.
Navigate with the KEY (+) or (-) until on the display Function:
Feed water–supply.
Press the KEY (P). The currently set option appears on the display,
e.g. Feed water-supply – internal.
To set the value to Feed water-supply – External, press the
KEY (P) again.
In order to save the setting and to leave the menu, press the KEY (S).
By again pressing the
KEY (S), you completely leave the menu and return
to the starting position of the display. At initial delivery, the
Feed water-supply is set to Internal.
Installation examples for connecting a water treatment unit
Example 1 – MELAdem 40
(1)
(7)
(2)
(4)
(6)
(8)
(11)
(3)
(5)
(9)
(10)
Purified water from the ion exchanger MELAdem
40
Drain via waste water container
Fig.10: Feed water-supply from the ion exchanger MELdem 40
(1)
Safeguard combination according to EN1717, consisting of return flow inhibitor and pipe aerator
(2)
Tap water adapter (reduction from 3/4“ to 1/4“)
(3)
Input side MELAdem 40, Ø 6 mm inlet hose
(4)
Ion exchanger MELAdem 40
(5)
Water tap (available on-site)
(6)
Waste water hose (Ø 8 mm)
(7)
Waste water container
(8)
Waste water connection
(9)
Feed water filter MELAdem
(10)
Connection of feed water (11)
Connection for water level monitoring in the waste water container
Function:
Feed water supply
Page 20
Chapter 3 – Initial Commissioning
20
Example 2 – MELAdem 47
(1)
(14)
(6)
(10)
(9)
(2)
(7)
(8)
(15)
(16)
(17)
(5)
(4)
(11)
(3)
(1
3)
(18)
(12)
Purified water from the reverse osmosis unit MELAdem
47
Drain via waste water container
Fig. 11: Feed water supply from the reverse osmosis unit MELAdem 47
(1)
Rear view Vacuklav 40 B+
(2)
Safeguard combination, consisting of return flow inhibitor and pipe aerator according to EN 1717
(3)
Y piece 3/4” (Art. No. 37300)
(4)
Cooling water inflow fitting
(5)
Leak monitor with shut-off valve and probe (Art.-No. 01056 – optional), recommended with fixed water connection
(6)
Connection for water level monitoring in the waste water container
(7)
Waste water connection
(8)
Connection of feed water
(9)
Waste water container
(10)
Water tap (available on-site)
(11)
T-piece with hose (Art.-No. 38605)
(12)
Feed water filter MELAdem
(13)
Cooling water outlet fitting
(14)
Reverse-osmosis unit MELAdem 47
(15)
Double hose nozzle with anti-flooding flap (Art. No. 37400)
(16)
Double-chamber trap (Art. No. 26635)
(17)
Water tap DN40 (available on-site)
(18)
Electrical connection
Page 21
Chapter 3 – Initial Commissioning
21
Compressed air connection
(only MELAquick 12+)
A compressed air connection is intended for MELAquick 12+. Connect a free end of the compressed air hose (Ø 6 mm) to the quick coupling piece of the steril e filter pre
-fitted to the steam sterilizer
and the
other end to the on-site compressed air supply. A coupling connector (NW
7.2) is included in the scope of de
livery of the steam sterilizer
. Do not
remove the sterile filter pre
-fitted to the steam steri liz er .
Danger
DIN EN 13060 prescribes that the steam sterilizer may not be operated
without the sterile filter integrated in the compressed air hose.
The
externally provided compressed air must meet the following
conditions:
Dried, condensate free, oil-free and filtered with a filter fineness
2 µm
Pressure range 3 ... 8 bar (43.5 ... 116 PSI)
For hose lengths greater than 4 m the minimum pressure must be
4 bar (58 PSI).
Standard volume flow 10 l/min. This is taken from the supply only
during drying.
For the connection of the compressed air hose, different manifolds and coupling plugs are available as optional accessories (see
Appendix A -
Accessories
).
Danger
If the steam sterilizer issues Error 42 "Drying -
drying pump" or is a program is aborted without the compressed air supply having been completed at this point, the program is viewed as not having been
completed successfully.
In such a case, the Universal-Program must be perform
ed five times
without a load (alternatively, three times with activated additional
drying). Drying may not be aborted. Only then is the
steam st erilizer
ready for operation.
The steam sterilizer
can only be operated correctly with a
permanently-connected compressed air supply.
Ensure that the compressed air supply shut-
off valve is open before
every program start.
Danger
If the steam sterilizer is operat ed w ithout a compr es sed air s upply ,
the stopper set to disable the drying (art.-no. 30986 included in the scope of delivery of the steam sterilizer) must have been installed before commissioning the steam sterilizer.
To install, follow the "Instructions for deactivating the drying function".
If the steam sterilizer is operated without a compressed air supply, in lacking a drying function, it does not conform to DIN EN 13060. The
steam sterilizer
no longer corresponds with the intended use
described in the operating manual!
Attention
According to DIN EN ISO 7396-1, the steam sterilizer may not be
connected to the supply network for medical compressed air, e.g. medical air technology.
Page 22
Chapter 3 – Initial Commissioning
22
Electrical connection
Provide the following electrical connections for the steam sterilizer:
Electric circuit with 220-240 V (max. voltage range 207 - 253 V) and
50/60 Hz (MELAquick 12+) and 50 Hz (MELAquick 12+ p, see name plate)
10 A separate fuse protection – at least Automat Type B
Protection from leakage current 30 mA
Danger of injury
The connections for electrical provision and water supply and
discharge must be set-up by trained personnel.
Non-observance could lead to a short-
circuit and/or fire and/or water
damage and/or electrical shock. This could result in serious injury.
Observe the following precautionary measures when handling the mains cable and power plug:
Danger of injury
Never splice or change the mains cable.
Never bend or twist the mains cable.
Never unplug by pulling on the power cable (socket and steam
sterilizer). Always take a grip on the plug.
Never place any heavy objects on the mains cable.
Never run the mains cable over places where it could be squeezed
(e.g. doors or windows).
Never lead the cable along a source of heat.
Do not use any nails, staples or similar objects to fixate a cable.
If the mains cable or the power plug is damaged, turn off the steam
sterilizer. Mains cables or power plugs may only be replaced by authorised persons.
If these points are not observed, the cable or plug can be damaged and/or a fire or electrical shock might occur. This could result in
serious injury.
Connecting the mains cable
Plug the mains plug into the socket and the IEC plug into the mains connector (see Fig. 1/(10)).
Attention
The IEC plug on the steam sterilizer or the mains socket must be
freely accessible after installation so that the steam sterilizer can be taken from the electricity supply at any time.
Preconditions for commissioning
The following prerequisites must be satisfied for initial start-up:
The feed water supply must be connected and filled with fresh feed
water of the appropriate quality and/or a water treatment unit must be connected as alternative.
The waste water container must be connected.
The compressed air line must be connected.
The power supply of the steam sterilizer must be assured.
Record of installation and s e t ti ng
up
PLEASE NOTE
As docu
mentation of proper setting up, installati on and comm iss ion ing as
well as for your warranty claim, the record of installation
and set up must
be filled out by the person responsible and a copy sent to MELAG.
Page 23
Chapter 3 – Initial Commissioning
23
Switch on the
steam sterilizer
Switch on mains
Turn the mains switch on to power the steam steriliz er (see Fig. 1/(3)).
After switching on
After switching on the steam sterilizer with the mains switch, the
display shows in alternation to the initial state the message
Unlocking door with button (+)
, if the door is closed.
PLEASE NOTE
In case of insufficiently filled and/or empty containers or not connected water treatment unit, a signal tone sounds and a respective message is shown on the
display.
After switching the steam sterilizer on, it require approx. 15 minutes to reach operational readiness, if the "autom. preheating" option is activated.
Open
ing the door
After the end of a program
After the end of a program, the request Acknowledge with button ’+’
is displayed. Press the KEY (+) to unlock the door. After
unlocking the door you can open the door and remove the sterilized items.
Turn the twist grip 90° counter-clockwise in order to open the sliding
door – minor resistance can be perceived just prior to the stop. Now slide the door completely to the back.
For connected output medium and
Immediate output Yes a logging
printout is additionally carried out (see page
38, Automatic immediate
log output
.)
PLEASE NOTE
Only if the twist grip is completely turned by 90°, the door can be opened:
If it was unintentionally attempted to open the door while the door was still locked, the twist grip must first be turned back until the groove is perpendicula
r to the longitudinal axis. Only then the door can be opened
again using the (+) key. Otherwise the locking mechanism jams.
Page 24
Chapter 3 – Initial Commissioning
24
Closing the door
When closing the door of the steam sterilizer, observe the following instructions:
Pull the sliding door on the twist grip forward to the end and lock the
door by turning the twist grip 90° clockwise.
In correctly locked condition, the groove in the door grip must be perpendicular to the longitudinal axis.
Attention
Never push the sliding door downwards while pulling it shut.
Otherwise, the door gasket could fold over or become damaged.
The result may be a leaking sterilization chamber as well as a subsequent malfunction message.
Page 25
Chapter 4 – Sterilization
25
Chapter 4 – Sterilization
In this chapter you learn
Which prerequisites should be created for a smoothly-running sterilization What you should observe when preparing the items to be sterilized How to correctly load the steam sterilizerWhich programs you should use for what How you start a programWhich stages a program runs throughHow to cancel a program How to recognise when the sterilization is successfully finished What you can do to improve the drying results What you must consider when removing the items to be sterilized
Create prerequisites
Create the prerequisites for a smooth and successf ul ster ilization process as follows:
Provide feed water
If feed water is supplied from the feed water container, then fill the
container with fresh feed water of appropriate quality (see page 44,
Use qualitatively high-grade feed water).
If you employ a water treatment unit, e.g. MELAdem 40, MELAdem
47, and the water feed is closed, then open it up. If the MELAdem 47 storage container should be empty, open the water feed tap at least an hour before sterilization begins.
Connect compressed air line
Make sure that the compressed air supply is ensured. This means
that the compressed air line is connected and that the shut-off valve
is
opened.
Switching on
If the steam sterilizer is still not switched on, switch it on at the mains switch. For this purpose see page
23, Switch on the steam sterilizer.
Prepare items to be sterilized
A significant prerequisite for safe disinfection and sterilization of
sterilizing
materials is the appropriate preparation, i.e. cleaning and
maintenance of the sterilizing mater ial s accor din g to the manufa ctur er' s instructions. Furthermore the materials, cleaning agents and processing procedure employed are of significan ce.
Page 26
Chapter 4 – Sterilization
26
Instruments
Observe the following instructions for treating used and brand new instruments:
Danger of infection!
Follow without fail the instructions of the instruments manufacturer
for treatment and sterilization, and observe the relevant standards and directives of your country (in Germany, e.g. →BGV A1, →RKI
und →DGSV).
Clean the instruments very thoroughly, for instance with the help of a
washer -disinfector.
After disinfection and cleaning, rinse the instruments if possible with demineralized or
distilled
water and subsequently thoroughly dry
the instruments with a clean, non-fuzzing cloth.
Employ only cleaning materials which are suitable for steam sterilization. Ask the manufacturer about the cleaning materials.
In case of non-observance, residual dirt can possibly be loosened up during sterilization in the steam pressure. Disinfection and cleaning agent residue lead to corrosion. Increased maintenance requirements and impairment of the function of the steam sterilizer
s
can be the consequence. Unsuitable cleaning materials, e.g. water-rejecting cleaning
materials or steam-impermeable oils, can result in unsterile instruments. This endangers your health and the health of your patients.
When using the following devices, observe without fail the preparation instructions of the instrument manufact urer:
Ultrasonic units
Maintenance appliances for hand pieces and angle pieces
Washer-disinfectors
Loading the
steam st erilizer
Only if the steam sterilizer is properly loaded can the sterilization be effective and the drying deliver good results.
Danger of burn
ing!
For loading the steam sterilizer:
Use the basket lifter to remove the universal basket.
Never touch the chamber or the door with unprotected hands. The
elements may be hot.
Failure to observe this requirement can result in burns.
Unwrapped instruments
Always use the universal basket for unwrapped instruments. It is possible to sterilize approx. 12 handpieces, turbines or other instruments up to a length of maximum 20 cm in this basket.
Unless specified differently by the manufacturer, hand pieces shoul
d
always be inserted into the steam sterilizer with the shaft, hence the drive side, pointing upwards.
For the sterilization of very small objects to be sterilized, e.g. in the endodontic area (e.g. Hedström files, Reamer, K files), use the semicircular sm
all part insert, which can be hooked into the universa l
basket at a greater or lower height.
Wrapping
The correct use of suitable wrapping is important for the success of the sterilization.
Attention
Employ only packaging materials and -systems (sterile barrier
systems), which satisfy the DIN EN IS0 11607-1 Standard.
You can employ soft wrapping such as clear-plastic sterilization packages, paper bags, sterilization paper or fleece.
Page 27
Chapter 4 – Sterilization
27
Soft sterilization wrapping
Soft sterilization wrapping must always be sterilized using the package holder. When using soft sterilization packages such as MELA
fol 1002. observe the following instructions:
Attention
Arrange →soft sterilization wrapping standing vertically in the
package holder to ensure that the paper side of the outer wrapping points to the outside.
The sterilizing wrappings must be clamped on both sides into the slots of the package holder. Only use respectively wide sterilizing wrapping (a width of 10 cm is recommended).
Wrapping may not be jammed between door gasket and sliding door
.
Fold the upper end of the wrapping overas needed (film side to the outside).
If the seal seam tears open during sterilization, then it might be that the wrapping was too small. If this is the case, re-wrap the instruments and sterilize them once again. The pac kag ing sh ould be filled to ¾ or the spacing for the sealing seam at 3 cm.
If the seal seam tears open during sterilization, extend the sealing impulse of the package sealing device or seal the wrappings with a double seam.
If the sterilizing wrappings are increasingly damaged when using the Prion-Program, you can reinforce the industrial seams of the packaging via sealing prior to sterilization.
Non-observance can ultimately lead to unsterile instruments and thereby to endangering the health of the patient and practice team.
Multiple wrapping
The steam sterilizer functions on the →fractionated flow method. The use of
multiple wrapping is not possible.
Mixed loads
Sterilization of mixed loads is not permitted.
Page 28
Chapter 4 – Sterilization
28
Selecting the program
Select by rotating between the initial state and the desired program with the program selection button (P).
Select the sterilization program according to how the
items to be
sterilized are wrapped. Furthermore, consider the resistance to high temperature of the items to be sterilized.
The following table shows which program to use for which items to be steriliz
ed.
Program
Wrapping
Especially suitable for
Sterilization at
Operating time1)
Drying
Load
Universal­Program S
single
wrapped
Solid instruments,
simple hollow items, dental transfer instruments
2)
134 ˚C
13-15 min.
ca. 7-10
min.
1 kg unwrapped/
270 g wrapped
Quick­Program S
unwrapped
only
Simple solid
instruments, simple hollow items, dental transfer instruments
2)
134 ˚C
6-8 min.
ca. 1 min.
1 kg
Prion­Program S
single
wrapped
Instruments where a
danger of infection by pathologically­modified proteins is suspected (e.g. Creutzfeld-Jacob, BSE), dental transfer instruments
2)
134 ˚C
28-30min.
ca. 7-10
min.
1 kg unwrapped/
270 g wrapped
1) Without drying and dependent on loading and insta lla tion cond itio ns su ch as mains volt a ge , air pressure, feed water supply
for example. We reserve the right to technical modifications. After a cold start, this can take up to 2
min longer. 2
) For further information please refer to the certificate of suitability in the appendix.
Select automatic pre-heating
Using automatic pre-heating, the chamber of the steam steril iz er is heat ed up to the program
-specific pre-heating temperature before a program start
or maintained at this temperature between two program runs. Prerequisite is that the
steam sterilizer remains continuously switched on.
The condensate formation at the chamber wall is reduced and drying is supported. The cycle times shorten.
The automatic pre
-heating is activated in the factory settings.
To alter this setting proceed as follows:
Select the Function
menu by short simultaneous pressing of the (+)
and (-) KEYS. The display shows Function: Last batch number..
Navigate with the KEY (+) or (-), up to the display
Function: autom. Pre-heating is displayed.
For confirmation press the KEY (P). The display shows the current
preset option, e.g. Pre-heating YES.
Press the KEY (P) once again, switch the display to Pre-heating
NO. The pre-heating is deactivated.
In order to end the menu Function: autom. Pre-heating
and return to the initial state, press the KEY (S) twice.
PLEASE NOTE
MELAG recommends the activation of automatic pre
-heating.
Function:
autom. Pre-heating
Page 29
Chapter 4 – Sterilization
29
Select additional drying
For difficult drying tasks you can extend the drying time of a program by 50% with the function Additional drying.
For this proceed as follows:
When starting the program, simultaneous ly press the KEY (S)
AND KEY (+).
The display shows
Additional drying selected. Then the program
sequence begins.
Starting the program
If you selected a program with the program selection button (P), in addition to the selected program, the sterilization temperature is displayed and whether the respective program is
suitable for wrapped or unwrapped
sterilizing materials.
Press the KEY (S), in order to start the program.
The steam sterilizer checks the transport of the →feed water and its
conductivity. If the Quick
-Program S is started, the warning text Warning only
unwrapped instruments
is shown on the display.
If the load exclusively contains unwrapped instruments, press the KEY
(S) once again to confirm and start the program.
Program run
After starting the program, you can follow the program run in the →display. Chamber temperature and pressure as well as the remaining duration until end of sterilization respectively the expired drying time is displayed.
Evacuation phase
In the evacuation phase the →fractionated flow method with pulsating, repeated steam entry and discharge, sucks the air out of the chamber and introduces saturated steam into the chambe r.
Depending on selected program and the current chamber temperature at program start, additional
fractionations follow.
Heating-up phase
The heating-up phase follows after the evacuation phase. A pressure discharge with a corresponding display is carried out at the end of the sterilization phase.
Sterilization phase
After achieving the sterilization parameters of pressure and temperature, the sterilization phase begins.
In alternation with pressure and temperature the remaining time of the sterilization phase is also displayed.
A pressure discharge with a corresponding display is carried out at the end of the sterilization phase.
Drying phase
The drying phase begins after the pressure discharge. The regular drying time is approx. 1
minute for the Quick-Pro gram S and
approx. 10 minutes for all other programs.
Additional drying
selected
Universal-Program S
134wrapped
1. Fractionation
0.69 bar 115°C
Sterilization
2 min., 12 s left
Current-drying
since 1’ 0.9 bar 85 °C
Page 30
Chapter 4 – Sterilization
30
Manual program termination
You can manually terminate a running program in any of the phases.
PLEASE NOTE
Do not interrupt a running program by switching off the mains switch! In case of non
-observance, an malfunction message of power loss is
displayed when the
steam sterilizer is switched back on.
Danger of burning!
Take heed that when opening the door after terminating a running program, hot water steam can escape out of the chamber, depending
on when the program was terminated.
Use the basket lifter to remove the insert basket and the package holder. Never touch the →sterilized items, the chamber or the door with unprotected hands. The elements are hot.
Failure to observe this requirement can result in burns.
Manual termination before
beginning of drying
If you terminate a program before the beginning of drying, then the →items to be sterilized are still unsterile.
To terminate the program, carry out the following steps:
Press the KEY (S).
Confirm the following prompt Stop program? Repeatedly press
"Stop" with the KEY (S).
PLEASE NOTE
The confirmation prompt is
shown for about 5 seconds on the display. If
the key is not pressed again, the program continues with the normal process.
Depending on the point in time of termination, a pressure release of the device is carried out. An appropiate display is shown on the screen.
After the pressure
release is carried out, there is a request for an
acknowledgement of the program termination. The screen alternately displays
Stop End and Acknowledge with
button ’
–’.
Press the KEY (-).
The message
Unlocking door with button ’+’ is displayed in
alternation with the previously
-selected program.
After pressing the
KEY
(+) you can open the door. If the door is not opened
within 40 seconds, the (+)
KEY must be pressed again.
A reference to
“Program stopped/load not sterile” is issued to the logging
printout.
Manual termination
during drying
In the drying phase you can cancel the program without the steam sterilizer
reporting a malfunction.
You must then expect insufficient drying, especially for wrapped
sterilized
items. Sufficient drying is a prerequisite for sterile storage.
Therefore, if possible, always let sterilization programs with wrapped items continue up to the end of the drying phase. In a Quick program, sterilized unwrapped instruments dry from their ow
n
heat after removal from the
steam sterilizer.
In the drying phase, the expired drying time is shown in alternation with the message
Immed. Removal Press "Stop" is displayed.
To terminate the program, carry out the following steps:
Press the KEY (S).
Confirm the following prompt Immediate removal? Repeatedly
press "Stop" with the KEY (S).
The termination is confirmed with
Drying stopped on the display.
Immediate removal
Press STOP
Page 31
Chapter 4 – Sterilization
31
PLEASE NOTE
The confirmation prompt is shown for about 5 seconds on the display. If the key is not
pressed again, the program continues with the normal
process.
After the ventilation of the chamber with a corresponding display text the message
Universal-Program run successfully follows in
alternation with
Last batch number. xx and quit with ’+’.
If a printer or another output medium is connected to the
steam sterilizer
and the
Immed. Output is set to YES, a reference to Drying stopped
is issued to the logging printout.
Sterilization phase i s finished
Sterilization phase
successfully finished
You can recognize on the display whether the sterilization phase completed successfully.
The remaining time of the sterilization phase is displayed in alternation with the
specification of the pressure and temperature.
Sterilization not successfully
finished
The sterilization is not successful if it was terminated by the operator or the system in the event of a malfunction.
Program termination by the
system
At a system abort, the system puts the chamber into a press urele ss stat e.
PLEASE NOTE
If the program is terminated by the operator, a warning text appears. If the program is aborted by the system, a malfunction message is displayed.
Drying phase
The steam sterilizer does an excellent job of drying the items to be sterilized. In the event of difficult drying tasks it might however be necessary to take the following measures to further improve drying:
Improve drying results
Always use the package holder for wrapped instruments.
Activate the additional drying function. Observe the section “A
dditional
drying” on page 29.
Sterilization
still 2 min, 12 s
Last batch number 7
Acknowl. with '+'
Page 32
Chapter 4 – Sterilization
32
Program is finished
At the end of a program run the chamber pressure is adapted to the ambient pressure. If the program was successfully finished, an appropriate
notification appears on the →display.
Immediate output
The report of the completed program is issued to the selected output medium for activated immediate output (see page
34, Chapter 5 –
Logging
).
Display batch number
Display last batch number
After every completed program you automatically see on the display the last batch number of the day.
You can also manually
display the last batch number:
Select the SETUP Menu Function
by short simultaneous pressing of
the KEYS (+) and (-). The display shows Function: Last batch number.
Press the KEY (P) to display the current batch number of the day.
In order to reach the starting position again, press the
KEY (S) twice.
Display total batch number
You can display the total number of →batches.
Select the SETUP Menu Function by short simul tane ous pressing of
the KEYS (+) and (-). The display shows Function: Last batch number.
Navigate with the KEY (+) or (-), up to the display where the submenu
Batch counter is shown.
Press the KEY (P).
You see the current counter reading of the total batch number on the display.
In order to reach the starting position again, press the KEY (S) twice.
Remove sterilized items
Danger of burning!
After the end of a program observe the following instructions for removing
the →sterilized items:
Never use force to open the door. The steam sterilizer could be
damaged and/or hot steam could escape.
Use the basket lifter to remove the universal basket.
Never touch the sterilized items, the chamber or the door with
unprotected hands. The elements are hot.
Failure to observe this requirement can result in burns.
Danger of infection!
Check the wrapping of the sterilized objects for damage when
removing them from the steam sterilizer.
If a wrapping is damaged, wrap the object again and re-sterilize it.
Non-observance can result in unsterile instruments. This endangers the health of the patients and the practice team.
Universal-Program S
run successfully
Last batch number
6
Batch counter
367
Page 33
Chapter 4 – Sterilization
33
Condensate residues on the
sterilized items
If you remove the →sterilized items from the steam sterilizer directly after the end of the program, it can happen that slight quantities of condensate moisture are found on the sterilized items.
According to →DIN 58953, Part 7: Smaller quantities of water which
are
found on the upper side of paper bags and clear pl
astic sterilization
packages are unobjectionable if they dry out within 30 minutes after removal from the
steam sterilizer.
PLEASE NOTE
Residual moisture dries especially quickly in the warm sterilization chamber. Thus, let wrapped instruments dry in the opened
steam sterilizer
for approx. 1 to 3 minutes.
Storage of sterilized it ems
Exclusively employ packaging which complies with standards for the sterilized materials. Do not store sterilized items in the preparation room. For the storage of the sterilized items, observe the criteria mentioned below.
Storage conditions
Sealed against dust e.g. in a closed instrument cabinet
Protected from damage on smooth surfaces
Protected from excessive fluctuations in temperature
Protected from moisture (e.g. alcohol, disinfectant)
Storage duration in accordance with the type of package wrapping
Storage time
The maximum storage life depends on the wrapping and the storage conditions. For
sterilized items which have been wrapped in compliance
with the standards (dust
-protected storage is assumed) – this time period
is up to six months.
Page 34
Chapter 5 – Logging
34
Chapter 5 – Logging
In this chapter you learn
Why and how you document batches
Which output media you can employ for the batch documentation
How to read logs correctly
How to set the date and time at the steam steriliz er
Batch documentation
The batch documentation is indispensable as documentation for the successfully completed sterilization process and as an obligatory measure of quality assurance
(Medical Devices Operator Ordinance).
The data, such as e.g. type of program as well as
batch and process
parameters of the completed programs, is store d in an inter n al log memory of the
steam sterilizer.
For the batch documentation you can transfer the internal log memory readout and the data onto diverse output media. This can be performed at the end of every program or at a later point, such as at the end of the day.
Capacity of the internal
log memory
The capacity of the internal storage suffices for about 40 logs. If the internal log memory is full, when starting the next program the respectively oldest log is automatically overwritten.
If a printer is connected and option
Immed. output "NO" is adjusted
(see page
38, Automatic immediate log output as well), a confirmation
prompt is issued prior to overwriting the saved record.
Output media
You can store the logs of the completed programs on the following output media and
archive them accordingly.
Log printer MELAprint 42/44
MELAflash CF card printer on CF card
Computer with the software MELAview/MELAtrace
(from software version 3)
MELAnet Box
PLEASE NOTE
For further information of the protocoll printer (for
example for the duration
of for he log prinouts) please refer to the respective operating instructions.
State of delivery
No option for the output of the logs is set in the initial delivery state of the steam sterilizer
. You can read below how to output the logs on the cited
media.
Page 35
Chapter 5 – Logging
35
Setting the date and time
Observe changing the clock
Date and time of the steam sterilizer must be correctly set for proper bat ch documentation. Observe the time re
-settings in autumn and spring
("daylight saving time"), since this is not reset automa tic ally S et date and time as described below:
Select the menu Function by simultaneous brief pressing of the
KEYS (+) and (-). The display shows Function: Last batch number..
Navigate with the KEY (+) or (-) in the Menu Function, up to the
display Function: Date/Time are displayed.
Press the KEY (P) in order to confirm. The current hour is displayed.
Select between the following setting possibilities using the (+) or (-)
Key: Hour, minute, second, day, month, year
In order to adjust the hour parameter, press the KEY (P) for the
confirmation.
The current value on the
display blinks.
The value can be increased or reduced with the KEYS (+) and (-).
In order to store the value, confirm with the KEY (P).
The current preset value on the display no longer blinks. In order to adjust the other parameters, proceed ana log ous ly .
After finishing the settings, press the KEY (S) to leave the menu.
On the display, the
Function: Date/Time are again displayed.
By pressing the (S)
KEY again you leave the menu completely and the
display shows its starting position again.
Use log printer as output medium
If you want to employ the log printer MELAprint 42/44 as the output medium, connect it to the
steam sterilizer as follows:
Connect printer
Connect the cables of the log printer to the printer connection (RS232
interface) on the rear side of the steam sterilizer.
Close the log printer MELA
print 42/44 according to the pertinent operating
manual.
Initialisation of the
log printer
MELAprint 42/44
In order to be able to print out logs on the connected log printer, you must initialise MELA
print 42/44 once by making the following setting at the
steam sterilizer
:
If the steam sterilizer is still not switched on, switch it on.
Wait until the display shows the basic settings (page 14, Control
panel).
Select the menu Function by simultaneous brief pressing of the
KEYS (+) and (-). The display shows Function: Last batch number..
Navigate with the KEY (+) or (-) in the Menu Function, up to the
display Function: Batch output is shown.
Press the KEY (P) to select the submenu Batch output –
Output
medium.
Press the KEY (P) again. The display shows Output medium – No
Output medium
, in case no printer was yet selected.
Function
Date/Time
Page 36
Chapter 5 – Logging
36
Navigate with the KEY (+) or (-), until the display shows as output
medium MELAprint.
For confirmation press the KEY (P). The display changes back to the
menu Batch output – Output medium.
Press the Key (S), in order to confirm and return to the menu
Function: Batch output.
Af
ter renewed pressing of the KEY (S) the display again shows the initial
state.
Test outputs on the
log printer
MELAprint 42/44
In order to verify the functional reliability of the log printer MELAprint 42/44 and its communication with the
steam sterilizer, a test output can be
carried out:
Select the menu Function by simultaneous brief pressing of the
KEYS (+) and (-). The display shows Function: Last batch number..
Navigate with the KEY (+) or (-) until on the display Function:
Batch output is shown.
Press the KEY (P). to sel ect the submenu Batch output –
Output
medium.
Press the KEY (P) in order to confirm.
Navigate with the KEY (+) or (-) until the display Test output is
shown.
In order to start the output, press the (P) KEY.
An appropriate display with Test output appears.
In order to cancel the test output or leave the menu after finished
printout, press the KEY (S). The display shows Function: Batch output.
By again pressing t
he KEY (S), you completely leave the menu and return
to the starting position of the display.
Employing MELA
flash as output medium
The MELAflash CF card printer serves for storage of sterilization logs on the MELA
flash CF card. The readout of the logs from the CF card to the
practice computer is carried out with the MELA
flash card reader.
In addition, you can use the
software MELAview/MELAtrace (from
software version 3)
software to read out log data.
PLEASE NOTE
You can download a current version of the software on the internet page of MELAG on www.melag.de in the download area.
Connect the MELAflash CF c
ard printer via the serial interface (similar to
the log printer MELA
print 42/44, see page 35).
On the
steam sterilizer, set the output media in the Function menu to
MELAflash. Detailed information can be found in the operating manual for the MELA
flash CF card printer.
Output medium
MELAprint
Test output
Page 37
Chapter 5 – Logging
37
MELAnet Box as output medium
The MELAnet Box enables the connection of the steam sterilizer to the computer network of the practice or clinic via an Ethernet interface.
The logs which are created during a program run can be saved by means of a
FTP server or over TCP via the MELAnet Box on the computer.
In the MELA
net Box there runs a small web server program, which
displays status information of the connecte d
steam sterilizer and settings
for the MELA
net
Box via a web browser. Thus, you can display the current
progress of a program sequence or current measured value s .
Connect the MELAflash CF card printer with the serial interface like the log printer MELA
print 42/44 (see page 35) and set at the steam sterilizer as
output medium
MELAflash in the SETUP menu. On the steam sterilizer,
set the output media in the
Function menu to MELAnet. Detailed
information can be found in the operating manual for the MELA
net Box.
Computer as output medium
Initialisation of the
computer
In order to be able to employ a computer as output medium, the computer must be connected via the serial interface to the
steam sterilizer. The
software
MELAview/MELAtrace (from software version 3) can be installed
for the log readout.
In order to register the computer at the steam sterilizer, proceed as described under
Use log printer as output medium on page 35.
Observe that for the output medium in the SETUP
menu Function you
must set at the
steam sterilizer as output medium Computer.
Reading out a text log on the computer
All text logs can be opened and printed out using a text editor, a word processing program or a spreadsheet program
. Graphic logs can only be
displayed with the MELAtrace/MELAview documentation software
.
Each text log (e.g. PRO, .STR, .STB etc.) must be linked with the text editor to enable the computer
to open them automatically with a text editor
.
The meanings of the e
ndings are outlined in the section Subsequent log
issue
(page 39). The following examples show how you can link the
Windows 7
editor with a specific text log.
Working in Windows Explorer, double click on the log file.
If the file ending is unfamiliar, Windows 7 will display the following
message:
Output medium
MELAnet+graphic data
Output medium
Computer
Page 38
Chapter 5 – Logging
38
Select "Select a program from a list of installed programs" and confirm
with "OK".
You can then open files with this ending via a double-
click in Windows
Editor.
Automatic immediate log output
If you immediately would like to output the associated report automatically on an output medium after the end of a program, use the function
Immediate output YES
. This is not preset in the initial delivery state.
Prerequisite for
automatic
immediate out
put of the logs
after the end of the program
The following prerequisites must be satisfied for the immediate log output after the end of a program:
The computer, the log printer MELAprint 42/44 or the MELAflash CF
card printer must be connected and initiated as output medium.
Proceed as follow in order to set the option for immediate log output after program end:
Switch on the steam sterilizer at the mains.
Select the menu Function by simultaneous brief pressing of the
KEYS (+) and (-). The display shows Function: Last batch number..
Navigate with the KEY (+) or (-) until on the display Function:
Batch output is shown and press the KEY (P).
Navigate with the KEY (+) or (-) until on the display the submenu
Immediate output YES/NO is shown.
Press the Key (P), in order to switch between Immediate output
NO and Immediate output YES.
In order to output logs immediately,
Immediate printout YES
must be
set.
In order to save the setting and to leave the menu again, press the
K
EY (S). The display shows Function: Batch output.
By again pressing the KEY (S), you leave the menu completely and
return back to the initial state.
PLEASE NOTE
If it is not possible to output a log report for the immediate output because for example the
activated output medium is not connected, a warning
message appears. MELAG recommends that you employ the immediate output of logs.
Immediate output
YES
Page 39
Chapter 5 – Logging
39
Subsequent log issue
You also have the option of outputting the logs subsequently and independently from the time of
program end. You can select whether
selected or all saved logs (up to 40 logs) should be printed. For this purpose the connected output medium is used, e.g. the log printer.
Output selected logs
In order to subsequently print selected logs of certain programs, proceed as follows:
Select the menu Function by simultaneous brief pressing of the
KEYS (+) and (-). The display shows Function: Last batch number..
Navigate with the KEY (+) or (-) until on the display Function:
Batch output is shown and press the KEY (P).
The Menu Batch output – Output medium is displayed.
Navigate with the KEY (+) or (-) until the display
Last cycle output: No. 40 is shown (as example No. 40).
Press the KEY (P). The current report number blinks.
In order to output a log of another cycle, navigate with the KEYS (+) or
(-) up to the desired number, here e.g. 25.
Press the KEY (P) in order to start the output of the selected program.
The display shows Output.
After finished printout, the display changes again to its previous setting
Last cycle output: No. 25
.
In order to output additional logs, repeat the last three steps.
In order to leave the submenu without outputting, pres s
the K
EY (S).
In order to leave the menu after the output, press the KEY (S).
The display shows the menu Function: Batch output.
By again pressing the
KEY (S), you completely leave the menu and return
back to the basic position of the display.
Output all saved logs
In order to subsequently output all saved logs, proceed as follows:
Select the menu Function by simultaneous brief pressing of the
KEYS (+) and (-). The display shows Function: Last batch number..
Navigate with the KEY (+) or (-) until on the display Function:
Batch output
is shown and press the KEY (P).
Navigate with the KEY (+) or (-), until the display Output stored
cycles is displayed.
Press the KEY (P) in order to start the output of the selected program.
The display shows Output.
After finishing the output the display again shows Output stored
cycles.
In case you would like to leave the submenu without printing, press
the K
EY (S).
PLEASE NOTE
A termination during the output on the log printer is only possible by disconnecting the unit at the mains switch or interrupting the power supply of the printer.
In order to leave the menu, likewise press the KEY (S). The display
shows the menu Function: Batch output.
By again pressing the
KEY (S), you completely leave the menu and return
to the starting position of the display.
Last cycle
output: No. 25
Output
stored cycles
Page 40
Chapter 5 – Logging
40
Delete saved logs
In order e.g. to suppress the warning text Protocol memory full for preset option
Immediate output No, the saved logs can be manually
deleted. In
the following example you can read how you can delete all saved logs.
Select the menu Function by simultaneous brief pressing of the
KEYS (+) and (-). The display shows Function: Last batch number..
Navigate with the KEY (+) or (-), up to the display
Function: Batch output is shown and press the K
EY (P).
Navigate with the KEY (+) or (-) until on the display All cycles
delete is shown.
Press the KEY (P) in order to delete all logs now.
In order to cancel the submenu without deleting, press the KEY (S).
In order to leave the menu after deletion, press the KEY (P). The
display shows Function: Batch output.
By again pressing the KEY (S), you completely leave the menu and
return to the starting position of the display.
Disp
lay protocol memory
With a connected and initialised printer or another output medium you can verify how many logs are already saved in the log memory of the steam sterilizer.
Proceed as follows:
Select the menu Function by simultaneous brief pressing of the
KEYS (+) and (-). The display shows Function: Last batch number..
Navigate with the KEY (+) or (-) until on the display Function:
Batch output is shown and press the KEY (P).
Navigate with the KEY (+) or (-), up to the display for the storage
occupancy (see left display figure) are shown.
In order to leave the menu again, press the KEY (S) twice.
Reading logs correctly
Header
The head of the program listing includes the general fundamental information for the
completed program such as for example current date,
the selected program, the daily batch number and the steam sterilizer type.
Program step values
During the program the program sequence phases are recorded with the accompanying values for steam
pressure, temperature and time (relative
to the program start).
Summary
The summary states whether the program was successfully concluded. Furthermore the values of the required sterilization time, the sterilization temperature and the pressure including
its maximum deviations are
displayed.
All cycles
delete
Allocated: 40
Free: 0
Page 41
Chapter 5 – Logging
41
---------------------------------------
MELAG MELAquick 12+
---------------------------------------
Header
Program : Universal-Program S
134°C wrapped Date : 13.03.2015 Time of day : 12:16:14 (Start) Batch number: 2 SN : 201512+p1017
---------------------------------------
Started program
Current date Time at program start Daily batch number Serial number
Preheating 133.5 °C AIN6: Conductivity 12 µS/cm
Preheating temperature conductivity of the feed water
Preheating 133.5 °C
AIN6: Conductivity 12 µS/cm
Program step Press. Temp. Time bar °C min
Start 0.02 127.6 00:00
1.Fractionation Steam entry 1.75 128.6 00:54 Press. release 0.19 127.4 01:00
2.Fractionation Steam entry 1.72 129.7 01:27 Press. release 0.22 118.0 01:35
3.Fractionation Steam entry 1.71 129.8 02:04 Press. release 0.25 108.7 02:14
4.Fractionation Steam entry 1.71 129.8 02:45 Press. release 0.25 108.7 02:58
5.Fractionation Steam entry 1.72 130.1 03:33 Press. release 0.24 108.2 03:49
6.Fractionation Steam entry 1.72 130.1 04:25 Press. release 0.23 107.3 04:42
7.Fractionation Steam entry 1.71 130.0 05:20 Press. release 0.25 108.2 05:36
8.Fractionation Steam entry 1.72 130.0 06:13 Press. release 0.24 108.8 06:30
9.Fractionation Steam entry 1.71 130.0 07:08 Press. release 0.23 108.5 07:26 Heat up 2.04 134.3 08:26 Steriliz.begin. 2.04 134.3 08:26 Steriliz.end 2.18 136.1 11:56 Press. release 0.16 108.7 12:10 Flow drying Drying begin. 0.16 108.7 12:10 Drying flow 0.12 103.1 12:16 Drying pumping 1.20 107.1 12:29 Drying end 1.21 123.1 20:06
End 0.02 123.0 20:11
Program step values
Program sequence phases with the accompanying values for pressure, temperature and time (relative to the program start)
------------------------------------------
Summary
PROGRAM PROPERLY EXECUTED!
Control message
Temperature : 136.0 +0.3 /-0.2 °C
Pressure : 2.18 +0.02/-0.02 bar Sterilizate time: 3 min 30 s Time of day : 12:36:25 (End)
---------------------------------------
Mean sterilization temperature with max. deviations
Mean sterilization pressure with max. deviations Sterilization time maintained Time at the end of the program
58 20150101 5.15 5.05
Information with total-cyc le co unter , fact or y ser ial number
and device software version n o.
Fig. 12: Example of a program log for a successful finished Universal-Program S
Page 42
Chapter 6 – Maintenance
42
Chapter 6 – Maintenance
In this chapter you learn
How to clean the steam sterilizer and which cleaning agents are suitable How to avoid the formation of spots How to exchange the door gasketWhat to observe in steam sterilizer maintenance
Cleaning
Weekly inspection of
chamber and
door gasket
Examine the chamber including door gasket and wiping rubber (page 26,
Loading the steam sterilizer
) once weekly for contaminants, deposits or
damage.
In case of contaminations
If you detect contaminations, clean the chamber with a soft, lint-free cl oth soaked in alcohol (spirit). Clean the wiping rubber and the
door gaske t
using a commercially available, gentle liquid cleaning agent. Check if the entire
door gasket is properly inserted in the groove. The
door
gasket
may not show any waves, otherwise push it back into the groove
using a rounded object. Next, start a sterilization program without loading.
Attention
When cleaning chamber and door gasket, observe the following:
Switch the steam sterilizer off before cleaning and pull the power plug from the wall socket.
Make sure that the chamber is not hot.
Use a soft and lint-free cloth.
Use chlorine- and vinegar-free cleaning agents.
Employ neutral liquid cleaning agents to clean the door gas k et.
No cleaning agents may get into the piping system which exits the
steam sterilizer chamber.
Do not use any rough objects such as pot cleaners made of metal or steel brushes.
In case of non-observance the cleaned surface could become scratched or damaged and the sealing surfaces not tight. This would promote dirt deposits and corrosion in the sterilization chamber.
Sieve tray of the chamber
The Sieve tray may only be removed and cleaned in case of coarse contaminations. The
sieve tray prevents dirt particles from depositing in
the bores of the
steam sterilizer
chamber and clogging them. Thus, ensure
that the
steam sterilizer chamber is clean prior to removing the sieve tray.
You can use a
n adhesive tape to remove the sieve tray.
A
strongly-soiled sieve tray can trigger malfunction messages
(see
Chapter 9 – Malfunctions).
Housing parts
Clean the housing parts with neutral liquid cleaners or spirit.
External feed water container
If you provide the feed water supply manually via the external feed water container, check for contamination when you refill the feed water. If necessary, clean the container with a cloth and fresh feed water
before
refilling.
Page 43
Chapter 6 – Maintenance
43
External waste water
container
Only completely empty the waste water container for cleaning. During running operation, always ensure that the cooling hose is covered with water when draining water.
Sterile filter
(only MELAquick 12
+)
Use a ph-neutral and chlorine-free cleaning fluid to remove soiling (e.g. dust) from the surface of sterile filter and the compressed air hose.
Check the filter medium regularly for discolouration. A light grey colouring of the filter medium does not influence the filter performance.
Danger
of injury
Other colourings or dampness in the filter housing points to insufficient
treatment of the compressed air. In this case, the sterile filter must be replaced immediately. Check whether the quality of the compressed air available corresponds to the requirements (see
Chapter 3 – Initial
Commissioning).
Insulating plug on the sterile
filter (only
MELAquick 12+ p)
Check the insulating plug on the sterile filter every four weeks. To do so, you can remove this using the plastic bracket and then remove
it from the
bracket. If necessary, clean the insulating plug in the sterile filter with clean water and allow it to dry before inserting.
Push the insulating plug some 2 mm over the rim of the passage into
the plastic bracket while rotating it slightly.
Place the plastic bracket in the opening in the side wall and turn it by
a full rotation so that the insulating plug sits securely in the opening of
the sterile filter.
Avoid formation of spots
Formation of spots due to
improperly cleaned
instruments
Only if you properly clean the instruments before sterilization, you can avoid the detachment of residues resulting from loading or instrument preparation under steam pressure dur ing the ster iliz ation. Detached dirt residues (e.g. residues of disinfectant
s) can clog the filter,
nozzles and valves of the
steam sterilizer and deposit as stains, spots or
discolorations on the instruments and in the chamber (see page
25,
Prepare items to be sterilized
).
Formation of spots due to
third-party rust
All steam-conducting parts of the steam sterilizer are made of non-rusting materials. This excludes the formation of rust caused by the
steam
sterilizer
. Should rust stains occur, then we are dealing with third-party
rust. If the instruments have been inco
rrectly prepared, rust can form even on
stainless steel instruments of leading manufacturers. Often a single rust
-
producing instrument already suffices for third
-party rust to form on the
other instruments or in the
steam sterilizer.
Remove third
-party rust from the instruments with a chlorine -free
stainless steel cleaning agent (see page
42, Cleaning) or send the
damaged instruments to the manufactur er for reco ndi tion ing.
Formation of spots due to
inadequate quality
of the feed water
The extent of the formation of spots on the instruments also depends on the quality of the
feed water employed for steam production.
Page 44
Chapter 6 – Maintenance
44
Use qualitativel y high-grade feed water
Attention
Observe the following when using feed water for the steam
sterilization:
Use only demineralised or distilled water according to DIN EN 13060, Appendix C.
In case of non-observance, stains o r damage o n the inst ru ments can arise and the functional readiness of the st eam ster ilizer can be adversely affected.
PLEASE NOTE
→Feed water which you can inexpensively produce with the ion exchanger MELA
dem 40 or with the revers osmosis process with MELAdem 47,
satisfies the requirements on the feed water.
Exchange the door
gasket
If the door gasket shrinks, is very wavy or shows cracks in the sealing lip, it is necessary to exchange the door
gasket.
The consequence can be leaks which can lead to the leakage of steam. Safe and successful sterilization may be impacted.
PLEASE NOTE
For
inserting the door gasket it is mandatory to observe the instructions
provided with the door
gasket spare part with the article no. 12550.
Remove the worn door gasket from the groove.
Insert the new door gasket in accordance with the instructions for
fitting the door gasket.
Maintenance
Attention
Maintenance should only be performed by trained customer services
technicians, or stockist technicians. Consult your specialist dealer or the nearest MELAG customer services point.
Maintain the specified servicing intervals.
Malfunctions at the steam sterilizer can occur on continuation of operation despite the servicing message.
For retention of value and
functional reliability
Regular maintenance is indispensable for the retention of value and the reliable practice operation of the
steam sterilizer.
All function and safety
-relevant components and electrical units are
checked during maintenance and replaced where necessary. Maintenance is carried out according to the maintenance instructions relevant for this steam sterilizer
.
Maintenance intervals
Regularly allow for maintenance every 3000 program cycles or every two years.
Page 45
Chapter 7 – Operating pauses
45
Chapter 7 – Operating pauses
In this chapter you learn
How quickly you can start sterilization programs in sequenceWhat you must observe with longer operating pauses How to put the steam sterilizer out of operation, transport it and re-start it
Sterilization fre quency
No pause times
required
Pause times between individual programs are not necessary. After the expiration or termination of drying and removal of the items to be sterilized, you can immediately re-load the steam sterilizer and start a program.
Pause times
Longer operating pauses
Close the sliding door, but do not lock the door by turning the twist grip by 90°, in order to preserve the
door gasket. Switch the steam sterilizer
off for longer operating pauses, e.g. overnight or on the weekend. Close, if available, the water feed
of the water treatment unit and the compressed
air supply.
Shut
-down
If you would like to shut down the steam sterilizer (put it out of operation) for a longer period of time, for instance because of holidays or a planned transport, proceed as follows:
Switch the steam sterilizer of f at the mains sw itch.
Pull the power plug from the wall socket.
Empty the feed water and waste water container.
Close, if available, the water feed of the water treatment unit and the
compressed air supply.
Transport
Danger of injury
Observe the following when carrying the steam sterilizer:
Use the supplied carrying straps to transport the steam sterilizer and attach them to the sides e.g. using package tape.
Never lift the steam steriliz er on the twist grip. Otherwise the locking mechanism will be damaged and functionality is not ensured any more.
Observe that the spacing between the underside of the housing floor
plate of the steam sterilizer and the installation surface is small.
Page 46
Chapter 7 – Operating pauses
46
Attention
Observe the following when transporting the steam sterilizer, e.g.
when moving or shipping the steam sterilizer or for the transport within the practice:
Empty the feed water and waste water container.
When using a water treatment unit, close the water feed and remov e
the hose connections on the back of the device.
In case of non-observance, damages to the steam sterili zer and errors may occur.
Recommissioning after relocation
At re-startup of the steam sterilizer after a change of locality, proceed as for an initial start
-up (see page 16, Chapter 3 – Initial Commissioning).
Page 47
Chapter 8 – Function test
47
Chapter 8 – Function test
In this chapter you learn
How the steam sterilizer automatically carries out the function test Which possibilities you have for the manual function test Which function tests should be conducted in daily operation How to display the water quality
Automatic function test
Process evaluation and
monitoring systems
The interaction of the sterilization-relevant parameters of pressure, temperature and time are continually automatically monitored by the electronic parameter control. The
process evaluation system of the steam sterilizer compares the
process parameters with each other during the program and monitors them with respect to their limit values. The monitoring system of the
steam sterilizer checks the dev ice
components with regard to their functional reliability and their plausible interaction. If the parameters exceed defined limit values, the
steam
sterilizer
outputs warning messages or malfunction messages. If
necessary, it interrupts the
program with a corresponding message.
When the program has ended successfully, the corresponding message will be issued on the
display.
Manual function test
You can track the program sequence based on the values shown on the
display. Furthermore on the basis of logs made for every program, you
can logically reconstruct whether a program was successful (see page
34,
Chapter
5 – Logging).
Checks in daily operation
Helix test body system
Batch-related tests
We recommend testing functional reliability of the steam sterilizer using a Helix test body system at least once a month.
For this purpose, specialist dealers offer specific dental test body systems as indicator and batch monitoring systems, e.g. Dental BMS (batch monitoring system) by GKE.
This consists of a test body, t
he Helix, and an
indicator strip,
and is used with the MELAquick 12+/MELAquick 12+ p in
the Universal
-Program S.
Prior to program start the
door gasket should be cleaned.
Page 48
Chapter 8 – Function test
48
Display water quality
The water quality can be displayed on the display of the switched on steam sterilizer
at any time - even while a program is running.
To do this keep the
KEY (-) pressed until the display shows the
c
onductivity. The conductivity is stated in µS/ cm.
As soo
n as you release the (-) KEY, the display returns to its previous
status (e.g. initial state).
Pre-heating temperature of
the chamber
If you press the KEY (-) twice shortly and the second time keep it pressed, the pre
-heating temperature of the chamber is shown instead of the
conductivity.
AIN6: Conductivity
7 µS/cm
AIN4: Temp. Preheat.
120°C
Page 49
Chapter 9 – Malfunctions
49
Chapter 9 – Malfunctions
In this chapter you learn
Which type of messages existHow to rea ct to errorsWhat you can do before calling the hotline What you can do if no display appears What you can do in case of poor drying results
Warning messages
Warning messages are displayed when necessary. Warning messages are not malfunction messages. They help to ensure malfunction
-free
operation and to recognize undesirable situations. Observe these warnings early in order to avoid
malfunctions.
Malfunction message
Malfunction messages are displayed if safe operation or sterilization safety is not guaranteed. These can appear on the
display shortly after switching on the steam
sterilizer
or while a program is running.
If a malfunction occurs during a program run, the program will be interrupted.
Danger of infection!
If a program is discontinued during the sterilization phase, the load is
unsterile. Repackage and repeat the sterilization for the respective
items to be sterilized, as needed.
Failure to observe these provisions can endanger the health of your patients and practice team.
The error message is then shown on the display in alternation with the current program phase. After the program abort or termination, the error message is shown on the display in alternation with the message
Acknowledge with button "
-" and Program stopped. In order to
delete the error message, press the
KEY (-).
Before you call
Follow the operating procedures shown in connection with a warning text or an
malfunction message on the display of the steam sterilizer. You
will find the most important
events in the following table. Possible causes
and appropriate operating instructions are listed for the events.
Page 50
Chapter 9 – Malfunctions
50
If you do not find the event in the table below or your efforts do not lead to success, turn to your specialist dealer or the
authorised MELAG
customer service centre in your vicinity. In order for us to help you, keep the serial number of your
steam sterilizer and a detailed error description
of the malfunction message ready.
Warning message
Possible cause
What you can do
Warning: Door open/
No start possible
Door switch is not closed on start
Adjust the twist grip into the proper position
90° perpendicular to the longitudinal axis.
Correct display: Door closed.
Warning: No feed water/
Refill feed water – No start possible
Only if feed water supply from
external feed water container: Not enough feed water in the
container, hose / cable not properly connected
Float switch blocke d
Check feed water level in the feed water
container, if necessary refill feed water. Check proper connection of the connection
hose / cable plug to the feed water container.
Check if float switch is blocked or jammed in the feed water container.
Check filter insert in the feed water container
for blockages.
Warning: No feed water/
Check feed water quality
The warning message appears
after program start. The built-in flow monitor doesn't close
PLEASE NOTE
At the first start-up or re-start-up, this message can occur once since the piping system is still not completely filled: restart.
Feed water supply from the feed water container
If the message appears repeatedly, have the unit checked by the MELAG customer service.
Feed water supply from MELAdem
40
Check water treatment unit, possibly open inflow to the unit.
If the message appears repeatedly, have the unit checked by the MELAG customer service.
Feed water-supply from MELdem
47
Check water treatment unit, possibly open inflow to the unit. When the storage tank is empty, restart after about 1 h.
If the message is repeated, have the water treatment unit checked by the MELAG
customer service.
Hose not correctly connected
Check hose for correct connection to the
steam sterilizer and the feed water container
and for kink-free installati on.
Warning – Empty the waste
water tank/No start possible
Waste water container is full
Empty the waste water container ensuring
that residual water (approx. 10 cm) will remain in the container (see page
42,
Chapter 6 – Maintenance).
If the steam sterilizer does not start af ter the acknowledgement, briefly switch the steam
sterilizer off and on.
Page 51
Chapter 9 – Malfunctions
51
Warning message
Possible cause
What you can do
Feed water quality bad/
Exchange cartridge/module
Conductivity of the feed water too
high 40µS
Still possible to start by repeatedly
pressing the KEY (S): Feed water supply from the feed water
container
Empty the feed water container, clean, rinse with fresh feed water and fill fresh feed water of the respective quality into the container.
Feed water supply from water treatment
unit
Mixed-bed resin pellets consumed
MELAdem 40:
Exchange the mixed-bed resin pellets (see operating manual for the water treatment
unit MELAdem 40).
Mixed-bed resin pellets in the after ion exchanger (3
rd
cartridge)
consumed
MELAdem 47:
Exchange the mixed-bed resin pellets (see operating manual for the water treatment unit MELAdem
47) and examine the water
treatment unit. On repeated occurrence, have maintenance
carried out by MELAG customer service or service of the specialist dealer. Possibly the pre- and activated carbon filter must also be
exchanged.
Mixed-bed resin pellets in reverse osmosis unit consu med
For different water treatment units:
Change the module/resin cartridge according to the manufacturer’s instructions.
On repeated occurrence, carry out maintenance.
PLEASE NOTE
Initiate program start after completion of the above-mentioned work. On the first start after maintenance of the water treatment unit, this warning text can appear once again since the inlet hose/measuring cell are still not completely flushed with fresh water.
Feed water quality
insufficient/ No start possible
Conductivity of the feed water too
high Conductivity ≥ 65 µS
Start no longer possible:
See warning message: Feed water quality insufficient/Exchange cartr idg e/module
Please wait
Preheating chamber
Message is displayed during
program start phase. The steam sterilizer has not reached the starting temperature yet.
The steam sterilizer starts automatically
after reaching the starting temperature.
Page 52
Chapter 9 – Malfunctions
52
Warning message
Possible cause
What you can do
Output medium is not ready
The steam sterilizer is operated
without output medium although an output medium is registered
Select the option in the menu Batch
output No output medium (see page 35, Initialisation of the log printer).
Output medium is not correctly
connected and/or CF card is not inserted
Check if the CF card of the CF card printer
is correctly inserted. Check the correct connection of the data cable to the steam sterilizer and at the
output medium.
Electric power supply of the log
printer or the CF card printer is interrupted
Log printer is "offline"
Ensure electric power supply, the red LED
"P" at the log printer MELAprint 42/44 must light red.
Set printer to "online" (press key "SEL" at the MELAprint
42/44, LED "SEL" must light
green).
Protocol memory full
The device-internal log memory is
allocated (capacity got up to 40 logs).
The message is shown at the start of a
program. Repeatedly press the KEY (S) to delete the
message, and the program starts. The
oldest log is then deleted.
An output medium is registered and
in the menu Batch output the option Immediate output NO is set.
Set printer to Immediate output YES
(see page 38, Automatic immediate log output). Delete internal log memory (see page 40 ,
Delete saved logs
In orde preset deleted In the f
), possibly output all saved logs beforehand (see page 39, Output all saved logs). Log off the output medium in the menu batch output output medium and
set the option no output medium.
Execute Service
The servicing message is activated,
the device has reached the prescribed batch number
The message is shown after every program
start. Repeatedly press the KEY (S) to delete the
message, and the program starts. Retain the message: press the (S) KEY twice
to start. Have maintenance carried out by MELAG
customer service or service of the specialist dealer.
PLEASE NOTE
Service counter is reset by customer service.
Warning/Battery empty
Monitoring of the device's internal
battery voltage determined a too
low voltage value.
The battery must be exchanged by MELAG
customer service or service of the specialist
dealer.
Page 53
Chapter 9 – Malfunctions
53
Malfunction message
Possible cause
What you can do
Fault 2: Steam generator
Steam sterilizer is overloa ded
Decreased heating power, since supply voltage too low
Loss of water as a result of leaks or of formation or collection of water
Steam generator overheated
Pay attention to the correct loading
quantities (see page 26, Loading the
steam sterilizer).
Check building-side electrical installation; operate steam sterilizer on a trial basis on another electric circuit.
Upon repeated occurrence, inform your specialist dealer.
Switch steam sterilizer off and have it cool down for 15 minutes.
Fault 4: Pressure release
Sieve tray contaminated
Drain hose is blocked
Clean sieve tray (see page 42, Cleaning).
Check waste water hose for kinks. On repeated occurrence notify your
specialist dealer.
Fault 8: Time bas e
Maximum difference between the
program cycle time and the internal computer clock is exceeded
Upon repeated occurrence, inform your
specialist dealer.
Fault 9: Door open!
The twist grip was turned during a
running program or was not properly closed, door contact opened during a program
Adjust the twist grip into the proper
position 90° perpendicular to the longitudinal axis. Correct display: Door closed.
Upon repeated occurrence, inform your specialist dealer.
Fault 10: Overheat. Steam
generator
Capillary tube regulator "level control"
is opened at program start (malfunction message immediat ely after start) or during a program run (until the end of the sterilization) the monitoring time until the capillary tube regulator is switched back (with water feed from feed water) is exceeded
After program termination and immediate
start, this error message can be repeated; Repeat program start after a pause time of two minutes.
Check feed water hose for kinks. Upon repeated occurrence, inform your
specialist dealer.
Fault 12: Door locking
Twist grip not 90° perpendicular to
the longitudinal axis of the steam sterilizer.
Properly turn twist grip until the stop into
the correct position. Upon repeated occurrence, inform your
specialist dealer.
Fault 14: No feed water
See warning text Warning no feed
water, however this error appears after the start of a program.
Fault 18: Sensor: ...Input: …
The device-internal inspectio n of the
sensors for temperature, pressure or conductivity resulted in a too large deviation, the message can occur when the device is switched on or during a program
Upon repeated occurrence, inform your
specialist dealer.
Fault 21: Preheating
The monitoring time from switching
on pre-heating until the achievement of the respective pre-heating temperature was exceeded, e.g. due to a too low supply voltage
On repeated occurrence selec t option
Automatic pre-heating NO ( see page
28, Select automatic pre-heating)
and inform your specialist dealer.
Page 54
Chapter 9 – Malfunctions
54
Malfunction message
Possible cause
What you can do
Fault 22: Overheat
Preheating
The maximum preheating
temperature was exceeded
On repeated occurrence select option
Automatic pre-heating No ( see page
28, Select automatic pre-heating)
and inform your specialist deal er s
Fault 23: Current
Monitoring time for the pressure
release in the outflow procedure during fractionations was exceeded
Sieve tray contaminated or de nted Waste water hose kinked
Clean sieve bottom and check waste water
hose (see page 44). Check installation of the waste water hose. Upon repeated occurrence, inform your
specialist dealer.
Fault 26: A/D converting
The maximum admissible deviation
of the computer-internal signal conditioning (A/D converting) was exceeded
Upon repeated occurrence, inform your
specialist dealer.
Fault 27: Temp.Sens. def
1.2
The maximum admissible deviation
between the two temperature sensors for the steam temperature was exceeded
Upon repeated occurrence, inform your
specialist dealer.
Fault 32: Power
loss/Sterilize sterile f ilt er
The loss of the operating voltage
occurred after the start of a program
The malfunction message occurs after the
operating voltage is again available. Check building-side installation. If no
defects are detectable, notify MELAG
customer service
Disconnection of the ste am st eriliz er
during a running program
Only terminate a running program with KEY
(S), see also page 30.
Fault 33: Pressure drop
The maximum switch -on tim e of the
steam generator for achieving the control pressure was exceeded
Check the door gasket.
Upon repeated occurrence, inform your
specialist dealer.
Fault 34: Sterilization TU1
Shortfall of minimum permissible sterilization temperature (temperature
sensor 1)
Operate steam sterilizer with smaller load.
Check door gasket for wear or damage, exchange as needed (see page 44).
Upon repeated occurrence, inform your specialist dealer.
Clean sieve bottom and check waste water hose (see page 44).
Fault 35: Sterilization TO1
Exceeding the maximum permissible
sterilization temperare (temperare
sensor 1)
Upon repeated occurrence, inform your
specialist dealer.
Fault 36: Sterilization PU
Shortfall of the minimum sterilization
pressure
Operate device with smaller load
Check door gasket for wear or damage, exchange as needed (see page 44).
Upon repeated occurrence, inform your
specialist dealer.
Fault 37: Sterilization PO
Exceeding the maximum permissible sterilization pressure
Upon repeated occurren ce, inf orm your specialist dealer.
Page 55
Chapter 9 – Malfunctions
55
Malfunction message
Possible cause
What you can do
Fault 38: Sterilization TD1
Maximum permitted difference
between theoretical temperature calculated from the pressure signal, and the temperature measured at
temperature sensor 1 was exceeded
Upon repeated occurrence, inform your
specialist dealer.
Fault 41: Current-drying
Sieve tray contaminated
Drain hose is kinked
Clean sieve bottom and check waste water
hose (see page 44). Check waste water hose for kinks. Upon repeated occurrence, inform your
specialist dealer.
Fault 42: Drying press.
pump
MELAquick 12+:
Compressed air supply not connected or is interrupted; inflow valve for the compressed air is closed.
Check compressed air connection for
possible kinks and proper connection, open inlet valve.
The steam sterilizer is ready for operation
only after the Universal-
Program has been run 5 times (or 3 times with activated additional drying) without a load and after
re-
connection of the compressed air
supply.
Upon repeated occurrence, inform your
stockist.
MELAquick 12+ p:
Check the insulating plug on the sterile
filter; if necessary clean the insulating plug in the sterile filter with clean water and allow it to dry before inserting.
Upon repeated occurrence, inform your stockist.
Fault 51: Sterilization 2
Shortfall of minimum permissible sterilization temperature (temperature
sensor 2)
Operate device with smaller load.
Check door gasket for wear and damage. Upon repeated occurrence, inform your
specialist dealer. Clean sieve bottom and check waste water
hose (see page 44).
Fault 52: Sterilization TO2
Exceeding the maximum permissible sterilization temperature (temperature
sensor 2)
Upon repeated occurrence, inform your
specialist dealer.
Also see Fault 35
Fault 53: Sterilization TD2
Maximum permitted difference
between theoretical temperature calculated from the pressure signal, and the temperature measured at
temperature sensor 2 was exceeded
Upon repeated occurrence, inform your
specialist dealer. Also see Fault 38
Page 56
Chapter 9 – Malfunctions
56
No display on the screen
After switching on the steam sterilizer the display remains empty.
What you can do
Check whether the power plug is correctly plugged in the socket and
the IEC plug on the steam sterilizer.
Check the supply voltage at the socket.
If necessary, exchange the device fuses on the rear side of the
steam
sterilizer.
Electrical Shock
Hazard
Pull the power plug from the power socket on the rear of the device
before replacing the device fuses.
Non-compliance can cause electric shock.
Heavy injuries will be the consequences.
Poor drying results
Apart from a proper device function, the drying process depends decisively on the correct setting up and loading of the
steam sterilizer.
What you can do
Do not overload the ste am ste riliz er (max. 1 kg load).
Activate the pre-heating (see page 28, Select automatic pre-
heating)
Employ the function Additional Drying (see page 29, Select
additional drying).
Open door in case of a power failure
Attention
Absolutely make sure that the pressure of the st eam ste r ilizer is
completely released:
No steam may escape the steam sterilizer.
Let the steam sterilizer absolutely cool down. Metal parts such as
door and chamber may be hot.
Non-observance may result in severe scalding/burns.
In the case that the door cannot be opened due to e.g. a
power failure,
proceed as follows taking into account the safety information mentioned above:
Switch the steam sterilizer of f at the mains sw itch and pul l the power
plug from the wall socket.
Insert a flat rounded object into the sideways slots of the steam
sterilizer (see
Fig. 1/12)) and push the object down.
Hold the object down and unlock the sliding door at the same time by
turning the twist grip counter-clockwise by 90° and opening the
sliding
door towards the back.
Page 57
Overview of programs
57
Overview of programs
MAIN MENU
hh:mm:ss
0.00 bar 89°C
Initial state
Quick-Program S
134°C unwrapped
Universal-Program S
134°C wrapped
Prion-Program S
134°C wrapped 20’
P P P
Function*
release
1st time press
*see next page
h
AIN6: Conductivity
5 µS/cm
+
2nd time press
h
AIN4: Temp_preheat.
120°C
release
Program sequence
START
Quick-Program S
run successfully
Last batch number x
Quit with button ’+’
Stop program
Press ‚Stop
Program runs
Manually terminate
during drying
Immediate removal?
Press ‚Stop’
Drying stopped
S
Manually terminate
before drying
Acknowledge
with button ‚–’
Unlocking door
with button ‚+’
Stop End
0.02 bar 88°C
Pressure release
1.5 bar 112°C
Programm
stopped
S
Stop program?
Press ‚Stop’
S
Additional drying
selected
Main Menu
Press K
EY
”Start/Stop” and (+) simultaneuously
K
EY “Start/Stop“: Stop and terminate a program
K
EY „Program“: "Enter/Confirm/Input“
S
P
Press (+) and (–) simultaneously to select the Function
menu
Select by keep pressing the
KEY (-)
Unlocking door with
KEY (+)
h
+
– S +
S
Only unwrapped
instruments
S
Page 58
Overview of programs
58
Menu: Function
K
EYS
(+) and (-): (next/previous) in the menu
K
EY “Program: “Enter/Confirm/Input
K
EY “Start/Stop”:”Terminate/Escape/Leave
K
EY “Start/Stop”: “Escape/Leave without saving
No
Yes
Press both keys simultaneously
autom.
Pre-heating
hh:mm:ss
0,00 bar 89°C
Function:
+
Initial state
S
Feed water
supply
+
+
Internal
External
S
S
P
+
S
P
Batch counter
S
Batch counter.
xx
Last
batch number
S
Last batch number
xx
P
P
+
S
Date/Time
Month: xx
Date/Time
Minute: xx
Date/Time
Second: xx
Date/Time
Day: xx
Date/Time
Year: xx
Date/Time
= s. o.
= s. o.
= s. o.
= s. o.
= s. o.
Date/Time
Hour: xx
Allocated: xx
Free: xx
Immediate output
Yes/No
Batch output
P P
+
S
P
Output medium*1)
P
+
S
P
Last cycle output:
No. xx
S
Stored cycles
output
P
S
All cycles
delete
P
S
P
Test output
+
*1) No output medium
MELAprint MELAflash MELAnet+graphic data
Computer
S P S P S P S
+
Main and submenus can always be selected by KEY (+) „next“ and
KEY (-) “previous” and you can always leave them
with
KEY (S).
P
P P P
Page 59
Technical tables
59
Technical tables
Nominal value tolerance s
Step
Press. P
Tolerance P Tol. P Tol.
2700 - 30/+160 Steam entry 1250 -390/+ 30 Steam flow off
n. F.
2700 - 50/+110 Steam entry 1250 -160/+ 50 Steam flow off 3050 - 50/+ 70 Heat up
3050 - 50/+ 70
Sterilization
start
3170
- 90/+ 90
Sterilization
1200 -140/+ 50 Pressure release
Quick-Progr. S
Universal-Progr. S
Prion-Progr. S
Fractio-
nation
1. F.
◄ same meaning as in Universal-Pr. S
All values in mbar
Pressure-time charts
Fig. 13: Pressure-time chart for Universal-Program S, 134 °C and 2.1 bar
Press. (mbar rel.)
Page 60
Technical tables
60
Fig. 14: Pressure-time chart for Quick-Program S, 134 ° C and 2.1 bar
Fig. 15: Pressure-time chart for Prion-Program S, 133 °C and 2.1 bar
Press. (mbar rel.)
Press. (mbar rel.)
Page 61
Technical tables
61
Feed water quality
Evaporation residue
10
mg/l
Silicon oxide, SiO2
1
mg/l
Iron
0.2
mg/l
Cadmium
0.005
mg/l
Lead
0.05
mg/l
Heavy metals except for those named
above
0.1
mg/l Chloride
2
mg/l
Phosphate
0.5
mg/l
pH value
5 – 7
Appearance
colourless; clean; without sediment
Hardness
0.02
mmol/l
Table 3: Minimum requirements to the feed water following the EN 13060, Appendix C
Page 62
Technical data
62
Technical data
Model name
MELAquick 12+
MELAquick 12+ p
Device dimensions (WxHxD)
19,5 x 43 x 47 cm
20,5 x 43 x 46 cm
Chamber (Ø x T)
9 x 20 cm
Volume (chamber)
1.35 l
Weight (empty)
23 kg
24 kg
Electrical connection
220-240V*, 50/60Hz, 9.5A 2050 W
220-240 V*, 50 Hz, 9.5 A 2050 W
Max. sound pressure level
66 dB(A) @1 m
69 dB(A) @ 1 m
Heat emission
c. 0.48 kW/h (1.74 MJ/h)
Surrounding temperature
5-40 °C (recommended max. 25 °C)
Relative humidity
max. 80 % with temperatures up to 31 °C
max. 50 % at 40 °C (in between decreasing in a linear fashion)
Max. altitutde
height to 2000 m
Installation category
II
Feed water quality
demineralised and/or distilled water according to DIN EN 13060,
Appendix C
Maximum load
1 kg unwrapped, 270 g single wrapped
Energy consumption during
standby
approx. 105 W Degree of soiling
Category 2
Degree of protection
(following IEC 60529)
IP20
EMC
Emitted interference and interference resistance according to EN 61326,
EN 60601-1 and -2 at 3 V/m
CE mark
CE 0197
*Observe the maximum voltage range of 207-253V
Page 63
Glossary
63
Glossary
aqua dem
demineralized water aqua dest
distilled water
Authorized persons
Medical specialist dealers, technician from depots or with customer service designated by MELAG, who were trained by MELAG
Batch
Collection of the items to be sterilized that together passed through one and the same sterilization program
BGV A1
Specifications from professional associations (Berufsgenossenschaftliche Vorschriften) – the pri ncipl es of prevention
CF card
Compact Flash card; A memory card for digital data in compact form; CF is a standardized term, i. e. these memory cards can be used in every device with a CF slot. The CF card can be read and possibly written by every device that supports the standard.
Condensate A liquid (e.g. water), which forms upon cooling from the vaporous state and thereby separates
Conductivity
is the opposite of electrical resistance; measured in micro­Siemens / centimetre (µS/cm); the greater the amount of dissolute matter in the water, the better it can conduct electrical current and thus the higher its conductivity. distilled water ideally has the conductivity zero
Conductivity measurement Measurement of the Conductivity
Corrosion Chemical change or destruction of metallic materials by water and chemicals
Delayed boiling
This is the phenomenon that in certain conditions liquids can be heated beyond their boiling point without boiling; this state is unstabile; a slight shock can cause a very rapid formation of large gas bubble which expands like an explosion
Demineralised water Also designated as aqua dem; water to a large extent without the minerals which occur in normal mineral or tap water; is obtained by ion exchange from normal tap water. It is employed here as feed water
DGSV
Deutsche Gesellschaft für Sterilgutversorgung (German Association for the Sterilized Equipment Ordinance). The DGSV training centres are specified in DIN 58946, part 6 as "Requirements of personnel".
DIN 58953 Norm – Sterilization, sterile goods supply
DIN EN 867-5 European standard: non-biological systems for use in sterilizers – Part 5: Stipulations of indicator systems and testing bodies for performance tests of small sterilizers of Type B and Type S
DIN EN 868-8 European standard: Packing materials and sytems for medical devices to be sterilized
DIN EN ISO 11140-1 European standard – Sterilization of products for health care – chemical indicators – Part 1: General requirements
DIN EN ISO 11607-1
Standard – Requirements on materials, sterile barrier systems and packaging systems; this standard is a harmonization of the DIN EN 868 Part 1 and the international standard DIN EN ISO 11607.
DIN EN 13060 European standard –Small steam sterilizers
DIN EN 285 European standard – sterilization – steam sterilizers – large sterilizers
Display Display panel on electronic devices for communicating information. Here: Graphic display of the control panel
Distilled water From lat. aqua destillata; also designated as aqua dest; water, which is largely freely of salts, organic materials and microorganisms, is obtained by distillation (evaporation and subsequent condensation) from normal tap water or pre-cleaned water. It is employed here as feed water
Dynamic pressure test of the sterilization chamber Serves to verify that the rate of the change of pressure occurring in the sterilization chamber during a sterilization cycle does not exceed a certain value, which could lead to damage of the wrapping material [DIN EN 13060]
Empty chamber test
Test without load; carried out in order to evaluate the performance of the sterilizer without the influence of the load; permits checking the maintained temperatures and pressures with respect to the intended settings [DIN EN 13060]
Evacuation Creation of a vacuum in a chamber
Feed water is required for the creation of water steam for the sterilization; typical values for the water quality according to DIN EN 285 or DIN EN 13060 – Appendix C
Flow-drying Drying, which removes moisture in the chamber using a mechanically generated air flow, is usually supported by heat
Fractionated flow method Overpressure method in which the air is expelled from the chamber and the items to be sterilized by alternating between steam entry and steam outlet.
FTP (File Transfer Protocol) is a data transmission procedure serving to transport data from the internet. These data can contain programs, files or information. Special FTP programs (FTP clients) serve for uploading any data on a server.
Heating-up period After switching on the steam sterilizer or after starting a sterilization program., it ist he time required for heating-up
Page 64
Glossary
64
the steam generator before the sterilization process starts; duration depends on the temperature of the sterilization process
Hollow body A
→Narrow lumen
Hollow body B
→Simple hollow items
Initialization
Creation of a certain initial state of the →software at start­up
Integrated steam generation The steam generator is located directly on the floor of the sterilization chamber.
Items to be sterilized Unsterile objects, materials still to be sterilized
Kontamination here: Soiling of the sterilized load with unwanted or harmful materials
LED
Light Emitting Diode: Semiconductor diode which lights up when powered by current. LEDs are predominantly employed for status displays in devices, for example to display hard disk access.
Lubricants
Instrument oil or instrument milk
Massive
without hollows or cavities, compact, sealed closed Mixed load
Wrapped and unwrapped materials to be sterilized within one load
MPBetreibV
Ordinance covering the installation, operation, application and maintenance of medical products according to § 3 of the Medical Devices Directive with the exception of medical products for clinical evaluation or performance evaluation.
Multiple wrapping e.g. instruments doubly sealed in foil or wrapped in foil are additionally found in one container or in a textile wrapped container
Narrow lumen One-sided open bodies, for which applies: 1 = L/D = 750 and L = 1500 mm or a double-sided open body for which applies: 2 = L/D = 1500 and L = 3000 mm and which does not correspond to the hollow body B L…hollow body length D…hollow body diam
Porous
Pervious to liquids and air, e.g. textiles
Process evaluation system
Also self-monitoring system – observes itsel f, acts during the programs, compares sensors with each other
Pulsating overpressure drying Drying method which removes moisture from the chamber and the items to be sterilized by alternating between air entry and discharge.
RKI Robert Koch Institute
Self-Monitoring-system
Process evaluation system
Single wrapping Wrapped once, e.g. instruments sealed in a foil ­contrasted to: Multiple wrapping
Simple hollow items
An article open on one side to which the following applies: 1 ≤ L/D ≤ 5 and D ≥ 5 mm or an article with an opening on both sides which is: 2 ≤ L/D ≤ 10 and D ≥ 5 L…length of hollow article D…diameter of hollow article [DIN EN 13060
Soft sterilization package e.g. paper bags or clear-plastic sterilization packages
Software non-material component of IT systems e.g. a computer program
Solid load – Check of porous partial load Serves to verify that for the values set on the control, the required sterilization conditions were achieved within the entire load. The load must reperesent the maximum measurements of solid instruments for whose sterilization a sterilizer has been designed according to [DIN EN 13060]
Stainless steel cleaning agent e.g. Sidol
Standards-compliant
Conformity with all the relevant standards
Sterile barrier systems
Closed minimum packaging which prevent the occurrence of microorganisms, e.g. bags closed with sealing, closed reusable containers, folded sterilization cloths
Sterilization chamber Interior of a sterilizer Accommodates the items to be sterilized
Sterilized ite ms Also designated as batch, if already successfully sterilized: sterile goods
TCP
(transmission control protocol) refers to a standard protocol for connecting computers and networks.
VDE Verband der Elektrotechnik, Elektronik und Informationstechnik e.V. Association for Electrical, Electronic Information Technologies
Page 65
Appendix A - Accessories
65
Appendix A - Accessories
Article
Article number*
MELAquick 12+
MELAquick 12+ p
Accessories
Universal basket including small part insert
00125
Package Holder
00126
Basket lifter
10003
Holder
00127
Frame
00128
Support with ISO adapters for 7 instru ment s
00122
Support with round holders for 7 instruments
00123
Water treatment
units
MELAdem 40 ion exchanger
01049
MELAdem 47 reverse-osmosis unit
01047
For documentation
MELAflash CF card printer incl. MELAflash CF
card and card reader
01039
MELAnet Box
40296
Log printer MELAprint 44
01144
Other
Cable for water level monitor in g, 5 m
21353
Compressed air
connection
Hose coupling, T-piece
38600
--
Y-piece
80220
--
Coupling
80230
--
Spare parts
Device fuse 20 A gRL
21481
Door gasket
12550
*All specified articles are available through specialist dealers
Page 66
66
Certificate of suitability
MELAquick®12+/MELAquick®12+ p
Manufacturer: MELAG Medizintechnik oHG Address: Geneststraße 6-10 10829 Berlin Country: Germany Product: St ea m s terilizer (steam sterilizer) Type of device: MELAquick
®
12+/12+ p
Classification: Class IIb medical device
We declare that the sterilizer specified above is suitable for the steam sterilization of the following instruments (wrapped and unwrapped) in accordance with DIN EN 13060 class S:
- W & H Synea WA-99LT
- W & H Alegra WE-99 LED G
- KaVo GentleSilence LUX 8000B
- KaVo IntraCompact 25 LHC
- Sirona T1 Line
- Sirona T1 Control
- Handpieces with comparable qualities (material, weight, geometry)
References to loading quantities and loading variations are outlined in the user manual and must be observed.
Manufacturer instructions regarding the medical devices stipulated for sterilization in accordance with DIN EN ISO 17664 must be observed.
Berlin, 08/01/2016
--------------------------------------------
Dr. Steffen Gebauer
(General Management)