We should like to extend our thanks for the expression of trust in our company which you have displayed through the
purchase of this MELAG product.
As an owner-run and operated family concern founded in 1951, we have a long history of successful specialization in
hygiene products for practice-based use. Our focus on innovation, quality and the highest standards of operational
reliability has established MELAG as the world’s leading manufacturer in the instrument treatment and hygiene field.
You, our customer are justified in your demand for the best products, quality and reliability. Providing "competence in hygiene"and "Quality – made in Germany", we guarantee that these demands will be met. Our certified quality
management systems is subject to close monitoring: one instrument to this end is our annual multi-day audit conducted
in accordance with ISO 13485 and ISO 9001. This guarantees that all MELAG products are manufactured and tested in
accordance with strict quality criteria.
The MELAG management and team.
Symbol
Explanation
Draws your attention to a situation, which if not avoided, can result in damage to the water
Draws your attention to important information.
Symbol
Explanation
see Chapter 2
Reference to another text section or diagram within this manual.
https://stomshop.pro
General notes
Please read this operating manual carefully before commissioning the mixed-bed resin cartridge. The operating manual
includes important safety information. The functionality and value-retention of this mixed-bed resin depends on the care
accorded to it.
Please store this operating manual carefully and in close proximity to your mixed-bed resin cartridge. It represents a
component of the product.
User group
This operating manual is addressed to doctors, their assistants and service departments.
Validity
This operating manual applies to the water treatment unit MELAdem 53/MELAdem 53 C.
Intended Use ................................................................................................................................................................... 5
Mode of functioning ......................................................................................................................................................... 5
Scope of delivery ............................................................................................................................................................. 6
Views of the device .......................................................................................................................................................... 6
Removing from the packaging ......................................................................................................................................... 7
Requirements of the set-up location ................................................................................................................................ 7
Space requirements......................................................................................................................................................... 7
Requirements for connection to the water pipe ............................................................................................................... 7
Connecting the MELAdem 53/MELAdem 53 C ................................................................................................................ 9
Installation example ....................................................................................................................................................... 11
Water distributor for connecting further devices ............................................................................................................ 15
Installing the water distributor ........................................................................................................................................ 15
Installation example ....................................................................................................................................................... 16
Combination with other MELAG devices ....................................................................................................................... 17
Transport and storage ................................................................................................................................................... 20
Technical Data .............................................................................................................................. 21
Appendix A – Accessories ............................................................................................................ 22
Appendix B – Spare parts ............................................................................................................. 23
Chapter 1 – Performance Specifications
https://stomshop.pro
Safety Instructions
When operating the mixed-bed resin cartridge, please observe the following safety instructions as well as
those contained in subsequent chapters.
Maintenance
Use only original replacement parts for maintenance.
Leaks
The water intake is to be closed upon the discovery of leaks. Check all hoses and hose
connections for leaks.
4
Chapter 1 – Performance Specifications
https://stomshop.pro
Chapter 1 – Performance Specifications
Intended Use
This mixed-bed resin cartridge works according to the ion exchange procedure. It is used for the complete
demineralization of water (DI water) of drinking water quality.
The MELAdem 53/MELAdem 53 C must always be operated with the MELAtherm 10 washer-disinfector.
Optionally, further devices can be connected, e.g. one or more steam sterilizers.
Mode of functioning
The mixed-bed resin cartridge is filled with a cation and anion resin which removes the salts dissolved in
water using the so-called adsorption process.
The mixed-bed resin is exhausted after absorbing a certain amount of salt and must be replaced and
regenerated (see Chapter 4 – Maintenance).
The capacity of a mixed-bed resin cartridge depends on the hardness of the local water supply
(see Technical Data).
5
Chapter 2 – Device Description
1
2
3
4
5 6 7
8
https://stomshop.pro
Chapter 2 – Device Description
Scope of delivery
Standard scope of delivery
2x mixed-bed resin cartridges (1 reserve)
1x MELAG transport packaging
1x Operating Manual
1x Connecting bend for the DI water connection
1x Bleed valve for decompression
1x Black rubber seal, 3 mm
4x Black rubber seal, 2 mm (2x as a reserve)
1x Water inflow hose, 2.5 m with a 180° arch and flat seal, green
1x Safety combination HD in accordance with EN 1717 with a mount and a green flat seal
1x Water inflow hose 80 cm
1x Sealable bag with cable tie
1x Water inflow distributor, Y-piece
1x Spanner (SW30)
Optionally
Water distributor for the connection of multiple devices in parallel
A leak detector (water stop) for the avoidance of water damage following leaks
Views of the device
Fig. 1: Views of the device
(1) Polypropylene head-ring
(2) Label für die Seriennummer und Praxisadresse
(3) 20 l stainless steel mixed-bed resin cartridge
(4) Polypropylene foot ring
(5) Connection to the cold water pipe of the local drinking water supply (raw water IN)
(6) Ventilation screw
(7) Screw connection opening
(8) Connection to the MELAtherm
10 DI water inflow hose (clean water OUT)
6
Chapter 3 – Commissioning
https://stomshop.pro
Chapter 3 – Commissioning
Removing from the packaging
1. Remove the two mixed-bed resin cartridges included in the scope of delivery from the packaging.
2. Store one of the two mixed-bed resin cartridges included in the scope of delivery as a reserve in
accordance with the storage conditions specified on page 20.
3. Retain the MELAG transport packaging (1 folded box, 2 foam inserts) in a safe and dry location.
These are required when returning the mixed-bed resin cartridge for regeneration.
4. Note the serial number of the mixed-bed resin cartridge on the label for control purposes. The
serial number is located on the type plate on the rear of the mixed-bed resin cartridge.
Fig. 2: Label for entering the serial number
Requirements of the set-up location
The following conditions must be given:
The installation location is clean and frost-free.
The mixed-bed resin cartridge is installed in a perpendicular position.
The installation location is freely accessible for the installation, operation and regular replacement
of the mixed-bed resin cartridge.
The hose connections must be freely accessible.
Space requirements
The space requirement of the mixed-bed resin cartridge corresponds to its dimensions plus a minimum of
25 cm for the safety combination EN 1717 and free access to the hose connections. If the MELAtherm
is to be installed as an installation device, this should be placed in an adjacent floor unit.
10
Requirements for connection to the water pipe
Installation of a cut-off valve with a back-flow preventer and a 3/4" external thread connection in the
proximity of the installation location. In order to provide a standard-conform connection for the mixed-bed
resin cartridge with back-flow preventer and pipe aerator independent of the on-site installation, we
recommend one of the following connections:
PLEASE NOTE
To prevent water damage, we recommend the use of a leak detector e.g. the MELAG water stop
(order no. 01056).
7
Chapter 3 – Commissioning
(1) Tap with back-flow preventer and pipe
(2)
(1) Ø 10 mm cold water supply (to the mixer tap)
(2)
(3)
(4)
https://stomshop.pro
Connection to the cold water supply
The water flow pressure of the on-site tap should be a minimum of 1.5 bars and may not exceed 10 bars.
Alternative connection 1
If a separate water connection (nominal width DN15 with 1/2" collar) is present / installed, a water tap with
a back-flow preventer and a pipe aerator is installed.
Fig. 3: Installation with back-flow preventer and pipe aerator
Alternative connection 2
If a cold water connection with an angle valve and a Ø 10 mm duct is present / installed, an additional
water tap with an integrated back-flow preventer and pipe aerator will be fitted directly to an existing angle
valve.
aerator (order no. 37310)
Cold water inflow hose
Additional water tap with an integrated
back-flow preventer and pipe aerator
Cold water inflow hose
Existing angle valve
Fig. 4: Installation with an angle valve
Protection with a safety combination in accordance with EN 1717
The connection of consumers to the drinking water system must be performed in accordance with EN
1717 so as to protect the drinking water supply from pollution. To this end, fit a safety combination of a
backflow preventer and a pipe aerator to the connection in accordance with EN 1717 part 4. The safety
combination included in the scope of delivery should be mounted on the wall, 25 cm over the mixed-bed
resin cartridge (see Fig. 6).