6
1.2 Safety regulations
Description
The transmitted-light bases TL RC™ and TL RCI™
fulfill the highest requirements for observation
and documentation of Leica stereo microscopes
of the M series. It contains a path-folding mirror,
a device for partial pupil illumination and relief
contrast generation, a ground-glass screen, an
additional condenser and Fresnel lenses. The
complete stand consists of:
– Transmitted-light base TL RC™ or TL RCI™
– Stand 300 mm or 500 mm long with focusing
drive, manual coarse/fine or motorized focus
– Glass stage plate, clear, 220×170×4mm
– Light source and light guide according to
requirements
Accessories:
– Gliding stage
– Leica MATS Thermocontrol System with
heating stage
– Polarization set
and many more (see Expansion diagram)
Intended uses
TL RC™ and TL RCI™ transmitted-light bases are
used in the assembly of Leica M series stereo
microscopes via tripod column and microscope
carrier. They can be combined with any cold light
source and light guide and allows you to observe
transparent specimen with relief contrast and
direct transmitted light. The TL RCI™ transmittedlight base posseses an integrated halogen lamp
that particularly supports control by Leica Application Suite (LAS) software.
Prohibited uses
If you use the transmitted-light base
TL RC™/TL RCI™, its components, and the accessories in a way other than that described in the
user manual, harm may be caused to people or
objects.
Never:
– Change, rebuild or take apart parts, if not
specifically instructed to in this manual.
– Let non-authorized persons open parts of the
instrument.
– Use the transmitted-light base TL RC™/TL RCI™
for investigations and operations on the human
eye.
The units or accessory components described in
the user manual have been checked for safety
and possible dangers. Always consult the responsible Leica office or the main plant in Wetzlar at
any time there is a need to adjust the unit or it is
combined with non Leica components which go
beyond the scope of this manual.
Unauthorized adjustment of the unit or improper
use renders the warranty null and void.
Place of use
– Only use the transmitted-light base
TL RC™/ TL RCI™ in closed, dust free rooms
and between +10°C and +40°C. Make sure that
the rooms are free of oil vapors or other chemical vapors and extreme humidity.
– Set up the electrical components at least 10 cm
from walls and combustible objects.
– Avoid large temperature fluctuations, direct
sunlight and vibrations. This can cause faulty
measurements or microphotographs.
– In warm, or warm and humid climatic zones,
the transmitted-light base TL RC™/ TL RCI™
requires particular care to prevent formation
of fungus.
Transmitted-light bases TL RC™ and TL RCI™ – Safety concept