laerdal Upright Resuscitator User Manual

Intended Use
Laerdal Upright Resuscitator (”Upright”) is a self-inflating, manual, reusable resuscitator intended for newborns and infants up to 10 kg body mass who require respiratory support.
Upright can provide supplemental oxygen only when used with the accessory Oxygen Kit.
Using Upright
Orientation: Upright is operated as a normal resuscitator, with the bag having a vertical stance over the mask.
Newborn Mask - Snap design: The mask fits with standard 15 mm inner-diameter conical connectors, as defined by ISO 5356-1. Check fit before use with other devices. When used with Upright, the mask attaches with a snap fit when pressed completely into place.
User Guide
Cat. no Description
856156 856157
8561733
846131
856155
Regulatory
0434
REF
15 mm
Dead space
-18 °C to 50 °C
-40 °C to 60 °C
<2.5 cm H
2
O at 5 LPM
<0.5 cm H
2
O at 5 LPM
>150 ml
4 ml (water volume)
Approximatelyo 72 mm x 85 mm x 217 mm
Approximately
20-08250 Rev B
User Guide
Laerdal
Upright Resuscitator
REUSABLE - AUTOCLAVABLE
Pressure release valve: Upright has a pressure release valve (“pop-off”) which limits the airway pressure to 30-45 cmH
2
O. A hissing sound can be heard when the valve opens. If higher airway pressure is needed, press downwards on the valve with the index finger while squeezing the bag.
Vomitus: If the patient valve becomes contaminated with vomitus, clear the valve by shaking Upright and forcefully squeezing the bag to expel the vomitus. If it does not clear, disassemble the patient valve (see page
2), rinse it in water and then reassemble it.
If any components are loose, tighten or reassemble the device and test in accordance with page 2.
Warnings
Upright should only be used by persons who have received sufficient training in its use. Incorrect operation of the Upright can be hazardous.
Do not use the Upright if you have any reason to be concerned about its functionality.
For proper function, ensure that Upright components are not mixed and confused with similar-looking non-Laerdal components. All Upright components are marked LAERDAL, as shown on page 2.
Cat. no. 8560533 QTY 1 each
Accessories and Spare parts
Important Product Information
Newborn Mask - Snap design – Size 0* Newborn Mask - Snap design – Size 1* * 10 pcs
Oxygen kit (Complete) includes Oxygen Adaptor, Valve, Reservoir Bag and Tubing
Oxygen Reservoir Bag and Tubing
Upright Valves and Membranes kit* (Lip Valve, Inlet Valve Disc Membrane) * 10 pcs
Upright Oxygen kit (Complete)
This product is in compliance with the essential requirements of Council Directive 93/42/EEC as ammended by Council Directive 2007/47/EC.
Technical Information
Meets ISO 10651-4:2002/EN ISO 10651-4:2009, Lung ventilators – Particular requirements for operator–powered resuscitators.
Patient port on Newborn Mask Snap design fit standard 15/22 mm connector.
Operating temperature
Storage temperature
Expiratory resistance
Tidal volume
Inspiratory resistance
External dimensions
(with Newborn Mask size 1)
Mass
(with Newborn Mask size 1)
190 grams
Hard plastic components: Polysulfone (PSU) Soft plastic components: Silicone rubber Spring: Stainless stee Not made with natural rubber latex.
Page 1
Manufacturer Information
Global Warranty: See www.laerdal.com
Manufactured in China for: Laerdal Medical AS, Tanke Svilandsgate 30 P.O. Box 377, 4002 Stavanger, Norway Tel : +47 51 51 17 00 Fax: +47 51 52 35 57 Email: laerdal.norway@laerdal.no
Pending US and international design registration. Laerdal® is a trademark or registered trademark of Laerdal Medical AS.
© 2015 Laerdal Medical AS, All rights reserved.
Includes Newborn Mask (Snap design) Size 0 and Size 1
Materials
3.
1
2
3
4
OR
2.
1.
1. 2. 3.
4
. 5.
1 2 3 5 6
4v
Boiling*
1
2
3
1. 2. 4. 3.
*Validated at approximately
sea-level pressure
Check that the silicone bag's lip is not twisted or tucked under.
4
User Guide
Laerdal Upright Resuscitator
Product overview
To dismantle, follow steps 1-6. To reassemble, follow the steps in reverse.
Pull Out Unscrew Lift Out
“LAERDAL” “LAERDAL”
Pull forcefully on tab to release
Pull Out
Do not disassemble parts beyond the steps shown. For proper function, make sure to only use parts marked LAERDAL (locations as shown above).
Cleaning and Disinfection procedure
Dismantle Clean Disinfect by one of the methods
Dry and Inspect
Assemble and Test
Always dismantle Upright before cleaning.
Wash all parts in a clean tray with clean water and mild soap.
Use a scrub or brush to remove any soil.
Rinse parts in clean water to remove all soil and soap.
Repeat above steps until parts are clean.
Boil all parts in clean water for 10 min.
Autoclaving
Sterilize by steam autoclaving at 136 °C and
2.0 kg/cm2 for 10-20min.
10 minutes
100 °C
Steam 136 °C 10-20 minutes
Dry all parts.
Visually inspect each part for damage and cleanliness / mineral deposits.
Remove damaged or unclean parts from service.
Reassemble Upright.
Test usings steps shown below.
Testing before use
Lip valve function Pressure release valve
Inlet valve opening
Product sealing
Squeeze the bag. Check that the yellow valve opens and closes with every squeeze.
Seal the mask against your hand. Squeeze the bag forcefully. Check that air is released from the pressure release valve.
Keep the mask sealed against your hand. Release the squeezed bag. Check that the bag re-expands without resistance.
Keep the mask sealed against your hand. Press the pressure release valve down. Squeeze the bag and check that there is no leakage.
Page 2
If any of the above tests fail: Dismantle Upright, inspect that components, reassemble Upright and repeat the complete “Testing before use” procedure (Section 3).
Reprocessing instructions